Multielement analysis of a municipal landfill leachatewith total reflection X-ray fluorescence (TXRF). A comparisonwith ICP-OES analytical results
Franco Cataldo
Received: 28 December 2011 / Published online: 17 April 2012
� Akademiai Kiado, Budapest, Hungary 2012
Abstract A complete analysis of a landfill leachate
coming from a landfill site of several years old was per-
formed with a total reflection X-ray fluorescence (TXRF)
spectrometer in comparison with an inductively coupled
plasma optical emission spectroscopy (ICP-OES). The
results of the two analytical techniques are compared and
advantages and drawbacks emphasized. The TXRF ana-
lytical technique appears a reliable, economic, rapid and
simpler technique for the everyday monitoring of the
composition of the landfill leachate before the purification
treatment and after the treatment to check the quality of the
resulting purified water. The TXRF and the ICP-OES
analytical techniques were also employed in the analysis of
three groundwater samples.
Keywords Landfill leachate � Groundwater � TXRF �ICP-OES � Multielement analysis
Introduction
Total reflection X-ray fluorescence (TXRF) is a very fas-
cinating analytical technique for elemental analysis even of
complex matrices. TXRF has been reviewed extensively
[1–8] showing that this analytical technique has reached a
remarkable success in elemental analysis in recent years.
TXRF provides significant advantages compared to other
methods of trace element analysis. For instance: it detects
exceptionally small mass quantities (ng or pg); it is mul-
tielemental, meaning that it can simultaneously (in one
spectrum) analyze almost all the elements above a certain
atomic numbers higher; it has the ability to perform the
direct analysis of solid deposits (like membranes); the time
of analysis is short (typically 100–400 s) and the sample
preparation is minimal.
The TXRF technique was initially adopted in the quality
control of semiconductor fabrication [9] but then was
extended in environmental analysis and especially in water
analysis [8, 10, 11] including rainwater analysis [12],
wastes and leachates [13]. Many efforts were dedicated to
the improvement of the TXRF analytical technique looking
at the calibration and sample preparation [14–17] or other
equipment and design aspects [18–20].
Examples of application of TXRF analytical technique
are not limited to water analysis, there are reports on the
use of this technique in the analysis of metals in coal [21],
trace elements in diesel soot [22], trace metals in sugar
cane spirits [23], sweat extractable metals from textile
fabrics [24]. TXRF is an ideal technique in the materials
analysis [25], glasses [26], materials inhomogeneity as in
the case of boron carbide [27], archaeological samples [28]
and medieval pigments [29]. TXRF can be applied to the
determination of certain metal ions, after their concentra-
tion on a selective membrane [30]. In biological matrices
the TXRF is used for the determination of trace metals in
blood and in hairs [31, 32], in the determination of trace
elements in human cancerous and healthy tissues [33], as
well as in the search of trace elements in the blood of seals
of the North Sea [34] and in lichens [35].
Despite this large applications of the TXRF technique,
coming back to water and wastewater analysis, it can be
affirmed that relatively little work has been made in the
analysis of heavily polluted waters [36, 37] and in partic-
ular the analysis of the polluted water known as landfill
leachate. The present work is dedicated to the analysis of
the water coming from a landfill leachate. The TXRF
F. Cataldo (&)
Bioenergy Srl, Via Casilina 1626/A, 00133 Rome, Italy
e-mail: [email protected]
123
J Radioanal Nucl Chem (2012) 293:119–126
DOI 10.1007/s10967-012-1777-z
analytical results will be compared with those produced
with an optical inductively coupled plasma spectrometer
(ICP) on the same landfill leachate sample.
Experimental
Some aspects of the TXRF spectrometer employed
in the present study
The TXRF spectrometer used in the present work, the S2
Picofox from Bruker, involves the total reflection of the
X-ray beam on a reflective medium like polished quartz
glass or polyacrylate glass disc (Fig. 1). The sample is
deposited as a thin film where the X-ray spot is focused and
the sample fluorescence is collected by a detector as shown
in Fig. 1. The X-ray tube of S2 Picofox is air cooled and
made in metal-ceramic with Mo or W target producing a
spot size of 1.2 9 0.1 mm2 with an energy of 35 W/mm2.
