+ All Categories
Home > Documents > Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will...

Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will...

Date post: 24-Sep-2020
Category:
Upload: others
View: 6 times
Download: 0 times
Share this document with a friend
37
Model : JY-8S Steel Bar Position Tester Operation Manual
Transcript
Page 1: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

Model : JY-8S

Steel Bar Position Tester

Operation Manual

Page 2: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

1

Content

I. Summary……………………………………………………………….2

II. Instrument configuration and maintenance……………………………5

III. Operation interface…………………………………………………...6

IV. Measurement………………………………………………………...20

V. Test Methods for Steel Bar Measurement……………………………26

VI. Software…………………………………………………………….28

Page 3: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

2

I. Summary

(I) Introduction

It is used to determine the position and distribution of steel bar, thickness of concrete

protection layer with known diameter of steel bar, and diameter of steel bar and

thickness of concrete protection layer when diameter of steel bar is unknown, for quality

inspection of new and old concrete reinforcement structures, such as bridge, tunnel,

and civil protection structure, etc. It can scan steel bar path. What’ more, it can also be

used to determine position of magnetic substance and conductive substance in the

nonmagnetic substance and non-conductive substance, such as cable, and pipes, etc,

in the wall. It is a nondestructive equipments with functions of automatic determination,

data storage, and data output.

(II) Main functions

1. To determine position and distribution of steel bar in the concrete reinforcement

structures.

2. To estimate diameter of steel bar and thickness of protection layer.

3. To determine thickness of protection layer of steel bar (known diameter).

4. To determine thickness of protection layer under a side and show grid.

5. To determine thickness of protection layer under a line and show sectional diagram

of steel bar.

6. Self-calibration of detector.

7. Store and browse test data.

8. Data output.

(III) Main characteristics

1. It can accurately locate the steel bar, and determine the diameter of steel bar and

thickness of protection layer.

2. It has functions of grid scan and sectional scan. It can accurately determine the

thickness of steel bar.

3. Let the sensor over and parallel to the steel bar, and it can determine the diameter

of steel bar and thickness of protection layer at this measurement state. It can

Page 4: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

3

make determination rapidly without changing measurement state.

4. It is precise in diameter of steel bar and thickness of protection layer determination.

5. It can show the grid and sectional diagram of steel bar intuitively.

6. The software interface is simple and it is easy to operate.

(IV) Main technical specification

1. Steel bar diameter range: Ф6mm~Ф50mm;

The common diameter of steel bar: Ф6, Ф8, Ф10, Ф12, Ф14, Ф16, Ф18, Ф20, Ф22,

Ф25, Ф28, Ф32, Ф36, Ф40, andФ50.

Thickness of protection layer nominal scale range:

Table 1 Nominal Scale Range of Thickness of Protection Layer Unit: mm

Diameter of

steel bar

First Nominal Scale Range Second Nominal Scale Range

Lower limit Upper limit Lower limit Upper limit

6 6 70 7 90

8 7 70 10 100

10 7 80 11 126

12 7 80 14 126

14 8 80 15 126

16 8 80 16 126

18 8 80 16 126

20 8 86 18 160

22 9 86 18 160

25 9 86 19 160

28 9 86 21 160

32 9 90 21 180

Page 5: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

4

2. Max. error of displayed value (Protection layer thickness)

Table 2 Max. error of displayed value (Protection layer thickness) Unit: mm

Max. errorNominal Scale Range of Protection Layer

First Nominal Scale Range Second Nominal Scale Range

±1 6~59 7~79

±2 60~69 80~119

±3 70~90 120~180

3. Diameter nominal scale range: Ф6mm~Ф32mm;

4. Max. error of displayed value of diameter of steel bar:

Table 3 Displayed Value of Diameter of Steel Bar Unit: mm

Dia 6 8 10 12 14 16 18 20 22 25 28 32

Max.error +2

0

±2 ±2 ±2 ±2 ±2 ±2 ±2 +3

-2±3 +4

-3

0

-4

5. Data storage capacity: 131000 pieces of measurement data;

6. Working environment:

(1) Ambient temperature: -10 ℃~+40 ℃;

(2) Relative humidity: <90% RH;

(3) Electromagnetic interfere: Without strong magnetic field;

7. Battery: 6 pieces of No.5 LR6 (Alkali) batteries, it can provide power continuously

for more than 35 hours;

8. Instrument weight:

(1) Main unit: 220 mm*150 mm*90mm; 772 g;

(2) Detector: 110 mm*55mm*28mm; 240 g;

(3) Small vehicle: 130 mm*95mm*68mm; 160 g;

(V) Working principle:

It is composed of main system, signal transmission system, signal collection system,

detector, and man-machine joint. The signal transmission system will generate excitation

signal to excite detector under the control of main unit. The detector will detect the steel

Page 6: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

5

bar and generate output signal, which will be changed into digital signal by signal

collection system. Then the digital signal will be processed by the main unit to determine

the position of steel bar, thickness of protection layer, and diameter of steel bar.

It has characteristics of automatic self-calibration, and self-adaptation. It has

advantage of high precision, easy operation, large storage capacity, and humanization

interface.

II. Instrument configuration and maintenance

(I) Instrument configuration

The instrument is composed of main unit, two signal cables, detector, and small

vehicle (scan dolly); See Fig.1.

