Operator’s Manual
16ch T/R Hand Wrist Coil
for GE 1.5T and 3.0T MRI Systems
Model Number:
GE QED
5768098‐2 (1.5T) /5561531‐2 (3.0T)
Q7000180 (1.5T) /Q7000152 (3.0T)
2 | P a g e 6000683 Rev. 2
Warranty and Liability
The responsibility for maintenance and management of the product after delivery resides with the customer who
has purchased the product. The warranty does not cover the following items, even during the warranty period:
Damage or loss due to misuse or abuse.
Damage or loss caused by Acts of God such as fires, earthquakes, floods, lightning, etc.
Damage or loss caused by failure to meet the specified conditions for this equipment, such as inadequate
power supply, improper installation, or unacceptable environmental conditions.
Damage due to changes or modifications made to the product.
In no event shall QED be liable for the following:
Damage loss or problems caused by relocation, modification, or repair performed by personnel not
explicitly authorized by QED.
Damage or loss that results from negligence or from ignoring the precautions and operating instructions
contained in this operation manual.
Transportation and Storage Conditions
NOTICE: THIS EQUIPMENT SHALL BE TRANSPORTED AND STORED UNDER THE FOLLOWING CONDITIONS:
1. Ambient temperature range of ‐40°C to +70°C
2. Relative humidity range of 10% to 100%
3. Atmospheric pressure range of 50 kPa to 106 kPa
Medical Device Directive
This product conforms to the requirements of council directive 93/42/EEC concerning medical devices when it
bears the following CE mark of conformity:
Authorized Representative in Europe:
Medical Device Safety Service GmbH (MDSS)
Schiffgraben 41
30175 Hannover
Germany
United States Federal Law
Caution: Federal law restricts this device to sale, distribution, and use by or on the order of a physician. The
device is limited by Federal Law to investigational use for indications not in the Indications Statement.
Issue Date: January 2018
3 | P a g e 6000683 Rev. 2
Introduction
This manual contains detailed information on the safety precautions, use and care of the 1.5T
16ch T/R Hand Wrist Coil (5768098‐2) and 3.0T 16ch T/R Hand Wrist Coil (5561531‐2). For
safety and accuracy in using the product, read this manual as well as the MRI system operation
manual carefully prior to operation of the product. This manual does not include instructions or
safety information on equipment not provided by QED. Please consult the original equipment
manufacturer for information regarding non‐QED equipment.
Compatibility
The 1.5T 16ch T/R Hand Wrist Coil is compatible with GE 1.5T MRI Systems and 3T 16ch T/R
Hand Wrist Coil is compatible with GE 3.0TMRI Systems.
User Profile
Operator – Radiological technologists, laboratory technologists, physicians (note, however, that
all applicable laws in the relevant country must be followed).
User training – No special training is required to use this coil (however, GE provides a
comprehensive training course for MRI systems in order to instruct operators on the correct use
of MRI systems).
Patient Information
Age, health, condition – No special limitations
Weight – 550 lbs. or less (consult the operation manual for the MRI system, and if the maximum
allowable patient's weight for the system is lower than that for this coil, priority must be given
to the maximum weight for the system).
