Newborn Screening for Critical Congenital Heart
Diseases(CCHD)
Lazaros Kochilas, MDAssociate Professor of PediatricsUniversity of Minnesota
Disclosures and Support
I will not discuss off label use and/or investigational use in my presentation
and
I have no conflicts of interest to disclose
Pediatric Grand Rounds
Lazaros Kochilas, MD
…however, I have an interest in the conflict of screening for CCHD!
Advocates:Families with CHDPed. Cardiologists
Possible opponents:Delivery HospitalsThird party payers
Neutral:Public Health analysts
• Identify need for screening for CCHD in the nursery
• Discuss evidence-based recommendations for newborn screening for
CCHD
• Describe efforts for screening implementation on state and national levels
• Discuss barriers for screening implementation
Learning objectives
Criteria for using and appraising screening• Population: - sufficiently high incidence of screened condition
- likely to be compliant
• Condition: - significant mortality / morbidity- known natural history with detectable presymptomatic period- treatment makes a difference when introduced early
• Test: - suitable (simple, safe, reliable, validated) - known distribution of values in diseased and non-diseased- acceptable validation process for (+) screens- widely available and acceptable
• Treatment: - acceptable, available, effective, agreement on whom to treat
• Program: - adequate staffing/facilities- program is acceptable and effective
- acceptable cost- quality management / ongoing re-evaluation
Selection Criteria for Newborn Screening Conceptual Framework
Definition - Identifiable at birth - Condition characteristics
Test characteristics – Treatment - Cost effectiveness
Concepts in screening
• All screening programs do harm; some do good as well
• Criteria for appraising screening
• Screening is a program not a test
• Assess opportunity cost
Wilson, J and Jungner, J: Principles and practice of screening for disease: WHO, 1968Gray, MJ: New concepts in screening; BJ Gen Practice, 2004, 54, 292-298
Disease
+ -Screeningadverse effects
- A B
+ C D
Particular challenges• Limitations of randomized controlled studies
• Limited evidence to assess benefit/risk ratio
• Need to calculate opportunity costs “Would you spend $$$ for screening for CHD?” vs. “If you had $$$ to spend for the field of CHD would you spend it for screening?”
• Public pressure for increasing sensitivity of the testing
• Challenge of definition
Nomination Form
HRSAAdministrative
Review
SACHDNC*
Evidence Review Group
Recommendations to the HHS Secretary
Process for addition of new conditions to the uniform panel of newborn screening
• Not adding to the NBS• Additional studies• Pilot study• Targeted screening• Add to NBS
Implementationworkgroup
*Secretary’s Advisory Committee on Heritable Disorders in Newborns and Children
oximetry screen 1st clinical reports
Larger prospective studyIn NY (Koppel)
Request to AHAFor recommendations
AHA commentEvidence not sufficient
Proposed bill in TNto mandate screening
Several European countries adopt pulse-ox screen
as standard of care
AHA/AAP statement comment
2002 2005
Large European prospective studies(Norway, Sweden, UK, Germany, Switzerland)
Timeline of pulse oximetry screen for CCHD
2008 2011
SACHDNCrecommended
adding pulse-ox screen
Definition of the primary target: Critical CHD (CCHD)
Requiring surgical, cath or pharmacologic intervention to
avoid death or end-organ damage
- Hypoplastic left heart syndrome (HLHS)
- Pulmonary atresia
- Tetralogy of Fallot (TOF)
- Total anomalous pulmonary venous return (TAPVR)
- Transposition of the great arteries (TGA)
- Tricuspid atresia
- Truncus arteriosus
Current Uniform Screening Panel
• 29 primary conditions
• 25 secondary targets
-20 metabolic detected by MS (AA, FAO, OA)- 3 Hg-pathies (S/S, S/β Thal, S/C)- 6 Others (BIOT, CAH, CF, CH, GALT, HEAR*)
-22 metabolic detected by MS (AA, FAO, OA)- 1 Hg-pathies variants- 6 Others (GAL-epimerase, GAL-kinase)
* Only point-of-care type of screening
CHD:
5-10 / 1,000 live births
1.4 cyanotic CHD / 1,000 live births
2 critical CHD / 1,000 live births
25,000 cases of CHD/yr in US
25% of infantile deaths
31% of neonatal deaths
Congenital Heart Diseases: The magnitude of the problem
http://www.cdc.gov/ncbddd/features/heartdefects-keyfindings2010.html
Heron, M., et al. (2009). Deaths: Final data for 2006. National Vital Statistics Reports, 57(14). U.S. CDC and Prevention.