Fig.1

(II) Preparation before use

Take out of the instrument from package. If you make thickness determination and

diameter detection, you can just connect the main unit with the detector. If you need make

steel bar scan, you should connect the main unit with the detector and small vehicle (scan

dolly). Then press the button to the initial interface.

(III) Battery change

If there is less power or the voltage is less than 6.5, it will show “less power” in the

interface. Please change these batteries.

Page 7: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

6

(IV) Notes

1. Please keep it away from any substance with strong magnetic field, such as

magnet, and transformer, etc.

2. If it is not used for a long time, please take out of batteries in the instrument to

avoid any damage to the circuit caused by battery leaking.

3. It is forbidden to disassemble the instrument without any permission.

III. Operation interface

Press any button in the initial interface to the function selection interface, see Fig.2. It

has items, such as thickness determination (to determine thickness of protection layer),

steel bar grid, diameter determination, data transmission, data browse, and data delete.

Press the Up or Down button to select an item, and then press the Enter button to the

corresponding interface.

Fig.2

Fig.3

(I) Thickness determination

The thickness determination interface is shown as Fig.3. First, preset project details.

The preset methods when the diameter (default value is 16 mm) and serial number (The

first number is 1) are known, are shown as following: Press the Left or Right button to

move cursor and press the Up or Down button to modify the value. Press the Enter button

Page 8: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

7

to confirm after setting. It will make self-calibration for detector automatically. The detector

should be in the air and far away from strong magnetic interfere. There will be word “Wait!”

shown on the interface. When the word “Wait” disappears, it indicates that the calibration

procedure is over. It will enter into the measurement state, see Fig.4.

Fig.4

The current signal value of detector will be shown at the right side of the signal item,

the quantity of test data of thickness measurement saved will be shown at the right side of

the storage item, and the thickness of protection layer of steel bar will be shown at the

right side of the thickness item. The length of black bar over the signal item indicates

distance between the detector and the overhead of steel bar. When the black bar

increases, it indicates the detector is more close to the overhead of steel bar. When the

black bar decreases, it indicates the detector is leaving away from the overhead of steel

bar. After the detector scan from the overhead of steel bar, the instrument will make an

alarm. The thickness value of protection layer of steel bar will be shown at the right side of

the thickness item at the same time. Press the Save button to save the test data, and then

the value at the right side of the storage item will add 1 automatically.

If you press the Back button during measurement procedure, it will return to project

information preset state. Press the Enter button to calibrate the detector. The detector

should be in the air and far away from steel bar and strong magnetic field. When the word

“Wait” disappears, the calibration procedure is over and you can continue making

measurement. Press the Convert (Change) button, it can convert between the first

nominal scale range and the second nominal scale range. Please press the Enter button

to calibrate the detector again after nominal scale rage change. Otherwise, the test data

will be incorrect.

Page 9: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

8

In order to find the overhead of steel bar, please scan the concrete reinforcement

structure roughly at first. When you hear the alarm, you can move the detector backward.

Because you may move the detector quickly at the first time, it may lose some data. Then

you should move the detector slowly when you move it backward after you hear the alarm.

When you hear the alarm the second time, the value at the right side of signal item will

change at the same time. Repeat these procedures until the value at the right side of

signal item is at the max. Then the center of the arrow (it is used to show the overhead of

steel bar) on the detector is the overhead of steel bar.

Note: When the thickness of protection layer is out of the lower limit of nominal scale

range of thickness of protection layer, the interface will show ≤XX. If the thickness of

protection layer is large than 50 mm when you make measurement using the first nominal

scale range, you can make measurement using the second nominal scale range for higher

precision.

(II) Steel bar scan

Please connect the small vehicle to the main unit and assemble the detector and

small vehicle (Insert the detector into the small vehicle) before steel bar scan. Press the

Up or Down button to select desired item in the steel bar scan interface (See Fig.5). Press

the Enter button to the corresponding interface.

Fig.5

Fig.6

Page 10: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

9

1. Grid scan

The grid scan interface is shown as Fig.6. Press the Up or Down button to select

suitable length and width. Press the Enter button to the corresponding interface.

Fig.7

The interface for steel bar grid of 0.5*0.5 m is shown as Fig.7. Please preset the

project information at first. This includes known lateral (Y) steel bar diameter and

longitudinal (X) steel bar diameter (the default value is X, Φ16mm and Y, Φ08 mm), and

project serial number (The first number is 3). Then press the Left or Right button to move

cursor, and press the Up or Down button to modify value. Press the Enter button after

setting. Then it will make calibration for the detector. The detector should be in the air and

far away from strong magnetic field. There will show word “Wait!” in the interface during

calibration. When the word disappears, it indicates the calibration procedure is over and it

is in the measurement state.

The positive direction of small vehicle is the side with a wheel and plug. The small

vehicle only can move in the positive direction during measurement. The number below

the distance item will increase during movement.

The number below the distance item indicates the horizontal distance between the

detector and the zero point. Its unit is mm. The number below the thickness item is the

thickness of protection layer of steel bar. The direction shown at the right side of direction

item is the scan direction of small vehicle, or steel bar direction got from design

information or experience. If there are not sure, please refer to Test Methods for Steel Bar

Measurement to determine the position of steel bar. When the direction is → at the right of

the direction item, you should determine the longitudinal steel bar. The measurement

point should be chosen at a cross point of steel grid, so that to avoid any affects to the

Page 11: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

10

determination of longitudinal steel bar caused by the lateral steel bar. Hold the small

vehicle and move it from left to right horizontally (The direction of longitudinal steel bar;

the speed of small vehicle should not exceed 20 mm/s). It will show a black block moving

from left to right horizontally in the interface. When there is an alarm, it indicates that there

is steel bar under the detector and the steel bar will be shown in the interface vertical to

the X-axis. The thickness of protection later will be shown below the thickness item.

