Medrad™ Intego Vial Shield Reference This document is not intended to replace the Medrad™ Intego Operation Manual. The Operation Manual should be considered the main source of information on the system and should be read by a user before operation. NOTE: Radiation Shielding performance is achieved by using the vial shield designed by Bayer (or its equivalent). WARNING: Patient injury may result from using vials with other than a 20 mm diameter cap. Use of vials with other than a 20 mm diameter cap may create a particulate hazard. WARNING: No modification of this equipment is allowed. WARNING: Biological contamination may occur if aseptic techniques are not followed. Follow facility guidelines to disinfect the vial shield. WARNING: Radiation Exposure Hazard. The Medrad™ Intego PET Infusion System Vial Shields must be transported in a container that complies with applicable transportation regulations. CAUTION: If using an absorbent disc, place no more than one disc in the vial shield before installing an RP vial. Additional absorbent discs may result in improper fit and potential damage to the vial. CAUTION: If using an absorbent disc, do not expose the absorbent disc to cleaning solutions or other fluids. Discard and replace the absorbent disc if it has been exposed to any fluid. Failure to replace the absorbent disc that has been exposed to fluid may result in improper fit and potential damage to the vial. CAUTION: If using an absorbent disc, discard and replace the absorbent disc after each use. Failure to replace the absorbent disc may result in improper fit and damage the vial. CAUTION: Do not place labels, tags, or other materials on the exterior of the vial shield surface. Application of labels, tags, or other materials may cause the vial shield to fit incorrectly in the needle insertion device and result in improper fit and functioning of the system. CAUTION: Ensure that the needle insertion guide is clean and free of debris. Debris may cause the vial shield to fit incorrectly in the needle insertion device and result in improper fit and functioning of the system. NOTE: Table 1 represents the vials and compatible vial shiels for use with the Medrad™ Intego PET Infusion System. For additional vials not specified in this table, please contact Bayer Service.
Transcript
Page 1: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Medrad™ Intego Vial Shield ReferenceThis document is not intended to replace the Medrad™ Intego Operation Manual. The Operation Manual should be considered the main source of information on the system and should be read by a user before operation.

NOTE: Radiation Shielding performance is achieved by using the vial shield designed by Bayer (or its equivalent).

WARNING: Patient injury may result from using vials with other than a 20 mm diameter cap. Use of vials with other than a 20 mm diameter cap may create a particulate hazard.

WARNING: No modification of this equipment is allowed.

WARNING: Biological contamination may occur if aseptic techniques are not followed. Follow facility guidelines to disinfect the vial shield.

WARNING: Radiation Exposure Hazard. The Medrad™ Intego PET Infusion System Vial Shields must be transported in a container that complies with applicable transportation regulations.

CAUTION: If using an absorbent disc, place no more than one disc in the vial shield before installing an RP vial. Additional absorbent discs may result in improper fit and potential damage to the vial.

CAUTION: If using an absorbent disc, do not expose the absorbent disc to cleaning solutions or other fluids. Discard and replace the absorbent disc if it has been exposed to any fluid. Failure to replace the absorbent disc that has been exposed to fluid may result in improper fit and potential damage to the vial.

CAUTION: If using an absorbent disc, discard and replace the absorbent disc after each use. Failure to replace the absorbent disc may result in improper fit and damage the vial.

CAUTION: Do not place labels, tags, or other materials on the exterior of the vial shield surface. Application of labels, tags, or other materials may cause the vial shield to fit incorrectly in the needle insertion device and result in improper fit and functioning of the system.

CAUTION: Ensure that the needle insertion guide is clean and free of debris. Debris may cause the vial shield to fit incorrectly in the needle insertion device and result in improper fit and functioning of the system.

NOTE: Table 1 represents the vials and compatible vial shiels for use with the Medrad™ Intego PET Infusion System. For additional vials not specified in this table, please contact Bayer Service.

Page 2: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20 mm 3038383 (18 lbs) (1.417 ± 0.020”) (2.472 ± 0.027”) (0.787”)

INT VS BLU 84135054 Blue 8.6 kg 20-25 mL 20R 30 ± 0.25 mm 55 ± 0.7 mm 20 mm 3038384 (18 lbs) (1.181 ± 0.009”) (2.165 ± 0.027”) (0.787”)

INT VS GLD 84194174 Gold 8.6 kg 20-25 mL 20H 32 ± 0.45 mm 58 ± 0.6 mm 20 mm 3038385 (18 lbs) (1.259 ± 0.017”) (2.285 ± 0.023”) (0.787”)

INT VS GRN 84193992 Green 8.6 kg 10-15 mL 10R 24 ± 0.2 mm 45 ± 0.5 mm 20 mm 3038382 (18 lbs) (0.944 ± 0.007”) (1.771 ± 0.020”) (0.787”)

INT VS BRN 84194085 Brown 8.6 kg 10-15 mL 15R 24 ± 0.25 mm 60 ± 0.5 mm 20 mm 3038386 (18 lbs) (0.944 ± 0.009”) (2.362 ± 0.020”) (0.787”)

25 ± 0.25 mm ----- (0.984 ± 0.009”)

