Chema Diagnostica Via Campania 2/4 I-60030 Monsano | An Italy T +39 0731 605064 | F +39 0731 605672 www.chema.com | [email protected] Mindray BS-120 / BS-200 / BS-300 / BS-380 / BS-400 rev 432.0.2 - 2013-01-15
Transcript

Chema DiagnosticaVia Campania 2/4I-60030 Monsano | An ItalyT +39 0731 605064 | F +39 0731 605672www.chema.com | [email protected]

Mindray BS-120 / BS-200 / BS-300 / BS-380 / BS-400rev 432.0.2 - 2013-01-15

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 3 of 41

Warning!

These programs should be used only as guideline.

To check the reagent correct setting on the

instrument, use only good quality control sera

and work accordingly to with good laboratory

practice. Instrumental factor must be checked.

Chema Diagnostica will not be responsible for

instrumental bad programming.

Mindray BS-120 / BS-200 / BS-300 / BS-380 / BS-400rev 432.0.2 - 2013-01-15

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 4 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: ALBNo.: *FULL NAME: ALBUMINSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: g/dlPRECISION: 0.1PRI. WAVE: 630 605SEC. WAVE: - 700SAMPLE VOLUME (mL): 3 2R1 (mL): 400 300R2 (mL): - -REAC. TIME FROM: 0 0 0 34 44REAC. TIME TO: 17 19 25 36 36INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 0.1TO: 6.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: ALBUMIN- Code BC 0100 - 0500 - 1000 - 1500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 5 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: ALPNo.: *FULL NAME: ALP DGKCSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 510 505SAMPLE VOLUME (mL): 5 5R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 5TO: 2500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: ALKALINEPHOSPHATASEFL DGKC - Codes AL F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 6 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: ALPNo.: *FULL NAME: ALP IFCCSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 700 700SAMPLE VOLUME (mL): 5 5R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 5TO: 2500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: ALKALINEPHOSPHATASEFLIFCC - Codes AF F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 7 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: AMYNo.: *FULL NAME: AMYLASESTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 700 700SAMPLE VOLUME (mL): 6 6R1 (mL): 250 250R2 (mL): - -REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 69 75INCUBATION TIME: - - - -BLANKING TIME: - - -

LINEARITY RANGEFROM: 3TO: 2000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: AMYLASEFL - Codes AM F060 / F120 / F245 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 8 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: AMYNo.: *FULL NAME: AMYLASE IFCCSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 510 505SAMPLE VOLUME (mL): 5 5R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 12 14 15 52 64INCUBATION TIME: 16 19 25 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 5TO: 1500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: AMYLASEEPSFL - Codes EA F080 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 9 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: AMYPNo.: *FULL NAME: ISOAMYLASE IFCCSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 510 505SAMPLE VOLUME (mL): 5 5R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 12 14 15 52 64INCUBATION TIME: 16 19 25 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 3TO: 2000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: PANCREATICISOAMYLASEFL - Code PA F080 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 10 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CO2No.: *FULL NAME: BICARBONATESTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: DECREASEUNIT: mmol/lPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 300 200R2 (mL): - -REAC. TIME FROM: 0 0 0 34 44REAC. TIME TO: 17 19 25 36 36INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 1TO: 50

