Subscriber access provided by FORDHAM UNIVERSITY
Analytical Chemistry is published by the American Chemical Society. 1155 SixteenthStreet N.W., Washington, DC 20036Published by American Chemical Society. Copyright © American Chemical Society.However, no copyright claim is made to original U.S. Government works, or worksproduced by employees of any Commonwealth realm Crown government in the courseof their duties.
Article
New Life of Ancient Pigments: Application inHigh Performance Optical Sensing Materials
Sergey M Borisov, Christian Würth, Ute Resch-Genger, and Ingo KlimantAnal. Chem., Just Accepted Manuscript • DOI: 10.1021/ac402275g • Publication Date (Web): 04 Sep 2013
Downloaded from http://pubs.acs.org on September 8, 2013
Just Accepted
“Just Accepted” manuscripts have been peer-reviewed and accepted for publication. They are postedonline prior to technical editing, formatting for publication and author proofing. The American ChemicalSociety provides “Just Accepted” as a free service to the research community to expedite thedissemination of scientific material as soon as possible after acceptance. “Just Accepted” manuscriptsappear in full in PDF format accompanied by an HTML abstract. “Just Accepted” manuscripts have beenfully peer reviewed, but should not be considered the official version of record. They are accessible to allreaders and citable by the Digital Object Identifier (DOI®). “Just Accepted” is an optional service offeredto authors. Therefore, the “Just Accepted” Web site may not include all articles that will be publishedin the journal. After a manuscript is technically edited and formatted, it will be removed from the “JustAccepted” Web site and published as an ASAP article. Note that technical editing may introduce minorchanges to the manuscript text and/or graphics which could affect content, and all legal disclaimersand ethical guidelines that apply to the journal pertain. ACS cannot be held responsible for errorsor consequences arising from the use of information contained in these “Just Accepted” manuscripts.
New Life of Ancient Pigments: Application in High Performance
Optical Sensing Materials
Sergey M. Borisov,*a
Christian Würth,b Ute Resch-Genger,
b and Ingo Klimant
a
a Institute of Analytical Chemistry and Food Chemistry, Graz University of Technology,
Stremayrgasse 9, 8010, Graz, Austria
bBAM Federal Institute for Materials Research and Testing, Division 1.10 – Biophotonics,
Richard-Willstätter-Str. 11, D-12489 Berlin, Germany
Abstract: Calcium, strontium and barium copper silicates are demonstrated to possess
valuable photophysical properties which make them particularly attractive for application in
optical chemosensors. Several examples of sensing materials based on these phosphors are
provided. Particularly, broad excitation and near infrared (NIR) emission makes them ideal
candidates for the preparation of ratiometric sensors based on absorption-based indicators.
Due to their excellent chemical and photochemical stability and high brightness, these
phosphors can serve as reference for fluorescent indicators to enable ratiometric intensity or
dually lifetime referenced measurements. Finally, the moderate temperature dependence of
the luminescence decay time enables intrinsic temperature compensation of the sensing
materials at ambient temperatures. The improved sensitivity at temperatures above 100 °C
makes these new materials promising candidates for high temperature thermographic
phosphors.
Introduction
Calcium copper silicate Egyptian Blue CaCuSi4O10 and barium copper silicate Han Blue
BaCuSi4O10 are ancient pigments which were used already thousands of years ago in the Old
Kingdom in Egypt and during the Western Zhou period in China, respectively.1 Their high
chemical stability is emphasized by the fact that even after such a long time, the colors of
ancient monuments and artifacts retained their original splendor. The luminescent properties
of numerous minerals are well known for many decades, but the NIR luminescence of
Egyptian blue and Han Blue remained astonishingly undiscovered until very recently. Accorsi
et al. reported the interesting photophysical properties of Egyptian blue (absorption in the
Page 1 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
2
green-red part of the spectrum, NIR emission at about 900 nm and good brightness) and
suggested that these properties can be useful for application of this pigment in optical
materials.2 Pozza et al. indicated that Han Blue is also luminescent in the near-infrared (NIR)
region.3 Very recently Johnson-McDaniel et al. showed that Egyptian Blue nanosheets can be
produced upon prolong treatment of the phosphor with hot water, thereby preserving its
favorable luminescence properties.4 However, despite these exciting reports, the phosphors
have not yet found any application in advanced optical materials.
Optical chemosensors represent a very important and constantly growing field of
analytical chemistry.5,6
They often provide distinct advantages over conventional analytical
techniques due to the absence of electromagnetic interferences, minimal invasiveness,
suitability for miniaturization and continuous monitoring of analyte concentration as well as
their typically lower costs. Apart from an indicator (which responds to the analyte of interest
by altering its optical properties such as absorption, luminescence intensity, luminescence
decay time, anisotropy etc.) and a polymer or a sol-gel matrix, optical chemosensors often
require additional components to ensure reliable sensing. For example, fluorescence intensity
alone is a rather ambiguous parameter, which is affected also by fluctuations of the excitation
light intensity and indicator concentration, and is thus rarely used. Fluorescence lifetime
measurements are much more reliable,7 but the required instrumentation is more expensive.