Alternatively, it is possible to adopt a microfocus tube as
the X-ray beam source, so that a spot of 0.05 9 0.05 mm2
is obtained with a specific energy of 4,000 W/mm2 and in
this configuration the detection limits are 5–10 times better
than the standard configuration and become comparable to
the KW TXRF spectrometers. As shown in Fig. 2, with the
Mo target it is possible to detect nearly all elements from
Na–U with the exclusion of Zr, Nb, Mo and Tc. In the
present work we have used the Mo target. However, it is
possible to mount the W target and in this other case the
range of detectable elements is comprised from K–U with
the exclusion of Hf, Ta, W, Re. With the W target it is
possible to detect Zr and Mo and it is possible to achieve
improved detection limits for Cd, Sn, Sb (\50 ppb) in
comparison to the Mo target. The monochromator of the S2
Picofox is a flat Mo/B4C multilayer with 80 % reflectivity.
The detector used in the S2 Picofox is a silicon drift
detector (10 mm2 XFlash�) having an energy resolution
\150 eV. This detector is extremely fast in its response
with no dead time. Furthermore it does not need any
external cooling for instance with liquid nitrogen since it is
Peltier-cooled.
Materials and equipment
Landfill leachate sample was obtained from a landfill
located in the central Italy. The multielemental analysis on
the landfill leachate samples was performed on the TXRF
spectrometer S2 Picofox from Bruker. As reference, the
same sample of landfill leachate was also analyzed
according to standard procedure on an ICP-OES. Use was
made of a Perkin-Elemer Optima 7300 V ICP-OES.
Fig. 1 The TXRF configuration involves the total reflection of the
X-ray beam on a reflective medium like polished quartz glass or
polyacrylate glass disc. The sample is deposited as a thin film where
the X-ray spot is focused
Fig. 2 Elements which can be
detected with the TXRF
spectrometer S2 Picofox with a
Mo target. Nearly all elements
can be detected from Na to U
with the exclusion of Zr, Nb,
Mo and Tc
120 F. Cataldo
123
Experimental TXRF apparatus
The energy dispersive S2 Picofox TXRF spectrometer from
Bruker used in the present work was equipped with a Mo-
tube and a 30 mm2 silicon drift detector. The microfocus
X-ray tube is air-cooled with a power of 50 W. As the
detector was Peltier cooled, no additional cooling unit (e.g.
liquid nitrogen) was needed. The total power consumption
of this benchtop system was only 180 W. No external
calibration is needed for quantitative measurements of
element concentrations ranging from high wt% to the low
ppb range. With S2 Picofox all elements from Al–U can be
detected, but even Na and Mg are detectable although with
some difficulty.
Sample preparation and measurements
Sample preparation for qualitative analysis was done by
transferring 10 ll sample material to a quartz glass sample
carrier with subsequent drying in vacuum. For quantitative
analysis 1 ml of sample was mixed with 10 ll of a Ga-
solution (10 mg/L) and 100 ll of a polyvinyl alcohol
solution. After thorough homogenization 10 ll sample
were transferred to quartz glass sample carriers and dried in
vacuum.
The samples were analyzed with the following TXRF
parameters: tube setting at 50 kV and 600 lAmp. Mea-
surement time employed 1,000 s.
Analysis and quantification
In general all elements from sodium to uranium (with the
exclusion of the elements zirconium, niobium, molybde-
num and technetium) can be analyzed by the S2 Picofox.
The TXRF analysis is based on internal standardization.
Therefore, an element, which is not present in the
sample, must be added for quantification purposes.
The complete process of analysis and quantification is
described by the following steps:
(a) Measurement of the complete spectrum. All detect-
able elements are measured simultaneously.
(b) All identified elements are be marked for further
quantification, which can be done manually or
automatically by the software.
(c) Spectral deconvolution.
(d) On the basis of the chosen elements, the software
performs the deconvolution of the spectra. The net
intensities of the element peaks are calculated with
regard to corrections of line overlaps, background
factors, escape peak correction etc.
