4008 H / 4008 S
Hemodialysis Machine
Technical Manual
Supplemental Instructions for the OCM Option
Part no.: M32 514 1
CareFreseniusMedical
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
2/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 0-1
Table of Contents
Section Page
1 Description of errors .................................................................................................... 1-11.1 Display messages ........................................................................................................... 1-31.2 Description of errors ....................................................................................................... 1-5
3 Adjustment instructions ............................................................................................... 3-1
3.1 Measuring equipment ..................................................................................................... 3-33.2 Flow diagram .................................................................................................................. 3-43.3 Calibration mode ............................................................................................................. 3-8
4 Calibration program ...................................................................................................... 4-1
6 Setup menu ................................................................................................................... 6-16.1 Overview ......................................................................................................................... 6-1
6.2 Main menu 4008 H/S ...................................................................................................... 6-2
8 Circuit diagrams ............................................................................................................ 8-18.1 Circuit diagrams and component layout LP 632, CPU 2 ............................................... 8-38.2 Circuit diagrams and component layout LP 633-5, Input board .................................... 8-19
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
3/75
0-2 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
4/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 1-1
Table of Contents 1 Description of Errors
Section Page
1.1 Display messages ......................................................................................................... 1-3
1.1.1 Informational messages .................................................................................................. 1-31.1.2 Warnings ......................................................................................................................... 1-4
1.2 Description of errors ..................................................................................................... 1-5
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
5/75
1-2 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
6/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 1-3
1.1 Displaymeldungen
1.1.1 Informative Meldungen
Clear.: XXXml/min Shows the last measured clearance value. Displayed for1 minute after a successful measurement.
Plasma-Na: XXXmmol/l Shows the plasma sodium concentration of the last singlemeasurement for 1 minute.
Kt/V: X.XX Shows the current Kt/V. Displayed for 1 minute when the UFgoal has been reached or if the OD senses clear fluid.
Kt: XX.Xl Displayed instead of “Kt/V:X.XX” if no valid value for Vurea
isavailable. Shows the current value of Kt. Displayed for1 minute when the UF goal has been achieved or if the ODsenses clear fluid.
Kt/V: Not available There are no Kt/V data available as the calculation of Kt/V wasaborted. Displayed for 1 minute when the UF goal has beenachieved or if the OD senses clear fluid.
Enter V(urea)? This message is displayed at the start of a dialysis treatmentprovided the operator has not already entered the V
urea on the
screen. This message will be displayed when the OCM isswitched on manually by the operator or after the optical de-tector has sensed blood, provided “OCM measurement ON”was preselected in the Setup. This message prompts the ope-rator to enter the patient’s volume of urea distribution. It will
only be displayed if the Vurea has not yet been entered. If the Vu-rea
has been changed during the OCM measurement, the Kt/Vwill be recalculated provided it is already available. This mes-sage will be cleared by pressing the Start/Reset key.
HDF/HF < > OCM An OCM and an HDF or HF treatment procedure cannot beperformed simultaneously. Displayed once if OCM is selectedwhile HDF/HF is active. If HDF/HF is selected while OCM is al-ready active, the OCM will be aborted and the message “OCMoff” will be displayed. The OCM can, however, be reselectedafter HDF/HF has been switched off.
SN < > OCM The OCM cannot be started as a Single Needle treatment (SNor SN click-clack) has been selected.
Profile< > OCM An OCM cannot be started in conjunction with a Na-/UF profile1, 5 or 6 and a treatment time of less than 180 minutes. Dis-played when the OCM is switched on and/or when selecting aprofile.
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
7/75
1-4 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
1.1.2 Warnings
The operator has attempted to enter a value for an OCM para-meter that is below the minimum limit.
OCM tech. deact./X The OCM cannot be started as the CD cells or the temperaturesensors are not properly calibrated or have been manually de-activated when operating the machine. Also displayed whenthe integral relationship of the adjustment measurement is out-side the limits. Displayed when the OCM is activated manuallyor when the OD senses dark fluid if the default setting for theOCM is set to active.
Inform the service technician of the error code (X).
OCM off/X The OCM was aborted due to a malfunction and will be deacti-vated until the end of the treatment.Inform the service technician of the error code (X).
OCM adjust Displays during an OCM adjustment.
OCM adjust intended Displays when the optical detector senses dark fluid and if anadjustment measurement is required on completion of thetreatment in progress.
Input value too low
Input value too high The operator has attempted to enter a value for an OCM para-meter that is above the maximum limit.