Continue moving the small vehicle, and when the small vehicle moves more than 500 mm,

there will be continuous alarm. Then press the Down button to change the direction to ↓.

Then you can make determination for the lateral steel bar. As the same, you should

choose the cross point as the measurement point to avoid any affects to lateral steel bar

caused by longitudinal steel bar. Hold the small vehicle and move it from up to down

horizontally. There will be a black vehicle moving from up to down in the interface. When

there is an alarm, the steel bar will be shown in the interface parallel to the X-axis, and the

thickness of protection layer will be shown below the thickness item. Continue moving the

cursor, and when it moves more than 500 mm, there will be continuous alarm. Then press

the Save button to store test data.

Press the Convert button during determination to convert between the first nominal

scale range( ) and the second nominal scale range ( ). Press the Enter button to

calibrate the detector after nominal scale range change. Otherwise, the test data is

incorrect. When the direction is →, press the Down button, and the direction will change to

↓. Then it will make determination for the lateral steel bar. Press the Save button to store

test data (The test data can be saved only when the direction is ↓). Press the Enter button

for self-calibration of detector. Please let the detector in the air and be away from steel bar

and strong magnetic field. When the word “Wait!” disappears, the calibration procedure is

over. Then you can continue making determination.

The difference of operation methods for 1.0*1.0 m, 2.0*2.0 m and 0.5*0.5 m is that

the moving distance of detector relative to the zero point in lateral and longitudinal

direction for 1.0*1.0 m and 2.0*2.0 m is ≤1.0m and ≤2.0m relatively. Other operation

methods are the same.

Page 12: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

11

No border scan is convenient for user making scan in wide range. There is no limit on

the scan area, so it can enlarge the scan area to 10*10 m. The displayed area for no

border scan in each screen is 2.0*2.0 m. When the distance is out of this range, it will

change displayed area automatically. The difference of operation methods between no

border scan and 0.5*0.5 m is that the moving distance of detector relative to the zero point

in lateral and longitudinal direction is less than 10 m for no border scan. Other operation

methods are the same.

Note: The moving speed of small vehicle should not be more than 20 mm/s for no

border scan, or some steel bars may not be measured and there will be a mess appears in

the display. When the approximate scan range is known, it is better to select scan with

limit border. Then it can show distribution of steel bar more intuitively.

2. Section scan

Section scan is used for measuring steel bars, which are distributed abreast. The

section scan interface is shown as Fig.8. Press the Up or Down button to select the

desired length. Press the Enter button to the measurement interface of selected length.

Fig.8

Fig.9

The interface for section scan of 0.5 m is shown as Fig.9. Please preset the project

details, diameter of longitudinal steel bar (Default value is 16 mm), and serial number (The

first number is 4). Press the Left or Right button to move cursor, and press the Up or Down

Page 13: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

12

button to modify value. Press the Back button to the function selection interface. Please

press the Enter button to confirm after setting and make self-calibration of detector. The

detector should be in the air and away from strong magnetic field during calibration. There

will be word “Wait!” shown on the interface during calibration. When the word “Wait!”

disappears, it indicates the calibration is over. Then it will enter into the measurement

state.

The number below the distance item indicates the relative distance between the

detector and the zero point. Its unit is mm. The number below the thickness item is the

thickness of protection layer of steel bar. The direction shown at the right side of direction

item is the scan direction of small vehicle. Please decide steel bar direction from design

information or experience. If it is not sure, please refer to Test Methods for Steel Bar

Measurement to determine the position and direction of steel bar. Hold the small vehicle

and move it from left to right horizontally at a constant speed not more than 20 mm/s (The

detector is parallel to the steel bar). It will show a black block moving from left to right

horizontally in the interface. When there is an alarm, it indicates that there is steel bar

under the detector and the steel bar will be shown in the interface. The thickness of

protection later will be shown below the thickness item. Continue moving the small vehicle,

and when the small vehicle moves ≥ 500 mm, there will be continuous alarm. Then press

the Save button for data storage.

Press the Convert button during determination to convert between the first nominal

scale range( ) and the second nominal scale range ( ). Press the Enter button to

calibrate the detector after nominal scale range change. Otherwise, the test data is

incorrect. Press the Save button to store test data. Press the Enter button for

self-calibration of detector. Please let the detector in the air and be away from steel bar

and strong magnetic field. When the word “Wait!” disappears, the calibration procedure is

over. Then you can continue making determination.

The difference of operation methods for 1.0 m, 2.0 m and 0.5 is that the moving

distance of detector relative to the zero point in lateral direction for 1.0 m and 2.0 m is

≤1.0m and ≤2.0m relatively. Other operation methods are the same.

Page 14: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

13

No border scan is convenient for user making scan in wide range. There is no limit on

the scan length, so it can enlarge the scan length to 10 m. The displayed area for no

border scan in each screen is 2.0 m. When the distance is out of this range, it will change

displayed area automatically. The difference of operation methods between no border

scan and 0.5 m is that the moving distance of detector relative to the zero point in lateral

and longitudinal direction is less than 10 m for no border scan. Other operation methods

are the same.

