MOVACOLOR MC-Balance Instruction Manual FOR SOFTWARE VERSION : V3.0 and higher. MANUAL VERSION : V1.6  June 2005 Sales/Service/Distribution by:
Transcript

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 1/27

MOVACOLOR MC-BalanceInstruction Manual

FOR SOFTWARE VERSION : V3.0 and higher.MANUAL VERSION : V1.6

 June 2005

Sales/Service/Distribution by:

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 2/27

2

INDEX

1. GENERAL INFORMATION _________________________________________________________ 3

1.1  Warranty ______________________________________________________________________ 3

1.2  Safety _________________________________________________________________________ 3

2. METERING PRINCIPLE ____________________________________________________________ 4

3. DOSING CYLINDERS ______________________________________________________________ 5

4. INSTALLATION ___________________________________________________________________ 6

4.1 Transport ________________________________________________________________________ 6

4.2 Mechanical Installation ____________________________________________________________ 6

4.3 Electrical Connections _____________________________________________________________ 7

5. OPERATION_______________________________________________________________________ 8

5.1 The Interface _____________________________________________________________________ 8

5.2 Service Menu  ____________________________________________________________________ 9

5.3 Calibration Procedure ____________________________________________________________ 11

5.4 Weight Check Menu: _____________________________________________________________ 12

5.5 Consumption registration (optional): _______________________________________________ 12

5.6 Configuration Menu: _____________________________________________________________ 13

5.7 Production Menu:  _______________________________________________________________ 14

5.8 Function Hopper empty detection  _________________________________________________ 15

5.9 Function Fast Calibration: _________________________________________________________ 15

5.10 Filling Systems (optional) ________________________________________________________ 16

6. SYSTEM PERFORMANCE _________________________________________________________ 19

7. TROUBLESHOOTING _____________________________________________________________ 21

APPENDIX A - Wiring Schematics_____________________________________________________ 23

APPENDIX B - Wiring Scheme Multiple Dosing Units ___________________________________ 24

APPENDIX C - Function of the Keys ___________________________________________________ 25

APPENDIX D - Technical Overview ___________________________________________________ 26

APPENDIX E - Quality Checklist ______________________________________________________ 27

APPENDIX F - Declaration of Conformity ______________________ Error! Bookmark not defined. 

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 3/27

1. GENERAL INFORMATION

Thank you for purchasing a Movacolor metering device. This manual is addressed to operatorsand specialized personnel taking care of the metering of dry additives. If you have any questionsabout our products, please contact your local agent or Movacolor.

1.1   Warranty* Movacolor guarantees all work/goods delivered by us to be free from defects in material or

workmanship under normal use and service.

* Movacolor’ obligation under this warranty being limited to making goods at it’s factoryany Movacolor product(s) which shall within three (3) years after delivery to the originalpurchaser be returned intact to Movacolor in the Netherlands.

* Our liability under this warranty will extend only to equipment that is returned to ourfactory PREPAID.

1.2  Safety

* Use the equipment only for what it is designed for. The metering of dry additives!

* Before switching on the unit for the first time, ensure that the main power voltage beingapplied is set on the voltage selector at the main power entry.

* Always switch off the Movacolor control cabinet and disconnect from electrical powerbefore performing maintenance.

* The symbol certifies that the machine conforms to the European Union regulations onminimum safety standards.

* Make sure all parts are securely fixed to your extruder or injection molding machine. 

3

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 4/27

2. METERING PRINCIPLE

Movacolor metering devices use the dosingcylinder®. This unique metering system combinedwith a very precise adjustable stepping motor gives an accurate and regular additive output.

Movacolor provides different neckpieces to suit your particular application. The most common

mounting of the neckpiece is in between the production machine and its hopper. The picturebelow shows a cut through of the standard type B neckpiece.

During operation the virgin material runs from the machine hopper through the neckpiece into themachine. Inside the neckpiece the virgin material flow is divided into two streams by the coverplate. In the space below the cover plate, the rotating cylinder is dosing additive.

