  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT ALBUMIN CATLOG BT 10.002.00 METODOLOGY COLORIMETRIC DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME ALBUMIN TYPE FINAL POINT FILTER 620 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 . CONTROL - 1 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT ALPHA-AMYLASE CATLOG BT 11.001.00 METODOLOGY KINETIC COLORIMETRICDETERMINATIONS 60

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME AMY TYPE KINETIC FILTER 405 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 6950 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.800 ABS HIGH 2.300 . CONTROL - 1 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT ALKALINE PHOSPHATASE

    CATLOG BT 11.005.00 METODOLOGY KINETIC COLORIMETRICDETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME ALP TYPE KINETIC FILTER 405 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 2757 SAMPLE BLANK NO RETARD 60 READING TIME 60 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT ALT/TGP CATLOG BT 11.008.00 METODOLOGY KINETIC UV DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME ALT/TGP TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 1746 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.800 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT AST/TGO CATLOG BT 11.007.00 METODOLOGY KINETIC UV DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME AST/TGO TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 1746 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.800 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT BILIRUBIN DIRECT CATLOG BT 10.003.00 METODOLOGY COLORIMETRIC DETERMINATIONS 250

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME BDI TYPE FINAL POINT FILTER 546 TEMPERATURE 25 UNIT mg/dL REAG BLANK NO ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.220

    ............- ABS HIGH 0.330 FACTOR 1.0 SAMPLE BLANK SI RETARD 5 READING TIME 2 REACTION VOL 900 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT BILIRUBIN TOTAL CATLOG BT 10.003.00 METODOLOGY COLORIMETRIC DETERMINATIONS 250

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME BTO TYPE FINAL POINT FILTER 546 TEMPERATURE 25 UNIT mg/dL REAG BLANK NO ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.220

    ............- ABS HIGH 0.330 FACTOR 1.0 SAMPLE BLANK SI RETARD 5 READING TIME 2 REACTION VOL 900 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CLCIUM ASX CATLOG BT 12.002.00 METODOLOGY ARSENAZO III DETERMINATIONS 200

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CALCIUM TYPE FINAL POINT FILTER 620 TEMPERATURE 25 UNIT mg/dL REAG BLANK Si ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.220

    ............- ABS HIGH 0.330 FACTOR 1.0 SAMPLE BLANK NO RETARD 5 READING TIME 2 REACTION VOL 900 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CLCIUM CATLOG BT 12.001.00 METODOLOGY CPC DETERMINATIONS 200

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix the Reagent A + Reagent B in a proportion 1:1 and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CALCIUM TYPE FINAL POINT FILTER 578 TEMPERATURE 25 UNIT mg/dL REAG BLANK Si ABS - LOW STANDARD

    ABS - HIGH 1 CALIB - NAME @ ............- NR 1 ............- CONC 0.000 ............- REP 2.300 ............- ABS LOW 1.00 ............- ABS HIGH No FACTOR 10 SAMPLE BLANK 2 RETARD 500 READING TIME 1 REACTION VOL * REPLICATES * REF. LOW 0.000 REF. HIGH 2.300 . * CONTROL - 1 * LOT * TARGET VALUE * LOW LIMIT HIGH LIMIT

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CK NAC CATLOG BT 11.002.00 METODOLOGY KINETIC UV DETERMINATIONS 40

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CK NAC TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 4127 SAMPLE BLANK NO RETARD 5 READING TIME 30 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CK MB CATLOG BT 11.003.00 METODOLOGY KINETIC UV DETERMINATIONS 40

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CK MB TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 8254 SAMPLE BLANK NO RETARD 5 READING TIME 120 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CHLORIDE CATLOG BT 12.003.00 METODOLOGY COLORIMTRIC DETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CHLORIDE TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mEq/L REAG BLANK Si ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1.00 SAMPLE BLANK NO RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CHOLESTEROL CATLOG BT 10.004.00 METODOLOGY ENZIMTIC

    COLORIMTRIC DETERMINATIONS 1000

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CHOL TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CHOLESTEROL HDL CATLOG BT 10.005.00 METODOLOGY PRECIPITATION +

    ENZIMATIC COLORIMTRIC

    DETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME HDL CHOL TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT CREATININE CATLOG BT 10.007.00 METODOLOGY COLORIMTRIC DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME CRE TYPE TWO POINTS FILTER 505 TEMPERATURE 37 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 12 READING TIME 60 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT GAMMA GT CATLOG BT 11.006.00 METODOLOGY KINTIC COLORIMTRICDETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME GGT TYPE KINETIC FILTER 405 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 2757 SAMPLE BLANK NO RETARD 60 READING TIME 60 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT GLUCOSE CATLOG BT 10.008.00 METODOLOGY ENZIMTIC

