Measurements of US stave using ESPI and capacitive sensors 30 th August 2012
Transcript
Page 1: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Measurements of US stave using ESPI and capacitive sensors

30th August 2012

Page 2: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

• Three main tests carried out:

1. Vibrational modes (ESPI)2. Cool down measurements Z direction (Capacitive sensors)3. Cool down measurements X direction (Capacitive sensors)

Page 3: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

US stave mounted in environmental chamber.

Invar panelCapacitive sensors

Cooling pipes

Markers showing positions of mounting brackets

End mount brackets

Page 4: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Isotropic View of the US stave on the Invar Panel

US stave

US Brackets

Fully Fixed Mounting fixture

2 Dimensional Fixed Mounting Fixture

Page 5: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

End View

Fully Fixed Mounting2 Dimensional Fixed Mounting

Page 6: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Fixed in 2 planes

L-Spacer

V groove spacer

The idea of constraining the 2nd fixed point which constrain in x and z directions only

Page 7: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Fixed point at x, y and z planes Fixed point at 2 planes

Page 8: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Constraint –ve X and Z planes

Constraint +ve X plane only

Page 9: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Vibrational modes (60 Hz)

Position of central bracket

Measurements of vibrational modes using ESPI (Room temperature)

Page 10: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Vibrational modes (119Hz)

Position of central bracket

Page 11: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Position of central bracket

Vibrational modes (185 Hz)

Page 12: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Cool down test measurements in Z direction using capacitive sensors

Schematic of US stave showing locations of capacitive sensors

Page 13: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012
Page 14: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012
Page 15: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100105110115120125130135

-20.0

0.0

20.0

40.0

60.0

80.0

100.0

120.0

140.0

160.0

180.0

200.0

Graph of Z plane data 20/8/2012 Sensor positions G,H and I

Sensor GSensor HSensor I

Time (Minutes)

Dist

ance

Mov

ed (u

m)

Page 16: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Conclusion: In the Z direction during cool down the whole stave is moving away from the sensors and it is also tilting more along the top edge. It is tending to tilt more along the right hand side than the left hand side.

Page 17: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

In order to counteract the movement of the right hand side of the stave in Z direction, blocks were used to minimise the amount of movement allowed, free movement in X direction maintained.

Page 18: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210

-40.0-30.0-20.0-10.0

0.010.020.030.040.050.060.070.080.090.0

100.0110.0120.0130.0140.0150.0160.0170.0180.0

Cool down test US stave 22 Aug 2012 (Right hand side of stave restricted in Z direction movement)

Sensor 1

Sensor 2

Sensor 3

Sensor 4

Sensor 5

Sensor 6

Time (Minutes)

Dis

tanc

e m

oved

(um

)

Sensor 1 = ASensor 2 = CSensor 3 = DSensor 4 = FSensor 5 = GSensor 6 = I

Page 19: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Schematic diagram of US stave showing capacitive sensor positions for x plane movement.

Bracket

Cooling pipe

Capacitive sensor

Sensor XSensor YSensor Z

Cool down test measurements in X direction using capacitive sensors

Page 20: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

-250.0-240.0-230.0-220.0-210.0-200.0-190.0-180.0-170.0-160.0-150.0-140.0-130.0-120.0-110.0-100.0

-90.0-80.0-70.0-60.0-50.0-40.0-30.0-20.0-10.0

0.010.020.030.0

Cool down test US stave 28 Aug 2012. X direction movement.

Sensor XSensor YSensor Z

Time (Minutes)

Dist

ance

mov

ed (u

m)

Page 21: Measurements of US stave using ESPI and capacitive sensors 30 th  August 2012

Provisional Conclusions

1. Normal modes don’t exhibit nodes at bracket locations.

2. Stave movement highly dependent on end supports.


Recommended