IAEAInternational Atomic Energy Agency
Slide set of 148 slides based on the chapter authored by
J.C. Lacefield
of the IAEA publication (ISBN 978-92-0-131010-1):
Diagnostic Radiology Physics:
A Handbook for Teachers and Students
Objective:
To familiarize the student with practical issues associated
with ultrasound imaging.
Chapter 13: Ultrasound Imaging
Slide set prepared
by E. Berry (Leeds, UK and
The Open University in
London)
IAEA
CHAPTER 13 TABLE OF CONTENTS
13.1. Introduction
13.2. Array System Principles
13.3. B-Mode Instrumentation and Signal Processing
13.4. Modern Imaging Methods
13.5. Colour Flow Imaging
13.6. Image Artifacts and Quality Assurance
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13. Slide 1 (02/148)
IAEA
13.1 INTRODUCTION13.1
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.1 Slide 1 (03/148)
IAEA
13.1 INTRODUCTION13.1
Introduction – conventional ultrasonography
� Images are acquired in reflection, or pulse-echo, mode
� An array of small piezoelectric elements transmits a
focused pulse along a specified line of sight known as a
scan line
� Echoes returning from the tissue are received by the same
array
• focused via the delay-and-sum beamforming process
• demodulated to obtain the magnitude, or envelope, of the echo
signal
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.1 Slide 2 (04/148)
IAEA
13.1 INTRODUCTION13.1
Introduction – A-mode
� The scanner measures the arrival time of the echoes
relative to the time the pulse was transmitted
• maps arrival time to distance from the array using an assumed
speed of sound
� One-dimensional A-mode (amplitude mode) format
• result of a single-pulse acquisition
• plot echo magnitude as a function of distance
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.1 Slide 3 (05/148)
IAEA
13.1 INTRODUCTION13.1
Introduction – B-mode
� Two-dimensional (2D) or three-dimensional (3D) B-mode
(brightness mode) image
• acquired by performing a large number of pulse-echo acquisitions
• incrementing the scan-line direction between each pulse-echo
operation
• this sweeps out a 2D or 3D field of view
� B-mode imaging because
• the echo magnitude from each point in the field of view is mapped
to the gray level, or Brightness, of the corresponding pixel in the
image
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.1 Slide 4 (06/148)
IAEA
13.2 ARRAY SYSTEM
PRINCIPLES13.2.1 ELECTRONIC FOCUSING AND BEAMSTEERING
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 1 (07/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
Beamsteering
� The array transducers employed by modern ultrasound
systems enable the use of high-speed electronic focusing
and beamsteering methods
• the basis of the high frame rates achieved by ultrasound
� Beamsteering increments the direction of the scan line to
sweep out the B-mode field of view
� Details of the beamsteering process differ slightly
depending which of the three major types of array is used:
• linear array
• curvilinear array
• phased array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 2 (08/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
Ultrasound array and B-mode image geometries.
(a) Linear array, (b) Curvilinear array, (c) Phased array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 3 (09/148)
IAEA
Beamsteering with linear array
(1 of 2)
� A linear array consists of up to
256 elements in a single row
� Each pulse-echo operation is
performed by selecting a small
subaperture of adjacent element
� The scan line is always directed
perpendicular to the array from
the centre of the active
subaperture
• along the axial dimension of the
image
� Scan line is stepped across the
face of the array between each
pulse-echo acquisition
Linear array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 4 (010/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
� Active subaperture: dark grey
� Inactive portion of array: light
grey
IAEA
Beamsteering with linear array
(2 of 2)
� Stepping of scan line achieved
by
• deactivating an element at one
end of the subaperture
• activating a new element at the
opposite end of the subaperture
� This procedure yields a
rectangular field of view whose
lateral width is comparable to the
length of the array
Linear array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 5 (011/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
� Active subaperture: dark grey
� Inactive portion of array: light
grey
IAEA
Beamsteering with curvilinear
array
� Curvilinear array functions in a
manner analogous to a linear
array, except the face of the
array is convex rather than
planar
� As the subaperture is stepped
across the array, the scan line
both translates laterally and
rotates in azimuth angle
� This implementation produces a
circular sector image with a wide
field of view at all depths
Curvilinear array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 6 (12/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
� Active subaperture: dark grey
� Inactive portion of array: light
grey
IAEA
Beamsteering with phased
array (1 of 4)
� Elements of a phased array are
also co-planar but
• smaller, less numerous, and
more closely spaced than the
elements of a linear array
� Each scan line originates from
the centre of the array and is
acquired using all of the
elements
� The azimuth angle of the scan
line, θ, is incremented by altering
the relative timing of the pulses
transmitted by each element
Phased array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 7 (13/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
� Active subaperture: dark grey
� Inactive portion of array: light
grey
� Azimuth angle, θ
IAEA
Beamsteering with phased
array (2 of 4)
� Circular sector field of view
� Spans as much as 90° in
azimuth angle
� Converges to a narrow field of
view at shallow depths, near
the top of an image
Phased array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 8 (14/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
� Active subaperture: dark grey
� Inactive portion of array: light
grey
� Azimuth angle, θ
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
Beamsteering with phased array (3 of 4)
� All array types are laterally focused: beam of a phased
array is steered using delay-and-sum beamformation
� During transmission
• ensure that the pulses transmitted from each element arrive
simultaneously at the focal point
• so result is constructive interference
� Relative firing time of element to firing time at centre of
aperture determined by
• computing the distance from each element to the focal point using
elementary trigonometry
• assume the speed of sound, c = 1540 ms-1 (average in soft tissue)
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 9 (15/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.1 Electronic Focusing and Beamsteering
Beamsteering with phased array (4 of 4)
� During reception
• similar delays are applied to temporally align the echo signals
received by each array element from the intended focal point
• the delayed signals are then summed
• echoes received from the focus add constructively
• this yields a beamformed receive signal
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.1 Slide 10 (16/148)
IAEA
13.2 ARRAY SYSTEM
PRINCIPLES13.2.2 ARRAY BEAM CHARACTERISTICS
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 1 (17/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Modelling Array Beam Characteristics
� The lateral beam patterns produced by a linear or phased
array can be modelled
• apply linear systems analysis to the Fourier transform relationship
between aperture and beam pattern that arises from the
Fraunhofer diffraction integral
� A single element can be viewed as a narrow rectangular
aperture of width d, represented mathematically by
rect(ξ/d)
• where ξ is the lateral dimension within the aperture plane
• the orthogonal dimension (η) is ignored
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 2 (18/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Aperture function
� To model an infinite array with element spacing s.
