Date post: | 21-Oct-2015 |
Category: |
Documents |
Upload: | gowri-krishna-chikkala |
View: | 14 times |
Download: | 0 times |
OBJECTIVE To provide Basic concept of measurement system Detail application methodology of all
statistical studies for variable and attribute data
Methods to meet QS9000 & ISO/TS-16949 requirements
IntroductionUses of measurement data are Decision to adjust a manufacturing process Determine the significant relationship
between two or more variables using statistical studies
The result of analytical studies is depends upon the Quality of data produced by measurement system
Why MSA required?
Defect prevention
Prediction
Statistics
Data
Inspection
SPC
MSA is mandatory prior to SPC
PURPOSE OF MSA Quantify Variation present in the
Measurement system Compare the Consistency of
inspectors Provide methods for Validation of
Measurement System
Measurement Systems Process to assign a number or decision to a characteristic Process
Appraiser Checking Method Instrument Environment
E.g, OD checking, Concentricity checking
Statistical properties of MS
Ideal Measurement system is the one Which produces always only the
correct measurements Always agree with the standard Statistical data produced over time
Statistical properties of MSProperties of “GOOD” Measurement Systems : Adequate discrimination & Sensitivity System is always under statistical control Variability of the measurement system is
small compared to Manufacturing process variation and Tolerance
Criteria for MS design selection Team members Purpose of use Frequency of use Part specification Sensitivity required Contact? Non contact/ Auto? Manual?………………
Statistical properties of MSClassification Measurement system errors are: Bias Linearity Stability Repeatability Reproducibility
Statistical properties of MSDetermination of these errors will help us to: Criteria to accept new MS Compare one against another To evaluate a suspect gage To evaluate the effectiveness of repair
QS9000 / ISO-TS16949 Requirements
Define a process for implementation of MS Identify all types of measurement systems Cover all systems covered in Control Plan Perform Appropriate statistical study Use MSA ( or Customer approved) Manual Obtain Customer approval for use of
methods other than MSA manual
QS9000 / ISO-TS16949 Requirements
Types of MS is based on L.C Range Construction Parameter Manufacturer
QS9000 / ISO-TS16949 Requirements
Vernier
L.C – 0.02 mm
Range – 0 – 200mm
Construction- Ordinary
Parameter- O.D
Vernier
L.C – 0.02 mm
Range – 0 – 200mm
Construction- Ordinary
Parameter- Width
Is it Same or different type of MS ?
QS9000 / ISO-TS16949 Requirements
Bore Dial Gage
L.C – 0.001 mm
Range – 18-35 mm
Construction- Ordinary
Parameter- Bore
Air gage
L.C – 0.001 mm
Range – ---
Construction- Dial
Parameter- Bore
Discrimination Ability to detect the small changes in
process Traditional rule
1/10th of Tolerance 1/3rd Considering the cost/ criticality
MSA/ SPC requirement 1/10th of Process variation Resolution 1/10th of Process variation
ACCURACY & PRECISION ACCURACY Closeness to reference
or master value Required where two or
more MS measuring a same characteristic
Same parameters are checked at Suppliers end or at Customer end
PRECISION Ability of MS to repeat
the same reading Required where MS is
repeatedly used to assess and adjust the process
In process inspection as per control plan
ACCURACY & PRECISIONACCURACY
Captured by
BIAS LINEARITY STABILITY
PRECISION
Captured by
REPEATABILITY REPRODUCIBILITY
BIAS - STEPS1. Capture reference value
• Use masters calibrated traceable to standards
• Measure by better instrument than the instrument being measured
• By Layout inspection
2. Measure the part/ master by Min. 10 times
BIAS - STEPS3. Calculate BIAS for individual reading &
draw histogram
4. Calculate , bias, t limits
APPENDIX C t distribution table
ACCEPTANCE CRITERIA
Zero must lie between t Limits of Bias distribution or
Bias should be within acceptable gage error as defined in Calibration Proc.
