Neo NaviGuide P E R F E C T I M P L A N T P L A N N I N G F O R B R I G H T S M I L E Neo NaviGuide Kit Drilling Protocol April 24, 2017
Transcript
Page 1: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Neo NaviGuide

P E R F E C T I M P L A N T P L A N N I N G F O R B R I G H T S M I L E

Neo NaviGuide Kit Drilling Protocol April 24, 2017

Page 2: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Fixture Stop

Mount

Ø3.0 Drill

+

Ø2.2 Point Drill

Initial Drill Ø2.2 Drill

800~1,200rpm / 30~45Ncm 50~100rpm / 30~45Ncm

(Option : 800~1,200rpm / 30~45Ncm)

30~50rpm / 30~45Ncm

Fixture Non Stop Mount

Drilling Speed & Torque

Cortical Tap IS-III Ø3.5

Cortical Tap IS-II Ø3.5

Fixture Driver Contra Angle

Fixture Driver Rachet

Precautions

1. Must use drill with cold saline.

2. Must use Tap Drill in the case of D1~2 bone. 3. In the case that Sleeve Offset is applied, use the Fixture Nonstop mount when placing implant. 4. Must check the insertion torque after removing the guide. In the case of low insertion torque value, insert the implant slightly more. On the other hand, if the torque value is over 40Ncm, counterturn about 45˚~90˚, then double-check the torque value. 5. Must use Bone Profile after removing the guide. 6. Recommended use of Fixture Driver is to use after removing the guide for measuring the torque value or for further inserting the implant. of Surgical Guide. As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 3: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Cortical Tap

Cortical Drill

Fixture Stop

Mount

Bone Profile

Ø3.0 Drill

Ø3.5 Drill

Ø4.0 Drill

Ø4.5 Drill

Ø4.3 Tissue Punch

Ø2.2 Point Drill

Bone Trimmer

Initial Drill Ø2.2 Drill

800~1,200rpm / 30~45Ncm 50~100rpm / 30~45Ncm

(Option : 800~1,200rpm / 30~45Ncm)

30~50rpm / 30~45Ncm 50rpm

Drilling Speed & Torque

Fixture NonStop Mount

R/W

Fixture Driver Contra Angle

Fixture Driver Rachet

Precautions

1. Must use drill with cold saline.

2. Must use Tap Drill in the case of D1~2 bone. 3. In the case that Sleeve Offset is applied, use the Fixture Nonstop mount when placing implant. 4. Must check the insertion torque after removing the guide. In the case of low insertion torque value, insert the implant slightly more. On the other hand, if the torque value is over 40Ncm, counterturn about 45˚~90˚, then double-check the torque value. 5. Must use Bone Profile after removing the guide. 6. Recommended use of Fixture Driver is to use after removing the guide for measuring the torque value or for further inserting the implant. of Surgical Guide. As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 4: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

1.2mm (Tip Length)

8.5mm (Fixture Length)

9mm (Sleeve Offset)

#

Ø3.5 x 8.5mm / Offset 0mm

= 18.7mm (Drill Length)

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

Ø3.0 8.5mm Drill

Ø3.5 Cortical Tap (Option* ) 1

Fixture Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 5: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

1.2mm (Tip Length)

8.5mm (Fixture Length)

= 20.2mm (Drill Length)

10.5mm (Sleeve Offset)

#

Ø3.5 x 8.5mm / Offset +1.5mm

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 10mm Drill

Ø3.0 10mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 6: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 21.7mm (Drill Length)

#

Ø3.5 x 8.5mm / Offset +3.0mm

1.2mm (Tip Length)

8.5mm (Fixture Length)

12mm (Sleeve Offset)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 11.5mm Drill

Ø3.0 11.5mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 7: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

1.2mm (Tip Length)

10mm (Fixture Length)

= 20.2mm (Drill Length)

9mm (Sleeve Offset)

#

Ø3.5 x 10mm / Offset 0mm

Ø3.5 Cortical Tap (Option* ) 1

Fixture Stop Mount

Ø2.2 10mm Drill

Ø3.0 10mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 8: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 21.7mm (Drill Length)