• convolve rect(ξ/d) with a sampling, or comb, function, comb(ξ/s)
� To produce a linear-systems description of a 1-D array
extending from ξ = −L/2 to ξ = L/2
• multiply by a broad rect function, rect(ξ/L)
� Result is A(ξ), known as the aperture function
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 3 (19/148)
( )
∗
=
LsdA
ξξξξ rectcombrect
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Lateral beam profile in focal plane
� Apply the Fraunhofer diffraction integral
� The lateral beam profile in the focal plane, U(x), is
proportional to the Fourier transform of A(ξ)
where sinc(a) = sin(πa)/(πa) and constant terms are omitted in the
interest of compact notation
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 4 (20/148)
( )
∗
∝
F
dxd
F
sxs
F
LxLxU
λλλsinccombsinc
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Lateral point spread function
� The same aperture function is usually employed for
transmission and reception
� So lateral point spread function PSF(x) is given by square
of the lateral beam profile in the focal plane, U(x)
� Can plot normalized point spread function |PSF(x)|/|PSF(0)|
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 5 (21/148)
( )2
sinccombsinc
∗
∝
F
dxd
F
sxs
F
LxLxPSF
λλλ
IAEA
Primary beam and main lobe
Lateral point-spread function,
PSF(x), of an unapodised linear
array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 6 (22/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
� f-number, F/L=1
� LR: lateral resolution
� PSL: peak side-lobe level
� Decibel scale
� The primary beam arises from
the sinc(Lx/λF) term in PSF
equation
� The largest central peak of
the primary beam is called the
main lobe of the PSF
IAEA
Lateral resolution
Lateral point-spread function,
PSF(x), of an unapodised linear
array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 7 (23/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
� f-number, F/L=1
� LR: lateral resolution
� PSL: peak side-lobe level
� Decibel scale
� Under the Rayleigh resolution
criterion
• lateral resolution of the
imaging system, LR, is equal
to the distance from the
maximum to the first zero of
the main lobe
� Setting sinc(Lx/λF) = 0 and
solving for x
� as previously found for a single-
element rectangular aperture
L
FLR
λ=
IAEA
Contrast resolution Lateral point-spread function,
PSF(x), of an unapodised linear
array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 8 (24/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
� Side lobes: Peaks
flanking the main lobe in
the primary beam
� PSL: peak side lobe level
• expressed in decibels
• indication of the contrast
resolution of the imaging
system
( )( )( )( )
=
mainlobe within max
mainlobe outside maxlog20 10
xPSF
xPSFPSL
IAEA
Contrast resolution Lateral point-spread function,
PSF(x), of an unapodised linear
array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 9 (25/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
� The image contrast expected for
a small anechoic target
� For the PSF shown here
• the largest side lobe arises from
the first off-axis peak of the
primary beam
• PSL = −26.6 dB
( )( )( )( )
=
mainlobe within max
mainlobe outside maxlog20 10
xPSF
xPSFPSL
IAEA
Apodisation of aperture
� Reduces the side lobe
levels
� Weight magnitudes of
the transmit and receive
signals
• on the outer elements
• during delay-and-sum
beamformation
• with a function that
decreases more gradually
than rect(ξ/L)
Lateral point-spread function of
a linear array with f-number
equal to one and Hamming
window apodisation
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 10 (26/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
IAEA
Apodisation of aperture
� E.g. Hamming window is
applied for apodisation
during both transmit and
receive beamforming
• peak side-lobe level is
reduced by about 56 dB
• the lateral resolution is
about twice as coarse as
the resolution of the
uniformly weighted
aperture
No apodisation
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 11 (27/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Hamming window apodisation
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Apodisation of aperture
� Array-based ultrasound systems almost always employ
some form of apodisation because contrast resolution is a
crucial design consideration for diagnostic tasks such as
lesion detection
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 12 (28/148)
IAEA
Grating lobes
Lateral point-spread function,
PSF(x), showing grating lobes
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 13 (29/148)
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
� f-number, F/L=1
� Unapodised
� Decibel scale
� Convolution of the sinc(Lx/λF)
term with the sampling
function produces copies of
the primary beam called
grating lobes
� Spaced by λF/s along the
lateral dimension of the focal
plane
( )2
sinccombsinc
∗
∝
F
dxd
F
sxs
F
LxLxPSF
λλλ
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 14 (30/148)
Grating lobes
� Under the paraxial approximation
� The azimuth angle, θm, of the mth grating lobe away from
the main lobe peak is
� Grating lobes produce echoes from off-axis targets that
appear as artifacts in images
• See section 13.6
≈ −
s
mm
λθ 1sin
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 15 (31/148)
Grating lobes and undersampling
� Mathematically analogous to the copies of spectral peaks
that are observed in the discrete-time frequency spectrum
of an undersampled signal
� In an array system, the aperture function is undersampled
if the element spacing, s, is too large
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 16 (32/148)
Grating lobes: linear array design rule
� For a linear array
• beam is focused but not steered
• grating lobes are avoided if the azimuth angle of the first-order
grating lobe is greater than π/2 radians
� If m = 1, solving for s yields the linear-array design rule for
element spacing: s ≤ λ
2sin 1 πλ
θ ≥
≈ −
s
mm
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 17 (33/148)
Grating lobes: phased array design rule
� For a phased array
• beam is steered
• most challenging scenario for grating lobe suppression arises
when the beam is steered as far as possible in azimuth angle
• i.e. parallel to the face of the array
� It can be shown that s ≤ λ/2� This is the phased-array design rule for element spacing
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 18 (34/148)
Elevation (out-of-plane) dimension of beam
� Analyse in a similar fashion to the lateral dimension
� Conventional linear and phased arrays possess only one
row of elements
• electronic focusing and beamsteering techniques, including
apodisation, cannot be applied in the elevation dimension
� Elevational focusing is performed by an acoustic lens
positioned at the face of the array
� Since the height of an element is substantially smaller than
the lateral aperture length
• spatial resolution in the elevation dimension is relatively coarse (of
the order of several millimetres)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 19 (35/148)
Depth of field in elevation dimension
� Depth of field is the interval over which the beam remains
in focus
� Depth of field is proportional to the wavelength multiplied
by the square of the f-number
� In view of the fixed focal distance of the lens, one benefit
of the weaker focusing in elevation is that the depth of field
is larger in the elevation dimension than the lateral
dimension
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 20 (36/148)
Electronic focusing in elevation dimension
� Electronic focusing in the elevation dimension is possible
using multirow linear arrays
• technology that emerged in the late 1990s
• elements of these arrays are divided into a small number of rows
(typically 5 to 9) in the elevation dimension
• permits electronic focusing, but not beamsteering, to be performed
in the elevation dimension
� Compared to a one-dimensional array with an elevation
lens, multirow arrays enable the depth of the elevational
focus to be changed
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.2 Array Beam Characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.2 Slide 21 (37/148)
Electronic focusing in elevation dimension: pros and cons
