+ All Categories
Home > Documents > MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL...

MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL...

Date post: 20-Apr-2020
Category:
Upload: others
View: 8 times
Download: 0 times
Share this document with a friend
155
MATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY By P. C. Weinrauch M.B., B.S. (Qld) A Thesis for the Degree of Master of Engineering School of Engineering Systems Queensland University of Technology August, 2006
Transcript
Page 1: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

MATERIAL PROPERTIES OF BILAMINAR

POLYMETHYLMETHACRYLATE CEMENT

MANTLES IN REVISION HIP

ARTHROPLASTY

By

P. C. Weinrauch M.B., B.S. (Qld)

A Thesis for the Degree of Master of Engineering

School of Engineering Systems

Queensland University of Technology

August, 2006

Page 2: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

ii

Keywords Hip, Arthroplasty, Revision, Cement, Polymethylmethacrylate, Adhesion.

Page 3: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

iii

Abstract Cement – within – Cement (C-C) revision techniques have been demonstrated to

reduce the complications associated with removal of secure cement from the femoral

canal during revision hip joint arthroplasty. Material failure at the interface between new

and old cement mantles represents a theoretical limitation of this technique. The

objectives of this thesis are to describe the variability in material properties of uniform

and bilaminar polymethylmethacrylate (PMMA) cement mantles in shear with respect to

duration of post-cure and the influence of commercial inclusion of antibiotics on

bilaminar cement mantle interfacial shear strength.

Uniform mantles of Surgical Simplex P and Antibiotic Simplex PMMA cements

demonstrated variability in ultimate shear stress to failure with respect to duration of

post-cure (p<0.001), however the variations were quantitatively small and unlikely to be

of clinical relevance. Bilaminar cement mantles were 15 – 20 percent weaker than

uniform mantles (p<0.001) and demonstrated similar time dependant material property

variations in shear (p<0.001). Bilaminar PMMA test specimens manufactured using

Antibiotic Simplex cement demonstrated equivalent ultimate shear stress to failure as

bilaminar specimens manufactured from Surgical Simplex (p=0.52). High C-C

interfacial strengths are demonstrated as early as one hour after cement application.

Interfacial adhesion by mechanisms other than mechanical interlock significantly

influence the bond formed between layered PMMA cements, with an important

contribution by diffusion based molecular interdigitation.

In the presence of a secure cement-bone interface, C-C femoral revision can be

recommended as a viable technique on the basis of the strong interfacial bond formed

between new and old cement mantles. The use of Antibiotic Simplex in C-C revision is

recommended as detrimental effects on the interfacial shear properties have not been

demonstrated with the commercial addition of Tobramycin.

Page 4: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

iv

Table of Contents Keywords ........................................................................................................................... ii

Abstract............................................................................................................................. iii

Table of Contents ............................................................................................................. iv

List of Figures.................................................................................................................. vii

List of Tables .................................................................................................................. xiv

List of Equations ............................................................................................................ xvi

List of Symbols .............................................................................................................. xvii

List of Abbreviations ...................................................................................................xviii

Glossary of Terms ........................................................................................................... xx

Statement of Originality ..............................................................................................xxiii

Acknowledgements....................................................................................................... xxiv

Dedication ...................................................................................................................... xxv

Chapter 1 Introduction .................................................................................................... 1

1.1 Overview of Research and Clinical Application............................................... 1

1.2 Research Objectives and Hypotheses................................................................ 3

1.3 Thesis Overview................................................................................................ 4

Chapter 2 Literature Review........................................................................................... 7

2.1 PMMA Cement in Primary and Revision Hip Joint Arthroplasty .................... 7

2.1.1 PMMA cement in primary total hip arthroplasty ........................................ 7

2.1.2 PMMA cement in hip joint hemiarthroplasty............................................ 10

2.1.3 Revision hip arthroplasty and cement removal ......................................... 16

2.2 Cement – Within - Cement Femoral Revision................................................ 23

2.2.1 Clinical application.................................................................................... 23

2.2.2 Radiographic assessment of the bone – cement interface ......................... 28

2.2.3 Prosthetic femoral components designed for C-C revision....................... 38

2.3 Composition and Material Properties of PMMA Cement............................... 41

2.4 Properties of PMMA Cement and Bilaminar Mantles in Shear...................... 50

2.5 Summary of Literature Review and Relevance to Thesis............................... 63

Chapter 3 Experimental Evaluation of the Shear Properties of PMMA cements ... 67

Page 5: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

v

3.1 Introduction ..................................................................................................... 67

3.2 Hypotheses ...................................................................................................... 67

3.3 Method ............................................................................................................ 67

3.3.1 Equipment design considerations and objectives ...................................... 67

3.3.2 Overview of experimental equipment and outline of applications............ 68

3.3.3 Test specimen manufacture ....................................................................... 69

3.3.4 Evaluation of PMMA shear properties...................................................... 73

3.3.5 Testing schedule and data analysis............................................................ 74

3.3.6 Testing method validation and shear modulus calibration ........................ 77

3.4 Results ............................................................................................................. 78

3.4.1 Sawbones reference and shear modulus calibration.................................. 78

3.4.2 Uniform PMMA test specimen shear properties ....................................... 79

3.5 Discussion ....................................................................................................... 81

3.6 Conclusions ..................................................................................................... 83

Chapter 4 Experimental Evaluation of the Shear Properties of Bilaminar Cement

Mantles ................................................................................................................. 85

4.1 Introduction ..................................................................................................... 85

4.2 Hypotheses ...................................................................................................... 85

4.3 Method ............................................................................................................ 86

4.3.1 Bilaminar test specimen manufacture and testing schedule ...................... 86

4.3.2 Bilaminar test specimen evaluation of shear properties ............................ 89

4.4 Results ............................................................................................................. 89

4.4.1 Surface characteristics & fracture evaluation by LM and SEM................ 89

4.4.2 Bilaminar PMMA test specimen shear properties..................................... 94

4.5 Discussion ....................................................................................................... 97

4.6 Conclusions ................................................................................................... 101

Chapter 5 Conclusions and Recommendations ......................................................... 105

5.1 Application of Research Findings to Clinical Practice ................................. 105

5.2 Further Research ........................................................................................... 105

Appendix 1 ..................................................................................................................... 109

Appendix 2 ..................................................................................................................... 117

Page 6: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

vi

Appendix 3 ..................................................................................................................... 119

References ...................................................................................................................... 131

Page 7: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

vii

List of Figures

Figure 2-1 Cumulative risk for revision of THA by cement status excluding infection

(Australian Orthopaedic Association National Joint Replacement Registry 2004)...9

Figure 2-2 Age and sex distribution for hip joint hemiarthroplasty (Australian

Orthopaedic Association National Joint Replacement Registry 2004)....................11

Figure 2-3 Cumulative risk for revision of UAM and CT prostheses (Australian

Orthopaedic Association National Joint Replacement Registry 2004)....................13

Figure 2-4 Increasing incidence of acetabular disease over time with patients managed

with hip hemiarthroplasty (Yamagata et al. 1987)...................................................15

Figure 2-5 Diagnosis requiring revision hip replacement (Canadian Institute for Health

Information 2004). ...................................................................................................17

Figure 2-6 Extraction of cemented femoral component in line of implant long axis

(Hozack 1998)..........................................................................................................19

Figure 2-7 Surgical technique of C-C femoral revision (Stryker International)..............24

Figure 2-8 Coronal plane Gruen Zones for the standardised assessment of the femoral

component bone – cement interface (Gruen et al. 1979). ........................................30

Figure 2-9 Gruen modes of radiographic failure of cemented femoral THA components

(for explanation see text) (Harkess 1999). ...............................................................31

Figure 2-10 Radiograph demonstrating normal subsidence of an Exeter stem, with

cement-prosthesis debonding visible in Gruen Zone 1 (courtesy of R. Crawford). 32

Figure 2-11 Normal subsidence of a collarless cemented polished double taper femoral

component (Yates et al. 2002). ................................................................................32

Figure 2-12 Gruen Mode IB failure (compound subsidence). Radiolucency in Zone 1

representing distal migration of the cement – prosthesis composite, with distal

radiolucency and pedestal formation (Harkess 1999)..............................................33

Figure 2-13 Gruen Mode II failure (medial stem pivot). Note varus migration of the stem

with separation of the cement – prosthesis interface in zones 1 and 2 (Harkess

1999). .......................................................................................................................34

Page 8: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

viii

Figure 2-14 Gruen Mode III failure (calcar pivot). Initial postoperative radiograph

depicted on the left for comparison. Progressive varus migration of the distal stem

demonstrated and osteolysis of the greater trochanter however minimal migration

of the proximal prosthesis consistent with pivoting about a proximal (calcar)

location (Harkess 1999). ..........................................................................................34

Figure 2-15 Gruen Mode IV failure (bending cantilever). Serial radiographs from time of

implantation from left to right. Left: Varus malposition of the stem at initial surgery.

Centre: Proximal cement – prosthesis lucency in zone 1 without abnormality in

distal cement fixation. Right: Fragmentation of proximal cement mantle in Gruen

Zone 7 with fatigue failure of prosthesis (Harkess 1999). .......................................35

Figure 2-16 Digital subtraction arthrogram demonstrating loosening of the femoral

implant with extensive extravasation of radio-opaque contrast about the cement –

bone interface and distal contrast leak to cortical perforation (Ovesen 2003).........37

Figure 2-17 Standard length Exeter stem with winged centraliser (left) and Exeter short

revision stem (125mm, +44 offset) with revision wingless centraliser (right)

(Stryker International 2005).....................................................................................40

Figure 2-18 Exeter Trauma Stem (ETS) monoblock hemiarthroplasty. Polished double

taper geometry without a prosthetic collar...............................................................40

Figure 2-19 Presentation of Antibiotic (Tobramycin) Simplex PMMA cement. ............42

Figure 2-20 Differences in molecular weight of PMMA chains due to alternate

sterilisation methods. Variation in the constituents of different cement preparations

produces differences in the material properties of the polymerised products (Kuehn

et al. 2005)................................................................................................................43

Figure 2-21 Residual monomer in Surgical Simplex PMMA Cement after initial

polymerisation (Kuhn 2000). ...................................................................................44

Figure 2-22 Increase in shear strength of Surgical Simplex P cement related to duration

of post-cure (Wilde et al. 1975). ..............................................................................45

Figure 2-23 Working curves for Simplex P Cement (Kuhn 2000). .................................47

Figure 2-24 Fatigue life of PMMA cements to cyclic compression are improved with

porosity reduction by vacuum preparation (Kuehn et al. 2005). .............................48

Page 9: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

ix

Figure 2-25 Schematic diagram of the PMMA shear testing apparatus used by

Greenwald et al. (1978). Cast PMMA cylinders with a known cross-sectional area

are subjected to shear between two plates similar to a guillotine. ...........................52

Figure 2-26 ASTM D732 - 02 Shear punch tool .............................................................53

Figure 2-27 Variables influencing the calculation of material properties by shear punch

tool (Guduru et al. 2005)..........................................................................................54

Figure 2-28 Shear strength of four orthopaedic PMMA cements with variable

concentrations of added gentamicin (Moran et al. 1979).........................................55

Figure 2-29 PMMA cement shear properties determination conducted by Black et al.

(1982). Interfacial shear strength was not assessed with this experiment................57

Figure 2-30 Porosity observed in the experimental PMMA cement mantle applied onto

concurrently polymerising cement. Heat is transferred to the experimental cement

mass during early polymerisation and porosity is induced by the trapping of

vaporised monomer (Black et al. 1982). ..................................................................58

Figure 2-31 Method of interfacial shear stress evaluation (Greenwald et al. 1978). .......58

Figure 2-32 ASTM C 273 – 00 Standard test method for shear properties of sandwich

core materials. ..........................................................................................................59

Figure 2-33 Interfacial ultimate shear strength shows marked weakening with blood

present at the interface prior to application of the second cement mantle, however

approaches the strength of a uniform mantle with suitable preparation (Greenwald

et al. 1978)................................................................................................................60

Figure 2-34 Interfacial ultimate shear strength between PMMA mantles decreases when

the second mantle is applied late. Late application is associated with a decreased

mixture weight due to the evaporation of monomer from the cement surface prior to

polymerisation (Greenwald et al. 1978)...................................................................62

Figure 2-35 Interfacial ultimate shear strength between PMMA cement mantles

decreases in proportion to monomer loss from the second mix of cement applied

(Greenwald et al. 1978)............................................................................................62

Figure 3-1 SPT demonstrating casting and testing modes of application. The central

casting rod is removed for the manufacture of solid PMMA cast cylinders............68

Page 10: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

x

Figure 3-2 Parting of PMMA discs from solid cylinder. With the use of suitable lathe tip

allowing adequate swarf clearance, temperature of the cutting tool typically

measured at 35.0 – 37.0oC, well below the glass transition temperature of

polymerised PMMA (see text). ................................................................................71

Figure 3-3 Facing of the test specimens was performed on both sides. With the use of

suitable lathe tip allowing adequate swarf clearance, maximum temperature of the

cutting tool typically measured 25.0-26.5oC, well below the glass transition

temperature of polymerised PMMA (see text).........................................................71

Figure 3-4 Confirmatory measurement of specimen thickness using digital verniers

calliper with precision of 10μm. ..............................................................................72

Figure 3-5 Dry post-curing of PMMA specimens at 37oC. .............................................72

Figure 3-6 Shear punch testing using Hounsfield unit.....................................................73

Figure 3-7 Sectioned isometric view of SPT device in testing mode. 2mm gap between

top plate and base plate are demonstrated................................................................74

Figure 3-8 Illustration of R1 and R2 measurements used in Equation 3.3, where R1 is the

radius of the SPT die and R2 is the punch radius. ....................................................76

Figure 3-9 Ultimate shear stress to failure of Antibiotic Simplex and Surgical Simplex P

cements with 95% confidence intervals marked. Similar variations in shear

properties related to duration of post-cure are observed for both cement

preparations. .............................................................................................................79

Figure 3-10 Calculated shear modulus of Antibiotic Simplex and Surgical Simplex P

cements with 95% confidence intervals marked......................................................80

Figure 3-11 Above: Schematic representation of SPT device in testing mode. R1 =

radius of die. R2 = radius of punch. Below: Magnification of the zone of shearing.

The portion of material undergoing shear is represented in light blue. Die-Punch

clearance was estimated to be 20μm by use of micrometers, however inaccuracy in

this clearance measurement would potentially create large errors in the assessment

of shear modulus by direct calculation. Indirect methods for shear modulus

determination were therefore employed...................................................................82

Figure 4-1 Simplex PMMA cement cylinder with central hole after removal of casting

rod. After post-curing of the PMMA cylinder, fresh PMMA cement was poured

Page 11: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xi

into the central hole, therefore creating a solid cement cylinder with outer “old”

cement and inner “new” cement. .............................................................................87

Figure 4-2 Completed 5mm thickness bilaminar PMMA test specimen. For illustration

purposes, coloured cement is used to differentiate the outer (old) cement mantle

from the inner (new) cement mantle. .......................................................................88

Figure 4-3 SEM of the interfacial surface of the old cement mantle prior to the

application of new cement. Smooth surface with minimal porosity is demonstrated.

Surface roughness of this interface measured by profilometry was 0.22μm Ra.

(Original magnification x27). ..................................................................................90

Figure 4-4 SEM of the interfacial surface of the old cement mantle prior to the

application of new cement. At high power magnification the surface profile of the

cement secondary to defects on the casting rod and irregularities created during the

process of rod extraction are appreciated on careful inspection (Original

magnification x1500). ..............................................................................................90

Figure 4-5 Interfacial region between inner PMMA and outer PMMA cement mantles

after SPT testing. Clean separation between the cement mantles is observed

(Original magnification x10). ..................................................................................91

Figure 4-6 Interfacial region of Surgical Simplex P after punch testing. Brittle fracture at

the interface is demonstrated (Original magnification x80). ...................................91

Figure 4-7 Transverse section SEM of a bilaminar Simplex – Simplex test specimen at

the interfacial region after SPT evaluation. Linear separation between new and old

cement mantles is demonstrated. Outer (old) cement mantle is located upper left

and inner (new) cement mantle is located lower right. (Original magnification

x110). .......................................................................................................................92

Figure 4-8 Florescent dye penetrant examination of transverse section of bilaminar

PMMA test specimen after SPT testing demonstrating linear cement through the

interfacial region. .....................................................................................................93

Figure 4-9 Florescent dye penetrant examination of bilaminar test specimen after SPT

testing demonstrating cement fracture through the interfacial region. ....................93

Figure 4-10 Ultimate shear strength of Surgical Simplex P compared to bilaminar test

specimens manufactured from the same material with 95% confidence intervals

Page 12: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xii

marked. Outer mantle of the bilaminar test specimens was post-cured for 14 days

prior to application of the new mantle and tested 0 – 90 days after new mantle

application. Bilaminar PMMA mantles demonstrate time dependant shear

properties similar to that of uniform mantles, however a comparative reduction in

ultimate strength by 15 to 20 percent was observed. ...............................................94

Figure 4-11 Calculated shear modulus for uniform mantle and bilaminar Surgical

Simplex P PMMA samples. 95% confidence intervals marked. .............................95

Figure 4-12 Interfacial shear strength bilaminar test specimens comparing cement

preparations with and without the commercial inclusion of Tobramicin antibiotic,

95% confidence intervals marked. ...........................................................................96

Figure 4-13 Interfacial shear modulus of bilaminar test specimens comparing cement

preparations with and without the commercial inclusion of Tobramicin antibiotic.

95% confidence intervals marked. ...........................................................................96

Figure 4-14 Contributions to adhesion between polymer solids: (a) mechanical

interlocking; (b) interdiffusion of chains; (c) electrical interactions; (d) chemical

interactions (Garbassi et al. 1998)............................................................................97

Figure 4-15 Hypothetical components of the adhesion force between PMMA bone

cement and metal surfaces (Bundy et al. 1987). ......................................................99

Figure A3.1 Exeter PMMA Cement Reamer……………………………. ……………………110

Figure A3.2 Reamer Tip. Note the difference between negative rake angle established at

the centre cutting area of the drill compared to positive helix angle of the drill

body………………………………………………………………………………..…………….110

Figure A3.3 Swarf repeatedly accumulates in front of leading edge of drill tip when

performing end cutting procedures. Note the flutes running down the body of the

bit are clear of debris due to favourable helix angle………………………..………….112

Figure A3.4 Macroscopic appearance of mantle reamed to depth to allow revision with

standard length stem. Reaming was conducted without use of irrigation or regular

clearing of swarf. Note the delaminated appearance of reamed surface…………..113

Figure A3.5 Fluorescent Dye Penetrant study on mantle reamed with irrigation and

regular swarf clearance. Well finished surface with increased roughness compared

Page 13: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xiii

to unreamed. Fine circumferential surface marks from the drilling tip are held by

the material. No mantle cracks identified………………………………………………..114

Figure A3.6 Fluorescent dye penetrant study on same mantle reamed without adequate

irrigation or swarf clearance. Very rough surface with appearance of cement

delamination. No mantle cracks identified………………………………………...…….114

Figure A3.7 SEM (Magnification x27) PMMA Cement mantle unreamed……….……..115

Figure A3.8 SEM (Magnification x27) PMMA Cement mantle reamed according to

manufacturer recommendations with use of irrigation and regular swarf clearance.

Roughened surface but well finished……………………………………………………..115

Figure A3.9 SEM (Magnification x27) PMMA Cement mantle reamed without irrigation

or adequate swarf clearance. Very rough surface produced by smearing of melted

PMMA produced during drilling procedure……………………………...……………..116

Figure A3.10 SEM (Magnification x1500) Unreamed PMMA cement mantle……..….116

Figure A3.11 SEM (Magnification x 1500) Reamed mantle using irrigation and regular

swarf clearance. Roughened surface with mantle integrity maintained………..….117

Figure A3.12 SEM (Magnification x1500) PMMA Cement mantle reamed without

adequate irrigation or swarf clearance. Smearing of melted PMMA can be

appreciated……………………………………………………………………………...…….. 117

Figure A3.13 SEM (Magnification x400) of same mantle displayed in Figure A3.12..118

Page 14: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xiv

List of Tables

Table 2-1 Swedish National Hip Arthroplasty Register percentage survivorship for

cemented and uncemented THA implants after 12 years with failure defined as

revision for any cause. ^ = indicates 10 year follow up data. * = insufficient data.

Statistical analysis by two tailed χ2 test (The Swedish National Hip Arthroplasty

Register 2004). ...........................................................................................................9

Table 2-2 Diagnosis requiring revision hip replacement for primary THR performed

1979 - 2003 (The Swedish National Hip Arthroplasty Register 2004)....................17

Table 2-3 Diagnosis requiring revision hip replacement (Australian Orthopaedic

Association National Joint Replacement Registry 2004).........................................18

Table 2-4 Components used in major THA revision (Australian Orthopaedic Association

National Joint Replacement Registry 2004). ...........................................................21

Table 2-5 Indications and requirements for C-C Revision Hip Arthroplasty. Refer to

Figure 2-9 and text for explanation of Gruen modes of cemented femoral

component failure.....................................................................................................26

Table 2-6 Advantages of C-C femoral revision. ..............................................................27

Table 2-7 Contraindications and disadvantages of C-C femoral revision. ......................28

Table 2-8 Radiographic features of femoral component loosening in cemented THA

(Harkess 1999). See Figure 2-8 and text for explanation of Gruen zones. ..............30

Table 2-9 Composition of four commonly used orthopaedic PMMA cements. MA =

Methacrylic acid; PS = Polystyrene; MMA = Methylmethacrylate (monomer).

Values expressed in wt% (Wright et al. 2001).........................................................43

Table 2-10 Factors influencing the extent and distribution of porosity within PMMA

cements.....................................................................................................................48

Table 2-11 Properties of a typical surgical PMMA cement preparation (Greenwald et al.

1978; Harper et al. 2000; Kuhn 2000; Kuhn et al. 2001; Kuehn et al. 2005). .........51

Table 2-12 Summary of literature: PMMA ultimate shear strength. Refer to text for

details. ......................................................................................................................56

Table 2-13 Summary of literature: PMMA C-C interfacial ultimate shear strength .......61

Page 15: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xv

Table 3-1 Testing schedule for determination of PMMA cement shear properties.........75

Table 3-2 Results of Sawbones reference sample testing. Expected shear strength and

modulus values obtained from manufacturer. * Values +/- 10%. (James 2005;

Pacific Research Laboratories 2005). ......................................................................78

Table 3-3 Tabulated data from Figures 3-9 and 3-10. .....................................................80

Table 4-1 Testing schedule of bilaminar test specimens for the determination of PMMA

interfacial shear properties. ......................................................................................88

Table 4-2 Tabulated data for C-C interfacial shear properties displayed in Figures 4-10

to 4-13. .....................................................................................................................97

Page 16: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xvi

List of Equations

Equation 3-1 Calculation of ultimate shear strength of the test specimens. A is the

surface area of the test specimens subjected to shearing force within the SPT device.

..................................................................................................................................74

Equation 3-2 Calculation of test specimen shear area. T is the thickness of the test

specimen (5.00mm) and Rave is the average of radius of the punch and die. ...........74

Equation 3-3 Average radius of punch and die. Measurements R1 and R2 are depicted in

Figure 3-8. ................................................................................................................74

Equation 3-4 Calculation of the shear modulus calibration ratio, x. Gradient of the force/

displacement data obtained by SPT is represented by ΔF/ΔD…………………………75

Equation 3-5 Calculation of shear modulus of test specimens using calibration ratio.....75

Page 17: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xvii

List of Symbols

A Area of test specimen subjected to shear stress..

D Displacement..

F Force. .

ΔF/ΔD Force/ Displacement gradient. .

G Shear modulus...

Ra Arithmetic mean surface roughness...

R1 Radius of shear punch apparatus die..

R2 Radius of shear punch apparatus punch..

Rave Average radius shear punch apparatus punch and die..

T Test specimen thickness..

τult Ultimate shear stress..

x Shear modulus calibration ratio..

