ISSN 0267-9477
Journal of Analytical Atomic Spectrometry
0267-9477(2010)25:9;1-9
HOT PAPERNiemax et al.Measurement of element mass distributions in particle ensembles applying ICP-OES
EDITORIALHaraguchi and FurutaAnalytical atomic spectrometry in Japan over the last 25 years
www.rsc.org/jaas Volume 25 | Number 9 | September 2010 | Pages 1361–1492
Celebrating
25 years
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online / Journal Homepage / Table of Contents for this issue
PAPER www.rsc.org/jaas | Journal of Analytical Atomic Spectrometry
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
Measurement of element mass distributions in particle ensembles applyingICP-OES†
Ayrat Murtazin,a Sebastian Grohb and Kay Niemax*a
Received 7th April 2010, Accepted 25th May 2010
DOI: 10.1039/c004946h
End-on inductively coupled plasma optical emission spectrometry (ICP-OES) of the relative mass
distributions in ensembles of spherical SiO2 particles with small and large size variations is presented.
The particles are introduced into the ICP by injection of monodisperse microdroplets generated from
diluted particle suspensions. The spectroscopic signals, taking into account the peak or the integrated
intensities of the Si 288.16 nm line, are compared with mass distribution measurements of
corresponding SiO2 particle ensembles performed by Scanning Electron Microscopy (SEM). It is found
that the relative mass distributions measured by ICP-OES are in good agreement with SEM
measurements if the spectroscopic signal to noise ratio is sufficiently large indicating that intensity
variations due to different particle trajectories in the ICP are less important than expected.
Introduction
Recently, it was demonstrated in our laboratory1 that the
element masses of spherical SiO2 and Au particles in the micro-
and nanometre size range can be measured with good accuracy
by ICP-OES. The particles were introduced by monodisperse
micro-droplets into the plasma and the line emission of the
atomized particles was compared with the emission obtained
from droplets of the same size as the droplets with particles but
made from respective element standard solutions of known
concentrations. While the mean intensity ratios measured with
particles and standard solutions were in very good agreement
with the beforehand known element masses in the particles and
solution droplets, it was unclear whether this new chemical
particle characterization technique can be used for rapid analyses
of the mass distributions of particles or size distributions if the
particles have spherical shape.
In the recent paper,1 the peak as well as the integrated
emission intensities from the elements of the atomized particles
were evaluated. Both quantities which are proportional to the
particle mass showed intensity distributions which might not
only be due to the size distributions of the spherical particles but
also to possible differences of the particle trajectories in the ICP
and to the intensity noise if the signals are close to the detection
limit. To separate the different contributions, the mass distri-
butions of the particle ensembles of interest have to be known
with good statistical accuracy. In the present paper,
aDepartment of Analytical Chemistry and Reference Materials, FederalInstitute for Materials Research and Testing (BAM),Richard-Willst€atter-Strasse 11, 12489 Berlin, Germany. E-mail: [email protected] f€ur Analytische Wissenschaften—ISAS at the TechnischeUniversit€at Dortmund, Bunsen-Kirchhoff-Strasse 11, 44139 Dortmund,Germany
† Research has been performed within the network Plasma-AnalyteInteraction Working Group (PAIWG). PAIWG is a collaborativeeffort of the University of Florida (Gainesville, USA) and the FederalInstitute for Materials Research and Testing (BAM, Berlin, Germany),jointly funded by NSF and DFG.
This journal is ª The Royal Society of Chemistry 2010
representative ensembles of spherical SiO2 particles with good
but also suboptimal size distributions are measured by Scanning
Electron Microscopy (SEM). The mass distributions obtained
by SEM are then compared with the intensity distributions
measured with respective representative particle ensembles by
ICP-OES.
It has to be noted that the introduction of monodisperse
droplets of standard solution into flames and plasmas has
already a long history. It has been studied by different research
groups in the past. We like to refer here to the pioneering
investigations of Hieftje et al.,2,3 Olesik and Hobbs,4,5 and Lazar
and Farnsworth.6,7 More details on earlier publications are pre-
sented in our recent paper.1
Experiment
The spectroscopic arrangement (see Fig. 1) and experimental
details are described in detail in a recent paper of this journal.1
Therefore, we refer to that paper and present here only the most
important features of the experiment.