OCM adjust OK Displays after successful completion of an OCM adjustment.
OCM adjust cancelled The OCM adjustment was not successfully completed and willbe repeated after the next treatment.
Kt/V too low Displays if the prescribed Kt/V cannot be reached at the end ofthe treatment with the current Kt/V.
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
8/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 1-5
Error message displayed Description
OCM OFF The OCM option will be deactivated until the machine is
switched off. – Power failure with battery backup
– CPU II error
OCM OFF/2 The OCM option will be deactivated until the machine isswitched off.One of the CD cells has measured a value below 11.5 mS/cmduring more than 3 consecutive OCM measurements.
– Check CD cells for proper connection
– P.C.B. LP 633-5 defective
OCM OFF/3 The OCM option will be deactivated until the machine isswitched off.
– System error
OCM OFF/5 The OCM option will be deactivated until the machine isswitched off.
A downward pulse occurs although an upward pulse was re-quested.
– Check the bibag/concentrate delivery – Check the concentrate pumps / mixing system
– Air in the dialysate
OCM OFF/6 The OCM option will be deactivated until the machine isswitched off.
– System error
OCM OFF/7 The OCM option will be deactivated until the machine isswitched off.
Implausibility of the ratio of blood flow to clearance or no appa-rent change on the CD cell during more than three consecutivemeasurements.
– Poor terminal connection of the CD cell
1.2 Description of errors
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
9/75
1-6 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
Error message displayed Description
OCM tech. deact./1 The OCM option is decalibrated and can only be reactivatedafter a successful pulse calibration.Deviation between the 8-bit and the 12-bit CD value of the pre-cell > 0.5 ms/cm for > 20 s.
– Disinfection – Check OCM calibration and calibrate, if necessary
OCM tech. deact./4 The OCM option is decalibrated and can only be reactivatedafter a successful pulse calibration.Three cyclic pressure holding tests (CPHT) showed a deviationof > 0.5 ms/cm between CD cell 7 and CD cell 110 for > 4.5 s.
– Disinfection – Check OCM calibration and calibrate, if necessary
OCM tech. deact./7 The OCM option is decalibrated and can only be reactivatedafter a successful pulse calibration.
More than 5 consecutive measurements showed a clearance ofmore than 1.3 times the blood flow or below zero.
– Check the CD cells for poor terminal connection. – Disinfect, if necessary – Check the calibration
OCM tech. deact./8 The OCM option is decalibrated and can only be reactivatedafter a successful pulse calibration.Zero pulse measurement failure or time of maximum pulse be-tween pre- and post-CD cell > 13 s.
– Check the hydraulic for proper flow
– Check OCM calibration and calibrate, if necessary
OCM tech. deact./9 The OCM option is decalibrated and can only be reactivatedafter a successful pulse calibration.The correction factor of the zero pulse measurement for upwardor downward pulses is outside the acceptable range of 0.8 to1.2.
– Perform a disinfection – Check OCM calibration and calibrate, if necessary
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
10/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 3-1
Table of Contents 3 Adjustment instructions
Section Page
3.1 Measuring equipment ................................................................................................... 3-3
3.2 Flow diagram ................................................................................................................. 3-43.2.1 Flow diagram 4008 H with OCM ..................................................................................... 3-43.2.2 Flow diagram 4008 S with OCM ..................................................................................... 3-6
3.3 Calibration mode ........................................................................................................... 3-83.3.1 Basic conditions .............................................................................................................. 3-83.3.2 Display messages (4008 H/S) ........................................................................................ 3-8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
11/75
3-2 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
12/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 3-3
3.1 Measuring equipment
CautionThe accuracy of the measuring equipment used during the calibration procedureis decisive for the accuracy of the OCM measurement.
The measuring equipment used for the calibration of the conductivity must havean accuracy of 0.05 mS/cm within a temperature range of 35 °C – 39 °C.We recommend to use the UMED universal measuring device which is availablefrom Fresenius (part no. M32 403 1).
