+ All Categories
Home > Documents > Module I-6 Temperature Instruments

Module I-6 Temperature Instruments

Date post: 03-Apr-2018
Category:
Upload: nabil160874
View: 219 times
Download: 0 times
Share this document with a friend

of 48

Transcript
  • 7/29/2019 Module I-6 Temperature Instruments

    1/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 1

    Practical Demonstration of

    TEMPERATURE INSTRUMENTS

    MODULE NO. : I-6

    MODULE SUBJ.: Temperature Instruments

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    2/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 2

    Tasks:

    I-6.1 Perform calibration on bimetallic dial type temperature gauge.

    I-6.2 Perform calibration on filled system temperature indicator.

    I-6.3 Perform service and calibration of a temperature switch..

    I-6.4 Check the accuracy of:

    a) Thermocouple output, and b) RTD output.

    I-6.5 Perform servicing and calibration of a pneumatic temp. transmitter.

    I-6.6 Perform servicing and calibration of an electronic temperature

    transmitter with RTD input.

    I-6.7 What are self regulating temperature valves? Function & application.

    I-6.8 Service and calibrate a pneumatic temperature controller.

    I-6.9 Perform calibration of an electronic transmitter using thermocouple

    input.

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.1

    Perform calibration on bimetallic dial type temperature

    gauge.

    Reference: None

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    3/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 3

    Materials: 1. Cleaning Rags, and2. Solvent.

    Equipment & Tools: 1. Tool Box, and2. Standard temperature bath with reference

    thermometer..

    Conditions: To be performed in instrument workshop..

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to select proper tools to perform this task, Describe different types of temperature gauges, Understand the concept of bimetallic temperature elements (flat and helical), Understand the purpose of using thermowells for temperature elements, Be able to perform calibration adjustments of dial temperature gauges, Describe an understanding to his trainer, and Write observations and procedures in his workbook.

    TASK No.: I-6.1 Continue

    Details:

    Bimetallic Thermometers

    A bimetallic thermometer is a device, which senses temperature with a helical or

    spiral bimetallic strip and displays the temperature on a dial for visual observation.

    The bimetallic temperature element expand when the temperature increases and

    contract when the temperature decreases. The increase length per unit length per

    degree of temperature increase is called the coefficient of thermal expansion for

    that material.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    4/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 4

    A bimetallic element can be formed in spiral or helix to increase the amount of

    motion available for a given temperature change.

    The spiral form of bimetallic element is convenient for housing in a circular flat

    case and is typically used is dial thermometers that measure ambient temperature.

    The helical form is well suited for housing in a narrow tube (stem) for insertion

    into a fluid directly or housing within a thermowell with a small bore.

    TASK No.: I-6.1Continue

    Figure I-6.1A, Typical Bimetallic

    Dial Thermometers Types

    Figure I-6.1B,Exploded andSectional View ofa Bimetallic DialThermometer

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    5/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 5

    Figure I-6.1A shows a typical bimetallic dial thermometer, one end of the helical

    element is welded to the bottom of the stem and the other end is welded to theshaft. The pointer is attached to the end of the shaft. The sensing portion of the

    thermometer is that portion of the stem that contains the element.

    The standard calibration accuracy for bimetallic thermometers is +/- 1% of the

    full-scale reading. Some thermometers allow single point calibration, which is

    accomplished by rotating the scale behind the pointer, either by rotating the case

    or by adjusting a gear from a screwdriver slot protruding through the case. Others

    allow no adjustment.

    TASK No.: I-6.1 Continue

    Figure i-6.1B illustrates exploded and sectional views of bimetallic dial

    thermometer, where the service is limited to replacement of the dial window glass

    if is broken, adjust or replacement of the pointer. An external adjustment screw is

    usually provided so that the thermometer can be calibrated at a single point, but

    there is usually no adjustment for span.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    6/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 6

    Bimetallic thermometers are available in convenient range increments for

    measurement between50 to 500 degree C.

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.2

    Perform calibration on filled system temperatureindicator

    Reference: None

    Materials: 3. Cleaning Rags, and4. Solvent.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    7/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 7

    Equipment & Tools: 3. Tool Box, and4. Standard temperature bath with reference

    thermometer..

    Conditions: To be performed in instrument workshop..

    Requirements By Trainee:

    To study the task and familiarise himself,

    Be able to select proper tools to perform this task, Understand the concept of filled system temperature elements, Understand the purpose of using thermowells for temperature elements, Be able to perform calibration adjustments of filled system temperature gauges, Describe an understanding to his trainer, and Write observations and procedures in his workbook.

    TASK No.: I-6.2 Continue

    Details:

    Thermometers

    Filled system thermometers have a bulb filled with an expanding substance,

    (usually an inert gas) and a dial, which is controlled by a bourbon tube. The bulb

    is connected by a capillary tube, which can be up to about 50 feet (15 meters)

    long, to the dial mechanism. Their accuracy is about the same as a bimetallic

    thermometer and they are much more expensive. Therefore, filled systemthermometers are not usually unless remote installation of the gauge is desired.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    8/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 8

    TASK No.: I-6.2 Continue

    Figure I-6.2A, Filled System Devices, Types of Compensating

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    9/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 9

    The indicating temperature of filled system thermometer are subject to significant

    error from elevation changes in the capillary and are relatively difficult to

    compensate and The reading may change somewhat with changes in ambient

    temperature. The errors thus induced can be minimised by a compensating

    mechanism, which senses the ambient temperature and automatically adjusts the

    temperature reading. Figure I-6.2A illustrates types of compensation of gas field

    system.