Monodisperse microdroplets are generated from particle
suspensions applying a commercial piezo-electric droplet
dispenser (Type MD-E-201 H with dispenser head MD-K-150,
microdrop Technologies, Norderstedt, Germany). The dispenser
head is equipped with a gas nozzle (dispenser gas flow) around
the 30 mm diameter orifice producing droplets with diameters of
about 50 mm for the chosen piezo-electric pulse parameters
(voltage and length). The variation of the diameter for successive
droplets is smaller than 1% as claimed by the manufacturer. For
visual control of droplet production, diameter and velocity, the
droplets are stroboscopically illuminated and measured with
a commercial microscope equipped with a CCD also delivered by
microdrop Technologies. The accuracy of the droplet diameter
measurements is about �3 mm, limited by the pixel size of the
CCD. The droplet generation rate was mostly 20 Hz. In very few
cases, it was 25 Hz and in one 50 Hz. Within the experimental
error there was no effect of the different droplet rates on the
measured line intensities as long as the dispenser gas flow was not
changed.
J. Anal. At. Spectrom., 2010, 25, 1395–1401 | 1395
Fig. 1 Experimental arrangement used for SiO2 particle characteriza-
tion and the time respective position dependent intensity of the Si 288.16
nm line.
Table 1 SiO2 particles provided by microparticles GmbH (Berlin) withRSD of the diameters measured with the Coulter Multisizer II
Product name Nominal diameter/mm RSD (%)
SiO2-F-L1516 1.06 7.6SiO2-F-L1513 1.11 4.4SiO2-F-B1060 1.16 4.2SiO2-F-L1280 1.55 2.3SiO2-F-L1561 2.06 2.2SiO2-82103 2.08 6.1
‡ Although the uncertainties of the particle diameters exceed the digitsafter the decimal points (see Table 1) they are not rounded for betterdiscrimination of the samples. Moreover, the diameters given arenominal diameters used by the production company.
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
Once the stability of droplet generation is established, the
droplets made from the particle suspensions are transported by
a dispenser gas flow of 0.21 L min�1 (Ar) directly through the
injector tube (diameter at the end: 1 mm) of the vertically
mounted ICP torch into the plasma. A silicone O-ring between
the dispenser head and the injector tube is applied to tighten the
units. The injector flow rate is lower than usually applied in ICP
spectrometry because the droplet transport is unstable for flow
rates larger than �0.3 L min�1. On the other hand, the residence
time of the droplets with particles in the ICP has to be as long as
possible so that atomization and plasma equilibration is
accomplished.8
The ICP with power supply unit was taken from a prototype
ICP-instrument (ELEMENT 1, Finnigan MAT, Bremen,
Germany). The operation conditions during the ICP-OES
measurements in Ar are as follows: 1000 W rf forward power, 16
L min�1 plasma gas flow rate, and 1 L min�1 auxiliary gas flow
rate. The spectral emission was collected end-on and focused
onto a quartz fiber guiding the radiation to a 1000 M SPEX
monochromator (Jobin-Yvon). The intensity at the mono-
chromator exit slit is detected by a fast photomultiplier (EMI
9789 QA). The signals were amplified by a Stanford Research
SR570 amplifier (time constant: 53 ms), digitized (National
Instruments DAQPad-6015), and processed.
The SiO2 particles were analyzed by tuning the Spex mono-
chromator to the Si(I) 288.16 nm emission line. Although it was
shown before that the line intensities measured with mono-
disperse microdroplet injection are very reproducible in time,1,8,9
1396 | J. Anal. At. Spectrom., 2010, 25, 1395–1401
the Ca(II) emission line at 393.36 nm was simultaneously recor-
ded in some of the experiments applying a second mono-
chromator,1 not shown in Fig. 1. A small amount of Ca solution
was added to the suspensions so that the Ca concentration was
�100 ng mL�1. The Ca(II) signals served as a marker for the
presence of droplets in the ICP and to monitor the transport
stability of the droplets.
The mass distributions of the SiO2 particle ensembles were
determined using a Quanta 200 F scanning electron microscope
(FEI). The evaluations of the SEM pictures regarding the particle
sizes were made applying the free scientific image processing
software ImageJ v1.42q.10 The SEM image was imported into the
program, the internal scale calibrated with the respective scales
of the SEM-pictures, and the threshold function of the program
applied to produce a binary black and white image. After
applying of the watershed function, the ‘‘Analyze Particles’’
option of the program was used for particle numbering and
displaying the outer contours of the particles. The particle shape
pictures were inspected at the computer screen and particles
excluded which obviously were fragments, attached to each other
(e.g., twin-particles), had artefacts due to the evaluation proce-
dure, or were not totally displayed because they were on the
brink of the SEM-picture. Fragments and twin-particles were in
particular observed in the particle suspensions with broad size
distributions. The volumes of the remaining particles were then
calculated assuming spherical symmetry. Finally, the Si mass
determined taking into account the SiO2 density of 1.85 g cm�3.