UMED Service CasePart number M32 403 1
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
13/75
3-4 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
3.2 Flow diagram
3.2.1 Flow diagram 4008 H with OCM
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
14/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 3-5
Legend
2 Temperature sensor3 Temperature sensor4 Temperature sensor5 Float switch6 Level sensor7 Conductivity cell
8 Blood leak detector9 Pressure transducer10 Reed contact, concentrate12 Reed contact, bicarbonate21 Flow pump22 UF pump23 Concentrate pump24 Dialyzer valve 124b Dialyzer valve 225 Bicarbonate pump26 Bypass valve29 Degassing pump30 Outlet valve
31 Balancing chamber valve 132 Balancing chamber valve 233 Balancing chamber valve 334 Balancing chamber valve 435 Balancing chamber valve 536 Balancing chamber valve 637 Balancing chamber valve 738 Balancing chamber valve 839 Negative pressure valve41 Water inlet valve43 Fill valve54 Heater rod61 Pressure reducing valve63 Water inlet filter65 Loading pressure valve66 Heater block66a Water inflow chamber66b Heater rod chamber66c Float chamber68 Balancing chamber71 Filter / concentrate72 Filter / bicarbonate73 Filter / dialysate ext.74 Filter / UF75 External flow indicator76 Filter / fill valve77 Heat exchanger
78 Relief valve84 Disinfectant valve85 Disinfectant connector86 Recirculation valve87 Drain valve88 Multifunction block88a Degassing chamber88b Secondary air separator88c Primary air separator89 Degassing orifice90a Rinse chamber / concentrate90b Rinse chamber / bicarbonate91 Rinse valve
92 Vent valve94 Concentrate suction tube
95 Bicarbonate suction tube97 Air separation pump99 Rinse valve100 Rinse valve102 Central concentrate delivery valve104 Central bicarbonate delivery valve
109 Temperature sensor110 Conductivity cell111 Hydrophobic filter112 Vent valve114 Dialysate filter115 Sensor Disinfection valve116 Sampling valve117 Check valve / concentrate118 Check valve / bicarbonate119 Filter / concentrate120 Filter / bicarbonate121 Central concentrate delivery connector122 Central bicarbonate delivery connector
123 Pressure switch for V102124 Pressure switch for V104130 biBag drain valve132 biBag conductivity cell133 biBag temperature sensor134 biBag pressure transducer136 biBag connector137 biBag micro-switch 1138 biBag micro-switch 2148 Filter / rinse valve V100149 Filter / rinse valve V99150 Filter / fill valve151 Orifice181 Temperature compensation plate 2182 Pressure transducer 2183 Test valve184 Filter / test valve185 Compressor188 Evacuation valve189 Retentate valve190 Online filter191 Online valve 3192 Online valve 2193 Online valve 1194 Rinse port195 Substituate port201 Air separator / concentrate
202 Level sensor / concentrate203 Air separator / bicarbonate204 Level sensor / bicarbonate205 Mixing point / concentrate / bicarbonate206 Buffer volume chamber210 Filter
Hydraulics measuring points:
A Reduced water inlet pressureB Balancing chamber loading pressureC Flow pump pressure
D Degassing pump pressure
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
15/75
3-6 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
3.2.2 Flow diagram 4008 S with OCM
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
16/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 3-7
Legende
2 Temperature sensor3 Temperature sensor4 Temperature sensor5 Float switch6 Level sensor7 Conductivity cell
8 Blood leak detector9 Pressure transducer10 Reed contact, concentrate12 Reed contact, bicarbonate21 Flow pump22 UF pump23 Concentrate pump24 Dialyzer valve 124b Dialyzer valve 225 Bicarbonate pump26 Bypass valve29 Degassing pump30 Outlet valve
31 Balancing chamber valve 132 Balancing chamber valve 233 Balancing chamber valve 334 Balancing chamber valve 435 Balancing chamber valve 536 Balancing chamber valve 637 Balancing chamber valve 738 Balancing chamber valve 839 Negative pressure valve41 Water inlet valve43 Fill valve54 Heater rod61 Pressure reducing valve63 Water inlet filter65 Loading pressure valve66 Heater block66a Water inflow chamber66b Heater rod chamber66c Float chamber68 Balancing chamber71 Filter / concentrate72 Filter / bicarbonate73 Filter / dialysate ext.74 Filter / UF75 External flow indicator76 Filter / fill valve77 Heat exchanger
78 Relief valve84 Disinfectant valve85 Disinfectant connector86 Recirculation valve87 Drain valve88 Multifunction block88a Degassing chamber88b Secondary air separator88c Primary air separator89 Degassing orifice90a Rinse chamber / concentrate90b Rinse chamber / bicarbonate91 Rinse valve
92 Vent valve94 Concentrate suction tube