CAUSE OF LARGE GAGE BIAS
• Error in Master
• Instrument
- Worn
- Made to wrong dimension
- Measuring wrong characteristic
- Improper use by operator
- Improper calibration
APPLICABILITY OF GAGE BIAS• Not a Problem like R&R
• Required where Accuracy to be captured
• Being Carried out as part of Calibration
• Applicable for Gages designed by the Organization
Linearity – Steps 1. Collect 5 parts / masters over the operating range of the gage through process variation
2. Capture reference value for the 5 parts3. Measure at least 12 times in a random manner to cover all 5 parts
4. Measure at least 12 times in a random manner to cover all 5 parts
Linearity – Steps 5. Calculate Average, BIAS and enter the data in a Spread sheet
6. Draw Best fit line for the average, Lower limit and Higher limit of Bias
Linearity – Acceptance criteriaZERO LINE MUST BE CONTAINED WITHIN THE TWO LIMIT LINES OR WITHIN ACC.CRITERIA OF CAL ERROR
0
+ VE
- VE
Upper t limits
Lower t limits
Is this MS acceptable ?
Causes for Gage Linearity•Error in Master
•Instrument
-Worn
-Made to wrong dimension
-Measuring wrong characteristic
-Improper use by operator
-Improper calibration
Applicability of Gage Linearity• To Capture Accuracy over the operating range to assure its suitability of gage for the entire process variation
• Required for Special gages designed by the Organization
• Can be performed as part of Calibration
Stability – How?No.of Samples/ Masters
1 or 3 at Minimum, Mid and Higher Specification limits
Subgroup Size 1 or 3 times over a period in the Morning, Noon and Evening
Frequency Daily,Weekly, Monthly based on the stability assessment of instrument
Charts One chart for each sample, X-MR , or X bar& R or X bar & S
Stability – Acceptance criteria• No out of Control conditions in Control charts with careful interpretation
• No drift
• No trend
• Zero must lie within two limits of bias variation
Stability – Applicability• To compare two or more measurement
systems measuring a same characteristic
e.g Bore dial gage & Air gage
• To establish or evaluate the Calibration Frequency
Constitution of total variation Total variation
Manufacturing process variation Part to part variation Within part variation
Measurement system variation Appraiser variation Equipment variation Appraiser part interaction
Methods to capture variationElements of Variation
Range Method
X bar & R Method
ANOVA Method
TV
PV
WIV
AV
EV
App & part interaction
Repeatability & ReproducibilityRepeatability
• One Appraiser
• One Equipment
• Same part
• Several trials
This Variation is represented by Equipment
Reproducibility
• Same equipment
•Same Parts
•Several trials
•Different AppraiserThis Variation is represented by Appraiser
Range method
No.of Samples 5 nos. gives 80 % chance of detecting a wrong MS.
10 nos gives 90% chance
No.of Appraisers Min. 2
Formula Sigma = R bar / d2*
Range method - ExerciseAppraiser
A1. 15.032. 15.043. 15.024. 15.055. 15.02
Appraiser B
15.0515.0415.0315.0515.02
Range
0.02
0.00
0.01
0.00
0.00
Tol = 0.05, TV(SPC)= 0.120
Range method - Exercise
Average R = 0.006
Std.deviation = 0.006 / 1.128 = 0.0053
where n=2 d2 is1.128, n=3 d2 is 1.693R&R ( 1 sigma) = 0.0053
% R&RTV = (0.0053)/ (0.120/6) * 100 = 26.5
Interpretation = MS IS NOT ACCEPTABLE
% R&RTOL = (0.0053)/ ( 0.050/6) * 100= 66.5
Range method - Exercise
COMPARE AGAINST TOLERANCE WHEN
• TV IS MORE THAN TOLERANCE
• WHERE SPC IS NOT IMPLEMENTED
COMPARE AGAINST TOTAL VARIATION WHEN
• TV IS LESS THAN TOLERANCE
• WHERE SPC IS IMPLEMENTED
Range method – Acceptance criteria
% R&R < 10 % OF TOL OR TV
% R&R 10 – 30 % OF TOL OR TV ACCEPTABLE SUBJECT TO ANALYSIS AND JUSTIFICATION W.R.T COST OF REPAIR AND CRITICALITY
Range method – Applicability
Being the Short study this can be used for Periodic verification of R&R to detect any changes rather than average range method
Being the short study it is suitable for studying MS where it is time consuming to measure one measurement e.g Concentricity, R/O
Average & Range MethodNo.of Samples 10 nos. Randomly selected
representing actual process variation
No.of Appraisers 3 From personnel actually performing on a day to day operation
No.of trials Min 2 . Preferably 3
Average & Range Method- steps for conducting study
1. Identify the parts from 1 – 10
2. Mark the Place where to check to eliminate Within part variation
3. Communicate the purpose of the study to all appraisers