#

Ø3.5 x 10mm / Offset +1.5mm

10.5mm (Sleeve Offset)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 11.5mm Drill

Ø3.0 11.5mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

1.2mm (Tip Length)

10mm (Fixture Length)

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 9: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 23.2mm (Drill Length)

#

Ø3.5 x 10mm / Offset +3.0mm

12mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 10: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 21.7mm (Drill Length)

#

Ø3.5 x 11.5mm / Offset 0mm

9mm (Sleeve Offset)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Stop Mount

Ø2.2 11.5mm Drill

Ø3.0 11.5mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

1.2mm (Tip Length)

11.5mm (Fixture Length)

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 11: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 23.2mm (Drill Length)

#

Ø3.5 x 11.5mm / Offset +1.5mm

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

11.5mm (Fixture Length)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 12: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 24.7mm (Drill Length)

#

Ø3.5 x 11.5mm / Offset +3.0mm

12mm (Sleeve Offset)

1.2mm (Tip Length)

11.5mm (Fixture Length)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 14.5mm Drill

Ø3.0 14.5mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 13: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 23.2mm (Drill Length)

#

Ø3.5 x 13mm / Offset 0mm

9mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Stop Mount

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 14: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 24.7mm (Drill Length)

#

Ø3.5 x 13mm / Offset +1.5mm

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 14.5mm Drill

Ø3.0 14.5mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 15: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

= 23.2mm (Drill Length)

#

Ø3.5 x 13mm / Offset 0mm

9mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Ø3.5 Cortical Tap (Option* ) 1

Fixture Stop Mount

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 Point Drill

Ø2.2 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone Use IS-II Tap for inserting IS-II Fixture Use IS-III Tap for inserting IS-III Fixture

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 16: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 8.5mm / Offset 0mm

= 18.7mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Fixture Stop Mount

Mount Remover (Option* ) 1

Ø4.3 Tissue Punch (Option)

Ø2.2 8.5mm Drill

Ø3.0 8.5mm Drill

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

Page 17: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 8.5mm / Offset +1.5mm

= 20.2mm (Drill Length)

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Bone Trimmer

1.5mm Offset (Option)

Fixture Non Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 10mm Drill

Ø3.0 10mm Drill

Ø2.2 7.3mm Drill

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 18: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 8.5mm / Offset +3.0mm

= 21.7mm (Drill Length)

12mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Bone Trimmer

3mm Offset (Option)

Fixture Non Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 11.5mm

Drill

Ø3.0 11.5mm

Drill

Ø2.2 7.3mm Drill

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 19: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 10mm / Offset 0mm

= 20.2mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Fixture Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 10mm Drill

Ø3.0 10mm Drill

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 20: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 10mm / Offset +1.5mm

= 21.7mm (Drill Length)

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Fixture Non Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 11.5mm

Drill

Ø3.0 11.5mm

Drill

Ø2.2 7.3mm Drill

Bone Trimmer

1.5mm Offset (Option)

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 21: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 10mm / Offset +3.0mm

= 23.2mm (Drill Length)

12mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Fixture Non Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 7.3mm Drill

Bone Trimmer

3mm Offset (Option)

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 22: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 11.5mm / Offset 0mm

= 20.2mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

11.5mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Fixture Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 11.5mm

Drill

Ø3.0 11.5mm

Drill

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 23: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 11.5mm / Offset +1.5mm

10.5mm (Sleeve Offset)

11.5mm (Fixture Length)

Mount Remover (Option* ) 1

= 22.2mm (Drill Length)

1.2mm (Tip Length)

Ø2.2 Point Drill

Fixture Non Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 7.3mm Drill

Bone Trimmer

1.5mm Offset (Option)

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 24: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø2.2 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø3.5 x 13mm / Offset 0mm

= 23.2mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Mount Remover (Option* ) 1

Ø2.2 Point Drill

Fixture Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 13mm Drill

Ø3.0 13mm Drill

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

*1 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. * Use of Tissue Punch and Bone Trimmer is optional. This may result in excessive removal of soft or hard tissue.