95 Bicarbonate suction tube97 Air separation pump99 Rinse valve100 Rinse valve102 Central concentrate delivery valve104 Central bicarbonate delivery valve
109 Temperature sensor110 Conductivity cell111 Hydrophobic filter112 Vent valve114 Dialysate filter115 Sensor Disinfection valve116 Sampling valve117 Check valve / concentrate118 Check valve / bicarbonate119 Filter / concentrate120 Filter / bicarbonate121 Central concentrate delivery connector122 Central bicarbonate delivery connector
123 Pressure switch for V102124 Pressure switch for V104130 biBag drain valve132 biBag conductivity cell133 biBag temperature sensor134 biBag pressure transducer136 biBag connector137 biBag micro-switch 1138 biBag micro-switch 2148 Filter / rinse valve V100149 Filter / rinse valve V99150 Filter / fill valve151 Orifice181 Temperature compensation plate 2182 Pressure transducer 2183 Test valve184 Filter / test valve185 Compressor188 Evacuation valve189 Retentate valve190 Online filter191 Online valve 3192 Online valve 2193 Online valve 1194 Rinse port195 Substituate port201 Air separator / concentrate
202 Level sensor / concentrate203 Air separator / bicarbonate204 Level sensor / bicarbonate205 Mixing point / concentrate / bicarbonate206 Buffer volume chamber210 Filter
Hydraulics measuring points:
A Reduced water inlet pressureB Balancing chamber loading pressureC Flow pump pressure
D Degassing pump pressure
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
17/75
3-8 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
3.3 Calibration mode
3.3.1 Basic conditions
– The hemodialysis machine must beswitched off.
– The service switch must be in the ON(up) position.
– Turn the hemodialysis machine on.
3.3.2 Display messages (4008 H/S)
7 8 9 0
4
1
5 6
2 3 C Best
Esc
CALIBRATION
004.7 Volt
011.9 Volt
023.6 Volt
020.4 Volt
12 V voltage (11.5 V – 12.5 V)
Watchdog supply voltage (4.5 V – 5.5 V)
Battery charging voltage (> 20 V)
24 V voltage (23 V – 25 V)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
18/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-1
CALIBRATION
Confirm
key
DIAGNOSTICS
MISCELLANEOUS
SETUP MENU
4 Calibration program
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
19/75
4-2 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
● Main menu
CALIBRATION
Confirmkey
CAL. ART. PRESSURE
CAL. ART. P_MODULE
CALIB. (B)-PUMP-RATE
CAL. VENOUS PRESSURE
CAL. VEN. P_MODULE
ADJ. UF-PUMP VOLUME
CAL. DEGAS. PRESSURE
CAL. FLOW 300 ml/min
CAL. FLOW 500 ml/min
CAL. FLOW 800 ml/min
CALIB. TEMPERATURE
Confirmkey see Part 11
CAL. MIXING SYSTEM
CALIB. CONDUCTIVITY
Confirmkey see Part 13
CAL. DIAL. PRESSURE
CALIBRATE BLD
CALIB. BIBAG VALUES
NOVRAM
back to main menu ?
Confirmkey
RESET FAILURE RECORD
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
20/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-3
● Part 11: Calibrating the dialysate temperature
CALIB. TEMPERATURE
Confirmkey
Temp. ADJUSTMENT
Confirmkey
see Part 11.1
Check TEMPERATURE
Confirmkey
see Part 11.2
OCM TEMP.-CheckConfirm
keysee Part 11.3
back to menu?
Confirmkey
NoteAccuracy of the external meter to be connected: ±0.2 °C.
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
21/75
4-4 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
Temp. ADJUSTMENT
Confirmkey
DATA STORED
Connect the externalreference instrument.
Connect the BIC suctiontube to the BIC container
or place it in waterof room temperature.
Tone Mutekey
approx. 3 sec
adj. temp to 37°CEsckey
Adjust the digital value inthe “UF Rate” window ➁ by
pressing the +/– keys.
Wait until the temperatureon the external reference
instrument indicates 37.0 °C.
DATA STORED
Machine in bypass
Tone Mutekey
approx. 3 sec
OCM TEMP. AdjustEsckey
Wait for the temperatureto stabilize (37 °C)
Indicated in "UF rate"window ➁
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
adj. temp to 37°C
XXXX ADC-digits
XXXX ADC-digits
XXXX ADC-digits
XXXX DAC-digits
➀ NTC3; 8 bit
➁ Control 8 bit
➂ NTC3; 12 bit
➃ NTC109; 8 bit,if NTC 109 is activated in SETUP
OCM TEMP.-Adjust
XXXX ADC-digits
XXXX ADC-digits
XXXX DAC-digits
XXXX °C
➀ NTC3; 12 bit
➁ NTC3 temperature
➂ NTC4; 12 bit
➃ Dynamic measurement range switching
1
2
3
4
● Part 11.1: Adjusting the dialysate temperature
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
22/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-5
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
● Part 11.2: Checking the dialysate temperature
Check TEMPERATURE
Confirmkey
use to select the
Temperature setting field
act temp = 37.0 °C
Repeat theprocedure
with differentvalues
(e.g. 35/39 °C)
Check whether thepreselected temperature
is achieved.Check: reference instrument,
display and 37.0 °CLED display correspondto 115/116 digits in the“UF Volume” window ➀.