4. Conduct the study in a Random manner covering all parts for required no.of trials through all appraisers 5. Monitor all appraisers are following the same method
Average & Range Method- steps for conducting study
6. Record the observations in a manner which is not seen by the appraisers – FORMAT-7. If any abnormal readings are observed ask the appraiser to repeat the reading
8. Maintain the samples till the completion of analysis
10. Calculate EV,AV,PV & TV and it Percentage against TOL or TV and ndc
9. Draw R chart and X bar chart and interpret
X bar & R chart Method – Acceptance criteria
% R&R < 10 % OF TOL OR TV
% R&R 10 – 30 % OF TOL OR TV ACCEPTABLE SUBJECT TO ANALYSIS AND JUSTIFICATION W.R.T COST OF REPAIR AND CRITICALITYNo. of distinct data categories
ndc > 5 where SPC is used
ndc >2 where no SPC is used
R chart interpretationCONDITION
One or more than one point of one appraiser out of UCL R
One or more than one point of all appraisers out of UCL R
In one part all appraisers points are out of UCL R
INTERPRETATION His method is
different from others
Measurement System is sensitive to appraisers skill
Part is deformed or Damaged
X chart interpretationCONDITION
More than 50 % of readings are out of control limits
INTERPRETATION
Measurement system is adequate enough to capture process variation
Causes for R&RWhen repeatability is large compared to
reproducibility
Instrument needs maintenance
Redesign gage for more rigidity
Improve clamping or location of gauging
Excessive within – part variation
( Conduct R & R with within part variation )
Causes for R&RWhen reproducibility is large compared to
repeatability
Appraisers need better gage use training
Need better operational definition
Incremental divisions on instrument are not readable
Need fixture to provide consistency in gage use.
X bar & R method - Applicability• To capture Precision, Mandatory before SPC study
• Measurement Systems which are used repeatedly on line to control the process based on the measurement data
• Where individual readings are affected by the precision of the instrument
• Applicable for all MS used in In process where the MS can be repeated
ATTRIBUTE MEASUREMENT
• COMPARES EACH PART TO A SPECIFIC SET OF LIMITS AND ACCEPTS THE PART IF THE LIMITS ARE SATISFIED
• IS DESIGNED TO ACCEPT/REJECT A SET OF MASTER PARTS
• CANNOT INDICATE HOW GOOD OR HOW BAD A PART IS,ONLY WHETHER THE PART IS ACCEPTED OR REJECTED (PASS/FAIL)
• VISUAL STANDARDS MAY RESULT IN 5 TO 7 DISTINCT
DATA CATEGORIES. MSA MANUAL DOES NOT PRESCRIBE ANY METHOD OF EVALUATION OF THE SAME
• GPC METHOD IS MOST APPROPRIATE ANALYSIS• RISK ANALYSIS METHOD IS CUMBERSOME, BUT IT IS PREFERRED BY SOME CUSTOMERS (Eg, FORD)
GPC KEY INSTRUCTIONS Select the gage to be studied Calibrate the gage to know the actual size of the
gage Understand the distribution and location of the
process Decide the on which side of specifications
samples to be taken and the measurement method Select samples (8) and repeat minimum 20 times
on a random manner and count no.of accepts ( a )
GPC KEY INSTRUCTIONS Select Minimum 8 samples out of that One sample on lower than specification, no.of
accepts a = 0 One sample on higher specification, no of accepts
a = 20 Six parts on the gray area where, a is more than
or equal to 1 and less than or equal to 19 Select sample repeat the study till we meet the
above criteria
GPC KEY INSTRUCTIONS Calculate Probabilities of acceptance using
formulae Plot Probabilities of acceptance as % in
normal probability paper Calculate Bias & Repeatability Conclude the decision on the acceptability
of gage
RISK ASSESSMENT – Key steps Take 50 parts representing entire process
variation having bad, marginally bad, good & marginally good parts
Mark as 1 to 50 Select 3 appraisers Conduct study in a random manner and
record the decisions, 1 as OK and 0 as Not ok
RISK ASSESSMENT – Key steps Get the reference value for all the 50 parts
using layout inspection/ variable instrument and decide its status as 0 or 1
Compare each trial of each inspector with the another inspector for their decision
Complete the cross tabulation table Calculate Kappa for
A Vs B, A vs. C , B Vs C A vs. Ref, B Vs Ref., C vs. Ref.
Risk analysis – Acceptance criteriaDecision Effectiveness
Acceptable for the appraiser
More than 90 %
Marginally acceptable for the appraiser
More than 80%
Unacceptable for the appraiser – Need improvement
Less than 80 %