� Valuable capability for the multi-focal imaging methods
(section 13.2.3)
� However, use of multirow arrays also increases the system
complexity
• arrays with smaller elements are more difficult to manufacture
• the scanner must include additional channels of transmitter and
receiver electronics
IAEA
13.2 ARRAY SYSTEM
PRINCIPLES13.2.3 MULTI-FOCAL IMAGING METHODS
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 1 (38/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.3 Multi-Focal Imaging Methods
Multi-focal imaging
� The control panel of a modern clinical scanner provides
the user with some flexibility for selecting the number and
depths of transmit foci
� Imaging with multiple transmit focal zones is achieved by
• acquiring each scan line repeatedly with the transmit beam
focused at a different depth along the scan line
• constructing a composite scan line which consists, at each point
along the line, of the pulse-echo sample acquired using the nearest
transmit focus
• if a multirow array is used, the elevation as well as the lateral focus
position can be changed for each pulse echo acquisition
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 2 (39/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.3 Multi-Focal Imaging Methods
Multi-focal imaging: Pros and cons
� Imaging with multiple transmit zones produces an image
with more consistent resolution throughout the field of view
� But this improvement comes at the expense of a reduced
frame rate
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 3 (40/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.3 Multi-Focal Imaging Methods
Dynamic receive focusing (1 of 2)
� All modern scanners also employ dynamic receive
focusing
� The arrival time of echoes at the receive aperture
corresponds to the depth of the scatterers that produced
the echoes
� So the receive focusing delays are updated in real time
such that the receive focus tracks the pulse as it
propagates along each scan line
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 4 (41/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.3 Multi-Focal Imaging Methods
Dynamic receive focusing (2 of 2)
� The receiver delays can be updated continuously as a
scan line is acquired
� So dynamic receive focusing means that the lateral width
of the receive beam (and its elevation width if a multirow
array is used) should be minimised everywhere in the field
of view
� Use of dynamic receive focusing has no effect on frame
rate since the receive focus is produced computationally
after a pulse-echo signal has been acquired
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 5 (42/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.3 Multi-Focal Imaging Methods
Frame rate
� The frame rate of an ultrasound system is determined by
• the lateral resolution
• the field of view
• the number of transmit focal zones
� In general, reduce the frame rate by imaging
• over a larger field of view
• with higher spatial resolution
• with a greater number of transmit focal zones
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 6 (43/148)
IAEA
13.2 ARRAY SYSTEM PRINCIPLES13.2.3 Multi-Focal Imaging Methods
Frame rate dependencies
� The total time to acquire one scan line can be
approximated by the round-trip pulse travel time to and
from the maximum depth in the image multiplied by the
number of transmit focal zones
� The lateral spacing of adjacent scan lines should be no
greater than one half of the lateral resolution at the focus
to ensure adequate spatial sampling
� The number of scan lines in one frame therefore depends
on the lateral field of view and the lateral resolution
� The time to acquire one frame is the product of the number
of scan lines and the total time to acquire one line Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.2.3 Slide 7 (44/148)
IAEA
13.3 B-MODE INSTRUMENTATION
AND SIGNAL PROCESSING13.3
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 1 (45/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Block diagram of a B-mode ultrasound imaging system
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 2 (46/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
B-mode ultrasound imaging (1 of 3)
� High-voltage excitation pulses are applied to the array by
the transmitter electronics to fire a pulse
� The transmit/receive switch then disconnects the array
from the transmitter electronics and connects it to the
receiver electronics, thereby isolating the receiver
electronics from the high-voltage excitation pulses
� The returning echo signals are immediately amplified,
digitized, and then combined via delay-and-sum
beamforming to produce a beamformed radio-frequency
(RF) signal
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 3 (47/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
B-mode ultrasound imaging (2 of 3)
� Time-gain compensation is applied to the beamformed RF
signal to compensate for attenuation of echoes arriving
from deeper targets
� Envelope detection is performed to obtain the magnitude
signal
� Signal is logarithmically compressed to maximize the
dynamic range of the image
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 4 (48/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
B-mode ultrasound imaging (3 of 3)
� Scan conversion is performed to map
• each sample of the magnitude signal to its 2D or 3D position in the
image
• the log-scaled magnitude values to gray levels
� Once the scan lines covering the entire field of view have
been acquired, the resulting image frame is displayed on
the scanner’s video monitor
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 5 (49/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Transmitter electronics
� Design requirements
• High bandwidth to enable use of short excitation pulses to obtain
high axial resolution
• An ability to operate at high power to drive the cable connecting
the scanner and transducer with 50−100 V excitation signals
� Some modern imaging methods employ coded transmit
pulses (section 13.4.3)
• require more sophisticated transmit electronics to support the use
of programmable excitation waveforms
• in conventional B-mode imaging a simple excitation waveform is
used and the spectral characteristics of the transmitted pulse are
determined primarily by the frequency response of the transducer
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 6 (50/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Receiver electronics
� Design requirements for digital processing
• sampling at sufficiently high frequency to sample the highest
frequency components of the RF signal
• e.g. 25-50 MHz depending on the pulse spectrum
• quantization using at least 12, and preferably 16, bits to represent
the RF waveform accurately
� Analogue-to-digital conversion is preceded by an anti-
aliasing low-pass filter
• the cut-off frequency of the anti-aliasing filter is often a significant
determinant of the ultimate axial resolution of the image
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 7 (51/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Beamforming
� Beamforming delays must be applied with high precision
to obtain a sharp focus
� The typical procedure is to interpolate the digitized RF
signals to a sampling frequency as high as 1 GHz
• which would enable focusing delays to be applied with 1 ns
precision
� Then downsample the time-shifted signals to the original
sampling frequency for subsequent processing
� Relatively simple interpolation methods must be used to
enable this step to be performed in real time for dynamic
receive focusing
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 8 (52/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Amplification
� Three types of amplification are applied to the
beamformed RF signal
� Constant gain
� Time-gain compensation (TGC)
� Lateral gain compensation
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 9 (53/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Constant gain
� A constant gain is applied to enable the user to adjust the
overall brightness of the image
� Often simply labelled “Gain” on the scanner’s control panel
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 10 (54/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Time-gain compensation (TGC)
