New Developments in New Developments in Proton Treatment Planning Systems Daniel Yeung
Transcript
Page 1: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

New Developments in New Developments in Proton Treatment Planning Systems

Daniel Yeung

Page 2: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Statement of Disclosure

Funded Research & Development:

Philips Medical Systems

IBA

Page 3: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Proton Planning Systemsg y

C i l S t A d i S tCommercial Systems Academic SystemsAnalytical/Semi-Analytical CMS - Xio MGH, HCL

Varian – Eclipse Proton PSIVarian – EyePlan ClatterbridgeOptivus - Odyssey Loma LindaDosigray OrsayDosigray Orsay

Others… Others…* Under Development Philips – Pinnacle* UF*Monte Carlo CMS*, Varian*,… MGH, DKFZ,…

Page 4: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Proton Pencil Beam Algorithmsg

Hong et al, MGH 1996

Szymanowski et al, Institut Curie 2001

Beam Model dosimetry in water inhomogeneity correction g y

• water equivalent thickness (wet) • HU stopping power

Page 5: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Proton Pencil Beam Algorithmg

Beam Model

CAX Depth Dose (DD) broad beam pristine peak sobp

Radial Spread (RS) multi coloumb scatter

• beamline: degrader, nozzle elements (wet only)• compensator: material dependent scattering powerp p g p• patient: wet only

gaussian kernelσ2

total = σ2line + σ2

comp + σ2patientσ total σ line σ comp σ patient

Page 6: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Proton Pencil Beam Algorithmg

Dose CalculationDose Calculation

Dose from a pencil beam

Convolve DD with RS for each pencil Sum dose from all pencils

Page 7: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Bortfeld Model of Pristine Peak

Analytical model of proton BP (up to ~ 200 MeV)• accounts for energy spread • empirical model of nuclear fragmentation (data fitting)empirical model of nuclear fragmentation (data fitting)• numeric depth dose calculation of fitted BP• assumption – range straggling ‘constant’ with depth

Page 8: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Pristine Peak - Analytical Model y

R=9 0 g/cm2R=5 86 g/cm2

R=21.86 g/cm2

R 9.0 g/cmR 5.86 g/cmV 8.0

V 8.1R=19.8 g/cm2

For R > 15 g/cm2,

Range straggling depth

E i B tf ld d l

Eclipse Model Bortfeld Model

Error in Bortfeld model

Page 9: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Pristine Peak – Bortfeld Model

R=19.8 g/cm2

R=28.4 g/cm2

Bortfeld Model Bortfeld - Enhanced

Page 10: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

SOBP Calc vs MeasurementCalc vs Measurement

Courtesy R. Slopsema

Page 11: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Beam Profile Calc vs MeasurementCalc vs Measurement

R=15.1 M=10.4

d=5 mm

d=10 cm

Option B5 - R=15.1 g/cm^2, M=10.4 g/cm^2, Air gap = 11.7cm, SSD=220.1, aperture: 15cmx15cm

1

1.11.2

1.3

umbr

a [c

m]

Eclipse convolved - @9.9 cm Measurement (av) - @9.9 cmEclipse convolved - @0.5 cm Measurement (av) - @0.5 cmEclipse convolved - @14.1 cm Measurement (av) - @14.1 cm

0.5

0.6

0.70.80.9

180

% L

ater

al P

enu

d=15 cm 0.3

0.4

0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00Air Gap [cm]

20%

-8

Courtesy R. Slopsema

Page 12: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Proton Dose Calc in water is generally accurate!

Page 13: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

…ok, what about Clinical Issues?

Physics Clinical

Page 14: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Match & Patch Fields

used to avoid OARs adjacent to target used to avoid OARs adjacent to target partition target into segments (sub-targets)

• sub-targets treated with ‘sub-beams’ • angle sub-beams to avoid OARs

combined with other fields for dose uniformity

Page 15: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Match Fields match fields abutting each other penumbra matching penumbra penumbra matching penumbra

Inf - LAT Sup - RAO

matchline

Page 16: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Patch Fields

th b t t ti l t t thru beam txt partial target residual txt with patch lateral penumbra (t beam) ‘matched’ with distal lateral penumbra (t-beam) matched with distal

falloff (p-beam) LPO beam (inferior) patched with SPO (superior) LPO beam (inferior) patched with SPO (superior)