Note: The moving speed of small vehicle should not be more than 20 mm/s for no

border scan, or some steel bars may not be measured and there will be a mess appears in

the display. When the approximate scan range is known, it is better to select scan with

limit border. Then it can show distribution of steel bar more intuitively.

(III) Diameter measurement

Fig.10

The diameter measurement interface is shown as Fig.10. Please preset the project

No. (The first number is 2) at first. Press the Left or Right button to move cursor, and press

the Up or Down button to modify value. Press the Enter button to confirm after setting and

make self-calibration of detector. Please let the detector in the air and keep it away from

strong magnetic field. There will be word “Wait!” appears in the interface during calibration.

When the word “Wait!” disappears, it indicates the calibration is over, see Fig.11. Then it

will enter into the measurement state.

Page 15: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

14

Fig.11

The current signal value of detector will be shown at the right side of the signal item,

the quantity of test data saved will be shown at the right side of the storage item, the

estimated diameter of steel bar is shown at the right side of the diameter item, and the

thickness of protection layer of steel bar will be shown at the right side of the thickness

item. The length of black bar over the signal item indicates distance between the detector

and the overhead of steel bar. When the black bar increases, it indicates the detector is

more close to the overhead of steel bar. When the black bar decreases, it indicates the

detector is leaving away from the overhead of steel bar.

In order to find the overhead of steel bar and accurately measure the diameter of

steel bar and thickness of protection layer, please scan the concrete reinforcement

structure roughly at first. When you hear the alarm, you can move the detector backward.

Because you may move the detector quickly at the first time, it may lose some data. Then

you should move the detector slowly when you move it backward after you hear the alarm.

When you hear the alarm the second time, the value at the right side of signal item will

change at the same time. Repeat this procedure until the value at the right side of signal

item is at the max and the black bar is longest. Then the center of the diamond (it is used

to show the overhead of steel bar) on the detector is the overhead of steel bar. Press the

Convert button and wait for a while, and then it will show the estimated diameter of steel

bar and thickness of protection layer, which will be shown at the right side of the diameter

item and thickness item in bold. Then you can press the Save button for data storage. The

value at the right side of storage will add 1 automatically. It indicates the quantity of data

Page 16: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

15

saved. You can continue making measurement now. Press the Back button to the project

details preset state.

Press the Enter button for self-calibration of probe during calibration. Please put the

probe in the air and keep it away from steel bar and strong magnetic field. When the word

“Wait!” in the interface disappears, it indicates the calibration is over. Then you can

continue making measurement.

Note: When the thickness of protection layer is less than the lower limit of the

thickness of protection layer in the table 4, the words “too thin” will be shown at the right

side of the diameter item. If the thickness of protection layer is larger than the higher limit

of the thickness of protection layer in the table4, the words “too thick” will be shown at the

right side of the diameter item. Then it can’t determine the diameter of steel bar.

Table 4 Nominal Scale Range of Diameter of Steel Bar Unit: mm

Diameter of steel bar The lower limit of thickness

of protection layer

The higher limit of thickness of

protection layer

6 7 60

8 10 60

10 11 65

12 14 65

14 15 65

16 16 65

18 16 65

20 18 65

22 18 65

25 19 65

28 21 65

32 21 65

(IV) Data transmission

The data transmission interface is shown as Fig.12. Press the Up or Down button to

select suitable data transmission mode. Press the Enter button to relative interface.

Page 17: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

16

Fig.12

1. USB transmission

The USB transmission interface is shown as Fig.13. Press the Enter button for data

transmission. Press the Back button to the function selection button without data

transmission. There will be words “Sending …” shown in the interface during transmission,

see Fig.14. There will be word “End” shown in the interface after transmission, see Fig.16.

Press any button to the function selection interface.

Fig.13

Fig.14

Fig.15

Page 18: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

17

Fig.16

2. Serial port transmission

The interface for serial port transmission is shown as Fig.15. Press the Enter button

for data transmission. Press the Back button to the function selection interface without

data transmission. There will be word “End” shown in the interface after data transmission,

see Fig.16. Press any button to the function selection interface.

(V) Data browse

The data browse interfaces are shown as Fig.17, Fig.18, Fig.19 and Fig.20. The left

side is the project No. area, and the left side is the data area of selected project. The item

with project No. in the mode of “1XXX” contains data of the thickness of protection layer

when the diameter of steel bar is known. The item with project No. in the mode of “2XXX”

contains data of the diameter of steel bar and the thickness of protection layer when the

diameter of steel bar is unknown. The item with project No. in the mode of “3XXX”

contains data of grid scan. The item with project No. in the mode of “4XXX” contains data

of section scan. Press the Up or Down button to select desired project number. The

number with arrow is the selected one, and the data shown in the data area at the right

side is the data of selected project.

Fig.17

Press the Left or Right button to browse data of selected project in different pages.

Press the Right button to browse data of current project in next page and press the Left

Page 19: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

18

button to browse data on the previous page. Press the Back button to the function

selection interface.

Press the Up or Down button to select project No. in the mode of “1XXX”, see Fig.17.

The known diameter item is the diameter you preset according to the design information.

The data below it is the data of measured thickness of protection layer.

Press the Up or Down button to select project No. in the mode of “2XXX”, see Fig.18.

The data area at the right side contains data of diameter of steel bar and thickness of

protection layer. The diameter of steel bar is corresponding to the thickness of protection

layer. The first data of a line in the thickness column is corresponding to the first data of

same line in the diameter column. The second data is the same.

