Measurement of Urine Cystine using Liquid Chromatography Tandem Mass Spectrometry Joanne E Wear, Brian G Keevil Department of Clinical Biochemistry Wythenshawe Hospital
Transcript
Page 1: Measurement of Urine Cystine using Liquid Chromatography ...

Measurement of Urine Cystine using Liquid

Chromatography Tandem Mass Spectrometry

Joanne E Wear, Brian G KeevilDepartment of Clinical

BiochemistryWythenshawe Hospital

Page 2: Measurement of Urine Cystine using Liquid Chromatography ...

Introduction• Cystine • Cystine transport• Cystinuria• Methods for the measurement of

cystine

Page 3: Measurement of Urine Cystine using Liquid Chromatography ...

Introduction cont.

• Measurement by HPLC-MS/MS– Sample preparation– HPLC– MS/MS – Assay characteristics

• Conclusion

Page 4: Measurement of Urine Cystine using Liquid Chromatography ...

Cystine• Cystine is a disulphide-linked homodimer of

cysteineCOOH-CHNH2-CH2-SH

COOH-CHNH2-CH2-S-S-CH2-CHNH2-COOH • Essential component of glutathione • Excreted via the kidneys• Reabsorbed at renal tubules using a sodium

independent heteromeric amino acid transporter which consists of a heavy subunit rBAT and a light subunit b0,+AT

Page 5: Measurement of Urine Cystine using Liquid Chromatography ...

N

N

C

S S

b0,+ transport system

Page 6: Measurement of Urine Cystine using Liquid Chromatography ...

Amino acid transport• The b0,+ transporter expressed on the cell

membrane in the epithelium of the renal tubules and brush border of the small intestine

• The dibasic amino acids lysine, arginine and ornithine are also transported in exchange for neutral amino acids

• Sequential mechanism – ternary complex with substrate bound at either side of membrane

Page 7: Measurement of Urine Cystine using Liquid Chromatography ...

Cystinuria• Autosomal recessive disease• Excessive excretion of cystine in the

urine• recurrent urolithiasis• Progressive renal failure• Incidence 1:7000

– Heterozygotes 1:20-200– Homozygotes 1:15000 (US)

Page 8: Measurement of Urine Cystine using Liquid Chromatography ...

Cystinuria cont.• Caused by defect in b0,+ system • Classification based on genotyping• Mutation in either SLC3A1 (rBAT) or

SLC7A9 (b0,+AT)

Cystinuria Subtype

Mutation

A Mutation in both copies of SLC3A1 B Mutation in both copies of SLC7A9

AB Mutation in one copy each of SLC3A1 and SLC7A9

Page 9: Measurement of Urine Cystine using Liquid Chromatography ...

Cystinuria cont.

• Cystinuria diagnosed when urine cystine >100 mg/day*

• Dibasic amino acids measured as 2nd line test

• Treatment – large fluid intake, urine alkalinisation (pH 7.5-8), thiol agents e.g. captopril

*Morton AR, Iliescu EA, Wilson JWL (2002). Nephrology :1. Investigation and treatment of recurrent kidney stones. CMAJ 166 (2) p213-218.

Page 10: Measurement of Urine Cystine using Liquid Chromatography ...

Methods for measurement of cystine

• Qualitative cyanide-nitroprusside method

• Thin layer chromatography• Reverse phase HPLC (cysteine)• GC with flame photometric detection• LC-MS - qualitative• Proton NMR

Page 11: Measurement of Urine Cystine using Liquid Chromatography ...

Sample preparation

• 10 µL centrifuged sample, standard or QC

• 30 µL d4 cystine internal standard (50 mg/L)

• 500 µL distilled water

Page 12: Measurement of Urine Cystine using Liquid Chromatography ...

HPLC

• Waters 2795 Alliance HT LC system• Phenomenex SecurityGuard 4 x 3

mm SCX column + Waters Atlantis C18 column 50 x 2.1 mm, 5 µm

• Flow rate 0.3 mL/minute• Elution time 0.89 minutes• Injection-to -injection time ~3.5

minutes

Page 13: Measurement of Urine Cystine using Liquid Chromatography ...

•Step-wise gradientHPLC cont.

A = 2 mmol/L Amm Ac, 0.1% (v/v) formic acidB = 2 mmol/L Amm Ac, 0.1% (v/v) formic acid in MeOHC = 100 mmol/L Amm Ac, 0.1% (v/v) formic acid

Mobile phase (% v/v) Time (minutes) A B C

0 75 5 20 0.2 5 5 90 0.4 75 5 20

Page 14: Measurement of Urine Cystine using Liquid Chromatography ...

Mass Spectrometry• Waters Quattro Micro Tandem

mass spectrometer with Z spray source

•ES+ mode, column temperature 50°C, desolvation gas flow 630 L/hour

Page 15: Measurement of Urine Cystine using Liquid Chromatography ...

Mass spectra for cystineParent ion

Daughter ion

Page 16: Measurement of Urine Cystine using Liquid Chromatography ...

Mass spectra for d4 cystine

Parent ion

Daughter ion

Page 17: Measurement of Urine Cystine using Liquid Chromatography ...

Ion suppresssiond4 cystine

Cystine

Page 18: Measurement of Urine Cystine using Liquid Chromatography ...

LLOQ

• Analyte response at least 5 times that of blank*

• Analyte peak response identifiable, discrete, reproducible with precision of 20%, accuracy of 80-120%

• LLOQ 10 mg/L, with CV of 2%, bias 115%.

*Guidelines for industry: Bioanalytical Method Validation. U.S. Department of Health and Human Services,Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CVM). May 2001

Page 19: Measurement of Urine Cystine using Liquid Chromatography ...

Chromatogram

Internal Standard(d4 cystine)

Cystine

Page 20: Measurement of Urine Cystine using Liquid Chromatography ...

Linearity

0

200

400

600

800

1000

0 200 400 600 800 1000

Concentration (mg/L)

LC

MS

/MS

re

spo

nse

• r2 = 0.9998, y = 1.0236x - 1.374

Page 21: Measurement of Urine Cystine using Liquid Chromatography ...

Precision and accuracy

• Determined using a minimum of 5 determinations per concentration level.

• Mean within 15% of theoretical value, CV <15%

Page 22: Measurement of Urine Cystine using Liquid Chromatography ...

Within batch precision and accuracy

Page 23: Measurement of Urine Cystine using Liquid Chromatography ...

Between batch precision and accuracy

Page 24: Measurement of Urine Cystine using Liquid Chromatography ...

Recovery and stability

• Mean recovery 103% (91-118%, n=9)• Bias after 5 days at room

temperature <7.5% (n=8)• Bias after 5 freeze-thaw cycles <10%

(n=8)• Post preparative stability measured -

bias <5% after 24 hours (n=9)

Page 25: Measurement of Urine Cystine using Liquid Chromatography ...

Stability of a single extract over 17 hours

0.6

0.7

0.8

0.9

1

0 50 100 150 200 250 300

Injection number

Pe

ak

are

a r

ati

o

Page 26: Measurement of Urine Cystine using Liquid Chromatography ...

Reference Interval• Normal values below 79 mg/day

(below 95th percentile)

Page 27: Measurement of Urine Cystine using Liquid Chromatography ...

Conclusion

• Simple, robust method for the measurement of cystine

• No derivitisation• Quick sample preparation• High throughput