� Time-dependent amplification applied to each scan line to
offset the effects of attenuation
� The reduction in signal intensity from attenuation decays
as an exponential function of propagation distance
• so the TGC should be an approximately exponential function of the
echo arrival time (hence the T in TGC)
� TGC slope is set separately in several depth bands
covering the field of view by adjusting the TGC to obtain
visually consistent brightness throughout the image
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 11 (55/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Lateral gain compensation
� Lateral gain compensation is applied to compensate for
shadowing or enhancement artifacts (see section 13.6)
that cause image brightness at a given depth to vary as a
function of lateral position
� The lateral gain is adjusted manually by the user in a
manner similar to the TGC
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 12 (56/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Computation of magnitude signal
� Modern scanners typically compute the magnitude signal
using the Hilbert transform
• applies an exact π/2 phase shift to the RF signal
• thereby estimating its quadrature component
� The instantaneous magnitude at each time sample is then
obtained by adding the original and Hilbert-transformed
signals in quadrature
� Prior to widespread use of digital processing
• envelope detection via analogue amplitude demodulation
• e.g. full-wave rectification followed by low-pass filtering of the
beamformed RF signal
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 13 (57/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Dynamic range of magnitude signal
� The magnitude signal is logarithmically compressed
� The dynamic range of the pulse-echo data can be greater
than 80 dB
• much greater than the 48 dB (= 20log10(256)) dynamic range of a
standard 256-gray-level display
� The scanner’s electronics could map the data to greater
than 256 gray levels
• little benefit to this approach as the human visual system has
limited sensitivity to subtle differences in brightness
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 14 (58/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Compression of magnitude signal
� An effective algorithm for logarithmic compression is to
• use the mean of the magnitude signal (averaged over the entire
field of view) as the reference magnitude
• convert each of the linearly scaled magnitude samples to decibels
with respect the reference value
� The dB values are linearly converted to gray levels such
that
• a magnitude of –X/2 dB maps to gray level 0 (black)
• 0 dB (i.e., the mean magnitude) maps to gray level 128
• X/2 dB maps to gray level 255 (white)
� X is the displayed dynamic range, typically 60 or 80 dB
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 15 (59/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Interpolation – Linear array
� For the rectangular images produced by a linear array
system
• spatial sampling of the original echo data is much coarser in the
lateral dimension compared to the axial dimension due to the
anisotropic spatial resolution of an ultrasound scanner
• if the scan lines are displayed directly, the rectangular aspect ratio
of the pixels produces visible banding artifacts in the images
� Images are laterally interpolated before they are displayed
to obtain a uniform pixel width in all dimensions
• a relatively simple one-dimensional interpolation, e.g. a cubic
spline, yields a large improvement in the appearance of the image
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 16 (60/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Interpolation – Phased array (1 of 2)
� Sector images are produced by a phased array system
• scan lines acquired in a polar format must be displayed on a
rectangular pixel grid
� Adjacent scan lines will pass through the same pixel near
the origin of the sector
• a single gray level can be assigned by averaging the magnitude
signals or use the maximum of the overlapping magnitude samples
� Adjacent scan lines may be separated by greater than the
pixel width near the base of the sector
• gaps between scan lines at deeper ranges are typically filled by
two-dimensional interpolation within the image plane
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 17 (61/148)
IAEA
13.3 B-MODE INSTRUMENTATION AND SIGNAL
PROCESSING 13.3
Interpolation – Phased array (2 of 2)
� The interpolation process also compensates for the fact
that locations of most of the samples in the original echo
data do not correspond exactly to centre of any pixel in the
rectangular display grid
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.3 Slide 18 (62/148)
IAEA
13.4 MODERN IMAGING
METHODS13.4
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4 Slide 1 (63/148)
IAEA
13.4 MODERN IMAGING METHODS13.4
Modern Imaging Methods
� 13.4.1 Contrast-Enhanced Imaging
� 13.4.2 Tissue Harmonic Imaging
� 13.4.3 Coded Excitation Imaging
� 13.4.4 Three- and Four-Dimensional Imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4 Slide 2 (64/148)
IAEA
13.4 MODERN IMAGING
METHODS13.4.1 CONTRAST-ENHANCED IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 1 (65/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Contrast-Enhanced Imaging
� Microbubbles are used as blood-pool contrast agents in
diagnostic ultrasound
• gas-filled
• encapsulated
• sizes typically ranging from 1 to 4 mm diameter
� A microbubble scatters ultrasound strongly despite its
small size
• because of the over three orders of magnitude difference between
the acoustic impedances of a gas and soft tissue
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 2 (66/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Size of microbubbles
� 1 -4 mm diameter microbubbles are smaller than red blood
cells
• so they are not trapped in capillary beds but are still too large to
extravasate
• their fundamental resonant frequency is in the 2-4 MHz range,
which is convenient for ultrasound imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 33 (67/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Encapsulation of microbubbles
� Microbubbles are encapsulated by a shell to prevent the
gas from dissolving into the blood
• the original “first-generation” microbubble formulation consisted of
air bubbles encapsulated in a relatively stiff shell material such as
albumin
• most contrast-enhanced exams are now performed with “second-
generation” microbubbles consisting of an inert perfluorocarbon
gas core, such as perfluoropropane (C3F8), encapsulated in a
phospholipid shell
� Circulating half-life of second-generation agents increased
• perfluorocarbon gas is insoluble in blood
• more deformable shell
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 4 (68/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Harmonics and microbubbles
� The key to forming contrast-enhanced ultrasound images
is to differentiate echoes produced by microbubbles from
echoes produced by tissue
• exploit the nonlinear scattering characteristics of microbubbles
� A microbubble’s spherical symmetry yields strong
scattering resonances
• at a fundamental frequency determined by the bubble’s radius and
shell properties
• resonances at harmonics (integer multiples) and subharmonics
(integer fractions) of the fundamental frequency
� So, seek echoes with harmonic spectral characteristics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 5 (69/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Non-linear ultrasound propagation
� During the initial development stages of contrast-
enhanced ultrasound, it was assumed that echo signals
containing harmonic spectral characteristics would
uniquely identify echoes from microbubbles
� Soft tissue was unexpectedly discovered to possess
nonlinear ultrasonic properties of its own
� This led to the development of tissue harmonic imaging
(section 13.4.2) and also motivated further research into
contrast-agent detection techniques
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 6 (70/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Pulse Inversion (1 of 2)
� Contrast-enhanced imaging may use a variation on a