Fig.18

Press the Up or Down button to select a project No. in the mode of “3XXX”, see

Fig.19.

Fig.19

The data area at the right side contains data of diameter of steel bar, thickness of

protection layer, and relative distance between the steel bar and the zero point. The No.

item contains the serial number of steel bar. The Xφ16 item indicates preset longitudinal

Page 20: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

19

diameter of steel bar for current project number. For the data below this item, the first line

is the measured thickness of protection layer of longitudinal steel bar and the second line

is the relative distance between steel bar and the zero point. The Yφ08 item indicates

preset lateral diameter of steel bar for current project number. For the data below this item,

the first line is the measured thickness of protection layer of lateral steel bar and the

second line is the relative distance between steel bar and the zero point.

Press the Up or Down button to select project No. in the mode of “4XXX”, see Fig.20.

The data area at the right side contains serial number of steel bar, thickness of protection

layer of steel bar, and relative distance between the steel bar and the zero point. The data

below the No. item is serial number of steel bar. The data below H/16 item is measured

thickness of protection layer when the diameter of steel bar (16 mm) is known. The data

below Dx item is the relative distance between steel bar and the zero point.

Fig.20

Fig.21

(VI) Data delete

The interface for data delete is shown as Fig.21. Press the Enter button to delete all

test data. It will return to the function selection interface after deletion. Press the Back

button to the function selection interface without data deletion.

Note: These data will not be recovered after deletion. Please think twice before

Page 21: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

20

deletion.

(VII) End

Press the button to turn off the instrument.

Note: In order to lower down impact to the screen, you can turn on the instrument

again at least 30 seconds after you turn off the instrument.

IV. Measurement

(I) Notes

1. When it enters into the measurement state (thickness measurement, diameter

measurement, and steel bar scan), it will automatically calibrate the detector. Then you

should let the detector in the air or keep it away from metal substances and other

magnetic substances.

2. Please make sure the measurement surface is flat so as to improve measurement

accuracy and to avoid any wrong determination.

3. Please try your best to move the detector at a constant speed during determination.

Don’t move it backward before you find a steel bar. Or you may get wrong determination

results.

4. The moving speed should not be more than 20 mm/s, or there will be greater

measurement error, even data missing.

5. When it runs continuously for a long time, in order to improve measurement

accuracy, you should let the detector in the air and keep it far away from steel bar, and

press the Enter button to reset it once every 5 minutes to eliminate measurement error (If

you have doubts on the determination results, you can make measurement again after

reset).

6. When you make determination for thickness of protection layer (thickness

measurement) with the diameter of steel bar is known, in order to ensure the accuracy of

thickness of protection layer measurement, the preset diameter should be equal to the

real diameter of steel bar. The response of detector is different for steel bar of different

diameter, so the test results may be difference when you make determination for a steel

Page 22: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

21

bar using different setting value of diameter.

7. The Convert button is used to change between the first nominal scale range and

the second nominal scale range. In order to improve the measurement accuracy, please

press the Enter button to reset the instrument once after change.

8. Please pay more attention to the moving direction of small vehicle. It is forbidden to

move it backwards before it finds the steel bar. Or this may cause wrong determination.

(II) Reset

1. Please let the detector in the air and keep it away from metal substance (It should

be at least 0.5 m distance from the metal substance).

2. Press the Enter button to reset the instrument in the interfaces of thickness

measurement, diameter measurement, and steel bar scan, so that to eliminate the effects

of environment.

3. After about 3 seconds, the word “Wait!” will disappear and the reset procedure is

over. Then you can make measurement normally. If there is strong magnetic field around,

it will take long time to reset the instrument.

Note: Don’t reset the instrument when the detector is close to the steel bar. Or the

test data will be incorrect.

(III) Single steel bar location, thickness of protection layer and diameter of steel bar

determination.

A. To determine thickness of protection layer when the diameter of steel bar is known

1. Take out of the instrument from package and connect the detector.

2. Turn on the power supply of the instrument and select function of thickness

measurement. Preset the diameter of steel bar as the known value got from design

information, and project No. of the measurement.

3. Let the detector in the air and keep it away from metals (at least 0.5 m). Press the

Enter button to reset it, so that to eliminate effects of environment.

4. Please decide the steel bar direction according to design information or experience.

If it is not sure, please make multi-point scan in two orthogonal directions to determine the

position of steel bar, see Fig.22.

Page 23: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

22

Fig.22

5. Put the detector on the surface of structure to be determined. Let the detector

parallel to the steel bar and move it at a constant speed (less than 20 mm/s) in the

direction vertical to the steel bar direction. When the detector reaches to the overhead of

steel bar to be determined, the instrument will make an alarm indicating there is steel bar

below. It can automatically show the thickness of protection layer. Press the Save button

to save test data under current project number. Move the detector in the contrary direction

slowly and observe the value at the right side of interface. When the value is the minimum

and the signal value is the maximum, the position of detector is the accurate position of

steel bar. The center of steel bar will coincide with the diamond on the detector.

6. Value judgment

When the thickness of protection layer is larger than a certain value, the signal of

detector will be very weak. Therefore, in order to lower down wrong determination, it is

better not to make judgment for the position of steel bar automatically by the instrument.

The user should decide the position of steel bar according to the change of current value.