All together 1.55 /1,000 live births
50% of deaths from CHD occur in 1st year and
50% of infantile deaths occur in 1st month of life
Timing of death from CHD
Boneva, R: Circulation. 2001;103:2376
Deaths from undiagnosed CHD in Wisconsin
• 345,573 births (2002-2006)
• 14 deaths during first 2 wks of life (1: 24,684)
Benton, N and Hokanson,J: Missed Congenital Heart Disease in NeonatesCongenital Heart Disease 2010; 5(3):292-296
Significant physiologic compromise from undiagnosed CHD
• 490 patients with undiagnosed critical CHD (2000-2003)
• 76 (15.5%) with significant physiologic compromise
• 33 (6.7%) preventable
• Incidence of potentially preventable 1:15,000-1:26,000
Schultz, A: Epidemiologic features of the presentation of critical CHD: implications for screeningPediatrics 2008; 121(4):751-757
Missed Critical Congenital Heart Diseases (CCHD)
Hoffman, J. It is time for routine neonatal screening by pulse oximetry. Neonatology 2011;99:1-9
Types of frequently unrecognized CCHD
Hoffman, J. It is time for routine neonatal screening by pulse oximetry. Neonatology 2011;99:1-9
Pulse oximetry as screening method- pulse oximetry measures the amount of O2Hgb in the arterial blood
- based on differential absorption of O2Hgb and RHgb
- coupled with ability to separate pulsatile from non-pulsatile
components
- non-invasive and painless
- accurate with newer generation oximeters
- “motion resistant” (SET) technology
- fast (<2 min) and reliable
- inexpensive
- peripheral perfusion index (PPI)
Distribution of O2 saturations in 24h newborns with newer generation pulse oximeters
de-Wahl Grannelli, A: Acta Paediatrica 2005;94:1590
Critical CHDNewborns
Peripheral perfusion index (PPI) for screening for critical left heart obstructive disease (LVOD)
PPI < 0.7 in at least one limb
suggesting of LVOD
OR 23.8 [95% CI (6.4-88.7)]
de-Wahl Grannelli, A: Acta Paediatrica 2007;96:1455
Mahle, W et al.: Role of Pulse Oximetry in Examining Newborns for Congenital Heart DiseaseA Scientific Statement From the AHA and AAP, Circulation 2009; 120:447-458
Studies examining pulse oximetry screening for CCHD
- 13 large studies ( > 200,000 newborns )
- Overall test performance >24h:
- Sensitivity 70% Specificity 99.9%
- False Positive Rate 0.035% False Negative Rate 0.01%
- Positive Predictive Value 47% Negative Predictive Value
99.9%
- overall improved detection rate vs physical exam alone
- improved outcomes?