Samples
Diluted suspension of monodisperse SiO2 particles was prepared
from commercially available suspensions and from suspensions
which are not for sale because of relatively broad particle
distributions. All suspensions were provided by microparticles
GmbH, Berlin, Germany. The product names, the nominal
diameters, as well as the relative standard deviations (RSD) of
the diameters as quoted by the company are given in Table 1. The
particle size distributions were determined by the production
company applying a Coulter Multisizer II instrument.11 The SiO2
suspensions with 1.16, 1.55, and 2.06 mm particle diameters had
already been used in the preceding paper1 while the additional
1.06 mm, 1.11 mm, and 2.08 mm suspensions were new.‡ The 1.11
mm particles had a small size distribution (RSD: 4.4%) while the
This journal is ª The Royal Society of Chemistry 2010
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
distributions of the 1.06 mm and 2.08 mm particles were relatively
broad (RSDs: 7.6% and 6.1%, respectively).
The undiluted suspensions with known particle numbers were
put in an ultrasonic bath for about 30 min to minimize
agglomeration and sedimentation of particles. In the next step,
the suspensions were diluted with deionized water in order to
have on average less than 1 particle in each of the 50 mm droplet
produced by the droplet generator. After each dilution step and
before analysis the samples were again treated in the ultrasonic
bath for �20 min. Once the suspensions were properly diluted
and the generated droplets visually controlled with the micro-
scope, the droplet dispenser was directly attached to the injector
of the ICP torch so that the droplets could be transported by the
dispenser gas flow into the plasma.
Results and discussion
The probability of particle inclusion in the monodisperse droplets
ICP-OES measurements of size or particle mass distributions in
suspensions are only possible if the probability of finding more
than one particle in the droplet is practically zero. In the recent
Fig. 2 The probability of finding SiO2 particles in the monodisperse 50 mm d
number in the suspension to the number of droplets representing the total volu
the open are theoretical data expected from the experimentally prepared partic
the particle sizes, and the number of droplets injected to the ICP are given in
This journal is ª The Royal Society of Chemistry 2010
paper on micro- and nanoparticle characterization1 was already
mentioned that the probability of finding particles in mono-
disperse droplets is approximately given by Poisson statistics.
Here, the probability of finding none (x¼ 0), one (x¼ 1) or more
particles (x ¼ 2, 3,.) in the droplets is given by
PlðxÞ ¼lx
x!e�l;
where l is the ratio of the number of particles in a suspension to
the number of droplets representing the total volume of the
suspension. The precondition for application of Poisson statistics
is that the particles are uniformly distributed in the suspension,
i.e., that there is no particle agglomeration and precipitation.
In order to investigate the validity of Poisson statistics for SiO2
particles in the size range up to about 2 mm, suspensions of
different particle sizes and different l were prepared applying the
dilution and ultrasonic preparation procedure described in the
Experiment section. Fig. 2 shows the comparison between
measured (full symbols) and calculated probabilities (open
symbols) taking into account Poisson statistics and the prepared
particle concentrations. As can be seen, the agreement between
the experimental and theoretical data points is very good,
roplets for diluted suspensions characterized by l, the ratio of the particle
me of the suspension. The full symbols represent experimental data, while
le concentrations in the suspensions. The formula of the Poisson statistics,
the insets.
J. Anal. At. Spectrom., 2010, 25, 1395–1401 | 1397
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
supporting the assumption that Poisson statistics is valid for
suspensions with SiO2 particles, at least for particles in the size
range 0.83–2.06 mm. The numbers of the injected droplets are
given in the insets of Fig. 2. They vary between 713 and 2556
droplets.
Absolute and relative mass distributions of SiO2 particle
ensembles with narrow and broad size dispersions
Fig. 3 shows SEM exposures of two different ensembles of SiO2
particles with narrow and broad size distributions. This
Fig. 3 SEM pictures of 2.06 (top) and 2.08 mm (bottom) SiO2 particle
ensembles showing the different dispersions of the particle samples.