    COLORIMTRIC DETERMINATIONS 1000

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME GLU TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT IRON CRX CATLOG BT 12.004.00 METODOLOGY COLORIMETRIC DETERMINATIONS 200

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME IRON TYPE FINAL POINT FILTER 620 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT LACTATE DEHYDROGENASE

    CATLOG BT 11.004.00 METODOLOGY KINTICO UV DETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME LDH TYPE KINETIC FILTER 340 TEMPERATURE 37 UNIT U/L REAG BLANK NO

    CALIB NAME NO

    FACTOR 8095 SAMPLE BLANK NO RETARD 60 READING TIME 60 REACTION VOL. 500 REPET. TEST 1 REF.LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT MAGNSIUM CATLOG BT 12.007.00 METODOLOGY MAGON DETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME MAG TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT PHSPHORUS UV CATLOG BT 12.006.00 METODOLOGY Molibidate UV DETERMINATIONS 100

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME PUV TYPE FINAL POINT FILTER 340 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT PROTEN TOTAL CATLOG BT 10.009.00 METODOLOGY COLORIMTRIC DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME PRO TYPE FINAL POINT FILTER 546 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT PROTEN URINRY CATLOG BT 10.016.00 METODOLOGY COLORIMTRIC DETERMINATIONS 200

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME PRO-U TYPE FINAL POINT FILTER 600 TEMPERATURE 25 UNIT g/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT TRIGLICRIDES CATLOG BT 10.010.00 METODOLOGY ENZIMTIC

    COLORIMTRIC DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME TGR TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT URA UV CATLOG BT 10.012.00 METODOLOGY KINTICO UV DETERMINATIONS 250

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME UUV TYPE TWO POINTS FILTER 340 TEMPERATURE 37 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 20 READING TIME 60 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.600 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT URIC ACID - MONOREAGENT

    CATLOG BT 10.001.00 METODOLOGY ENZYMATIC

    COLORIMETRIC DETERMINATIONS 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Reagents ready to use. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME UAC TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUCT URIC ACID - TWOREAGENT

    CATLOG BT 10.014.00 METODOLOGY ENZYMATIC

    COLORIMETRIC DETERMINATIONS 250

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Mix 4 parts of Reagent A with 1 part of Reagent B and homogenize it gently. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME UAC TYPE FINAL POINT FILTER 505 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *

  • Equipment Applications

    MICROLAB 200

    TRADEMARKS REVISION 00 - APR/2006 CUSTOMER TECHNICAL SERVICE BioTcnica Ind e Com. Ltda.

    Tel / Fax: +55 35 3214-4646 Varginha MG Brasil. Site: www.biotecnica.ind.br / e-mail: [email protected]

    PRODUTO UREA CATLOGO BT 10.013.00 METODOLOGIA ENZYMATIC

    COLORIMETRIC N DETERMINAES 500

    - The number of determinations was calculated NOT taking into account the dead volume of the equipment flask. - Its recommended to read the Instructions before using this product. Some complementary information for the realization of the test are described joint to the Automation Protocol below:

    REAGENT PREPARATION Please follow the Instructions of the kit. CALIBRATOR CONTROLS Autocal H CAT BT 13.002.00

    To evaluate the performance of the test its sugested the use of the BioTcnica Controls: - Quantinorm CAT BT 13.003.00 - Quantialt CAT BT 13.004.00

    SPECIAL OBSERVATIONS None

    # Parameter defined by the user. @ Insert the concentration of the calibrator according to the indication in the label of Autocal H - CAT BT 13.002.00

    TEST NAME URE TYPE FINAL POINT FILTER 578 TEMPERATURE 25 UNIT mg/dL REAG BLANK SI ABS - LOW 0.000 ABS - HIGH 2.300 CALIB - NAME STANDARD

    ............- NR 1

    ............- CONC @

    ............- REP 1

    ............- ABS LOW 0.000

    ............- ABS HIGH 2.300 FACTOR 1 SAMPLE BLANK No RETARD 5 READING TIME 2 REACTION VOL 500 REPLICATES 1 REF. LOW * REF. HIGH * ABS LOW 0.000 ABS HIGH 2.300 LOT * TARGET VALUE * LOW LIMIT * HIGH LIMIT *