Thus ratiometric wavelength referencing, requiring either two spectrally distinguishable
absorption (excitation) or emission bands, is rather popular, particularly in (microscopic)
imaging applications.8,9
Since most probes emit only at a single wavelength, typically a
second dye excitable at the same wavelength but emitting in a different part of the spectrum is
added for signal referencing. Alternatively, referencing of the fluorescence intensity can also
be realized with the help of the so-called Dual Lifetime Referencing (DLR) approach.10-13
DLR relies on the use of a phosphorescent reference which has a significantly longer decay
time than the fluorescent indicator (typically µs-ms and ns time domains, respectively). In the
frequency domain, for example, the overall phase angle is a function of the amplitudes and the
phase angles of both luminescent components. Recently, we reported a new approach for
referencing absorption-based indicators.14
In this scheme, a broadband absorbing or a
broadband emitting phosphor acts as a second luminophore, the luminescence intensity of
which is modulated by changes in the absorption of the indicator due to the inner filter effect.
Finally, it is well known that the response of all optical chemosensors can be affected by
temperature as the optical properties of most indicators and fluorophores are temperature-
dependent. Although temperature can be measured independently with e.g., a thermocouple, it
Page 2 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
3
is often more attractive to use an optical probe for this purpose. In this case, both parameters,
i.e., analyte concentration and temperature, are measured at the same place which improves
the robustness of the sensor against fluctuations in temperature and enables miniaturization
(e.g., fiber-optic microsensor). A number of luminescent metal complexes15-17
and
thermographic phosphors18-21
have been employed so far in optical temperature sensors and in
dually- or multi-analyte sensing materials operating at ambient temperature.22-24
However,
high temperature thermographic phosphors are also of much interest.25
In this contribution, we will demonstrate that Egyptian blue, Han blue and strontium
copper silicate represent highly versatile materials for application in optical sensors.
Particularly, it will be shown that these phosphors are promising as references for fluorescent
indicators utilizing both ratiometric and DLR schemes as well as for absorption-based
indicators and can also serve as internal temperature probes.
Experimental part
Materials. Egyptian Blue and Han Blue were obtained from Kremer Pigmente (Aichstetten,
Germany). Ethyl cellulose (ethoxyl content 49%), m-Cresol purple, calcium carbonate,
strontium carbonate, trimethylsilyl chloride, tetraoctylammonium hydroxide solution
(TOAOH, 20 % in methanol) were obtained from Sigma-Aldrich. Silicon oxide (99.9 %, 1µm
particles), vinyl-terminated polydimethylsiloxane (viscosity 1000 cSt.), methylhydrosiloxane
dimethylsiloxane copolymer (viscosity 25-35 cSt.), tetravinyltetramethyl cyclotetrasiloxane
and platinum divinyltetramethyldisiloxane complex were purchased from ABCR (Karlsruhe,
Germany). Basic copper carbonate was obtained from Fischer Scientific. The solvents were
from VWR (Austria). Barium carbonate and the buffer salts N-cyclohexyl-2-
aminoethanesulfonic acid, 2-(N-morpholino)ethanesulfonic acid and N-cyclohexyl-3-
aminopropanesulfonic acid were supplied by Carl Roth (Germany). Polyurethane hydrogel
(HydromedTM
D4) was purchased from AdvanSource biomaterials (Wilmington, USA).
Poly(ethylene terephthalate) (PET) support Melinex 505 was from Pütz (Taunusstein,
Germany).The gases for the calibration – carbon dioxide and nitrogen were supplied by Linde
(Austria).
[5-(4-hydroxyphenyl)-3-phenyl-1H-pyrrol-2-yl]-[5-phenyl-3-phenylpyrrol-2-ylidene]amine
“aza-BODIPY” was prepared as reported previously.26
Page 3 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
4
High temperature solid state synthesis of the phosphors. The synthesis is exemplified for
calcium copper silicate (SrCuSi4O10). The phosphor was prepared via a high temperature solid
state reaction using silicon oxide, strontium carbonate and basic copper carbonate.
Stoichiometric amounts of these substances were thoroughly homogenized in an agate mortar
and sintered at 1000 °C for 24 h. The resulting material was homogenized to powder in an
agate mortar. The samples of Egyptian Blue (CaCuSi4O10) and Han Blue (BaCuSi4O10) were
prepared analogously using calcium carbonate and barium carbonate, respectively.
Lipophilization of Egyptian blue. For application in the sensing materials, Egyptian blue
was ground to microparticles (1-5 µm) in a ball mill. The surface of the particles was rendered
lipophilic via silanization using the following procedure: 1.6 g of Egyptian blue
microparticles were dispersed in 5 mL of anhydrous tetrahydrofuran and 0.5 mL of
trimethylsilyl chloride were added. The suspension was stirred for 30 min, the sediment was
separated via centrifugation, washed four times with acetone and twice with ethanol, and
dried.