Page 18: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xviii

List of Abbreviations

AB Antibiotic (Tobramycin)

ANOVA Analysis of variance

ASTM American Society of Testing Methods

BPO Dibenzoyl peroxide

C-C Cement-within-cement (revision technique)

CDH Congenital Dysplasia of Hip

CI Confidence Interval

CT Cemented Thompson (monoblock hemiarthroplasty)

DMpT Dimethyl-para-toluidine

ESRS Exeter Short Revision Stem (125mm; +44 Offset)

ETS Exeter Trauma Stem (monoblock hemiarthroplasty)

FEA Finite Element Analysis

Ga-67 Gallium 67 citrate radio – isotope

g/cc Grams per cubic centimetre

HR Hazard Ratio

In-111 DTPA Indium -111 diethylenetriaminepentacetic acid radio –isotope

LM Light Microscopy

LSD Least Significant Difference

LV Low Viscosity

MA Methacrylic Acid

Mag Magnification

MMA Methylmethacrylate (monomer)

MPa Mega Pascals

PCF Pounds per Cubic Feet (measurement of density)

PMMA Polymethylmethacylate

PPM Parts Per Million

PS Polystyrene

PSI Pounds per Square Inch

Page 19: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xix

RR Risk Ratio

SEM Scanning Electron Microscopy

SPT Shear Punch Testing

Tc-99 Technetium 99 phosphate radio-isotope

THA Total Hip Arthroplasty

TKA Total Knee Arthroplasty

RPM Revolutions Per Minute

UAM Uncemented Austin Moore (monoblock hemiarthroplasty)

W/W Weight/ Weight proportion

Page 20: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xx

Glossary of Terms

Axial Perpendicular to the long axis.

Bilaminar Comprising two mantles; applied to C-C revision THA relates to a

cement mantle with layered new and old cement components.

Cancellous Bone with lattice like structure found within the central part of the

proximal femur and other anatomical regions. Surrounded by

cortical bone.

Centraliser Cap placed on the distal end of a femoral prosthesis which assists

in positioning the tip of the prosthesis within the middle of the

bone and allows the stem to subside.

Coronal Anatomical plane situated in the direction of the coronal suture

and at right angles to both the sagittal and axial planes.

Cortex Outer layer of dense hard bone.

Distal Remote; farther away from any point of reference; opposed to

proximal.

Glass transition (Tg) Temperature at which the amorphous phase of a polymer is

converted between glass and rubbery states.

Hazard Ratio (HR) The hazard ratio in survival analysis is a summary of the

difference between two survival curves, representing the

reduction in the risk of death on treatment compared to control,

over the period of follow-up. It is a form of relative risk.

Helix angle Angle made by the advance of the thread as it wraps around a

cylinder; Applied to drill bits and machine tools relates to the

angle between the long axis of the tool and spiral of the flutes.

Hemiarthroplasty Prosthetic hip replacement where the acetabulum is not resurfaced.

The metal femoral component articulates with cartilage.

Interfacial Boundary between two solid masses; Applied to C-C revision

relates to the boundary between new and old cement mantles.

In Vitro Outside the body; within an artificial environment.

Page 21: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xxi

In Vivo Within the body.

Liner Articulating insert for an uncemented acetabular prosthesis.

Housed within a metal case referred to as a shell.

Loosening Mechanical instability of prosthetic components.

Modular Composed by separate pieces which can be assembled or

disassembled; applied to hip arthroplasty relates to femoral

prostheses with separate stem and head components or an

uncemented acetabular prosthesis with separate shell and liner

components. Opposite of monoblock.

Monoblock Prosthesis unable to be assembled or disassembled. Opposite of

modular.

Necrosis Sum of the morphological changes indicative of cell death.

Point angle Angle between cutting edges at the tip of a drill bit or machine

tool. Standard point angle for an all-purpose drill bit is 118

degrees.

Proximal Nearest; closer to any point of reference; opposed to distal.

Rake angle Angle between the leading edge of a cutting tool and a

perpendicular to the surface being cut.

Ream Dilation of a canal by means of a power or hand instrument.

Risk Ratio (RR) Probability of the disease in the risk group divided by the

probability of the disease in the non-risk group.

Shell Outer metal portion of an uncemented THA acetabular component

which interfaces with bone on the outer aspect and accepts an

articulating insert known as a liner.

Stem Femoral prosthesis.

Swarf Particulate debris or shavings created during procedures such as

drilling, reaming or turning.

Tobramycin Fractionated and purified aminoglycoside antibiotic which has

bactericidal activity against many gram-negative and some gram-

positive organisms. Heat stable antibiotic constituent of some

Antibiotic Simplex PMMA cement preparations.

Page 22: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xxii

Uniform mantle PMMA cement mantle manufactured from a single cement type in

a single preparation. As opposed to bilaminar.

Unipolar Single articulation; Applied to hip joint hemiarthroplasty relates

to a prosthesis which articulates between the prosthetic head and

acetabulum only. Compared to a bipolar prosthesis which has a

second articulation built into the prosthetic head.

Van Der Waals A weak attractive force between atoms or non-polar molecules

caused by a temporary change in dipole moment arising from a

brief shift of orbital electrons to one side of one atom or molecule,

creating a similar shift in adjacent atoms or molecules.

Varus Deformity in which angulation of the limb or part is toward the

midline of the body.

Page 23: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xxiii

Statement of Originality

The work contained in this thesis has not been previously submitted for a degree

or diploma at any other higher institution. This thesis contains no material previously

published or written by another person except where due reference is made.

Signature: _______________________________________

Patrick Weinrauch

Date: _______________________________________

Page 24: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xxiv

Acknowledgements

I owe gratitude to many for their assistance in preparing this manuscript. My

family Megan, Lachlan and Harrison have again supported me through yet another

endeavour, and provided me with perspective to my efforts. My supervisor Professor

Ross Crawford has remained an exceptional role model who has demonstrated trust in

my ability and enthusiasm for my research project. The “engine room” vis. Cameron

Bell and Lance Wilson – I have thoroughly enjoyed working with, both from

professional and personal perspectives. Between us, the gap between clinical

orthopaedics and engineering is small. My old friend Dr O has provided practical advice

and comic entertainment during the year which has proven invaluable. Cameron Lutton

and Ben Goss have consistently provided me with answers regarding polymer chemistry

and the properties of bone cement and I am indebted to them for their patience and

tuition. Jon James I have to thank for assisting in the design and then expeditiously

manufacturing the testing apparatus used for this research. Sarah Whitehouse has

assisted with statistical analysis.

I am also indebted to the Queensland Training Committee of the Australian

Orthopaedic Association for supporting my research endeavours. I wish also to thank the

Orthopaedic Research Foundation of the Australian Orthopaedic Association for

financial support via the Foundation Fellowship. I acknowledge support from Stryker

Australia by the provision of cements used in this research.

Page 25: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

xxv

Dedication

To my great mate Paul McCarthy.

Wish you were here.

Page 26: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

1

Chapter 1

Introduction

1.1 Overview of Research and Clinical Application

Aseptic loosening of prosthetic components is the leading cause for revision of

Total Hip Arthroplasty (THA) (Australian Orthopaedic Association National Joint

Replacement Registry 2004; Canadian Institute for Health Information 2004; The

Swedish National Hip Arthroplasty Register 2004). Revision of a cemented hip

arthroplasty traditionally requires the removal of failed prosthetic components and

extraction of both loose and secure cement. While extraction of cemented femoral stems

may be conducted by selective cement removal and the application of a retrograde force,

removal of retained cement from the femoral canal during revision THA remains a time-

consuming and hazardous procedure (Halawa et al. 1978; Choy et al. 1987; Turner et al.

1987; Larson et al. 1988; Gray 1992; Rosenstein et al. 1992; Gardiner et al. 1993;

Hozack 1998; Harkess 1999) . Proximally the cement may be easily accessible, however

the surrounding cortex may be thin and fragile thereby susceptible to damage during

cement extraction. Access to the cement mantle distal to the metaphyseal flare becomes

increasingly difficult, sometimes necessitating exposures such as an extended

trochanteric osteotomy or distal cortical windowing (Hozack 1998; Harkess 1999). Even

within major arthroplasty centres, cement removal results in intraoperative femoral

fracture in approximately 8% of revisions (Gardiner et al. 1993). Other risks associated

with cement removal include cortical perforation, loss of bone stock, excessive blood

loss and increased duration of anaesthesia (Gray 1992; Hozack 1998).

Multiple techniques and instrumentation systems have been developed to assist

in the removal of retained cement from the femoral canal, including the use of hand and

power tools, ultrasonic devices and lasers (Choy et al. 1987; Gray 1992; Schurman et al.

1992; Brooks et al. 1993; Gardiner et al. 1993; Cautilli et al. 1994; Hozack 1998).

Femoral cortical perforations occur in approximately 10% of cases using high speed

cement removal instruments (Turner et al. 1987). Such perforations often occur in an

uncontrolled manner producing large gaps in the cortex, rendering the femur at risk of

Page 27: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

2

postoperative fracture (Larson et al. 1988). Cortical perforations commonly occur

distally and posteriorly within the femur, representing a difficult region to reconstruct

(Gray 1992; Harkess 1999).

In addition to the intraoperative risks associated with the removal of secure

cement, extraction of a well fixed mantle is associated with the unavoidable

simultaneous removal of cancellous bone, thereby compromising the bone quality

required for stable implantation of the revision THA stem. The ultimate shear stress to

failure at the interfacial region between bone and cement is determined by the strength

of the cancellous bone, and damage sustained during cement removal results in a 30

percent reduction in the ultimate shear strength of bone-cement interface in revision

femoral arthroplasty (Halawa et al. 1978; Rosenstein et al. 1992). Traditional cemented

revision by unwarranted extraction of a stable cement mantle may therefore compromise

the performance of the subsequent revision implant.

C-C femoral revision refers to the technique of extracting the femoral prosthesis

with retention of stable cement and subsequent re-cementing of the revision stem into

the old mantle. Retention of the primary cement may be undertaken with no mantle

modification, or alternatively portions of the cement mantle may be removed as required

either proximally or distally (Hozack 1998; Harkess 1999; Hubble et al. 2005).

Excellent early to mid term clinical results have been described using this technique for

appropriate indications (Lieberman et al. 1993; Barrack et al. 1995; McCallum et al.

1995; Hubble et al. 2005).

Laboratory testing of the interfacial shear strength between new and old cement

mantles has been used to evaluate the mechanical integrity of bilaminar cement mantles

created in C-C revision THA. With correct surgical technique including the provision of

a clean and dry old cement mantle with exclusion of blood and debris at the interfacial

region, acceptable interfacial strength may be obtained between layered cement mantles

(Gruen et al. 1976; Greenwald et al. 1978; Li et al. 1996). Duration of post-cure between

PMMA cement mantles is important as PMMA exhibits variability in material properties

related to time (Wilde et al. 1975; Lee et al. 1977; Lee et al. 1978) and the energy

required to separate adhesive bonds formed between solid polymer masses is known to

increase with duration of contact (Wool 1991). It would be reasonable therefore to

Page 28: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

3

postulate that the interfacial region of a bilaminar cement mantle will demonstrate

variability in interfacial shear strength related to duration of post-cure.

1.2 Research Objectives and Hypotheses

Objectives of this research are to describe:

1. Shear properties of Surgical Simplex P and Tobramycin Antibiotic Simplex

PMMA cements (Stryker International; Rutherford, NJ), and to monitor the

changes with respect to duration of post-cure;

2. Material properties of the interfacial region between new and old cement

mantles in shear, and to monitor the changes with respect to duration of

interfacial post-cure;

3. Influence of Tobramycin antibiotic present within Antibiotic Simplex on C-C

interfacial shear properties; and

4. Mechanisms of adhesion between layered PMMA cement mantles.

Specific hypotheses relating to the shear properties of uniform and bilaminar PMMA

cement mantles discussed within this thesis are:

1. Surgical Simplex and Tobramycin Antibiotic Simplex PMMA cements have

material shear properties which demonstrate variability with respect to duration

of post-cure.

2. Antibiotic Simplex PMMA cement demonstrates equivalent shear properties to

Surgical Simplex P.

3. The ultimate shear strength of the interfacial region between bilaminar cement

mantles approximates that of uniform mantles.

4. Bilaminar PMMA cement mantles demonstrate variation in the interfacial shear

properties with duration of interfacial post-cure.

5. Antibiotic Simplex when used in bilaminar cement mantles results in weakening

at the interfacial region.

6. Interfacial adhesion by mechanisms other than mechanical interlock play a

significant role in the bond formed between new and old PMMA cements.

Page 29: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

4

1.3 Thesis Overview

Chapter 2 Literature Review is separated in to five main topics vis. PMMA

Cement in Primary and Revision Hip Joint Arthroplasty, Cement-Within-Cement

Femoral Revision, Composition and Material Properties of PMMA cement, Evaluation

of the Material Properties of PMMA Cement and Bilaminar Mantles in Shear and

Context of Thesis to Literature Review. Section 2.1 PMMA Cement in Primary and

Revision Hip Joint Arthroplasty describes the clinical application of PMMA cements in

current orthopaedic practice. Rationale is provided for the use of cemented femoral

components in preference to uncemented in both primary THA and hip joint

hemiarthroplasty. The aetiology of revision THA and issues relating to cement removal

from the femur during revision are discussed. Section 2.2 Cement-Within-Cement

Femoral Revision describes the clinical indications and requirements for the conduct of

C-C femoral prosthesis revision. Preoperative planning for C-C revision by radiographic

assessment of the bone-cement interface is discussed. Operative techniques and implants

facilitating the conduct of C-C femoral revision are presented. Section 2.3 Composition

and Material Properties of PMMA Cement provides an overview of commercial PMMA

acrylic cements used in hip joint arthroplasty, with particular reference to Surgical

Simplex P. Intraoperative factors influencing the material properties of PMMA cements

are discussed. Section 2.4. Evaluation of the Material Properties of PMMA Cement and

Bilaminar Mantles in Shear describes die-punch testing and other techniques for the

investigation of cements in shear. Previous research relating to the shear properties of

PMMA cement and bilaminar interfaces is presented. Section 2.5 Summary of Literature

Review and Relevance to Thesis discusses the rationale of the selected thesis objectives

and hypotheses with respect to the available literature.

Chapter 3 Experimental Evaluation of the Shear Properties of PMMA Cement

describes material properties of PMMA cements in shear, with specific reference to the

variation in material properties with respect to duration of post-cure. These experiments

provide data to validate the die-punch testing method and to supply reference data for

interpretation of results obtained for bilaminar PMMA cement test experimentation.

Page 30: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

5

Chapter 4 Experimental Evaluation of the Shear Properties of Bilaminar Cement

Mantles describes the material properties of cement – cement interfaces in shear.

Objectives selected for this experimentation are based on issues relevant to the clinical

practice of C-C revision arthroplasty. Variation in the shear properties of C-C interfaces

with respect to duration of post-cure and the influence of the commercial inclusion of

antibiotics on interfacial shear properties is evaluated. Mechanisms of interfacial

adhesion between layered PMMA cement mantles are discussed.

Main conclusions and recommendations of the thesis are provided in Chapter 5.

Page 31: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

6

Page 32: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

7

Chapter 2

Literature Review

2.1 PMMA Cement in Primary and Revision Hip Joint Arthroplasty

Section 2.1 describes the clinical application of PMMA cements in current orthopaedic

practice. Rationale is provided for the use of cemented femoral components in

preference to uncemented in both primary THA and hip joint hemiarthroplasty. The

aetiology of revision THA and issues relating to cement removal from the femur during

revision are discussed.

2.1.1 PMMA cement in primary total hip arthroplasty

Polymethylmethacylate (PMMA) acrylic cements have been regularly used to

secure hip arthroplasty prostheses since the popularisation of joint replacement by Sir

John Charnley in the 1960’s (Charnley 1960; Ege et al. 2001; Smith 2005). The

functions of PMMA cement are to enable stable fixation of prosthetic components by the

anchoring of implant to bone, to provide an elastic buffer for the uniform transfer of load

from prosthesis to bone and as a delivery system for the local administration of

antibiotics (Kuehn et al. 2005).

Hip joint arthroplasty is one of the most successful and cost effective surgical

interventions in medicine, with approximately 27 000 procedures performed in Australia

per annum (Garrellick et al. 1998; Australian Orthopaedic Association National Joint

Replacement Registry 2004). Cemented conventional and hybrid total hip arthroplasty

(THA) prostheses constitute 49.6% of primary replacements performed in Australia,

with the Exeter cemented stem being the most frequently implanted primary femoral

component considering both cemented and uncemented fixation methods (Australian

Orthopaedic Association National Joint Replacement Registry 2004). Cemented and

hybrid implants constitute greater then 90% of primary THA procedures in both Sweden

and Norway (Department of Orthopaedic Surgery Haukeland University Hospital 2004;

The Swedish National Hip Arthroplasty Register 2004).

Page 33: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

8

Cemented THA continues to demonstrate exceptional long term durability and

performance. The Swedish National Hip Arthroplasty Register has now recorded the

results of almost 230 000 primary THA procedures since 1979. With implant failure

defined as revision, cemented implants in the Swedish Hip Registry demonstrate 90.6%

12 year survival, which compares favourably to the outcomes observed with uncemented

THA implants, demonstrating only 74.5% survival at 12 years. Superior survivorship of

cemented THA implants compared to uncemented is similarly demonstrated in both the

Australian (Figure 2-1) and Norwegian joint arthroplasty registers.

While there is an increasing trend for use of uncemented THA in younger and

higher demand patients, improved prosthetic survivorship for uncemented implants in

these groups has not been demonstrated, and substantial evidence exists to suggest

otherwise. The Swedish Hip Arthroplasty Register shows that for male patients under

the age of 50 years, representing the youngest and highest demand patient cohort,

cemented THA has a 83.7% 12-year survivorship, compared to 71.5% for uncemented

prostheses (p<0.001; χ squared analysis). Indeed, the Swedish arthroplasty register

demonstrates superior prosthetic survivorship for cemented implants in all patients

regardless of age or gender (Table 2-1).

Despite strong evidence to support the use of PMMA cement as a method of

fixation in THA, the use of cemented implants is decreasing in favour of uncemented

technology (Jones et al. 1987; Australian Orthopaedic Association National Joint

Replacement Registry 2004; The Swedish National Hip Arthroplasty Register 2004).

Uncemented implants avoid the potential adverse intraoperative cardiovascular events

associated with pressurised cement implantation, although with appropriate surgical

technique and anaesthetic management, the significance of PMMA cement on

cardiovascular function has been questioned (Yasuda et al. 1975; d'Hollander et al.

1979; Dahl et al. 1992; Karlsson et al. 1995; Breusch 2001). Uncemented femoral

components characteristically have reduced inventory of instrumentation required for

implantation and are quicker to implant thereby reducing anaesthetic time. Uncemented

acetabular components also allow the use of alternative bearing surfaces to polyethylene,

Page 34: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

9

Figure 2-1 Cumulative risk for revision of THA by cement status excluding

infection (Australian Orthopaedic Association National Joint Replacement Registry

2004).

Cemented (%) Uncemented(%) P

All Implants 90.6 74.5 <0.001Age < 50, Male 83.7 71.5 <0.001Age < 50, Female 74.8 67.6 <0.001Age 50 - 59, Male 82.4 81.5 0.52Age 50 - 59, Female 88.8 71.9 <0.001Age 60 - 75, Male 89.4 81.1 <0.001Age 60 - 75, Female 92.7 91.9 0.62Age >75, Male 93.5 * *Age >75, Female 96.4 * *

Table 2-1 Swedish National Hip Arthroplasty Register percentage survivorship for

cemented and uncemented THA implants after 12 years with failure defined as

revision for any cause. ^ = indicates 10 year follow up data. * = insufficient data.

Statistical analysis by two tailed χ2 test (The Swedish National Hip Arthroplasty

Register 2004).

Figure removed for digital reproduction.

hoshiko
Rectangle
halla
This table is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 35: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

10

the wear debris of which have been implicated in osteolysis about joint arthroplasty

implants. Furthermore, in revision situations with a solidly fixed uncemented acetabular

shell, modification of the bearing surface material or geometry is made possible by liner

exchange without disruption of the implant – bone interface (Australian Orthopaedic

Association National Joint Replacement Registry 2004). The exothermic reaction of

PMMA polymerisation has not been demonstrated to result in temperature elevations

significant enough to result in bone necrosis during THA (Jefferiss et al. 1975; Reckling

et al. 1976; Sih et al. 1980; Schultz et al. 1987; Kuehn et al. 2005).

2.1.2 PMMA cement in hip joint hemiarthroplasty

Hemiarthroplasty prostheses are principally used for the management of

displaced intracapsular hip fractures in the elderly, with greater than 95% of

hemiarthroplasty procedures in Australia conducted for this indication (Gebhard et al.

1992; Australian Orthopaedic Association National Joint Replacement Registry 2004).

The demographic features of patients requiring hip joint hemiarthroplasty are somewhat

different to THA, with patients on average being 14 years older (mean age 83 years) and

75% of patients being female (Figure 2-2). In Australia unipolar articulations are used in

72.1% of hemiarthroplasty procedures, with monoblock stems accounting for 83% of all

unipolar implants. Currently 45.8% of hip hemiarthroplasty prostheses recorded in the

Australian National Joint Registry are cemented, however the use of PMMA cement is

increasing due to the poor outcomes observed with uncemented hemiarthroplasty

implants (Australian Orthopaedic Association National Joint Replacement Registry

2003; Australian Orthopaedic Association National Joint Replacement Registry 2004;

Weinrauch et al. 2005; Weinrauch et al. 2006).

Elderly patients sustaining fractures of the proximal femur frequently have

significant concomitant medical conditions resulting in high mortality rates after

operative management (Parvizi et al. 2004). The 1 year mortality rate after the diagnosis

of femoral neck fracture across multiple studies is approximately 30%, with the majority

of deaths within the first 6 months of surgery (Kenzora et al. 1984; Lu - Yao et al. 1994;

McAuliffe et al. 2004). Patients surviving the first 6 – 12 months after surgical

management of hip fractures however have an equivalent mortality rate when compared

Page 36: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

11

Figure 2-2 Age and sex distribution for hip joint hemiarthroplasty (Australian

Orthopaedic Association National Joint Replacement Registry 2004).

to matched controls (Fitts et al. 1959; Gordon 1971; Kenzora et al. 1984; Taine et al.

1985). Despite high mortality rates associated with the operative management of hip

fractures, hemiarthroplasty endoprosthesis implantation has been demonstrated to be a

highly cost effective treatment in terms of quality adjusted life years gained (Parker et al.

1992).

Excluding infection, common diagnoses necessitating the revision of

hemiarthroplasty prostheses are early aseptic loosening and secondary acetabular

arthrosis (D'Arcy et al. 1976; Devas et al. 1983; Dixon et al. 2004). Early revision due to

failure on the femoral side of hip hemiarthroplasty has a strong association with the use

of uncemented implants (Wrighton et al. 1971; Lausten et al. 1982; Sonne-Holm et al.

1982; Dorr et al. 1986; Gebhard et al. 1992; Warwick et al. 1998; Khan et al. 2002;

Australian Orthopaedic Association National Joint Replacement Registry 2004; Parker

et al. 2004). Multiple studies have demonstrated uncemented monoblock

Figure removed for digital reproduction.

Page 37: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

12

hemiarthroplasty of the hip joint to be technically demanding, with a high rate of

intraoperative error contributing to early failure of the prosthesis (Kwok et al. 1982;

Sharif et al. 2002; Weinrauch 2004; Yau et al. 2004; Weinrauch et al. 2005; Weinrauch

et al. 2006). Parker and Gurusamy (2004) in a meta-analysis of fifteen trials involving

1670 patients concluded uncemented prostheses were associated with a higher risk of

failure to regain mobility (HR 0.60; 95% CI 0.44 to 0.82) and a higher incidence of

postoperative pain at one year (HR 0.51; 95% CI 0.31 to 0.81) (Parker et al. 2004). Khan

et al. (2002) reviewed the results of 244 patients managed with the Austin Moore

prosthesis, with and without the use of PMMA cement. At 32 – 36 month follow up,

patients who were managed without cement had greater pain (p=0.003), reduced walking

ability (p=0.002), increased use of walking aids (p=0.003) and reduced capacity for

activities of daily living (p=0.009). The use of cemented hemiarthroplasty in the elderly

was supported by these findings (Khan et al. 2002). Foster et al. (2005) in a review of

244 patients undergoing Uncemented Austin Moore (UAM) or Cemented Thompson

(CT) prosthesis implantation for femoral neck fractures found a 7% periprosthetic

fracture rate over a 2 year period in patients managed with the UAM (Foster et al. 2005).

There were no periprosthetic fractures observed in 174 patients managed with the CT

over the same time period (p=0.002) and the periprosthetic fracture rate was found to be

independent of age or gender. Sharif & Parker (2002) in a review of 12 month outcomes

after UAM implantation in 243 patients found 25.1 % of patients had residual hip pain

and 7% required revision within one year for aseptic loosening (Sharif et al. 2002).

