    BP-150ELECTROSURGICAL

    USER MANUAL

    Transmai Equipamentos Mdicos Hospitalares Ltda.

    Av. Maria Estela, 33 - Jardim Maria Estela

    CEP: 04180-010 - So Paulo SP - Brasil

    + 55 11 2335.1000

    Fax extension 210

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    2/37

    I N D E X

    Chapter 1 Introduction

    Note of property ............................................................................................ . 3General precautions ........................................................................... . . 3

    Operational precautions . . 3

    Notices . 4

    Basic care resume . . . 5

    Symbology ........................................................................................................ 5

    Danger ........................................................ . 6

    Chapter 2 Recognizing Equipment

    General information . . 8

    Packing and transport . . 13

    Selection and use of accessories .... . . 14

    Accessories accompanying Electrosurgical BP-150 . . 15

    Accessories Description .. . . 16

    Biocompatibility . . . 18

    Identifying front panel ................................................................... . . 19

    Identifying rear panel .................................................................................... . 20

    Knowing general functioning ... . 20

    Chapter 3 Use Instructions

    Instructions for use .. 23

    Recommendations during surgical operation ... . 24

    Remarks ............. 25

    Chapter 4 Care and Maintenance

    Cleaning .. . . 27

    General care and sterilization ............................................................... . . 27

    Interferences . . . . 29

    Maintenance ...................................................................................................... 30

    Inspection and preventive maintenance . 30Problems and solutions ........................................................................... . . 31

    Conclusion .. . . 32

    Technical specification ........................................................................... . . 32

    Bibliography . 34

    Guarantee 35

    Certificate of Guarantee .......................................................................... . . 36

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    3/37

    Chapter 1

    INTRODUCTION

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    4/37

    Chapter1 Introduction_______________________________________________________

    NOTE OF PROPERTY

    All information contained in this document is property of Transmai and no part of the whole

    of this publication may be duplicated without permission in writing from Transmai. Until thedate of this publication all efforts were directed to make possible the most accurate

    information in this manual.

    Transmai reserves the rights of making any changes when necessary in handbook or product

    without prior notice, aiming at improvement of product.

    GENERAL PRECAUTIONS

    ELECTRIC SHOCK RISK: Do not remove equipment lid. Besides dangerous inner

    tensions, there exists the risk of damage to patient protection system. No inner part of

    equipment could be repaired without knowledge, technical documentation and specific

    training carried out by engineering section of the company.

    This equipment must be used by qualified staff. Operator must be acquainted with

    information contained in this manual before using this electrosurgical.

    OPERATIONAL PRECAUTIONS

    Safe and properly use of electro surgery depends so much on factors only under control of

    operator.

    Really important is that operating instructions coming with or any other equipment be read,

    understood and followed judiciously.

    Electro surgery uses RF (radio frequency) to cut and coagulate tissues and therefore there will

    always be higher or lower intensity of sparks on the place, depending on power used.

    Consequently it is inherently dangerous its use near anesthetics, fluids or flammable objects.

    Precautions must be taken in the sense of restricting use of flammable material in places of

    electro surgery, being them in form of anesthetic skin conditioners or those naturally

    produced within cavities of body.

    Electro surgery units must be carefully used when in presence of internal or external

    pacemaker, since interference of electro surgery current may make them enter into

    asynchronous mode or could block them completely. When doubts turn out, consultpacemaker manufacturer and/or cardiology department.

    During surgery, do not let patient be in contact directly with ground connected metallic

    objects, such as auxiliary table for instruments, surgery table structure, etc. Diverse probes

    and electrodes used in monitoring devices of images and stimulation may provide way for

    high frequency current, independently of their 60Hz isolation or whether operated by battery.

    Risks of electro surgery burns can be reduced by placing these electrodes and probes as far as

    possible from surgery place and neutral plate.

    Transmai BP-150 3

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    5/37

    Chapter 1 Introduction______________________________________________________

    Contacts skin to skin, like arm to patient trunk for example, can be avoided placing a 10cm

    piece dry bandage and by keeping always attention to points like this one, of easy

    accumulation of liquids after asepsis of patient.

    Before using active accessories, disposables and neutral plate, read instructions, care andprecautions which come with them.

    Note: Keep active accessories far from patient, if they are not in use. Accessories

    labeled as disposable, are of one way use. Do not use or sterilize them again.

    When using coagulator vacuum cleaner, make sure that outer part of suction tube is free of

    blood and mucus. Cleaning fault of suction tube could open way to conduction of electric

    current, which may cause burns accidentally.

    Note: To avoid possibility of burns to user, turn off scalpel always before flexing or

    giving new form to suction tube. Do not dip tube into conducting liquids.

    NOTICES

    NOTICE 1: Electrical hazardous outlet. This equipment must be operated by qualified

    professional only.

    NOTICE 2: Check pens and neutral plate cords constantly. (KEEP ALWAYS A SPARE SET).

    NOTICE 3: Activation tone is an important safety characteristic. Do not let it inaudible.

    NOTICE 4: Check always perfect mechanical and electric contacts among

    EQUIPMENTCORD/NEUTRAL PLATE AND PATIENT, this is of

    fundamental importance for protection and prevention against accidents.

    NOTICE 5: Put always conductive paste between neutral plate and patient and keep its whole

    surface in firm contact with patient.

    NOTICE 6: Potential conditions of risk may occur when similar accessories are used, mainly

    monopole pens and tweezers.

    NOTICE 7: Check always position of selector key: Cut, Blend 1, Blend 2 and Blend 3;

    monopole or bipolar as one of these functions always inhibits any other.

    NOTICE 8: Power cord has to be connected to socket with ground, no adapter plugs or

    extensions must be used.

    NOTICE 9: Despite our neutral plate has bigger area than those required by standards, NEVER

    reduce size or contact area of neutral plate in order to ensure safety to patient.

    NOTICE 10: Fails on local electric network or in scalpel may cause undesired increase or

    decrease of power outlet.

    __________________________________________________________________Transmai BP-150 4

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    6/37

    Chapter1- Introduction_______________________________________________________

    NOTICE 11: While working, electrosurgical of high frequency may wrongly influence upon

    other

    electronic equipments.

    NOTICE 12: Electrical hazard: Do not remove upper lld of equipment. Call our technicalservices authorized network.