Preparation of the pH sensors
250 mg of hydrogel D4, 0.5 mg of the pH indicator aza-BODIPY and 250 mg of Egyptian
blue lipophilic microparticles were dissolved/dispersed in 2.5 g of an isopropanol:H2O
mixture (9:1 v/v). The “cocktail” was knife-coated on a polyethylene terephthalate support to
give about 8 µm-thick sensor layers after solvent evaporation. Alternatively, the fiber-optic
sensors were prepared by coating the “cocktail” on a distal end of a PMMA waveguide (∅ 1
mm, length 1 m, Ratioplast, Lübbecke, Germany).
Preparation of the CO2 sensors
100 µL of tetraoctylammonium hydroxide were added to 2 mg of m-Cresol Purple and the
resulting solution was saturated with carbon dioxide. In the second vial, 0.2 g of ethyl
cellulose were dissolved in a mixture of 1.52 g of ethanol and 2.28 g of toluene. The content
of both vials was combined and the obtained “cocktail” was knife-coated on a polyethylene
terephthalate support and was allowed to dry for 1 h at room temperature. The thickness of the
layer after solvent evaporation was estimated to be about 3.5 µm.
A “cocktail” of silicone primers was obtained by mixing 800 mg of vinyl-terminated
polydimethylsiloxane, 32 µl of methylhydrosiloxane dimethylsiloxane copolymer, 2 µl of
tetravinyltetramethyl cyclotetrasiloxane and 800 mg of hexane. Micrometer-sized lipophilic
Page 4 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
5
particles of Egyptian Blue (400 mg) were dispersed in the “cocktail”. Finally, 4 µl of the
platinum complex catalyst were added and the composition was knife-coated onto the CO2-
sensitive layer (75 µm spacer) and was allowed to cross-link at 60 °C for 20 min.
Measurements
The absorption spectra were acquired on a Cary 50 UV-VIS spectrophotometer from Varian.
Emission spectra of individual phosphors were measured in front face geometry with a
calibrated FSP 920 fluorescence spectrometer from Edinburgh Instruments equipped with
double monochromators and a liquid nitrogen cooled NIR-sensitive InP/InGaAs
photomultiplier from Hamamatsu (R5509-72). All measurements were performed with
polarizers in the excitation (0°) and the emission channel (54.7°). All emission spectra
presented are corrected for the spectral responsivity of the instrument´s detection channel
(instrument characterization traceable to radiometric units, i.e., to the spectral radiance
scale.27,28
Luminescence excitation spectra and emission spectra of the sensing materials were
acquired on a Fluorolog 3 fluorescence spectrometer (Horiba) equipped with a
photomultiplier R2658 (Hamamatsu) optimized for the spectral range of 300-1050 nm. About
10 µm thick layers of the phosphors in polystyrene (1:1 w/w) on a glass support were used for
the measurement of the excitation spectra in order to minimize wavelength-dependent
scattering effects. The excitation spectra were corrected for the wavelength-dependent output
of the Xe arc lamp.
Planar carbon dioxide and pH optodes were read out with a Fluorolog 3 spectrometer.
A home made flow-through cell was used. In case of the carbon dioxide sensors, the gas
mixtures were adjusted using Red-y gas flow controllers (Vögtlin Instruments AG, Aesch,
Switzerland). The gas mixtures were adjusted to 100 % relative humidity. The pH of the
buffer solutions was controlled by a digital pH meter (InoLab pH/ion, WTW GmbH & Co.
KG, Germany) calibrated at 25 °C with standard buffers of pH 7.0 and pH 4.0 (WTW GmbH
& Co. KG). The buffers were adjusted to constant ionic strength (IS = 0.15 M) using sodium
chloride as the background electrolyte. The temperature was kept constant at 37 °C using a
cryostat ThermoHaake DC50 (Thermo Fisher Scientific Inc). The fiber-optic pH sensors were
read-out with a Firesting oxygen meter from Pyroscience (Aachen, Germany) which is
equipped with a read 625 nm LED and a long-pass RG 9 filter from Schott. A modulation
frequency of 1.5 kHz was used.
The carbon dioxide sensors were also read out using a dually-emitting LED (590/650
nm) from PUR-LED®
GmbH & Co. KG (Selzen, Germany). A short-pass Calflex X filter
Page 5 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
6
from Linos was used in front of the LED. The luminescence was detected at a modulation
frequency of 916 Hz using a lock-in amplifier from PreSens (Germany) equipped with a
photodiode and a long-pass RG-9 filter from Schott (Germany).
The temperature sensitivity of the phosphors was investigated using a Firesting-mini
device from Pyroscience. A glass waveguide (∅ 3 mm, length 15 cm) was used as an
extender. The distal end of the waveguide was brought into direct contact with the phosphor
powder kept in a 4 mm glass vial. The vial was heated in a home-made metal block using a C-
MAG HS 7 heating plate from IKA®-Werke GmbH & Co. KG (Staufen, Germany). The
modulation frequency of 1.5 kHz was used to access the decay times.
Results and Discussion
Photophysical properties
Luminescence excitation and emission spectra of Egyptian blue, Han blue and strontium
copper silicate are shown in Fig. 1. The phosphors have a broad absorption / excitation band
located in the green-red part of the spectrum which is attributed to 2B1g →
2A1g and
2B1g →
2Eg transitions.