Page 17: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Lacrimal Gland CarcinomaPTV 50 4PTV 50.4

partition into sup + inf targetsb i t

sup

inf

brainstem

inf

PTV50 4 5 fi ld ith t h & t h

LPO-Inf (spare optics & BS)SPO-patch

PTV50.4: 5 fields with match & patch Target FieldsPTV LAO LPO50%

PTV(Inf) LPO-InfPTV(Sup) LAO-match SPO-patch

Page 18: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Lacrimal Gland CarcinomaPTV50 4 – 5 FieldsPTV50.4 5 Fields

Inf - LPO Sup – LAO match Sup – SPO patch

LPO LAOLPO LAO

Page 19: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Patch Field Selection

Patch Field Angle Selection optimal geometric coverage (G = 230 ) avoid inhomogeneity along path (G = 205 )

Patch line Patch line

Gantry 230 Gantry 205

Page 20: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Patch Field – Beam Angle Selectiong

Gantry 230° Gantry 205°

G=205°

G 230°G=230°

Page 21: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Distal Blockingg

selective pullback of range to spare OARs pullback achieved with added compensator potential pitfalls setup error or motion may nullify sparing ‘simple’ distal blocking may compromise

target coverage b t f h assess robustness of approach

Page 22: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Distal Blockinggadded compensator

distal blockTarget

p

OAR

range pullback

Page 23: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Distal Blockingg

underdose

Whole BS Partial BS

PTV

Whole BS Partial BS

BS

RPO FieldPartial BS

Page 24: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Clinical Tools

I t t Cli i l kflIntegrate Clinical workflow clinical database web-based applications mu model physics qa plan evaluation

quality assurance clinical efficiency & efficacy

Page 25: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Plan/Dvh Evaluation/

Dvh statistics for CTV, PTV, OARs

Courtesy R. Malyapa, C.McKenzie, Z. Li

, ,are extracted manually from plots

Page 26: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

RT PACSTPS

Treatment Plans Plan, Dose , ROI Information

Clinical RT-PACSTPS Information Server

Active Forms

SharePointServerPhysics

• Plan• Contouring • DVH

Q/A Forms

Clinicians

• Prescription

Tx Delivery

Electronic Q/A Process:Electronic Q/A Process:Upon export of Upon export of txtx plan to RTplan to RT--PACS, clinical Q/A forms are generated. PACS, clinical Q/A forms are generated. Tx. DeliveryActive forms are sent via email to the personnel on the list . Active forms are sent via email to the personnel on the list . Contents of active forms is stored in a Clinical Information Database.Contents of active forms is stored in a Clinical Information Database.

Courtesy V. Frouhar

Page 27: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Electronic Q/A Process:Electronic Q/A Process:U f l RTU f l RT PACS i li i l Q/APACS i li i l Q/AUpon expert of treatment plan to RTUpon expert of treatment plan to RT--PACS, appropriate clinical Q/A PACS, appropriate clinical Q/A forms are generated. Active forms are sent via email to the personnel forms are generated. Active forms are sent via email to the personnel on the list . Contents of active forms is stored in a Clinical on the list . Contents of active forms is stored in a Clinical Information Database.Information Database.

Page 28: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4
Page 29: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4
Page 30: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Integration of Clinical Workflow

Patient Information

R&VHIS

External Connectivity

IGRT LinacTPSPhysics Calc. Clinical

D t M t Machine Data

TPSCT SimPET/MRQuality

Assurance

RXDelivery Setup

Data Management&

Workflow

Clinical I f ti

PhysicianPhysicist

Information DosimetristTherapist

Page 31: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Golden Beam Data

TPS commissioning time consuming share beam data among gantries (institutions)?s a e bea data a o g ga t es ( st tut o s)? golden beam data set accuracy requirements on modeling parameters accuracy requirements on modeling parameters

• pristine depth dose & sobp• effective source size• virtual sad• effective sad

Page 32: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Golden Beam Data

max error in penumbra of ±0.5mm in air, at isocenter

Courtesy R. Slopsema

Page 33: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Golden Beam Data

Page 34: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Summaryy

Analytical Proton Algorithms provide accurate dose Analytical Proton Algorithms provide accurate dose model

Proton specific clinical planning issues requires p p g qvigilance

Dose plan is a snap shot of dose distribution Dose delivered depends on uncertainties in range, setup, organ motion, etc.

Select beams and parameters to minimi e ncertainties Select beams and parameters to minimize uncertainties Tools to integrate clinical workflow are essential Golden beam data looks feasible Golden beam data looks feasible

Page 35: New Developments in Proton Treatment Planning Systems Daniel … · 2009-05-12 · Beam Profile Calc vs Measurement R=15.1 M=10.4 d=5 mm d=10 cm Option B5 - R=15.1 g/cm^2, M=10.4

Thanks!