    Figure I-6.2B, FilledSystem Thermometers

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    10/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 10

    TASK No.: I-6.2 Continue

    As shown in figure I-6.2C an industrial duratemp thermometer components and

    replacement parts.

    Calibration:

    To calibrate filled system thermometer; arrange calibration set-up using reference

    accurate thermometer with heating bath to compare between them. Consult your

    trainer for details.

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.3

    Figure I-6.2C, Filled System Thermometer Components

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    11/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 11

    Perform service and calibration of a temperature

    switch

    Reference: None

    Materials: 1. Cleaning rags.

    Equipment & Tools: 1. Tool Box, and2. Digital Multimeter / Digital Voltmeter (DVM).

    Conditions: To be performed at workshop.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe the operation principle of temperature switches, Be able to identify the main parts of a mechanical temperature switch, Understand deadband error of a mechanical temperature switch, Understand the advantages of a solid-state temperature switch, Perform periodic and corrective maintenance of temperature switches, Draw / Sketch calibration set-up of the solid-state temperature switch, Discuss an understanding with his trainer, and Write observation and procedures in his workbook.

    TASK No.: I-6.3Continue

    Details:

    Electric Temperature Switches

    An electric temperature switch is a device, which causes a contact to open or close

    with a change in temperature. Most switches can be used as either high

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    12/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 12

    temperature or low temperature sensors, depending on how they are calibrated and

    electrically connected.

    Most mechanically operated temperature switches use a vapour-filled system or a

    liquid-filled system to operate pressure switch. Gas-filled systems generally do not

    develop enough power for switch use. Figure I-6.3A shows vapour filled systemmechanically operated temperature switch.

    Filled system switches are available for both local and remote mounting. The local

    mounting type has the bulb rigidly attached to the switch mechanism and housing.

    The assembly has a threaded connection so that it can be screwed into and be

    supported by a thermowell. The remote mounting type has the bulb connection to

    the switch mechanism by a capillary tube from 6 feet (2 meters) to 25 feet (8

    meters) or more. The switch cannot be separated from the bulb in the field.

    One disadvantage of mechanically actuated temperature switches is significantdeadband. When temperature switch trips on raising or falling temperature, it does

    not reset at exactly the same temperature as it tripped. The difference is called the

    deadband or reset. Sometimes the required trip point is very near the normal

    operating temperature. In these cases, a solid-state temperature switch should be

    used.

    Solid-state temperature switches use a resistance temperature detector (RTD), a

    thermocouple, or a thermistor to detect the temperature and contain the required

    amplifiers and other electronic circuitry to activate a relay or solid-state output

    device at the set temperature. Figure I-6.3B shows a solid-state switch for controlroom mounting.

    The output from the temperature switch is usually either a microswitch or a relay.

    The contact configuration is most often either single-pole, double threw (SPDT) or

    double-pole, double-threw (DPDT).

    TASK No.: I-6.3 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    13/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 13

    TASK No.: I-6.3 Continue

    Figure I-6.3A, Mechanically Operated Temperature Switch

    Figure I-6.3B,Typical Solid State

    Temperature Switch

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    14/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 14

    A double-pole type switch or snap-disk mechanism is needed if two circuits, such

    as a shutdown and alarm that annunciates that shutdown are to be actuated by the

    same sensor. Multiple temperature set adjustments are if one temperature switch

    needs to be actuated at more than one set-point. Switch housing must be suitable

    for hazardous environment. Figure I-6.3C shows temperature switches with

    various types of housing.

    TASK No.: I-6.3 Continue

    Figure I-6.3C, Temperature Switch Housing for Various Installations

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    15/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 15

    Pneumatic Temperature Switches

    A pneumatic temperature switch is a device, which senses temperature and

    actuates a valve to supply or vent a gas or air signal. Pneumatic switches are not as

    abundant as electric switches, but they are still available from several sources.

    Some manufacturers will substitute an air relay for the microswitch in a

    mechanically actuated temperature switch.

    The pneumatic temperature switches, which are essentially converted electrical

    switches use the same principles of operation as do their electrical counterparts.

    The ones designed for the pneumatic system are often called temperature valves.

    They are manufactured with both two-way valves and three-way valves. The two-

    way type, are designed so that a vent port opens when the temperature exceeds the

    set point, the three-way valve type either connects the receiver device to the

    pressure source or vents it, depending on the temperature. These units use the

    bimetallic principle. A typical temperature valve is shown in figure I-6.3D.

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.4

    Check the accuracy of:

    a) Thermocouple output, and b) RTD output

    Figure I-6.3D, Typical Temperature Valve

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    16/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 16

    Reference: None

    Materials: Thermocouple element and RTD element

    Equipment & Tools: 1. Tool Box,2. Bath heater with reference temperature

    thermometer, and

    3. Digital Multimeter.

    Conditions: To be performed at workshop.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to select proper tools to perform this task, Understand the concept of thermocouple and RTD as temperature sensors, Be able to recognise using of the reference temperature tables for

    thermocouples and RTDs.