Tolerance: ±0.5 °C
ACT temp = XX.X °C
Esckey
approx. 3 sec
ACKNOWLEDGED
Tone mutekey
Specify a temperature bypressing the +/ – keys.
set temp = XX.X °C
act. temp = 37.0 °C
XXXX ADC-digits
XXX.X °C
XXXX ADC-digits
Alpha
display
Temperature
adjustment
➀ NTC3; 8 bit
➁
➂ NTC109; temperature
➃ NTC109; 8 bit
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
23/75
4-6 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
● Part 11.3: Checking the dialysate temperature for OCM
OCM TEMP.-Check
Confirmkey
use to select the
Temperature setting field
act temp = 37.0 °C
Repeat theprocedure
with differentvalues
(e.g. 35/39 °C)
Check whether thepreselected temperature
is achieved.Check: Display,
37.0 °C bargraph displayTolerance of the
displayed temperaturein the UF rate field ➁
and the UF time field ➃:±0.2 °C
ACT temp = XX.X °C
Esckey
approx. 3 sec
ACKNOWLEDGED
Tone Mutekey
Specify a temperature bypressing the +/ – keys.
set temp = XX.X °C
Open the shunt coverto switch the machine
to the bypass mode
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
OCM Temp.-Check
XXXX ADC-digits
XX.XX °C
XXXX ADC-digits
XX.XX °C
Alpha
display
Temperature
adjustment
➀ NTC3; 12 bit
➁ NTC3; temperature
➂ NTC4; 12 bit
➃ NTC4; temperature
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
24/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-7
● Part 13: Calibrating the conductivity
CALIB. CONDUCTIVITY
Confirmkey
CONDUCTIVITY Set
Confirmkey
see Part 13.1
TEMP/COND COMP Calib
Confirmkey
see Part 13.3
OCM PULSE calib.
Confirmkey
see Part 13.4
CONDUCTIVITY Check
Confirmkey
see Part 13.5
OCM COND.-Check
Confirmkey
see Part 13.7
TEMP/COND COMP CheckConfirm
keysee Part 13.8
back to menu?
Confirmkey
COND. with Concentr. see Part 13.2Confirmkey
COND Check with ConcConfirm
keysee Part 13.6
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
25/75
4-8 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
● Part 13.1: Setting the conductivity
CONDUCTIVITY Set
Confirmkey
CD set: 12.80 mS/cm
CD set: 12.80 mS/cm
CD set: 15.70 mS/cm
Esckey
Use tomove to the
Concentrate setting field.
Use +/ – to reducethe concentrate pump
volume to get to a CD inthe lower display range.
Conc. set = ±0 %If no key ispressed forapprox. 4 sec
Tone Mutekey
ACKNOWLEDGED
approx. 3 sec
Wait for the CD to stabilize.Compare the CD with theexternal reference meter.
In case of a deviation, adjust the dis-play by pressing the +/ – keys.
Tone Mutekey
ACKNOWLEDGED
approx. 3 sec
Use tomove to the
Concentrate setting field.
Use +/ – to raisethe concentrate pump
volume to get to a CD inthe lower display range.
conc. set = –X %If no key ispressed forapprox. 4 sec
e.g.:
Tone Mutekey
Conc. set. = +10 %
ACKNOWLEDGED
approx. 3 sec
CD set: XX.X mS/cm
Wait for the CD to stabilize.Compare the CD with the
external reference instrument.In case of a deviation, adjust the dis-
play by pressing the +/ – keys.
Tone Mutekey
DATA STORED
approx. 3 sec
Conc. set. = –X %
CD set: 12.80 mS/cm
XXXX ADC-digits[ADC-digits (OCM)]
XXXX steps(concentrate pump)
XXXX steps(bicarbonate pump)
Concentrate
setting
XXXX ADC-digits[ADC-digits (OCM)]
Bicarbonate
setting
Alpha
display
➀ CD cell 7; 12 bit / 8 bit alternating
➁
➂
➃ CD cell 110; 12 bit/ 8 bit alternating
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
26/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-9
● Part 13.2: Setting the conductivity with concentrate
NoteAn adjustment of the conductivity using a calibration concentrate is currentlynot possible, since a suitable calibration concentrate is not available.