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 25: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 7.3mm / Offset 0mm

Bone Profile

(Option* ) 3

= 17.5mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

7.3mm (Fixture Length)

Ø2.2 Point Drill

Fixture Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

Ø3.0 7.3mm Drill

Ø3.5 7.3mm Drill

Ø4.0 Cortical Tap (Option* ) 1

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 26: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 7.3mm / Offset +1.5mm

Bone Profile

(Option* ) 3

= 18.7mm (Drill Length)

10.5mm (Sleeve Offset)

0.9mm (Tip Length)

7.3mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 8.5mm Drill

Ø3.5 8.5mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 8.5mm Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 27: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 7.3mm / Offset +3.0mm

Bone Profile

(Option* ) 3

= 20.2mm (Drill Length)

12mm (Sleeve Offset)

0.9mm (Tip Length)

7.3mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 10mm Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 28: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 8.5mm / Offset 0mm

Bone Profile

(Option* ) 3

= 18.7mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 8.5mm Drill

Ø3.5 8.5mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Ø2.2 8.5mm Drill

Bone Trimmer Standard (Option)

Fixture Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 29: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 8.5mm / Offset +1.5mm

Bone Profile

(Option* ) 3

= 20.2mm (Drill Length)

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 10mm Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 30: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 8.5mm / Offset +3.0mm

Bone Profile

(Option* ) 3

= 21.7mm (Drill Length)

12mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 11.5mm

Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 31: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 10mm / Offset 0mm

Bone Profile

(Option* ) 3

= 18.7mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Ø2.2 10mm Drill

Bone Trimmer Standard (Option)

Fixture Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 32: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 10mm / Offset +1.5mm

Bone Profile

(Option* ) 3

10.5mm (Sleeve Offset)

= 21.7mm (Drill Length) 1.2mm (Tip Length)

10mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 11.5mm

Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 33: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 10mm / Offset +3.0mm

Bone Profile

(Option* ) 3

12mm (Sleeve Offset)

= 23.2mm (Drill Length) 1.2mm (Tip Length)

10mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 13mm Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 34: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 11.5mm / Offset 0mm

Bone Profile

(Option* ) 3

= 21.7mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

11.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø4.0 Cortical Tap (Option* ) 1

Ø2.2 11.5mm

Drill

Bone Trimmer Standard (Option)

Fixture Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 35: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 11.5mm / Offset +1.5mm

Bone Profile

(Option* ) 3

10.5mm (Sleeve Offset)

= 23.2mm (Drill Length) 1.2mm (Tip Length)

11.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 13mm Drill

Fixture Non Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 36: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

Mount Remover (Option* ) 2

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.0 x 13mm / Offset 0mm

Bone Profile

(Option* ) 3

= 23.2mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø4.0 Cortical Tap (Option* ) 1

Ø2.2 13mm Drill

Bone Trimmer Standard (Option)

Fixture Stop Mount

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 37: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

2

R/W

#

Ø4.5 x 7.3mm / Offset 0mm

Mount Remover (Option* ) 2

= 17.5mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

7.3mm (Fixture Length)

Bone Profile

(Option* ) 3

Ø2.2 Point Drill

Fixture Stop Mount

Ø4.3 Tissue Punch (Option)

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

Ø3.0 7.3mm Drill

Ø3.5 7.3mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Ø4.0 7.3mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 38: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 7.3mm / Offset +1.5mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 18.7mm (Drill Length)

10.5mm (Sleeve Offset)

0.9mm (Tip Length)

7.3mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 8.5mm Drill

Ø3.5 8.5mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 8.5mm Drill

Fixture Non Stop Mount

Ø4.0 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 39: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 7.3mm / Offset +3.0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 20.2mm (Drill Length)

12mm (Sleeve Offset)

0.9mm (Tip Length)

7.3mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 10mm Drill

Fixture Non Stop Mount

Ø4.0 10mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 40: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 8.5mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 18.7mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 8.5mm Drill

Ø3.5 8.5mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Ø2.2 8.5mm Drill

Bone Trimmer Standard (Option)