2 Excitation due to
2B1g →
2B2g transition occurs in the NIR part of the
spectrum. As follows from Fig. 1, the excitation bands in the visible region are very similar
for all phosphors, but a small bathochromic shift (∼ 10 nm, 280 cm-1
) of the 2B1g →
2A1g band
is observed on going from calcium copper silicate to barium copper silicate. The NIR
excitation is affected more significantly and shifts from 806 nm for calcium copper silicate to
842 nm (530 cm-1
) for barium copper silicate.
All phosphors emit in the NIR above 800 nm (Fig. 1). Similarly to the NIR excitation,
the emission associated with the 2B2g →
2B1g transition shifts bathochromically from calcium
to barium copper silicates (λmax = 909, 914 and 948 nm for calcium, strontium and barium
copper silicates, respectively). Precise determination of the luminescence quantum yields in
the NIR region, which is not possible with a spectrofluorometer in a 0°/90° or front face
measurement geometry due to size-dependent scattering of the micrometer-sized phosphor
particles and requires an integrating sphere setup, was beyond the scope of this study.27
Previous results suggest a luminescence quantum yield of 10.5 % for Egyptian blue.2
Measurements under red excitation (625 nm LED) and a photodiode as a detector revealed
that the brightnesses of the strontium copper silicate and Han blue are comparable to that of
Egyptian blue. It should be kept in mind here, however, that the brightness equals the product
of the luminescence quantum yield and the molar absorption coefficient. Considering a rather
Page 6 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
7
high amount of the chromophore in the phosphors, which is about 10-20-fold higher than for
typical phosphors (e.g., chromium(III)-activated hosts such as yttrium aluminum garnet,
yttrium aluminum borate, spinel, aluminum oxide (ruby) or manganese(IV)-activated
magnesium fluorogermanate; typical doping of 0.5-5% of the activator), the overall brightness
of our materials is excellent. This is particularly true for thin layers of the phosphor materials
since in this case a significantly more efficient absorption can compensate for lower quantum
yields. For example, excitation of a thin sensing layer of a mixture of chromium(III)-activated
gadolinium aluminum borate (Cr-GAB) and Egyptian blue (25 % w/w of each phosphor in
polystyrene) with red light (600 nm, corresponding to the excitation maxima of both
phosphors) results in about 3-fold higher signals for Egyptian blue compared to Cr-GAB.
500 600 700 800 900 1000 1100 1200
0.0
0.2
0.4
0.6
0.8
1.0
500 600 700 800 900 1000 1100 1200
0.0
0.2
0.4
0.6
0.8
1.0
Norm
. L
um
inesc
en
ce in
ten
sity
CaCuSi4O10 SrCuSi
4O10 BaCuSi
4O10
Wavelength, nm
Excitation Emission
Figure 1. Excitation (lines) and emission spectra (dashed lines) of Egyptian blue (black),
strontium copper silicate (red) and Han blue (blue line). The insert shows photographic
images of the measured µm-sized phosphor particles, i.e., the powdered pigments under
ambient light.
High homogeneity of the sensing materials is important to ensure reproducibility of the
calibration and sensor response. This is particularly true if the phosphors or other luminescent
species are used for referencing purposes in the composite materials. Therefore, the phosphors
were ground to microparticles (∅ 1-5 µm) using a ball mill to enable a better reproducible
sensor production. Although grinding does not appreciably affect the optical properties of
these phosphors, we note a slight shortening of the decay time of the ground phosphors
compared to the larger crystals obtained after sintering. For example, the decay times of the
Egyptian Blue at 20 °C were 159 µs and 126 µs after sintering and after grinding,
respectively. For comparison, the decay time of commercially available Egyptian blue powder
(Kremer Pigmente) was measured to be 139 µs. The decrease of the decay time can be
explained by accumulation of the defects induced by mechanical stress during grinding.
Page 7 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
8
Ratiometric pH-sensing materials
The broad excitation bands in the visible part of the spectrum of the phosphors (Fig. 1) are
ideally compatible with the absorption of a large number of fluorescent indicators and very
favorable for ratiometric (2 wavelength) referencing of luminescence intensity. Since
fluorescent indicators typically possess rather small Stokes shifts, the emission of the
phosphors at 900 -1100 nm can be easily spectrally separated from that of the analyte-
responsive probe. High chemical and photochemical stability of the phosphors is of course
also advantageous. Fig. 2 shows the absorption and emission spectra of the
tetraphenylazadipyrromethene pH indicator (“aza-Bodipy”)26
and the emission spectra of the
pH-sensitive material. The sensor is prepared by dissolving/dispersing the pH indicator and
Egyptian blue micropowder in a polyurethane hydrogel D4. In order to render the surface of
the microparticles stable in aqueous media (hydrogel) end-capping with trimethylsilyl groups
was performed. The indicator and the phosphor are excitable with the red light. Although both
emit in the NIR part of the spectrum, the emissions are well separated. As expected, the
fluorescence of the pH probe is quenched as the pH increases but the luminescence of
Egyptian blue is not affected by pH. A minor increase of the luminescence intensity of the
phosphor (about 10 % for going from pH 5.2 to 10.3) is due to a change in absorption of the
pH indicator and thus, a decrease of the inner-filter effect. In fact, the absorption maxima of
the protonated and deprotonated forms of the pH indicator are located at 687 and 743 nm,
respectively (Fig. 2a), and the absorption at the excitation wavelength (625 nm) is about 3-
times more efficient for the protonated form. Evidently, the system is ideal for ratiometric
emission measurements. The apparent pKa value calculated from the respective calibration
curve for the luminescence intensity ratio is 7.53 which is adequate for measurements in the
physiological probes and for application in seawater. It should be mentioned that simple
synthetic modification allow to tune the pKa value of the indicator in a broad range.26
Considering the spectral properties of the two other phosphors (Fig. 1), they can be used for
ratiometric referencing of other indicators with even more bathochromically shifted excitation
and emission spectra.