The judgment of users is called as value judgment. Observe the value shown at the right

side of the interface. When the value changes from large to small, it indicates the detector

is close to the steel bar gradually. When the value changes from small to large, it indicates

the detector is leaving away from steel bar. Move the detector slowly in a reciprocating

way. When the value is the smallest and the signal value is the largest, the position is the

correct position of steel bar.

B. To determine the thickness of protection layer and the diameter of steel bar when

the diameter of steel bar is unknown.

It can make determination for the thickness of protection layer and the diameter of

Page 24: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

23

steel bar at the same time when the diameter of steel bar is unknown.

1. Take out of the instrument and connect the detector.

2. Turn on the power supply and select the diameter measurement item. Preset the

project No. of this determination.

3. Let the detector in the air and keep it away from metal substances (at least 0.5 m).

Press the Enter button to reset the instrument, so that to eliminate the effects of

environment.

4. Please decide the steel bar direction according to design information or experience.

If it is not sure, please make multi-point scan in two orthogonal directions to determine the

position of steel bar, see Fig.23.

Fig.23

5. Put the detector on the surface of structure to be determined. Let the detector

parallel to the steel bar and move it at a constant speed (less than 20 mm/s) in the

direction vertical to the steel bar direction. When the detector reaches to the overhead of

steel bar to be determined, the instrument will make an alarm indicating there is steel bar

below. Move the detector in the contrary direction slowly in a reciprocating way and

observe the signal value in the interface. When the signal value is the maximum, the

position of detector is the accurate position of steel bar. Keep the detector stable, and

press the Convert button. It can show the diameter of steel bar and thickness of protection

layer after a while. Their values will be shown at the right side of the diameter item and the

thickness item in bold. Press the Save button to save test data under current project

number.

(IV) Location of several steel bars and thickness of protection layer determination

In general, the steel bar distribution modes for several steel bars adopts several

Page 25: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

24

abreast distributed steel bars (main steel bars) with ties (such as beams and poles) or

web type (such as boards and walls).

A. Steel bar grid determination

1. Take out of the instrument and connect the small vehicle and the detector.

2. Turn on the power supply of the instrument and select the steel bar scan item.

Then select the grid scan item and select suitable size. Preset the diameter of steel bar as

the known value get from design details. If the diameter of steel bar is unknown, please

set it as the default value. Preset the project No. of the determination.

3. Let the probe in the air and keep it away from metal substance and strong

magnetic field (at least 0.5 m). Press the Enter button to reset the instrument to eliminate

the effects of environment.

4. To decide the direction of steel bar according to design details or experience. If it is

not sure, please refer to Test Methods for Steel Bar Measurement to determine the

direction of lateral steel bar and longitudinal steel bar of steel bar grid, as well as their

position.

5. Put the small vehicle and detector on the surface to be determined. Let the

detector parallel to the steel bar to be determined and move it at a constant speed less

than 20 mm/s. Please move the small vehicle as the direction → shown at the right side of

the direction item. Please determine the longitudinal steel bar at first. The measurement

point should be at the cross point of steel bars to avoid any affects to the longitudinal steel

bar caused by lateral steel bar. Hold the small vehicle and move it from left to right (The

speed should not be more than 20 mm/s). There will be a black block moving from left to

right in the interface. When you hear an alarm, it indicates the detector finds the steel bar

and the measured steel bar will be shown on the interface as a line vertical to X axis, and

the thickness of protection layer of steel bar will be shown below the thickness item.

Continue moving the small vehicle to right. When the moving distance is ≥500/1000/2000,

there will be continuous alarm appears. Then you can press the Down button to change

the direction to ↓. It can make determination for the lateral steel bar now. You should

select the cross point of steel bars as the measurement point to avoid any affects to lateral

steel bar caused by longitudinal steel bar. Hold the small vehicle and move it from up to

Page 26: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

25

down. There will be a small black block moving from up to down in the interface. When

you hear an alarm, the measured steel bar will be shown in the interface in the mode of a

line parallel to the X axis and the thickness of protection layer of steel bar is shown below

the thickness item in the interface. Continue moving the small vehicle, when the moving

distance is ≥ 500/1000/2000, there will be continuous alarm appears. Then you can press

the save button to save test data under current project number. The project No. will add 1

and you can continue making determination using the new project number. Press the

Back button to the scan mode selection interface.

B. Section steel bar measurement

1. Take out of the instrument and connect the small vehicle and detector.

2. Turn on the power supply of the instrument and select the steel bar scan item.

Select the section scan item and select suitable size next. Preset the diameter of steel bar

as the known diameter of steel bar got from design details. If it is unknown, you can preset

it as default value. Then preset the project No. of the determination.

3. Let the detector in the air and keep it far away from metal substance and strong

magnetic field (At least 0.5 m). Then press the Enter button to reset the instrument to

eliminate effects of environment.

4. To decide the direction of steel bar according to the design details or experience. If

it is not sure, you should determine the direction and position of steel bar using the

instrument at first.

5. Put the small vehicle and detector on the surface to be determined. Let the

detector parallel to the steel bar to be determined and move it at a constant speed less

than 20 mm/s. Please move the small vehicle as the direction shown at the right side of

the direction item, so that to determine the steel bar vertical to the moving direction of

small vehicle. There will be a black block moving from left to right in the interface. When

you hear an alarm, the measured steel bar will be shown on the interface in the mode of

small black block and the thickness of protection layer of steel bar will be shown below the

thickness item. Continue moving the small vehicle to right. When the moving distance is

≥500/1000/2000, there will be continuous alarm appears. Then you can press the save

button to save test data under current project number. Press the Back button to the scan

Page 27: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

26

mode selection interface.