    BASIC CARE RESUME

    TransmaiElectrosurgical BP-150, despite of its sophisticated technology, having three types

    of electric current and several other resources, has an extremely simple handling, but working

    together with other equipment may complicate its operation and every kind of care should be

    taken in order to avoid accidents to both patient and user.

    SYMBOLOGY

    - Attention: Outlets which produce physiologic effects to patient and

    user

    - Risk of electric shock/dangerous tension

    - CF type part / Outlet protected against defibrillator discharge

    - Equipment with isolated outletF

    - Neat cut

    - Blend 1

    -Blend 2

    - Blend 3

    Transmai BP--150 5

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    7/37

    Chapter1- Introduction_______________________________________________________

    - Fragile content Handle with care

    - Maximum piling

    - Right position for storing and transport (This side up)

    - Should be transported and stored away from rain or high humidity

    - Top temperature for transport and storing

    - Case should not be exposed to direct sunlight (protected against sunlight)

    - Maximum weight upon case during transport and storing

    -20C

    75C

    3

    17kg

    DANGER

    EXPLOSION HAZARD

    DO NOT USE NEAR FLAMMABLE GASES AND ANESTHETICS

    _________________________________________________________________Transmai BP-150 6

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    8/37

    Chapter 2

    RECOGNIZING EQUIPMENT

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    9/37

    Chapter 2- Recognizing equipment______________________________________________

    GENERAL INFORMATION

    FUNCTION AND APPLICATION OF ELETROSURGICAL EQUIPMENT

    Electrosurgical equipment is a generator of high frequency currents which are applied to makesurgical incisions, destroying cells of tissues by means of high frequency current. There is no

    need of physical pressure, procedures are quicker, accurate and controlled. The aim of this

    technique is to explode tissue cells by heating their intern liquids. Thence rupture occurs

    and cutting effect of tissue is obtained.

    Essentially, equipment generate outlets of three kinds, one that makes cuts of tissues, a second

    one which makes coagulation of blood vessels and a third one combining cut and coagulation

    (called blend) which makes possible cut on tissues and coagulation of vessels at the same

    time.

    As described above, for every application we have the respective operational mode whichpresents the best results in application and desired surgical effects.

    Following items describe electro surgery principles, differences between electrosurgical

    equipments, electricity effects on tissues, operating modes, monopole surgery, bipolar surgery

    and surgical effects.

    ELECTROSURGERY PRINCIPLE:

    In electrosurgical equipment, electric power in inner main source is turned into an electric

    current of radio frequency. This high frequency current flows through one cord till an active

    electrode (pen). At contact point with tissue it occurs high concentration of energy within

    small area, which produces desired electronic effect (cut tissue or carry out hemostasis). Thiscurrent passes through patient down to neutral plate, which has big area of contact to patient,

    reducing current concentration (avoiding burns). Plate in connection with scalpel, circuit then

    is closed.

    High frequency is used, because frequency under 100kHz will stimulate muscles and nerves.

    Too low frequency would electrocute patient. Frequencies above 4MHz could perfectly be

    used in electro surgery, but electric phenomena like capacitive, inductive reactances and skin

    effect become significant, making it difficult the use of cords for these applications.

    _________________________________________________________________Transmai BP-150 8

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    10/37

    Chapter 2- Recognizing equipment_____________________________________________

    DIFERENCE BETWEEN ISOLATED AND GROUND REFERRED ELECTROSURGICALS:

    In ground referred electrosurgical, electric current passes from active electrode through

    patient and returns through neutral plate, however it also could return through any grounded

    object.

    Isolated electrosurgical are safer than ground referred equipments, because current

    theoretically returns by neutral plate to the negative part of inner isolator transformer of

    equipment. If neutral plate is not duly connected, this current will not find a low resistance

    way through other grounded objects. Practically, in isolated electrosurgical small escape

    current will freely circulate through this alternative way.

    Alternative ways are opened when patient leans against grounded objects. If these objects

    have small contact area, this will provoke burns to patient, much more serious with ground

    referred than with isolated electrosurgical.

    EFFECTS OF ELECTRICITY ON TISSUES:

    It is notorious that low frequencies cause several problems to patient; therefore

    Electrosurgical BP-150 operates with frequencies four times above the range in which

    stimuli to muscles and nerves do not happen.

    Intensity and distribution of heat through tissues may be changed by varying operation mode,

    cut, blend and coagulation. Electrosurgical provides adequate power for every resistance in

    safe and efficient way. It should always be observed that different tissues have different

    resistances and conductivities, and that he position of neutral plate in relation with active

    electrode and contact with patient will also interfere on more effective power values.

    Tissue Typical Resistance (Ohms)Blood 160

    Muscle, Heart, Kidney 200

    Liver, Spleen, Pancreas 300

    Lung 1000

    Adipose Tissue 3300

    Tissue Resistance Table for Frequencies from 300kHz to 3MHz

    ELECTROSURGICAL OPERATION MODES:

    Cut: Resection or dissection of biologic tissues due to passage of a high frequencyand intensity current.

    Blend: Combination of cut and coagulation effects.

    Coagulation: Sealing of small blood vessels or biologic tissues due to high frequency current.

    _________________________________________________________________Transmai BP-150 9

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    11/37

    Chapter 2- Recognizing equipment_____________________________________________

    MONOPOLE SURGERY: Patient mostly constitutes he electric circuit. Active lector de is

    responsible for conduction of electric current, which goes

    through patient body and returns to electrosurgical by neutral plate. In theory neutral plate

    does not produce heating effect or burns to tissues because electric current is less concentratedin neutral plate.

    BIPOLAR SURGERY: A nippers type electrode is mostly used in this surgery, where

    both poles are isolated from ground (isolated scalpels) and act

    directly on tissue. As active and neutral poles are placed in same nippers, when activated,

    current only circulates by tissue located between poles.

    ADVANTAGEOUS USE OF BIPOLAR OUTLET:

    1 - Reduces risk to patient, since current does not go through patient body.2 - Reduces hurting of tissues around region where operation is being done. Important

    its use in facial restoring surgery, neurosurgery, etc.

    3 - Due to concentrated characteristics of dissipation and dissection, produces a more

    efficient coagulation of blood vessels, because on humid tissues, with low resistance,

    delivers more potency with lower tensions, when in perfect contact with tissue.