    Demonstrate the accuracy procedure of thermocouple and RTD, Discuss an understanding with his trainer, and Write observation and procedures in his workbook.

    TASK No.: I-6.4 Continue

    Details:

    Thermocouples

    A thermocouple is a junction of dissimilar metals used to measure temperature.

    When two different metals come in contact with each other, thermal energy is

    converted into electrical energy. Any two metals can be used and the amount of

    electrical energy created is a direct function of the absolute temperature except in

    circumstances. Also the amount of energy converted depends on the metals

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    17/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 17

    selected. Certain combinations of metals have been identified which create enough

    energy in a sufficiently linear manner so that they can be used to measure

    temperature with a high degree of accuracy.

    Thermocouple wires of the selected metals, are joined together to make electrical

    contact. They can be twisted, welded, soldered or even wrapped under the samescrew. The electrical limitations are that the junction, including any third metal,

    must be at the temperature to be measured, the wires must be insulated from each

    other from the junction to be receiver, and if the junction is grounded, there must

    be no other ground. The only physical limitation is that the wires must be able to

    stand the environment to which they are subjected.

    Connection of thermocouples does present some difficulty because when the

    selected metals are connected to any other metal, such as copper wire, another

    thermocouple is created and the temperature of this connection will affect the

    measurement as much as the temperature of the primary junction.

    TASK No.: I-6.4 Continue

    Figure I-6.4A shows a thermocouple constructed from copper (Cu) wire and a

    copper-nickel alloy wire named constantan (c) connected to a voltmeter made of

    copper. The constant an wire must be connected an wire must be connected to

    copper somewhere in addition to the thermocouple to complete the circuit, but this

    will form another thermocouple. This connection is made so that the temperature

    can be held constant at a known temperature and is called the cold junction or

    reference junction (J2). A temperature that is easy to create and duplicate is that of

    a bath formed by pure water and the ice that water, 32F (0C) by holding the

    reference junction at the ice point, the temperature of the primary junction (J1) can

    be found by measuring the voltage it creates in reference to the voltage created by

    the reference junction.

    Figure I-6.4A,

    Copper-Constantanthermocouplecircuit with anexternal reference

    junction

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    18/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 18

    Tables of the voltage created at temperature versus the ice point are published by

    the Untied National Bureau of Standards (NBS) and are used worldwide. By

    measuring the thermocouple voltage, the temperature can be found from the table.

    If the reference junction is located where the temperature is known or can be

    measured accurately, then the junction voltage for this temperature can be addedto the measured voltage to find the temperature of the primary junction. All of the

    connections and the measurement are made to a thermally conductive, but

    electrical insulating material known as the isothermal block. This block is usually

    in the instrument case, but in large installations is sometimes done elsewhere by

    minimising the thermocouple wire. If the temperature is computed circuit it is

    known as software compensation. If an electronic circuit is used to correct the

    reading, it is known as hardware compensation or an electronic ice point.

    The thermocouple is connected to the isothermal block by wire made from thesame metals as the thermocouple, called thermocouple extension wire. A

    thermocouple extension wire is usually a shielded, twisted pair with the shield

    grounded at the instrument to minimise interference pickup. Terminal strips

    constructed of thermocouple material are available and should be used if

    intermediate connections are required. Only a few mili-volts are produced by a

    thermocouple, so careful attention to proper wiring and shielding is essential to

    good measurement.

    TASK No.: I-6.4 Continue

    Figure I-6.4B,HardwarecompensatedThermocouple

    Assembly.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    19/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 19

    Some thermocouple assemblies are manufactured so that the thermocouple makes

    electrical contact with the sheath (called ground junction), and some are

    manufactured where the thermocouple is electrically insulated from the sheath

    (called ungrounded junction). A third option, is where the thermocouple extends

    slightly beyond the sheath (called exposed junction). Exposed junction offer the

    fastest response, but are not used in oil and gas processing because they are

    subject to physical damage.

    TASK No.: I-6.4 Continue

    They would need to be installed without a thermowell to take advantage of this

    faster response. Grounded junction offer faster response than ungrounded

    junctions because the contact area which provides the electrical connection also

    provides good thermal conduction. Also, grounding at thermocouple provides the

    most nearly symmetrical circuit, which reduced interference picked up by the

    wires to a minimum. Grounded thermocouples should be selected unless other

    components of the circuit require that, the ground be at some other point, or the

    process fluid and piping are not at ground potential.grounding any measurement

    loop at more than one point will usually cause measurement errors because ofpotential difference in the grounding system.

    These errors are more acute with low voltage signals such as generated by

    thermocouples. These statements do not preclude grounding the extension wire

    shield at the receiver, which is recommended.

    The most common and least expensive thermocouple is iron versus constant an

    (ISA type J). Type can be used for measurement from -320F (-195C) to1400F

    Figure I-6.4C,Three Basic Typesof ThermocoupleAssembly

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    20/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 20

    (760C), but is normally limited to 32 to 1000F (0-500C). Type J is usually

    furnished when no specific type is specified.

    Chromal versus alumal thermocouples (ISA type K) offer better corrosion

    resistance. Type K can be used for-310F (-190C) to 2500F (1370C) but

    usually limited to 32 to 2000F (0 to 1000C) Type K does not produce as muchout put as type

    Copper versus constantan thermocouples (ISA type T), are usually used when

    temperatures below zero are to be measured. While the usable range for type T,

    310 to 750 F (- 190 to 400 C), is the same for the lower limit and less for the

    higher limit than for types J and K, the recommended range is -290 to 700F (-180

    to 370C). The materials used in type T behave more predictably at low

    temperatures than those used for types J and K.