Fixture Stop Mount

Ø4.0 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 41: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 8.5mm / Offset +1.5mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 20.2mm (Drill Length)

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 10mm Drill

Fixture Non Stop Mount

Ø4.0 10mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 42: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 8.5mm / Offset +3.0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 21.7mm (Drill Length)

12mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 11.5mm

Drill

Fixture Non Stop Mount

Ø4.0 11.5mm

Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 43: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 10mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

9mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Fixture Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Ø2.2 10mm Drill

Bone Trimmer Standard (Option)

Ø4.0 10mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 44: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 10mm / Offset +1.5mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

10.5mm (Sleeve Offset)

= 21.7mm (Drill Length) 1.2mm (Tip Length)

10mm (Fixture Length)

Fixture Non Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 11.5mm

Drill

Ø4.0 11.5mm

Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 45: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 10mm / Offset +3.0mm

Mount Remover (Option* ) 2

12mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Fixture Non Stop Mount

Bone Profile

(Option* ) 3

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 13mm Drill

Ø4.0 13mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 46: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 11.5mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

9mm (Sleeve Offset)

1.2mm (Tip Length)

11.5mm (Fixture Length)

Fixture Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø4.5 Cortical Tap (Option* ) 1

Ø2.2 11.5mm

Drill

Bone Trimmer Standard (Option)

Ø4.0 11.5mm

Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 47: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 11.5mm / Offset +1.5mm

Mount Remover (Option* ) 2

10.5mm (Sleeve Offset)

= 23.2mm (Drill Length) 1.2mm (Tip Length)

11.5mm (Fixture Length)

Fixture Non Stop Mount

Bone Profile

(Option* ) 3

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 13mm Drill

Ø4.0 13mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 48: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø3.5 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø4.5 x 13mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

9mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Fixture Stop Mount

= 23.2mm (Drill Length)

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø4.5 Cortical Tap (Option* ) 1

Ø2.2 13mm Drill

Bone Trimmer Standard (Option)

Ø4.0 13mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 49: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 7.3mm / Offset 0mm

2

Mount Remover (Option* ) 2

= 17.5mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

7.3mm (Fixture Length)

Bone Profile

(Option* ) 3

Fixture Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch (Option)

Ø2.2 7.3mm Drill

Bone Trimmer Standard (Option)

Ø3.0 7.3mm Drill

Ø3.5 7.3mm Drill

Ø4.0 7.3mm Drill

Ø5.0 Cortical Tap (Option* ) 1

Ø4.5 7.3mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 50: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 7.3mm / Offset +1.5mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 18.7mm (Drill Length)

10.5mm (Sleeve Offset)

0.9mm (Tip Length)

7.3mm (Fixture Length)

Ø5.0 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 8.5mm Drill

Ø3.5 8.5mm Drill

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 8.5mm Drill

Ø4.0 8.5mm Drill

Ø4.5 8.5mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 51: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 7.3mm / Offset +3.0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 20.2mm (Drill Length)

12mm (Sleeve Offset)

0.9mm (Tip Length)

7.3mm (Fixture Length)

Ø5.0 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 10mm Drill

Ø4.0 10mm Drill

Ø4.5 10mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 52: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 8.5mm / Offset 0mm

Mount Remover (Option* ) 2

= 18.7mm (Drill Length)

9mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø5.0 Cortical Tap (Option* ) 1

Fixture Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 8.5mm Drill

Ø3.5 8.5mm Drill

Ø2.2 8.5mm Drill

Bone Trimmer Standard (Option)

Ø4.0 8.5mm Drill

Ø4.5 8.5mm Drill

Bone Profile

(Option* ) 3

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 53: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 8.5mm / Offset +1.5mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 20.2mm (Drill Length)

10.5mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø5.0 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 10mm Drill

Ø4.0 10mm Drill

Ø4.5 10mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 54: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 8.5mm / Offset +3.0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

= 21.7mm (Drill Length)

12mm (Sleeve Offset)

1.2mm (Tip Length)

8.5mm (Fixture Length)