(e) Calculation of concentrations.
(f) The element concentration is calculated by the simple
formula:
Cx ¼ Nx=Sx½ � Nis=Sis½ ��1Cis
where N is the net intensity, S the relative sensitivity and C
the concentration—each either of the analyte x or the
internal standard is, as indicated. The S2 Picofox can
measure the elements Al–Y using spectral K-lines. Ele-
ments Ru–U are measured using L-lines. The typical lower
limits of detection (LLDs) of the S2 Picofox for many
elements is close to or below 1 lg/L.
Results and discussion
TXRF as the simplest analytical technique
for the monitoring of the metal ions in a leachate
Landfill leachate varies widely in composition depending
on the age of the landfill and the type of wastes stored in
the landfill. A young leachate is acidic while a late-type,
older leachate is basic and rich in ammonia and ammonium
ions. A landfill leachate is usually dark in color and con-
tains both dissolved and suspended material. The genera-
tion of leachate is caused principally by the rain waters
percolating through waste deposited in a landfill. The
percolating water passing through the decomposing wastes
becomes contaminated and once it flows out of the waste
material it is termed leachate. A further contribution to the
total leachate volume is produced during the decomposi-
tion of carbonaceous material producing a wide range of
other products including methane, carbon dioxide and a
complex mixture of organic acids, humic and fulvic acids,
aldehydes, alcohols and simple sugars. In actively
decomposing waste the temperature rises and the pH falls
rapidly and many metal ions which are relatively insoluble
at neutral pH can become dissolved in the developing
leachate. Leachate is corrosive and reacts with materials
that are not themselves prone to dissolve in tap water such
as fire ash, cement based building materials, iron and steel
wastes, aluminium wastes and gypsum based materials
changing the chemical composition of the leachate [38].
Therefore, one of the main concern regarding the landfill
leachate is the difficulty of treatment and disposal because
of the relatively large content of metals and especially
transition metals. In developed countries the municipal
wastes are no more mixed with industrial wastes in the
landfilling practice and therefore the heavy metals content
in the leachate is reduced in comparison to previous
practices when the mixing of different types of wastes
having different origin was allowed. Notwithstanding these
Multielement analysis 121
123
restrictions, the landfill leachate remains a waste which
poses serious risks of pollution of the environment and in
particular of the groundwater. Furthermore, the transition
metals content is variable in the leachate, depending from
the degree of ‘‘maturation’’ or aging of both the landfill site
and the resulting leachate [38]. Moreover, the concentra-
tion of the transition metals are also varying seasonally and
this is linked to the frequency of the rainfall which can vary
considerably from month to month. In order to be able to
know in advance the current content of transition metals in
a certain leachate, to decide what could be the most
appropriate treatment to which submit the leachate, but
also to monitor the effectiveness of the treatment in the
removal of the transition metals from leachate [39], it is
necessary to have available a simple, fast and effective
analytical technique which does not require any sample
preparation before the analysis without any need of using
reagents and with the possibility to analyze simultaneously
all elements of interest. The best analytical technique
which should fulfill all the mentioned wishes is the TXRF.
In the present work we have analyzed a sample of
landfill leachate coming from a landfill having about
6–7 years old. The purpose of the present study is to show
that the last generation of TXRF afford reliable analytical
results also on extremely complex matrices like a landfill
leachate. The validation of the results of our analysis with
TXRF was made using an ICP-OES classical analysis.
The complete chemical analysis results of the landfill
leachate sample
The leachate sample was taken from a landfill in September
2011. The chemical oxygen demand of the sample (COD)
was 5,550 mg/L, a value quite typical of a leachate coming
from a landfill site having several years old. The pH was
basic, 8.7 and the electrical conductivity of the leachate
was found at 10,880 lS/cm, indicating a quite high level of
electrolytes. The basic pH is attributable to a high ammonia
and ammonium ion concentration which can exceed easily
200 mg/L. The ionic chromatographic analysis of the
leachate sample (after filtration) revealed also a concen-
tration of 0.5 mg/L of formiate ion (HCOO-) and 70 mg/L
of acetate ion (CH3COO-). Other ions found with ionic
chromatography are: nitrites (NO2-) 0.9 mg/L, nitrates
(NO3-) 6.7 mg/L, phosphate (PO4
3-) 31.6 mg/L and sul-
phate (SO42-) 109.3 mg/L.