4 | P a g e 6000683 Rev. 2
Table of Contents
Introduction .................................................................................................................................................. 3 Compatibility ................................................................................................................................................. 3 User Profile ................................................................................................................................................... 3 Patient Information ....................................................................................................................................... 3 Table of Contents .......................................................................................................................................... 4 Chapter 1 – 16ch T/R Hand Wrist Coil Components ..................................................................................... 5 Chapter 2 – Safety ......................................................................................................................................... 7 Symbols ..................................................................................................................................................... 7 Indications ................................................................................................................................................. 8 Contraindications ...................................................................................................................................... 8 Precautions ............................................................................................................................................... 8 Cautions – RF Coil ...................................................................................................................................... 9 Emergency Procedures ........................................................................................................................... 10
Chapter 3 – TR Port Location ...................................................................................................................... 11 TR Port Location ...................................................................................................................................... 11
Chapter 4 – Quality Assurance .................................................................................................................... 11 Scanner Verification ................................................................................................................................ 11 Signal to Noise Ratio (SNR) Test .............................................................................................................. 11 Multi‐Coil Quality Assurance (MCQA) Tool ............................................................................................. 18 Using MCQA Viewer ................................................................................................................................ 21
Chapter 5 – Coil Setup and Use .................................................................................................................. 22 Determine Scan Position ......................................................................................................................... 22 Positioning the 16ch T/R Hand Wrist Coil: Horizontal Base .................................................................... 25 Positioning the 16ch T/R Hand Wrist Coil: Vertical Base ........................................................................ 27 Pad Configuration ................................................................................................................................... 30 Position the Patient: Horizontal Base ...................................................................................................... 31 Position the Patient: Vertical Base .......................................................................................................... 33 Lock the Coil ............................................................................................................................................ 35 Landmark ................................................................................................................................................ 36
Chapter 6 – Cleaning, Maintenance, Service, and Disposal ........................................................................ 39 Cleaning the RF Coil ................................................................................................................................ 39 Disinfection ............................................................................................................................................. 39 Maintenance ........................................................................................................................................... 40 Service ..................................................................................................................................................... 40 Disposal ................................................................................................................................................... 40
5 | P a g e 6000683 Rev. 2
Chapter 1 – 16ch T/R Hand Wrist Coil Components
The 16ch T/R Hand Wrist Coil is shipped with the parts shown below. Upon receipt, please
ensure that all parts are included in the shipment.
7
4
5
6
8
9
10
1
3
2
10
6 | P a g e 6000683 Rev. 2
Item # Description Qty GE Part # QED Part #
1 16ch T/R Hand Wrist Coil 1
5768098‐2 (1.5T) /
5561531‐2 (3.0T)
Q7000180 (1.5T) /
Q7000152 (3.0T)
2 16ch T/R Hand Wrist Coil – Horizontal Baseplate 1 5561531‐4 2001768
3 16ch T/R Hand Wrist Coil – Side‐mount Baseplate 1 5561531‐5 2001769
4 16ch T/R Hand Wrist Coil – Posterior Liner Pad 1 5561531‐6 3004567
5 16ch T/R Hand Wrist Coil – Anterior Liner/Phantom Position Pad
1 5561531‐7 3004566
6 16ch T/R Hand Wrist Coil – Palm Pad 1 5561531‐15 3004964
7 16ch T/R Hand Wrist Coil – Wedge Pad 1 5561531‐8 3004751
8 16ch T/R Hand Wrist Coil – Elbow/Arm Pad 1 5561531‐9 3004607
9 16ch T/R Hand Wrist Coil – Wrist Coil Filler Pad 1 5561531‐10 3004716
10 16ch T/R Hand Wrist Coil – Side‐mount Base Pad 2 5561531‐11 3004612
Coil weight: 3.9kg (8.5lb)
7 | P a g e
Chapt
Th
w
Symbo
e
ter 2 –
his section de
when this coil
When us
safety in
ols
Symbol
Safety
escribes the g
is used.
sing the MRI s
nformation de
Number
0434A
1641
5.4.3
5172
5333
3082
2497
6192
5.1.2
2493
2498
N/A
0632
general preca
system, also
escribed in th
Standa
ISO 700IEC 604
ISO 700IEC 604
ISO 1522
ISO 700IEC 604
ISO 700IEC 604
ISO 700IEC 604
ISO 700IEC 604
ISO 700IEC 604
ISO 1522
ISO 700IEC 604
ISO 700IEC 604
N/A
ISO 700IEC 604
utions and sa
refer to the c
e operation m
rd Title,
00 17
Cautithe doperato avo
00 17
Operainstru
23‐1 Operainstru
00 17
Class
00 17
Type
00 17
Manu
00 17
Date
00 17
RF Co
23‐1 Autho
00 17
Catal
00 17
Seria
ETL L
00 17
Temp
afety informa
contraindicati
manual for th
Meaning
on, caution isevice and/or ator awareneoid undesirab
ator's manuauctions befor
ator’s manuauctions befor
II equipment
BF applied pa
ufacturer
of Manufactu
oil, Transmit a
orized Repres
og Number
l Number
isted (Canada
perature limit
tion that mus
ions, precaut
he MRI system
s necessary wthe situation
ess or operatoble conseque
al, Consult ope operating t
al, Consult elee operating t
t
art
ure
and Receive
sentative in E
a & USA)
t
6000683 R
st be observe
ions, and oth
m.