INT VS BLK 84180211 Black 8.6 kg 10-15 mL 10H 25.4 ± 0.4 mm 53.5 ± 0.5 mm 20 mm 3038387 (18 lbs) (1.944 ± 0.009”) (2.106 ± 0.020”) (0.787”) 25.5 ± 0.25 mm ----- (1.003 ± 0.009”) 21.25 ± 0.2 mm 51 ± 0.5 mm

84179876 (0.837 ± 0.007”) (2.008 ± 0.020”)

INT VS VLT 3037589

Violet 8.6 kg 10 mL -----

21.75 ± 0.25 mm 52 ± 0.5 mm 20 mm

(18 lbs)

(0.856 ± 0.009”) (2.047 ± 0.020”) (0.787”)

Table 1: Vial Shield and Vial Specifications - Summary

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Page 3: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Table 2: Vial Shield and Vial Specifications - Silver

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20 mm 3038383 (18 lbs) (1.417 ± 0.020”) (2.472 ± 0.027”) (0.787”)

Page 4: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Table 3: Vial Shield and Vial Specifications - Blue

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

INT VS BLU 84135054 Blue 8.6 kg 20-25 mL 20R 30 ± 0.25 mm 55 ± 0.7 mm 20 mm 3038384 (18 lbs) (1.181 ± 0.009”) (2.165 ± 0.027”) (0.787”)

Page 5: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Table 4: Vial Shield and Vial Specifications - Gold

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

INT VS GLD 84194174 Gold 8.6 kg 20-25 mL 20H 32 ± 0.45 mm 58 ± 0.6 mm 20 mm 3038385 (18 lbs) (1.259 ± 0.017”) (2.285 ± 0.023”) (0.787”)

Page 6: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Table 5: Vial Shield and Vial Specifications - Green

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

INT VS GRN 84193992 Green 8.6 kg 10-15 mL 10R 24 ± 0.2 mm 45 ± 0.5 mm 20 mm 3038382 (18 lbs) (0.944 ± 0.007”) (1.771 ± 0.020”) (0.787”)

Page 7: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Table 6: Vial Shield and Vial Specifications - Brown

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

INT VS BRN 84194085 Brown 8.6 kg 10-15 mL 15R 24 ± 0.25 mm 60 ± 0.5 mm 20 mm 3038386 (18 lbs) (0.944 ± 0.009”) (2.362 ± 0.020”) (0.787”)

Page 8: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

25 ± 0.25 mm ----- (0.984 ± 0.009”)

INT VS BLK 84180211 Black 8.6 kg 10-15 mL 10H 25.4 ± 0.4 mm 53.5 ± 0.5 mm 20 mm 3038387 (18 lbs) (1.944 ± 0.009”) (2.106 ± 0.020”) (0.787”) 25.5 ± 0.25 mm ----- (1.003 ± 0.009”)

Table 7: Vial Shield and Vial Specifications - Black

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Page 9: Medrad™ Intego Vial Shield Reference - Bayer in Radiology · PDF fileISO 8362-1:2003 ISO 8362-4:2003 INT VS SLV 84136212 Silver 8.6 kg 30-35 mL 30H 36 ± 0.5 mm 62.8 ± 0.7 mm 20

Table 8: Vial Shield and Vial Specifications - Violet

Source: Medrad™ Intego PET Infusion System operator manual – Appendix D

Vial Shield Vial(s)

Catalog Bayer Part Collar Weight Volume Size Body Diameter Glass Height CapNumber Number Color Range Specification Diameter ISO 8362-1:2003 ISO 8362-4:2003

21.25 ± 0.2 mm 51 ± 0.5 mm

84179876 (0.837 ± 0.007”) (2.008 ± 0.020”)

INT VS VLT 3037589

Violet 8.6 kg 10 mL -----

21.75 ± 0.25 mm 52 ± 0.5 mm 20 mm

(18 lbs)

(0.856 ± 0.009”) (2.047 ± 0.020”) (0.787”)

20.0 mm (0.787”)20,0 mm (0,787”)

52.0 mm 0.5 mm(2.047” 0.020”)

±±

±±

52,0 mm 0,5 mm(2,047” 0,020”)

21.75 mm 0.25 mm (0.856 0.009”)± ±± ±21,75 mm 0,25 mm (0,856 0,009”)

21.25 mm 0.2 mm (0.837 0.007”)± ±± ±21,25 mm 0,2 mm (0,837 0,007”)

20.0 mm (0.787”)20,0 mm (0,787”)

51.0 mm 0.5 mm(2.008” 0.020”)

±±

±±

51,0 mm 0,5 mm(2,008” 0,020”)

RadiologyBayer HealthCare Medical Care

Bayer HealthCare LLC 100 Bayer BoulevardP.O. Box 915Whippany, NJ 07981

www.radiologysolutions.bayer.com

Bayer Medical Care Inc. 1 Bayer DriveIndianola, PA 15051-0780 USA

Customer Service/Orders1-800-633-7231

Customer Service Fax1-412-767-4120

Bayer reserves the right to modify the specifications and features described herein, or discontinue manufacture of the product described at any time without prior notice or obligation. Please contact your authorized Bayer HealthCare representative for the most current information.

Bayer, the Bayer Cross, and Medrad™ are trademarks of the Bayer group of companies.

© 2014-2015 Bayer

PP-INT-US-0005 May 2015


Recommended