As described in the experimental section, the landfill
leachate sample was directly dried on the support for the
TXRF analysis and spiked with the internal standard (a Ga
solution) without any further treatment and without any
filtration and mineralization. This aspect is a really great
advantage of TXRF over the necessity to filtrate or, better
to mineralize the sample needed to be submitted for
analysis in the ICP-OES which does not accept the injec-
tion of the raw sample. Indeed, the typical procedure for
the ICP-OES sample preparation before injection into the
plasma torch, requires the mineralization. The leachate
sample is heated in a microwave oven in an autoclave in
presence of nitric and hydrochloric acid for 15 min. The
resulting mineralized leachate is then injected in the ICP-
OES torch. Our feeling is that the TXRF offers much less
alteration of the sample, which can be analyzed simply
after drying without the aggressive degradation led by
nitric and hydrochloric acid which in their turn may add
some undesired transition metal ion contamination. The
analytical results of the ICP-OES in comparison with the
TXRF results are shown in Table 1. For the TXRF results
Table 1 Landfill leachate elemental composition ICP-OES versus
TXRF
ICP-OES (mg/L) TXRF (mg/L) TXRF LLD (mg/L)
B 4.94
F 2.9
Na 300 1104 32.8
Mg 45 27.5 7.4
Al 0.348
Si 35
P 21 7.5 0.587
S 70 49.3 0.379
Cl 2519 1862 0.223
K 450 762 0.066
Ca 154 79.9 0.038
Ti 0.861 0.715 0.018
V 0.069 0.049 0.014
Cr 0.351 0.426 0.011
Mn 0.75 0.944 0.0089
Fe 5.9 7.1 0.0073
Co 0.074
Ni 0.179 0.21 0.0038
Cu 0.03 0.028 0.0033
Zn 0.306 0.538 0.0029
As 0.223 0.232 0.0023
Se 0.0009
Br 8.4 10.1 0.0021
Rb 8.6 7.1 0.0026
Sr 4.5 5.7 0.0019
Cd 0.0008
Mo 0.012
Sn 0.271
Sb 0.258
Ba 3.85 2.7 0.029
Hg 0.001 0.0026 \0.003
Pb 0.045 0.008 \0.003
LLD lowest detectable limit by TXRF
122 F. Cataldo
123
in the last column in the right are reported also the LLD
values in mg/L. The data reported on Table 1 were
obtained from the fluorescence spectra shown in Figs. 3, 4.
Figure 3 displays the detail of the TXRF spectrum taken in
the lower energy range while Fig. 4 displays the spectrum
taken in the higher energy range.
Table 1 shows that the elemental concentration in the
landfill leachate measured with the TXRF are in reasonable
agreement with the ICP-OES results. Such agreement can
be better appreciated at first glance in the log/log graph
shown in Fig. 5 where a straight line can be obtained by
putting in ordinate the analytical results of the ICP-OES
and in abscissa the analytical results of the TXRF. It is
remarkable that also sodium and magnesium were detected
by the TXRF in reasonable agreement with the ICP-OES
results although, as shown in Fig. 2, these two elements are
rather difficult to detect with TXRF. The agreement
between the ICP-OES results and the TXRF results is
simply impressive for the transition metals like for instance
Ti, V, Cr, Mn, Fe, Ni, Cu, Zn and As but also for Br, Rb,
Sr, Ba. Instead the TXRF seem to give some underesti-
mation of the following light elements P, S, Cl and Ca in
the landfill leachate matrix if we take as reference the
values measured with the ICP-OES.