multi-pulse imaging method such as pulse inversion
� Contrast-enhanced images produced in this fashion reveal
abnormalities in vascular function via regional differences
in the timing and spatial patterns of contrast enhancement
� Each scan line is acquired twice in close succession
• first by transmitting a standard pulse
• then by transmitting an inverted pulse (e.g. the original pulse
multiplied by –1)
� Resulting echo signals are added
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 7 (71/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Pulse Inversion (2 of 2)
� Tissue component of the second echo signal is
approximately an inverted copy of the first echo signal
• the echoes from tissue cancel when the two received signals are
summed
� Harmonic components of the microbubble echoes are not
inverted in the second echo signal
� The summation step
• cancels the fundamental frequency components of the microbubble
echoes
• maintains the harmonic components
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 8 (72/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Destruction-replenishment imaging (1 of 2)
� Another important approach to contrast-enhanced imaging
� A sequence of high-mechanical-index (MI) transmit pulses
is used to fragment all microbubbles in the field of view
� Contrast enhancement kinetics are measured as new
microbubbles flow into the region of interest
• contrast replenishment can be imaged using a low MI technique,
such as pulse inversion, that does not destroy the agents
� Biophysical models used to estimate parameters such as
• blood volume
• transit time through the region of interest
• perfusion
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 9 (73/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.1 Contrast-Enhanced Imaging
Destruction-replenishment imaging (2 of 2)
� Destruction-replenishment imaging is arguably the
contrast-enhanced ultrasound method with the greatest
biomedical value because the functional parameters
estimated with this technique are most similar to the
perfusion parameters measured using modalities such as
dynamic contrast-enhanced CT or MRI
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.1 Slide 10 (74/148)
IAEA
13.4 MODERN IMAGING
METHODS13.4.2 TISSUE HARMONIC IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.2 Slide 1 (75/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.2 Tissue Harmonic Imaging
Tissue harmonic imaging or native tissue harmonic imaging
� Harmonic spectral components are generated as an
ultrasound pulse propagates through tissue
• from modulation of the sound speed of the tissue by pressure wave
� Substantial harmonics are only produced at the transmit
focus of a diagnostic imaging system, where the pulse
intensity is greatest
� The received signal can be bandpass filtered to construct
an image showing only the harmonic component of the
echoes
• need sufficiently high bandwidth transducer to detect echoes at the
second harmonic of the transmit frequency
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.2 Slide 2 (76/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.2 Tissue Harmonic Imaging
Tissue harmonic imaging – resolution and penetration
� Partially offsets the trade-off between spatial resolution
and penetration depth encountered in conventional B-
mode imaging
• lower frequency transmit pulse experiences less severe frequency-
dependent attenuation than the returning second-harmonic echoes
• the lateral width of the receiver focus is narrower than the transmit
beam.
� The resulting image exhibits lateral resolution and
penetration intermediate between those observed in
• conventional imaging at the fundamental frequency and
• conventional imaging at twice that frequency
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.2 Slide 3 (77/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.2 Tissue Harmonic Imaging
Tissue harmonic imaging – pros and cons
� Tissue harmonic imaging has also been shown
experimentally to reduce clutter artifacts compared to
conventional B-mode imaging
� Benefits are achieved at the expense of reduced axial
resolution
• the passband of the transducer’s frequency response must be
divided into fundamental and harmonic segments rather than using
the entire bandwidth to generate the shortest possible transmit
pulse
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.2 Slide 4 (78/148)
IAEA
13.4 MODERN IMAGING
METHODS13.4.3 CODED EXCITATION IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.3 Slide 1 (79/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.3 Coded Excitation Imaging
Coded Excitation Imaging
� Technique was developed to increase the penetration
depth of ultrasound imaging systems
� Also enables high-frame-rate (e.g., several hundred
frames per second) imaging
� Transmit a relatively long duration signal such as
• a chirp (a sinusoid with increasing or decreasing instantaneous
frequency)
• a pulse modulated code, in which a sinusoid is switched on and off
in a specific temporal sequence to create a binary code
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.3 Slide 2 (80/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.3 Coded Excitation Imaging
Coded Excitation Imaging and penetration depth
� Use of a pulse code spreads the transmitted energy over
the pulse duration
� The signal-to-noise ratio and hence the penetration depth
can be increased without exceeding patient exposure
regulatory requirements
� Matched filtering techniques are used during reception to
deconvolve the transmitted code from the echo signal
� Axial resolution is close to that achieved by conventional
B-mode imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.3 Slide 3 (81/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.3 Coded Excitation Imaging
Coded Excitation Imaging – high-frame-rate
� Simultaneously transmit multiple orthogonal pulse codes
steered to different azimuth angles
� The echoes produced by each of these transmit beams will
interfere at the receive aperture
� Matched filtering by the receiver discriminates among the
echoes produced by each pulse code, enabling multiple
scan lines to be acquired simultaneously
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.3 Slide 4 (82/148)
IAEA
13.4 MODERN IMAGING
METHODS13.4.4 THREE- AND FOUR-DIMENSIONAL IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 1 (83/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Three- and Four-Dimensional Imaging
� Advances in ultrasound technology have enabled
conventional 2D B-mode imaging to be supplemented, and
in some applications supplanted, by
• 3D (or volumetric) imaging
• 4D imaging, in which a temporal sequence of 3D images is
presented as a ciné loop
� There are three primary methods of acquiring three-
dimensional images that present differing combinations of
advantages, disadvantages, and technical complexity
• mechanical scanning, freehand scanning and real-time imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 2 (84/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Mechanical scanning
� Mechanically translate or rotate a linear or curvilinear array
transducer with the motion directed out of the 2D image
plane
� A sequence of 2D images is acquired at regular linear or
rotational increments
� A 3D image volume can readily be reconstructed from the
set of 2D images because the spatial relationship among
the images is known in advance
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 3 (85/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Mechanical scanning – pros and cons
� Relatively uncomplicated to implement
� For stationary structures, provides the highest image
quality
• large-aperture or higher frequency linear arrays can be used
• the transducer motion is not operator dependent
� One 3D image volume can require several seconds to
acquire
• the translation or rotation speed of the transducer should be slow
compared to the 2-D frame rate
• extensive respiratory and cardiac gating required if it is used to
image dynamic structures such as the beating heart
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 4 (86/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Freehand scanning