Additive is added direct to the center of the virgin material flow, just before it enters theproduction machine. This is a great advantage over metering devices that use batch pre-mixingbecause pre-mixing can actual cause material separation. Separation of materials results in anirregular additive flow into the production machine.

4

1.  Neckpiece2.  Dosing cylinder3.  Cover plate4.  Virgin material5.  To production

machine 

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 5/27

3. DOSING CYLINDERS

Depending on the application a different dosing cylinder might be needed.Use the following table to roughly determine the best system for the application. For more detailedinformation please contact your agent or Movacolor.

Cylinder type Granular materials* Powder materials** Accuracy Max. output ***

G YES YES +25 kg/h

GL YES NO ++2 kg/h

H NO YES ++2 kg/h

A-21Feed screw

YES YES +/- 40 kg/h

A-30Feed screw

YES YES +/- 150 kg/h

* Granular material can either be normal or micro but has to be free flowing and non-static.

** Powder material must be free flowing, not sticky and non-static.

*** Max output is strongly relating to bulk density, the mentioned values are average.

Each dosing cylinder has a code, which indicates the type. 

G cylinders code H cylinders code

GN 4 HN 5GK 6 HK 7GNL 8 HBK 3

GBK 1 HS HGKL 2GKBL 0GS G 

5

Code

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 6/27

4. INSTALLATION

4.1 Transport

To protect the Movacolor unit against damage during transport, the unit is packed in a cartboardbox filled with polyurethane foam. In the event of damage to the unit, please contact your local

agent or Movacolor immediately. Delivery terms are Ex-Works Sneek, The Netherlands. Buyer isresponsible for the transport. Movacolor cannot be hold liable for any damage during transport.

4.2 Mechanical Installation

All mechanical parts are pre-assembled, making installation quick and simple.

1.  Disconnect the hopper assembly from weighing platform by loosening the screw at the sideand pulling the hopper assembly backwards.

2.  Install the neckpiece directly on top of the entrance of the production machine. Make surethat the complete unit is mounted horizontally and fixed securely. Handle the weighingplatform with care!

3.  Connect the hopper assembly back to the weighing platform. Also assure that theearthwire from the hopper loader hose is connected to the load frame.

4.  Assure proper earthing from the unit to the neckpiece, normally this is assured by makingthe connection with the bolds.

6

Hopper assembly

Weighing platform

Mount the unit in a 90°-degree angle to themachine barrel. As shownin the picture

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 7/27

4.3 Electrical Connections

The MC-Balance needs an input signal from the production machine in order to operate. 

Three different input signals can be used to control the MC-Balance.

1.  A potential free relay contact.

Use the white and brown wire for the potential free contact.

2.  A relay signal up to 24 Volt DC.Relay signals can be used for an injection molding machine or an extruder that has nottacho signal. In case of a powered relay signal connect the white wire to +24 VDC and theyellow wire to - side.

3.  A tacho signal up to 24 Volt DC.This is used when the MC-Balance needs to be connected to an extruder that has agenerator that produces a voltage linear to the extruder speed. When using a Tachogenerator signal, shorten the white and brown wire, connect green to +24 VDC and yellowto other contact of the generator. See Paragraph 5,5 for information about the configurationof the tacho parameters.

The only other electrical installation necessary is connecting the cables to the control cabinet.

See APPENDIX A and B  for wiring schematics.

7

CAUTIONMake sure the right voltage is applied tothe control cabinet. The control cabinet hasa sticker to indicate the voltage setting. 

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 8/27

5. OPERATION

5.1 The Interface

In this manual a button will be depicted as follows [name of button]. The software in the balance has an easy to use menu structure. To switch between the different

menus press [MENU]. The [<<] and [>>] can be used to browse the menu. Questions in the menusshould be answered with [YES] or [NO]. The numerical keyboard is used to enter other data. Tomove the cursor from one place to the other when entering data with the numerical keyboard us

the [ 

], [ 

], [→

] and [↓

] buttons. Whenever using the numerical keyboard press [E] to confirm thechanges. Use [C] to discard warning messages from the screen. [ON] is used to switch the motoron or in standby position and [OFF] to switch the motor off.

MC-Balance interface

Appendix C is a list of all buttons and button combinations and their functions.