COND. with Concentr.
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
27/75
4-10 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
● Part 13.3: Setting the temperature/conductivity compensation
If the conductivity isoutside the limits:use to select
the concentrate settingfield
TEMP/COND COMP Calib.
Confirmkey
approx. 3 sec
act temp = XX.XX °CEsckey
conc. set = +X %
ACKNOWLEDGED
Tone Mutekey
approx. 3 sec
act temp = XX.XX°C
If no key ispressed forapprox. 4 sec
conc. set = ±0 %
Use +/ – to reduce/increasethe concentrate pump volume
until the conductivity isinside the alarm limits
13.9 – 14.5 mS/cm
Verify that the presettemperature of 35 °C is achieved
and that it has stabilized.Check: Display,
LED bargraph displayTolerance: ±0.5 °C
ACKNOWLEDGED
approx. 3 sec
act temp = XX.XX °C
DATA STORED
Tone Mutekey
Verifythat the conductivity of
both measuring cells in the"UF rate" (➁) and the "UF time"(➃)
window is identical and that it iswithin 13.9 to 14.5 mS/cm.
Tone Mutekey
Verify that the presettemperature of 39 °C is achieved
and that it has stabilized.Check: Display,
LED bargraph displayTolerance: ±0.5 °C
Machine in bypass
act temp = XX.XX °C
XX.XX mS/cm
Temperature
adjustment
XX.XX mS/cm
XX.XX mS/cm
XX.XX mS/cm
Concentrate
setting
Bicarbonate
setting
act temp = XX.XX °C
XX.XX mS/cm
XX.XX mS/cm
XX.XX mS/cm
XX.XX mS/cm
➀ CD cell 7,Value in mS/cm after confirmation of 35°C
➁ CD cell 7, current value in mS/cm
➂ CD cell 110,Value in mS/cm after confirmation of 35°C
➃
CD cell 110, current value in mS/cm
Temperature
adjustment
Alpha
display
Alpha
display
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
28/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-11
● Part 13.4: OCM pulse calibration
OCM PULSE calibr.
Confirmkey
DATA STORED
Tone Mutekey
approx. 3 sec
Start PULSE calib.?
Tone Mutekey
Esckey
OCM PULSE UP activ
Positive OCM pulseduration
Esckey
OCM PULSE DOWN activ
Negative OCM pulseduration
Esckey
OCM PULSE finished
approx. 3 sec
Tone muteLED is flashing.
Info sound
pos. Dialy: –XXXX.X neg. Dialy: XXXX.X
alternating
Tolerance per pulse: ±25
Verify that theconductivity is
within13.9 to 14.6 msand wait for the
conductivity to stabilize.
NoteIf the value is outside ±25,check the conductivity mea-suring system of the dialysissystem.
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
OCM PULSE UP activ
XXXX Seconds
XXXX Index
XXXX
XXXX mS/cm
➀ Time elapsed since the pulse calibrationwas started
➁ Counter
➂Concentrate setting
➃ CD cell 7,compensated CD value
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
29/75
4-12 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
● Part 13.5: Conductivity check
CONDUCTIVITY Check
Confirmkey
act. cond: XX.X mS/cm
use to selectthe Concentrate
setting field
Esckey
Verify the conductivity.Check:
Reference meter, display,LED bargraph display
conc. set = ± 0%
Use the +/ – keysto reduce the
concentrate pump volume
approx. 3 sec
act. cond: XX.X mS/cm
Repeat theprocedure withdifferent values
If no key ispressed forapprox. 4 sec
conc. set = –6 %
ACKNOWLEDGED
Tone Mutekey
e.g.:
➀ CD cell 7; 12 bit / 8 bit alternating➁
➂
➃ CD cell 110; 12 bit / 8 bit alternating
act. cond: XX.X mS/cm
XXXX
XXXX
XXXX
XXXX
ADC-digits[ADC-digits (OCM)]
steps(concentrate pump)
steps(bicarbonate pump)
Concentrate
setting
ADC-digits[ADC-digits (OCM)]
Temperature
adjustment
Bicarbonate
settingAlpha
display
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
30/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-13
● Part 13.6: Conductivity test with concentrate
NoteA calibration of the conductivity using a calibration concentrate is currently notpossible, since a suitable calibration concentrate is not available.