V. Test Methods for Steel Bar Measurement

In general, the distribution of steel bar in the concrete reinforcement structure adopts

several abreast distributed steel bars with ties (such as beam, and pole) or web type

distribution (such as board, and wall). But the position of steel bar in the concrete is not

sure. Therefore, in order to improve measurement efficiency and measurement accuracy,

we should make determination as per some rules. Now we have established suitable

methods for steel bar measurement from large quantity of experiences.

(I) Get design information

Get the design details of concrete reinforcement structure to be determined and

decide the approximate position, direction, and diameter of steel bar in the structure.

Preset the diameter of steel bar as the design value. If you could not get these details, you

can preset the diameter of steel bar as the default value. Then you can measure the

diameter of steel bar and thickness of protection layer using grid scan, section scan, and

diameter measurement.

(II) To decide the measurement area

Please select an area as the measurement area on the concrete reinforcement

structure to be determined. Please select a smoother surface to improve measurement

accuracy.

(III) To determine the position of main steel bar (The steel bar at the upper layer)

Select an initial measurement point and make scan along the vertical direction of

main steel bar (for beam, and pole, etc) or vertical direction of steel bar at the upper layer

(for board, and wall, etc) to determine the position of main steel bar or steel bar at the

upper layer. Then move horizontally for certain distance and make other scan, see Fig.24.

Connect measured points during two scan procedures using lines. Notes: If the scan line

is just over the ties of main steel bar or the steel bar in the lower layer, see Fig.25, it may

not find the steel bar or get a wrong position of steel bar. Then the position of steel bar

when you repeat the scan procedure may have great deviation. You should move the scan

line for certain distance equal to twice of diameter of steel bar and make scan again.

Page 28: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

27

Fig.24

Fig.25

Fig.26

(IV) To determine the position of ties (or lower layer bar)

Make scan in the middle of two measured steel bars along the vertical direction of ties

(or lower layer bar) to determine the position of ties (or lower layer bar). Then make scan

in the middle of another two measured steel bars and connect measurement points during

two scan procedures using lines, see Fig.26.

V. To determine the thickness of protection layer and diameter of steel bar

The methods to determine thickness of protection layer when the diameter of steel

bar is known: Select the thickness measurement item. Preset the project No. and

diameter of steel bar. Make scan in the middle of two ties (or lower layer bar) along the

vertical direction of main steel bar (upper layer bar) to determine the thickness of

protection layer of main steel bar (upper layer bar). Make scan in the middle of two main

Page 29: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

28

steel bars (upper layer bars) along the vertical direction of ties (lower layer bar) to

determine the thickness of protection layer of ties (lower layer bar). Please note to preset

the state of steel bar grid.

The methods to determine thickness of protection layer and diameter of steel bar:

Select diameter measurement item. Preset the project No. and make scan in the middle of

two ties (lower layer bar) along the vertical direction of main steel bar (upper layer bar) to

determine the position of main steel bar (upper layer bar). Then put the detector parallel to

overhead of steel bar to be determined and make determination for the diameter of steel

bar and thickness of protection layer. Make scan in the middle of two main steel bars

(upper layer bar) along the vertical direction of ties (lower layer bar) to determine the

position of ties (lower layer bar). Then let the detector parallel to overhead of steel bar to

be determined and preset state of steel bar grid to make determination for the diameter of

steel bar and thickness of protection layer.

VI. Software

(I) The software is used for further process of test data. It can run on the computer

with operation system of Windows95/98/Me/2000/NT/XP.

(II) Installation

The installation methods of this software are the same as that of normal Windows

software.

1. Open the software CD. Double click the Setup icon in the software folder. The

computer will copy file at first and start installation after copy. Click Next to continue

installation and click Cancel to quit from installation interface. Click Continue to the

installation interface and click exit to exit from software installation.

2. Click the View to change installation path in the installation interface. Click Next to

start installation. Click Previous to previous interface. Click Cancel to quit from installation.

3. It will need to install USB driver during installation procedure. Click INSTALL to

install USB driver. Then use USB cable to connect the instrument with the computer. Click

Enter to end USD driver installation.

4. Click Finish to complete installation.

Page 30: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

29

(III) Software structure and functions

1. Software interface

The software interface mainly includes the following parts: Menu column, Tool column,

Control panel, and Window area.

Menu column is composed of menu items. Click each item and there will be a

dropdown menu. The submenu items of four menu items contain all functions of the

software. When a menu item is in gray color, it indicates the item could not be used at

current state.

Fig.27

Tool column is composed of buttons. Each button is corresponding to a normal

function. Though you can realize these function using menu items, it is convenient to use

buttons in the tool column. Let the mouse stay at a button for a while, it will show the

function of button automatically. When the button is in gray color, it indicates the button

could not be used at current state. The tool column can realize functions of Open, Save,

Print, Preview, Save as bmp file, Save as txt file, create Word report, create Excel report,

data transmission, show or hide control panel, insert data before current line, insert data

after current line, delete data, and about, etc. Its functions are the same at that of menu

items.

Control panel: The user can complete browse operation for test data.

Window area: It is used to show the data and diagram relative to current project No.

you selected in the project No. list.

(1) Open: Open the file you need to process or browse.