Fig. 4 Size evaluation of SEM exposures using the ImageJ v1.42q
software. The section was taken from the SEM picture of the 2.08 mm
SiO2 particles in Fig. 3.
1398 | J. Anal. At. Spectrom., 2010, 25, 1395–1401
difference can already be seen by visual inspection. While the
2.06 mm particles at the top of the figure are obviously quite
uniform in size, the 2.08 mm particles at the bottom have slightly
different diameters. Moreover, there are particle fractions and
several particles in the 2.08 mm ensemble which consist of two
particles which are glued together (twin-particles).
The size and finally the mass distributions of both classes of
particles were evaluated applying the image processing software
ImageJ v1.42q as described in the Experiment section. For
example, Fig. 4 displays the contours of 2.08 mm particles in four
sub-section of Fig. 3 (bottom). The contours which are used for
the calculation of the particle volume are result of the image
processing program as described above. The particles are
numbered because this allows discriminating individual particles
from evaluation. As can be seen in Fig. 3, the particles are
predominantly rounded. However, one can immediately observe
fragments (see, e.g., particle 790 and 814 in Fig. 4b and c,
respectively), twin particles (703/708 in Fig. 4a), or artefacts due
to the evaluation which appear as small protuberances (e.g., 698,
768, and 859 in Fig. 4a, b, and c, respectively) and incomplete
contours (95 in Fig. 4d). Fragments, twin-particles and particles
with incomplete contours were excluded from evaluation while
particles having contours with only small protuberances were
taken into account when the estimated errors in particle mass
determination were only small.
The Si mass distributions of 2.06 and 2.08 mm SiO2 particle
ensembles evaluated from SEM pictures are plotted in form of
histograms in the upper parts of Fig. 5 and 6, respectively. Three
sub-ensembles were individually evaluated for each particle sort.
Their individual data with the respective RSDs are given in
Table 2 together with the values for the total ensembles measured.
Histograms of the peak as well as of the integrated intensities
measured with 2.06 and 2.08 mm particle ensembles by ICP-OES
are presented in the lower parts of Fig. 5 and 6, respectively. It
can be seen that the shapes of the relative mass distributions
measured by OES are very similar to the absolute mass distri-
butions determined by SEM if we neglect the intensity distribu-
tions at low and high values. The peaks at �7 mA and �9 mA ms
in Fig. 5 and �9 mA and �8 mA ms in Fig. 6 are due to
contamination by smaller SiO2 particles which had been depos-
ited in the dispenser head from earlier measurements. In all later
measurements the dispenser head was also very carefully cleaned
before analysis to avoid particle contaminations from earlier
measurements. Therefore, peaks at lower intensities due to
contamination by other particle ensembles could not be observed
in later measurements (see, e.g., Fig. 7). In case of the 2.08 mm
particles the intensities at high values are due to twin particles of
different mass, while small fragment particles contribute to the
low intensities. The disregard of fragment signals as well as of
twin-particles is allowed in the comparison of ICP and SEM
measurements because both sorts of particles were not taken into
account in the evaluation of the SEM pictures. In addition to the
histograms in Fig. 5 and 6, the results of the ICP-OES
measurements are given in Table 3a and b, respectively. The
tables do not only present the results of the total ensembles
measured but also of the five sub-ensembles measured with 2.06
and 2.08 mm particles.
Comparing the data given in Tables 2 and 3, the relative
standard deviations of the mass distributions measured by SEM
This journal is ª The Royal Society of Chemistry 2010
Fig. 5 Comparison of the absolute Si mass distribution for the 2.06 mm SiO2 particles measured by SEM (top) and the relative mass distributions
derived from the peak as well as from the integrated intensities of the Si 288.16 nm line (bottom).
Fig. 6 Comparison of the absolute Si mass distribution for the 2.08 mm SiO2 particles measured by SEM (top) and the relative mass distributions
derived from the peak as well as from the integrated intensities of the Si 288.16 nm line (bottom).
This journal is ª The Royal Society of Chemistry 2010 J. Anal. At. Spectrom., 2010, 25, 1395–1401 | 1399
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
Table 2 Determination of the Si mass of (a) 2.06 mm, (b) 2.08 mm, and(c) 1.16 mm SiO2 particles evaluating three sub-ensembles each from SEMpictures
Nominaldiameter/mm No. Particles m(Si)/pg RSD (%)
(a) 2.06 1 489 4.17 8.22 551 4.18 6.63 807 4.07 7.5Total 1847 4.13 7.5
(b) 2.08 1 497 4.03 18.92 761 4.06 17.53 716 4.01 16.5Total 1974 4.07 17.6
(c) 1.16 1 945 0.83 10.92 895 0.81 10.63 1179 0.80 11.0Total 3019 0.81 10.9
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
and by ICP-OES are close together for both sorts of particles.