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: BICARBONATEFL- Codes BR F060 / F245 / F400 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 11 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: BILTNo.: *FULL NAME: BILIRUBIN TOTALSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.1PRI. WAVE: 510 505SEC. WAVE: 700 700SAMPLE VOLUME (mL): 10 10R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 17 19 25 60 79INCUBATION TIME: 16 12 15 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 0.1TO: 20.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: BILIRUBINTOTAL FL - Codes DT F125 / F500 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT A Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 12 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: BILDNo.: *FULL NAME: BILIRUBIN DIRECTSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.1PRI. WAVE: 546 546SEC. WAVE: 700 700SAMPLE VOLUME (mL): 10 10R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 17 19 25 60 79INCUBATION TIME: 16 12 15 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 0.1TO: 13.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: BILIRUBINDIRECT FL - Codes DD F125 / F500 CH Preparation: R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT A Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 13 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CANo.: *FULL NAME: CALCIUMSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.1PRI. WAVE: 630 660SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 300 200R2 (mL): - -REAC. TIME FROM: 0 0 0 34 44REAC. TIME TO: 17 19 25 36 36INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 0.1TO: 20.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: CALCIUMASX - Code CA 0100 - 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 14 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CANo.: *FULL NAME: CALCIUMSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.1PRI. WAVE: 578 570SEC. WAVE: 700 700SAMPLE VOLUME (mL): 7 5R1 (mL): 200 150R2 (mL): 200 150REAC. TIME FROM: 0 0 0 38 63REAC. TIME TO: 17 19 25 40 67INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 0.1TO: 20.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: CALCIUM - Code CA 0305 - 0505 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 15 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CHOLNo.: *FULL NAME: CHOLESTEROLSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 510 505SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 300 200R2 (mL): - -REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 20TO: 700

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: CHOLESTEROLFL - Codes CT F100 / F400 / 100F CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 16 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: HDL-CNo.: *FULL NAME: HDL-CSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 578 605SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 225 225R2 (mL): 75 75REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 17 19 25 60 79INCUBATION TIME: 16 16 19 -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 5TO: 220

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: HDLDIRECTFL - Code HD F080 / F245 / F400 CH Preparation: R1 - USE LIQUID RFADY TO USE REAGENT A R2 - USE LIQUID RFADY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 17 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: LDL-CNo.: *FULL NAME: LDL-CSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 578 605SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 225 225R2 (mL): 75 75REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 17 19 25 60 79INCUBATION TIME: 16 16 19 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 5TO: 400

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: LDLDIRECTFL - Code DL F080 CH Preparation: R1 - USE LIQUID RFADY TO USE REAGENT A R2 - USE LIQUID RFADY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 18 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CHENo.: *FULL NAME: CHOLINESTERASESTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: DECREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 700 700SAMPLE VOLUME (mL): 5 5R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 100TO: 25000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: CHOLINESTERASEFL(DGKC)- Codes CH F096 - F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 19 of 41

Application: CK-NACFL - Codes CK F060 / F120 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CKNo.: *FULL NAME: CKSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 546 546SAMPLE VOLUME (mL): 5 5R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 10 12 15 51 63REAC. TIME TO: 18 23 30 60 80INCUBATION TIME: 16 12 15 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 5TO: 2000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 20 of 41

Application: CK-MBFL - Code MB F060 / F120 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CKNo.: *FULL NAME: CKSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 546 546SAMPLE VOLUME (mL): 10 10R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 10 12 15 51 63REAC. TIME TO: 18 23 30 60 80INCUBATION TIME: 16 12 15 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 3TO: 1000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 21 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CUNo.: *FULL NAME: COPPERSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 578 570SEC. WAVE: 700 700SAMPLE VOLUME (mL): 20 20R1 (mL): 150 150R2 (mL): 150 150REAC. TIME FROM: 0 0 0 38 63REAC. TIME TO: 17 19 25 40 67INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 5TO: 500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: COPPER- Code CU 0100 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: ROOM TEMPERATURE (Warning! The reagent A could precipitate if refrigerate) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 22 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CREANo.: *FULL NAME: CREATININESTANDARD No.: 2REAC. TYPE: FIXED TIMEDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.01PRI. WAVE: 510 505SEC. WAVE: 670 700SAMPLE VOLUME (mL): 18 16R1 (mL): 180 160R2 (mL): 180 160REAC. TIME FROM: 2 2 2 38 45REAC. TIME TO: 8 8 12 48 56INCUBATION TIME: 16 5 5 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 0.1TO: 20.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: CREATININE- Code CR 0500 / 1000 / 1500 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 23 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: CREA-ENo.: *FULL NAME: CREAT ENZSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.01PRI. WAVE: 546 546SEC. WAVE: 700 700SAMPLE VOLUME (mL): 6 6R1 (mL): 180 180R2 (mL): 60 60REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 17 19 25 60 79INCUBATION TIME: 16 12 15 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 0.1TO: 20.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: CREATININEEFL- Code CE F125 / F375 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 24 of 41