Page 8 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
9
700 750 800 850 900 950
0
100
200
300
400
500
5 6 7 8 9 10
0.0
0.5
1.0
1.5
2.0
550 600 650 700 750 800 850
0.0
0.1
0.2
0.3
0.4
0.5
550 600 650 700 750 800 850
0.0
0.2
0.4
0.6
0.8
1.0
inte
nsi
ty,
a.u
.L
um
ines
cen
ce
Wavelength, nm
5.2 6.84
7.27 7.86
8.22 8.59
9.01 10.28
ra
tio
71
5/9
10
nm
Lu
min
escen
ce I
nte
nsi
ty
pH
pKa = 7.53
N
N
NB
FF
O
N
N
NB
FF
HO
No
rm.
Inte
nsi
ty
Ab
sorp
tio
n
Wavelength, nm
Figure 2. A: Spectral properties of the aza-Bodipy pH indicator in hydrogel D4 (black thick
line – absorption spectrum at pH 5.2; blue thin line – absorption spectrum at pH 10.3; red
dotted line – emission spectrum at pH 5.2 and λexc = 625 nm); B: pH dependence of the
emission spectra of the sensing material consisting of the pH indicator and µm-sized Egyptian
blue particles (37 °C, 150 mM ionic strength, λexc = 625 nm) dissolved/dispersed in hydrogel
D4; C: Respective calibration for the ratio of the luminescence intensities (squares) and a
sigmoidal fit (line).
Dually lifetime referenced pH sensors
Dual Lifetime Referencing (DLR) represents another technique for referencing the
fluorescence of indicators. Here, an inert analyte-insensitive luminescent material is dispersed
in the polymer matrix along with the analyte-responsive fluorescent indicator. A reference
suitable for this application should be chemically and photochemically inert and should
possess a relatively long luminescence decay time (typically in µs – ms time domain). This
technique can be used both in the time domain12
and in the frequency domain.11,29
In the
frequency domain, the overall phase shift is measured; both the fast-decaying fluorescence of
the indicator and the long-lived luminescence of the reference contribute to this value. The
Page 9 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
10
most frequently used DLR references are hence metal complexes .30,31
Due to their
disadvantageous liability to quenching by oxygen, these dyes, however, have to be
encapsulated in gas-blocking polymers (e.g., polyacrylonitrile)11,30
which may be challenging.
Application of phosphors as DLR-reference materials is rather rare.32
As we demonstrate
here, Egyptian Blue and other copper silicate phosphors represent excellent DLR references
since they combine bright long-lived luminescence, excellent (photo)stability, and inertness to
oxygen. Fig. 3 shows the calibration curve for a fiber-optic pH sensor. The material used to
coat the plastic PMMA fiber is identical to that employed in the planar optodes with
ratiometric read-out. The fluorescence amplitude of the indicator is proportional to the
difference of the cotangents of the overall phase shift and the phase shift of the reference:30
ref
indref
A
A
sin
1cotΦcotΦ ⋅
Φ=−
ref
where Φ and Φref are the phase angles of the overall signal and of the luminescent reference,
respectively; Aind and Aref – are the amplitudes of the fluorescent indicator and of the
luminescent reference, respectively. As follows from Fig. 3, the apparent pKa value obtained
for the DLR-sensor (7.47) is very close to that calculated from the ratiometric read-out. The
individual sensors have slightly varying ratios of the fluorescent dye and the reference and are
not calibration-free (Fig. 3, phase shift plots). Thus, a two point calibration is recommended
(low pH and high pH buffer). The same refers also to the ratiometric sensor.
5 6 7 8 9 10
10
20
30
40
50
5 6 7 8 9 10
0.0
0.2
0.4
0.6
0.8
1.0
Norm
ali
zed
Cot ΦΦ ΦΦ
-Co
t ΦΦ ΦΦ0
Ph
ase
sh
ift ΦΦ ΦΦ
, °
pH
pKa = 7.47
Figure 3. Calibration plots for the fiber-optic DLR-pH sensor (37 °C, 150 mM ionic strength)
based on an aza-Bodipy pH indicator and Egyptian blue in hydrogel D4. Symbols represent
experimental data and lines sigmoidal fits. The phase shifts for three individual sensors are
shown. The values CotΦ-CotΦref were normalized to enable the comparison of all individual
sensors. The inserts show the photographic images of a fiber-optic sensor without LED
illumination and during the readout.