While the RSDs of the peak and integrated intensities were 7.6%
and 6.2% for the 2.06 mm SiO2 particles and 18.5% and 19.1% for
2.08 mm SiO2 particles, the RSDs of the mass measurements by
SEM were 7.5% and 17.6%, respectively.
Fig. 7 is presenting the results for the 1.16 mm SiO2 particle
suspension (see Table 1). Here, the agreement between the SEM
and ICP-OES distributions is not as good as for the larger 2.06
and 2.08 mm SiO2 particles shown in Fig. 5 and 6. The relative
mass distributions measured by ICP-OES are slightly broader
than the distribution evaluated from SEM pictures. Further-
more, the secondary peak structure at 9.5 pg in the SEM mass
plot, which was observed in all SEM evaluations of sub-ensem-
bles, is obviously smeared in both intensity plots. The relative
Fig. 7 Comparison of the absolute Si mass distribution for 1.16 mm SiO2 part
derived from the peak as well as from the integrated intensities of the Si 288.
1400 | J. Anal. At. Spectrom., 2010, 25, 1395–1401
standard deviations of the spectroscopic measurements were 17%
and 16% whereas a RSD of �11% was found for the SEM
distribution. Table 2c and Table 3c are presenting the results
obtained for sub- and total ensembles by SEM and ICP-OES,
respectively. The reason for the difference and for the absence of
the secondary mass peak in the spectroscopic measurements is
the relatively small signal to noise ratio (S/N ¼ 14) which
significantly broaden the distribution of the 1.16 mm particles. It
means that such small S/N cannot be neglected in the evaluation
of the relative mass distribution. In cases of the 2.06 and 2.08 mm
particles the S/N was about 130.
It is interesting to note that the RSD of the relative mass
distributions of the 2.06 and 2.08 mm particles obtained from the
spectroscopic peak intensity measurements (about 7.6% and
18.5%, respectively) are in reasonable agreement with the relative
standard deviations of the mass distribution which can be
calculated taking into account the RSDs of the diameters
measured by microparticles GmbH using the Coulter Multisizer
II instrument (6.6% and 18% for 2.06 and 2.08 mm, respectively).
However, the RSD of the mass distribution of 1.16 mm particles
calculated from the RSD of the diameter in Table 1 is �12.5%.
This value is significantly smaller than the RSD derived from the
intensity measurements (�17%, see Table 3c) but in rough
agreement with the RSD of the SEM measurement (�11%). As
pointed out above, the larger RSD of the spectroscopic
measurements is due to the contribution by the intensity noise.
The peak intensity measurements were taken for comparison
because they are generally more robust than measurements of the
integrated intensity at least for the evaluation of solid particles.
The reason is that the length of emission signals from particles is
sensitively dependent on the transport gas flow as shown
icle ensembles measured by SEM (top) and the relative mass distributions
16 nm line (bottom).
This journal is ª The Royal Society of Chemistry 2010
Table 3 The peak and integrated intensities of the Si 288.16 nm linemeasured by end-on ICP-OES of monodisperse microdroplets of verydiluted suspensions of (a) 2.06 mm, (b) 2.08 mm, and (c) 1.16 mm SiO2
particles
Nominaldiameter/mm Droplets Peaks
Peak intensity Integrated intensity
h/mA RSD (%) A/mA ms RSD (%)
(a) 2.06 25 613 911 55.0 6.5 70.6 5.525 575 553 53.7 7.5 71.0 6.428 250 303 51.1 8.4 71.9 7.425 550 116 52.4 7.7 72.1 6.413 088 15 52.4 10.1 70.0 9.6
118 076 1898 53.8 7.6 71.0 6.2(b) 2.08 10 205 524 53.1 20.5 43.6 20.4
11 940 1064 56.2 19.0 46.5 18.710 230 1078 57.5 17.4 50.4 17.610 345 999 57.5 17.8 51.2 18.210 250 604 55.4 17.7 48.9 18.152 970 4269 56.3 18.5 48.6 19.1
(c) 1.16 10 175 308 6.8 18 6.2 1510 070 293 7.3 16 5.5 1310 740 345 7.2 17 5.7 1510 045 305 6.7 16 6.6 1410 075 328 6.9 15 6.7 1451 105 1579 7.0 17 6.1 16
Publ
ishe
d on
18
June
201
0. D
ownl
oade
d by
Uni
vers
ity o
f V
irgi
nia
on 0
1/10
/201
3 08
:33:
28.