� A conventional linear or curvilinear array is translated or
rotated manually by the sonographer
� Transducer motion
• measured by an external electromagnetic or optical tracking
system that is synchronized with the 2D image acquisition, and the
transducer position measurements are used to position each 2D
image plane within the reconstructed 3D volume, or
• estimated directly from the image sequence using cross-correlation
to estimate motion within the 2D plane and using the rate of
decorrelation of consecutive frames to estimate out-of-plane
motion
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 5 (87/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Freehand scanning – pros and cons
� The quality of the reconstructed image will depend on the
ability of the sonographer to sweep the transducer in a
regular, smooth pattern so that the 3D image volume is
uniformly filled in by the constituent 2D images
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 6 (88/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Real-time imaging
� The most technologically sophisticated method of 3D
imaging
� Two approaches
• 2D phased array or
• 2D array of slightly larger elements
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 7 (89/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Real-time imaging with 2D phased array
� Phased array consists of a matrix of small (ideally < λ/2 on
each side) square elements
� This enables phased array beamsteering in both the
azimuth and elevation angles
� The resulting image volume is a 3D pyramidal sector that
can be considered an extension of the circular sector
produced by a conventional 1D phased array
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 8 (90/148)
IAEA
13.4 MODERN IMAGING METHODS13.4.4 Three- and Four-Dimensional Imaging
Real-time imaging with 2D array
� 2D array of slightly larger elements than in phased array
� Acquire images in a manner comparable to a conventional
linear array system
� Each scan line is acquired using a 2D subaperture of
active elements that are focused straight ahead
� The subaperture is stepped in both dimensions across the
face of the array to build up a 3D rectilinear image volume
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.4.4 Slide 9 (91/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5 Slide 1 (92/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5
Colour Flow Imaging
� Doppler-based methods for blood-flow imaging can be
viewed as extensions of the pulsed-wave Doppler method
in which Doppler processing is applied to a large number
of sample volumes to produce 2D or 3D images of blood
flow
� 13.5.1 Flow Imaging Modes
• Colour Doppler
• Power Doppler
• Duplex / Triplex Doppler
� 13.5.2 Tissue Doppler Imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5 Slide 2 (93/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 FLOW IMAGING MODES
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 1 (94/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Colour Doppler display
� One of two principal formats for displaying the resulting
flow images
� Colour Doppler uses a red and blue colour scale to
represent the mean axial velocity in each sample volume
• flow towards the transducer shown in red
• flow away from the transducer shown in blue
• the velocity magnitude mapped to the colour intensity
� The colour pixels are superimposed on a B-mode image
such that the echo magnitude from tissue volumes
containing no detectable flow is displayed in gray scale
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 2 (95/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Colour Doppler
� Used when imaging the heart and major blood vessels
• in applications for which mean flow velocity is a diagnostically
useful parameter
� Acquisition of the B-mode image must be interleaved with
acquisition of the Doppler data
• so the colour Doppler frame rate is always less than the frame rate
of conventional B-mode imaging
• and the velocity estimates in a colour Doppler image are prone to
aliasing because the Doppler pulse repetition frequency is limited
by the need to also acquire B-mode scan lines
• the spectral wrap-around produced by aliasing appears as sudden
changes in pixel colour, and thus the apparent direction of flow
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 3 (96/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Power Doppler display
� The second of the two principal formats for displaying the
resulting flow images
� Power Doppler uses a red-to-orange-to-yellow colour
scale to represent the total power in the Doppler spectrum
at each sample volume
• the lowest powers are displayed in red
• the highest powers are displayed in yellow
� As in colour Doppler, the B-mode signal is displayed in
gray scale for tissue volumes with no detectable flow
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 4 (97/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Power Doppler
� Doppler power is in theory proportional to the
concentration of moving blood cells in the sample volume
� Power Doppler is typically used for applications where
blood volume is a diagnostically useful parameter
• such as tumour imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 5 (98/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Power Doppler – pros and cons
� Power Doppler images contain no information about flow
direction or velocity, but
� Doppler power does not depend on the Doppler angle
• so power Doppler provides a more continuous display of tortuous
vessels than colour Doppler
� Aliasing does not affect the total power in the Doppler
spectrum
• so power Doppler images are not susceptible to aliasing artifacts
� Integrated power is less affected by low SNR than is the
mean Doppler frequency
• so power Doppler better for imaging small, slow-flow vessels
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 6 (99/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Triplex Doppler
� Colour Doppler and pulsed-wave Doppler are sometimes
combined in a mode known as triplex Doppler
• the B-mode information included in the colour Doppler image
represents the third component of the triplex
� In triplex Doppler, acquisition of a pulsed-wave Doppler
spectrum from a user selected sample volume is
interleaved with acquisition of a colour Doppler image
� The colour flow image and the Doppler spectrum are
displayed side-by-side
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 7 (100/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.1 Flow Imaging Modes
Triplex Doppler – the advantage
� This mode provides both
• the 2D spatial information of a colour flow image and,
• for the pulsed-wave sample volume, the higher maximum velocity
and velocity resolution of a conventional pulsed Doppler exam
� Triplex Doppler can be considered an extension of duplex
Doppler, a scanning mode that predates colour flow
imaging in which a B-mode image and a pulsed Doppler
spectrum are acquired and displayed simultaneously
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.1 Slide 8 (101/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.2 TISSUE DOPPLER IMAGING
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.2 Slide 1 (102/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.2 Tissue Doppler Imaging
Tissue Doppler Imaging – low-velocity motion
� Conventional Doppler systems discriminate blood flow
from soft tissue motion based on velocity
• including continuous-wave Doppler, pulsed-wave Doppler and
colour flow imaging systems
• a Doppler system assumes blood flow is concentrated at
intermediate and high velocities while scatterers moving at low
velocities correspond to soft tissue
• discrimination is achieved by applying a high-pass wall filter, also
known as a clutter filter, to eliminate low-Doppler-frequency
components from the Doppler signal
� In Tissue Doppler, the wall filter is replaced with a low-
pass filter so only low-velocity motion is displayed
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.2 Slide 2 (103/148)
IAEA
13.5 COLOUR FLOW IMAGING 13.5.2 Tissue Doppler Imaging
Tissue Doppler Imaging
� Useful in applications that require measurements of soft
tissue motion
• diagnosis of regional abnormalities in ventricular wall motion, e.g.