    Chromel versus constant an thermocouples (ISA type E) provide the largest

    voltage change per temperature change for standard thermocouples. An output of

    40 millivolts at 1000F can be compared to 30 mv for type J and 22 mv for type

    K. type E can be used for 320 to 1830F (-195 to 1000() and is recommended

    for32 to 1600F (0 to 870F).

    TASK No.: I-6.4 Continue

    Some sources extend this range downward to -100F (-73 C), but type T is

    generally considered a better choice for below freezing temperature. Type E has

    more tendency to change characteristics with time than types J, K and T.

    These four types of thermocouples comprise the base metal thermocouples.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    21/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 21

    Figure I-6.4D, Output versus temperature curves for the four types ofbase metal thermocouples. (Types J, K, T and E)

    Other thermocouple types, called the noble metal types are available for

    measurements where the base metal types are not suitable. They are made from

    expensive metals such as platinum, rhodium, iridium and tungsten thus are moreexpensive. Also, they do not provide as much output as the base metal types.

    These noble metal thermocouples are used in laboratories, for molten metals and

    other applications, but are rarely used in production facilities.

    TASK No.: I-6.4 Continue

    Figure I-6.4E,TypicalInstallation ofThermocouple orRTD Element

    into Thermowell

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    22/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 22

    Thermocouple Accuracy Check:

    Figure I-6.4F shows thermocouple accuracy check set-up. Every six month

    thermocouple sensor must be checked against the standard conversion table.

    Thermocouple has no repair or maintenance procedure.

    TASK No.: I-6.4 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    23/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 23

    Check procedure

    1. Unscrew the thermocouple head cover and disconnect the electrical wires totransmitter.

    2. Using V adjustable wrench to remove the thermocouple element from thethennowell.

    3. Immerse the element in a regulated thermal bath with a reference temp.indicator.

    4. Using accurate digital multivolt meter to record measured mv values across theleads wires of the thermocouple at different temp. values (5 points at least).

    5. Compare the record data with the standard data table for the specifiedthermocouple type.

    TASK No.: I-6.4 Continue

    Resistance Temperature Detectors (RTDs)

    The resistance of a conductor usually increase as the temperature increase. If the

    properties of that conductor are known, the temperature can be calculated from the

    measured resistance. A resistance temperature detector (RTD) is a conductor of

    known characteristics constructed for insertion into the medium for temperature

    measurement. Any conductor can be used to construct an RTD, but a few have

    Figure I-6.4F,Typical BenchSet-up forThermocouple

    Accuracy Check.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    24/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 24

    been identified as having more described characteristics than others. The

    characteristics, which are desired, include.

    1. Stability in the temperature range to be measured. The material must not melt,corrode, embrittle or change electrical characteristics when subjected to the

    environment in which it will operate.2. Linearity. The resistance change with temperature should be as liner as

    possible over the range of interest to simplify readout.

    3. High resistively. Less material is needed to manufacture an RTD with aspecified resistance when the material has a high characteristic resistively.

    4. Workability. The material must be suitable for configuring for insertion intothe media.

    The materials which, have been identified as having acceptable characteristics are

    copper, nickel, tungsten and platinum.

    Copper has good linearity, workability, and is able up to 250F (120C), but has

    low resistively, thus either a long conductor or one with a very small cross-

    sectional area is required for a reasonable resistance. Nickel and nickel alloys have

    high resistively, good stability and good workability, but have poor linearity.

    Tungsten is brittle and difficult to work with. Platinum has been accepted as the

    material which best fits all the criteria and has been generally accepted for

    industrial measurement between -300 and 1200 F (-150 and 650C).

    The effect of resistance inherent in the lead wires of the RTD circuit on the

    temperature measurement can be minimised by increasing the resistance of the

    sensor; however, the size of the sensor will also be increased. RTDS are

    commercially available, with resistance from 50 to 1000 ohms at 32F (0C) and

    increase resistance 0.385 ohms for every C of temperature raise.

    TASK No.: I-6.4 Continue

    Chemically pure platinum has a resistance of.392 ohms per C for a 100 ohm RTD

    in accordance with the American (A) standard.

    When the resistance of the RTD is found by measurement, the temperature can be

    calculated:

    C = (Ohms reading -100)/0.385

    The accuracy of this calculation is determined primarily by the accuracy of the

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    25/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 25

    reading. Modem instruments can measure resistance very accurately and the

    temperature can be determined precisely if the resistance of the connecting circuit

    is insignificant or is known. Unfortunately, this resistance usually not negligible or

    known for most practical circuits. The wire that usually used (16 AWG stranded

    copper) has a resistance of approximately 4 ohms per 100 feet (305 m). If it is

    assumed that the RTD is connected to the instrument by a 625- foot cable as

    shown in Figure, the total resistance will be 5 ohms larger than the RTDresistance, which will cause a 23.4 F (13C) error. Furthermore, copper wire has

    a temperature coefficient of about 0.0039 ohms /C/ so the reading will vary about

    a degree for every 20 change in ambient temperature. These errors can be

    compensated by measuring the resistance of every loop and keeping track of the

    ambient temperature. A compromise connection method for RTDS that uses three

    wires and a balanced bridge circuit is shown above. For this circuit, Rl and R2 are

    selected to be the same resistance so that the voltage at the negative terminal of

    the voltmeter is one half o the supply voltage. R3 is selected to be the same

    resistance as the RTD at the base temperature, 100 ohms if 0C is used as the base.