Jadhav et al. (1996) reviewed the 12 – 48 month results of 40 patients managed with the

UAM prosthesis and demonstrated 70% had pain of non infective origin, with calcar

resorption detected in 85% of implantations and radiologic evidence of stem migration

in the majority of cases (Jadhav et al. 1996). The Australian National Joint Registry

demonstrates that the UAM has a significantly higher early revision rate compared to the

CT prosthesis, with over 6% of implantations requiring revision within 2.5 years (HR

2.89; 95% CI 1.8 – 4.6; p<0.0001) (Figure 2-3). The Australian registry data also

suggest that the Thompson prostheses when inserted without PMMA cement has a

similar failure rate to the uncemented Austin Moore, but if an Austin Moore prosthesis is

cemented then the incidence of early revision is similar to that seen with the Cemented

Page 38: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

13

Figure 2-3 Cumulative risk for revision of UAM and CT prostheses (Australian

Orthopaedic Association National Joint Replacement Registry 2004).

Thompson (Australian Orthopaedic Association National Joint Replacement Registry

2004).The Australian Joint Registry concludes on the basis of these observations that

continued use of the UAM prosthesis cannot be justified and recommends the use of

PMMA cement as the critical factor influencing results after hemiarthroplasty of the hip

joint (Australian Orthopaedic Association National Joint Replacement Registry 2004).

On the basis of the extensive literature supporting the use of PMMA cement in hip joint

hemiarthroplasty it would therefore appear that the choice of which hemiarthroplasty

prosthesis to use is somewhat less important than the decision to use cement. In

conclusion, femoral side failure after hip joint hemiarthroplasty is a complication largely

avoidable with the utilisation of PMMA cement.

Despite concerns regarding the poor functional outcome and increased revision

rates associated with uncemented hemiarthroplasty prostheses, apprehension about the

systemic effects of PMMA acrylic cement implantation in the elderly patient continues

Figure removed for digital reproduction.

Page 39: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

14

to influence prosthesis selection. Cement insertion has been demonstrated to adversely

effect pulmonary and cardiovascular function during the conduct of surgery and the

immediate postoperative period, which may be poorly tolerated in the elderly with pre-

existing co morbidity (Ries et al. 1993; Pitto et al. 1999). Parvizi et al. (1999) in a

review of 38488 hip arthroplasty procedures reported 23 intraoperative deaths due to

irreversible cardiorespiratory disturbances initiated during cementing. 12 of the 23

deaths were in patients undergoing cemented hemiarthroplasty, with the incidence of

intraoperative death contributed by the use of acrylic bone cement during

hemiarthroplasty found to be less than 0.2%. No intraoperative deaths were recorded in

any of the 15411 uncemented hip replacements during the 28 year period. Elderly

patients with pre-existing cardiovascular conditions undergoing arthroplasty for the

management of hip fractures were identified as patients at risk of intraoperative death

associated with the use of cement (Parvizi et al. 1999). These findings are consistent

with a review of 1118 hemiarthroplasty procedures conducted in Queensland public

hospitals, which demonstrated the intraoperative mortality rate attributable to the use of

PMMA cement to be 1/ 738 (Weinrauch et al. 2006). Patterson et al (1991) in a review

of 7 intraoperative cardiac arrests occurring during hip arthroplasty implicated advanced

age, osteoporotic bone and larger volumes of cement as risk factors during the

cementing process (Patterson et al. 1991). Offset against the quantitatively small risk

associated with the use of acrylic bone cement however are the previously outlined

advantages demonstrated for its use in hemiarthroplasty of the hip joint and the high

mortality risk associated with revision arthroplasty. The risk of intraoperative

complications related to the use of cement may be reduced by canal preparation using

pulsatile lavage, venting the femur, avoiding excessive pressurization during insertion

and adequate preoperative hydration (Patterson et al. 1991; Parvizi et al. 1999; Pitto et al.

1999; Breusch 2001).

For secure hemiarthroplasty prostheses, development of symptomatic acetabular

disease including acetabular erosion and protrusio represent common causes for

conversion to THA. Warwick et al. (1998) in a review of 56 patients requiring revision

of failed hemiarthroplasty to THA at an average of 33 months after primary implantation,

found the mode of hemiarthroplasty failure to be femoral loosening in 21 patients

Page 40: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

15

(37.5%), acetabular erosion in 26 (46.5%) and combined femoral loosening and

acetabular disease in 5 (9%). Failure by femoral component loosening presented earlier

than acetabular erosion, and was associated with the use of uncemented

hemiarthroplasty prostheses (Warwick et al. 1998). It is important to note that over half

of hemiarthroplasty failures in this series were for the diagnosis of acetabular erosion,

and in over 80% of these cases the femoral component was secure. Revision arthroplasty

in this situation is made significantly more challenging due to the requirement to remove

a firmly secure femoral component from osteoporotic bone. Philips (1989) reviewed the

results of 72 subcapital fractures of the femoral neck managed with CT prosthesis 3 to

14 years after the procedure. Development of symptomatic acetabular erosion was

associated with high levels of functional activity as defined by the University of

California at Los Angeles hip rating system. 34 of 38 (89.5%) patients scoring 4 or

greater (capable of mild activity, including walking, limited housework and shopping)

were diagnosed with acetabular erosion, compared to 0/ 34 scoring 3 or less (p<0.001).

Radiographic evidence of acetabular erosion was also associated with pain and disability

during activity. Of the 57 hips displaying no radiographic evidence of loosening of the

hemiarthroplasty prosthesis, 21 displayed radiographic evidence of advanced acetabular

erosion, all of which were symptomatic. No patient with a radiographically well fixed

stem without evidence of acetabular erosion was symptomatic (Phillips 1989). Yamagata

(1987) demonstrated the incidence of acetabular wear after monopolar and bipolar

hemiarthroplasty increases over the first 5 years after surgery (Figure 2-4). Acetabular

Figure 2-4 Increasing incidence of acetabular disease over time with patients

managed with hip hemiarthroplasty (Yamagata et al. 1987)

Figure removed for digital reproduction.

Page 41: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

16

wear was also associated with osteoporosis (p<0.05) and poor acetabular fit due to

inappropriately undersized prosthesis selection (p<0.01) (Yamagata et al. 1987).

Similarly other authors have demonstrated the risk of developing symptomatic

acetabular wear is higher in patients of younger age and higher functional demands

(D'Arcy et al. 1976; Kofoed et al. 1983).

2.1.3 Revision hip arthroplasty and cement removal

Diagnoses leading to the requirement of revision arthroplasty within the Swedish,

Australian and Canadian joint arthroplasty registers are presented in Table 2-2, Table

2-3 and Figure 2-5. Aseptic loosening of prosthetic components is consistently the

leading cause for revision of THA. Aseptic loosening clinically presents with increasing

pain, frequently of a mechanical or activity related nature, typically many years after the

primary surgery. A complete review of the radiographic assessment of THA implant

stability with cemented femoral components is presented in Chapter 2.2.2. Radiological

findings may include progressive radiolucency at the bone cement interface or between

the bone and implant for uncemented prostheses. Radionuclear studies may reveal

abnormal uptake of Tc99MDP consistent with increased bone turnover about the

prosthesis due to bone remodelling. Arthrography of the effected joint may show an

abnormal tracking of contrast medium between the bone–cement interfaces. Aspiration

of the joint for bacterial culture analysis is frequently required to assist in the exclusion

of prosthetic loosening secondary to infection. At time of revision, a loose implant is

unstable to surgical manipulation, but still may be difficult to revise due to sections of

cement or porous surface remaining well fixed to bone. Histological findings of aseptic

loosening include formation of a fibrous tissue interface between the bone and prosthesis,

allowing relative micro motion of the implant (Spinelli 1976; Linder et al. 1983).

Revision THA procedures in the Australian National Joint Registry are

classified as major or minor. Major revision is defined as removal or replacement of a

device which interfaces with bone, considering either the acetabular or femoral

components. Minor revision by comparison is defined as replacement of a prosthetic

device not interfacing with bone, for example exchange of the prosthetic head of a

Page 42: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

17

Diagnosis Number PercentAseptic Loosening 13581 76.0Deep Infection 1292 7.2Dislocation 1176 6.6Fracture 966 5.4Technical Error 447 2.5Implant Fracture 276 1.5Miscellaneous 86 0.5Pain Only 57 0.3

Total 17881 100.0

Table 2-2 Diagnosis requiring revision hip replacement for primary THR

performed 1979 - 2003 (The Swedish National Hip Arthroplasty Register 2004).

Figure 2-5 Diagnosis requiring revision hip replacement (Canadian Institute for

Health Information 2004).

Figure removed for digital reproduction.

hoshiko
Rectangle
halla
This table is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 43: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

18

Diagnosis Number PercentAseptic Loosening 5575 49.5Dislocation 1688 15.0Lysis 1133 10.1Fracture 937 8.3Infection 855 7.6Wear Acetabulum 375 3.3Implant Breakage 291 2.7Pain 208 1.8

Total 11261 100.0

Table 2-3 Diagnosis requiring revision hip replacement (Australian Orthopaedic

Association National Joint Replacement Registry 2004).

modular femoral stem or liner exchange with an uncemented acetabular component

(Australian Orthopaedic Association National Joint Replacement Registry 2004).

Revision of cemented hip arthroplasty traditionally requires extraction of both

loose and secure cement after removal of the failed components. Removal of a cemented

femoral prosthesis may be considered in two steps, vis. extraction of the prosthesis and

disruption of the cement – bone interface. For taper design stems without porous coating,

removal of the component is easily completed by the selective removal of overhanging

cement from the shoulder of the stem, followed by application of a retrograde force in

line with the femoral component (Figure 2-6). In contrast however, removal of femoral

cement during revision THA is a time-consuming and hazardous procedure. Proximally

the cement is easily visualised, however the surrounding cortex may be thin and fragile

because of osteolysis. Distally the cement mantle becomes increasingly difficult to

visualise, particularly beyond the metaphyseal flare, sometimes requiring extensile

exposures such as extended trochanteric osteotomy or distal cortical perforation for

adequate access (Hozack 1998; Harkess 1999). Even within major arthroplasty centres,

cement removal results in intraoperative femoral fracture in approximately 8% of

revisions (Gardiner et al. 1993). Other risks associated with cement removal include

cortical perforation, loss of cortical bone stock, excessive blood loss and increased

duration of anaesthesia (Gray 1992; Hozack 1998).

hoshiko
Rectangle
halla
This table is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 44: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

19

Figure 2-6 Extraction of cemented femoral component in line of implant long axis

(Hozack 1998).

Multiple techniques and instrumentation systems have been developed to assist

in the removal of retained cement from the femoral canal, including the use of hand and

power tools, ultrasonic devices and lasers (Choy et al. 1987; Gray 1992; Schurman et al.

1992; Brooks et al. 1993; Gardiner et al. 1993; Cautilli et al. 1994; Hozack 1998).

Femoral cortical perforations occur in approximately 10% of cases using high speed

cement removal instruments (Turner et al. 1987). Such perforations often occur in an

uncontrolled manner producing large gaps in the cortex, rendering the femur at risk of

postoperative fracture (Larson et al. 1988). Cortical perforations commonly occur

distally and posteriorly within the femur, representing a difficult region to reconstruct

for anatomical and surgical exposure perspectives (Gray 1992; Harkess 1999).

Combined use of hand and ultrasonic devices reduces the risk of cortical perforation to

4%, however superficial bone burns are sustained in 9% of cases (Gardiner et al. 1993).

Sustained temperature elevation of magnitude high enough to result in clinically

significant thermal bone necrosis at the bone-cement interface is not produced with

Figure removed for digital reproduction.

Page 45: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

20

appropriate use of ultrasonic cement removal tools however (Brooks et al. 1993). Fume

emissions resulting from ultrasonic removal of PMMA cement contain at most 1 part per

million MMA, well below the National Institute for Occupational Safety and Health

recommended exposure limit (Cautilli et al. 1994). Nd:YAG laser disintegration of

cement remains experimental, and is associated with the production of trace

concentrations of aromatic hydrocarbons (Choy et al. 1987).

Schurman and Maloney (1992) describe an interesting technique of femoral

cement removal taking advantage of the high cement-cement interfacial shear strengths

obtained with suitable preparation of the old PMMA cement mantle. After removal of

the femoral component by use of extractor, the retained cement is prepared by wire

brushing, irrigation and drying. A fresh layer of PMMA cement is introduced in a

retrograde manner with the use of a thin nozzle cement gun, and a threaded rod is placed

within the polymerising mass. After hardening of the new cement mantle has occurred,

the threaded rod is screwed out leaving a bilaminar cement mantle with an inner

threaded portion. Extraction rods are then screwed into the centre mantle in 2.5cm

increments, and cement is extracted by the use of a slap hammer. Segmental extraction

of the outer cement is thereby achieved. Using this technique Schuerman and Maloney

(1992) report the results of fifteen revisions, including cases in which the cement was

assessed as firmly secured at the bone–cement interface. No femoral fractures or

perforations were sustained, and complete removal of cement was obtained in 75% of

patients. It is interesting to note that for this technique to be successful, the shear

strength of the interface between new and old cement must be stronger than the bone

cement interface, and greater than the tensile strength of PMMA cement at the point of

controlled mantle fracture (Schurman et al. 1992).

The 2004 Australian Joint Registry demonstrates that 35.8% of major revision

THA procedures are for isolated acetabular revision, as compared to isolated femoral

component revision conducted in only 20.8 % of cases (Table 2-4). This observation

suggests that femoral reconstruction in primary THA in general maintains superior

survivorship to that of acetabular reconstruction. For cemented implants this conclusion

is also supported by the Swedish National Hip Arthroplasty Register and the published

long term survivorship series of cemented THA prostheses (Fowler et al. 1988; Barrack

Page 46: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

21

et al. 1992; Mulroy et al. 1995; Williams et al. 2002; The Swedish National Hip

Arthroplasty Register 2004). Revision THA for isolated acetabular revision may

reasonably expected to increase over time, due increasing frequency of THA

implantation and primary cemented femoral implants providing reproducible long term

results (Australian Orthopaedic Association National Joint Replacement Registry 2004).

Component Used Number Percent Femoral and Acetabular 3178 38.2Acetabular Only 2974 35.8Femoral Only 1730 20.8Cement Spacer 168 2Removal Prosthesis 131 1.6Bipolar Head & Femoral Comp. 119 1.4Reinsertion of Components 9 0.1

Table 2-4 Components used in major THA revision (Australian Orthopaedic

Association National Joint Replacement Registry 2004).

During isolated acetabular revision, surgical access is made more difficult by the

position of the femoral neck and prosthetic head with respect to the socket within the

surgical field (Amstutz et al. 1982). Options to improve surgical exposure in these

situations usually include soft tissue and capsular releases, allowing recession of the

femoral head and neck into a position suitable to allow adequate access for

instrumentation to permit acetabular reconstruction. By performing additional capsular

releases a pocket is created, and with the use of retractors the femoral component may

be positioned and restrained into an appropriate location. Aggressive soft tissue

dissection may however predispose the patient to postoperative prothetic instability

(Goldstein et al. 2001). Furthermore, during complex acetabular reconstruction

significant osteolysis may be encountered, and forceful retraction by leverage upon the

acetabular rim may risk intraoperative periprosthetic fracture. Another alternative to

improve acetabular access with the use modular femoral components is for removal of

the prosthetic head and subsequent exchange, however this exposes the morse taper to

damage during the conduct of surgery and the retained femoral neck may still obscure

hoshiko
Rectangle
halla
This table is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 47: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

22

access to the acetabulum. Isolated acetabular revision with a cemented stem and secure

PMMA mantle in situ may alternatively be facilitated by the removal of the entire

femoral component, with retention of the well-fixed cement (Archibald et al. 1988).

Extraction of the femoral component while minimising disruption of the cement mantle

during revision is assisted by the use of polished taper stem designs during the primary

arthroplasty. In this situation the femoral component may be easily extracted by the

selective removal of cement overhanging the shoulder of the prosthesis in proximal and

lateral regions, followed by in-line withdrawal of the stem by mallet and punch or slap-

hammer device. In theory this technique should be classified as a minor revision

according to Australian National Arthroplasty Registry guidelines, as the cement – bone

interface has not been disrupted. Removal of the entire femoral component with

subsequent cement – within – cement (C-C) revision allows optimal exposure of the

acetabulum with minimisation of the soft tissue dissection and retraction required for

femoral positioning. After acetabular reconstruction has been completed femoral

revision may be conducted by replacing the original stem into the old mantle or selecting

a new femoral prosthesis (Archibald et al. 1988). Replacing the original stem has some

theoretical disadvantages however and is infrequently performed, as the original

prosthesis may have accumulated fatigue or become damaged during the process of

extraction therefore predisposing to delayed failure by stem fracture. New femoral

components should therefore be used when performing using C-C revision (Personal

communication, R. Ling 2005). Within the tolerances of the manufacturing process,

minor differences in geometry exist between matched stems of the same design, size and

offset (Stryker Howmedica Osteonics 2002). For this reason, selection of a smaller size

stem is recommended for C-C revision, with the relative reduction in stem size

determining the thickness of the new cement mantle.

Page 48: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

23

2.2 Cement – Within - Cement Femoral Revision

Section 2.2 describes the clinical indications and requirements for the conduct of C-C

femoral prosthesis revision. Preoperative planning for C-C revision by radiographic

assessment of the bone-cement interface is discussed. Operative techniques and implants

facilitating the conduct of C-C femoral revision are presented.

2.2.1 Clinical application

Cement-within-cement (C-C) femoral revision refers to the technique of

extracting a well fixed stem with retention of a stable cement mantle, and subsequent

implantation of a cemented revision stem into the old cement mantle. The operative

techniques for C-C revision femoral arthroplasty are illustrated in Figure 2-7. Important

points to note in the revision technique are provision of a clean and dry cement mantle,

the early retrograde insertion of PMMA cement by gun and the use of a femoral

component with revision centraliser to allow normal stem subsidence with polished taper

stem designs (Gruen et al. 1976; Greenwald et al. 1978; McCallum et al. 1995; Li et al.

1996; Flivik et al. 1997; Park et al. 2001; Hubble et al. 2005; Stryker International 2005).

Indications and prerequisite requirements for the conduct of C-C femoral

revision arthroplasty are presented in Table 2-5. The primary cement mantle may be

retained with no modification, or alternatively portions of the cement mantle may be

removed as required either proximally or distally (Hozack 1998; Harkess 1999).

The rationale of cement-cement femoral revision is to avoid the complications

associated with unnecessary cement removal during revision hip arthroplasty as

previously described in Chapter 2.1.4. Advantages of C-C techniques for femoral

component revision are detailed in Table 2-6. C-C revision is assisted by the use of

polished taper design femoral components in primary arthroplasty, as the stem geometry

Page 49: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

24

Figure 2-7 Surgical technique of C-C femoral revision (Stryker International).

Figure removed for digital reproduction.

Page 50: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

25

and absence of firm adhesion to the cement mantle facilitates stem extraction without

damaging the cement-bone interface (Archibald et al. 1988; Hozack 1998).

Traditional cemented revision by extraction of the entire cement mantle

regardless of the stability of the bone – cement interface potentially compromises the

performance of the cemented revision implant. Halawa et al. (1978) demonstrated the

ultimate shear stress to failure at the interfacial region between bone and cement was

limited by the strength of cancellous bone. Extraction of secure cement requires the

simultaneous removal of cancellous bone, compromising the host surface available for

cement interdigitation during cemented revision THA. For this reason, the ultimate shear

strength of bone cement interface after revision arthroplasty is 30 per cent weaker than

primary interfaces (Rosenstein et al. 1992).

C-C revision is not an applicable technique in the majority of revision hip

arthroplasty procedures due to the presence of osteolysis, infection or femoral cement

loosening (Barrack et al. 1995; McCallum et al. 1995). Contraindications and

disadvantages for C-C femoral revision are presented in Table 2-7.

Only limited clinical studies are available describing the results obtained after C-

C revision. Liebermann et al. (1993) reviewed the results of 19 revision THA procedures

using this technique at an average of 64 months. Despite cement retention techniques,

femoral perforation was sustained in 2 cases in which the original stem was in varus and

limited cement extraction was attempted, thus highlighting the difficulties involved in

cement removal during femoral revision procedures. At final review no C-C femoral

component had required revision for loosening and all were stable on radiographic

evaluation. The authors recommended C-C revision for selected patients with good bone

stock and a secure cement mantle (Lieberman et al. 1993). McCallum and Hozak (1995)

reviewed the results of 15 patients with Barrack grade A or B femoral cement mantles

managed with C-C revision using ultrasonic tools to reshape the existing cement mantle

(Barrack et al. 1992). No femoral perforations or fractures were sustained, and no patient

required extensile approaches for cement removal. Radiographic assessment 4 years

after revision revealed no evidence of loosening in any femoral component or adverse

effect on the bone-cement interface. One patient required re-revision for dislocation at 1

Page 51: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

26

Table 2-5 Indications and requirements for C-C Revision Hip Arthroplasty. Refer

to Figure 2-9 and text for explanation of Gruen modes of cemented femoral

component failure.

Indications for C-C Revision Hip Arthroplasty

1. Facilitate surgical exposure in complex acetabular revision

2. Fractured femoral component

3. Correction of leg length discrepancy after hip arthroplasty

4. Recurrent dislocation of THA where prosthetic femoral anteversion is

assessed as contributory

5. Revision of monoblock femoral components to modular systems to

enable adjustment of head size or bearing articulation, usually in

combination with acetabular revision

6. Adjustment of femoral offset

7. Conversion of cemented unipolar hemiarthroplasty to THA

8. Gruen Type 1a failure in composite beam femoral component

9. Significant complication associated with cement removal anticipated

Requirements for C-C Revision Hip Arthroplasty

1. Cemented primary and revision femoral components

2. Stable cement mantle and bone-cement interface

3. Technique facilitated by polished taper design primary stem and

specifically designed C-C revision femoral components

4. Ability to extract femoral component without jeopardising mechanical

integrity of bone – cement interface

Page 52: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

27

Table 2-6 Advantages of C-C femoral revision.

year and was managed again with C-C revision. The authors recommended C-C revision

with limited cement mantle modification using ultrasonic instruments as a suitable

technique for management of fractured femoral components, obtaining surgical access

during isolated acetabular revision and correction of stem malposition or articular

bearing modification (McCallum et al. 1995).

The largest clinical series of C-C revisions has recently been reported by Hubble

et al. (2005). Since 1989, 354 cases of C-C femoral revision using the Exeter stem have

been conducted at the Princess Elizabeth Orthopaedic Centre, Exeter England. Clinical

and radiographic results of 175 cases beyond 5 years from revision were presented, with

no patients lost to review. Radiographic analysis revealed no cases of femoral loosening

or osteolysis, and on no occasion was femoral component re-revision required for

subsequent aseptic loosening. The authors conclude C-C revision demonstrates

acceptable outcomes at 5 years and should be considered a valuable approach to revision

hip surgery.

Advantages of C-C Femoral Revision

1. Extensile exposures for cement removal not required

2. Equipment inventory required for revision procedure reduced

3. Cancellous bone retained

4. Allows immediate patient mobility and postoperative rehabilitation

5. Decrease risks associated with cement removal

a. Cortical perforation

b. Blood loss

c. Thermal bone necrosis

d. Intraoperative fracture

6. High bone-cement interface strength maintained

7. Decrease operative time

Page 53: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

28

Table 2-7 Contraindications and disadvantages of C-C femoral revision.

2.2.2 Radiographic assessment of the bone – cement interface

Radiographic evaluation provides valuable preoperative information to assist in

the planning of revision THA. Failure of the bone – cement interface may be predicted

by careful radiographic assessment, and is therefore essential prior to the conduct of C-C

femoral revision, as failure at this interface represents an absolute contraindication to

this technique (Archibald et al. 1988; Lieberman et al. 1993).

Contraindications for C-C Femoral Revision

(Relative and Absolute)

1. Prosthetic joint infection

2. Loose femoral component bone - cement interface

a. Preoperative imaging (see Chapter 2.2.2)

b. Intraoperative assessment

3. Revision using uncemented femoral prosthesis

4. Mechanical integrity of femur questionable, where femoral impaction

allografting or long stemmed revision prosthesis is planned

a. Previous femoral perforation with cement leak

b. Impending fracture

c. Osteolysis

Disadvantages of C-C Femoral Revision

1. Assessment of bone-cement interface integrity

2. Limits selection of revision femoral component

a. Cemented

b. Stem geometry and length

3. Interface between new and old cement - potential mechanism of failure

4. Accumulative fatigue in old cement mantle

Page 54: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

29

Clinical loosening of prosthetic cemented implants in THA typically occurs at

the bone – cement interface. Stem migration at the prosthesis – cement interface is not

considered evidence of mechanical failure of modern polished taper design femoral

components, which are intended to debond and subside within a viscoelastic cement

mantle (Fowler et al. 1988; Williams et al. 2002). Radiographic assessment allows early

detection of subclinical prosthetic failure, as symptomatic loosening of THA

components is often preceded by evidence on plain radiographs (Gruen et al. 1979;

Weissman 1983). Preoperative radiographic assessment of the bone – cement interface

integrity is therefore universally conducted during the initial evaluation of the painful

THA (Evans et al. 1992).