Analysis of three groundwater samples
Three groundwater samples were taken in three different
weeks in September 2011 and tested also with TXRF
against the ICP-OES. The results are reported in Table 2
and the spectra of these samples are again reported in
Figs. 3, 4. Of course, in the case of groundwater samples
the concentrations of the transition metals are considerably
lower than those found in the leachate. At these concen-
trations the agreement between the TXRF and the ICP-
OES is really impressive for all elements (see Table 2).
Fig. 3 TXRF (low energy range from 0.8 to 4.5 keV). The orange intense
curve is due to the landfill leachate sample while the three curves at the
bottom of the figure with lower intensity are due from three samples
(respectively blue green and red) of groundwater used as reference. Note
the high content of chlorine, potassium and calcium in the leachate sample.
(Abscissa in keV and ordinate in 91000 Counts). (Color figure online)
Multielement analysis 123
123
Fig. 4 TXRF (high energy range from about 5 to 15 keV). The
orange intense curve is due to the landfill leachate sample while the
three curves at the bottom of the figure with lower intensity are due
from three samples (respectively blue green and red) of groundwater
used as reference. (Abscissa in keV and ordinate in 91000 Counts).
(Color figure online)
Fig. 5 Log/log plot of the
elemental concentration
measured in a landfill leachate
sample with ICP-OES and with
TXRF
124 F. Cataldo
123
First of all it must be underlined that even the light ele-
ments show a good agreement between the two methods
while for the heavier elements the TXRF is giving a con-
centration which instead in most cases is not quantified by
the ICP-OES method being beyond the LLD of the latter
analytical methodology. It must be emphasized that in the
analysis of not polluted groundwater also the sample to be
used in the ICP-OES should not be mineralized and
probably it is this similar treatment of the samples for the
two analytical techniques that permit to avoid any alter-
ation of the samples and a better analytical agreement as
shown in Table 2.
Conclusions
A landfill leachate coming from a landfill site of several
years old was fully analyzed for all elements with two
Table 2 Groundwater elemental composition ICP-OES versus TXRF
Sample #1 Sample #1 Sample #1 Sample #2 Sample #2 Sample #2 Sample #3 Sample #3 Sample #3
ICP-OES
(mg/L)
TXRF
(mg/L)
TXRF LLD
(mg/L)
ICP-OES
(mg/L)
TXRF
(mg/L)
TXRF LLD
(mg/L)
ICP-OES
(mg/L)
TXRF
(mg/L)
TXRF LLD
(mg/L)
B 0.01 0.01 \0.01
F 0.36 0.36 0.19
Na 11.3 18.4 8.6 11.3 25.5 9.36 12.5 22.1 13.7
Mg 13.6 10.9 1.8 13.5 10.7 2.01 24.8 8.67 3.05
Al \0.001 \0.001 \0.001
P 0.166 0.152 0.125 0.183 0.319 0.144 0.036 1.00 0.22
S 1.85 1.67 0.074 2.60 1.503 0.082 1.35 2.88 0.133
Cl 8.69 13.4 0.043 9.9 10.7 0.05 8.04 12.36 0.0790
K 7.65 12.1 0.0016 7.95 9.41 0.025 10.0 10.45 0.0040
Ca 27.2 29.4 0.0092 50.9 62.14 0.01 187 168.56 0.0140
Ti 0.01 0.0044 \LLD 0.0046 \LLD 0.0062
V 0.003 \LLD 0.0037 0.005 0.0063 0.0038 \0.001 0.0086 0.0051
Cr \0.001 \LLD 0.0029 \0.001 \LLD 0.003 \0.001 \LLD 0.0039
Mn 0.003 0.0081 0.0022 0.045 0.059 0.0023 0.077 0.015 0.0030