when operatinn described neor action in ornces
erating he device
ectronic operahe device
EU
Rev. 2
ed
er
ng eeds rder
ating
8 | P a g e
Indicat
Th
T/
of
Contra
N
Precau
e
Symbol
tions
he 1.5T 16ch
/R Hand Wris
f the hand an
aindicatio
one.
utions
Patien
Patien
Patien
young
Patien
Patien
sensiti
failure
Number
2620
2621
W017
N/A
T/R Hand Wr
st Coil is inten
nd or wrist tha
ons
ts with increa
ts who are un
ts with an ina
children)
ts with loss o
ts who have
ive to increas
e, or impaired
Standa
ISO 700IEC 604
ISO 700IEC 604
ISO 2440ISO 8528
EN5041EU2012/1
rist Coil is inte
nded for use w
at can be inte
ased likelihoo
nconscious, h
ability to mai
of feeling in an
difficulty regu
es in body te
d perspiration
rd Title,
00 17
Humi
00 17
Atmo
09‐2 8‐13
Warn
19 8/EU
The ushoulBy encorreconsehealthinappFor mreturnconsuthe pr
ended for use
with GE 3.0T
erpreted by a
od of seizures
heavily sedate
ntain reliable
ny body part
ulating their b
mperature (f
n)
Meaning
idity limitatio
ospheric press
ning; Hot surf
use of this symld not be treansuring that thctly, you will hequences for h, which coulpropriate wastmore detailed n and recyclinult the supplieroduct.
e with GE 1.5T
MR systems,
trained phys
s or claustrop
ed, or in a con
e communicat
body temper
for example, p
on
sure limitatio
face
mbol indicateated as househis product ishelp prevent the environmd otherwise bte handling ofinformation c
ng of this proder from whom
T MR system
to produce d
sician.
phobia
nfused menta
tions (for exa
ature or who
patients with
6000683 R
on
es that this proehold waste. disposed of potential neg
ment and humbe caused by f this product
concerning theduct, please
m you purchas
s and 3.0T 16
diagnostic ima
al state
mple, infants
o are particula
fever, cardia
Rev. 2
oduct
ative man
. e
sed
6ch
ages
s or
arly
ac
9 | P a g e 6000683 Rev. 2
Cautions – RF Coil
Do not place any disconnected devices (RF coils, cables, etc.) in the gantry during
scanning.
Connect only the designated RF coils to the RF coil connection port.
Do not use a defective RF coil, especially if the outer covering has been damaged or if
metal parts are exposed.
Do not attempt to change or modify the coil.
Do not cross or loop coil cables.
Ensure that the patient does not come into
direct contact with the coil cables.
Do not allow the patient to form a loop with
any body parts. Use pads to ensure that the patient’s hands and legs do not touch the
coil, MRI system, patient table, or another body part that may form a loop.
Do not allow the patient or RF coil to touch any part of the MRI system. Use pads to
separate the patient from the bore, if necessary.
Stop the scan immediately if the patient complains of warming, tingling, stinging, or
similar sensations. Contact a physician before continuing with the scan.
Ensure that the coil does not come into contact with liquids, such as water or
medications.
10 | P a g e 6000683 Rev. 2
If a coil is found to be defective, stop using the coil immediately and contact your GE
representative.
Use only the accessories described in this manual with the coil.
Emergency Procedures
In case of an emergency during the scan, stop the scan immediately, remove the patient from
the room, and obtain medical assistance, if necessary.
11 | P a g e 6000683 Rev. 2
Chapter 3 – TR Port Location
TR Port Location
The 16ch T/R Hand Wrist Coil is a Transmit and Receive coil. To properly use the coil, ensure the
system interface connector is connected to P‐port on the system. Consult the system user
manual to locate the port that supports both transmit and receive (P1 on 60 cm or 70 cm curved
or detachable tables and P2 on 70cm fixed table systems).
Chapter 4 – Quality Assurance
Scanner Verification
Perform system level Signal to Noise (SNR) Test. Refer to Service Methods CD; System Level
Procedures; Functional Checks; Signal to Noise (SNR) Test.