Radiographic assessment of prosthetic stability after THA requires reproducible

and accurate radiographic technique and diagnostic criteria. Clinical and radiographic

criteria used to define loosening significantly influences the documented correlation

between radiographs and operative findings between published studies (Brand et al.

1986). Radiographic criteria defining loosening of cemented femoral and acetabular

components and the uncemented acetabulum are presently well established, however

assessment of the uncemented femoral stem remains problematic (Hodgkinson et al.

1988; Ranawat et al. 1995; Berger et al. 1996; Ritter et al. 1999; Udomkiat et al. 2001).

Radiographic features of cemented component loosening commonly used in clinical

practice are presented in Table 2-8.

Gruen et al. (1979) reviewed the radiographic appearances of 301 cemented

Charnley prostheses in 389 patients at 6 months to 6 years post implantation. Using a

system of delineated sections about the femoral component for zonal evaluation of

radiographic loosening (Figure 2-8), four predominant modes of cemented stem

loosening were described (Figure 2-9). Mode I failure refers to pistoning of the

prosthesis, either inside a fixed cement mantle (Mode Ia) or composite pistoning of the

cement and prosthesis (Mode Ib). Mode Ia failure was associated with distal cement

mantle fractures due to end bearing of the prosthesis. With regard to Mode Ia failure, the

differences in mechanical behavior between a taper-slip femoral component such as the

Exeter stem and composite beam prostheses require consideration. Composite beam

Page 55: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

30

Table 2-8 Radiographic features of femoral component loosening in cemented THA

(Harkess 1999). See Figure 2-8 and text for explanation of Gruen zones.

Figure 2-8 Coronal plane Gruen Zones for the standardised assessment of the

femoral component bone – cement interface (Gruen et al. 1979).

Figure removed for digital reproduction.

halla
This table is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 56: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

31

cemented prostheses are not intended to subside within the surrounding cement mantle,

and debonding of the cement – prosthesis interface in this situation may be considered

evidence of mechanical failure of these stems (Gruen et al. 1979). Subsidence is

however the anticipated behavior for polished tapered collarless stem designs, and when

observed on postoperative radiographs should not be misinterpreted as failure (Fowler et

al. 1988; Williams et al. 2002; Yates et al. 2002). Subsidence of a taper – slip design

stem is readily observed at the prosthesis – cement interface in Zone 1 as demonstrated

in Figure 2-10 and is usually evident 12 months after initial implantation. The normal

subsidence of taper – slip design stems is approximately 2mm in the first two years after

implantation, after which minimal subsidence is observed (Figure 2-11).

Gruen Mode Ib failure (Figure 2-12) is characterized by disruption of the

mechanical integrity between cement and bone. This mode is characterized by a

radiolucent line around most or all of the cement mantle and is the most frequent mode

of failure observed with Charnley prostheses (Gruen et al. 1979).

Gruen failure modes II – IV are associated with varus migration of the prosthesis.

Mode II failure (Medial stem pivot - Figure 2-13) is characterized by medial

Figure 2-9 Gruen modes of radiographic failure of cemented femoral THA

components (for explanation see text) (Harkess 1999).

Figure removed for digital reproduction.

Page 57: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

32

Figure 2-10 Radiograph demonstrating normal subsidence of an Exeter stem, with

cement-prosthesis debonding visible in Gruen Zone 1 (courtesy of R. Crawford).

Figure 2-11 Normal subsidence of a collarless cemented polished double taper

femoral component (Yates et al. 2002).

Figure removed for digital reproduction.

Page 58: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

33

Figure 2-12 Gruen Mode IB failure (compound subsidence). Radiolucency in Zone

1 representing distal migration of the cement – prosthesis composite, with distal

radiolucency and pedestal formation (Harkess 1999).

migration of the proximal stem coupled with lateral migration at the tip. Mode III failure

(Calcar pivot - Figure 2-14) is due to insufficient distal support of the prosthesis, with

“windshield wiper” lateral migration of the distal stem despite adequate proximal

support. Type IV failure (Cantilever bending - Figure 2-15) is characterized by

insufficient proximal and medial support in the calcar region, with subsequent medial

migration of the proximal stem while the distal prosthesis remains rigidly fixed.

O’Neil & Harris (1984) examined the correlation between preoperative

radiographs and operative findings at revision THA in 61 patients. Radiographic

diagnosis of femoral loosening was made on the observation of prosthetic migration or

the presence of a continuous bone – cement radiolucent zone surrounding the prosthesis

measuring at least 2mm in thickness in any region. Operative criteria for the diagnosis of

stem loosing were defined as motion of the prosthesis with the manual application of

Zone 1 Lucency

Pedestal formation

Figure removed for digital reproduction.

Page 59: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

34

Figure 2-13 Gruen Mode II failure (medial stem pivot). Note varus migration of the

stem with separation of the cement – prosthesis interface in zones 1 and 2 (Harkess

1999).

Figure 2-14 Gruen Mode III failure (calcar pivot). Initial postoperative radiograph

depicted on the left for comparison. Progressive varus migration of the distal stem

demonstrated and osteolysis of the greater trochanter however minimal migration

of the proximal prosthesis consistent with pivoting about a proximal (calcar)

location (Harkess 1999).

Separation of cement and

prosthesis

Varus migration

Figure removed for digital reproduction.

Figure removed for digital reproduction.

Page 60: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

35

Figure 2-15 Gruen Mode IV failure (bending cantilever). Serial radiographs from

time of implantation from left to right. Left: Varus malposition of the stem at initial

surgery. Centre: Proximal cement – prosthesis lucency in zone 1 without

abnormality in distal cement fixation. Right: Fragmentation of proximal cement

mantle in Gruen Zone 7 with fatigue failure of prosthesis (Harkess 1999).

strong rotational and varus force without the use of a mallet. 47 of 61 femoral

components were assessed as loose at revision, of which 42 were predicted by

preoperative radiographs. The sensitivity and specificity for plain radiographs was found

to be 89% and 92% respectively. The authors concluded that plain radiographs are

accurate with respect to the assessment of cemented femoral components, however the

majority of patients in their series demonstrated prosthetic migration and the reliability

of bone – cement interface radiolucency assessment was not demonstrated (O'Neill et al.

1984).

Lyons et al. (1985) in a similar study to O’Neil & Harris (1984) examined the

correlation between femoral component loosening at revision THA and preoperative

plain radiography in 45 THA femoral components. Radiographic evaluation was

demonstrated to be 76% sensitive with 100% specificity for the diagnosis of femoral

loosening. Varus migration of the stem, fracture of the prosthesis or cement, endosteal

Figure removed for digital reproduction.

Page 61: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

36

bone resorption, bone - cement radiolucency greater than 2mm and large cystic defects

in the PMMA mantle were strongly correlated with loosening at time of revision. The

authors noted that progressive widening of the bone cement interface beyond 2mm was

highly suggestive of clinical loosening, however extensive radiolucency involving

multiple Gruen zones is required for the radiographic diagnosis of a loose femoral

component (Lyons et al. 1985).

Hendrix et al. (1983) in a retrieval study at revision THA demonstrated 79%

sensitivity and 100% specificity for radiographic assessment of the cemented stem in 31

patients using bone – cement interface lucency 2mm or greater in thickness as the

criteria for radiographic loosening (Hendrix et al. 1983).

Arthrography provides another modality for the preoperative assessment of bone

– cement integrity. Arthrography is performed by the installation of radiographic liquid

contrast into the hip joint and inspection for penetration of the dye into the region of the

bone cement interface, indicative of loosening (Figure 2-16). Digital subtraction and

post-ambulation techniques can be applied to increase the sensitivity of arthrography

(Maus et al. 1987; Hardy et al. 1998). Aspiration of fluid for the hip joint can also be

undertaken during the conduct of arthrography for the assessment of deep prosthetic

infection (O'Neill et al. 1984). The diagnostic accuracy of contrast arthrography in

isolation is questionable for the diagnosis of loosening with uncemented prostheses and

should conducted in conjunction with other modalities (Swan et al. 1991).

Similar to plain radiography, various diagnostic criteria have been described for

the diagnosis of loosening of femoral components with arthrography. Arthrography

across multiple studies has been demonstrated to have approximately 90% sensitivity

and 85% specificity for the diagnosis of femoral component loosening. (Evans et al.

1992). Maus et al. (1987) reviewed the results of 178 hip arthrograms in 170 patients,

with 97 undergoing revision THA. Of the 62 cases where contrast was detected in the

bone – cement interface below the level of the intertrochanteric line, all except one were

loose at revision. Overall sensitivity was 96% with a specificity of 92%. The authors

recommended the use of digital subtraction and routine intra-articular infiltration of

bupivicaine to assist in the diagnostic assessment of loose THA implants. The diagnostic

Page 62: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

37

value of intra-articular local anaesthetic injections for the diagnosis of hip pathology

prior to primary joint arthroplasty has been well established (Crawford et al. 1998).

Due to the high diagnostic accuracy of plain radiography in the assessment of

cemented femoral component loosening, O’Neill et al. (1984) support arthrography only

in the situation where prosthetic migration has not been demonstrated, particularly of the

acetabular component. Arthrography in the preoperative assessment of the painful THA

is recommended when inconclusive information is obtained from serial plain

radiographs (Ovesen et al. 2003).

Figure 2-16 Digital subtraction arthrogram demonstrating loosening of the femoral

implant with extensive extravasation of radio-opaque contrast about the cement –

bone interface and distal contrast leak to cortical perforation (Ovesen 2003).

Radionuclear imaging techniques are also frequently used in the preoperative

assessment of the painful THA. Standard bone scans involve the intravenous

administration of technetium-99m (Tc-99) isotope linked to phosphorus, which is taken

up by the skeleton in accordance with the rate of bone mineralisation (Weissman 1983).

The diagnostic sensitivity and specificity for femoral component loosening in THA is

lower for Tc-99 radionuclear imaging when compared to arthrography (Evans et al.

Figure removed for digital reproduction.

Page 63: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

38

1992; Ovesen et al. 2003). Tc-99 scans normally demonstrate increased uptake in the

perioperative period, with up to 20% of cemented prostheses having false positive

results at 12 months, and almost all uncemented acetabular and femoral components

displaying abnormal uptake at 2 years (Evans et al. 1992).

Gallium – 67 (Ga-67) citrate isotope scanning is often used in combination with

Tc-99 bone scans to assist in the differentiation of septic and aseptic loosening. Ga-67

uptake in inflammatory conditions is predominantly by leukocytes, and incongruent or

increased uptake relative to Tc-99 scanning is suggestive of infection or less commonly

inflammatory disorders (Weissman 1983). Indium -111 diethylenetriaminepentacetic

acid (DTPA) radio-isotope scanning, infrequently used in Australia due to availability,

may be used as an alternative to Ga-67 scanning (Evans et al. 1992).

Due to the diagnostic limitations of Tc-99 bone scanning and other nuclear

radiology studies, their routine use in the assessment of the painful THA is not

recommended (Ovesen et al. 2003). Tc-99 bone scans in isolation do not usually provide

additional information over plain radiography. The value of Tc-99 scans remains in

combination with the use of Ga-67 imaging of the loose THA, to differentiate aseptic

and septic processes. In either of these situations however, the patient has a symptomatic

arthroplasty usually requiring surgical intervention, and the diagnosis of infection can be

made more effectively by other methods such as haematological inflammatory markers,

preoperative joint aspiration, intraoperative frozen section, tissue gram stain and

postoperative cultures. The role of nuclear radiography in the assessment of the painful

THA is therefore limited.

2.2.3 Prosthetic femoral components designed for C-C revision

C-C revision femoral arthroplasty is facilitated by particular design features of

both the primary and revision stems. Primary stems with a highly polished surface

minimise the interfacial bond between cement and prosthesis, and in combination with a

straight double taper geometry, prosthesis removal is facilitated with minimal disruption

of the cement mantle (Archibald et al. 1988; Stryker International 2005 B). Collar – free

stems are in general easier to revise, as the presence of a prosthetic collar may impede

Page 64: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

39

access to the bone-cement interface should removal of portions of the cement mantle be

required (Hozack 1998).

Selection of suitable femoral components for C-C revision requires consideration

of both stem length and offset. Should the revision stem selected be of equal length to

the primary stem removed, limited cement mantle modification by distal reaming is

required to allow implantation of a centraliser to enable normal postoperative subsidence.

Distal cement reamers have recently been developed for this application, and the results

of a pilot study on the Exeter PMMA Cement Reamer are presented in Appendix 3

(Appendix 3 – Technical Report in the Exeter PMMA Cement Reamer).

Selective removal of distally located cement risks femoral perforation and may

also result in the unnecessary removal of well fixed cement. Theoretically, during the

process of limited cement removal retained adjacent cement may also be damaged,

predisposing to cement mantle failure. Selecting a stem of decreased length minimises or

eliminates the requirement for distal cement mantle modification. Smaller length stems

however frequently have a corresponding decrease in femoral offset which will result in

poor soft tissue balancing and early impingement, predisposing to joint instability. The

Exeter Short Revision Stem (ESRS) (Stryker International; Rutherford, NJ) addresses

both length and offset issues in C-C revision femoral arthroplasty by providing a stem

25mm shorter than standard primary Exeter stems with a standard (+44mm) offset

(Figure 2-17). In the long axis of the implant, the ESRS measures 125mm from the distal

tip to shoulder, or approximately 140mm to the centre of hip rotation. The ESRS with

attached revision centraliser will fit within the secure cement mantle of a standard length

Exeter stem or other standard length cemented stem designs with minimal or no distal

mantle modification.

The Exeter Trauma Stem (ETS) (Stryker International; Rutherford, NJ) is a

polished double taper monoblock hemiarthroplasty for cemented use, with +40mm

femoral offset and stem size 0.5 (between sizes 0 and 1) relative to the existing Exeter

inventory (Figure 2-18). Consistent with other unipolar hemiarthroplasty prostheses, the

principal indication for implantation of the ETS is displaced intracapsular fractures of

the femoral neck, particularly in elderly patients. As described in Section 2.1.2,

common diagnoses

Page 65: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

40

Figure 2-17 Standard length Exeter stem with winged centraliser (left) and Exeter

short revision stem (125mm, +44 offset) with revision wingless centraliser (right)

(Stryker International 2005).

Figure 2-18 Exeter Trauma Stem (ETS) monoblock hemiarthroplasty. Unipolar

polished double taper geometry without a prosthetic collar.

Figure removed for digital reproduction.

Page 66: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

41

requiring revision of monopolar hemiarthroplasty prostheses are early aseptic loosening

and secondary acetabular disease (D'Arcy et al. 1976; Devas et al. 1983). As the ETS is

designed for cemented implantation, the risk of revision due to early aseptic loosening is

significantly lower than uncemented hemiarthroplasty prostheses (Section 2.1.2).

However should a stable ETS require revision due to symptomatic secondary acetabular

disease, the design features assist simplified revision by C-C techniques. Due to the

straight double taper geometry and polished surface finish, the ETS may be easily

removed by “tap out” with minimal or no disruption of the existing cement mantle

(Stryker International 2005 B). By comparison, the Cemented Thompson (CT)

prosthesis is much more difficult to revise in situations of symptomatic acetabular

disease, as the curved stem geometry requires significant cement removal and access to

the cement-bone interface is impeded by the presence of a prosthetic collar (Archibald et

al. 1988).

2.3 Composition and Material Properties of PMMA Cement

Section 2.3 provides an overview of the composition and material properties of PMMA

acrylic cements used in hip joint arthroplasty, with particular reference to Surgical

Simplex. Intraoperative factors influencing the material properties of PMMA cements

are discussed. The material properties of PMMA cements in shear are presented

separately in Section 2.4.

Surgical PMMA cement is presented for use in two parts, vis. powder and liquid

components (Figure 2-19). Typically the liquid is greater than 95% methylmethacrylate

(MMA) monomer, with the remaining ingredients consisting of the polymerisation

activator dimethyl-para-toluidine (DMpT) and a stabilising agent hydroquinone. MMA

monomer is an ester of methacrylic acid with polymerisable double bond. MMA is a

translucent colourless liquid with a boiling point of 100oC. It is only slightly water

soluble and has an intense odour, detectable when present greater than 0.2 ppm. MMA is

a mild skin irritant and has the potential to induce hypersensitivity reactions in humans

(Kuehn et al. 2005). The primary constituent of cement powder is

Page 67: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

42

polymethylmethacrylate (PMMA), which may also include the addition of a copolymer

such as styrene. The powder component also includes the polymerisation initiator

dibenzoylperoxide (BPO) and a radio-opacifying agent such as zirconium dioxide or

barium sulphate. Some PMMA cement powder preparations contain antibiotic agents.

The powder and liquid compositions of four PMMA cements commonly used in

Australian orthopaedic practice are presented in Table 2-9. Variations in the composition

and manufacture of the powder and liquid constituents between surgical PMMA

preparations results in differences in the working behaviour and material properties of

the polymerised cements (Kuehn et al. 2005) (Figure 2-20).

Figure 2-19 Presentation of Antibiotic (Tobramycin) Simplex PMMA cement.

When the liquid and powder components are mixed during the conduct of

surgery, the PMMA polymer powder takes up the monomer liquid by physical processes

known as swelling and dissolution. Swelling and dissolution are responsible for the

initial viscosity and working characteristics of the cement, and is influenced by the

cement composition and powder to monomer ratio (Kuehn et al. 2005). PMMA cement

Page 68: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

43

CMW-1 Palacos R Simplex P Zimmer Reg.Powder Benzoyl Peroxide 2.6 0.5-1.6 1.2 0.75BaSO4 9.1 10 10ZrO2 14.9PMMA 88.3 89.3PMMA/MA 83.5-84.7PMMA/PS 82.3

Liquid MMA 98.7 97.9 97.5 97.3Dimethyl toluidine 0.4 2.1 2.5 2.7Hydroquinone (ppm) 15-20 64 75 75

Table 2-9 Composition of four commonly used orthopaedic PMMA cements. MA =

Methacrylic acid; PS = Polystyrene; MMA = Methylmethacrylate (monomer).

Values expressed in wt% (Wright et al. 2001).

Figure 2-20 Differences in molecular weight of PMMA chains due to alternate

sterilisation methods. Variation in the constituents of different cement preparations

produces differences in the material properties of the polymerised products (Kuehn

et al. 2005).

Figure removed for digital reproduction.

hoshiko
Rectangle
halla
This table is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 69: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

44

cures by radical polymerisation, which begins with the interaction of DMpT activator

from the liquid component and BPO initiator from the powder component. This

initiation reaction between DMpT and BPO produces free radicals which react with the

double bond of MMA, resulting in the polymerisation of MMA monomer and polymer

chains in an exothermic reaction. PMMA polymer chains increase in length by the

addition of monomer at one end only. When the “growing end” of two polymer chains

connect, no addition of monomer is possible and the resultant polymer chain can no

longer increase in size (Kuehn et al. 2005). After the initial polymerisation reaction,

typically less than 5 percent of the monomer remains unpolymerised. Limited

polymerisation may still continue however, with a small detectible reduction in the

residual monomer present in the hardened cement over the first two weeks (Figure 2-21).

Continued polymerisation is partly responsible for variations in material properties of

PMMA cement related to duration of post-cure, particularly within the first 7 days of

preparation (Lee et al. 1973; Wilde et al. 1975) (Figure 2-22).

Figure 2-21 Residual monomer in Surgical Simplex PMMA Cement after initial

polymerisation (Kuhn 2000).

Figure removed for digital reproduction.

Page 70: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

45

Figure 2-22 Increase in shear strength of Surgical Simplex P cement related to

duration of post-cure (Wilde et al. 1975).

Working properties of PMMA cements after mixing of powder and liquid

constituents may be described in four phases (Kuhn 2001; ISO 5833 2002). The

described phases characterising the early behaviour of PMMA cement are:

Phase I Mixing Phase

Time required to mix constituents into homogeneous consistency.

Phase II Waiting Phase

Dough sticky to touch with an unpowdered latex glove.

Phase III Working (Application) Phase

Cement no longer sticky. Suitable for implantation. ISO 5833

“dough time” refers to the onset of working phase.

Phase IV Hardening Phase

Dough can no longer be joined smoothly. Complete hardening

occurs shortly after (latter known as total setting time).

Figure removed for digital reproduction.

Page 71: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

46

Commercial PMMA cement preparations have variable working characteristics,

and are influenced by the ambient temperature and humidity in which the cement is

prepared. The working curves characterising the behaviour of Surgical Simplex P

cement with respect to initial curing time and temperature are illustrated in Figure 2-23.

According to the working characteristics observed during initial preparation,

PMMA cements are classified as high, medium or low viscosity (Kuhn 2001). High

viscosity cements have a short wetting phase and quickly lose their stickiness, entering

the working phase quickly after mixing. The viscosity of “high viscosity” preparations

remains mostly unchanged during the working phase, and slowly increases after entering

the hardening phase. Examples of high viscosity cements include CMW 1, CMW 2,

Palacos R and Palamed (Kuhn 2000). By comparison, low viscosity cements remain

sticky for longer (3 minutes or greater), however when entering the working phase

viscosity rapidly increases, with a narrower transition time between working and

hardening phases. Examples of low viscosity cements include CMW 3, Palacos LV and

Zimmer LVC (Kuhn 2000). Low viscosity cements are generally used in clinical

practice for application by gun under pressure to increase interdigitation into cancellous

bone, thereby increasing the shear strength of the bone cement interface (MacDonald et

al. 1993). Medium viscosity PMMA cements have a low viscous wetting phase and as a

general rule are workable by 3 minutes. During the working phase medium viscosity

cements behave similarly to high viscosity preparations, with viscosity remaining mostly

unchanged or slowly increasing until the hardening phase. Examples of medium

viscosity cements include Antibiotic Simplex, Surgical Simplex P and Zimmer dough-

type radiolucent (Kuhn 2000).

During polymerisation, conversion of MMA monomer molecules to polymeric

form results in decreased intermolecular distance between MMA molecules.

Polymerisation volume reduction is therefore observed with surgical PMMA cement

preparations. Pure MMA has a theoretical volume reduction capacity of 21%. Surgical

cement preparations however contain pre-polymerised PMMA in powder form, which

accounts for approximately two-thirds of the total MMA. The maximum polymerisation

volume reduction of surgical PMMA cement preparations is therefore 6 – 7% (Kuehn et

al. 2005). Multiple factors have been demonstrated to influence the extent and

Page 72: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

47

distribution of polymerisation volume changes within PMMA cements in clinical

practice. Polymerisation volume reduction is significantly limited or even reversed by

the presence of air inclusions and porosity (Charnley 1970; De Wijn et al. 1975; Muller

et al. 2002). Porosity within polymerised PMMA may be classified as macropores (pore

diameter greater than 1.0mm) or micropores (pore diameter 0.1 to 1.0mm diameter).

Reduction in porosity with the use of vacuum mixing increases both static and fatigue

mechanical strength of PMMA cements compared to hand mixing (Lidgren et al. 1984;

Alkire et al. 1987; Wixson et al. 1987; Wang et al. 2001; Kuehn et al. 2005). The

influence of vacuum mixing and porosity reduction on the fatigue resistance in cyclic

compression is illustrated in Figure 2-24. Numerous variables influence the effects of

vacuum mixing on porosity, including cement viscosity and volume, amount of suction

applied, stirrer geometry, mixing speed and method of transfer from mixing apparatus to

gun (Kuehn et al. 2005).

Figure 2-23 Working curves for Surgical Simplex P Cement (Kuhn 2000).

Figure removed for digital reproduction.

Page 73: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

48

Figure 2-24 Cyclic compression fatigue life of PMMA cements are improved with

porosity reduction by vacuum preparation (Kuehn et al. 2005).

Table 2-10 Factors influencing the extent and distribution of porosity within

PMMA cements.