COND Check with Conc
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
31/75
4-14 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
● Part 13.7: OCM conductivity check
OCM COND.-Check
Confirmkey
OCM Cond.-Check
Use toselect the
concentrate setting field
Esckey
Verify the conductivity.Check:
Reference meter, display,LED bargraph display
conc. set = ± 0%
Use the +/ – keysto reduce the
concentrate pump volume.
approx. 3 sec
OCM Cond.-Check
Repeat theprocedure withdifferent values
If no key ispressed forapprox. 4 sec
conc. set = –6 %
ACKNOWLEDGED
Tone Mutekey
e.g.:
➀ CD cell 7; 12 bit➁ CD cell 7
➂ CD cell 110;12 bit
➃ CD cell 110
OCM Cond.-Check
XXXX ADC-digits (OCM)
XX.XX mS/cm
XXXX ADC-digits (OCM)
XX.XX mS/cm
Alpha
display
Concentrate
setting
Temperature
adjustment
Bicarbonate
setting
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
32/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 4-15
● Part 13.8: Temperature/conductivity compensation check
The individualconductivities and their
compensation factors aredisplayed in the fields
(see 4008 H/S)
TEMP/COND COMP Check
Confirmkey
act temp = XX.XX °C
The conductivity cellscan be crosschecked
by changing theconcentration
or temperature.
Esckey
➀ CD cell 7, compensated
➁ CD cell 7, compensation factor
➂ CD cell 110, compensated
➃ CD cell 110, compensation factor
7 8 9 0
4
1
5 6
2 3 C Conf
Esc
1
2
3
4
act temp = XX.XX °C
XX.XX mS/cm
XXXX mS/(cm•°C)
XX.XX mS/cm
XXXX mS/(cm•°C)
Alpha
display
Concentrate
setting
Temperature
adjustment
Bicarbonate
setting
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
33/75
4-16 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
34/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 6-1
6.1 Overview
CALIBRATION
DIAGNOSTICS
MISCELLANEOUS
SETUP MENU
Confirmkey
6 Setup Menu
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
35/75
6-2 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
6.2 Main menu 4008 H/S
SETUP MENU
Confirmkey
SET ALARM/WARN TIME
SETUP CLEANING PGM
INFO SOUND (C-PGM)
SETUP DILUTION
SET CONDUCT. LIMIT
SET DIAL PARAMETERS
AUTOM. SN-START
ACTIV. MONIT_NTC109
ACTIV. STD UF-DATA
SET STD. PRIME-TIME
SOUND I/O-SWITCH
SET CENTRAL-DELIVERY
STORE DEFAULT VALUES
BACK TO MAIN MENU ?
T1 –TEST AUTOSTART
SET RINSE –VOLUME
BPR/UFR –WARNING
SET KEY-CLICK
HAEMOGLOBIN UNIT
OCM SETTINGSConfirm
keysee Part 9
ONLINE plus SETTINGS
CONFIRMKEY
CALC. CUMUL. BLOOD-VOL
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
36/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 6-3
● Part 9: OCM Settings
● Part 9.1: Activating the OCM measurement
OCM SETTINGS
Confirmkey
OCM MEASUREMENT
Confirmkey
see Part 9.1
OCM ZERO MEASUREMENT
Confirmkey
see Part 9.2
OCM DELAY TIME
Confirmkey
see Part 9.3
OCM CLR CORR FACTOR
Confirmkey
see Part 9.4
OCM BASELINE DIFFConfirm
keysee Part 9.5
back to menu?
Confirmkey
OCM INTEGRAL VALUEConfirm
keysee Part 9.6
OCM KT/V WARNLEVELConfirm
keysee Part 9.7
OCM MEASUREMENT
Confirmkey
OCM Measurement: OFF
+/ –
Tone Mutekey
OCM Measurement: ON
DATA STORED
Esckey
Tone Mutekey DATA STORED
Esckey
Afterapprox. 3 sec
Afterapprox. 3 sec
+/ –
(Default: OFF)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
37/75
6-4 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
● Part 9.2: Activating the OCM zero measurement
OCM ZERO MEASUREMENT
Confirmkey
OCM Zero Measurement: OFF
Tone MutekeyOCM Zero Measurement: ON DATA STORED
Esckey
Tone Mutekey DATA STORED
Esckey
Afterapprox. 3 sec
Afterapprox. 3 sec
+/ –
+/ –
(Default: ON)
NoteIf the OCM zero measurement is set to “OFF”, the OCM option will be deacti-vated.When the OCM zero measurement is reactivated (ON) it is imperative to per-form an OCM pulse calibration.
