MEASUREMENT & AUTOMATION MYOMED MYOGRAPH SYSTEM SET-UP GUIDE DOC-263 Measurement and Automation Set-up Guide Rev. 1.0 Copyright © 2012 All Rights Reserved Catamount Research and Development, Inc. P.O. Box 1100 St. Albans, Vermont 05478
Transcript
Page 1: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

MEASUREMENT & AUTOMATION

MYOMED MYOGRAPH SYSTEM

SET-UP GUIDE

DOC-263 Measurement and Automation Set-up Guide Rev. 1.0

Copyright © 2012 All Rights Reserved Catamount Research and Development, Inc. P.O. Box 1100 St. Albans, Vermont 05478

Page 2: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

ii

Table of Contents

Chapter 1 ............................................................................................................. 1

Overview................................................................................................................................. 1

Instructions ......................................................................................................................... 1

Appendix A ........................................................................................................ 11

Contact Information ............................................................................................................ 11

Page 3: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 1 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

CHAPTER 1

Overview

In the this guide you will find step-by-step guide using Measurement & Automation Explorer to

setup and troubleshoot Force Transducers in MYOMED Myograph System.

Instructions

1.) Locate the Measurement & Automation icon on your desk top.

Figure 1 - Measurement & Automation Icon

Page 4: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 2 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

2.) If there is no icon on your desktop, you may alternatively find the program in the Start

Menu, under Programs: National Instruments: Measurement & Automation.

Figure 2 – Alternative Program Location

Page 5: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 3 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

3.) Open Measurement & Automation. The Measurement & Automation Explorer window

will open.

Figure 3 – Measurement & Automation Explorer

Page 6: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 4 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

4.) Click on the Devices and Interfaces in the directory tree on the lefts side of the

Measurement & Automation Explorer .

Figure 4 – Devices and Interfaces

Page 7: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 5 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

5.) Continue down the directory tree and open Traditional NI-DAQ (Legacy) Devices folder.

Figure 5 – Traditional NI-DAQ (Legacy) Devices

6.) Select PCI-6023E (Device 1) to bring up the device options in the center explorer.

Figure 6 – PCI-6023E (Device 1)

Page 8: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 6 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

7.) Now, click Test Panels… tab located at the top of the explorer. The Test Panel Window

will now open. Before proceeding, confirm the Device Name in the test panel window is

PCI-6023E.

Figure 7 – Test Panels

Figure 8 – Confirm Device Name

Page 9: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 7 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

8.) Verify that the Analog Input tab is selected and Channel 0 is showing in the Channel

Drop Down Menu Box.

Figure 9 – Analog Input Tab

NOTE: Channel 0 in the Test Panel window corresponds to the Channel 1 force transducer on

the myograph. Likewise, Channel 1 in the Test Panel window corresponds to the Channel 2 force

transducer on the myograph. Also note that the units of the graph in the Test Panel window are

Volts and NOT grams. The purpose of this procedure is to test the viability of transducers and

not to calibrate them.

Page 10: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 8 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

9.) To test Channel 1 transducer, ensure sure Channel 0 is selected, and slightly tap on the

device. You should see a similar reading to this.

Figure 10 – Test Reading

Page 11: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 9 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

10.) Next, using the up arrow button in the Channel Drop Down Box select Channel 1

(Channel 2 on the myograph).

Figure 11 – Test Reading

11.) Lightly tap the Channel 2 transducer on the myograph to display a voltage reading the

Test Panel Window.

Figure 12 – Test Reading

Page 12: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 10 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

12.) Repeat these steps for Channels 0 through 7 in the Test Panel Window (Channels 1

through 8 on the myograph.

If you experience Clipping or No Signal at all, there is a problem with the transducer block itself

or the connectivity with the system.

Signal Clipping

Figure 13 – Test Reading

No signal

Figure 14 – Test Panels

Page 13: Measurement & Automation - Med Associates Inc. · CATAMOUNT RESEARCH AND DEVELOPMENT MEASUREMENTS & AUTOMATION ii Table of Contents Chapter 1..... 1

CAT AMO UNT RESE ARCH AN D DEVELOPME NT MEASURE ME NTS & AUTOM A TIO N

- 11 -

DOC-263 Rev 1.0 Copyright © 2012

Catamount Research and Development

APPENDIX A

Contact Information

Catamount Research and Development, Inc. P.O. Box 1100 St. Albans, Vermont 05478 www.catamountresearch.com 802.527.2364 [email protected]