    DISADVANTAGES:

    1 - Contact of nippers with tissue, mainly with dry tissue, it could stick to nippers.

    2 - Bipolar outlet always has a reduced potency, thus is less efficient when cutting and

    also needs special electrode to cut and not coagulate.

    Despite of smaller scape current in bipolar mode, the same cares should be

    taken respecting patient isolation.

    _________________________________________________________________Transmai BP-150 10

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    12/37

    Chapter 2- Recognizing equipment_____________________________________________

    BASIC SURGICAL EFFECTS MADE BY ELECTROSURGICAL EQUIPMENTS:

    Cut, incision, resection (act of separating surgically): objective

    is to heat cellules as quick as to transform them in steam,leaving a cavity in cellular matrix. Since heat is dissipated into

    steam, this avoids that neighboring cellules undergo the same

    effect. Therefore, it is important that adequate power be used to

    obtain smooth cut, because as power increases, tension increases

    the same way and the higher tension the more sparks and effects

    will occur on neighboring cellules.

    Drying, coagulation, hemostasis (act of stanching bleeding):

    Drying could be provided with any of these modes: CUT,

    BLEND 1 or BLEND 3. Current circulates by tissue, making

    water evaporate slowly, differently from cut where cellule

    evaporates faster. Normally BLEND 3 is used for drying

    because electrode is in close contact with tissue avoids that

    sparkling necroses tissue beyond the desired effect.

    When CUT is used, work should be done with reduced power to avoid cutting tissue, and

    when cutting with hemostasis is needed, BLEND mode will be more efficient.

    BLEND 3 can be used to seal vessels, however in humid parts Bipolar outet, as already said,

    could be more efficient.

    Fulguration, coagulation: BLEND 3 being a short sinusoidal

    pulse of peak high voltage, sparks to tissue without cutting

    effect. This happens because heating effect is intermittent.

    Temperature of water in cellules rises slowly until their

    carbonization. Coagulation involves drying so as fulguration

    which differs from drying in several aspects. One of them is

    that fulguration produces necrosis any place where sparks are

    discharged, while in drying sparkling is concentrated on the

    area of contact with electrode.

    NEUTRAL PLATE: When monopole outlet is in use, neutral plate functions as returningway for scalpel. Plate should be very much bigger than active

    electrode and good contact should be kept with patient (whole plate), the nearest possible to

    region where surgery will take place, smooth surface of body very well irrigated by blood

    vessels. Same should not be placed near ECG electrodes, cables or places where liquids

    could accumulate, metallic implants, skin bends, bone bulges, hair in excess, etc.

    __________________________________________________________________Transmai BP-150 11

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    13/37

    Chapter 2- Recognizing equipment_____________________________________________

    BURNS: Intensity of burns will depend on density of escape current and happen if neutral

    plate is too small, if there is no good contact to patient, if there are contacts

    patient to ground or through a very high capacitance to ground, or patient wears earrings,necklaces or metallic objects entering in contact with any grounded part, if neutral plate is

    placed on very haired area, it will not have good contact with patient.

    Currently most of ECG monitors are isolated, reducing any possibilities of burns to patient,

    but if monitor or cords are with escape to ground, as ECG electrodes are smaller than neutral

    plate, burns may occur in region of ECG electrodes.

    Synthetic underwear may produce sparks due to static electricity and subsequent ignition of

    flammable anesthetics or fluids used for skin cleaning, or electric shock onto another

    person.

    A current will flow over surface of patient body if something facilitates way, thus when usingflammable liquids in patient disinfection, remove liquids completely. With obese people,

    having skin bends or folds able to accumulate liquids, extra care should be taken. With slim

    people may also happen above effect, for example, when hip bones press strongly patient skin

    and form an isolated point where sweat and liquids accumulate and take long time to

    evaporate.

    USE OF LOOP TYPE ELECTRODE: When electrodes of this type are employed, some

    effects can be obtained by changing loop

    diameter or thickness of loop wire. Enlarging loop diameter will have smaller cutting effect,

    while increasing diameter of loop wire will have advantage of not damaging easily (burn)

    electrode. Nevertheless, to have same results of cutting when using electrode with thinnerwire, power will have to be slightly increased. Power increase should be soft, since high

    power facilitates cutting, but can also burns electrode.

    PATIENT WITH PACEMAKER: Electrosurgical units must be used with care if

    internal or external pacemakers are present because

    interference of electrosurgical current may mislead them into asynchronous mode, or could

    block completely their effects. When any doubt comes up consult pacemaker manufacturer

    and/or cardiology department.

    1 Before beginning the surgery check again all connections of active cords, neutral plate, to

    avoid sparkling between metals.

    2 - Whenever possible, use original instruments always.

    3 - If monopole instruments are employed, place neutral plate the nearest possible to surgery

    area and always do not let pass the current which circulates from active to neutral

    plate near heart.

    Notice 1: Monitor always patients with pacemaker during surgery.

    Notice 2: Keep always a defibrillator ready, during surgery of patient using

    pacemaker.

    _________________________________________________________________Transmai BP-150 12

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    14/37

    Chapter 2- Recognizing equipment______________________________________________

    NEUROMUSCULAR ESTIMULATION:

    Radio frequency is used to prevent neuromuscular stimulation. Since muscles and nerves

    have no significant response ability to alternate currents, in frequencies above 100kHz, nevercould be imagined that muscles stimulation could ever be a problem with any scalpel

    operating above said frequency, when drying tissue, this is essentially true. Nevertheless,

    when making cut or fulguration, spark complicates the problem, as electronic noise producer.

    Sparkling toward tissue is a random process and whenever power increases and decreases,

    probability of spark jumping to tissue falls. This fault of sparkling and variation of sparkling

    current produce low tensions in all DC spectrum, higher or lower than fundamental frequency

    of scalpel. In case that this current flows through patient, it will cause stimulation.

    In case of sparkling among metals on any point of outlet circuit, components of low frequency

    will be generated. This could happen when there is no good contact in some connections

    (active and neutral) and could result in contractions or even in pains to awake patient.

    If any kind of neuromuscular stimulation happens, observe following items:

    1- Check all scalpel connections, return and active electrodes, to find possible

    sparkling among metals.

    2- Check grounding of surgical center and other equipments.