Fe 0.01 0.024 0.0017 0.014 0.0044 0.0018 0.015 0.016 0.0024
Co \0.001 \0.001 \0.001
Ni \0.001 0.0023 0.0012 \0.001 0.0017 0.0013 \0.001 0.0069 0.0016
Cu \0.001 0.04 0.0011 \0.001 0.0033 0.0011 \0.001 0.0058 0.0014
Zn \0.001 0.209 0.0009 \0.011 0.534 0.00098 0.019 0.346 0.0012
As \0.001 0.83 0.0006 \0.001 0.0012 0.00065 \0.001 0.0013 0.0009
Se \0.001 \0.001 \0.001
Br 0.04 0.04 0.0006 0.035 0.00058 0.112 0.0008
Rb 0.112 0.0007 0.102 0.00071 0.112 0.0009
Sr 0.44 0.677 0.0005 0.48 0.684 0.00053 0.69 1.013 0.0007
Mo \0.001 \0.001 \0.001
Cd \0.001 \0.001 \0.001
Sn \0.001 \0.001 \0.001
Sb \0.001 \0.001 \0.001
Ba 0.12 0.129 0.00066 0.12 0.133 0.0007 0.12 0.14 0.0094
Hg \0.001 \LLD 0.00073 \0.001 \LLD 0.00076 \0.001 \LLD 0.0010
Pb \0.001 0.003 0.00076 \0.001 0.0026 0.00079 \0.001 \LLD 0.0010
NO2 0.37a 0.19a 0.14a
NO3 15.2a 14.8a 12.7a
PO4 0.51a 0.47b 0.56a 0.94b 0.11a 3.06b
SO4 5.55a 4.98b 7.8a 4.5b 4.06a 8.64b
a Measured with ionic chromatographyb Calculated from the concentration of P or S
Multielement analysis 125
123
analytical techniques based respectively in TXRF and on
ICP-OES. A reasonable agreement between the analytical
results obtained with the two techniques was found for
light elements and a good agreement for the higher atomic
weight elements. The great advantage of TXRF is in the
sample preparation. Practically there is no sample prepa-
ration, only drying on a support the raw landfill leachate
after the addition of an internal standard. On the other hand
the ICP-OES techniques requires the acid digestion and
thermal mineralization of the landfill leachate before the
injection in the plasma torch. This operation is certainly
time consuming and if not properly conducted may lead to
alterations and contaminations of the original sample. Such
an event is rather improbable with the TXRF technique
where the sample is used raw. Additional advantages of the
TXRF technique is the compactness of the bechtop spec-
trometer which in principle is even portable and the very
low energy consumption. Instead, the ICP-OES analytical
technique requires a rather large equipment with large
consumption of energy and also an important consumption
of Ar gas which instead is not required at all for the TXRF
spectrometer.
The goal of this work was to show that the quick and
everyday monitoring of a landfill leachate can be achieved
simply, economically and with reliability in the analytical
data with a TXRF spectrometer. Such a monitoring is
needed either to check the analytes in the leachate which
are changing continuously due to the type of wastes put in
the landfill and mainly due to the changes in the atmo-
spheric conditions from rainy seasons to dry seasons.
Furthermore, after the leachate treatment [39], it is needed
to check again the analytical composition of the purified
water and again the TXRF provides a very quick and
reliable response to this requirement.
Acknowledgments The author is indebted with Dr. Hagen Stosnach
from Bruker for the chemical analysis made at the S2 Picofox TXRF
spectrometer and also with Mr. Cristian Vailati also from Bruker for
the helpful discussion of the results.
