Symptoms of the problem?
When a cycle is launched the power does not work, the solenoid valve does not click.
When a cycle is started the solenoid valve clicks but the gas does not flow.
The oxygen consumption is too high.
Yes No
Home
Is the red “OUT” indicator light on the timer lit up?
Yes No
Home
Is the green "POWER" light on the timer flashing?
Yes No
Home
Is the timer in "C" mode?
Yes No
Home
Is the supply voltage correct?
(24, 110 or 230V AC depending on the model)
Yes No
Home
Is the equipment connected to the regulator properly?
(sometimes the connection is poor due to oxygenation)
Yes No
Home
Is the bottle open and set to the correct pressure?
Yes No
Home
Remove the pipe from the outlet.
When you begin a cycle, does gas flow freely?
Yes No
Home
Remove the pipe from the diffuser inlet.
When you begin a cycle, does gas flow freely?
Yes No
Home
Remove the valve.
When you begin a cycle, does gas flow freely?
Yes No
Home
Is the valve installed in the circuit?
Check the power supply.
If ok, either the timer or control button is defective. Contact our after-sales service for a replacement.
Did this help to solve your problem?
Yes No
Home
Switch the timer to "C" mode then try again.
Did this help to solve your problem?
Yes No
Home
Check that there is no rust on the plug pins, if not consult an electrician.
Did this help to solve your problem?
Yes No
Home
The timer is defective. Contact our after-sales service for a replacement.
Did this help to solve your problem?
Yes No
Home
The solenoid valve or the solenoid valve coil is defective. Contact our after-sales service for a replacement.
Did this help to solve your problem?
Yes No
Home
Clean or replace (recommended) the couplings and try again.
Did this help to solve your problem?
Yes No
Home
Set the pressure to 1 or 3 bars depending on use
Try again.
Did this help to solve your problem?
Yes No
Home
The solenoid valve or the solenoid valve coil is defective.
Contact our after-sales service for a replacement.
Did this help to solve your problem?
Yes No
Home
A pipe is blocked. Replace it.
Did this help to solve your problem?
Yes No
Home
The valve is defective. Replace it.
Did this help to solve your problem?
Yes No
Home
The ceramic element is clogged. Clean or replace it.
Did this help to solve your problem?
Yes No
Home
Replace the valve and try again.
Did this help to solve your problem?
Yes No
Home
Check that there are no leaks in the circuit.
See attached procedure.
Did this help to solve your problem?
Yes No
Home
Thank you for using Vivelys' online help.We hope that we have met your expectations, and are available to help if you need
any further information.
Please do not hesitate to contact us by email or telephone:
+33 (0)4 67 85 68 [email protected]
Home
Do you need more help?Were you unable to solve the problem using the online troubleshooting guide?
Please contact our technical support team by email or telephone:
+33 (0)4 67 85 68 [email protected]
Home