    PACKING AND TRANSPORT

    Important:

    Electrosurgical packing is appropriate for safe transport. Keep it for eventual transportation

    in case of necessity.

    Characteristics of packed product:

    - Keep always packed product in right direction as indicated on case

    - Keep always packed product in humidity protected place

    - Keep always packed product in covered place

    - Place maximum of 3 packed units above each case

    - Storing temperature from 20C to +75C

    - Maximum weight distributed over packed product of 17kg.

    Packed product can be transported by any means of transportation, under Transmai

    Equipamentos Mdicos Ltda. recommendations as above. Transportation of any kind should

    be carried out complying with its original packing.

    _________________________________________________________________Transmai BP-150 13

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    15/37

    Chapter 2- Recognizing equipment_____________________________________________

    SELECTION AND USE OF ACCESSORIES

    Use always original accessories supplied or indicated by Transmai even those registered and

    approved by standards.As to form of utilization of accessories, please see Chapter - Technical Information.

    Neutral Plate Cord: Select according to patient, adult or infantile, preferably disposable type

    High frequency peak tension (isolation): Over 5k V

    High frequency peak current: Over 2 A

    Neutral Plate: Select according to patient, adult or infantile

    High frequency peak tension (isolation): Over 5k V

    High frequency peak current: Over 2 A

    Material: Stainless steel AISI 316L, AISI 420 or for medical use

    Standard Pen (Low Surgery): Selection made according to isolation and current

    High frequency peak tension (isolation): Over 5k V

    High frequency peak current: Over 2 A

    Reusable Bipolar Cord: Selection made according to isolation and current

    High frequency peak tension (isolation): Over 5k V

    High frequency peak current: Over 2 A

    Bipolar Nippers: Selection made according to isolation and current

    High frequency peak tension (isolation): Over 5k V

    High frequency peak current: Over 2 A

    Material: Stainless steel AISI 316L , AISI 420 or for medical use

    Electrodes: Selection made according to isolation, current and for each type of incision.

    High frequency peak tension (isolation): Over 5k V

    High frequency peak current: Over 2 A

    Material: Stainless steel AISI 316L , AISI 316 or for medical use

    Single Pedal: Should have IPX8 protection

    Isolation tension: Over 5kV

    Peak current (DC): Over 100mA

    Operation tension (DC): from 12 to 24 Vdc

    Power Cord: Double isolation with 3 pins (one for ground and protection)

    Minimum operation temperature: 70C

    Operation tension (minimum): 300Vac

    Isolation between conductors (minimum): 750Vac

    Minimum current: 12A

    _________________________________________________________________

    Transmai BP-150 14

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    16/37

    Chapter 2- Recognizing equipment______________________________________________

    ACESSORIES ACCOMPANYING ELECTROSURGICAL BP-150

    01 Neutral Plate, Stainless steel (150 x 100 x 0,5) mm- Ref. PN-100

    01 Neutral Plate Cord- Ref. CPN-400

    01 Standard Pen - Low surgery Ref. CPB-100

    01 Kit electrodes Low surgery Ref. KEB-100

    01 Electrode, straight blade type, small size (67mm) Ref. EFR-67

    01 Electrode, ball type (=2,1mm)- Ref. EB-2.1

    01 Electrode, ball type (=4,2mm)- Ref. EB-4.2

    01 Electrode, loop type, small size (=4,5mm)- Ref. EAL-4.5

    01 Electrode, needle type (85mm)- Ref. EA-85

    01 Electrode, depilation needle type (66mm)- Ref. EA-66

    01 Pedal, single- Ref. PS-100

    01 User Manual 01 Guarantee Certificate

    OPTIONAL ACCESSORIES

    Electrode, curve blade type, small size (67mm)- Ref. EFC-67

    Electrode, curve blade type, big size (83mm)- Ref. EFC-83

    Electrode, straight blade type, big size (100mm)- Ref. EFR-100

    Electrode, ball type (=6,0mm)- Ref. EB-6

    Electrode, ball type (=7,5mm)- Ref. EB-7.5

    Electrode, loop type, big size (

    =9,0mm)- Ref. EAL-9 Bipolar Cord, reusable- Ref. CB-400

    Bipolar Nippers, isolated, bayonet type- Ref. PBB-400

    Bipolar Nippers, isolated, curve type- Ref. PBC-400

    Bipolar Nippers, isolated, straight type- Ref. PBR-400

    Monopole Nippers, hemostatic- Ref. PMH-400

    Services Manual

    _________________________________________________________________Transmai BP-150 15

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    17/37

    Chapter 2- Recognizing equipment_____________________________________________

    ACESSORIES DESCRIPTION

    Neutral Plate, Stainless (150x100x0,5)mm-Ref. PN-100 (Accompanying part)

    Neutral Plate Cord Ref. CPN-400

    (Accompanying part)

    Kit of Electrodes, Low Surgery Ref. KEB

    (Accompanying part)

    A- Electrode, straight blade type, small (67mm)-

    Ref. EFR-67B- Electrode, ball type (=2,1mm)- Ref. EB-2.1

    C- Electrode, ball type (=4,2mm)- Ref. EB-4.2

    D- Electrode, loop type, small (=4,5mm)-Ref. EAL-4.5

    E- Electrode, needle type (85mm)- Ref. EA-85

    F- Electrode, depilation needle type (66mm)- Ref. EA-66

    Pedal, Single Ref. PS-100

    (Accompanying part)

    _________________________________________________________________Transmai BP-150 16

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    18/37

    Chapter 2- Recognizing equipment_____________________________________________

    Electrodes (Optional)

    A- Electrode, curve blade type, small (67mm)- Ref. EFC-67B- Electrode, curve blade type, big (83mm)-Ref. EFC-83

    C- Electrode, straight blade type, big (100mm)-Ref. EFR-100

    D- Electrode, ball type (=6,0mm)- Ref. EB-6

    E- Electrode, ball type (=7,5mm)- Ref. EB-7.5

    F- Electrode, loop type, big (=9,0mm)- Ref. EAL-9

    Standard Pen Low Surgery - Ref. CPB-100(Accompanying part)

    Bipolar Cord, Reusable- Ref. CB-400

    (Optional)

    _________________________________________________________________Transmai BP-150 17

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    19/37

    Chapter 2- Recognizing equipment_____________________________________________

    Nippers (Optional):

    Please verify whether case contents all these items and whether they are in good conditions. In case ofany visible damage regarding any above items, please contact Transmai.