8

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 9/27

9

5.2 Service Menu

To make the service menu available, keep [MENU] pressed while switching the main power on.All data entered in the Service menu is saved in the memory and remains in the memory evenwhen the unit is shut off or unplugged.

1.  Press [MENU] until the service menu appears.

2.  Press [>>] to enter the menu.

3.  Language code screen shows. Fill in the desired language code and press [E].

CODE LANGUAGE

01 English

02 Francais

03 Deutsch

04 Español

05 Nederlands

06 Português

07 Türkçe

08 Italiano

4.  The MC-Balance software version will appear. Press [>>].

5.  Factory default screen shows. Changing to factory defaults will reset the memory of the E-

EPROM. Answer [NO] unless you are sure to clear all previous setting.

6.  Filling System screen shows. Press [YES] or [NO] to enable or disable a filling system.

7.  If the previous question is answered with [YES] the filling system type has to be selected.Press [NO]  to switch between MV and ME system. Press [YES] to confirm one of thesystems. An extra filling system menu will be enabled.

8.  Motor configuration screen. Motor type 315 is automatically set. Press [YES]  or [>>]  tocontinue. When 30 mm feed screw is selected in the configuration menu, the unit willautomatically change to motor type 580.

10. Alarm connector output can be set. Press [NO] to switch between Alarm and Dosing. Press[YES] to confirm one of the options. When set to alarm, 24 Volt is applied to this connectorto power an external alarm signal. Dosing mode is only used if you demand a "motorrunning control signal". This signal is equivalent to the motor control LED on the front ofthe control cabinet and gives you a 24 Volt DC output signal.

11. Device address setting : In the case of using a PC to communicate with your MC-Balance,enter the communication address of your MC-Balance. Acknowledge the setting bypressing [E]. By default the address is set to 01.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 10/27

10

12. Fast calibration screen: In most circumstances you should answer this question with [No].Only in certain rare cases when working in injection molding mode where output isextremely low it should be answered with [YES]. If answered with [YES] an extra functionwill be activated in the production menu.This function will enable the user to perform a fast offline calibration. See Paragraph 5.8 fordetailed information about this function.

13. Switch main power off and on again to make the service menu unavailable to unauthorizedpersonnel.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 11/27

11

5.3 Calibration ProcedureAlways  perform a calibration after the machine is installed. To make the calibration menuavailable, keep [MENU] pressed while switching the main power on. For an accurate calibrationmake sure that the MC-Balance is powered on for at least 15 minutes and that the system is stableduring calibration. Perform the calibration with an empty hopper, the motor and lid in place.

1.  Press [MENU] until calibration menu appears.

2.  Press [>>] to enter the menu.

3.  The current weight is being displayed. Press [>>].

4.  Press [YES] to start calibration.

5.  After approximately 60 seconds the Start mass is fetched.

6.  Place the Movacolor reference weight of 500 grams on top of the hopper. Press [YES]  tostart the second part of the calibration procedure.

7.  After approximately 120 seconds the Calibration will be completed. Remove the referenceweight from the hopper.

8.  Your MC Balance is calibrated. Press [YES] to continue.

9.  Switch main power off and on again to make the calibration menu unavailable tounauthorized personnel.

You can check the unit for proper calibration with the weight check function.(see chapter 5.4)

NOTE. Movacolor advises to recalibrate the unit once every 6 months to maintain properoperation.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 12/27

12

5.4 Weight Check Menu:

With the weight check function you can check the correct function of the MC-Balance weighingsystem. It also gives you the opportunity to weigh objects up to 8 kilogram. This is useful todetermine a shot weight.

1.  Press [MENU] until the weight check menu appears.

2.  Press [>>] to enter the menu.

3.  Acknowledge that you want to weigh an object by pressing [YES].Please wait while the unit tare his weight.

4.  Place the object on the hopper or platform and the weight will be shown on the screen.The upper line shows the weight of the object.The line below shows the weight of the material inside the hopper.