Page 10 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
11
Ratiometric inner filter effect sensors
Very recently, we presented a new scheme for the referencing of absorption-based
indicators.14
Here, a reference acts a secondary emitter, the luminescence intensity of which is
modulated by an indicator with an analyte-responsive absorption via inner filter effects (in
absorption and /or emission). A suitable reference for this application should feature either a
broad excitation spectrum or a broad emission spectrum which overlap with at least one
absorbing form of the indicator. Such a scheme allows also a ratiometric readout. In addition
to the general versatility of this approach, which can utilize all types of broad available
colorimetric indicators and probes, and its inertness to fluorescence quenchers, the use of
phosphors with long luminescence decay times as reference allows complete elimination of
autofluorescence originating from the sample, indicator and/or optical components of the
detection system. As follows from Fig. 1, the optical properties of Egyptian blue and other
copper silicate phosphors make them ideal references for this new sensing scheme. Currently
the number of available NIR colorimetric probes is rather limited, particularly those absorbing
above 850 nm, but the situation is likely to change in the future. We therefore demonstrate
here only the applicability of the phosphors for this versatile sensing scheme using an
indicator-modulated inner filter effect in excitation, see Fig. 4. The broad luminescence
excitation spectra of the phosphors almost perfectly overlap with the absorption of the
deprotonated form of triphenylmethane indicators applied as pH, carbon dioxide, and
ammonia probes. A two layer carbon dioxide sensor was realized, with the first layer
containing m-Cresol purple and tetraoctylammonium hydroxide dissolved in ethylcellulose
and the second layer Egyptian blue microparticles dispersed in silicone rubber (Fig. 4a). In
this sensor design, silicone rubber acts not only as a matrix for the immobilization of the
reference phosphor but also as a permeation-selective membrane which protects the sensitive
layer from protons and other ions. As can be observed in Fig. 4c, the excitation spectrum of
the sensing material is dependent on pCO2, thereby reflecting the modulation of the excitation
light intensity by the absorption of m-Cresol purple (Fig. 4b) as change in the emission
intensity of the phosphor. This also demonstrates that ratiometric readout in luminescence
excitation is possible.
Dually emitting LEDs are available matching both the excitation wavelength of the
phosphors and the indicator absorption. Thus, a very compact, simple, and inexpensive setup
for the ratiometric readout of the optical sensors can be realized (Fig. 4a), using a dually
emitting LED, two optical filters, and a silicon photodiode. Fig. 4d shows the calibration plot
Page 11 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
12
obtained using a dually emitting red-orange (650/590 nm) LED. Relatively high standard
deviations are observed since the thickness of both layers can vary significantly for individual
sensors. Thus, the sensors realized according to the above scheme are not calibration-free, and
2 point calibration is required before the measurement. Nevertheless, the combination of
Egyptian blue (or other copper silicate phosphors) with absorption-based indicators represent
an excellent way to realize simple and robust optical sensors for various analytes. Indeed, the
pH and ammonia sensors based on triphenylmethane dyes can be designed in a similar
manner. Referencing of irreversible colorimetric probes can also be performed. Notably, very
large effective shift between excitation and emission of the phosphors (> 200 nm) favors
elimination of potential interferences such as autofluorescence and Raman scattering. Further
simplification can include a photodiode with an integrated daylight filter (e.g. BPW 34 FAS
available from Osram), rendering other optical filters unnecessary. Alternatively, the filter-
less set-up can be realized by eliminating potential interferences with time-resolved
measurement.
It is interesting to compare the properties of the sensing materials based on the new
phosphors which employ three different read-out schemes presented above (Table 1).
Generally, the material relying on the inner filter effect-modulated excitation read-out is
significantly more robust towards potential interferences (fluorescence quenchers, background
fluorescence from the sample) which can be even further minimized in the time-resolved
measurement. Along with the DLR scheme this method requires only a very simple
instrumentation with few optical components. Indeed, only a simple long-pass filter is
necessary for the inner filter read-out, a long-pass and a short-pass filter for the DLR method,
and at least three filters (short-pass for the excitation, a long-pass and a band pass for the
emission, and an additional dichroic mirror) for the ratiometric scheme. On the other hand, the
variation in the layer thickness only minor affects the calibration in the DLR and ratiometric
scheme, but more severely in the inner-effect read-out.
Page 12 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
13
Figure 4. A: Cross-section of the carbon dioxide sensor and optical setup for the readout of
the sensor with a dually emitting LED; B: Absorption spectra of m-Cresol purple (in ethyl
cellulose) in the deprotonated (low pCO2, 1) and protonated (high pCO2, 2) forms and the
excitation spectrum of Egyptian blue (3); C: Excitation spectra (λem = 900 nm) of the carbon
dioxide sensor at varying pCO2 (25 °C) and the emission spectra of the dually emitting LED
(590/650 nm); D: calibration plot obtained for the excitation of the carbon dioxide sensor with
a dually emitting LED, the emission monitored at λ > 730 nm (RG 9 filter) with a photodiode
(25 °C) Average values for 3 individual sensors are shown. The chemical structure of the
indicator used here is shown in Fig. 4D.