    For this circuit, it is important that wire A and wire B have the same resistance.The usual practice is to run the three wires as a shielded raid, thus they will all be

    the same length and the same resistance within manufacturing tolerance.

    TASK No.: I-6.4 Continue

    Figure I-6.4H, Three-wire RTD circuit with a balanced bridge

    Figure I-6.4G,Two-wire RTD

    Circuit

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    26/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 26

    At the base condition, the positive terminal will also see one-half of the supply

    voltage and the reading will be zero. If 5.2 volts is used to power the bridge, the

    voltage will be 2.6 at each terminal of the voltmeter at the base temperature. When

    the temperature of the RTD is raised one degree C, the voltage reading will

    increase to one millivolt. The symmetry will be upset as the reading moves away

    from the base temperature and the one millivolt per degree will not continue to beexact, but various schemes of completion are available to give an acceptable

    reading.

    The proceeding paragraphs are intended to explain the basis of two, three and four

    wire RTD connections. The selection of resistors and compensation schemes are

    left to the manufacturer of the instrument, but the facilities engineer selects which

    of the connection methods to use. The three wire method is the proper selection

    for virtually all production facility applications.

    Resistance temperature detectors (RTDS) are the most frequently used electronictemperature sensors for production facilities.

    The industry has standardised on RTDS that are calibrated to Din standard 43760

    which is also known as the European standard RTDS which meet this standard

    measure 100 ohms at 0C, are made of platinum and exhibit a resistance increase

    of 0.385 ohms per C temperature increase. Another standard, called the American

    Standard, is available but is not in wide use.

    TASK No.: I-6.4 Continue

    RTDS are usually purchased as a probe assembly consisting of the RTD sensor

    installed in a type 304 stainless steel sheath. The sheath is held in the thermowell

    by a fitting which is threaded on both ends for attachment to the thermowell and

    the head so that the tip of the sheath touches or is very near the end of the well.

    The preferred method of attachment of the sheath to the fitting is with a spring

    assembly which allows the fitting to be screwed into the thermowell as the spring

    is compressed. The spring holds the sheath firmly against the bottom of the well

    for good heat transfer. Another method is to sliver solder the sheath into the fitting

    which makes a good firm assembly, but requires a small clearance from the

    bottom of the well. The third popular method is with a comparison fitting so that

    the sheath can be pushed against the bottom of the well after the fitting is screwed

    into the well. The compression nut is then tightened to hold the sheath. The

    compression fitting

    allows use of a universal probe in different lengths of thermowells.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    27/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 27

    The head of the assembly is a chamber where the leads from the RTD and the

    leads to the receiver instrument can be terminated and connected to each other.

    RTD Accuracy Check

    Figure I-6.4I shows RTD accuracy check set-up. Every six month RTD sensormust be checked against the standard conversion table. RTD element has no repair

    or maintenance procedure.

    Accuracy Check Procedure:

    1. Adjust the multimeter to OHM position.2. Measure the resistance value of the RTD at workshop temperature, suppose it

    is 20 c.

    3. From reference table find the RTD value (107.79 ohms) and check themultimeter reading to be the same value.

    4. Prepare the temperature bath maintained at 100 c.TASK No.: I-6.4 Continue

    5. Immerse the sensing part of the RTD in the bath, when the multimeter readingis stabilised, note the reading.

    6. Check the value of RTD at 100 c. from the table and the multimeter reading tobe the same (138.5 ohms).

    7. Increase the temperature of the temperature bath gradually to 200 c.8. When the multimeter reading is stabilised, note the reading.9. Check the value of the RTD at 200 c. from the table, and the multimeter

    reading to be the same (175.84 ohms).

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    28/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 28

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.5

    Perform servicing and calibration of a pneumatic

    temperature. transmitter

    Reference: OJT Instructor to arrange reference catalogue / Servicemanual for pneumatic temperature transmitter model

    relevant to each working area.

    Materials: 1. Cleaning rags, and2. Solvent.

    Equipment & Tools: 1. Tool Box,2. Stirred heated control bath,3. Standard pneumatic calibrator, and

    Figure I-6.4I, Typical Bench Set-upfor RTD Accuracy Check.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    29/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 29

    4. Standard output gauge.Conditions: Work permit.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe the main parts of pneumatic temperature. transmitter, Understand the principle of operation of pneumatic transmitter, Describe wiring connections of pneumatic temperature transmitter, Describe the procedure for calibrating a pneumatic transmitter,

    Perform periodic maintenance and troubleshooting of a pneumatic transmitter, Discuss an understanding to his trainer, and Write observations and procedures in his workbook.

    TASK No.: I-6.5 Continue

    Details:

    Temperature Transmitters

    Temperature transmitters are used when it is necessary to convert the signal from

    a temperature sensor to one of the standard signals for transmission over a long

    distance or interface with other instruments. The signal is usually 4 to 20 mA. for

    electronic transmission and 3 to 15 psig for pneumatic if a transmitter is used.