5.  Press [MENU] to leave the weight check menu or [<<] to weigh another object.

NOTE. The reference weight result should be 500 grams ± 2. After removing the reference weight

the display should read 0 grams ± 2. If this is not the case calibration may be wrong or it indicatessome problem with the weighing platform. Refer to the troubleshooting section if needed.

5.5 Consumption Registration (optional)

The consumption menu will be visible when the ME or MV filling system is enabled in the servicemenu. Without use of the optional ME or MV filling system, accurate working of the consumption

registration is not supported.

With the Consumption function it is possible view the total quantity of material dosed by the MC-Balance. The consumption is saved in the memory and remains in the memory even when the unitis shut off or unplugged.

1.  Press [MENU] until the Consumption menu appears.

2.  Press [>>] to enter the Consumption menu. The total quantity of dosed material is shown.

3.  To reset the consumption registration to zero, press [>>] and answer the question with YES.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 13/27

5.5 Configuration Menu:Configuration data is normally entered once and it remains the same for subsequent operation.Configuration data can only be changed when the motor is switched off. A warning will bedisplayed when there is an attempt to change the configuration data when the motor is running. Just press [C] to discard the warning. All data entered in the Configuration menu is saved in thememory and remains in memory even when the unit is shut off or unplugged.

1.  Press [MENU] until the Configuration Menu appears.

2.  Press [>>] to enter the configuration menu.

3.  Select the type of cylinder used. Press [NO] to switch between the available options. Press[YES]  to confirm the selection. Choosing the right option lets the MC-Balance reach thedesired setpoint faster. See Chapter 3 for the available cylinder options.

4.  Select the type of material used. Press [NO] to switch between the available options. Press[YES] to confirm the selection.

5.  Set the type of production. Switch between Injection molding and Extrusion by pressing[NO]. Press [YES] to confirm the selection.

6.  Set the input signal mode. Switch between Relay, Timer and Tacho* by pressing [NO].Press [YES] to confirm the selection.If Relay is selected in the Injection Molding mode, the next screen will show the Inputsignal Filter. In case the relay-time of the Injection Molding machine is fluctuating, thisfunction will calculate the average relay time of the entered number of shots. The calculatedaverage relay-time will be automatic used as metering time. The average relay-time can becalculated of min. 3 shots and max. 50 shots. Advised is to use the (standard) factorydefault of 3 shots.

7.  Select the way the motor is controlled. Press [NO] to switch between RPM and Gravimetric.The gravimetric mode allows direct entry of the additive percentage that needs to be dosed.Calibration is not needed because the unit automatically adjusts its speed based on loss inweight information. This is the preferred mode. The RPM mode can be used when there issomething wrong with the weighing system.

8.  Press [YES] when ready to exit the configuration menu.

* Tacho If tacho is selected the next screen will show the tacho parameters. Volt DC represents the voltageat which the extruder has maximum output. Output represents the maximum output of the

extruder in kg/h. Set the values with the numerical keyboard and acknowledge with [E]. Themaximum voltage that can be applied to the MC-Balance is 24 volt DC. If the tacho generator has ahigher then 24 V DC voltage output at the maximum extruder output capacity, the tacho voltagehas to be reduced to 24 volt DC. See the diagram.

Rx (kilo-Ohm) = 3,7 x (Max. tacho output VDC – 2,5) – 80,6

If the extruder stops when connected to the metering unit, an isolated signal converter is needed.Contact Movacolor for more information.

13

I nputconnector

Resi st or ( Rx)C-pin +  TachogeneratorEXTRUDER

 A-pin -

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 14/27

14

5.6 Production Menu:

Production data entered in the Production menu can be altered when the motor is running. Alldata entered in the Production menu is saved in the memory and remains in memory even whenthe unit is shut off or unplugged.

1.  Press [MENU]  until the Production Menu appears. This screen also shows the type of

cylinder and material used. Check if this is the correct configuration.

2.  Press [>>] to enter the Production Menu.

3.  Fill in all necessary production data using the numerical keyboard and press [E]  toacknowledge the data. If the unit is configured to run extrusion with tacho input you haveto fill in the additive percentage and extruder output capacity. If configured to runextrusion with a relay input the extruder capacity and the percentage additive has to be set.If configured to run injection molding with a relay input the shot weight and the additivepercentage has to be set. It is not neccesary to set the time in relay mode unless the Fastcalibration function is used. When timer is used, the dosing time has to be set too.