Application: GAMMA-GTFL - Codes GT F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: GGTNo.: *FULL NAME: GAMMA GTSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 1PRI. WAVE: 405 412SEC. WAVE: 670 700SAMPLE VOLUME (mL): 20 20R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 3TO: 800

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 25 of 41

Application: GLUCOSEFL - Code GL F400 / 100F CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: GLUNo.: *FULL NAME: GLUCOSESTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 510 505SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 300 200R2 (mL): - -REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 10TO: 500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 26 of 41

Application: GLUCOSEUVFL - Code GL F601 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: GLUNo.: *FULL NAME: GLUCOSESTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 660 660SAMPLE VOLUME (mL): 3 3R1 (mL): 240 240R2 (mL): 60 60REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 10TO: 1000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 27 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: ASTNo.: *FULL NAME: GOT-ASTSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: DECREASEUNIT: U/LPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 405 412SAMPLE VOLUME (mL): 20 20R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 3TO: 500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: GOT/ASTFL - Codes GO F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 28 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: ALTNo.: *FULL NAME: GPT-ALTSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: DECREASEUNIT: U/LPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 405 412SAMPLE VOLUME (mL): 20 20R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 3TO: 500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: GPT/ALTFL - Codes GP F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 29 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: FENo.: *FULL NAME: IRONSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 546 570SEC. WAVE: 700 700SAMPLE VOLUME (mL): 30 30R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 17 19 25 60 79INCUBATION TIME: 16 12 15 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 10TO: 1000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: IRONFZ - Code FE F245 / F400 CH Preparation: R1 - USE LIQUID RFADY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B (preparation: read insert sheet) Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 30 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: LDHNo.: *FULL NAME: LDHSTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: DECREASEUNIT: U/LPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 670 700SAMPLE VOLUME (mL): 3 3R1 (mL): 240 240R2 (mL): 60 60REAC. TIME FROM: 3 4 5 41 50REAC. TIME TO: 13 15 20 56 69INCUBATION TIME: 16 3 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 20TO: 4000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: LDHFL - Codes LD F060 / F120 / F245 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 31 of 41

Application: LIPASEFL - Code LP F060 / LP F125 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: LIPNo.: *FULL NAME: LIPASESTANDARD No.: 2REAC. TYPE: KINETICDIRECTION: INCREASEUNIT: U/LPRECISION: 0.1PRI. WAVE: 578 570SEC. WAVE: 670 700SAMPLE VOLUME (mL): 4 4R1 (mL): 200 200R2 (mL): 50 50REAC. TIME FROM: 6 5 8 41 50REAC. TIME TO: 10 12 16 56 69INCUBATION TIME: 2 8 10 -BLANKING TIME: - - -

LINEARITY RANGEFROM: 0TO: 60

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 32 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: MGNo.: *FULL NAME: MAGNESIUMSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mEq/lPRECISION: 0.01PRI. WAVE: 510 505SEC. WAVE: 700 700SAMPLE VOLUME (mL): 4 3R1 (mL): 200 150R2 (mL): 200 150REAC. TIME FROM: 0 0 0 38 63REAC. TIME TO: 17 19 25 40 67INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 0.1TO: 8.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: MAGNESIUM- Code MG 0200 / 0500 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 33 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: MGNo.: *FULL NAME: MAGNESIUMSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mEq/lPRECISION: 0.01PRI. WAVE: 546 546SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 3R1 (mL): 300 300R2 (mL): - -REAC. TIME FROM: 0 0 0 38 63REAC. TIME TO: 17 19 25 40 67INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 0.1TO: 6.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: MAGNESIUMXL - Codes MX 0300 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE (2-8°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 34 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: PHNo.: *FULL NAME: PHOSPHORUSSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.01PRI. WAVE: 340 340SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 3R1 (mL): 300 300R2 (mL): - -REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 0.1TO: 20.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: PHOSPHORUS - Code PH 0100 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 35 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: HSPNo.: *FULL NAME: PROTEINS HSSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 578 605SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 400 300R2 (mL): - -REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 5TO: 500