Factors influencing the extent and distribution of porosity within PMMA cements

Air inclusions produced during mixing of PMMA components

Air dissolved within liquid monomer component prior to mixing

Air inclusions created during cement transfer to application device or gun

Vacuum mixing (reduced porosity)

Prosthesis temperature (distribution of porosity effected)

Application of new cement onto a polymerising mass (increased porosity)

Figure removed for digital reproduction.

Page 74: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

49

Factors influencing the extent and distribution of porosity within PMMA cements are

listed in Table 2-10. For hand mixed cement, the total dynamic volume change of the

polymerising mass has been determined as between 3.4% of shrinkage to 3 – 5%

expansion. Vacuum mixing reduces the porosity of the cement and therefore results in

greater polymerisation volume reduction when compared to hand mixed cements,

with up to 6% volume reduction observed after vacuum mixing (Alkire et al. 1987;

Muller et al. 2002). The greater volume reduction in vacuum mixed cement has been

suggested as an explanation for the higher rate of early prosthetic revision with the use

of vacuum mixing observed in the Swedish Hip Registry shortly after this technique

became commercially available (Muller et al. 2002). Latest results from the Swedish Hip

Registry demonstrate the risk ratio for early revision when using vacuum mixed cements

is reduced to 0.74 (Wang et al. 2001).

Porosity distribution and polymerisation volume reduction may not be uniform

within a cement mantle. Preheating the femoral component results in more rapid

polymerisation of the cement in contact with the prosthesis, with reduced porosity

demonstrated at the stem-cement interface and associated improvements in the material

properties of the cement in contact with the prosthesis (Bishop et al. 1996; Parks et al.

1998; Lesaka et al. 2003). This finding somewhat contradicts those of Black et al. (1982)

who demonstrated increased porosity in cements polymerised when in contact with hot

surfaces and considered this to be secondary to trapping of evaporated monomer. In vivo

water absorption into cement also results in a limited delayed volume expansion of the

polymerised mass (Kuehn et al. 2005).

Methods for the characterisation of the material properties of hardened surgical

PMMA cements are defined by ISO 5833 and ASTM standard F 451 – 99a. A useful

reference text comparing the material properties of commercially available PMMA

cements is provided by Kuhn (2000). A summary of typical material properties for

Surgical Simplex P Cement is provided in Table 2-11.

Page 75: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

50

2.4 Properties of PMMA Cement and Bilaminar Mantles in Shear

Section 2.4 describes die-punch testing and other techniques for the investigation of

cements in shear. Previous research relating to the shear properties of PMMA cement

and bilaminar interfaces is presented.

Material failure at the interface between new and old cement mantles represents

a theoretical limitation of C-C revision techniques. Finite element analysis demonstrates

variability in the direction and magnitude of the stress vectors applied to the femoral

mantle related to the type of prosthesis, mechanism of loading and location within the

mantle (Bell 2005). Although no finite element analysis (FEA) or experimental models

are currently available to characterize cement mantle stress after C-C revision, it is likely

that the stress applied to interfacial region between layered cement mantles in vivo is a

combination of shear and compression. As compared to compressive loading, shearing

parallel to the interfacial region is more likely to result in dissociation between layered

cement mantles. Although cyclic fatigue testing in hip simulators would more closely

represent the interfacial loading conditions experienced after C-C revision in vivo, static

loading in pure shear represents the best method of providing reproducible data able to

be clearly interpreted independent of confounding factors prosthesis geometry, femoral

anatomy, mantle thickness, loading regime and anatomical position within the cement

mantle. It is for these reasons that pure shear testing is an appropriate and reproducible

laboratory based model for the investigation of C-C interfacial failure in vivo and has

been utilised by a number of authors (Gruen et al. 1976; Greenwald et al. 1978; Li et al.

1996). In order to interpret the results of C-C interfacial shear strength, an understanding

of the shear properties of the base material must first be understood.

Variable methods for the determination of shear properties of PMMA cement

and similar materials have been described (Greenwald et al. 1978; ASTM B 769 - 94

1994; Farshad et al. 1997; ASTM C 273-00 2000; ASTM D 732-02 2002; Guduru et al.

2005; Lutton et al. 2005). ISO and ASTM standards for the testing of PMMA cement do

not define a uniform method for shear strength evaluation (ASTM F 451 - 99a 1999;

ISO 5833 2002). Greenwald (1978) describes the casting of PMMA cylinders which are

placed within a testing rig which guillotines the material in a transverse direction relative

Page 76: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

51

Table 2-11 Properties of a typical surgical PMMA cement preparation (Greenwald

et al. 1978; Harper et al. 2000; Kuhn 2000; Kuhn et al. 2001; Kuehn et al. 2005).

to the cylinder long axis, allowing shear stress determination as the cylinder cross-

sectional area is known (Figure 2-25). Moran et al. (1978) and Black et al. (1982)

employed similar methods for the shear testing of PMMA cement. More recently

however, Demian et al. (1998) and Kuehn et al. (2005) in review papers on the

regulatory perspectives and characterisation of material properties of orthopaedic bone

cements, cite ASTM D 732 as the appropriate standard for the testing of PMMA in shear.

Characteristics of Surgical Simplex P Cement

Medium viscosity

Barium sulphate radio-opacifier (10%)

Powder contains styrene copolymer (29.4g per 40.0g powder)

Sterilised by γ irradiation

Recommended mixing sequence: powder then liquid

Long working time

Molecular weight less than 350 000 Da

Relatively low impact strength

Material Properties of a Surgical Simplex P PMMA Cement

Ultimate Compressive Strength 80.1 MPa

Bending Strength (ISO 5833) 67.1 MPa

Bending Modulus (ISO 5833) 2643 MPa

Tensile Strength 50.1MPa

Shear Strength 41.2 MPa

Glass Transition Temperature 97oC (dry post-cure 37oC 24hrs)

Glass Transition Temperature 70oC (wet post-cure 37oC 8 weeks)

Impact Strength 3.9 kJ/m2

Page 77: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

52

This is supported by the United States Food and Drug Administration recommendations

for industry guidance into PMMA cement control (Centre for Devices and Radiological

Health 2002). ASTM D 732 – 02 (2002) describes the standard test method for shear

strength testing of plastics by punch tool. The test method describes shear punch testing

(SPT) for the evaluation of shear strength of organic plastics in sheets and moulded discs

between 1.27 and 12.7mm in thickness. Design of the ASTM D 732 – 02 punch tool is

illustrated in Figure 2-26. A similar standardised test method has been described for the

shear testing of aluminium alloys (ASTM B 769 - 94 1994).

Guduru et al. (2005) provides a detailed analysis of factors influencing the

application of punch testing in the evaluation of material shear properties, including

machine – punch compliance, die – punch clearance and specimen thickness. For the

testing of relatively soft materials such as PMMA, Guduru et al.(2005) demonstrated

Figure 2-25 Schematic diagram of the PMMA shear testing apparatus used by

Greenwald et al. (1978). Cast PMMA cylinders with a known cross-sectional area

are subjected to shear between two plates similar to a guillotine.

Figure removed for digital reproduction.

Page 78: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

53

Figure 2-26 ASTM D732 - 02 Shear punch tool

that correction for compliance of the testing apparatus was not required, however the use

of a retaining ring to constrain the test specimens is recommended. When performing

SPT the tolerance between the die and punch is critical in the interpretation of the data

obtained. Deep drawing (forming) occurs when the clearance is excessive, and if

inadequate clearance is provided then true shear conditions are not achieved (Guduru et

al. 2005). Furthermore, the calculation of shear stress from force data obtained within

the laboratory is dependent on the geometry of both punch and die (Young et al. 2002;

Guduru et al. 2005). This relationship is presented in Figure 2-27. Gurudu et al. (2005)

Figure removed for digital reproduction.

Page 79: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

54

Figure 2-27 Variables influencing the calculation of material properties by shear

punch tool (Guduru et al. 2005).

recommends a die punch clearance between 5μm and 20μm as suitable for materials

which demonstrate an ultimate shear stress to failure less than 100MPa.

A summary of the available literature concerning the ultimate shear strength of

PMMA cement is presented in Table 2-12. Greenwald & Wilde (1973) examined the

effect of the addition of the radio-opacifier barium on the shear properties of Surgical

Simplex P. Addition of 10% barium sulphate did not significantly alter the ultimate

shear strength of PMMA. Wilde et al. (1975) determined the ultimate shear stress of

Surgical Simplex P to be 5 762 psi (39.72 MPa). Moran et al. (1979) using a similar

testing technique examined the effect of antibiotic Gentamicin addition on the ultimate

shear stress of Palacos R, with comparison to other readily available PMMA cements.

Increasing concentrations of gentamicin added to the mixture were associated with

statistically significant reductions in the ultimate shear stress of Palacos R. The addition

of 2.0g of gentamicin per 40g cement mix reduced the ultimate shear strength from

41.9Mpa to 37.5Mpa. The ultimate shear stress of gentamicin impregnated Palacos R

greatly exceeded that demonstrated for the bone–cement interface in all situations (<

Figure removed for digital reproduction.

Page 80: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

55

4Mpa), with the authors concluding that failure of cancellous bone is the limiting factor

relevant to shear failure in a joint arthroplasty model (Moran et al. 1979) (Figure 2-28).

The shear strength of PMMA Simplex P cement can be increased by 59% with the

addition of 6.6% w/w graphite fibre without significantly influencing the ultimate

compressive stress, however carbon reinforced PMMA cements are not available for

clinical use (Saha et al. 1979).

Figure 2-28 Shear strength of four orthopaedic PMMA cements with variable

concentrations of added gentamicin (Moran et al. 1979).

Black & Greenwald (1982) examined the weakening of Simplex P when applied

to a cement mantle undergoing concurrent polymerisation. Bilaminar cement mantles

were created by the initial production of a PMMA base, with the sequential application

of new cement at differing time periods while the original PMMA base was still

undergoing polymerisation. The second layer of cement was then machined into

cylinders to enable shear testing in a manner similar to that previously employed by

Wilde & Greenwald, and compared to the material properties of the PMMA base as the

Figure removed for digital reproduction.

Page 81: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

56

Table 2-12 Summary of literature: PMMA ultimate shear strength. Refer to text

for details.

control (Figure 2-29). Reduction in the ultimate shear strength of Simplex P was

demonstrated when it was applied to a polymerising PMMA cement base, with time

periods of 2 minutes, 5 minutes and 7 minutes between mixing the base and

experimental PMMA mantles. No weakening of the secondary cement mantle was

observed however when applied 20 minutes after polymerisation of the PMMA cement

base. This structural weakening correlated with increased porosity observed within the

secondary cement mantle. The authors suggest the application of new cement onto a

Author Cement MPa Comment

Astleford (1975) Simplex – P 24.7 Strain rate 10-3 rad/ sec

Simplex – P 32.0 Strain rate 10-1 rad/ sec

Black (1982) Simplex 44.8 Exptl 20 min; Density 1.055g/ml

Simplex 15.2 Exptl 7 min; Density 0.554 g/ml

Simplex 8.6 Exptl 5 min; Density 0.542 g/ml

Simplex 15.5 Exptl 2 min; Density 0.708 g/ml

Greenwald (1973) Simplex 33.4 Without radio-opacifier

Simplex 35.1 10% Barium Sulphate added

Greenwald (1978) Simplex 41.2 Uniform cylinders

Li (1996) Palacos LV 40.0 Uniform cylinders

Moran (1979) Simplex 43.5 Without antibiotic

CMW 43.0 Without radio-opacifier

CMW 42.5 10% Barium Sulphate

Sulfix 6 42.5

Palacos R 41.9 Without addition of Gentamicin

Saha (1979) Simplex P 30.0

Simplex P 48.0 6.6% w/w 6mm graphite fibres

Wilde (1975) Simplex 39.7

Page 82: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

57

polymerising PMMA base adversely affects the material properties of the secondary

cement mantle because of the exothermic environment to which the secondary mantle is

exposed, creating cement voids due to trapped air or vaporising monomer, therefore

increasing the porosity and reducing the material density (Figure 2-30). When applied to

a PMMA base 20 minutes old, more complete polymerisation of the initial mantle had

occurred with a relative reduction in temperature, and the ultimate shear strength of the

newly applied cement mantle in shear was found to be unaffected (44.9 MPa) (Black et

al. 1982).

Figure 2-29 PMMA cement shear properties determination conducted by Black et

al. (1982). Interfacial shear strength was not assessed with this experiment.

Accurate determination of the shear properties at the interface between new and

old cement requires the development of a testing method and instrumentation with

precision greater than that required for the testing of uniform cement mantles.

Greenwald et al. (1978) created bilaminar cement mantles by first pouring PMMA into

cylindrical moulds. Old cement mantles were post-cured for 7 days at 21oC prior to

reinsertion into the casting mould and subsequent casting of new cement to create

bilaminar specimens (Figure 2-31). The bilaminar cylinders were then placed within a

blunt guillotine shear testing device as previously described (Figure 2-25). This process

of test bilaminar specimen preparation and shear stress determination is presented in

Figure 2-31. As the surface area of contact at the interfacial region was known, the

ultimate shear stress for the interface could be calculated. In addition, the influence of

different methods in surface preparation of the old mantle was investigated, including

the effect of surface roughening and the inclusion of blood at the interface.

Figure removed for digital reproduction.

Page 83: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

58

Figure 2-30 Porosity observed in the experimental PMMA cement mantle applied

onto concurrently polymerising cement. Heat is transferred to the experimental

cement mass during early polymerisation and porosity is induced by the trapping

of vaporised monomer (Black et al. 1982).

Figure 2-31 Method of interfacial shear stress evaluation (Greenwald et al. 1978).

ASTM C 273-00 describes a standardised test method for the determination of

the shear properties of sandwich core materials. The testing apparatus requires the

application of the test material onto opposing rigid mounting plates which are attached

to swivels to adjust the line of force application normal to the interface between the test

materials (Figure 2-32). Limitations of the test method prevent the useful application of

ASTM C 273-00 to the evaluation of bilaminar PMMA cement mantles. Reproducibility

of the test specimen geometry, particularly of the second mantle applied, and the

relatively complex specimen geometry required is not conducive to the preparation of

Figure removed for digital reproduction.

Figure removed for digital reproduction.

Page 84: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

59

multiple standardised test specimens for repeated evaluation. The testing apparatus also

requires the rigid application of the test materials onto a metal backing, which creates

two additional failure regions. The ASTM C 273 - 00 testing method has not been

validated and experimental bias has not been determined (ASTM C 273-00 2000). For

these reasons ASTM C 273 – 00 is unsuitable for the evaluation of interfacial shear

stress of bilaminar PMMA cement mantles.

Figure 2-32 ASTM C 273 – 00 Standard test method for shear properties of

sandwich core materials.

A summary of the available literature regarding the ultimate shear strength of

PMMA cement is presented in Table 2-13 (Gruen et al. 1976; Greenwald et al. 1978; Li

et al. 1996). Greenwald et al. (1978) demonstrated the interfacial strength of bilaminar

Figure removed for digital reproduction.

Page 85: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

60

cement mantles approximated that of uniform mantles when the second mantle was

applied within 3 minutes of initiating mixture, the surfaces were clean and rasping was

undertaken to improve mechanical interlock of the surfaces (Figure 2-33). Weakening of

the C-C interface was demonstrated with the late application of cement, which correlated

with monomer loss by evaporation from the polymerising cement mass (Lee et al. 1973;

Greenwald et al. 1978). The authors concluded that high concentrations of PMMA

monomer within the freshly applied cement dough are associated with improved

adhesion to the pre-polymerised PMMA cement mantle. As monomer loss by

evaporation increases with time to application of the fresh cement, this explained the

observed increase in interfacial shear stress values when cement was applied early

(Figure 2-34, Figure 2-35).

Figure 2-33 Interfacial ultimate shear strength shows marked weakening with

blood present at the interface prior to application of the second cement mantle,

however approaches the strength of a uniform mantle with suitable preparation

(Greenwald et al. 1978).

Figure removed for digital reproduction.

Page 86: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

61

Table 2-13 Summary of literature: PMMA C-C interfacial ultimate shear strength

Author MPa Comment

Greenwald (1978)1 34.8 Clean/ dry interface. New cement applied at 3 min.

38.5 Rasped interface. New cement applied at 3 min.

25.8 Blood at interface. New cement applied at 3 min.

32.3 Clean and dry interface. New cement applied at 4 min.

22.4 Clean and dry interface. New cement applied at 9 min.

Gruen (1976)2 21.9 New cement applied at 2 minutes post mixture.

19.5 New cement applied at 3.5 minutes post mixture.

Li (1996)3 39.0 Clean/ dry interface.

1.2 Blood/ debris at interface.

1 Old cement mantle post-cured for 7 days at 21oC prior to application of fresh

PMMA cement. Bilaminar samples post-cured for 7 days at 21oC prior to testing.

Surgical Simplex P.

2 Old cement mantle post-cured 24 hours prior to fresh PMMA application. Post-

cure temperature not specified. Duration of post-cure of bilaminar specimens not

specified. Surgical Simplex P.

3 Old cement mantle post-cured for 4 days at 37oC prior to application of fresh

PMMA cement. New cement applied when “doughy”. Bilaminar samples post-cured

for 4 days at 37oC prior to testing. Palacos LV.

Page 87: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

62

Figure 2-34 Interfacial ultimate shear stress between PMMA mantles decreases

when the second mantle is applied late. Late application is associated with a

decreased mixture mass due to the evaporation of monomer from the cement

surface prior to polymerisation (Greenwald et al. 1978).

Figure 2-35 Interfacial ultimate shear strength between PMMA cement mantles

decreases in proportion to monomer loss from the second mix of cement applied

(Greenwald et al. 1978).

Figure removed for digital reproduction.

Figure removed for digital reproduction.

Page 88: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

63

Li et al. (1996) found the ultimate shear stress of cleaned bilaminar Palacos LV

cement mantles to be 39 MPa, approximating that of uniform mantles (40MPa). The

inclusion of a thin layer of blood with clots and marrow debris significantly reduced the

ultimate interfacial shear strength (6.2MPa). The authors argue that in clinical practice

the removal of all contaminants from the interface cannot be reliably achieved, and

therefore all cement should be removed.

2.5 Summary of Literature Review and Relevance to Thesis

Section 2.5 discusses the rationale of the selected thesis objectives and hypotheses

presented in Chapter 1 with respect to the available literature.

Cemented THA demonstrates exceptional durability and reliable long term

clinical results in numerous series. Compared to uncemented prostheses, the use of

PMMA cement also provides superior outcomes for patients managed with unipolar

hemiarthroplasty devices for the treatment of femoral neck fractures. Failure of

cemented femoral prostheses in hip joint arthroplasty is predominantly by aseptic

loosening, with symptomatic acetabular wear being an additional significant cause of

failure of hip joint hemiarthroplasty. Revision of cemented hip joint arthroplasty

traditionally involves the extraction of both loose and well fixed PMMA cement. The

removal of secure cement is associated with significant intraoperative complications and

may also jeopardise the stability of the revision implant due to the associated extraction

of cancellous bone attached to the primary cement mantle. In the presence of a secure

cement-bone interface, C-C femoral revision by retention of well fixed cement has been

demonstrated to reduce the complications associated with preparation of the femoral

canal. The suitability of the bone-cement interface for the conduct of C-C revision can

be reliably predicted preoperatively by careful evaluation of plain radiographs and the

adjunctive use of arthrography in selected cases. Limited clinical studies of short to

medium term duration support the technique of C-C revision, and this has lead to the

design of prostheses specific to this application such as the Exeter Short Revision Stem

and the Exeter Trauma Stem.

Page 89: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

64

Static loading in pure shear represents an appropriate method of providing

reproducible data on C-C interfacial properties which is able to be interpreted

independent of confounding factors such as prosthesis geometry, femoral anatomy,

mantle thickness, loading regime and position related stress variations within the cement

mantle. It is for these reasons that static shear testing has been utilised by a number of

authors for the investigation of C-C interfacial failure in vitro. High interfacial shear

strengths between layered PMMA cements have been demonstrated, however critical to

the technique is adequate preparation of the old cement mantle by provision of a clean,

dry surface which is free from debris such as blood and fat. Mechanical roughening of

the old cement mantle has been demonstrated to result in a small increase in the

interfacial shear strength between layered PMMA cements. C-C adhesion is enhanced

by the application of fresh cement dough with a high monomer concentration shortly

after mixture of powder and liquid components.

Description of the material properties between layered cement mantles remains

incomplete. To date the literature characterising the shear properties of C-C interfaces

has been limited to old cement mantles with a post-cure duration of 7 days or less prior

to the application of fresh cement. These specimens still contain relatively high

concentrations of unpolymerised monomer which may potentially enhance interfacial

adhesion with the freshly applied PMMA cement dough. Furthermore, PMMA cements

demonstrate variability in material properties with respect to duration of post-cure

predominantly within the first 14 days of initial polymerisation. In clinical practice

however aseptic loosening of prosthetic components is a process which takes many

years to occur, and typically the primary cement mantle has post-cured in situ for many

years prior to revision. It is of clinical relevance therefore to characterise the interfacial

adhesion of fresh PMMA cement to an old cement mantle with material properties

similar to that seen in the retained primary cement mantle during C-C revision. For

experimentation of this nature old cement mantles should be post-cured for a minimum

of 14 days. Furthermore, the literature to date has limited the testing of C-C interfaces to

bilaminar specimens which have been post-cured for 7 days or less after the application

of fresh cement. As the monomer content and material properties within the new cement

mantle are expected to change beyond this time period, what effect will this have on the

Page 90: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

65

C-C adhesion? Will the shear properties between layered PMMA cements be altered

with maturation of the freshly applied cement? Does the interfacial region display

variations in shear strength behaviour which could impact on recommendations for

postoperative rehabilitation? Characterisation of the material properties and bilaminar

PMMA cement mantles in shear with respect to duration of post-cure is a clinically

important subject which remains largely unanswered.

Cemented revision hip arthroplasty in modern practice is usually conducted with

the use of antibiotic PMMA cement preparations. The literature to date has not yet

assessed the influence of the commercial addition of antibiotics on cement- cement

adhesion. Finally, a description of the mechanisms of adhesion between layered PMMA

cements has not been provided in the literature. These issues form the final subjects for

evaluation within this thesis.

Page 91: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

66

Page 92: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

67

Chapter 3

Experimental Evaluation of the Shear Properties

of PMMA cements

3.1 Introduction

Chapter 3 describes the experimental evaluation of the shear properties of

PMMA cement, with specific reference to the variation in shear properties with respect

to duration of post-cure. These experiments provide data to validate the testing method

and to supply reference data for the interpretation of results obtained for bilaminar

PMMA cement test samples.

3.2 Hypotheses

1. The shear properties of Surgical Simplex and Tobramycin Antibiotic Simplex

PMMA cements (Stryker, Rutherford, NJ) have variability with respect to

duration of post-cure.

2. Antibiotic Simplex (Tobramycin) PMMA cement demonstrates equivalent shear

properties to Surgical Simplex P.

3.3 Method

3.3.1 Equipment design considerations and objectives

Objectives of the equipment design for this experimentation are to:

1. Enable testing of PMMA cements in shear;

2. Enable the accurate manufacture of test specimens with reproducible geometry;

3. Facilitate the accurate production of multiple specimens;

4. Minimise the wastage of consumables;

Page 93: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

68

5. Enable future testing of the interfacial shear properties between bilaminar cement

mantles;

6. Allow modification of test specimen characteristics in accordance with

experimental objectives (cement type, duration of post-cure); and

7. Enable future testing of the shear fatigue properties PMMA cement with minimal

or no modification to testing apparatus.

3.3.2 Overview of experimental equipment and outline of applications

Design of the shear punch testing (SPT) device developed for this research is

illustrated in Appendix 1 and Figure 3-1. The SPT device has two modes of application,

vis. “Casting” and “Testing.” Casting mode refers to utilisation of the device as a mould,

into which PMMA is poured after mixing of liquid and powder constituents shortly after

mixing the cement constituents. This casting process enables the production of PMMA

cylinders from which individual test specimens are manufactured. The process of test

specimen manufacture is discussed further in Section 3.3.3.

Figure 3-1 SPT demonstrating casting and testing modes of application. The

central casting rod is removed for the manufacture of solid PMMA cast cylinders.

Figure removed for digital reproduction.

Page 94: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

69

“Testing mode” refers to the utilisation of the SPT device as a die-punch for the

application of shear force onto individual test specimens. Opposite faces of the base

plate and top plates have design features specific to the application of either testing or

casting mode (Appendix 1). For testing mode the surfaces of both top plate and base

plate display 37.9mm diameter recesses which correspond to the outer geometry of the

test specimens. These recesses centrally locate the specimens relative to the zone of

punch. Two 3mm diameter stainless steel locating pins align the base and top plates so

that the centre holes through which the punch passes during testing are aligned. The

process of specimen testing is discussed further in Section 3.3.4.