4.  Switch on the motor by pressing [ON].

5.  The “Dosing cylinder filled?” question will appear if this is the first time [ON] is pressedafter the MC-Balance system has been reset. Filling the cylinder before a production startwill make the unit adjust to the desired setpoint faster.

6.  In case the system is waiting for an input signal the green LED flashes. If the LED is lightedcontinuously the stepping motor is running.

7.  Your MC-BALANCE is in operation now.

8.  Within a few minutes, actual production data is displayed in the production menu screens.

NOTE 1. It is advised to monitor the unit the first few minutes after a start-up to check properoperation.

NOTE 2. It is advised to switch the system [ON] when the extruder or injection molding machineis running at the desired speed or cycle time.

NOTE 3. Changing one parameter in the production menu will cause the balance to adjust to the

changes but it will not reset totally. Changing both production parameters within 10 seconds aftereach other will cause the Balance algorithm to reset. This is necessary for the system to adjustquickly to these big changes in the settings. Switching the power off and on will also cause thealgorithm to reset. Pressing [OFF] will only cause the algorithm to pause.

NOTE 4. A Capacity deviation alarm will occur when the system is not running withinspecification for some period of time. When a capacity deviation alarm occurs it can be stopped bypressing the [Function] followed by [1]. Check the troubleshooting section for causes and possiblesolutions.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 15/27

5.7 Function Hopper empty detection

The system will give an “Hopper empty” alarm if there is less then 700 grams of material in thehopper. For this system to work correctly it is necessary that calibration is done with an emptyhopper and the hopper lid in place. This system will always be active.

When a hopper empty alarm occurs it can be stopped by pressing [Function] followed by [1]. It isnot necessary to stop production. Make sure to refill the hopper to make the empty detectionactive again.

5.8 Function Fast Calibration:

In some circumstances where the dosing rate is extremely low, the fast calibration option may beuseful because it allows the MC-Balance to calibrate faster and without the need of the productionmachine producing products that are not within spec. The fast calibration procedure is as follows:

1.  Fast calibration should be enabled in the service menu.

2.  After “Dosing cylinder filled?” in the production menu, a fast calibration options appears.When using fast calibration in combination with relay mode, make sure that the dosingtime has been set in the production menu.

3.  Answer [YES] if you want to perform a fast calibration.

4.  The “Pull machine back for offline dosing. Ready? Y” question will appear in the screen. Itis advisable to do the calibration offline to prevent the dosing of additive into the

production machine during calibration. Place the unit at 90° on the weighing platform andpress [YES] when ready. The unit will start calibrating towards the desired output.

5.  When the calibration is ready, mount the dosing unit in the production position. And press[YES]  to continue. The unit will automatically start the production as soon as there is aninput signal.

15

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 16/27

5.9 Filling Systems (optional)

The balance can be used with a Movacolor MV or ME automatic filling system. The control cabinethas to be fitted with two extra connectors to connect the filling system in order to operate thesystem. Contact your local agent or Movacolor for more information about the filling systems. Toenable the filling system, select the system of choice in the service menu. A new filling system

menu will come available.

ME-system, pneumatic ejector hopper loader for dust free granules, configuration:

1.  Press [MENU] until the ME Menu appears.

2.  Press [>>] to enter the ME Menu.

3.  Set the “Fill Alarm” time. Use the numerical keyboard and press [E] to acknowledge.

4.  Set the “Fill time”. Use the numerical keyboard and press [E] to acknowledge. During this

time the ME system is transporting material.

5.  The “ME switched off during alarm?” question appears. Press [YES] or [NO].

6.  Switch the system off or on by pressing [YES] or [NO].

16

1

2

   P  n  e  u  m  a   t   i  c  v  a   l  v  e Ejector pipe

Output connector

Compressed airinletHose:Ø6x1mm

Compressedair outlet

Hose:ø32x3.9mmL=3mtr

Hose:

ø25x3.8mmL=63cm

ME coupling

Ø25x3.8 L=25mm inserted to makeconnection with ejector pipe.