The Swedish prospective study (2004-2007)de-Wahl Grannelli, A: BMJ 2009; 338
W. Göteland Contemporary cohort p-value
Total 46,693 108,604
Screened 39,821 -
CCHD Prenatal Dx Clinically / PE (only) Pulse Ox Undiagnosed Deaths
622
41 / 51950
1099
72-
280
0.0025*0.16*
Sensitivity 62% N/A
False positives / FPR 69 (0.17%) N/A
Specificity 99.8% N/A
False negatives / FNR 11 (0.03%) N/A
Screening in Minnesota: Region’s Hospital, Saint Paul
• Location: Urban (Twin Cities metropolitan area)• Number of deliveries: 2,500/year• Majority of pregnancies receive prenatal care• No high risk pregnancies
• Start of implementation: 2005
• Implemented as new standard of care – No consent
• Timing: date of discharge (most cases≥ 24h)
• Cut off for normal ≥ 95%
• Location of probe: foot
• Type of probe: reusable
• No consent
• No formal f/u procedure
Newborn screen by pulse oximetry(Data provided by Larry Condon, MD)
Total patients 12,462
Total ECHOs 34 (0.27 %)
Indications
Murmur 22
Syndromes 7
IDM 2
Abnormal screens 3 (0.024%)
CCHD 2 (0.016 %)
False positives 1
False negatives 0
Screening Results
• asymptomatic newborns in well baby nursery ≥ 24h of age
• newer generation pulse oximeters
• motion resistant technology measuring functional O2 saturation
• accuracy (±2% root mean square error)
• both single use and reusable probes acceptable
• two sites: right arm and a foot
• cut-off for positive screen: < 95% and │O2Sats RA- F│> 3 points (3 times)
• any value <90% is abnormal
• cut-off values for areas in high altitude not defined
• screening incorporated with nursery’s practice and other screening activities
Practical considerations
Screening Protocol in well baby nursery for asymptomatic newborns >24h or shortly before discharge
check pulse ox in right arm and foot in room air
Re-screen in 1h
90-94% in both sites or arm-foot difference > 3 points
PASSED TEST
90-94% in both sites or arm - foot difference >3 points
< 90% in either siteIrrespective or difference
PCP notified (FURTHER ACTION)
≥ 95% or higher in either site and arm – foot difference ≤ 3 points
90-94 % in both sites or arm-foot difference > 3 points
Re-screen in 1h POSITIVE SCREEN
POSITIVE SCREEN
DISCHARGE
Response suggestions for primary caretaker to failed pulse oximetry screening
for Critical Congenital Heart Diseases (CCHD)
1. Confirm accuracy of reading
2. If <24h consider deferring discharge and repeat test at 24h or later; otherwise, follow same algorithm as in older infants
3. Perform additional clinical evaluation to assess for non-cardiac causes
4. Echocardiogram (in-center, transfer, telemedicine)
Failed Pulse ox screening: one value equal or less than 90%
OR
three times <95% in both sites or absolute difference of >3 points
Comparative cost of screening
• CCHD: 8,000 cases per year in the US
• Screen will cause additional 8,000 ECHOs per year
• Cost: $9,000 (5.5-29K) per asymptomatic case diagnosed
• Metabolic disorders: 6,400 cases per year in the US
Cost of metabolic screen $110 / infant X 4 million/year = $440M / yr
or $68,750 per patient diagnosed
Expected annual activity for Minnesota
• Minnesota birth rate: 70,000
• Distribution: 60% access to pediatric cardiology services
(50% metro and 10% non-metro cities)
• Expected positive screens: 46
• Expected false positive screens: 25
• Need for transfer: 18
Estimated financial cost• Nursing time: 5-10 min / screen $200K-$400K• Pulse oximeter equipment: $100K• Pulse oximeter supplies: $200K (reusable)
$700K (single use)• Echocardiography cost: $70K-200K• Transport costs: $50K
• Estimated total cost: $620K-$950K • Additional hospital time / hidden costs: ?
* Annual Budget - Minnesota Department of Health (MDH)
$500M / year ($7.5 M/ year on Newborn Screening)
Barriers in implementation• Regulatory (informed consent, reusable probes)
• Cost (equipment, supplies, personnel, transfers, days in hospital)
• Health care personnel: nursing staff
community practitioners
echo technicians
pediatric cardiologists
• Equipment: pulse oximeters / ECHO machines & probes
• Infrastructure & Accessibility
• Medico-legal concern
• Skepticism
• CCHD causes significant morbidity and mortality
• Newborn screening for CCHD with pulse oximetry is promising
• Physical examination still useful
• Early diagnosis of CCHD may improve outcomes
• Guidelines for implementation of screening have been defined
• Future refinements likely
Pulse Ox screening for CCHD is coming soon to
a nursery close to you!
Summary