Signal to Noise Ratio (SNR) Test
Tools/Fixtures Required
Description GE Part # QED Part # Qty
1.5T Unified Cubical Phantom 5342681 N/A 1
16ch T/R Hand Wrist Coil –Horizontal Baseplate 5561531‐4 2001768 1
16ch T/R Hand Wrist Coil – Anterior Liner/Phantom Position Pad
5561531‐7 3004566 1
Coil and Phantom Setup
1. Record the serial number of the coil(s) being used, as well as software build version (from testrecord or getver).
2. Remove any other surface coils (if present) from the cradle.
12 | P a g e 6000683 Rev. 2
3. Transport the coil to the patient cradle. Be sure to carry the coil with both hands by the handle on the base.
4. Place the coil onto the patient cradle. Note that the bore direction arrow pictured below should be pointing towards the bore.
13 | P a g
5. Toas
g e
o avoid loopss shown belo
s, route any ew.
Do not c
Ensure t
xcess cable u
cross or loop
that patient d
sing the cable
coil cables.
does not com
e routing clip
e into direct c
ps attached to
contact with
6000683 R
o the system c
the coil cable
Rev. 2
cable
es.
14 | P a g
6. Ccmco
*: For ref
7. LaisTud
g e
onnect the com curved or donnector aro
ference only,
andmark the s required, roturn the knob esired positio
oil connector detachable taund such that
applies to bo
coil at the cetate knobs inagain to the on.
to the approables and P2 ot it exhibits th
oth 1.5T and
nter landmarto unlocked plock position
opriate Transmon 70cm fixedhe LOCKED po
3.0T
rk (hand/wrisposition, as shto secure the
*
mit Port of thd table systemosition, see p
st mode) as shhown below, e coil in place
e system (P1 ms). Turn thepicture on righ
hown below. to achieve dee once the co
6000683 R
on 60 cm or e end of the Pht.
If coil adjustmesired alignmil has reached
*
Rev. 2
70 P‐Port
ment ment. d the
15 | P a g e 6000683 Rev. 2
Unlock
Lock
8. Open the coil by sliding the latch forward and pulling up on the anterior.
16 | P a g e 6000683 Rev. 2
9. Place Anterior Liner Pad (3004566) on anterior of coil.
10. Place the Unified Cubical Phantom (5342681) into the coil as shown below. Ensure the bottom
edge of phantom aligns with FOV markings on the coil.
11. Close the coil, ensuring that the anterior latch release clicks into place.
17 | P a g e 6000683 Rev. 2
12. Re‐confirm the landmark the coil at the center‐most landmarks shown below and move coil to isocenter.
18 | P a g e 6000683 Rev. 2
Multi‐Coil Quality Assurance (MCQA) Tool
All RF coil related tests must be run on a system that is well calibrated. EPIWP (White Pixel from
install in spec) shall pass.
Test ID Parameter Description Expected Result
1 EPIWP in spec PASS
To initiate MCQA:
1. From Common Service Desktop (CSD), go to Service Browser and select [Image Quality]
“Multi‐Coil QA Tool” and then “Click here to start this tool” as shown in Figure 1.
Figure 1
Note: If a “No valid MCR‐V (or MCR2/3)” warning (Figure 2) pops up select [Yes] and proceed
with test. MCR‐V diagnostics must be run before turning over system to customer.
Figure 2
19 | P a g e 6000683 Rev. 2
The current coil field will be automatically filled in (Figure 3), based on the CoilID of the coil
connected to the LPCA. Enter the serial number of the coil being tested in the Coil Serial number
field.
2. Click on [Start] to begin the automated test as shown in Figure 3. Depending on the number
of test locations (complexity of the coil) the test may take from 3 to 5 minutes.
Figure 3
3. Upon start‐up, a Note stating, “Phantom placement and coil landmarking are critical for
repeatable results” will appear. If the landmark has been set correctly and there are no air
bubbles in the phantom, click [Yes] to continue. (Figure 4).
Figure 4
004
20 | P a g e 6000683 Rev. 2
Note: The Status window of the MCQA Tool GUI will continuously update to give information on
what the tool is doing at any point in time. A time bar (Figure 5) will appear, showing
approximate total test time, elapsed time and percent complete.