following a myocardial infarction or in heart failure patients
� Tissue Doppler can be performed at a single sample
volume in a manner analogous to pulsed-wave Doppler
• the Doppler spectrogram attributed to tissue motion is displayed
� Or, tissue Doppler measurements can be performed over
a 2D region of interest in a manner analogous to colour
Doppler
• to produce an image of mean Doppler frequency or axial velocity
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.5.2 Slide 3 (104/148)
IAEA
13.6 IMAGE ARTIFACTS AND
QUALITY ASSURANCE 13.6
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6 Slide 1 (105/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6
IMAGE ARTIFACTS AND QUALITY ASSURANCE
� 13.6.1 B-mode Image Artifacts
� 13.6.2 Speckle
� 13.6.3 Quality Assurance Phantoms and Methods
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6 Slide 2 (106/148)
IAEA
13.6 IMAGE ARTIFACTS AND
QUALITY ASSURANCE 13.6.1 B-MODE IMAGE ARTIFACTS
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 1 (107/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6
IMAGE ARTIFACTS AND QUALITY ASSURANCE
� 13.6.1 B-mode Image Artifacts
� 13.6.2 Speckle
� 13.6.3 Quality Assurance Phantoms and Methods
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 2 (108/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
B-Mode Image Artifacts
Many common B-mode image artifacts can be understood in
terms of fundamental physical concepts
� Reverberation artifact (comet-tail artifact)
� Refraction artifact
� Mirror image artifact
� Shadowing and enhancement artifacts
� Sound-speed artifact
� Side-lobe artifact
� Grating-lobe artifact
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 3 (109/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Reverberation artifact
� Transmitted pulse bounces back-and-forth between two
strongly reflecting interfaces
• each time the reverberating pulse returns to the top interface, a
portion of the acoustic intensity is transmitted through the interface
and continues to the transducer
• these pulses appear in the image as copies of the deeper
boundary separated by a distance equal to the thickness of the
object causing the reverberation
� The image intensity of reverberations decays as a function
of depth
• due to the cumulative multiplication of the pulse intensity by the
reflection and transmission coefficients at the object’s boundaries
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 4 (110/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Reverberation artifact in practice
� Reduction of image intensity with depth means that the
artifact usually obscures only a limited region below the
reverberating object
� Reverberation artifacts are often produced by large
calcifications or metallic foreign bodies
� Sonographers describe the sequence of bright echoes
below a reverberating object as a comet-tail artifact and
use the artifact as an aid to indentifying such hard
inclusions
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 5 (111/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Refraction artifact
� Produced by transmission through specular interfaces
• refraction occurs when a pulse is obliquely incident on a boundary
between two tissues with differing sound speed
• the change in propagation direction of the pulse at the boundary
deflects the pulse away from the intended direction of the scan line
• if the refracted pulse subsequently encounters a strong reflector,
the resulting echo can be refracted back toward the transducer at
the overlying specular boundary
� The refracted echo will be displayed at a point along the
intended direction of the scan line
• the reflecting object will be displayed at an incorrect lateral position
in the image
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 6 (112/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Mirror image artifact
� Produced at a very strongly reflecting (i.e., R→ 1)
specular interface
� Consider a transmit pulse that is redirected by a specular
reflector as described by Snell’s law
• echoes from the redirected pulse are scattered back to the
specular reflector, which reflects them back to the transducer
� Those echoes will be displayed
• along the direction of the original scan line
• behind the specular reflector due to their longer round-trip travel
time
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 7 (113/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Mirror image artifact in practice
� The most striking mirror image artifacts are produced by
the diaphragm when imaging the liver with the scan plane
oriented cranially
� The redirected echoes will cause features to appear in the
image above the diaphragm in the lungs
• at locations from which no signal would be expected due to the
difficulty of coupling ultrasound into an air-filled organ
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 8 (114/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Shadowing and enhancement artifacts (1 of 2)
� Produced by localized variations in the attenuation
coefficient
� If the transmitted pulse traverses a feature that attenuates
more strongly than the surrounding tissue
• pulses incident on features located behind that feature will possess
a lower intensity than expected
• the tissue behind the strongly attenuating feature will thus appear
darker than the laterally adjacent tissue in the B-mode image
• this is shadowing
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 9 (115/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Shadowing and enhancement artifacts (2 of 2)
� Produced by localized variations in the attenuation
coefficient
� If the transmitted pulse traverses a feature that attenuates
less than the surrounding tissue
• pulses incident on features located behind that feature will possess
a higher intensity than expected
• the tissue behind the weakly attenuating feature will thus appear
brighter than expected in the B-mode image
• this is enhancement
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 10 (116/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Shadowing and enhancement artifacts in practice
� Neither time-gain compensation nor lateral gain
compensation is effective for eliminating shadowing or
enhancement artifacts
• time-gain compensation curve is the same along all scan lines
• lateral gain compensation is the same at all depths in the image
� In cancer imaging applications, the relative attenuation in a
lesion can be correlated with whether the lesion is benign
or malignant, so sonographers sometimes consider
shadowing or enhancement artifacts to be diagnostically
useful observations
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 11 (117/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Sound-speed artifacts
� Localized variations in the speed of sound can cause
reflectors to be displayed at incorrect depths in the image
� A region with elevated sound speed
• will cause echoes backscattered from behind that inclusion to
arrive at the receiver sooner than echoes from the same depth
along other scan lines
• scatterers behind the high-sound-speed region will appear to be
closer to the transducer than their true position
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 12 (118/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Sound-speed artifacts in practice
� The sound speeds of most soft tissues are clustered near
1540 m/s
• geometric distortion of the image due to sound speed variations is
rarely perceptible
� Sound-speed artifacts are a cause for concern when
ultrasound is employed in image-guided interventional
procedures
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 13 (119/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Side-lobe artifacts