Page 13 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
14
Table 1. Comparison between different read-out schemes used in this work.
Sensing scheme
Property
Ratiometric
(2-λ) emission
Dual Lifetime
Referencing
Inner filter effect-
modulated
excitation
Influence of the fluorescence
quenchers on the calibration
strong strong no influence
Influence of the thickness of
the sensing layer on the
calibration
minor minor large
Influence of background
fluorescence on the calibration
minor minor negligible
Possibility of the time-
resolved measurements
not possible not possible possible
Versatility of formats (planar
optodes, paints, microsensors
etc.)
high high low
Complexity of optical set-up moderate low very low
Intrinsic optical temperature
compensation
possible (more
complex
calibration
algorithms)
possible (more
complex
calibration
algorithms)
possible
(calibration is not
affected)
Phosphors as optical temperature probes
The response of almost every optical sensor is temperature dependent. Hence, the temperature
of the sensor environment should be known or be kept constant in order to obtain reliable
data. To circumvent this additional source of measurement uncertainties, either a constant
monitoring of the ambient temperature is required or a tool that enables an intrinsic
temperature compensation of the optical sensor. As follows from Fig. 5, which
shows the temperature dependence of the luminescence decay time (a) and luminescence
intensity (c) of the phosphors, our NIR-emissive phosphors are also excellently suited for this
task. Evidently, both parameters decrease with temperature for all materials. The temperature
sensitivity of all phosphors (Fig. 5b) is moderate at ambient temperature (dτ/dT 0.25, 0.24 and
0.21 %/K at 25 °C for Han blue, strontium copper silicate and Egyptian blue, respectively).
These values are significantly lower than those for recently reported chromium(III)-activated
yttrium aluminum borate (0.8%/K)21
but are comparable to e.g. Ruby (∼0.22 %/K)20
or
manganese(IV)-activated magnesium fluorogermanate (∼0.25 %/K)33
which are used as
thermographic phopshors. Interestingly, the dynamic range of the phosphors is significantly
different: Han blue operates at lower temperatures than Egyptian blue and strontium copper
Page 14 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
15
silicate. The luminescence of Han blue is almost completely quenched at 220 °C but the other
phosphors are still luminescent at temperatures above 300 °C. Importantly, the sensitivity
significantly improves at higher temperatures (Fig. 5c). For example, the sensitivities as high
as 1 %/K are achieved for Han blue, strontium copper silicate and Egyptian blue at 154, 283
and 303 °C, respectively. About 30-34 % of the original luminescence intensity is retained at
these temperatures which insures good signal-to-noise ratios. These observations underline
the excellent suitability of the investigated materials as thermographic phosphors operating at
100-400 °C. Importantly, no hysteresis of the calibration was observed and the plots were
identical for the cycles of increasing and decreasing temperature. Despite moderate
sensitivities at ambient temperatures, the phosphors enable the temperature compensation of
optical sensors (e.g. pO2, pH, pCO2 etc.). They are particularly attractive as temperature
probes for ratiometric sensing relying on absorption-based indicators since variations of the
luminescence intensity due to temperature-induced quenching are not relevant in this sensing
scheme.
0
50
100
150
200
0
1
2
3
4
-50 0 50 100 150 200 250 300 350 4000.0
0.2
0.4
0.6
0.8
1.0
Dec
ay
tim
e, µ
s (a)
(c)
(b)
dττ ττ/
dT
, %
/K
Han Blue
SrCuSi4O
10
Egyptian Blue
Norm
. In
ten
sity
, a
.u.
Temperature, °C
Page 15 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
16
Figure 5. Temperature dependence of the luminescence decay time (a) and luminescence
intensity (c) and the temperature sensitivity of the luminescence decay time (b) for the
phosphors.
Conclusions
We have demonstrated that calcium/strontium/barium copper silicates represent very
promising phosphors for application in optical sensing materials. Particularly, they can be
used as inert references for ratiometric and DLR readouts in combination with NIR
fluorescent indicators and as secondary emitters for the ratiometric luminescence readout of
absorption-based indicators. The latter scheme provides unique possibility of using very
simple and inexpensive instrumentation for the sensor readout. The phosphors favorably
compare to other materials used as references in optical sensors due to the combination of
several unique features such as broad excitation spectra in green-red part of the
electromagnetic spectrum, strong NIR emission, long luminescence decay times, high
chemical and photochemical stability, and inertness to oxygen. It is also important to note that
all the phosphors can be easily prepared from low cost materials and Egyptian blue and Han
blue are also commercially available. The phosphors can be additionally used to optically
access temperature for such sensors (albeit with moderate sensitivity at ambient temperatures)
or as high temperature thermographic phosphors if used without other additives. It can be
concluded that the ancient pigments finally found a new life as components of high
performance sensing materials. Other potential applications of these exciting phosphors in
optical materials are yet to be discovered.
AUTHOR INFORMATION
Corresponding author. Tel.: +43 316 873 32516. Fax: +43 316 873 32502. E-mail:
References
(1) Berke, H. Chem. Soc. Rev. 2007, 36, 15–30.