System set-up and connections for MC-Balance

Compressed air supply: 4-8 bar [58-116 psi] Air consumption: 15-25  m3/h [250-417 L/min] 

Specifications

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 17/27

17

Description

The ME (Movacolor Ejector) system is designed to transport dust free masterbatches, regrind andother additives. The ME system uses an compressed air ejector to create suction and transport thematerial. All Movacolor hoppers come equipped with a connection pipe to which the materialtransport hose can be connected. Supplied with the ME system is a special hopper lid with aperforated plate sieve. The sieve is necessary to let air escape from the hopper. The ME system is

NOT suitable for free flowing POWDER alike materials.

General installation instructions

•  Use oil-free compressed air at 4 to 8 bar.

•  Before starting production fill the hopper at least to 1/3 of its capacity to prevent material frombeing blown through the dosing cylinder the first time the ME system starts.

•  With the MC-Balance (software version 2.6 and later) the hopper empty detection is based onthe hopper weight. In order for this system to work correctly the loadcell has to be calibratedcorrectly and with all hardware (hoses, lid , etc.) attached to an empty hopper.

•  Good grounding of the parts is required to ensure proper operation.

Setting filling time and filling alarm.

Filling time Function: When it is detected that the hopper is running empty the filling system is switched on.The filling system will start loading when the weight in the hopper is 800 grams or lessThe filling time determines the duration of filling after the level of 800 gram is reached.This means that if the filling time is set to zero the loading system is only switched on as long asthe weight of the hopper is less than 800 grams. When the filling time is set to 10 seconds theloading system will continue to transport material for the duration of 10 seconds after the level of800 gram of material is reached.

Setting: The filling time setting depends on several parameters. Advised is to start with 10 secondsand adjust the time if necessary.

Filling alarm Function: The function of the filling alarm is to warn the operator that the hopper is not beingfilled although the hopper loader system is switched on. An empty material drum or loss of airpressure are possible causes of a filling alarm.Setting: The filling alarm should be set 2 to 3 seconds longer than the filling time. 

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 18/27

18

MV-system, pneumatic vacuum hopper loader for powder like materials, configuration:

4.  Press [MENU] until the MV Menu appears.

5.  Press [>>] to enter the MV Menu.

6.  Set the “Vacuum time”. Use the numerical keyboard and press [E] to acknowledge.

7.  Set the “Empty time”. Use the numerical keyboard and press [E] to acknowledge. This isthe time needed for the cyclone to empty itself.

8.  Set the “Alarm after nr. of cycles” parameter. Use the numerical keyboard and press [E] toacknowledge.

9.  Switch the system off or on by pressing [YES] or [NO].

The “Hopper loader empty” alarm should be switched off by pressing [FUNCTION] followed by[1].

Refer to the manual of the filling system for installation instructions.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 19/27

19

6. SYSTEM PERFORMANCE

System performance can be characterized by the time it takes the unit to reach the desired setpoint,the accuracy of the setpoint and the regularity of the material output.

The algorithm is self-adjusting to the conditions and because the conditions vary, it cannot bepredicted how long it will take the unit to adjust itself and reach a setpoint with certainaccuracy. 

The following variables influence system performance:

1.  Material properties. Easy flowing, non-sticky and non-static material that comes in theform of small granules or powder can be dosed very accurate and regular. The accuracyand regularity of material output drops with increased granule size. However this is only aproblem with extreme low outputs.

2.  Periodical cleaning of the dosing cylinder and seals is necessary for proper operation.

3.  Extreme vibrations and shocks influence system performance noticeably. Normally thesystem will be able to compensate for vibrations and shocks.

4.  The MC-Balance algorithm needs a certain time to weigh material loss and adjust the RPMaccordingly. This time depends largely on the setpoint and the above mentioned twovariables. The table below shows minimum times that the unit needs to come, guaranteed,within 10% of the setpoint for different setpoints. The time mentioned is the time thatmotor is running and that the unit is metering additive. The system constantly adjusts itselfto reach the best possible accuracy for current conditions. Over time it can reach an

accuracy within±

 1%.