Figure 5
When the test is complete, test results display on the screen (Figure 6). The PASS/FAIL status
shows PASS if all coil elements are functioning properly. The MCQA Tool GUI displays “Fail” for
one of the following possible reasons but not limited to:
• Bad Coil Element
• Incorrect phantom used for the test (Unified Cubical Phantom 5342681 should be used)
• Incorrect positioning/placement of the phantom
More information on the MCQA test can be found on the MR service methods DVD or website
via the path: Troubleshooting ‐> System ‐> Multi‐Coil Quality Assurance Tool
Figure 6
4. Click on [Quit] button to exit MCQA Tool.
21 | P a g
Using M
In
1
N
R
2
g e
MCQA Vi
n case if the re
. In the MCQ
file select
Note: The Resu
esults shown
. Select the
Viewer to
Test ID
1
iewer
esults are to
QA Tool wind
[View Report
ults Viewer w
on the tool G
ISNR option a
view the resu
D
be viewed at
dow select File
t Details] to re
will open as sh
GUI will also b
and the ISNR
ults.
Paramet
EPIWP in
a later stage
e, Open Resu
eview the res
hown in Figure
be listed acros
Figure 7
Specs check
ter Descriptio
n spec
follow the be
lts File and se
sults.
e 7. The Resu
ss the top of t
box in the m
on Expecte
PASS
elow steps:
elect the desi
ults file name
the viewer.
iddle portion
ed Result
6000683 R
ired coil resul
and Pass/Fai
of the Result
Rev. 2
lts
il
ts
22 | P a g e 6000683 Rev. 2
Chapter 5 – Coil Setup and Use
Determine Scan Position
The 16ch T/R Hand Wrist Coil is equipped with two bases, designed to image the patient either
at the patient’s side (vertical base) or over the patient’s head (horizontal base). Determine
optimal scan position based on patient size, comfort, and scan preference.
Horizontal Base
Vertical Base
23 | P a g
To
re
H
V
g e
o switch base
espective bas
orizontal Bas
Vertical Base
es, while hold
es below:
Note: Do not
se
ding the coil, f
t switch bases
firmly push on
s while the pa
n the coil rele
atient is in the
ease lever, sh
e coil.
6000683 R
hown on the
Rev. 2
24 | P a g e 6000683 Rev. 2
Then, install on desired base by aligning and locking coil into base, as shown below.
Horizontal Base
Vertical Base
25 | P a g e 6000683 Rev. 2
Positioning the 16ch T/R Hand Wrist Coil: Horizontal Base
1. Remove any other surface coils (if present) from the patient cradle.
2. Transport the coil to the patient cradle. Be sure to carry the coil with both hands by the handles
on the base.
3. Place the coil onto the patient cradle. Note that the bore direction arrow pictured below should be pointing towards the bore.
4. To avoid loops and patient contact, route any excess cable using the cable routing clips attached
to the system cable as shown below.
26 | P a g
5. ConnemanuLOCK
*
g e
ect the coil coual for TR PortED position, s
*: For referen
Do no
Ensur
onnector to tt Location) Tsee picture on
nce only, appl
ot cross or loo
e that patient
he appropriaurn the end on right.
lies to both 1
op coil cables.
t does not co
te Transmit Pof the P‐Port
.5T and 3.0T
*
.
ome into direc
Port of the sysconnector ar
ct contact wit
stem. (Refer ound such th
6000683 R
th the coil cab
to system ushat it exhibits
*
Rev. 2
bles.
ser the
27 | P a g
Positio
1
Curved
Standar
g e
oning the
. Set the ver
on the fee
change the
Table – 60cm
rd Flat Table
16ch T/R
The 16multipthe Ve
rtical base fee
t indicate wh
e setting, firm
m bore
– 70cm bore
R Hand W
6ch T/R Handple systems. Iertical Base m
et to the posi
at side shoul
mly grasp the
with remova
Wrist Coil:
d Wrist Coil sun order to pro
must be set ac
ition required
d be facing o
feet shown b
able table
: Vertical
upports crossovide optimizccordingly.
d for the syste
ut for the app
below, and ro
Base
‐platform comzed coil and p
em being use
propriate pat
otate to desire
6000683 R
mpatibility acpatient positio
ed. The marki
tient table. To
ed position.