3.3.3 Test specimen manufacture

Test specimens were manufactured by initially casting PMMA cement cylinders

with a standardised geometry from which individual test specimens were subsequently

produced. Cement preparation was undertaken at 1 atmosphere with a room temperature

of 22 ± 1oC. PMMA cement powder and liquid were mixed in an open bowl by hand

with proportions according to manufacturer recommendations. Cement was mixed at

frequency of 1Hz for 60 seconds, and immediately transferred into the casting mould.

Early pouring of the cement into mould ensured suitably low enough viscosity to enable

the cement to completely adopt the desired geometry prior to polymerisation. The

casting mould and cement were then transferred to a warming cabinet, maintaining

temperature at 37.0 ± 0.5oC. After the cement had hardened, the cylinder was extracted

from the casting mould and stored in a dry environment at 37.0 ± 0.5oC. The

temperature of the casting mould prior to application of cement was measured as 22 ±

1oC using a QM7222 Digitech Infrared Thermometer.

Individual test specimens were manufactured from the cast cylinders by parting

and facing the material on lathes. Heating of PMMA during parting and facing

procedures, particularly above the glass transition temperature (70oC), was considered

undesirable as this would potentially alter the geometry and material properties of the

test specimens. A method of individual test specimen production was therefore

Page 95: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

70

developed to minimise heating of the PMMA cylinders during specimen manufacture.

This technique included the selection of suitable cutting tools with appropriate cross

feed rate and lathe speed. Parting of individual specimens was undertaken using a lathe

fitted with four jaw self centring chuck and 3mm Iscar double ended parting tip (product

number: DGN 3102J; IC 354; P20-P40) (Figure 3-2). Parting was undertaken at 650rpm

with a cross feed rate of 300mm per minute. Temperatures at the cutting face during the

parting of specimens were measured perpendicular to the test specimen using a QM7222

Digitech Infrared Thermometer typically ranged from 35.0 – 37.0 oC. The maximum

temperature recorded during the parting of any specimen was 42.1 oC. Inspection of the

PMMA swarf did not show evidence of material melting. Facing was then conducted on

both sides of every specimen using a lathe fitted with a three jaw high precision self

centring chuck and linear scale positioning system with precision of 0.01mm (Figure

3-3). The cutting tool employed for facing was a Seco ground and polished tip usually

used for aluminium product fabrication (product number: TPUR110304R-95, HX).

Facing was undertaken at 1100rpm with a cross feed rate of 300mm per minute.

Temperatures recorded during the facing of specimens were measured by QM7222

Digitech Infrared Thermometer, and typically ranged from 25.0 – 26.5 oC. The

maximum temperature recorded during the facing of any specimen was 27.0 oC, and

swarf inspection did not reveal evidence of PMMA melting during facing procedures.

Parting and facing of the cylinder on lathes resulted in the creation of flat cement discs

of 5mm thickness. The thickness of the discs was confirmed with the use of a digital

verniers calliper with a precision of ± 0.01mm (Figure 3-4). The outer surface of the

cylinder was not reduced during the process of turning, and therefore the cement discs

created retained the same radial geometry as when removed from the original casting

mould. In this manner it was possible to produce a large number of test specimens with

highly reproducible geometry. A small number of test specimens were found to have

macroscopic porosity, defined as a cement mantle surface defect greater than 2mm in

diameter approximating the region of shear testing, and these were discarded due to

concerns regarding potential influence on the shear properties. Test specimens were

stored at 37oC in a dry environment (Figure 3-5).

Page 96: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

71

Figure 3-2 Parting of PMMA discs from solid cylinder. With the use of suitable

lathe tip allowing adequate swarf clearance, temperature of the cutting tool

typically measured at 35.0 – 37.0oC, well below the glass transition temperature of

polymerised PMMA (see text).

Figure 3-3 Facing of the test specimens was performed on both sides. With the use

of suitable lathe tip allowing adequate swarf clearance, maximum temperature of

the cutting tool typically measured 25.0-26.5oC, well below the glass transition

temperature of polymerised PMMA (see text).

Page 97: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

72

Figure 3-4 Confirmatory measurement of specimen thickness using digital verniers

calliper with precision of 10μm.

Figure 3-5 Dry post-curing of PMMA specimens at 37oC.

Page 98: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

73

3.3.4 Evaluation of PMMA shear properties

Determination of the ultimate shear strength of individual specimens was

undertaken using the experimental apparatus described above in “testing” mode (Figure

3-6, Appendix 1). PMMA test discs were placed into the base plate and top plate

37.9mm retaining rings, corresponding to the outer geometry of the test specimens. This

enabled the test specimens to be centrally located relative to the zone of punch testing

within the SPT apparatus. The 37.9mm test specimen recesses on top and base plates

were each 1.5mm in depth, so that when a 5mm thickness specimen was placed into the

device a 2mm gap between the plates is created (Figure 3-7). This enabled a

compressive preload to be to the outer section of the disc with the use of four lag bolts

passing through gliding holes drilled in the top plate and screwing into the base plate.

Deformation (flexing) of the test specimens during the application of shear force by the

punch device was therefore minimised.

Figure 3-6 Shear punch testing using Hounsfield unit.

Page 99: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

74

The central SPT punch was advanced using a Hounsfield H25K-S UTM

(Hounsfield Test Equipment, Salfords Redhill, UK) with cross head speed of 1mm per

minute until the ultimate shear strength of the PMMA disc was exceeded. The load cell

employed had a maximum load range of 25 000N, and typically failure of the 5mm

thickness PMMA test specimens occurred at 40% – 60% of this range. Calibration

details of the load cell employed are presented in Appendix 2. The radial tolerance

between SPT punch was measured to be 20 μm with the use of internal and external

micrometers. Shear testing was conducted at 37.0 ± 2.0oC.

Figure 3-7 Sectioned isometric view of SPT device in testing mode. 2mm gap

between top plate and base plate are demonstrated.

3.3.5 Testing schedule and data analysis

Testing was conducted on uniform PMMA cement mantle specimens using

Antibiotic Simplex (Tobramycin) and Surgical Simplex P (Stryker Inc., Rutherford NJ)

cements. Specimens were post-cured prior to SPT evaluation according to the schedule

presented in Table 3-1.

Figure removed for digital reproduction.

Page 100: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

75

Cement Type Age (Days) Number Specimens

Surgical Simplex P 0 (1 Hour) 9Surgical Simplex P 1 9Surgical Simplex P 7 8Surgical Simplex P 14 14Surgical Simplex P 30 9Surgical Simplex P 90 13

Antibiotic Simplex 0 (1 Hour) 8Antibiotic Simplex 1 8Antibiotic Simplex 7 8Antibiotic Simplex 14 15Antibiotic Simplex 30 9Antibiotic Simplex 90 5

Table 3-1 Testing schedule for determination of PMMA cement shear properties.

Force required to exceed the ultimate shear strength of the test specimens was

recorded using QMat software (Hounsfield Test Equipment, Salfords Redhill, UK) and

exported into Microsoft Excel (Version 2003; Microsoft Corporation) for further

analysis. The ultimate shear strength of the PMMA material was determined using

Equations 3.1 to 3.3. Equation 3.1 demonstrates the calculation of test specimen ultimate

shear strength, where A is the surface area of the test specimen subjected to shearing

force within the SPT device. Equation 3.2 calculates the test specimen shear area (A),

where Rave is the average radius of the SPT die (R1) and punch (R2), and T is the

thickness of the test specimen (5.00mm). Equation 3.3 demonstrates the calculation of

Rave using R1 and R2 values demonstrated in Figure 3-8.

Calculation of the force/ displacement gradient within the elastic deformation

region of the test specimens was determined using a software specific to this application

written using Matlab 7.0.4 (The MathWorks, Massachusetts, US). Determination of the

shear modulus using the force/ displacement gradient is discussed in Section 3.3.6 and

Section 3.5.

Page 101: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

76

Statistical analysis was conducted using Minitab 14 and SPSS for Windows 11.5

statistical software. Parametric and non parametric methods were utilised depending on

the data set and are specified with presentation of results in Section 3.4.

ultFA

τ =

2 aveA R Tπ=

1 2

2ave

R RR +=

Figure 3-8 Illustration of R1 and R2 measurements used in Equation 3.3, where R1

is the radius of the SPT die and R2 is the punch radius.

Equation 3.1

Equation 3.2

Equation 3.3

Page 102: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

77

3.3.6 Testing method validation and shear modulus calibration

Testing of 20 pcf (0.32 g/cc)and 40 pcf (0.64 g/cc) solid rigid polyurethane foam

specimens (Sawbones, Pacific Research Laboratories Inc, Washington USA) with

known shear properties was conducted at 22oC in order to validate the experimental

technique and to calibrate the SPT device to allow the evaluation of shear modulus. The

individual foam specimens were manufactured from solid block material using the

turning and facing methods described in Section 3.3.3, resulting in specimens of

identical geometry to the PMMA test specimens (37.9mm diameter x 5mm thickness

discs). Foam reference specimens were tested using the SPT device as described in

Section 3.3.4. The ratio of the experimental force/ displacement gradient to the known

shear modulus for the sawbones specimens was calculated in order to provide a

calibration ratio (x) for the calculation of shear modulus of the PMMA test specimens

(Equation 3.4 and Equation 3.5). Equation 3.4 demonstrates calculation of the shear

modulus calibration ratio (x), where ΔF/ΔD is the gradient of the force/ displacement

plot and G is the shear modulus. Equation 3.5 calculates the shear modulus of test

specimens using calibration ratio (x) obtained in Equation 3.4. Rationale for this method

of test specimen shear modulus calculation is further discussed in Section 3.5.

foam

foam foam

FxD GΔ

specimenspecimen

specimen

x FGD

=ΔΔ

Equation 3.5

Equation 3.4

Page 103: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

78

3.4 Results

3.4.1 Sawbones reference and shear modulus calibration

Comparison of expected and experimental shear properties of 20 pcf and 40 pcf

Sawbones solid rigid polyurethane foam are presented in Table 3-2. Ultimate shear

strength of the sawbones material is quoted as having ±10% variability according to the

manufacturer (James 2005). Experimental shear strength measurements obtained for the

reference sawbones material closely correlated with the expected values. The shear

modulus calibration ratios (x) calculated for both 20 pcf and 40 pcf foams were similar,

demonstrating the reproducibility of the method of shear properties evaluation. The

shear modulus calibration ratio obtained for 40 pcf foam was selected as the reference

standard for calculation of shear modulus of the PMMA test specimens as 40 pcf foam

had a higher shear modulus which more closely represented that of PMMA and the

force/ displacement gradient measurements obtained for 40 pcf foam were more

reproducible as demonstrated by a lower percentage variation between samples.

20 pcf Foam 40 pcf Foam

Expected shear strength (MPa)* 4.50 14.00Experimental shear strength (MPa) 5.33 17.84Shear strength 95% C.I. (MPa) 4.91 - 5.37 17.52 - 18.16

Expected shear modulus (MPa)* 67.00 187.00Experimental force displacement (N/mm) 2503.55 7992.42Force/ displacement 95% C.I. (N/mm) 2294.76 - 2712.34 7845.01 - 8139.83Shear modulus calibration ratio (X) 0.02676 0.0234

Table 3-2 Results of Sawbones reference sample testing. Expected shear strength

and modulus values obtained from manufacturer. * Values +/- 10%. (James 2005;

Pacific Research Laboratories 2005).

Page 104: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

79

3.4.2 Uniform PMMA test specimen shear properties

The ultimate shear strength of Antibiotic Simplex and Surgical Simplex P

cements were comparable (p=0.13; two-way ANOVA), and both variations in ultimate

shear strength with respect to duration of post-cure (p<0.001) (Figure 3-9, Table 3-3). A

quantitatively small reduction in ultimate shear strength of approximately 5 MPa was

observed for both cements at days 7 (p<0.05; LSD post hoc analysis) and 14 (p<0.001),

which recovered by day 30 to initial values (Figure 3-9, Table 3-3). Variability in the

force displacement gradient and derived shear modulus was observed to be less than five

percent for both cement types over the 90 day time period (Figure 3-10, Table 3-3). The

average shear modulus of Surgical Simplex over the 90 day time period was 442 MPa.

Antibiotic Simplex cement did not demonstrate a statistically significant difference in

the shear modulus compared to Surgical Simplex (446 MPa; p=0.10 Mann Whitney U

test).

Ultimate Shear Strength.

0

10

20

30

40

50

60

0 1 7 14 30 90Days

MPa

Antibiotic Simplex Simplex

Figure 3-9 Ultimate shear stress to failure of Antibiotic Simplex and Surgical

Simplex P cements with 95% confidence intervals marked. Similar variations in

shear properties related to duration of post-cure are observed for both cement

preparations.

Page 105: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

80

PMMA Shear Modulus.

0

100

200

300

400

500

0 1 7 14 30 90Days

MPa

Antibiotic Simplex Simplex

Figure 3-10 Calculated shear modulus of Antibiotic Simplex and Surgical Simplex

P cements with 95% confidence intervals marked.

Cement Type Age (Days) Strength (Mpa) 95% CI Modulus (Mpa) 95% CI

Ab Simplex 0 50.51 1.86 444.30 11.33Ab Simplex 1 49.77 1.80 444.93 7.97Ab Simplex 7 48.75 1.58 461.66 10.00Ab Simplex 14 44.98 1.27 448.58 8.92Ab Simplex 30 48.83 1.73 459.10 6.60Ab Simplex 90 46.62 1.69 424.12 9.26

Simplex 0 47.45 0.93 438.96 7.13Simplex 1 49.29 0.77 442.85 6.42Simplex 7 44.85 1.53 435.13 12.67Simplex 14 43.00 1.40 414.90 9.36Simplex 30 47.09 1.99 445.08 7.14Simplex 90 50.59 1.01 451.50 7.48

Table 3-3 Tabulated data from Figures 3-9 and 3-10.

Page 106: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

81

3.5 Discussion

This experiment demonstrates the application of punch testing for the materials

evaluation of uniform cement mantles in shear. The method of shear properties

evaluation described within this research is modelled on ASTM D 732 and observes

recommendations made by Guduru et al. (2005). Determination of the ultimate shear

stress and modulus of PMMA cements using the described testing technique provides

highly reproducible results which are comparable to that previously described in the

literature.

Shear modulus was approximated by indirect methods using materials of known

shear modulus for a calibration standard as this was considered the most accurate

method of shear modulus calculation using data obtained from punch testing. Although

direct calculation of shear modulus is possible by measuring the punch displacement and

die punch clearance (Figure 3-11), this would provide only a gross approximation at best

as this technique is limited by the accuracy of measurements taken to determine the

clearance of the SPT device. Radial clearance of the SPT device is estimated to be 20μm

by subtraction of the punch and die radii obtained by the use of internal and external

micrometers. The concern with using this measurement for the direct calculation of

shear modulus relates to the accuracy of the clearance measurement and precision of the

micrometers used. Quantitatively small errors in the measurement of either the punch or

die would result in proportionally large errors in the calculated shear modulus due to the

small clearance margin of the SPT device. By using a reference material with known

material properties to correlate force/ displacement data to shear modulus, precise

calculation of the die-punch clearance is not required. This indirect method of shear

modulus determination is therefore preferable when using a die-punch apparatus with

small radial tolerances. Validation of the precision of shear modulus calculation by this

indirect method is provided by the high correlation between shear modulus calibration

ratios obtained for both 20 pcf and 40 pcf reference specimens.

This experiment demonstrates variability in the ultimate shear strength of

Surgical Simplex P and Antibiotic Simplex PMMA cements with duration of post-cure.

Although the variability in shear strength observed was quantitatively small, due to the

precision of the testing method employed, changes in shear strength related to duration

Page 107: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

82

Figure 3-11 Above: Schematic representation of SPT device in testing mode. R1 =

radius of die. R2 = radius of punch. Below: Magnification of the zone of shearing.

The portion of material undergoing shear is represented in light blue. Die-Punch

clearance was estimated to be 20μm by use of micrometers, however inaccuracy in

this clearance measurement would potentially create large errors in the assessment

of shear modulus by direct calculation. Indirect methods for shear modulus

determination were therefore employed.

Page 108: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

83

of post-cure were found to be statistically significant. As relative weakening of the

PMMA cement preparations was only observed at the 7 and 14 day time intervals, and

quantitatively the reduction was small (approximately 5MPa or 10% of initial values), it

is unlikely that variations in shear strength of PMMA cements related to duration of

post-cure are clinically relevant. The characterisation of the shear properties of uniform

PMMA cement mantles with respect to duration of post-cure serves as a useful reference

for investigation of the interfacial shear strength of bilaminar cement mantles.

3.6 Conclusions

Surgical Simplex P and Antibiotic Simplex (Tobramycin) PMMA cements

demonstrate variability in ultimate shear strength with respect to duration of post cure,

however the magnitude of these variations are small and are unlikely to be of clinical

relevance. The manufacturer’s addition of Tobramycin antibiotic does not significantly

alter the shear properties of Simplex PMMA cement. The developed method of material

properties evaluation using a SPT apparatus modelled on ASTM D732 provides an

accurate and reproducible means of calculation of ultimate shear strength and shear

modulus estimation for surgical PMMA cements.

Page 109: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

84

Page 110: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

85

Chapter 4

Experimental Evaluation of the Shear Properties

of Bilaminar Cement Mantles

4.1 Introduction

Chapter 4 describes experimental investigations relating to the ultimate shear

strength of cement – cement interfaces. Three main objectives were selected for this

experimentation based on issues relevant to the clinical practice of cement – cement

revision arthroplasty:

1. How does the shear strength of bilaminar cement mantles compare to uniform

mantles?

2. Does the interfacial region display time dependant variations in shear strength

behaviour which could influence recommendations for postoperative weight

bearing?

3. Given that antibiotic loaded cements are preferentially used in revision hip

arthroplasty procedures, does the use of antibiotic PMMA cement in primary or

revision hip arthroplasty result in weakening of the C-C interfacial region?

4.2 Hypotheses

1. The ultimate shear strength of the interfacial region between bilaminar cement

mantles approximates that of uniform mantles.

2. Bilaminar PMMA cement mantles demonstrate variation in the interfacial shear

properties with duration of post-cure.

3. Antibiotic Simplex when used in bilaminar cement mantles results in weakening

at the interfacial region.

4. Interfacial adhesion by mechanisms other than mechanical interlock play a

significant role in the bond formed between new and old cement mantles.

Page 111: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

86

4.3 Method

4.3.1 Bilaminar test specimen manufacture and testing schedule

Bilaminar PMMA cylinders containing layered cement were manufactured by

inserting a smooth 316L stainless steel casting rod into the centre of the casting mould

prior to the pouring of cement. The casting rod had a diameter of 19.1mm and surface

roughness approximating that of a polished femoral stem used in primary hip

arthroplasty (<0.2 μm Ra). Both ends of the outer casting mould and central casting rod

were located and constrained by recesses in the casting mould base plate and top plate.

In this way the spatial location of the casting rod was accurately and reproducibly

centralised throughout the long axis casting cylinder. Cement constituents for the outer

PMMA mantle were mixed in an open bowl by hand at 1Hz for 60 seconds at 1ATM

and immediately poured into the casting mould including the centrally located casting

rod. The temperature of the casting mould and rod prior to application of cement was

measured as 22 ± 1oC using a QM7222 Digitech Infrared Thermometer. Early pouring

of the cement into the mould ensured suitably low viscosity and enabled the cement to

completely adopt the desired geometry prior to polymerisation. After the cement had

cured for a minimum duration of 30 minutes, the casting rod was extracted from the

PMMA cylinder using a Hounsfield testing device with a cross head speed of 500mm

per minute. This process enabled the production of cast PMMA cylinders with a central

hole of diameter and central location corresponding to the casting rod (Figure 4-1).

PMMA cylinders were removed from the casting mould and stored in a dry environment

at 37.0 ± 0.5oC for a variable duration according to the testing schedule presented in

Table 4-1. Surface roughness of the old cement mantle interfacial region after casting

rod extraction was measured in representative samples using a surface profilometer

(Taylor Hobson Surtronic 3+; Leicester England). Light microscopy (LM) and scanning

electron microscopy (SEM) were used to visually assess the interfacial surface of the

outer cement mantle in representative samples.

After appropriate post-curing of the outer cement mantle (Table 4-1), the fresh

PMMA cement was prepared as previously described and poured into the central portion

Page 112: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

87

Figure 4-1 Simplex PMMA cement cylinder with central hole after removal of

casting rod. After post-curing of the PMMA cylinder, fresh PMMA cement was

poured into the central hole, therefore creating a solid cement cylinder with outer

“old” cement and inner “new” cement.

of the hollow cement cylinder, creating a bilaminar PMMA cylinder with the freshly

applied cement surrounded by an old cement mantle. Pouring of the second PMMA

mantle was undertaken 90 seconds after initiating mixture of the liquid and powder

constituents, as this was judged to be representative of the earliest time in which cement

could be applied in a clinical setting during C-C revision. The bilaminar PMMA

cylinders were then stored in a dry environment at 37.0 ± 0.5oC for post-cure of variable

duration in accordance with the testing schedule (Table 4-1).

Individual test specimens were manufactured from the cast cylinders by parting

and facing the material on lathes as described in Section 3.3.3. This resulted in

individual bilaminar test specimens of 5mm thickness (Figure 3-4, Figure 4-2). LM and

SEM examination of the bilaminar PMMA test specimens was undertaken in

representative samples to evaluate the interfacial region between the new and old cement

mantles.

Page 113: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

88

Group PMMA Cement (Old) Age (Days) α PMMA Cement (New) Age (Days) β Number Specimens

1 Surgical Simplex P 14 Surgical Simplex P 0 (1 Hour) 162 Surgical Simplex P 14 Surgical Simplex P 1 173 Surgical Simplex P 14 Surgical Simplex P 7 124 Surgical Simplex P 14 Surgical Simplex P 90 145 Surgical Simplex P 90 Surgical Simplex P 1 156 Antibiotic Simplex 14 Antibiotic Simplex 1 18

α Age of outer (old) cement mantle prior to new cement applicationβ Age of bilaminar specimens prior to shear testing

Table 4-1 Testing schedule of bilaminar test specimens for the determination of

PMMA interfacial shear properties.

Figure 4-2 Completed 5mm thickness bilaminar PMMA test specimen. For

illustration purposes, coloured cement is used to differentiate the outer (old)

cement mantle from the inner (new) cement mantle.

Page 114: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

89

4.3.2 Bilaminar test specimen evaluation of shear properties

Determination of the ultimate shear strength of individual specimens was

undertaken using the SPT apparatus as described in Chapter 3 in “testing” mode.

For LM and SEM examination of the fracture surface in two planes, sectioning of

the bilaminar discs was conducted using a water irrigated microtome (Microslice 2,

Malvern Instruments Ltd, Worcestershire UK) fitted with a 280 grit diamond annular

blade (Van Moppes Super Abrasives; Gloucestershire UK). Detection of mantle cracks

under LM was assisted with the use of fluorescent dye penetrant (Ardrox 970P25;

SureChem, Girraween Australia) and an ultraviolet light source (Magnaflux ZB24A220/

240; Swindon Wiltshire UK).

Shear testing of PMMA cement against polished 316L stainless steel with an

equivalent surface roughness to the interfacial region of the outer cement mantle

(0.22μm Ra) was conducted in order to provide a reference for further assessment of the

mechanisms of C-C adhesion. For this test a steel mould with a 19.1mm central hole was

filled with PMMA cement and post-cured for 24 hours. Cement preparation and shear

testing at the cement-metal interface was conducted as described previously. This test

was conducted 5 times utilising the same metal mould.

4.4 Results

4.4.1 Surface characteristics & fracture evaluation by LM and SEM

Roughness of the interfacial surface of the outer (old) cement mantle after

casting rod extraction prior to the application of fresh cement was 0.22 μm Ra (95% CI

0.20 – 0.24 μm Ra). LM and SEM of the interfacial surface of the old cement mantle

prior to the application of new cement showed a smooth surface with minimal porosity

(Figure 4-3 and Figure 4-4).

Examination of the bilaminar PMMA specimens using LM and SEM prior to

punch testing was unable to detect a crack or easily identifiable interface plane between

the new and old cement mantles. After punch testing both LM and SEM revealed clean

separation between the new and old cement mantles with a regular linear fracture at the

interfacial region, and is illustrated in Figure 4-5 to Figure 4-9.