Setpoint in grams per second 10% time in seconds

0,2 147

0,5 57

0,8 36

5.  Under “normal” circumstances the unit will be more accurate than 10% after the firstadjustment of the RPM.

6.  Before the unit makes its first RPM adjustment it might be running already very close to the

desired setpoint because it uses a cylinder and material reference system to determine thefirst RPM setting. This accuracy however can not be guaranteed because materialproperties can vary substantially from material to material.

7.  With injection molding the shot to shot accuracy depends, besides the variables mentionedso far, on the shot time. If big and heavy granules have to be dosed in a very short time thiswill influence the shot accuracy and repeatability because one granule can make adifference of a few percent to the shot weight.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 20/27

20

8.  An unstable relay or tacho signal has a negative effect on the accuracy, repeatability andspeed of the system because it will adjust to these parameter changes.

9.  A long cycle time combined with low dosage per shot can result in a slow system.

Capacity deviation alarm!A Capacity alarm will occur when the system is not running within specification for some period.When a capacity deviation alarm occurs it can be stopped by pressing the [Function] followed by[1]. Check the troubleshooting section for causes and possible solutions.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 21/27

21

7. TROUBLESHOOTING 

 Problem : The balance does not come into specification or a capacity deviation alarm occurred .

Possible causes:

1.  Check if all cables are connected correctly.

2.  Check if the hopper assembly is fixed tightly to the weighing platform and that theneckpiece is fixed tightly to the production machine.

3.  Check if the dosing cylinder is tightly fixed to the motor shaft.

4.  Excessive build up of material on the dosing cylinder may have impact on proper dosing.To avoid this, be sure that the seals and dosing cylinder are clean.

5.  Check if there is no tension on the cables connected to the weighing platform.

6.  Use the weight check function with the reference weight to determine the correct functionof the weighing system.

7.  If the weight check gives the correct result, check if the material flow into the cylinder hasbeen blocked.

8.  Another cause may be an obstruction to the weighing system. Check if there is at least ≈ 1mm space between all the pointed screws and the loadcell frame or platform and thatthere is no material or dirt blocking the movement.

9.  In case of a water cooled neckpiece, check if there is material build up around the dosingcylinder and the water cooled pipe that can obstruct the free movement of weighingsystem.

10.. If none of the above seems to be the problem, recalibrate the system and try the weighcheck again.

 Problem : The balance does come into specification but seems to be slow.

Possible causes:

1.  Extreme vibrations and shocks to the system.

2.  Extreme low setpoint. See Chapter 6 : SYSTEM PERFORMANCE. 

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 22/27

22

Frequently asked questions  Question : The alarm is displayed on the screen but no alarm signal is given? E.g: Motor on

while trying to change configuration data.  Answer : Go into the Service Menu and switch Output mode to alarm.

  Question : The keyboard is locked. How to unlock the keyboard?  Answer : Press [Function], release and press [0] to unlock the keyboard.

  Question : How to make an alarm from the MC-BALANCE audible or visible?  Answer : Use the connector Alarm on the control cabinet to provide a signal of 24 V DC to any

device that can produce a visible or audible alarm.

  Question : The Configuration menu is not displayed. How to enable the configuration menu?  Answer : Press [Function] release this key and then press [6] to unlock the configuration menu.

  Question : How to disable the fast calibration option?  Answer : Change the setting in the service menu.

  Question : Help!  Answer : For all technical questions please contact your local agent, or Movacolor.

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 23/27

APPENDIX A - Wiring Schematics 

(optional) 

23

A = Br ownB = Yel l owD = Gr een

E = Whi t e

A = Yel l owC = Gr eenE = Br ownF = Whi t e

B = -E = +24 Vol t DC

B = -E = +24 Vol t DC

Alarm light(optional)

Output Valve(optional)

Input(Start signal)

Motor

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 24/27

APPENDIX B – Wiring Scheme Multiple Dosing Units 

24

Unit x:

Unit 3:

Unit 1:

Unit 2:

F = White

E = Brown

INPUT Wiring Scheme:

For connection of multiple dosing units to the Tachogenerator of an

Extruder.