    BIOCOMPATIBILITY

    All parts of equipment, applied parts, accessories, conductive gel, are materials and products

    biocompatible. Until present were not detected nor even received information about any

    problem of incompatibility. Technical information about incompatibility of materials used in

    the manufacture of accessories belong to our technical patrimonial files and are available for

    consultation of users.

    ___________________________________________________________________Transmai BP-150 18

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    20/37

    Chapter 2- Recognizing equipment______________________________________________

    IDENTIFYING FRONT PANEL

    Picture below shows front panel and functioning description of controls as follows:

    ON/OFF SWITCH: Enables turning ON/OFF equipment. And have inbuilt light

    signaling ON function.

    MONOPOLE KNOB: Enables adjusting ideal intensity of power to make pure cut and

    blend, according to position of four keys.

    BIPOLAR KNOB: Enables adjusting ideal intensity of power to make pure cut and

    blend, according to position of four keys.

    MONO KEY: Defines operation mode, monopole (yellow led) or bipolar

    (green led).

    CUT KEY: Provides smooth cutting with little hemostasis.

    BLEND 1 KEY: Inhibits former function and provides smooth cut with more

    hemostasis (Mix of cut and coagulation).

    BLEND 2 KEY: Inhibits former function and provides smooth cut with morehemostasis than Blend l.

    BLEND 3 KEY: Inhibits former function and provides smooth cut with morehemostasis than Blend 2.

    ACTIVE INSCRIPTION: Terminal where surgical pen is connected.

    NEUTRAL PLATE INSCRIPTION: Connector for neutral plate cord.

    PEDALCONNECTOR: To connect pedal plug completely.

    ___________________________________________________________________Transmai BP-150 19

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    21/37

    Chapter 2- Recognizing equipment______________________________________________

    VOLUME KEY: Selects sound level to low, high or off.

    BIPOLAR INSCRIPTION: Two terminals without polarity to be used with bipolar

    nippers of high isolation. This outlet provides coagulation

    of high efficiency and potency, because is only limited to the two poles of bipolar nippers. .

    IDENTIFYING REAR PANEL

    VOLTAGE SELECTOR SWITCH: Located on rear panel, enables equipment to work

    with 110 or 220V network. (FOR SAFETY, EQUIPMENT COMES OUT FACTORY WITH 220V).

    NETWORK CONNECTOR AND GENRAL FUSE: Connection for feeding cord and

    general fuse holder with 2 fuses in use.

    FUSE HOLDER LOW: With 1 fuse in each (phase) for logic circuit protection.

    FEEDING CONNECTOR: To connect electrosurgical to electric network.

    KNOWING GENERAL FUNCTIONING

    GENERAL FUNCTIONING: Turning on equipment, wait for 1 second approx. so

    that the whole system starts out. To increase or reduce

    outlet power, turn knob as desired. Take care in moving knob, as scales correspond to 1/10 ofmaximum power outlet and little movements increase potency in significant manner.

    When outlet is furnished by pedal, electrosurgical will not accept any command, change in

    selection of operational mode (Monopole, Bipolar, Cut or Blends), thus to change operational

    mode it is necessary that pedal is disabled during this process to the effect of protection to

    patient and user.

    Selection of Monopole or Bipolar modes is made by pressing once on MODE key and

    respective led will turn on and remain switched on until operational mode changes.

    ___________________________________________________________________Transmai BP-150 20

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    22/37

    Chapter 2- Recognizing equipment_____________________________________________

    In any function of Monopole mode, when working with pedal yellow indicator led lights on

    and respective sound will be heard (unless silenced). If Bipolar mode is selected, when pedal

    is working green indicator led lights on and same sound as in monopole mode will be heard

    (unless silenced).

    To set power outlet, press pedal, then a characteristic sound is heard and the led related to

    monopole or bipolar function will be lighted.

    Electrosurgical BP-150 memorizes the last situation before turning off, that is, if bipolar

    mode and blend 3 were selected when it was turned off, bipolar mode and blend 3 set again

    when electrosurgical is on. This also happens in case of black-out.

    ALARM:

    Electrosurgical BP-150 has a protection circuit which limits power outlet of scalpel to safe

    levels of operation. When alarm occurs, a sound signal shoots, ALARM signal winks andpower outlet falls to zero. To get out of this situation operator has to turn off equipment and

    turn on it again.

    Unique fail alarm happens in the occurrence of some kind of failure where outlet power

    exceeds safe levels, or could not be limited. In this situation ALARM led winks and alarm

    sound starts. To return to normal condition of work, operator has to turn off and turn on

    electrosurgical again.

    If in monopole mode any fail of connection of neutral plate occurs, sound signal shoots and

    ALARM led winks. The reverse of power excess alarm, as above mentioned, when problem

    of neutral plate is solved, sound is silenced, led ALARM stops winking and equipment returns

    to normal functions.

    SOUND:

    Electrosurgical is furnished with two different sounds, one indicates that outlet was energized

    (by pressing pedal) and another of alarm.

    The volume of sound signal of energized outlet may be adjusted in 0, medium and maximum

    by pressing VOLUME key, while sound signal of alarm can not be altered, that means fixed

    volume adjusted by factory.

    ___________________________________________________________________Transmai BP-150 21

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    23/37

    Chapter 3

    USE INSTRUCTIONS

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    24/37

    Chapter 3- Use Instructions____________________________________________________

    INSTRUCTIONS FOR USE

    1. Before connecting equipment, check whether power of network is the same as indicated by

    voltage selector switch located at the rear of equipment. If power does not match, turn switchto right voltage and change protection fuses, caring about placing adequate fuses.

    220Volts 3,15General

    110Volts 5

    2. For operator safety check whether equipment is connected to a real ground, in case of

    doubt, this connection must be made preferably by authorized electrician.

    3. Plug neutral plate cord pin to connector with inscription Neutral Plate. Next, place

    neutral plate in contact with patient skin, if possible near the local of surgical operation. To

    avoid bad contact of plate to patient, it is advisable the use ofTransmai Electro conductive

    Paste which is conditioned in packing of easy handling. Spill paste all over plate surface and

    next put same in contact with patient.

    4. Avoid liquid concentrations on any part of body after asepsis and even in course of

    surgery.