Page 115: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

90

Figure 4-3 SEM of the interfacial surface of the old cement mantle prior to the

application of new cement. Smooth surface with minimal porosity is demonstrated.

Surface roughness of this interface measured by profilometry was 0.22μm Ra.

(Original magnification x27).

Figure 4-4 SEM of the interfacial surface of the old cement mantle prior to the

application of new cement. At high power magnification the surface profile of the

cement secondary to defects on the casting rod and irregularities created during

the process of rod extraction are appreciated on careful inspection (Original

magnification x1500).

Page 116: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

91

Figure 4-5 LM of interfacial region between inner PMMA and outer PMMA

cement mantles after SPT testing. Clean separation between the cement mantles is

observed (Original magnification x10).

Figure 4-6 SEM of interfacial region between Surgical Simplex P cements after

punch testing. Brittle fracture at the interface is demonstrated (Original

magnification x80).

Figure removed for digital reproduction.

Page 117: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

92

Figure 4-7 Transverse section SEM of a bilaminar Simplex – Simplex test specimen

at the interfacial region after SPT evaluation. Linear separation between new and

old cement mantles is demonstrated. Outer (old) cement mantle is located upper

left and inner (new) cement mantle is located lower right. (Original magnification

x110).

Outer Mantle (old cement)

Inner Mantle (new cement)

Figure removed for digital reproduction.

Page 118: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

93

Figure 4-8 LM florescent dye penetrant enhanced transverse section examination

of a bilaminar PMMA test specimen after SPT testing demonstrating linear cement

fracture through the interfacial region.

Figure 4-9 LM florescent dye penetrant enhanced examination plan view of a

bilaminar test specimen after SPT testing demonstrating cement fracture through

the interfacial region.

Page 119: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

94

4.4.2 Bilaminar PMMA test specimen shear properties

Bilaminar cement mantles demonstrated a 15 to 20 percent decrease in shear

strength over all time periods evaluated when compared to uniform PMMA cement

mantles (p<0.001; two-way ANOVA) (Figure 4-10, Table 4-2). The ultimate shear

strength of bilaminar cement specimens was observed to temporarily decrease by 8.2

MPa (19.4%) seven days after initial cure (p<0.001; two-way ANOVA). Specimens

with an outer cement mantle aged for 90 days prior to the application of fresh cement

had equivalent shear properties to specimens with an outer mantle aged for only 14 days

(ultimate shear strength 43.75 MPa; 95% CI 41.47 - 46.03 MPa; p=0.17). Variability in

shear modulus

Simplex Ultimate Shear Strength. Uniform v Bilaminar Mantles

0

10

20

30

40

50

60

0 1 7 90Days

MPa

.

Uniform Mantle Cement - Cement

Figure 4-10 Ultimate shear strength of Surgical Simplex P compared to bilaminar

test specimens manufactured from the same material with 95% confidence

intervals marked. Outer mantle of the bilaminar test specimens was post-cured for

14 days prior to application of the new mantle and tested 0 – 90 days after new

mantle application. Bilaminar PMMA mantles demonstrate time dependant shear

properties similar to that of uniform mantles, however a comparative reduction in

ultimate strength by 15 to 20 percent was observed.

Page 120: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

95

of bilaminar Simplex test specimens was observed to be less than five percent over the

90 day time period (Figure 4-11). The average shear modulus of bilaminar surgical

simplex was 443.6 MPa (95% CI 438.9 – 448.3 MPa) over the time period. No

difference in the shear modulus was observed between bilaminar and uniform cement

mantles (p>0.1; two-way ANOVA).

Bilaminar PMMA test specimens manufactured using Antibiotic Simplex for

both cement mantles demonstrated equivalent ultimate shear strength (p=0.52; Kruskall-

Wallis test) and modulus to specimens using Surgical Simplex for both cement mantles

(Figure 4-12, Figure 4-13, Table 4-2).

The ultimate shear strength of Surgical Simplex P cement against 316L stainless

steel (0.22μm Ra) was 1.2 MPa (95% CI 0.6 – 1.8 MPa).

Simplex Shear Modulus. Uniform v Bilaminar Mantles.

0

50

100

150

200

250

300

350

400

450

500

0 1 7 90Days

MPa

.

Uniform Mantle Cement - Cement

Figure 4-11 Calculated shear modulus for uniform mantle and bilaminar Surgical

Simplex P PMMA samples. 95% confidence intervals marked.

Page 121: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

96

Interfacial Shear Strength. Plain v Antibiotic Cements.

0

10

20

30

40

50

60

Simplex - Simplex Ab Simplex - Ab Simplex

Cement Type

MPa

.

Figure 4-12 Interfacial shear strength bilaminar test specimens comparing cement

preparations with and without the commercial inclusion of Tobramicin antibiotic,

95% confidence intervals marked.

Interfacial Shear Modulus. Plain v Antibiotic Cements.

0

100

200

300

400

500

Simplex - Simplex Ab Simplex - Ab Simplex

Cement Type

MPa

.

Figure 4-13 Interfacial shear modulus of bilaminar test specimens comparing

cement preparations with and without the commercial inclusion of Tobramicin

antibiotic. 95% confidence intervals marked.

Page 122: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

97

Group Outer Age (Days) Inner Age (Days) Strength (Mpa) 95% CI Modulus (Mpa) 95% CI

1 Simplex 14 Simplex 0 36.27 1.34 427.84 8.352 Simplex 14 Simplex 1 42.67 0.99 448.36 5.933 Simplex 14 Simplex 7 34.40 1.30 438.40 6.904 Simplex 14 Simplex 90 43.36 1.22 459.93 8.265 Simplex 90 Simplex 1 43.75 1.15 457.96 7.286 Ab Simplex 14 Ab Simplex 1 42.19 1.16 451.25 6.68

Table 4-2 Tabulated data for C-C interfacial shear properties displayed in Figures

4-10 to 4-13.

4.5 Discussion

Mechanisms of adhesion between polymer solids include surface mechanical

interlocking, polymer chain inter-diffusion, chemical bonding and electrostatic

interactions (Garbassi et al. 1998) (Figure 4-14). It is generally accepted that the

different mechanisms of adhesion make a variable contribution to the strength of the

interface between bonded polymers. In this study the interfacial surface roughness of the

outer (old) cement mantle after casting rod extraction was 0.22 μm Ra (95% CI 0.20 –

0.24 μm Ra). This surface roughness is of particular clinical relevance as it approximates

Figure 4-14 Contributions to adhesion between polymer solids: (a) mechanical

interlocking; (b) interdiffusion of chains; (c) electrical interactions; (d) chemical

interactions (Garbassi et al. 1998).

Figure removed for digital reproduction.

Page 123: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

98

that of a PMMA cement mantle after removal of a polished cemented stem without

further intraoperative modification such as rasping (<0.25 μm Ra; Weinrauch 2005:

Unpublished data, Queensland University of Technology). Despite the smooth surface

finish reducing the opportunity for mechanical interlock between the cement mantles

(Bundy et al. 1987; Chen et al. 1999), high C-C interfacial shear stress values were

observed in this study and also by previous authors (Gruen et al. 1976; Greenwald et al.

1978; Li et al. 1996). The observed C-C interfacial shear stress (approximately 40MPa)

greatly exceeded the cement-metal interfacial shear stress (1.2 MPa) despite similar

surface roughness. The limited role of adhesion by mechanical interlock between

PMMA and metals of roughness similar to that used in this study has been well

established in the literature (Beaumont et al. 1977; Raab et al. 1981; Bundy et al. 1987;

Chen et al. 1999). Interfacial adhesion by mechanisms other than mechanical interlock

must therefore play a significant role in the bond formed between new and old PMMA

cement mantles. Bundy et al. (1987) noted the adhesion of PMMA cement to metals

paradoxically increased with decreasing surface roughness beyond a threshold limit

(Figure 4-15). On the basis of these observations, Bundy et al. (1987) hypothesised that

with decreasing surface roughness, adhesion by mechanical interlock decreases however

there is increased adhesion by atomic (or chemical) interactions such as Van der Waals

forces (Bundy et al. 1987). At the interfacial region between C-C mantles only limited

chemical bonding is possible as the residual monomer content within aged orthopaedic

cements is characteristically less than 1 percent, and terminated PMMA chains within

the old mantle possess few reactive sites for further polymerisation with the monomer of

the freshly applied cement (Kuhn 2000; Kuehn et al. 2005). This would therefore

suggest that the predominant adhesion mechanisms resulting in firm bonding between

new and old cement are interfacial inter-diffusion and possibly electrostatic adhesion.

Adhesion by inter-diffusion or “molecular interdigitation” is obtained when the addition

of fresh cement dough containing relatively high concentrations of monomer causes

swelling and separation of the PMMA chains at the interfacial surface of the old cement

mantle. This allows limited diffusion of the monomer from the freshly applied cement

between the PMMA chains of the old cement mantle, and subsequent polymerisation

results in the formation of PMMA chains which bridge the interfacial region. This

Page 124: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

99

mechanism would explain the findings of previous authors correlating monomer

concentration with interfacial shear strength between bilaminar cement mantles (Gruen

et al. 1976; Greenwald et al. 1978). Molecular interdigitation between the cement

mantles mechanism also explains the absence of a clearly defined interface between

cement mantles with LM and SEM examination prior to shear testing.

This experiment demonstrates C-C interfacial strength results similar to that

obtained by previous authors (Gruen et al. 1976; Greenwald et al. 1978; Li et al. 1996).

Bilaminar mantles demonstrated a reduction in ultimate shear strength of only 15 – 20 %

compared to of uniform mantles across all time periods, with interfacial stress to failure

34.4 – 43.7 MPa depending on the duration of post-cure. Variation in interfacial shear

strength with respect to the duration of specimen post-cure was observed for bilaminar

test samples. After initial maturation and strengthening of the interfacial region within

the first 24 hours of the application of fresh cement, a temporary weakening of

Figure 4-15 Hypothetical components of the adhesion force between PMMA bone

cement and metal surfaces (Bundy et al. 1987).

Figure removed for digital reproduction.

Page 125: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

100

8.2 MPa (19.4%) was observed at day seven. For bilaminar test specimens with an old

mantle with at least 2 weeks of age, variability in the interfacial shear strength was

demonstrated to be most dependent on the duration of post-cure of the freshly applied

cement (new mantle). Similar patterns of shear strength variability were observed for

uniform mantle test specimens and that observed for bilaminar mantle test specimens

with an equivalent duration post-cure of the new cement mantle. Duration of outer (old)

cement mantle post-cure prior to application of new cement (2 or 12 weeks) was not

demonstrated to influence the interfacial shear strength between cement mantles. These

observations provide additional support for the importance of adhesion by molecular

interdigitation between layered PMMA cements, as the PMMA chains bridging the

interfacial region are polymerisation products of the freshly applied cement dough.

Commercial addition of Tobramycin to Simplex cement in this study was not

associated with a reduction in interfacial shear strength between the new and old cement

mantles. As infection risk is significantly higher for THA revision than primary surgery

(Poss et al. 1984), and the use of antibiotic loaded cement has been demonstrated to

reduce infection rates resulting in superior prosthetic survivorship in primary THA

(Havelin et al. 2000), the use of antibiotic PMMA cement in C-C revision is

recommended.

In this experiment the outer (old) cement mantle was clean, dry and free from

debris prior to the application of fresh cement, resulting in a high interfacial shear

strength consistent with previous investigations (Gruen et al. 1976; Greenwald et al.

1978; Li et al. 1996). Li et al. (1996) found the ultimate shear stress of cleaned bilaminar

Palacos LV cement mantles to be 39 MPa, however the inclusion of a thin layer of blood

with clots and marrow debris was associated with a significant reduction in ultimate

interfacial shear strength to 6.2MPa. Li et al. (1996) argues that in clinical practice the

removal of all contaminants from the interface cannot be reliably achieved, and therefore

all cement should be removed. The suggestion that adequately clean and dry mantles

cannot be achieved is challenged by the extensive clinical experience with this technique

as reported by Hubble et al. (2005) and the favourable clinical results described by

multiple authors (Lieberman et al. 1993; McCallum et al. 1995; Hubble et al. 2005). The

Page 126: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

101

results obtained in this study are clinically applicable only after appropriate surgical

technique with attention paid to the preparation of the old cement mantle (Greenwald et

al. 1978; Li et al. 1996; Stryker International 2005).

When considering the overall clinical significance of the shear strength of

bilaminar cement interfaces in the context of revision THA, the relative strength at bone-

cement interface must be considered. Moran et al. (1979) using SPT methods

determined the bone – cement interfacial shear strength to be less than 5 MPa (Moran et

al. 1979). Utilisation of modern cementing techniques with distal plugging, canal lavage

and retrograde fill of low viscosity cement using syringe pressurisation results in

improved cement interdigitation, however higher bone-cement interfacial shear strength

consistently remains less than 10 MPa in all regions of the femur (Halawa et al. 1978;

MacDonald et al. 1993). Rosenstein et al. (1992) found the bone-cement interfacial shear

strength after revision was reduced 30 per cent compared to primary implantation due to

compromise in cancellous bone stock sustained during cement removal. The importance

of these findings is that even though bilaminar cement mantles demonstrate lower

ultimate shear strength compared to uniform mantles, the strength still greatly exceeds

that of the bone – cement interface, particularly in revision THA where the bone stock is

compromised. Conversely it may be argued that while the bone-cement interface has

some biological potential for repair, fatigue at the C-C interface is accumulative, and

therefore the standard by which C-C interfaces should be compared is to a uniform

cement mantle. An ideal bilaminar cement mantle in this regard should therefore have

the same material properties as a uniform mantle. Attention to appropriate surgical

technique to maximise the strength of the C-C interfacial region therefore remains

important, even though the C-C interfacial shear strength significantly exceeds that of

the bone-cement interface.

4.6 Conclusions

Bilaminar cement mantles demonstrate minor variations in interfacial shear

strength related to the duration of post cure of the inner (new) cement mantle. Even

without surface roughening of the old cement mantle, the ultimate shear strength at the

Page 127: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

102

interfacial region of bilaminar cement mantles is approximately 80 - 85 percent of

uniform mantles when the new cement is applied within 90 seconds after mixture of the

liquid and powder constituents. High C-C interfacial strengths are demonstrated as early

as one hour after cement application, therefore not precluding early postoperative

rehabilitation and weight bearing. Interfacial adhesion by mechanisms other than

mechanical interlock play a significant role in the bond formed between new and old

cement mantles, and the contribution of diffusion based molecular interdigitation is

emphasised.

C-C femoral revision is recommended as a viable technique on the basis of the

strong interfacial bond formed between new and old cement mantles. The use of

Antibiotic Simplex in C-C revision is recommended as detrimental effects on the

interfacial shear properties have not been demonstrated with the commercial addition of

Tobramycin.

Page 128: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

103

Page 129: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

104

Page 130: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

105

Chapter 5

Conclusions and Recommendations

5.1 Application of Research Findings to Clinical Practice

Surgical Simplex P and Antibiotic Simplex (Tobramycin) PMMA cements

demonstrate variability in ultimate shear strength with reference to duration of post cure,

however the magnitude of these variations are small and probably of no clinical

relevance. The manufacturer’s addition of Tobramycin antibiotic does not significantly

alter the shear properties of Simplex PMMA cement. Bilaminar cement mantles

demonstrate variability in interfacial shear properties related to the duration of post cure

of the freshly applied cement mantle. Without surface roughening of the old cement

mantle, when the new cement is applied early after mixture of the liquid and powder

constituents the ultimate shear strength at the interfacial region between cement mantles

is approximately 80 - 85 percent of uniform mantles. High C-C interfacial strengths are

demonstrated as early as one hour after cement application, therefore not precluding

early postoperative rehabilitation and weight bearing. Interfacial adhesion by

mechanisms other than mechanical interlock play a significant role in the bond formed

between new and old cement PMMA cement mantles, with an important contribution by

diffusion based molecular interdigitation.

C-C femoral revision is recommended as a viable technique on the basis of the

strong interfacial bond formed between new and old cement mantles. The use of

Antibiotic Simplex in C-C revision is recommended as detrimental effects on the

interfacial shear properties have not been demonstrated with the commercial addition of

Tobramycin.

5.2 Further Research

Various aspects of C-C interfacial shear properties have been evaluated within

this thesis and within the literature. To date the longest clinical follow up of any series of

Page 131: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

106

C-C femoral revisions is 5 years, and the results thus far are highly favourable for the

technique. In vitro testing conducted within this study and reported within the literature

however demonstrates that the material properties of bilaminar cement mantles are

inferior to uniform mantles. What is not known presently is if the reduction in material

properties observed in bilaminar cement mantles in vitro will become clinically

significant beyond 5 years in vivo. Ultimately this will not be determined adequately

until the current cohort of patients in the Exeter series is assessed at 10 years or greater

post C-C revision. Laboratory based tests can however be utilised to predict the long

term clinical outcome, and in particular fatigue testing of C-C interfaces in a simulated

THA revision would be of value. Within a suitably controlled experimental design,

fatigue limit testing of bilaminar cement mantles using SPT techniques would also

provide useful information of clinical relevance. Valuable insight will be obtained by

finite element modelling of the stress distribution within the cement mantle and at the

interfacial region after C-C femoral revision.

Presently the majority of research relevant to C-C femoral revision has been

conducted using Surgical Simplex P and Antibiotic Simplex cements. Variation in the

types of cement used and preparation techniques, and the effect of dissimilar C-C

combinations has not been evaluated. Interfacial shear testing of bilaminar cement

mantles using various combinations of cement preparations, with and without vacuum

mixing, would provide useful clinical reassurance that C-C revision may be considered

regardless of what cement preparation was used for the primary arthroplasty.

Page 132: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

107

Page 133: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

108

Page 134: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

109

Appendix 1 Design Drawings of Shear Testing Device

.

Page 135: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

110

Page 136: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

111

Page 137: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

112

Page 138: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

113

Page 139: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

114

Page 140: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

115

Page 141: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

116

Page 142: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

117

Appendix 2 Calibration of Hounsfield Load Cell

Load cell serial number: 705421

Calibration date: 9 Sep 2004

Next calibration due: 9 Sep 2006

Compression range: 25 000 N

Grade A from: 1 000 N

Grade B from: 300 N

Grade C from: 160 N

Tension range: 25 000 N

Grade A from: 160 N

Australian Calibrating Services Pty Ltd

Report number: QAC/401 B-A-1

Page 143: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

118

Page 144: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

119

Appendix 3 Technical Report on Exeter PMMA Cement Reamer

Weinrauch P, Beach T, Goss B, Lutton C, Bell C & Crawford R

Queensland University of Technology

School of Engineering Systems, Brisbane Australia

halla
This article is not available online. Please consult the hardcopy thesis available from the QUT Library
Page 145: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

131

References Alkire, M.J., Dabezies, E.J. and Hastings, P.R. (1987). High vacuum as a method of reducing porosity of polymethylmethacrylate. Orthopedics 10(11): 1533-9.

Amstutz, H.C., Ma, S.M., Jinnah, R.H. and Mai, L. (1982). Revision of aseptic loose total hip arthroplasties. Clin Orthop 170: 21 - 33.

Archibald, D.A., Protheroe, K., Stother, I.G. and Campbell, A. (1988). A simple technique for acetabular revision: brief report. J Bone Joint Surg Br 70(5): 838.

ASTM B 769 - 94 (1994). Standard Test Method for Shear Testing of Aluminum Alloys.

ASTM C 273-00 (2000). Standard Test Method for Shear Properties of Sandwich Core Materials.

ASTM D 732-02 (2002). Standard Test Method for Shear Strength of Plastics by Punch Tool.

ASTM F 451 - 99a (1999). Standard Specification for Acrylic Bone Cement.

Australian Orthopaedic Association National Joint Replacement Registry (2003). Annual Report 2003. Adelaide, AOA.

Australian Orthopaedic Association National Joint Replacement Registry (2004). Annual Report 2004. Adelaide, AOA.

Barrack, R.L. and Folgueras, A.L. (1995). Revision Total Hip Arthroplasty. The Femoral Component. J Am Acad Orthop Surg 3: 79-85.

Barrack, R.L., Mulroy, R.D., Jr. and Harris, W.H. (1992). Improved cementing techniques and femoral component loosening in young patients with hip arthroplasty. A 12-year radiographic review. J Bone Joint Surg Br 74(3): 385-9.

Beaumont, P.W.R. and Plumpton, B. (1977). The strength of acrylic bone cements and acrylic cement-stainless steel interfaces, part 2: the shear strength of an acrylic cement - stainless steel interface. J Mater Sci 12: 1853 - 6.

Bell, C.G. (2005). A Finite Element and Experimental Investigation of the Femoral Component Mechanics in a Total Hip Arthroplasty. Engineering Systems. Brisbane, Queensland University of Technology.

Berger, R., Fletcher, F., Donaldson, T., Wasielewski, R., Peterson, M. and Rubash, H.E. (1996). Dynamic Test to Diagnose Loose Uncemented Femoral Total Hip Components. Clin Orthop 330: 115 - 23.

Bishop, N.E., Ferguson, S. and Tepic, S. (1996). Porosity reduction in bone cement at the cement-stem interface. J Bone Joint Surg Br 78(3): 349-56.

Page 146: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

132

Black, J.D., Jr. and Greenwald, A.S. (1982). Structural weakening of layered acrylic bone cement. Clin Orthop(171): 94-6.

Brand, R.A., Pedersen, D.R. and Yoder, S.A. (1986). How Definition of "Loosening" Affects the Incidence of Loose Total Hip Reconstructions. Clin Orthop 210: 185 - 91.

Breusch, S.J. (2001). Cementing Technique in Total Hip Replacement: Factors Influencing Survival of Femoral Components. Bone Cements and Cementing Technique. G. H. I. M. Walenkamp and D. W. Murray. Berlin, Springer.

Brooks, A.T., Nelson, C., Stewart, C.L., Skinner, R.A. and Siems, M.L. (1993). Effect of an Ultrasonic Device on Temperatures Generated in Bone and on Bone-Cement Structure. J Arthroplasty 8(4): 413 - 8.

Bundy, K.J. and Penn, R.W. (1987). The effect of surface preparation on metal/bone cement interfacial strength. J Biomed Mater Res 21(6): 773-805.

Canadian Institute for Health Information (2004). Canadian Joint Replacement Registry 2004 Annual Report.

Cautilli, G.P. and Hozack, W.J. (1994). Analysis of Fume Emission From Ultrasonic Removal of Methymethacrylate Cement in Revision Hip Surgery. J Arthroplasty 9(3): 305 - 6.

Centre for Devices and Radiological Health (2002). Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Cement; Guidance for Industry and FDA. Rockville, Food and Drug Administration.

Charnley, J. (1960). Anchorage of the femoral head prosthesis to the shaft of the femur. J Bone Joint Surg Br 42(1): 28 - 30.

Charnley, J. (1970). Acrylic cement in orthopaedic surgery. Edinburgh and London, E & S Livingstone.

Chen, C.Q., Scott, W. and Barker, T.M. (1999). Effect of metal surface topography on mechanical bonding at simulated total hip stem-cement interfaces. J Biomed Mater Res 48(4): 440-6.

Choy, D.S.J., Kaminow, I.P., Kaplan, M., Case, R.B. and Zickel, R. (1987). Experiemtal Nd:YAG Laser Disintegration of Methylmethacrylate. Analysis of the Gaseous Products. Clin Orthop 215: 287 - 8.

Crawford, R.W., Gie, G.A., Ling, R.S. and Murray, D.W. (1998). Diagnostic value of intra-articular anaesthetic in primary osteoarthritis of the hip. J Bone Joint Surg Br 80(2): 279-81.

Page 147: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

133

Dahl, O.E., Johnsen, H., Kierulf, P., Molnar, I., Ro, J.S., Vinje, A. and Mowinckel, P. (1992). Intrapulmonary thrombin generation and its relation to monomethylmethacrylate plasma levels during hip arthroplasty. Acta Anaesthesiol Scand 36(4): 331-5.

D'Arcy, J. and Devas, M. (1976). Treatment of fractures of the femoral neck by replacement with the Thompson prosthesis. J Bone Joint Surg Br 58(3): 279-86.

De Wijn, J.R., Driessens, F.C. and Slooff, T.J. (1975). Dimensional behavior of curing bone cement masses. J Biomed Mater Res 9(4): 99-103.

Department of Orthopaedic Surgery Haukeland University Hospital (2004). The Norweigan Arthroplasty Register.