Rev. 2
cross on,
ings
o
28 | P a g
Extende
2
3
4
g e
ed Flat Table
v
. Remove an
. Transport
handles on
. Place the cshould be
‐ 70cm bore
Note: Incorre
vertical base
ny other surfa
the coil to th
n the base.
coil onto the pointing tow
with fixed ta
ect system ba
is set up corr
ace coils (if p
e patient crad
patient cradlewards the bore
ble
ase setup cou
rectly for the
resent) from
dle. Be sure t
e. Note that te.
ld result in po
correspondin
the patient c
to carry the c
the bore dire
oor image qu
ng system.
cradle.
coil with both
ction arrow p
6000683 R
ality. Ensure
h hands by the
pictured below
Rev. 2
the
e
w
29 | P a g
5
*
6
g e
. Connect thuser manuexhibits th
: For referenc
. To avoid lo
attached t
he coil conneual for TR Porthe LOCKED po
ce only, applie
oops and patio the system
ctor to the apt Location) Tosition, see pi
es to both 1.5
ient contact, cable as show
ppropriate TrTurn the end octure on righ
5T and 3.0T
route any excwn below.
ansmit Port oof the P‐Port t.
cess cable usi
*
of the systemconnector ar
ing the cable
6000683 R
. (Refer to syround such th
routing clips
*
Rev. 2
ystem hat it
30 | P a g
Pad Co
1
g e
onfigurati
. Various pa
to provide
Do no
Ensur
ion
ads are suppli
patient comf
ot cross or loo
e that patient
ed with the 1
fort.
7
7
op coil cables.
t does not co
16ch T/R Han
1
2
.
ome into direc
d Wrist Coil t
3
4
ct contact wit
to minimize m
5
6
6000683 R
th the coil cab
motion artifac
Rev. 2
bles.
ct and
31 | P a g e 6000683 Rev. 2
Number Descrption Qty GE Part Number
QED Part Number
1 16ch T/R Hand Wrist Coil – Posterior Liner Pad 1 5561531‐6 3004567
2 16ch T/R Hand Wrist Coil – Anterior Liner/Phantom Position Pad
1 5561531‐7 3004566
3 16ch T/R Hand Wrist Coil – Palm Pad 1 5561531‐15 3004964
4 16ch T/R Hand Wrist Coil – Wedge Pad 1 5561531‐8 3004751
5 16ch T/R Hand Wrist Coil – Elbow/Arm Pad 1 5561531‐9 3004607
6 16ch T/R Hand Wrist Coil – Wrist Coil Filler Pad 1 5561531‐10 3004716
7 16ch T/R Hand Wrist Coil – Side‐mount Base Pad 2 5561531‐11 3004612
Position the Patient: Horizontal Base
1. The 16ch T/R Hand Wrist Coil comes with a variety of pads to facilitate patient comfort.
Below is an example of the recommended layout for the horizontal orientation:
32 | P a g e 6000683 Rev. 2
2. Position the patient’s hand into the coil. Use marks on coil to aid in positioning as shown
below. If necessary, use Wedge and/or Palm Pads to immobilize the patient’s hand/wrist
and to ensure patient comfort.
33 | P a g e 6000683 Rev. 2
Position the Patient: Vertical Base
1. The 16ch T/R Hand Wrist Coil comes with a variety of pads to facilitate patient comfort. Below is
an example of the recommended layout for the vertical orientation:
2. Position the patient’s hand into the coil. Use marks on coil to aid in positioning the patient in
the coil as shown below. If necessary, use Wedge and/or Palm Pads to immobilize the patient’s
hand/wrist and to ensure patient comfort.
34 | P a g e 6000683 Rev. 2
35 | P a g e 6000683 Rev. 2
Lock the Coil
1. Close the coil, making sure not to pinch the patient, gown, or bedding material between the coil
halves. This could cause patient injury, poor image quality, or possibly result in damage to the
coil. Push the anterior half of the coil down until it “clicks” into place.
36 | P a g e 6000683 Rev. 2
Landmark
1. The 16ch T/R Hand Wrist Coil has 3 landmarks as shown below. These correspond to three
different coil modes: Hand Only (8‐channel mode), Hand/Wrist (16‐channel mode), and Wrist
Only (8‐channel mode). Select landmark based on desired target anatomy.