C = Green

A = YellowTachogenerator

0-24 VDC !

F = White

E = Brown

C = Green

A = Yellow

F = White

E = Brown

C = Green

A = Yellow

F = White

E = Brown

C = Green

A = Yellow

+

INPUT Wiring Scheme:

For connection of multiple dosing units to an Injection Molding Machine

or Extruder use potential free relay contacts.

Unit 1:

E = Brown (+24V)

F = White ( 0V )

Unit 2:

E = Brown (+24V)

F = White ( 0V )

Unit 3:

E = Brown (+24V)

F = White ( 0V )

Unit x:

E = Brown (+24V)

F = White ( 0V )(max. 10 units)

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 25/27

25

APPENDIX C - Function of the Keys 

[Yes] [No] Yes and No

[Menu] Switches between the several Menu modes like Production,Configuration, Service or Filling systems.

[>>] [<<] Next-previous screen in the current menu.

[On] Motor running or in standby mode.

[Off] Motor Off

[Function] + [0] Keyboard lock, repeat for unlock

[Function] + [1] Reset Alarm.

[Function] + [4] Fill Dosing cylinder (10 seconds at 50 RPM)

[Function] + [6] Production Menu only on

[Function] + [7] Production Menu only off

[Function] + [8] Lock Configuration menu

[Function] + [9] Unlock Configuration menu

[C] Correction, RESET warning  on display.

[E] Entering of data

NOTE :[Function + key] means first press [Function], release this key and then press within one secondthe other key !!

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 26/27

 

APPENDIX D - Technical Overview 

General:Controls Digital control via keyboard with back-lid LCD display.

Motor Stepper-motor (brush less) max 2 A.Inputsignal(s)

Potential-free contacts (relay)24 VDC contacts (relay)Tacho generator input V DC (extruder)

Injection molding Setting of additive % and shot weight.Automatic metering time adjustment to injection cycle.

Extrusion Setting of additive % and “Tacho” synchronization

Communication RS-232 and 485 port

Hopper Removable, stainless steel 6-liter

Machine flange Standard type B.

Volt / frequency 110 Volt 60 Hertz – 220 Volt 50 Hertz

Weighing

Loadcell 20 kg max, automatic temperature compensated, IP-63

Weighing, electronics 20 bits resolution A/D converter with full digital filtering

Optional parts:

Filling systems ME for normal and micro pellets. (ejector system)MV for powder material and micro pellets. (vacuum system)

Machine flanges Water cooled type(s) for PET applicationsMulti-machine type(s)

Extra connectors Level sensor input.Alarm output (24 VDC)Extra output (24 VDC)

26

www.movacolor.com

TheNetherlands

SERIALNO.

TYPENO.

the BALANCEPatent nr:1015439

 

7/17/2019 MC Balance 2002 Manual English V3.0

http://slidepdf.com/reader/full/mc-balance-2002-manual-english-v30 27/27

APPENDIX E - Quality Checklist 

MOVACOLOR MC-BALANCE QUALITY CHECKLIST

Date of production : / / 2006Movacolor order number :

Controller serial number :Motor serial number :

Check original order with internal packing list.

Ordercode : …………………………

Cylinder check  Type G/GL/H/auger21/auger30  Check for cylinder wrench

Steppermotor check  Check motor rotation direction 

  Check cable and connectors 

  Check sealing rings (2x) 

Dosing house check  Check slide valve  Check fitting of hopper clamp  Check sealing ring

Hopper check  Check sightglass

  Check lid in combination with fasteners

Neckpiece check  Check sightglass  Check fitting of dosing house  Perform max capacity load cell test  Check fastening of all calibration pins and ‘ LOCKTITE ’ them

Controller check- Service menu configuration  Check software version: ……………  Check language: ……………  Check communication address: (01 =default)

 Functional test  Perform calibration  Perform functional test with motor  Check weight check menu for correct display of reference weight  Check output/alarm on controller

Complete unit check  Check appropriate mains cable ……………..  Check Voltage Selector setting ……………..

  Check input cable  Check appropriate users manual ……………


Recommended