    5. Check always PATIENT / NEUTRAL PLATE / EQUIPMENT perfect contact and avoid

    patient contacts with metallic plates on bed or any surgical instrument during surgery.

    6. Connect pen to terminal with inscription Active.

    7. Connect pedal plug to front panel with inscription Pedal.

    8. Select type if electric current in accordance with surgical operation: cut or blend.

    9. Adjust current intensity in cut or blends in accordance with surgical operation, starting

    always with intensity at zero level and increasing gradually to obtain desired results. Never

    begin process with pre-selected level unless it be known for certain such level for such

    operation.

    10. Turn on equipment and check visual parts like pedal, pen cord, neutral plate, etc.

    (Special attention to last two items, since they are not checked by the test circuit of equipment).

    ___________________________________________________________________Transmai BP-150 23

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    25/37

    Chapter 3- Use Instructions____________________________________________________

    RECOMMENDATIONS DURING SURGICAL OPERATION

    1. Check constantly perfect electric and mechanical contact between neutral plate and patient

    and existence of conducting gel all over neutral plate surface.

    2. Check quality of isolation between surgical table and patient, placing any isolating

    material, like pillows of 3 or 4 cm thick between patient and all metallic parts (including

    surgical room table, metallic rails, stirrups, etc.) except for neutral plate which never must be

    isolated.

    3. Avoid skin to skin contact (for example: arms to patient thorax. Dry well these points

    and set pieces of dry bandage to avoid new accumulations during surgery).

    4. Never connect cardiac monitor ground to neutral plate even if such procedure increases

    quality of signal in screen, this is a safety measure for patient.

    5. Keep power adjustments at the lowest possible level to increase safety for both patient and

    user (since that the power be enough for that procedure).

    6. Remove scabs (dark crusts resulting from mortification to parts of tissue), which stick to

    electrode points. This will help to increase surgical effect without the need of increasing

    power of equipment.

    7. When using numerous accessories, keep conductor wires separated, do not interlace, pile

    or tie them up together, this could be dangerous and causes interferences.

    8. Keep active accessories away from patient when they are not in use.

    9. Avoid unnecessary and extended activation of scalpel, as soon as any alteration is

    perceived in power normally used (it may be caused by movement of patient over neutral

    plate).

    10. On simultaneous use of physiological monitoring equipments in same patient, any

    monitoring electrodes should be placed as far as possible away from surgical electrodes.

    Needle electrodes are not recommended.

    11 In patients bearing cardiac pacemaker or other active implants, exist potential risks of

    interferences in functioning or damages caused to them. Consult always and/or ask for

    approval from qualified persons.

    12. Occasional faults of electronic component parts may occur due to unwilling conduction

    like overcharge, lightning, fall of equipment to floor, sterilization of accessories at

    temperature above 130C and so on. It is important to have always available spare equipment

    and accessories.

    13. Use the shortest possible cord to avoid loops.

    14. Surgeon must always wear isolating gloves, mainly when using manual command

    scalpels, metallic instruments (scissors, nippers) and in bipolar surgery.

    _______________________________________________________________Transmai BP-150 24

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    26/37

    Chapter 3- Use instructions________________________________________________

    REMARKS

    Effective and safe application of Transmai Electrosurgical BP-150 very much depends ongrade of operator experience with this equipment particularly and on diverse surgery

    techniques.

    Therefore, recommendation is made for utilization of one piece fresh and moist raw meat, left

    to warm at ambient temperature (the nearest possible to human body temperature, as chilled

    meat delays effects of electric current).

    Place meat piece over neutral plate duly connected to scalpel.

    Put electrode in contact with meat piece, to be cut or coagulated, only after this, pedal will

    be set in motion, this way carbonization of tissue will be avoided.

    Increase and decrease gradually between minimum and maximum on scale of power

    intensity control in the eight electric current modalities.

    MONOPOLE (Yellow knob): Cut, Blend1, Blend2 and Blend3

    BIPOLAR (Green knob): Cut, Blend1, Blend2 and Blend3

    Observe effects of each one over meat in diverse power ranges until becoming acquainted

    with this electrosurgical model.

    _______________________________________________________________Transmai BP-150 25

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    27/37

    Chapter 4

    CARE AND MAINTENANCE

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    28/37

    Chapter 4- Care and maintenance____________________________________________

    CLEANING

    Equipment must be kept clean and dust free. Turn it off before cleaning. Use a soft cloth,

    moistened with a solution of water and neutral soap or isopropyl alcohol (isopropanol)followed by a dry cloth.

    Never use abrasive materials because they could compromise cabinet and principally

    polycarbonate displays.

    Never sink it in liquids.

    Electrodes and nippers should be re-sterilized in autoclave at temperatures below 130C after

    every surgical operation.

    Do not try sterilization in autoclave of accessories such as neutral plate, pedal or any other not

    indicated for this condition. Disinfect using specific products for this application, such as

    isopropyl alcohol (blood and tissues), enzymatic detergents (blood and tissues) glutaricaldehydes (use for disinfection of rubber parts, PVC, ETC.). When using flammable

    products for disinfection or sterilization, remove all excess and be certain that all is absolutely

    dry to avoid combustion of product during use.

    In the event of entering liquids in excess into electrosurgical, let it out of operation, for a

    period of 24hs minimum, disconnected from electric network, in ventilated place away from

    direct sunlight.

    In case of fall, send it to authorized technical assistance for maintenance and/or general

    inspection of equipment.

    GENERAL CARE AND STERILIZATION

    Transmai Electrosurgical BP-150 must be kept clean and dust free. Turn it off before

    cleaning. Use a soft cloth, moistened with a solution of water and neutral soap,

    followed by a dry cloth.

    Never use abrasive materials because they could compromise cabinet and principally

    polycarbonate displays.

    Never disconnect accessories by pulling cord. Pull always by connector.

    In case of disinfection by autoclave, verify whether accessory is appropriate forprocedure of this kind..

    BP-150 accessories can be cleaned with a humid cloth or sponge. If necessary, pedal,

    plate cord and standard pen could be disinfected by rubbing alcohol 70%, soft

    bleaching solution or any other disinfectant. Optional pen, neutral plate and electrodes

    can be sterilized in autoclave or with ethylene oxide (ETO).

    Note: Standard pen should be sterilized with ethylene oxide (ETO).

