What is sterilization?
DefinitionsSterile: Free from any living
organisms
Sterilization: Process of killing orremoving microorganisms from aproduct to insure that it is sterile
Sterilizing Agents Type of agents Chemical Physical Moist heat Dry heat
Most clinics use physical agents
Steam Sterilization: Autoclave Pressurized device that uses heat, steam and pressure to achieve
sterilization or decontamination
Steam – Advantages Simple process: The only two parameters to control are
time and temperature. Highly reliable and easilycontrolled. Widespread capability: All hospitals and many other
heath care facilities have steam sterilizers. Excellent process for reusable medical devices that are
not adversely affected by temperatures >121C (up to 135C). Excellent process for liquids that are not heat sensitive.
Steam – Disadvantages Comparatively high temperature required: Many
products and packaging materials cannot toleratetemperatures of >121C. Generally not appropriate for most single use disposable
medical devices produced in high volumes and sold assterile.
Principles of Steam Sterilization
u Sterilization: the complete destruction of allforms of microbial life, including bacterial spores
u Steam under pressure reaches high temperatures
Principles of Steam Sterilization (cont.)
u The “killing power” of steam is due to its latent heatof vaporization
80 calories: 1 L water to boiling540 calories: 1 L boiling water to steam
Action of Steamu Steam contacts a cooler surface, condenses, causing
a huge decrease in volume and setting up a negativepressure that draws more steam
u Condensation occurs as long as there is atemperature differential
u Action of steam ensures: Surface heating,penetration, and protein coagulation
Items that CAN be Autoclaved: Cultures and stocks of infectious material Culture dishes and related devices Discarded live and attenuated vaccines Contaminated solid items such as: petri dishes,
eppendorf tips, pipettes, gloves, paper towel Items for sterilizations such as; glassware media,
aqueous solutions
DO NOT Autoclave: Any liquid containing: Toxic, volatile, or corrosive chemicals e.x. bleach, formalin, glutaraldehyde Check MSDS before hand
Cancer drugs Do not break down under high temperatures Highly toxic
Radioisotopes
Parts of a Steam Sterilizer Control panel with graph Door handle Outer metal jacket-like chamber built around the
inner chamber Inner chamber Series of filters Temperature measurement of steam sensor
Components of a Steam Sterilizer Source of steam- generate own steam or rely on steam
from outside source Preheat metal jacket build around chamber for outside
sources reduces total cycle time Three main pipes1. A pipe with inline filters to bring steam to the sterilizer
and the chamber2. A drain pipe for steam, air and water from the chamber3. A pipe to deliver filtered air to the chamber at eh end of
the cycle
Steam Jacket: Surrounds the inner chamber and storessteam until pressure has been obtained.
Inner Chamber: Items to be sterilized are loaded intothis compartment without touching.
Construction of Steam Sterilizers Valves: Pressure regulating: Maintains the pressure of 15 to 20
psi in the jacket.
Safety Valve: Prevents excess pressure build-up.
Steam trap: If pressure to high, valve opens and releasespressure.
Construction of Steam Sterilizers Gauges: Located on front of sterilizer. Reading chamber pressure Reading Jacket pressure.Safety Steam-lock Door:
Located in front of chamberWeakest part of sterilizer
How to Autoclave1. Which cycle to use2. Preparing your items for autoclaving3. Loading the autoclave4. Choosing the cycle settings5. Unloading the autoclave6. Aborting Autoclave
How to Autoclave- Which cycle touse
Autoclaves run 3 types of cycle programs
The type of cycle depends on what is being autoclaved:
Liquid/Slow exhaust * For autoclaving liquids* Prevents liquids from boiling over
Solid/Gravity * Best for unwrapped solid items(ie glassware)
Solid/Vacuum * Best for wrapped solid items
How to Autoclave - Preparing your items Sign into log book Keeps track of autoclave use for
maintenance records
Use personal protective equipment! Eye protection Heat resistant gloves Lab coat
How to Autoclave - Preparing your items Use a primary container Container comes into direct contact with the contaminated or non-
sterilized material or fluid Do not fill more than 75% of holding capacity
How to Autoclave - Preparing your items Use a primary container (cont’d) Must NOT be a tightly sealed container (might explode) Primary container must permit heat (steam) penetration Loosen screw caps or use self venting caps Cap open containers with aluminum foil or muslin If using plastic waste bags, leave a small opening
Loosen screw capsSmall opening