References
1. Klockenkamper R, Von Bohlen A (1996) X-ray Spectrom 25:156
2. Klockenkamper R (1997) Total reflection X-ray analysis. Wiley,
New York
3. Wobrauschek P (1998) J Anal Atom Spectrom 13:333
4. Klockenkamper R, Von Bohlen A (2001) Spectrochim Acta
Atomic Spectrosc 56:2005
5. Hoefler H, Streli C, Wobrauschek P, Ovari M, Zaray G (2006)
Spectrochim Acta Atomic Spectrosc 61:1135
6. Streli C (2006) Appl Spectrosc Rev 41:473
7. Stosnach H (2005) Powder Diffr 20:141
8. Stosnach H (2005) Anal Sci 21:873
9. Weiss C, Knoth J, Schwenke H, Geisler H, Lerche J, Schulz R,
Ullrich HJ (2000) Microchim Acta 133:65
10. Tavares GA, Almeida E, de Oliveira JGG, Bendassolli JA, Na-
scimento Filho VF (2011) J Radioanal Nucl Chem 287:377
11. Kinimura S, Kaway J (2007) Anal Sci 23:1185
12. Montero Alvarez A, Estevez Alvarez JR, Padilla Alvarez R
(2007) J Radioanal Nucl Chem 273:427
13. Landsberger S, Kaminsky M, Basunia M, Iskander FY (2000) J
Radioanal Nucl Chem 244:35
14. Bennun L, Sanhueza V (2010) Anal Sci 26:331
15. Yamaguchi H, Itoh S, Igarashi S, Naitoh K, Hasegawa R (1998)
Anal Sci 14:909
16. Dargie M, Markowicz A, Tajani A, Valkovic V (1997) Fresenius
J Anal Chem 357:589
17. Ruiz RF (2008) Anal Chem 80:8372
18. Wobrauschek P, Streli C, Kregsamer P, Meirer F, Jokubonis C,
Markowicz A, Wegrzynek D, Chinea-Cano E (2008) Spectro-
chim Acta 63:1404
19. Tian YH, Liu K, Wu XR, Wang RG (1997) J Radioanal Nucl
Chem 217:243
20. Padilla Alvarez R, Chinea Cano E, Estevez Alvarez JR, Greaves
ED (1999) J Radioanal Nucl Chem 240:517
21. Amartaivan T, Greaves ED, Bernasconi G, Wobrauschek P
(1997) J Radioanal Nucl Chem 220:249
22. Vilhunen JK, von Bohlen A, Schmeling M, Rantanen L, Mi-
kkonen S, Klockenkamper R, Klockow D (1999) Microchim Acta
131:219
23. Cunha e Silva RM, Almeida E, Valencia EPE, Nascimento Filho
VF (2004) J Radioanal Nucl Chem 260:3
24. Dogan M, Soylak M, Elci L, von Bohlen A (2002) Microchim
Acta 138:77
25. Hoffmann P, Martin Hein TM, Scheuer V, Lieser KH (1990)
Mikrochim Acta 11:305
26. Nishiwaki Y, Shimoyama M, Nakanishi T, Ninomiya T, Nakai I
(2006) Anal Sci 22:1297
27. Kadenkin A, Amberger M, Fittschen UEA, Broekaert JAC (2011)
Microchim Acta 173:183
28. Fernandez-Ruiz R, Garcia-Heras M (2008) Spectrochim Acta
Atom Spectrosc 63:975
29. Wehling B, Vandenabeele P, Moens L, Klockenkamper R, von
Bohlen A, Van Hooydonk G, de Reu M (1999) Microchim Acta
130:253
30. Koulouridakis PE, Domazos EA, Galani-Nikolakaki SM, Kalli-
thrakas-Kontos NG (2004) Microchim Acta 146:245
31. Martinez T, Lartigue J, Avila-Perez P, Zarazua G, Navarrete M,
Tejeda S, Ramırez A (2004) J Radioanal Nucl Chem 259:511
32. Khuder A, Bakir MA, Karjou J, Sawan MKh (2007) J Radioanal
Nucl Chem 273:435
33. Magalhaes T, von Bohlen A, Carvalho ML, Becker M (2006)
Spectrochim Acta Part B Atom Spectrosc 61:1185
34. Griesel S, Mundry R, Kakuschke A, Fonfara S, Siebert U, Prange
A (2006) Spectrochim Acta Part B Atom Spectrosc 61:1158
35. Nimis PL, Lazzarin G, Lazzarin A, Skert N (2000) Sci Total
Environ 255:97
36. Espinoza-Quinones FR, Modenes AN, Palacio SM, Lorenz EK,
Oliveira AP (2011) Water Sci Technol 63:1506
37. Marguı E, Tapias JC, Casas A, Hidalgo M, Queralt I (2010)
Chemosphere 80:263
38. Baun DL, Christensen TH (2004) Waste Manag Res 22:3
39. Cataldo F, Angelini G (2012) Ozone Sci Eng. Submitted for
publication
126 F. Cataldo
123