(2) Accorsi, G.; Verri, G.; Bolognesi, M.; Armaroli, N.; Clementi, C.; Miliani, C.; Romani,
A. Chem. Commun. 2009, 3392–3394.
(3) Pozza, G.; Ajò, D.; Chiari, G.; De Zuane, F.; Favaro, M. J. Cult. Herit. 2000, 1, 393–
398.
Page 16 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
17
(4) Johnson-McDaniel, D.; Barrett, C. A.; Sharafi, A.; Salguero, T. T. J. Am. Chem. Soc.
2012.
(5) McDonagh, C.; Burke, C. S.; MacCraith, B. D. Chem. Rev. 2008, 108, 400–422.
(6) Wolfbeis, O. S. J. Mater. Chem. 2005, 15, 2657–2669.
(7) Becker, W. J. Microsc. 2012, 247, 119–136.
(8) Yang, Y.; Zhao, Q.; Feng, W.; Li, F. Chem. Rev. 2013, 113, 192–270.
(9) Kikuchi, K. Chem. Soc. Rev. 2010, 39, 2048–2053.
(10) Schäferling, M. Angew. Chem. Int. Ed. 2012, 51, 3532–3554.
(11) Huber, C.; Klimant, I.; Krause, C.; Wolfbeis, O. S. Anal. Chem. 2001, 73, 2097–2103.
(12) Liebsch, G.; Klimant, I.; Krause, C.; Wolfbeis, O. S. Anal. Chem. 2001, 73, 4354–4363.
(13) Mayr, T.; Klimant, I.; Wolfbeis, O. S.; Werner, T. Anal. Chim. Acta 2002, 462, 1–10.
(14) Borisov, S. M.; Klimant, I. Anal. Chim. Acta 2013, 787, 219–225.
(15) Liebsch, G.; Klimant, I.; Wolfbeis, O. S. Adv. Mater. 1999, 11, 1296–1299.
(16) Borisov, S. M.; Wolfbeis, O. S. Anal. Chem. 2006, 78, 5094–5101.
(17) Khalil, G. E.; Lau, K.; Phelan, G. D.; Carlson, B.; Gouterman, M.; Callis, J. B.; Dalton,
L. R. Rev. Sci. Instrum. 2004, 75, 192–206.
(18) Kissel, T.; Baum, E.; Dreizler, A.; Brübach, J. Appl. Phys. B Lasers Opt. 2009, 96, 731–
734.
(19) Aizawa, H.; Ohishi, N.; Ogawa, S.; Watanabe, E.; Katsumata, T.; Komuro, S.;
Morikawa, T.; Toba, E. Rev. Sci. Instrum. 2002, 73, 3089–3092.
(20) Aizawa, H.; Ohishi, N.; Ogawa, S.; Katsumata, T.; Komuro, S.; Morikawa, T.; Toba, E.
Rev. Sci. Instrum. 2002, 73, 3656–3658.
(21) Borisov, S. M.; Gatterer, K.; Bitschnau, B.; Klimant, I. J. Phys. Chem. C 2010, 114,
9118–9124.
(22) Nagl, S.; Wolfbeis, O. S. Analyst 2007, 132, 507–511.
(23) Stich, M. I. J.; Fischer, L. H.; Wolfbeis, O. S. Chem. Soc. Rev. 2010, 39, 3102–3114.
(24) Borisov, S. M.; Seifner, R.; Klimant, I. Anal. Bioanal. Chem. 2011, 400, 2463–2474.
(25) Khalid, A. H.; Kontis, K. Sensors 2008, 8, 5673–5744.
(26) Jokic, T.; Borisov, S. M.; Saf, R.; Nielsen, D. A.; Kühl, M.; Klimant, I. Anal. Chem.
2012, 84, 6723–6730.
(27) Wurth, C.; Grabolle, M.; Pauli, J.; Spieles, M.; Resch-Genger, U. Anal. Chem. 2011, 83,
3431–3439.
(28) Resch-Genger, U.; DeRose, P. C. Pure Appl. Chem. 2012, 84, 1815–1835.
(29) Waich, K.; Borisov, S.; Mayr, T.; Klimant, I. Sensors Actuators B Chem. 2009, 139,
132–138.
(30) Bultzingslowen, C. von; McEvoy, A. K.; McDonagh, C.; MacCraith, B. D.; Klimant, I.;
Krause, C.; Wolfbeis, O. S. Analyst 2002, 127, 1478–1483.
(31) Huber, C.; Klimant, I.; Krause, C.; Werner, T.; Mayr, T.; Wolfbeis, O. S. Fresenius J.
Anal. Chem. 2000, 368, 196–202.
(32) Borisov, S. M.; Gatterer, K.; Klimant, I. Analyst 2010, 135, 1711–1717.
(33) Brübach, J.; Dreizler, A.; Janicka, J. Meas. Sci. Technol. 2007, 18, 764–770.
Page 17 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960
18
Graphical Table of Content Figure
Page 18 of 18
ACS Paragon Plus Environment
Analytical Chemistry
123456789101112131415161718192021222324252627282930313233343536373839404142434445464748495051525354555657585960