Hand Only Mode Hand/Wrist Mode Wrist Only Mode
Hand Only Mode
Hand/Wrist Mode
Wrist Only Mode
Hand Elements
Wrist Elements
37 | P a g
2. If pto
U
Lo
g e
coil adjustmeosition, as sho secure the c
Unlock
ock
Note: Ensu
setup. The
ent is requireown below, tcoil in place o
ure the horizo
e coil may sh
ed for horizonto achieve deonce the coil h
ontal base is
ift during sca
ntal base confsired alignmehas reached t
locked after a
nning, which
figuration, roent. Turn the the desired po
any adjustme
could result
tate knobs inknob again toosition.
ent during lan
in poor imag
6000683 R
nto unlocked o the lock po
ndmark
e quality.
Rev. 2
sition
38 | P a g e 6000683 Rev. 2
3. Advance the patient into the magnet and landmark the coil using the reference marks on the
top of the 16ch T/R Hand Wrist Coil for the desired imaging mode.
39 | P a g e 6000683 Rev. 2
Chapter 6 – Cleaning, Maintenance, Service, and Disposal
Cleaning the RF Coil
Caution: Do not pour cleaning solution directly onto the coil or accessories.
Caution: Do not sterilize the coil or accessories.
Caution: Do not apply cleaning solution to electrical contacts.
The RF Coil and patient comfort pads must be cleaned after each use using the following
procedure:
1. Disconnect RF coil from the MRI scanner before coil cleaning.
2. Wipe off any dirt on the coil surface using a dry cloth. If dirt is difficult to remove, clean it
according to the procedures described below.
3. Wipe with a cloth that has been dampened in a solution of 10% bleach and 90% tap water,
or 70% ethanol and 30% tap water.
4. Should the coil need to be returned to GE Healthcare for service, wipe it down with a 10%
bleach solution (as described above) to minimize risk of exposure to potentially infectious
agents.
5. Dispose of any materials used to clean the coil and the pads according to all federal, state,
and local regulations.
Disinfection
If disinfection of the RF coil or patient comfort pads is necessary, clean as described above then
use the following procedure:
Pre‐Disinfection Steps:
1. Wet all surfaces with CaviCide (using spray applicator or using towelettes for certain
surfaces such as those close to electrical contacts; do not apply cleaning solution to
electrical contacts). Ensure all surfaces are visibly wet and remain wetted for a minimum of
30 seconds.
2. Use a soft nylon bristle brush and/or additional cleaner/disinfectant towelettes to loosen
hardened or difficult to remove debris or bioburden. Apply additional cleaner/disinfectant
(using spray applicator or using towelettes for certain surfaces such as those close to
electrical contacts) to areas subjected to any previous brushing or wiping. Ensure these
previously brushed or wiped areas remain visibly wetted with cleaner/disinfectant for a
minimum of 30 seconds.
3. Wipe surfaces with clean paper towels to remove debris.
40 | P a g e 6000683 Rev. 2
4. Discard used brushes, used cleaner/disinfectant towelettes and used paper towels.
5. Repeat steps 1 through 4.
6. If debris remains on the surfaces, repeat pre‐disinfection steps.
Disinfection Steps:
1. Apply CaviCide (using spray applicator or using towelettes for certain surfaces such as those
close to electrical contacts) directly to pre‐cleaned surfaces and ensure all surfaces are wet
and remain wetted for a minimum of two (2) minutes. Do not apply cleaning solution to
electrical contacts.
2. Wipe with clean paper towels to remove residual cleaner/disinfectant.
3. Discard used cleaner/disinfectant towelettes and used paper towels.
Allow coil and accessories to dry before use.
Maintenance
No regularly scheduled maintenance is required for the RF coil.
Service
Please contact your GE representative with questions regarding service of the RF coil.
Disposal
Please contact your GE representative with questions regarding the return or disposal of the RF
coil.
41 | P a g e 6000683 Rev. 2
THIS PAGE LEFT INTENTIONALLY BLANK
42 | P a g e 6000683 Rev. 2
Manufacturer:
Quality Electrodynamics, LLC. 6655 Beta Drive, Suite 100 Mayfield Village, OH 44143 U.S.A.
www.qualityelectrodynamics.com
Distributor:
GE Medical Systems, LLC Turkey Importer Details:
GE Medical Systems Turkey Ltd. Sti. Esentepe Mah. Harman Sok. No: 8 34394 Sisli – Istanbul Turkey