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: PROTEINSHS- Code HS 0100 / 0500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 36 of 41

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: TPNo.: *FULL NAME: PROTEINS TOTALSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: g/dlPRECISION: 0.1PRI. WAVE: 546 546SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 3R1 (mL): 300 300R2 (mL): - -REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 1.0TO: 12.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Application: TOTALPROTEINS- Codes TP 0100 / 0500 / 1000 / 1500 CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: ROOM TEMPERATURE OR REFRIGERATE (2-30°C) Stability: AS INDICATED IN THE LABEL

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 37 of 41

Application: TRIGLYCERIDESFL - Code TR F100 / F400 / 100F CH Preparation: R1 - LIQUID READY TO USE SINGLE REAGENT Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: TGNo.: *FULL NAME: TRIGLYCERIDESSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 546 546SEC. WAVE: 700 700SAMPLE VOLUME (mL): 3 2R1 (mL): 300 200R2 (mL): - -REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 10TO: 1000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 38 of 41

Application: UREAUVFL - Codes AZ F080 / F245 / F400 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: UREANo.: *FULL NAME: UREASTANDARD No.: 2REAC. TYPE: FIXED TIMEDIRECTION: DECREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 340 340SEC. WAVE: 670 700SAMPLE VOLUME (mL): 4 3R1 (mL): 320 240R2 (mL): 80 60REAC. TIME FROM: 2 3 3 39 47REAC. TIME TO: 7 9 11 47 59INCUBATION TIME: - - - -BLANKING TIME: - - -

LINEARITY RANGEFROM: 1TO: 300

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 39 of 41

Application: URICACIDAOXFL - Codes AX F100 / F250 / F600 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: UANo.: *FULL NAME: URIC ACIDSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.1PRI. WAVE: 546 546SEC. WAVE: 670 700SAMPLE VOLUME (mL): 4 3R1 (mL): 320 240R2 (mL): 80 60REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 14 15 20 60 79INCUBATION TIME: - - - -BLANKING TIME: - 33-35 40-42

LINEARITY RANGEFROM: 0.1TO: 35.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 40 of 41

Application: URICACIDTFL - Code AU F250 / F402 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: UANo.: *FULL NAME: URIC ACIDSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 0.1PRI. WAVE: 546 546SEC. WAVE: 670 700SAMPLE VOLUME (mL): 4 3R1 (mL): 320 240R2 (mL): 80 60REAC. TIME FROM: -1 -1 -1 58 77REAC. TIME TO: 14 15 20 60 79INCUBATION TIME: - - - -BLANKING TIME: - 33-35 40-42

LINEARITY RANGEFROM: 0.1TO: 25.0

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def

Analyzer: Mindray BS-120 BS-200 BS-300 BS-380 BS-400

Chema Diagnostica - Guidelines for automatic analyzers - Mindray BS series

IUS.432.0.2 Directions for use Edition 2013-01-15 page 41 of 41

Application: ZINC - Code ZN 0125 CH Preparation: R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B Storage: REFRIGERATE AT 2-8°C Stability: AS INDICATED IN THE LABEL

PARAMETER BS-120 BS-200 BS-300 BS-380 BS-400TEST NAME: ZNNo.: *FULL NAME: ZINCSTANDARD No.: 2REAC. TYPE: ENDPOINTDIRECTION: INCREASEUNIT: mg/dlPRECISION: 1PRI. WAVE: 578 570SEC. WAVE: 670 700SAMPLE VOLUME (mL): 15 15R1 (mL): 240 240R2 (mL): 60 60REAC. TIME FROM: 0 0 0 58 77REAC. TIME TO: 33 38 50 60 79INCUBATION TIME: - - - -BLANKING TIME: - 8-10 10-12

LINEARITY RANGEFROM: 5TO: 1000

SUBSTRATE LIMIT: -FACTOR: -

CALIBRATIONRULE: two-point linearREPLICATES: 3

* user def


Recommended