    Other signals can be used if required by the receiver, but these are the most

    common and should be used if possible.

    It is also possible to bring a temperature measurement into a control room without

    using a transmitter, a thermocouple / RTD can be wired directly to an instrument

    in the control room and this is acceptable practice. Figure I-6.5A illustratesFoxboro 12A series pneumatic temperature transmitter, which is a force-balance

    instrument, that continuously measures temperature and transmits it as a

    proportional 3 to 15 psi air pressure output signal.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    30/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 30

    TASK No.: I-6.5 Continue

    Principle of operation

    Figure I-6.5A, Pneumatic Temperature

    Transmitter, Foxboro 12A Series

    Figure I-6.5B, Pneumatic TemperatureTransmitter Principle of Operation

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    31/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 31

    As illustrated in figure I-6.5B, any change in the sensor temperature of the gas-

    filled thermal element causes a change in the gas pressure and, therefore, a change

    in the force being applied to the bottom of the force bar. The force bar pivots

    about a cross-flexure, and any motion of the force bar causes a change in the

    clearance between the nozzle and the top of the force bar. This produces a changein the output pressure from the relay to the feedback bellows, until the force

    exerted by the bellows balances the force exerted by the thermal system. The

    output pressure, which establishes the force balance, is the transmitted pneumatic

    signal and, therefore, is proportional to the measured temperature. The signal is

    transmitted to a pneumatic receiver to record, indicate, and/or control.

    TASK No.: I-6.5 Continue

    Calibration

    Figure I-6.5C illustrates bench calibration set-up of a pneumatic temperature

    transmitter, follow the detailed calibration procedure listed in the reference

    catalogue or service manual. Consult your trainer.

    Maintenance and Servicing

    Maintenance and servicing of pneumatic temperature transmitter are limited to

    clean or replace its parts, such as:

    Figure I-6.5C, Pneumatic TemperatureTransmitter Calibration Set-up

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    32/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 32

    1. Supply Air Filter blow out at least once a day,2. Replace Screen Filter of the process inlet,3. Clean Nozzle Assembly,4. Clean booster relay Restrictor,5. Replace thermal element,6. Replace compensating bellows,TASK No.: I-6.5 Continue

    7. Replace Booster Relay,8. Install or Adjust derivative unit,9. Change Range Bar, and10.Adjusting Flexure Cap Screw.Disassembly;

    Normal servicing of the transmitter does not require the removal of any parts other

    than those already mentioned. Further disassembly is not recommended because

    of possible loss of accuracy or damage to the transmitter, detailed servicing

    procedures are mentioned in maintenance section of the selected transmitter

    model. Consult your trainer.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    33/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 33

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.6

    Perform servicing and calibration of an electronic

    temperature transmitter with RTD input.

    Reference: OJT Instructor to arrange reference catalogue / servicemanual for electronic temperature transmitter model

    relevant to each working area.

    Materials:

    Equipment & Tools: 1. Tool Box,2. Standard dc power supply 24 Vdc at 35 mA,3. Resistance decade box, and4. Digital multimeter.

    Conditions: Work permit.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe the main parts of RTD input electronic temp. transmitter, Understand the principle of operation of RTD input transmitter, Describe wiring connections of RTD input transmitter, Describe the procedure for calibrating an RTD transmitter,

    Perform periodic maintenance and troubleshooting of an RTD transmitter, Discuss an understanding to his trainer, and Write observations and procedures in his workbook.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    34/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 34

    TASK No.: I-6.6 Continue

    Details:

    RTD input Electronic Temperature Transmitter

    Temperature transmitters for new installations are predominantly electronic with 4

    to 20 mA. Outputs and inputs from thermocouples or RTDS. These transmitters

    can be mounted in the field and on the thermowell or in the field on a support and

    connected to the sensor by a cable.

    Temperature transmitter mounted in the field must be protected from the elements

    by an appropriate housing. A weatherproof (INEMA 4) housing is adequate for m

    most applications, even in Division 2 hazardous area because there are no arching

    contacts in a typical temperature transmitter. An explosion proof (NEMA 7)

    housing is required for Division I area unless the installation is certifiedintrinsically safe. The energy level required in temperature transmitters is such

    that they can be used in intrinsically safe installations if isolated from the power

    supply and receiver by approved barriers.

    TASK No.: I-6.6 Continue

    Figure I-6.6A, Electronic Temperature Transmitter with RTD Input

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    35/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 35

    Figure I-6.6A, illustrates the main parts of an RTD temperature transmitter, which

    are:

    RTD sensor and thermowell,

    Transmitter electronic housing includes; Range board, Amplifier board and Output

    board.

    Figure I-6.6B, illustrates field-wiring connections of 3-wire RTD input

    temperature transmitter. The dc power supply to be regulated at 24 Vdc grounded

    at negative side terminal. The transmitter will operate with current signal loop 4-

    20 mA proportional to the calibrated temperature range.

    Theory of Operation

    Figure I-6.6C, illustrates RTD input electronic temperature transmitter functional

    block diagram The specific operation of the different functional blocks described

    below:

    TASK No.: I-6.6 Continue

    RTD: Is a platinum temperature sensor, 100 Ohms at zero degree C.