It is a dialog box of standard Windows type. Select the file you would like to open from

the Range item, and file type from the file type box. Input the file name from file name box

or select a file from the file list box. Then press the Open button to open the file.

(2) Save: It will save test data to a file.

Select a folder from Save as box and input the file name in the file name box. Then

press the Save button to save it.

Page 31: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

30

(3) Save as TXT file: Save test data to a txt file.

(4) Save as BMP file: Save section diagram of steel bar or grid diagram of steel bar to

a BMP file.

Fig.28

(5) Create Word report: It will run Microsoft Word automatically (you computer should

have Microsoft Word software, or it could not continue) and create report in the Word

format. It will ask you to input relative parameters for the report, see Fig.28. It contains

memory function for these parameters. It will use relative parameters you preset at last

time.

(6) Create Excel report: It will run Microsoft Excel automatically (You computer should

have Microsoft Excel software, or it could not continue) and create Report in the Excel

format.

(7) Save as: Save test data in other path and file name.

(8) Print: To print our content in current file.

(9) Preview: To show printing effects before printing.

(10) Print set: To preset parameters for printing, see Fig.29. It contains memory

function for these parameters. It will use relative parameters you preset at last time.

Page 32: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

31

Fig.29

Fig.30

(11) Exit: Exit from the software.

2. Data transmission

It is used to send test data from the instrument to the computer for further process

(See Fig.30).

(1) File path: To preset file path to save data got from the instrument to a computer.

(2) Port type: To select communication medium (USB port or serial port);

(3) Equipment No./Port No.: For transmission from USB port, it is used to select the

instrument for transmission (One computer can connect to several pieces of instruments).

The default value is Device1. For serial port transmission, the default value is COM1.

(4) State: It is used for serial port transmission. It can show the communication state

Page 33: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

32

of serial port.

(5) Send: It is different in operation for serial port transmission and USB port

transmission.

For serial port transmission, press the button after you connect the computer with the

instrument using serial port cable and preset relative parameters. The computer is in the

waiting state. Then you should use the transmission function of the instrument for data

transmission.

For USB port transmission, press the button after you connect the computer with the

instrument using serial port cable and preset relative parameters, and then it can make

data transmission automatically.

Fig.31

3. View Menu

It can show or hide tool column, state column, and control panel (See Fig.31). If there

is before a item, the item will be shown. Or it will be close.

(1) Control panel (Ctrl Panel): see Fig.32. The user can select suitable measurement

type from optional items. There are three measurement types, signal point measurement,

section measurement, and grid measurement.

Page 34: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

33

Fig.32

a. Parameter area

When you select a project No. from project No. list, its parameters will be shown in

the parameter area. If there is before a project No. in the list, it indicates printing,

creating report, and creating report form, etc will involve this project.

Project name: to show the name of current project.

Project No.: to show serial number of current project.

Design thickness: read design thickness;

Page 35: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

34

Upper limit of deviation: The upper limit of thickness deviation;

Lower limit of deviation: The lower limit of thickness deviation; please note the input

value should be negative.

Known diameter: To show the known diameter of steel bar when you make thickness

measurement.

Data (group) quantity: To show quantity of test data or data group.

Max. thickness: The max. value of thickness measurement data of current project.

Min. thickness: The min. value of thickness measurement data of current project.

When you double click current project No. in the project No. list, it will show dialog

box as Fig.33. You can modify the project name, design thickness, and upper limit of

deviation and lower limit of deviation.

Fig.33

b. Data area

Data display: The data area is used to show test data of the thickness of protection

layer for current project in the project list and judge whether they can meet the

requirements.

Thickness of protection layer judgment: When you input the design thickness of

protection layer and deviation tolerance, click the measurement area, and then the

software will judge whether it can meet your requirements. There will be “√” before data,

which can meet the requirements, and “Χ” before data, which can’t meet the

requirements.

(2) Section measurement

Page 36: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

35

When you select the section measurement item, it will appear an interface shown as

Fig.34. It will show test data of current project in the data area of control panel, and show

position diagram of steel bar got from section measurement in the window at the right side.

The Y coordinates will show the thickness of protection layer, and the X coordinates will

show the position of steel bar. A black point indicates a steel bar.

Fig.34

The methods for project name modification are the same as the methods for single

point measurement. You can also add or delete test data of steel bar. The methods are

described as following:

a. Add test data of steel bar. Select a line of test data. Press the button on the

tool column to insert a line of test data before the selected line, or press the button on

the tool column to insert a line of test data below selected line. Then modify data in the

line added to add test data of a steel bar.

b. Modify test data of steel bar. You can also make modification for measured test

data directly.

c. Delete test data of steel bar. Select test data of a steel bar in the data area. Press

the button on the tool column to delete test data and the steel bar shown on the

Page 37: Model: JY-8S SteelBarPositionTester OperationManual · 5 bar and generate output signal, which will be changed into digital signal by signal collectionsystem.Thenthedigitalsignalwillbeprocessedbythemainunittodetermine

36

diagram.

(3) Grid measurement

When you select the grid measurement item, it will appear a interface shown as

Fig.35. It will show test data of current project in the data area of control panel, and show

the position diagram got from grid measurement in the window at the right side. The Y

coordinates indicates position of steel bar got from vertical measurement and the X

coordinates indicates position of steel bar got from horizontal measurement. The black

lines indicate steel bars.

Fig.35


Recommended