View Article Online
recently.8 Therefore, small changes affect the integrated intensity
measurements stronger than the peak intensity measurements.
This is also the reason why the RSDs of the integrated intensity
measurements of the 1.16 mm sub-ensembles were smaller than
the RSD of the total ensemble (see Table 3c). There was a much
larger scatter of the measured integrated intensities than of the
peak intensities for the 1.16 mm sub-ensembles. Similar condi-
tions were also found in the measurements of other particle
ensembles (see, e.g., the scatter of the peak and integrated
intensities of the 2.08 mm sub-ensembles in Table 3b).
On the basis of the present experimental investigation it can be
stated that ICP-OES with microdroplet introduction can be used
for rapid evaluation of the mass distribution of micro- and nano-
particles in liquids. The only requirements for such measurements
are (i) high particle dilutions to exclude signals from two particles
in a droplet, (ii) ultrasonic pretreatment for homogenization of
the suspension and to avoid particle agglomeration and sedi-
mentation, and (iii) sufficiently large S/N ratios of the measured
line intensities to neglect intensity noise.
It is interesting that the variation of the emission line intensity
due to different particle trajectories are obviously of minor
importance. This seems to be plausible because of the relative
large inertia of microdroplets which assures a reproducible and
straight gas transport into the ICP. The situation would be
This journal is ª The Royal Society of Chemistry 2010
certainly different if the SiO2 particles would be introduced as
airborne particles. In particular very small particles are sensitive
to turbulences in the gas transport leading to signal variations
because of difference of the particle trajectories in the plasma.
Conclusion
The comparison of absolute particle mass distribution deter-
mined by the evaluation of SEM images and relative mass
distributions measured by ICP-OES of monodisperse micro-
droplets including these particles revealed that the spectroscopic
method can be applied as long as the S/N ratio of the line
intensities is sufficiently high. Absolute mass distributions can be
given if the ICP-OES signals are calibrated by the line intensities
obtained from monodisperse droplets of the same size but made
from element standard solutions of known concentrations as has
been demonstrated recently.1 To what extend the analytical
figures of merit of the new spectroscopic technique for particle
mass measurements can be transferred to ICP-MS has to be
investigated. ICP-MS would allow characterizing element masses
of particles with diameters below 100 nm because of its higher
detection power compared with ICP-OES.
Acknowledgements
Project funding by the Deutsche Forschungsgemeinschaft
(DFG) is gratefully acknowledged. The authors like to thank Dr
Harald Fiedler of the microparticles GmbH in Berlin for
providing suspensions of SiO2 particles with exceptionally broad
particle size distributions and Dr Alex von Bohlen for the SEM
pictures of the SiO2 particles.
References
1 C. C. Garcia, S. Groh, A. Murtazin, V. Horvatic and K. Niemax,J. Anal. At. Spectrom., 2010, 25, 645.
2 C. B. Boss and G. M. Hieftje, Anal. Chem., 1979, 51, 1897.3 R. E. Russo, R. Withnell and G. M. Hieftje, Appl. Spectrosc., 1981,
35, 531.4 J. W. Olesik and S. E. Hobbs, Anal. Chem., 1994, 66, 3371.5 J. W. Olesik, Appl. Spectrosc., 1997, 51, A158.6 A. C. Lazar and P. B. Farnsworth, Appl. Spectrosc., 1997, 51, 617.7 A. C. Lazar and P. B. Farnsworth, Appl. Spectrosc., 1999, 53, 457.8 S. Groh, C. C. Garcia, A. Murtazin, V. Horvatic and K. Niemax,
Spectrochim. Acta, Part B, 2009, 64, 247.9 S. Groh, P. K. Diwakar, C. C. Garcia, A. Murtazin, D. W. Hahn and
K. Niemax, Anal. Chem., 2010, 82, 2568.10 http://rsbweb.nih.gov/ij/.11 http://www.beckmancoulter.com/coultercounter/product_multisizer3.
jsp.
J. Anal. At. Spectrom., 2010, 25, 1395–1401 | 1401