    Check all accessories before every surgical operation.

    _______________________________________________________________Transmai BP-150 27

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    29/37

    Chapter 4- Care and maintenance____________________________________________

    Permanent Neutral Plate: Monopole outlet functions as a return of current

    to scalpel. It should be much bigger that active

    electrode and good contact with patient (plate in whole) should be kept, the

    nearest possible of surgery work, an even part of body very well irrigated byblood vessels.

    It should not be placed adjacently to ECG electrode and cord, places where

    accumulation of liquids may occur, metallic implants, skin bends, bone bulges, hair in

    excess, etc.

    In case of using together with ECG monitors, it is advisable

    that neutral plate be placed according to picture at left.

    When monitoring with ECG electrodes, care should be taken

    for application on coronary region.

    Keep neutral plate and active electrode cords the shortest

    possible and without contact with ECG cord.

    During surgery operation it should be avoided that cords

    remain rolled up and in contact with grounded metallic parts.

    Blade type electrodes are more frequently used in cut

    incision, resection (to separate surgically).

    _______________________________________________________________Transmai BP-150 28

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    30/37

    Chapter 4- Care and maintenance____________________________________________

    Ball type electrodes are more frequently used in dissection, coagulation, hemostasis

    (hemorrhage interruption) and fulguration.

    Loop type electrodes are frequently used to dissection polyps and in need of takingsamples of tissues.

    Needle type electrodes are frequently used in depilation.

    Elimination of disposable products, equipment residues and accessories:

    When using parts or accessories and residues enter in direct contact with blood (invasive) or

    patients with infectious-contagious diseases, said parts must be discarded in recipients proper

    for discarding cut/perforating objects.

    In the event of doubts use always recipients proper for discarding cut-perforating objects or

    consult competent persons to take a decision, but never cast those objects into commongarbage.

    INTERFERENCES

    Function principle of electrosurgical makes that same be an emitter source of noises. Armors

    and filters were projected to minimize these effects. Nevertheless, during surgical operation,

    contact of active electrode with patient tissue sparkling happen which will produce noise and

    active electrode will function as an antenna irradiating it in all directions. Therefore, before

    surgical operation, care should de taken in checking equipments more influenced by

    interference.

    INTERFERENCES IN PACEMAKERS

    a) Before beginning surgical operation, check again all connections of active cords and

    neutral plate to avoid sparkling among metals.

    b) If possible, use always a ground specific for this equipment.

    c) When using monopole instruments, put neutral plate in the nearest possible position

    to local of surgery works, always avoiding that the current which circulates from

    active to neutral plate do not pass by the proximities of heart.

    _______________________________________________________________Transmai BP-150 29

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    31/37

    Chapter 4- Care and maintenance____________________________________________

    Notice 1: Monitor always patient with pacemaker during surgical operation.

    Notice 2: Keep always a defibrillator, ready to use, during surgical operation in

    patient with pacemaker.INTERFERENCE IN MONITOR (ECG):

    a) Check chassis ground connections of both equipments.

    b) Check all electric equipments in surgical room, respecting defective grounding.

    c) When there are different equipments grounded to different grounds, a potential difference

    may appear and monitor could respond to same.

    MAINTENANCE

    Electrosurgical BP-150 should be only repaired by Transmai technical Service or authorizedstaff.. Inadequate maintenance may compromise safety of patient.

    Preventive maintenance and calibration of equipment must be done annually.

    Check always correct work of all equipments involved in surgical operation and observe basic

    rules recommended for electro surgery.

    The factory is not responsible for wrong use of this equipment and advises against

    manipulation by persons with lack of wide and total knowledge of electro surgery techniques.

    Fuses replacement: General Fuse Turn off general key on front panel of equipment,

    disconnect power cord from network and equipment, then unlock fuse holder with a

    screwdriver and replace them in accordance with table below.

    220Volts 3,15AGeneral

    110Volts 5

    After replacement carry out basic inspection of electrosurgical and in the event of doubts or

    whether burning of fuses persists, contact authorized technical assistance or Transmai

    directly.

    INSPECTION AND PREVENTIVE MAINTENANCE

    ACCESSORIES: Before any surgical proceeding all accessories should be submitted tovisual and operative review. This instrumental inspection should be

    carried out by the company itself through maintenance staff.

    Neutral Plate Cord and Permanent Neutral Plate:

    Visual- Integrity of cord, connector and neutral plate. Instrumental- Adjust multimeter for

    reading resistance at lowest scale. Resistance should remain below 10Ohms.

    Operative- Connect to scalpel and alarm and alarm should not go off.

    _______________________________________________________________Transmai BP-150 30

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    32/37

    Chapter 4- Care and maintenance____________________________________________

    Operating Pedal:

    Visual- Entireness of cord and connector. Instrumental- With multimeter check continuity

    and operating.

    Operative- Connect to scalpel (adjusted at 0% in all modes), press and verify operatingthrough sound and visual signals concerning each operational mode.

    Standard Pen, Autoclavable:

    Visual- Entireness of cord and connector. Insert one of electrodes to pen and check its perfect

    and firm setting. Fit in electrosurgical and check firmness of setting.

    Cord and Nippers, Bipolar:

    Visual- Entireness of cord and connector. Check perfect fitting (firm) of nippers to cord and

    cord to electrosurgical.

    Instrumental- With multimeter check continuity of cords.

    ELECTROSURGICAL: Before any surgical proceeding equipment will be checked andpreventive maintenance is to be made every year.

    Visual- Check entireness of cabinet, perfect fitting of power cord and accessories.

    PROBLEMS AND SOLUTIONS

    Problem Possible Cause Solution

    Burning of fuse Replace fuse

    Power cord Replace power cord1- Turned on it does not work /

    no signals are lightedLack of electrical energy Solve problem

    2- Turned on and light of

    On/Off key is not on.Internal neon lamp Replace lamp

    Power circuit

    Control circuitContact Technical Assistance

    3- Power is very weak

    Pen cord rupture Replace pen cord

    Control potentiometer Replace potentiometer

    Accessories Replace accessories

    4- Power very much reduced at

    same point already selected

    before

    Power circuit Contact Technical Assistance

    5- No sound Loudspeaker Loudspeaker

    Keyboard Replace keyboard6- No current at outlet

    Pedal Replace or repair pedal

    Power circuit

    Control CircuitContact Technical Assistance

    _________________________________________________________________Transmai BP-150 31

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    33/37

    Chapter 4- Care and maintenance____________________________________________

    Note: Any other doubt or problem not mentioned above, please contact Transmai or

    Authorized Technical Service.