Devas, M. and Hinves, B. (1983). Prevention of acetabular erosion after hemiarthroplasty for fractured neck of femur. J Bone Joint Surg Br 65(5): 548-51.

d'Hollander, A., Monteny, E., Hooghe, L., Camu, F., Donkerwolcke, M. and Brauman, H. (1979). Cardiovascular effects of methyl methacrylate monomer. Surg Gynecol Obstet 149(1): 61-4.

Dixon, S. and Bannister, G. (2004). Cemented bipolar hemiarthroplasty for displaced intracapsular fracture in the mobile active elderly patient. Injury 35(2): 152 - 6.

Dorr, L., Glousman, R., Sew Hoy, A.L., Vanis, R. and Chandler, R. (1986). Treatment of Femoral Neck Fractures With Total Hip Replacement Versus Cemented and Non Cemented Hemiarthroplasty. J Arthroplasty 21 - 28.

Ege, W. and Kuhn, K.D. (2001). Industrial Development of Bone Cement - Twenty Five Years of Experience. Bone Cements and Cementing Technique. G. H. I. M. Walenkamp and D. W. Murray. Berlin, Springer-Verlag.

Evans, B.G. and Cuckler, J.M. (1992). Evaluation of the painful total hip arthroplasty. Orthop Clin N Am 23(2): 303 - 11.

Farshad, M., Wildenberg, M.W. and Flueler, P. (1997). Determination of shear modulus and poisson's ratio of polymers and foams by the anticlastic plate bending method. Mater Struct 30: 377 - 82.

Fitts, W.T., Lehr, H.B., Schor, S. and Roberts, B. (1959). Life expectancy after fracture of the hip. Surg Gynae Obst 108: 7 - 12.

Flivik, G., Yuan, X., Ryd, L., Juliusson, R. and Lidgren, L. (1997). Effects of lamination on the strength of bone cement. Acta Orthop Scand 68(1): 55-8.

Foster, A.P., Thompson, N.W., Wong, J. and Charlwood, A.P. (2005). Periprosthetic femoral fractures - a comparison between cemented and uncemented hemiarthroplasties. Injury 36(3): 424-9.

Page 148: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

134

Fowler, J.L., Gie, G. and Ling, R.S. (1988). Experience with the Exeter Total Hip Replacement Since 1970. Orthop Clin N Am 19(3): 477 - 89.

Garbassi, F., Morra, M. and Occhiello, E. (1998). Polymer Surfaces - From Physics to Technology. West Sussex, England, John Wiley & Sons.

Gardiner, R., Hozack, W.J., Nelson, C. and Keating, E.M. (1993). Revision Total Hip Arthroplasty Using Ultrasonically Driven Tools. A Clinical Evaluation. J Arthroplasty 8(5): 517 - 21.

Garrellick, G., Malchau, H., Herberts, P., Hansson, E., Axelsson, H. and Hansson, T. (1998). Life expectancy and cost utility after total hip replacement. Clin Orthop 346: 141 - 51.

Gebhard, J.S., Amstutz, H.C., Zinar, D.M. and Dorey, F.J. (1992). A comparison of total hip arthroplasty and hemiarthroplasty for treatment of acute fracture of the femoral neck. Clin Orthop Relat Res(282): 123-31.

Goldstein, W.M., Gleason, T.F., Kopplin, M. and Branson, J.J. (2001). Prevalence of dislocation after total hip arthroplasty through a posterolateral approach with partial capsulotomy and capsulorrhaphy. J Bone Joint Surg Am 83-A Suppl 2(Pt 1): 2-7.

Gordon, P.C. (1971). The probability of death following a fracture of the hip. Can Med Assoc J 10: 47 - 62.

Gray, F.B. (1992). Total hip revision arthroplasty. Prosthesis and Cement Removal Techniques. Orthop Clin N Am 23(2): 313 - 9.

Greenwald, A.S., Narten, N.C. and Wilde, A.H. (1978). Points in the technique of recementing in the revision of an implant arthroplasty. J Bone Joint Surg Br 60(1): 107-10.

Gruen, T.A., Markolf, K.L. and Amstutz, H.C. (1976). Effects of laminations and blood entrapment on the strength of acrylic bone cement. Clin Orthop Relat Res(119): 250-5.

Gruen, T.A., McNeice, G.M. and Amstutz, H.C. (1979). Modes of Failure of Cemented Stem - type Femoral Components. Clin Orthop 141: 17 - 27.

Guduru, R., Daling, K., Kishore, R., Scattergood, R., Koch, C. and Murty, K. (2005). Evaluation of mechanical properties using shear-punch testing. Materials Science and Engineering A(395): 307-14.

Halawa, M., Lee, A.J.C., Ling, R.S.M. and Vangala, S.S. (1978). The Shear Strength of Trabecular Bone from the Femur, and some Factors Affecting the Shear Strength of the Cement-Bone Interface. Arch Orthop Trauma Surg 92: 19 - 30.

Hardy, D.C., Reinus, W.R., Totty, W.G. and Keyser, C.K. (1998). Artthrography after total hip arthroplasty: utility of postambulation radiographs. Skeletal radiology 17: 20 -3.

Page 149: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

135

Harkess, J.W. (1999). Arthroplasty of Hip. Campbell's Operative Orthopaedics. S. T. Canale, Mosby.

Harper, E.J. and Bonfield, W. (2000). Tensile cahracteristics of Ten Commercial Acrylic Bone Cements. J Biomed Mater Res B Appl Biomater 53: 741 - 6.

Havelin, L.I., Espehaug, B., Lie, S.A., Engesaeter, L.B., Furnes, O. and Vollset, S.E. (2000). Prospective Studies of Hip Prostheses and Cements. A Presentation of the Norwegian Arthroplasty Register 1987 - 1999. 67th Annual Meeting of the American Academy of Orthopaedic Surgeons, Orlando, Florida, USA.

Hendrix, R.W., Wixson, R.L., Rana, N.A. and Rogers, L.F. (1983). Arthrography after Total Hip Arthroplasty: A Modified Technique Used in the Diagnosis of Pain. Radiology 148(3): 647 - 52.

Hodgkinson, J.P., Shelley, P. and Wroblewski, B.M. (1988). The Correlation Between Roentgenographic Appearance and Operative Findings at he Bone - Cement Junction of the Socket in Charnley Low Friction Arthroplasties. Clin Orthop 228: 105 - 9.

Hozack, W.J. (1998). Removal of Components and Cement. The Adult Hip. J. J. Callaghan, A. Rosenberg and H. E. Rubash. Philadelphia, Lippincott-Raven Publishers. 2.

Hubble, M., Duncan, W.W., Timperley, A.J., Gie, G. and Howell, J. (2005). Cement in Cement Femoral Revision with the Exeter Hip. Australian Orthopaedic Association 65th Annual Scientific Meeting, Perth, Australia.

ISO 5833 (2002). Implants for surgery - Acrylic resin cements.

Jadhav, A.P., Kulkarni, S.S., Vaidya, S.V., Divekar, M.M. and Suralkar, S.P. (1996). Results of Austin Moore replacement. J Postgrad Med 42(2): 33-8.

James, J. (2005). Tolerance of Shear Properties Measurements: Sawbones 20 and 40 pcf Soild Rigid Polurethane Foam. P. C. Weinrauch.

Jefferiss, C.D., Lee, A.J. and Ling, R.S. (1975). Thermal aspects of self - curing polymethylmethacylate. J Bone Joint Surg Br 57(4): 511 - 8.

Jones, L.C. and Hungerford, D.S. (1987). Cement Disease. Clin Orthop 225: 192 - 206.

Karlsson, J., Wendling, W., Chen, D., Zelinsky, J., Jeevanandam, V., Hellman, S. and Carlsson, C. (1995). Methylmethacrylate monomer produces direct relaxation of vascular smooth muscle in vitro. Acta Anaesthesiol Scand 39(5): 685-9.

Kenzora, J.E., McCarthy, R.E., Lowell, J.D. and Sledge, C.B. (1984). Hip fracture mortality. Relation to age, treatment, preoperative illness, time of surgery and complications. Clin Orthop 186: 45 - 55.

Page 150: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

136

Khan, R.J., MacDowell, A., Crossman, P., Datta, A., Jallali, N., Arch, B.N. and Keene, G.S. (2002). Cemented or uncemented hemiarthroplasty for displaced intracapsular femoral neck fractures. Int Orthop 26(4): 229-32.

Kofoed, H. and Kofod, J. (1983). Moore prosthesis in the treatment of fresh femoral neck fractures. A critical review with special attention to secondary acetabular degeneration. Injury 14(6): 531-40.

Kuehn, K.D., Ege, W. and Gopp, U. (2005). Acrylic bone cements: composition and properties. Orthop Clin N Am 36: 17 - 28.

Kuehn, K.D., Ege, W. and Gopp, U. (2005). Acrylic bone cements: mechanical and physical properties. Orthop Clin N Am 36: 29 -39.

Kuhn, K.D. (2000). Bone Cements. Up -to-Date Comparison of Physical and Chemical Properties of Commercial Materials. Berlin, Springer - Verlag.

Kuhn, K.D. (2001). Handling Properties of Polymethylmethacrylate Cements. Bone Cements and Cementing Technique. G. H. I. M. Walenkamp and D. W. Murray. Berlin, Springer-Verlag.

Kuhn, K.D., Specht, R., Ege, W. and Kock, H.J. (2001). Mechanical Properties of Bone Cements. Bone Cements and Cementing Technique. G. H. I. M. Walenkamp and D. W. Murray. Berlin, Springer-Verlag.

Kwok, D. and Cruess, R.L. (1982). A Retrospective Study of Moore and Thompson Hemiathroplasty. Clinical Orthopaedics and Related Research 159: 179 - 85.

Larson, J.E., Chao, E.Y., Fitzgerald, R.H. and Kavanagh, B.F. (1988). Bypassing femoral cortical defects. Trans Orthop Res Soc 13: 525 - 8.

Lausten, G.S. and Vedel, P. (1982). Cementing v. not cementing the Monk endoprosthesis. Injury 13(6): 484-8.

Lee, A.J., Ling, R.S. and Vangala, S.S. (1977). The mechanical properties of bone cements. J Med Eng Technol 1(3): 137-40.

Lee, A.J., Ling, R.S. and Vangala, S.S. (1978). Some Clinically Relevant Variables Affecting the Mechanical Behaviour of Bone Cement. Arch Orthop Trauma Surg 92: 1 - 18.

Lee, A.J., Ling, R.S. and Wrighton, J.D. (1973). Some Properties of Polymethylmethacrylate with Reference to Its Use in Orthopaedic Surgery. Clin Orthop 95: 281 - 7.

Lesaka, K., Jaffe, W.L. and Kummer, F.J. (2003). Effects of preheating of hip prostheses on the stem-cement interface. J Bone Joint Surg Am 85-A(3): 421-7.

Page 151: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

137

Li, P.L., Ingle, P.J. and Dowell, J.K. (1996). Cement-within-cement revision hip arthroplasty; should it be done? J Bone Joint Surg Br 78(5): 809-11.

Lidgren, L., Drar, H. and Moller, J. (1984). Strength of polymethylmethacrylate increased by vacuum mixing. Acta Orthop Scand 55: 536 - 41.

Lieberman, J.R., Moeckel, B.H., Evans, B.G., Salvati, E.A. and Ranawat, C.S. (1993). Cement-within-cement revision hip arthroplasty. J Bone Joint Surg Br 75(6): 869-71.

Linder, L., Lindberg, L. and Carlsson, A. (1983). Aseptic Loosening of Hip Prostheses: A Histologic and Enzyme Histochmical Study. Clin Orthop 175: 93 - 104.

Lu - Yao, G.L., Keller, R.B., Littenberg, B. and Wennberg, J.E. (1994). Outcomes after displaced fractures of the femoral neck. A meta-analysis of one hundred and six published reports. J Bone Joint Surg Am 76: 15 - 25.

Lutton, C., Goss, B., Gerritis, B. and Crawford, R. (2005). Cement-on-Cement Adhesion in Revision Hip Arthroplasty. QUT, Department of Mechanical Engineering.

Lyons, C.W., Berquest, T.H., Lyons, J.C., Rand, J.A. and Brown, M.L. (1985). Evaluation of Radiographic Findngs in Painful Hip Arthroplasties. Clin Orthop 195: 239 - 51.

MacDonald, W., Swarts, E. and Beaver, R. (1993). Penetration and Shear Strength of Cement - Bone Interfaces In Vivo. Clin Orthop 286: 283 - 8.

Maus, T.P., Berquist, T.H., Bender, C.E. and Rand, J.A. (1987). Arthrographic Study of Painfrul Total Hip Arthroplasty: Refined Criteria. Radiology 162(3): 721 - 7.

McAuliffe, M., Ivers, R., Whitehouse, S. and Crawford, R. (2004). The unsolved fracture: A scoring system to direct the treatment of displaced subcapital fractures in patients over 60. QAOA, Couran Cove, Queensland.

McCallum, J.D., 3rd and Hozack, W.J. (1995). Recementing a femoral component into a stable cement mantle using ultrasonic tools. Clin Orthop(319): 232-7.

Moran, J.M., Greenwald, A.S. and Matejczyk, M.B. (1979). Effect of gentamicin on shear and interface strengths of bone cement. Clin Orthop(141): 96-101.

Muller, S.D., Green, S.M. and McCaskie, A.W. (2002). The dynamic volume changes of polymerising polymethyl methacrylate bone cement. Acta Orthop Scand 73(6): 684-7.

Mulroy, W.F., Estok, D.M. and Harris, W.H. (1995). Total hip arthroplasty with use of so-called second-generation cementing techniques. A fifteen-year-average follow-up study. J Bone Joint Surg Am 77(12): 1845-52.

Page 152: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

138

O'Neill, R.A. and Harris, W.H. (1984). Failed Total Hip Replacement: Assessment by Plain Radiographs, Arthrograms, and Aspiration of the Hip Joiint. J Bone Joint Surg Am 66(4): 540 -6.

Ovesen, O., Riegels-Nielsen, P., Lindequist, S., Jensen, I., Munkner, T., Torfing, T. and Marving, J. (2003). The Diagnostic Value of Digital Subtraction Arthrography and Radionuclide Bone Scan in Revision Hip Arthroplasty. J Arthroplasty 18(6): 735 - 40.

Pacific Research Laboratories (2005). Typical Properties of Sawbones Solid Rigid Polyurethane Foam.

Park, S.H., Silva, M., Park, J.S., Ebramzadeh, E. and Schmalzried, T.P. (2001). Cement - Cement Interface Strength: Influence of Time to Apposition. J Biomed Mater Res (Appl Biomater) 58: 741 - 6.

Parker, M.J. and Gurusamay, K. (2004). Arthroplasties (with and without bone cement) for proximal femoral fractures in adults (Review). The Cochrane Database of Systematic Reviews.

Parker, M.J., Myles, J.W., Anand, J.K. and Drewett, R. (1992). Cost-benefit analysis of hip fracture treatment. J Bone Joint Surg Br 74: 261 - 4.

Parks, M.L., Walsh, H.A., Salvati, E.A. and Li, S. (1998). Effect of increasing temperature on the properties of four bone cements. Clin Orthop Relat Res(355): 238-48.

Parvizi, J., Ereth, M.H. and Lewallen, D.G. (2004). Thirty-day mortality following hip arthroplasty for acute fracture. J Bone Joint Surg Am 86-A(9): 1983-8.

Parvizi, J., Holiday, A.D., Ereth, M.H. and Lewallen, D.G. (1999). The Frank Stinchfield Award. Sudden death during primary hip arthroplasty. Clin Orthop Relat Res(369): 39-48.

Patterson, B.M., Healey, J.H., Cornell, C.N. and Sharrock, N.E. (1991). Cardiac arrest during hip arthroplasty with a cemented long-stem component. A report of seven cases. J Bone Joint Surg Am 73(2): 271-7.

Phillips, T.W. (1989). Thompson hemiarthroplasty and acetabular erosion. J Bone Joint Surg Am 71(6): 913-7.

Pitto, R.P. and Koessler, M. (1999). The risk of fat embolism during cemented total hip replacement in the elderly patient. Chir Organi Mov 84(2): 119-28.

Pitto, R.P., Koessler, M. and Kuehle, J.W. (1999). Comparison of fixation of the femoral component without cement and fixation with use of a bone-vacuum cementing technique for the prevention of fat embolism during total hip arthroplasty. A prospective, randomized clinical trial. J Bone Joint Surg Am 81(6): 831-43.

Page 153: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

139

Poss, R., Thornhill, T.S., Ewald, F.C., Thomas, W.H., Batte, N.J. and Sledge, C.B. (1984). Factors influencing the incidence and outcome of infection following total joint arthroplasty. Clin Orthop Relat Res(182): 117-26.

Raab, S., Ahmed, A.M. and Provan, J.W. (1981). The quasistatic and fatigue performance of the implant/bone cement interface. J Biomed Mater Res 15(2): 159 - 82.

Ranawat, C.S., Deshmukh, R.G., Peters, L.E. and Umlas, M.E. (1995). Prediction of th eLong - Term Durability of All - Polyethylene Cemented Sockets. Clin Orthop 317: 89 - 105.

Reckling, F.W. and Dillon, W.L. (1976). The Bone - Cement Interface Temperature during Total Joint Replacement. J Bone Joint Surg Am 59(1): 80 - 2.

Ries, M.D., Lynch, F., Rauscher, L.A., Richman, J., Mick, C. and Gomez, M. (1993). Pulmonary function during and after total hip replacement. Findings in patients who have insertion of a femoral component with and without cement. J Bone Joint Surg Am 75(4): 581-7.

Ritter, M.A., Zhou, H., Keating, C.M., Keating, E.M., Faris, P.M., Meding, J.B. and Berend, M.E. (1999). Radiological factors influencing femoral and acetabular faliure in cemented Charnley total hip arthroplasties. J Bone Joint Surg Br 81(6): 982 - 6.

Rosenstein, A., MacDonald, W., Iliadis, A. and McLardy-Smith, P. (1992). Revision of cemented fixation and cement-bone interface strength. Proc Inst Mech Eng [H] 206(1): 47-9.

Saha, S. and Warman, M.L. (1979). Compressive and shear properties of graphite fibre reinforced bone cement. Orthop Trans(3): 169.

Schultz, R.J., Johnstone, A.D. and Krishnamurthy, S. (1987). Thermal effects of polymerisation of methyl-methacrylate on small tubular bones. International Orthopaedics 11: 277 - 82.

Schurman, D.J. and Maloney, W.J. (1992). Segmental cement extraction at revision total hip arthroplasty. Clin Orthop(285): 158-63.

Sharif, K.M. and Parker, M.J. (2002). Austin Moore hemiarthroplasty: technical aspects and their effects on outcome, in patients with fractures of the neck of femur. Injury 33(5): 419-22.

Sih, G.C. and Connelly, G.M. (1980). The effect of thickness and pressure on the curing of PMMA bone cement for the total hip replacement. Biomechanics 13.

Smith, D.C. (2005). The genesis and evolution of acrylic bone cement. Orthop Clin N Am 36: 1 - 10.

Page 154: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

140

Sonne-Holm, S., Walter, S. and Jensen, J.S. (1982). Moore hemi-arthroplasty with and without bone cement in femoral neck fractures. A clinical controlled trial. Acta Orthop Scand 53(6): 953-6.

Spinelli, R. (1976). A study of the interface between bone and acrylic cement by scanning electron microscopy. Ital J Orthop Traumatol 2(1): 103 - 15.

Stryker Howmedica Osteonics (2002). Design drawings of Exeter V40 +44 Offset No. 3 long stem 200 - 260mm: 1 - 4.

Stryker International (2005). Cement - Cement Revision Technique.

Stryker International (2005 B). Exeter trauma stem surgical technique. Exeter, UK, Princess Elizabeth Orthopaedic Centre.

Swan, J.S., Braunstein, E.M., Wellman, H.N. and Capello, W. (1991). Contrast and nuclear arthrography in loosening of the uncemented hip prosthesis. Skeletal Radiol 20(1): 15-9.

Taine, W.H. and Armour, P.C. (1985). Primary Total Hip Replacement for Displaced Subcapital Fractures of the Femur. J Bone Joint Surg Br 67: 214 - 17.

The Swedish National Hip Arthroplasty Register (2004). Annual Report 2003.

Turner, R.H., Mattingly, D.A. and Scheller, A. (1987). Femoral revision total hip arthroplasty using a long-stem femoral component. J Arthroplasty 2: 247.

Udomkiat, P., Wan, Z. and Dorr, L. (2001). Comparison of Preoperative Radiograps and Intraoperative Finidings of Fixation of Hemispheric Porous - Coated Sockets. J Bone Joint Surg Am 83(12): 1865 - 70.

Wang, J.S. and Kjellson, F. (2001). Bone Cement Porosity in Vacuum Mixing Systems. Bone Cements and Cementing Technique. G. H. I. M. Walenkamp and D. W. Murray. Berlin, Springer.

Warwick, D., Hubble, M., Sarris, I. and Strange, J. (1998). Revision of failed hemiarthroplasty for fractures at the hip. Int Orthop 22(3): 165-8.

Weinrauch, P.C. (2004). Intraoperative error during implantation of the uncemented unipolar Austin Moore hemiarthroplasty. 2004 Australian Orthopaedic Association Queensland Branch Meeting, Couran Cove, Queensland Australia.

Weinrauch, P.C., Moore, D., Shooter, D., Wilkinson, M., Dedy, N., Bonrath, E., Jabur, M., McMeniman, T., Whitehouse, S. and Crawford, R. (2006). Early prosthetic complications after Unipolar Hemiarthroplasty. ANZ Journal of Surgery 76(6): 432 - 5.

Weinrauch, P.C., Moore, W., Shooter, D., Wilkinson, M., Bonrath, E., Dedy, N., McMeniman, T., Jabur, M., Whitehouse, S. and Crawford, R. (2005). Early Prosthetic

Page 155: MATERIAL PROPERTIES OF BILAMINAR ...eprints.qut.edu.au/16286/1/Patrick_Weinrauch_Thesis.pdfMATERIAL PROPERTIES OF BILAMINAR POLYMETHYLMETHACRYLATE CEMENT MANTLES IN REVISION HIP ARTHROPLASTY

141

Complications After Uncemented Austin Moore and Cemented Thompson Hemiarthroplasty. A Multicentre Review of 1118 Patients. 65th Australian Orthopaedic Association Annual Scientific Meeting, Perth, Western Australia.

Weissman, B.N. (1983). The Radiology of Total Joint Replacement. Orthop Clin N Am 14(1): 171 - 91.

Wilde, A.H. and Greenwald, A.S. (1975). Shear strength of self-curing acrylic cement. Clin Orthop(106): 126-30.

Williams, H.D.W., Browne, G., Gie, G., Ling, R.S., Timperley, A.J. and Wendover, N.A. (2002). The Exeter universal cemented femoral component at 8 to 12 years. J Bone Joint Surg Br 84(3): 324 - 34.

Wixson, R.L., Lautenschlager, E.P. and Novak, M.K. (1987). Vacuum mixing of acrylic bone cement. J Arthroplasty(2): 141 - 9.

Wool, R.P. (1991). Welding, Tack amd Green Strength of Polymers. Fundamentals of Adhesion. L. H. Lee. New York, Plenum Press: 207 - 46.

Wright, T. and Li, S. (2001). Biomaterials. Orthopedic Basic Science. Biology and Biomechanics of the Musculoskeletal System. J. A. Buckwalter, T. A. Einhorn and S. R. Simon, American Academy of Orthopedic Surgeons.

Wrighton, J.D. and Woodyard, J.E. (1971). Prosthetic replacement for subcapital fractures of the femur: a comparative survey. Injury 2(4): 287-93.

Yamagata, M., Chao, E.Y., Ilstrup, D.M., Melton, L.J., 3rd, Coventry, M.B. and Stauffer, R.N. (1987). Fixed-head and bipolar hip endoprostheses. A retrospective clinical and roentgenographic study. J Arthroplasty 2(4): 327-41.

Yasuda, I. and Iwatsuki, K. (1975). Direct effects of acryl bone cement monomer on isolated heart muscle. Tohoku J Exp Med 117(1): 93-7.

Yates, P., Gobel, D. and Bannister, G. (2002). Collarless polished tapered stem: clinical and radiological follow-up over 5 years. J Arthroplasty 17(2): 189-95.

Yau, W.P. and Chiu, K.Y. (2004). Critical radiological analysis after Austin Moore hemiarthroplasty. Injury 35(10): 1020-4.

Young, W.C. and Budynas, R.G. (2002). Roark's formulas for stress and strain. Singapore, McGraw-Hill.


Recommended