    Resistance Bridge: This bridge converts the resistance versus temperature

    relationship of the sensor to a millivolt versus resistance signal.

    Modulation, ac Amplification, Demodulation: The differential signal from the

    bridge is converted to an ac signal. This signal is then amplified and converted

    Figure I-6.6B, Electronic TemperatureTransmitter Typical Wiring Connection

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    36/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 36

    back to a dc signal, and this to ensure that the signal will not be affected by

    ambient temperature changes.

    Dc Amplification, Current Control: The dc signal is further amplified to drive a

    transistor that controls a current signal that proportional to sensor temperature.

    Voltage Regulation: A voltage regulator circuit provides a stable voltage to

    ensure the signal is independent of supply voltage and load resistance variations.

    Calibration

    The transmitter is calibrated at the specified range on the nameplate, figure I-6.6D

    illustrates calibration set-up of RTD input electronic temperature transmitter. To

    re-calibrate this transmitter, reference catalogue or service manual has the detailed

    procedure for calibrating an RTD transmitter. Consult your trainer.

    TASK No.: I-6.6 Continue

    Figure I-6.6C, Electronic Temperature

    Transmitter Functional Block Diagram

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    37/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 37

    Maintenance

    Repair:

    In case of transmitters failure, the first step is to determine whether the fault lies

    with the sensors or the transmitters electronics, to repair or replace the faulty

    device.

    RTD Test:A platinum RTD with an ice-point resistance of 100 Ohms should read

    approximately as shown in reference R vs. T table. Consult your trainer.

    Electronics Assembly:

    The transmitter is designed for easy replacement of its plug-in, modular circuit

    boards. A malfunction can be most easily isolated by substituting boards one at a

    time until the unit functions properly.

    Detailed procedures for disassembly and reassemble of transmitters electronics

    are listed in the reference service manual. Consult your trainer.

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.7

    What are self-regulating temperature valves? Function

    & application

    Reference:

    Figure I-6.6D, Electronic Temperature

    Transmitter Calibration Set-up

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    38/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 38

    Materials: 1.Cleaning rags, and

    2. Solvent.

    Equipment & Tools: 1. Tool Box, and2. Digital Multimeter orstandard output gauge.

    Conditions: Work permit.

    Requirements by Trainee:

    To study the task and familiarise himself, Understand, zero value of a level transmitter (pneumatic / electronic), Demonstrate zero adjustment of a level transmitter, Be able to perform field zero check of a level transmitter, Discuss an understanding with his trainer, and Write observation in his workbook.

    TASK No.: I-6.7 Continue

    Details:

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    39/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 39

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.8

    Service and calibrate a pneumatic temperature

    controller

    Reference: OJT Instructor to arrange reference catalogue / Servicemanual for pneumatic temperature indicating

    controller model relevant to each working area.

    Materials: 1. Cleaning Rags.Equipment & Tools: 1. Pneumatic calibrator,

    2. Standard test gauges,3. Service/ Repair Kit, and4. Tool Box.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    40/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 40

    Conditions: Work permit

    Requirements by Trainee:

    To study the task and familiarise himself, Understand principle of operation of a pneumatic temperature controller, Describe the effect of the controllers modes (PID) on the output signal, Be able to perform periodic adjustments / calibration of a pneumatic indicating

    controller,

    To perform P.M, service, parts replacement of an indicating controller, Draw/ Sketch calibration set-up in his workbook, Discuss an understanding to his trainer, and Write observations and procedures in his workbook.

    TASK No.: I-6.8 Continue

    Details:

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    41/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 41

    Temperature Controllers

    A temperature controller is a device which senses temperature and manipulates an

    end device to control that temperature. The sensor is one of temperature sensors

    and the end device can be control valve to control any process variable or other

    device.

    TASK No.: I-6.8 Continue

    Pneumatic temperature controller supplied by 20 psi regulated air pressure and the

    output signal is 3 to 15 psi. Full featured, temperature controllers offer more

    precise control than the simple type. The integral and differential control modes in

    addition to the proportional mode will allow stable operation in fast processes

    where simple controller would oscillate between no output and full output. Figure

    Figure I-6.8A, PID Pneumatic TemperatureController Components Location

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    42/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 42

    I-6.8A illustrates a full featured, pneumatic temperature controller, in which use

    an external temperature sensor.

    Principle of Operation

    As shown in figure I-6.8B A PID pneumatic temperature controller. This

    temperature controller operate on the motion balance principle; motion from a

    pneumatic feedback unit balances the motion from a process measuring element.

    When the temperature at the sensing element increases, the bourdon spring uncoils

    and moves the process pointer to the right and the baffle-actuating pin to the left.

    The movement of the pin lowers the baffle to decrease the nozzle-baffle gap and

    increase the nozzle-back pressure. This pressure is fed to chamber A of the output

    relay.

    As the pressure in chamber A increases, the diaphragm assembly moves the relay

    stem downward closing the vent port and opening the air supply port to increasethe output. The output increases until it balances the downward force on the

    diaphragm assembly.

    By proportional response, the output pressure is fed to the follow-up bellows and

    raises the baffle-actuating pin and baffle and the output change is proportional to a

    change in process measurement.

    Reset response, automatically returns the process variable to the set-point after a

    sustained load change. This is accomplished by opposing the action of the follow-

    up bellows with a reset bellows.