    CONCLUSION

    Check always correct work of all equipments involved in surgical operation and observe basic

    rules recommended for electro surgery.

    The factory is not responsible for wrong use of this equipment and advises against

    manipulation by persons with lack of wide and total knowledge of electro surgery techniques.

    TECHNICAL SPECIFICATION

    Monopole Power: Cut: 150W Charge: 300 Ohms

    Blend1: 100W Charge: 300 Ohms

    Blend2: 75W Charge: 300 OhmsBlend3: 20 Charge: 300 Ohms

    Bipolar Power: Cut: 70W Charge: 100 Ohms

    Blend1: 50W Charge: 100 Ohms

    Blend2: 30W Charge: 100 Ohms

    Blend3: 20W Charge: 100 Ohms

    Potential could vary with regard to nominal values in 20% or 10W(at 110Vac), whatever be

    higher.

    Operating Frequency: 480 kHz

    Feeding: 110/220 Vac (Automatic) 580VA - 50/60Hz

    Adequate connection to electrical network CISPR 11

    RF ESCAPE CURRENT: Bipolar :

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    34/37

    Chapter 4- Care and maintenance____________________________________________

    CLASSIFICATION OF PRODUCT:

    Type of protection against electric shock: Class I Equipment

    Grade of protection against electric shock: Applied part of CF type defibrillation-proof. Grade of protection against harmful water penetration: IPX0 Equipment

    Grade of safe application in presence of flammable anesthetic mix with air, oxygen or

    nitrous oxide: non adequate

    Mode of operation: Continuous with intermittent charge Time ON = 10seconds

    - Time OFF = 30seconds

    Physiological effects: Equipment causer of physiological effects to patient and/or operator

    Documentation: Through users request and under company regulations electrical diagrams

    and component parts list will be available.

    Ambient Temperature: 0 - 50o

    C

    Transport and storage:

    Electrosurgical is packed with an extra protection of expanded polyurethane, to absorb

    vibrations and impacts during transport. Keep this material for future transportation.

    In the event of flakes as vibration and impact absorbers, the first packing should be a plastic

    bag to avoid powder going inside of equipment and piling up and maximum weight upon case

    would be lesser than specifications.

    Specifications Electrosurgical only

    Maximum piling 03 units

    Transport and storing temperature -20 to +75C

    Weight upon case during transport and storing 17kg

    Relative humidity of storing 10 to 90%

    OPEN TENSION X SCALE: BP-150

    _______________________________________________________________Transmai BP-150 33

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    35/37

    Chapter 4- Care and maintenance________________________________________________

    POTENTIAL x SCALE x CHARGE CURVES:

    Measures: High: 14,4cmWidth: 23,0cm

    Depth: 23,5cm

    Weight: 3,5kg

    BIBLIOGRAPHY

    1- Patient Safety in the Use of Electro surgical Devices

    U.Pohl

    OR male-nurse, Westphalian Wilhelms University, Mnster

    2- Patient Safety in the Use of Electro surgical Devices

    Wiedner-Heil

    DBfk Verband

    3- Principles of Biomedical Instrumentation and Measurement

    Richard Aston

    Merrill Publishing Company

    _______________________________________________________________Transmai BP-150 34

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    36/37

    Guarantee

    GUARANTEE

    MedGyn Products Inc. assures to owner-consumer of equipment herein identified, guarantee

    against manufacturing defects as long as proved by technician authorized by Transmai, for aperiod of 365 days counted from date of purchase by first buyer-consumer on, for product

    described on invoice.

    Transmai Equipamentos Mdicos Ltda. declares this guarantee invalid for all intents and

    purposes if this equipment suffers any damage caused by accidents, nature elements

    (lightning, floods, collapses, fall, bad use, etc.), use disagreed with Instruction Manual, for

    connection to unsuitable electric wiring or subjected to excessive fluctuations, or in the event

    of presenting violation signals, repairs by technician non authorized by Transmai.

    Note that guarantee will be invalid if purchaser does not present the invoice evidencing

    purchasing equipment or if said document contents erasures or alterations.

    Transmai Equipamentos Mdicos Ltda. is obliged to provide the services above referred.

    Owner-consumer will only be responsible for cost and risks of transport (there and back) of

    equipment to manufacturer.

    ___________________________________________________________________________

    This product is exclusively destined to medic-hospital use

    Transmai Equipamentos Mdicos Hospitalares Ltda.

    Av. Maria Estela, 33 - Jardim Maria Estela

    CEP: 04180-010 - So Paulo SP - Brasil

    + 55 11 2335.1000

    Fax extension 210

    __________________________________________________________________Transmai BP-150 35

  • 7/28/2019 Manual do Usurio BP-150 (ESU-150W) ingles

    37/37

    Guarantee_________________________________________________________________

    CERTIFICATE OF GUARANTEE

    This equipment has guarantee valid for 1 (one) year, counted from date of acquisition on,

    against manufacturing defects under following conditions:

    A-) Devolution of duly filled in Notice-Card to factory, within 10 (ten) days after

    purchase.

    B-) Correct and careful use with specified voltage only.

    C-) Repairs only by authorized distributors or manufacturer.

    D-) Presentation of supporting bill of sale together with this certificate dully filled in

    eventual need of repairs.

    In the event of repairing in factory send together with equipment:

    A-) Distributors: Consignment Note for repairs. (None other document will be accepted).

    ATTENTION:

    B-) Hospitals, Doctors, Clinics, Charities: Letter on heading paper containing name and

    signature of owner or responsible officer, address, phone number, city and state, register

    codes and indicating if previous budget is desired.

    Distributor:__________________________________________________________________

    Address:____________________________________________________________________

    City:_____________________________________ State:_____________________________

    Equipment:_____________________________________________ N__________________

    Invoice:_____________________________________ Date:___________________________

    Transmai Equipamentos Mdicos Hospitalares Ltda.

    Av. Maria Estela, 33 - Jardim Maria EstelaCEP: 04180-010 - So Paulo SP - Brasil

    + 55 11 2335.1000

    Fax extension 210