Modification to Log-Ampposition measurement in TT2-TT10Modification to Log-Ampposition measurement in TT2-TT10
BDI day 2003BDI day 2003T Bogey AB/BDI-EMT Bogey AB/BDI-EM
BDI day 2003 2
OutlineOutline
TTPOS principle Recall of problems Solutions implemented in 2003 Things to be done for 2004 Conclusions
BDI day 2003 3
PrinciplePrinciple
PS
SPS
TT2
TT10
BPCL100618
BPCL100509
FT16UDC151
FT16UDC211
FT16UDC251
FT16UDC311
FT16UDC346
FT16UDC368
BPCL100407
BPCL100208
BPCL101302
BPCL101502
BPCL101902
BPCL102073
BPCL102302
BPCL102402
BPCL102573
BPCL102739
BPCL102911
BPCL102963
BMU02T
BMU10T
Trigger
Trigger
BDI day 2003 4
PrinciplePrinciple
Pickup
BP FilterLogAmp
BP FilterLogAmp
BP FilterLogAmp
BP FilterLogAmp
Up-Down
Up
Down
Left
Right
Left-Right
100 / 600 m
Op Amp Integrator ADC
Op Amp Integrator ADC
VMECrate
Tunnel Auxiliary Building
Vertical
Horizontal
Auto Trigger
BDI day 2003 5
Recall of problemsRecall of problems
Auto trigger BMU10T● Position● Lead ions (stripped)
Radiation damage (102963) Position signal with no beam Loading settings when SSC changes Remote diagnostics
BDI day 2003 6
Auto trigger BMU10Twith lead ions (stripped)Auto trigger BMU10Twith lead ions (stripped)
2002: Auto trigger works with Pb53+
(unstripped) but does not work with Pb82+ (stripped)
2003: A new electronics for the auto trigger
module solved the problem.
BDI day 2003 7
Autotrigger position for BMU10TAutotrigger position for BMU10T
1013
1015
1019
1020
1023
1024
1025
1027
1029
1030
BMU10T
2003 Problem
2004
Solution
BDI day 2003 8
Radiations damages (102963)Radiations damages (102963)
The monitor is very close to the dump
During 2002 run ~ 23 E+3 gray on the electronics module
We moved the electronics away from the BPM
No more radiation damage during 2003 (but still a problem on the vertical plane)
BDI day 2003 9
Position signal with no beamPosition signal with no beam
Without beam Log-Amp still gives an output signal
No information of intensity
Autotrigger module gives a hardware interrupt to validate the data
A software algorithm check the data
BDI day 2003 10
Loading settingsLoading settings
When SSC changes MCR have to call experts to load settings
Stephen Jackson has implemented an option on the PCR Operational Software - BI applications web page to help operation
For a new SSC specialist still needed
BDI day 2003 11
BDI day 2003 12
BDI day 2003 13
Remote timing diagnosticRemote timing diagnostic
Already exists but requires expert
Solution has to be implemented for operators
BDI day 2003 14
ConclusionsConclusions
TT2/10 experience helps us for TT40/41 CNGS electronics
Each Log-Amp will have its own autotrigger
Position and Intensity available Possibility to use the same electronics
in TT2/10 upgrade