How to Autoclave - Preparing your items Use a secondary container Used to contain any spills The sides of the secondary container must be sufficiently high to
contain any spill that may occur Tray MUST be autoclave safe
For liquids add water
How to Autoclave - Preparing your items Do not over load primary or secondary containers Do not fill more than 75% to allow expansion without overflow
Do not pack or compress contents
Unautoclaved Autoclaved (regular) Autoclaved (due to beingcompressed & sealed)
How to Autoclave - Preparing your items Use temperature sensitive tape Will indicate that high temperature has been achieved Will not prove that decontamination or sterilization was successful Will assist in keeping track of autoclave and non autoclaved items
Before After
How to Autoclave - Loading the autoclave NEVER autoclave liquid loads using a solid load cycle
liquids may boil over
Be cautious if autoclave was recently used Rack may be hot Use heat protective gloves
Make sure tray is locked on dolly Rack may slide out unexpectedly if not locked
How to Autoclave - Loading the autoclave Farquharson: Tighten door so that prongs are
fully extended If noise and/or steam escapes,
abort the cycle and tighten thedoor more
Lumbers Keep the door button pushed until
the ready prompt is on the screen
How to Autoclave – Choosing the Cycle For Liquids: 20 mins / litre of liquid, 5 mins per additional litre
For Solids: Glassware (empty): 15 mins Instruments (utensils): 30 mins Biohazardous Garbage: at least 30 mins per full bag Use biological test strips to optimize duration
How to Autoclave – Unloading autoclave
Wait for autoclave to state END CYCLEbefore opening door
When opening, stand away from dooropening Make sure no one is standing by door
opening
How to Autoclave – Aborting Only qualified personnel should attempt to troubleshoot an autoclave
Farquharson Large autoclave: push abort button Small autoclave: need to manually advance through the autoclave cycle
Lumbers Push abort button
Autoclave: PerformanceIndicators How to know if autoclave is functioning correctly:
Physical - Annual testing by certified technician- Pressure, Temperature, Cycle times, recorded on paper
Chemical - Heat sensitive autoclave tape- Not an indicator of successful sterilization, useful to keep track ofautoclaved and unautoclaved items
Biological - Tests ability of autoclave to sterilize effectively- Bacillus stearothermophilus spore strips often used because theyare resistant to steam sterilization.
- EZ Test (SGM Biotech) (Fisher Sci #29801 074)- 3M Attest Rapid Readout Biological Indicators- Steris Verify Integrator Laminated and EO Integrators
DART - Daily Air Removal Test, verifies complete air removal and checksfor leaks.
Biological Indicator Controls/SporeTests
Bacillus Stearothermophilus:is used strictly for Steamand dry heat sterilizers.
Tests ability of sterilizer to kill microorganisms.
Biological Indicator Must be incubated: Follow manufacturers instructions
Use correct temperature 131-140 degrees F.
Use special incubator
Incubate for 48 hours or follow manufacturersinstructions
Biological Indicator Record results: Negative- no color change from original.
Positive- color changed is usually amber in color.
Take sterilizer out of service Report to supervisor Recall all items sterilized in sterilizer for
last 24 hour period
Types of Steam Sterilizers Gravity Air Displacement Sterilizer- using gravity to
remove air from the inner chamber Prevacuum Steam Sterilizer- uses vacuum in two
cycles to remove air from the inner chamber
Types of Steam Sterilizers
1. Gravity displacement.
a. Definition: Gravity pushes air through thepackages and down through the drain. Sterilizationbegins when steam passes the thermometer andreaches the desired temperature.
Gravity Displacement Settings: Temperature- 250 degrees F. to 254 Degrees F.
Pressure- 15 to 17 pounds per square inch(PSI)
Exposure time- 30 minutes for wrapped instruments.Minimum 15 minutes.
•Pre-vacuum (high temperature)Sterilizer
Definition: Air is completely evacuated from thechamber by a vacuum. The steam-injector helpseliminate the air out of packages. Steam thenpenetrates the packages on all surfaces.
Pre-vacuumSettings:
1. Temperature: 270 degrees F. to 276 degrees F.
2. Pressure: 27 PSI
3. Exposure Time: 15 to 30 minutes.
Sterilization: How is it validated?Steam IQ, OQ, PQ of steam sterilizer Temperature profiles of product during sterilization
cycles (using thermocouples) Fractional and/or half cycles with product & B.I.s (G.stearothermophilus) Then sterility testing of product & B.I.s
References
ANSI/AAMI/ISO standards and TIRs available atwww.aami.org USP 25 – order from www.usp.org PBL Compliance Guides Sterility Assurance Compliance Assessing Biocompatibility Online at www.pacificbiolabs.com