Ø5.0 Cortical Tap (Option* ) 1

Fixture Non Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 11.5mm

Drill

Ø4.0 11.5mm

Drill

Ø4.5 11.5mm

Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 55: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 10mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

9mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Fixture Stop Mount

Ø5.0 Cortical Tap (Option* ) 1

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 10mm Drill

Ø3.5 10mm Drill

Ø2.2 10mm Drill

Bone Trimmer Standard (Option)

Ø4.0 10mm Drill

Ø4.5 10mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 56: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

R/W

#

Ø5.0 x 10mm / Offset +1.5mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

10.5mm (Sleeve Offset)

= 21.7mm (Drill Length) 1.2mm (Tip Length)

10mm (Fixture Length)

Fixture Non Stop Mount

Ø5.0 Cortical Tap (Option* ) 1

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 11.5mm

Drill

Ø4.0 11.5mm

Drill

Ø4.5 11.5mm

Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

Page 57: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 10mm / Offset +3.0mm

Mount Remover (Option* ) 2

12mm (Sleeve Offset)

1.2mm (Tip Length)

10mm (Fixture Length)

Fixture Non Stop Mount

Bone Profile

(Option* ) 3

Ø5.0 Cortical Tap (Option* ) 1

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Bone Trimmer

3.0mm Offset (Option)

Ø2.2 13mm Drill

Ø4.0 13mm Drill

Ø4.5 13mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 58: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 11.5mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

9mm (Sleeve Offset)

1.2mm (Tip Length)

11.5mm (Fixture Length)

Fixture Stop Mount

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 11.5mm

Drill

Ø3.5 11.5mm

Drill

Ø2.2 11.5mm

Drill

Bone Trimmer Standard (Option)

Ø4.0 11.5mm

Drill

Ø4.5 11.5mm

Drill

Ø5.0 Cortical Tap (Option* ) 1

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 59: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 11.5mm / Offset +1.5mm

Mount Remover (Option* ) 2

10.5mm (Sleeve Offset)

= 23.2mm (Drill Length) 1.2mm (Tip Length)

11.5mm (Fixture Length)

Bone Profile

(Option* ) 3

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Bone Trimmer

1.5mm Offset (Option)

Ø2.2 13mm Drill

Ø4.0 13mm Drill

Ø4.5 13mm Drill

Fixture Non Stop Mount

Ø5.0 Cortical Tap (Option* ) 1

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 60: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

※ In soft(D4) bone or in condition of getting initial fixation at implant apex, Ø4.0 drill is the final drill ※ The fixture bigger than ∅4.5 diameter is recommended to be placed in posterior area

R/W

#

Ø5.0 x 13mm / Offset 0mm

Mount Remover (Option* ) 2

Bone Profile

(Option* ) 3

9mm (Sleeve Offset)

1.2mm (Tip Length)

13mm (Fixture Length)

Fixture Stop Mount

= 23.2mm (Drill Length)

Ø5.0 Cortical Tap (Option* ) 1

Ø2.2 Point Drill

Ø4.3 Tissue Punch

Ø2.2 7.3mm Drill

Ø3.0 13mm Drill

Ø3.5 13mm Drill

Ø2.2 13mm Drill

Bone Trimmer Standard (Option)

Ø4.0 13mm Drill

Ø4.5 13mm Drill

*1 Use Tap Drill in the case of D1 ~ D2 bone *2 Use Mount Remover for removing the Fixture Mount from placed implant Remove the Fixture Mount from the implant by first removing the screw of the Mount, and then connecting the Mount Remover and rotating it clockwise. *3 Use Bone Profile after removing the guide

* As the Sleeve Offset Marking is only applied for the standard 9mm, please be cautious when the Sleeve Offset is applied.

Page 61: Neo NaviGuide - NeoBiotechnas.neobiotech.co.kr/170430/PN2_NeoNaviGuide_Drilling...Cortical Tap Cortical DrillØ2.2 Fixture Stop Mount Bone Profile Ø3.0Tissue Drill Ø3.5 DrillPunch

GUIDE NeoNaviGuide

PM / GUIDE PM TEAM

neobiotech.co.kr │ +82-2-582-2885


Recommended