● Part 9.3: Setting the OCM delay time
OCM DELAY TIME
Delay Time: XX s
Use +/ – to set the valueshown in the table on the right
(1 – 70 seconds)
Confirmkey
Tone Mutekey
DATA STORED
After approx. 3 sec
Esckey
Machine Hydraulic unit type Time
4008 H AutoPRIME™ 15 sec4008 H ONLINEplus ™ 65 sec
4008 S AutoPRIME™ 18 sec
4008 S ONLINEplus ™ 67 sec
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
38/75
Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM) 6-5
● Part 9.4: Setting the OCM correction factor
OCM CLR CORR FACTOR
Confirmkey
ClrCorr Factor: 1.000
The value displayedis the
default value
Tone Mutekey
DATA STORED
After approx. 3 sec
Esckey
NoteDo not change the default value.
NoteIn 4008 H/S machines it is generally possible to define a correction factor forall clearance measurements. This feature has been provided to be able toadapt a machine at its place of installation to latest knowledge gained fromscientific discussion that has not been completed yet. Based on present know-ledge the correction factor for Fresenius Polysulfone membranes must be ad- justed to “1” (factory setting). The accuracies specified are only applicable forFresenius Polysulfone membranes in combination with a correction factor thatwas set to “1”.
● Part 9.5: Setting the OCM baseline difference
OCM BASELINE DIFF
Base Diff.: 0.015 ms/cm
The value displayedis the
default value
Confirmkey
Tone Mutekey
DATA STORED
After approx. 3 sec
Esckey
NoteDo not change the default value.
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
39/75
6-6 Fresenius Medical Care 4008 OCM Supplement 2/05.00 (TM)
● Part 9.6: Setting the OCM integral value
OCM INTEGRAL VALUE
Integral value: 55
The value displayedis the
default value
Confirmkey
Tone Mutekey
DATA STORED
After approx. 3 sec
Esckey
NoteDo not change the default value.
● Part 9.7: Setting the OCM KT/V warning level
OCM KT/V WARNLEVEL
Kt/V Warnlevel: 85 %
Use +/ – to set thewarning level
(Default: 85)Range: 0 – 99 %0 = no warning
Confirmkey
Tone Mutekey
DATA STORED
After approx. 3 sec
Esckey
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
40/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-1
Table of Contents 8 Circuit diagrams
Section Page
8.1 Circuit diagrams and component layout LP 632, CPU 2 .......................................... 8-3
8.2 Circuit diagrams and component layout LP 633-5, Input board ............................. 8-19
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
41/75
8-2 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
42/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-3
LP 632, CPU 2Component layout
8.1 Circuit diagrams and component layoutLP 632, CPU2
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
43/75
8-4 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
44/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-5
LP 632, CPU 2Circuit diagram 1/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
45/75
8-6 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
46/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-7
LP 632, CPU 2Circuit diagram 2/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
47/75
8-8 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
48/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-9
LP 632, CPU 2Circuit diagram 3/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
49/75
8-10 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
50/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-11
LP 632, CPU 2Circuit diagram 4/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
51/75
8-12 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
52/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-13
LP 632, CPU 2Circuit diagram 5/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
53/75
8-14 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
54/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-15
LP 632, CPU 2Circuit diagram 6/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
55/75
8-16 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
56/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-17
LP 632, CPU 2Circuit diagram 7/7
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
57/75
8-18 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
58/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-19
LP 633-5Component layout
8.2 Circuit diagrams and component layoutLP 633-5, Input board
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
59/75
8-20 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
60/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-21
LP 633Circuit diagram 1/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
61/75
8-22 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
62/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-23
LP 633Circuit diagram 2/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
63/75
8-24 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
64/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-25
LP 633Circuit diagram 3/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
65/75
8-26 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
66/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-27
LP 633Circuit diagram 4/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
67/75
8-28 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
68/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-29
LP 633Circuit diagram 5/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
69/75
8-30 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
70/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-31
LP 633Circuit diagram 6/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
71/75
8-32 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
72/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-33
LP 633Circuit diagram 7/8
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
73/75
8-34 Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM)
8/19/2019 MANUAL TECNICO-MAQUINA DE DIALISIS-4008H-4008S.pdf
74/75
Fresenius Medical Care 4008 OCM Supplement 2/10.01 (TM) 8-35
LP 633Circuit diagram 8/8