    Pre-act response, reduces the offset caused by a process disturbance as well as

    reduce the recovery time following the disturbance. This is accomplished by

    feeding the output pressure to the follow-up bellows through a needle valve.

    The controller can be set for either direct or reverse action by positioning the gain

    dial.

    TASK No.: I-6.8 Continue

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    43/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 43

    TASK No.: I-6.8 Continue

    Calibration

    Calibration adjustments of pneumatic temperature controller are limited to

    perform the following:

    Figure I-6.8B, Schematic Diagram of PID

    Pneumatic Temperature Controller

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    44/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 44

    1. Process pointer calibration; zero adjustment, span adjustment and linearityadjustment.

    2. Pneumatic-set pointer calibration; zero adjustment and span adjustment.3.

    Controller alignment; fixed high controller, differential-gap controller,

    proportional controller, proportional plus reset controller, proportional plus

    pre-act controller and proportional plus reset plus pre-act controller.

    4.Nozzle-height adjustment.Detailed procedures to perform the above adjustments are mentioned in the

    reference service manual of the relevant pneumatic temperature controller model.

    Consult your trainer.

    Periodic Servicing

    If the air supply is clean and dry the instrument should be serviced once a year. If

    the air supply is dirty or oily, more frequent servicing may be required. Servicing

    the pneumatic controller is limited to clean / repair / replace its parts; such as:

    1. Clean nozzle-tip and baffle-surface,2. Clean or replace the output relay,3. Clean the manual regulator,4. Clean reset and ore-act restrictors, and5. Replace internal O-Ring seal of auto-manual switch.Detailed procedures to perform the above services are mentioned in the reference

    service manual of the relevant pneumatic temperature controller model. Consult

    your trainer.

    TASK No.: I-6.8 Continue

    Troubleshooting:

    Reference catalogue or service manual of the pneumatic temperature controller

    model used has the details of possible causes ofcontrollers problems and actions

    to be taken to overcome these problems. Consult your trainer.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    45/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 45

    MODULE No.: I-6 Temperature Instruments

    TASK No.: I-6.9

    Perform calibration of an electronic transmitter using

    thermocouple input.

    Reference: OJT Instructor to arrange reference catalogue / service

    manual for Thermocouple input electronic temperaturetransmitter model relevant to each working area.

    Materials: None

    Equipment & Tools: 1. Tool Box,2. Standard dc power supply 24 Vdc at 35 mA,

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    46/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 46

    3. A stirred ice bath,4. Thermocouple reference junction, and5. Digital multimeter.

    Conditions: Work permit.

    Requirements By Trainee:

    To study the task and familiarise himself, Be able to describe the main parts of thermocouple input temp. transmitter, Understand the principle of operation of thermocouple input transmitter, Describe wiring connections of thermocouple input transmitter,

    Describe the procedure for calibrating a thermocouple input transmitter, Perform periodic maintenance and troubleshooting of a thermocouple

    transmitter,

    Discuss an understanding to his trainer, and Write observations and procedures in his workbook.

    TASK No.: I-6.9 Continue

    Details:

    Figure I-6.9A Elect. Temp. Transmitter withThermocouple Input Functional Block Diagram

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    47/48

    TASK DETAILING MANUAL

    Module I-6 www.arfanali.webs.com Page 47

    Theory of Operation

    Figure I-6.9A, illustrates thermocouple input electronic temperature transmitterfunctional block diagram The specific operation of the different functional blocks

    described below:

    Thermocouple: A thermocouple consists of two specific dissimilar metals joined

    at the measurement site, the thermocouple produces a millivolt-level signal

    proportional to process temperature.

    Bridge: Cold-junction compensation for the thermocouple is provided by two

    compensation resistors, the output from these resistors and the temperature-

    sensitive simulates that of a thermocouple temperature.

    Modulation, ac Amplification, Demodulation, dc Amplification, Current

    Control: Identical to that of RTD input transmitters.

    TASK No.: I-6.9 Continue

    Isolation: Isolated power to run the unit, is provided by a dc-to-ac converter,

    which feeds a transformer. On the secondary side of this transformer, diodes

    rectify that ac to provide dc power to the circuitry.

    Voltage Regulation: Regulates the voltage across the dc/ac converter, to protecttransmitters electronics against damage due to reverse power hook up.

    http://www.arfanali.webs.com/http://www.arfanali.webs.com/http://www.arfanali.webs.com/
  • 7/29/2019 Module I-6 Temperature Instruments

    48/48

    TASK DETAILING MANUAL

    Maintenance

    Repair:

    In case of transmitters failure, the first step is to determine whether the fault lies

    with the sensors or the transmitters electronics, to repair or replace the faulty

    device.

    TASK No.: I-6.9 Continue

    Thermocouple Test:

    A thermocouple could be tested with a reference junction relevant THE

    thermocouple type and should read approximately as shown in reference mV vs. T

    table. Consult your trainer.

    Electronics Assembly:

    The transmitter is designed for easy replacement of its plug-in, modular circuit

    boards. A malfunction can be most easily isolated by substituting boards one at a

    time until the unit functions properly.

    Detailed procedures for disassembly and reassemble of transmitters electronics

    are listed in the reference service manual. Consult your trainer.

    Figure I-6.9B, Temperature Transmitter

    Thermocouple Input Calibration Set-up


Recommended