� For a small anechoic feature such as a fluid-filled cyst,
• when acquiring scan lines passing through the cyst, the side lobes
of the beam may extend outside the cyst into surrounding tissue
• a portion of the intensity contained in the side lobes can be
backscattered toward the transducer
• echoes received as a result of scattering from the side lobes will be
displayed along the beam axis of the scan line, i.e., within the cyst
� The cyst thus appears to be a weakly scattering lesion
rather than an anechoic lesion
� Its image contrast is lower than it would be in the absence
of the side-lobe artifact
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 14 (120/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Side-lobe artifacts and contrast resolution
� Side lobes reduce image contrast
� This reasoning is the basis for interpreting the peak side-
lobe level
as an approximate measure of the contrast resolution of
an ultrasound system
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 15 (121/148)
( )( )( )( )
=
mainlobe within max
mainlobe outside maxlog20 10
xPSF
xPSFPSL
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Grating-lobe artifacts (1 of 2)
� Grating lobes, if present, can produce B-mode artifacts in
a manner analogous to a side-lobe artifact
• the main lobe of the beam is propagating through tissue with
moderate scattering strength
• while a grating lobe is incident upon a strongly reflecting feature
� The echo produced by the grating lobe will be displayed
along the axis of the scan line
� Most array systems are designed to prevent the formation
of grating lobes, so grating-lobe artifacts are relatively
uncommon
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 16 (122/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.1 B-Mode Image Artifacts
Grating-lobe artifacts (2 of 2)
� The grating lobe pattern is usually symmetric about the
main lobe axis, so grating-lobe artifacts tend to be visually
symmetric as well
• the strong reflector will be displayed brightly in the image at its
correct position
• less intense copies of the reflector will be displayed at equal lateral
distances to the left and right of the reflector
• the artifactual depictions of the reflector occur on scan lines in
which the m = 1 and m = −1 grating lobes are directed toward the
reflector.
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.1 Slide 17 (123/148)
≈ −
s
mm
λθ 1sin
IAEA
13.6 IMAGE ARTIFACTS AND
QUALITY ASSURANCE 13.6.2 SPECKLE
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 1 (124/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6
IMAGE ARTIFACTS AND QUALITY ASSURANCE
� 13.6.1 B-mode Image Artifacts
� 13.6.2 Speckle
� 13.6.3 Quality Assurance Phantoms and Methods
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 2 (125/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Speckle
� The granular or mottled texture observed in B-mode
images
� Sometimes considered another type of artifact, but speckle
is also a fundamental characteristic of a B-mode image
and therefore merits special consideration
� The formation of speckle can be understood using a
random-walk model that was originally developed in optics
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 3 (126/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Random-walk model (1 of 4)
� At any instant in time, the ultrasound system receives
echoes from multiple unresolved tissue structures located
in a resolution volume defined by the three-dimensional
point-spread function
� In the narrowband limit, each of these echoes can be
represented by a phasor with a distinct magnitude and
phase
� The instantaneous value of the received RF signal is equal
to the coherent sum of those phasors
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 4 (127/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Random-walk model (2 of 4)
� If all of the scatterers are similar structures
• the magnitude of each phasor can be modelled as a Gaussian
random variable
� If the scatterers are positioned randomly throughout the
resolution volume
• the phase of the individual echoes can be modelled as a uniformly
distributed random variable from –π to π
� The coherent summation can be visualized graphically as
a random walk in the complex plane
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 5 (128/148)
IAEA
Random-walk model (3 of 4)
� The coherent summation can
be visualized graphically as a
random walk in the complex
plane
� The thin black phasors
represent echoes from
individual scatterers
� The thick gray phasor is the
coherent sum, s, of those
echoes.
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 6 (129/148)
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
IAEA
Random-walk model (4 of 4)
� If there are at least 10
scatterers in the resolution
volume, the magnitude of the
phasor sum, which
corresponds to the envelope-
detected echo signal, follows
the Rayleigh probability
density function
� The Rayleigh scale parameter
equals one in this example,
so the mean value of |s| is
Rayleigh probability density
function for the magnitude of a
coherently summed speckle
signal, |s|
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 7 (130/148)
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
25.12 ≈π
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Speckle treated as noise
� Speckle in a B-mode image is thus a random process that
is sometimes compared to noise
� If all of the forgoing assumptions are satisfied
• the ratio of the mean to the standard deviation of the envelope-
detected signal, which is termed the point signal-to-noise ratio, is a
constant ≈ 1.91
• the histogram of the signal provides little information about the
tissue beyond its mean scattering strength
• the size of the individual speckle grains, i.e., the spatial
autocorrelation length of the speckle pattern, is entirely determined
by the point-spread function and so also carries little information
about the tissue
• . Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 8 (131/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Speckle treated as noise: speckle reduction
� In applications such as lesion detection, the gray-level
fluctuations due to speckle can obscure low-contrast
lesions
� Considerable effort has been devoted to developing
speckle reduction methods
� Spatial compounding, in which several images of a region
of interest are acquired from different angles and
averaged, is the most successful speckle reduction
method and is now implemented under various trade
names on most modern scanners
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 9 (132/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Speckle treated as signal (1 of 2)
� The comparison of speckle to noise can be misleading
� Speckle is the coherent sum of all of the echoes scattered
from the interior of a tissue structure, so the speckle is the
signal in most of the image
� If the transducer and the tissue are both stationary, the
speckle pattern, in contrast to any electronic noise in the
image, is constant
� Importantly, the conditions concerning the scatterers that
allow it to be treated as noise are met only in simulated
images and phantoms
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 10 (133/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Speckle treated as signal (2 of 2)
� All tissues possess some degree of spatial organization to
be able to perform their biological functions
• the condition of randomly positioned scatterers is strictly met only
in simulated images and tissue-mimicking phantoms
� There may also be
• fewer than scatterers per resolution volume
• two or more populations of scatterers in real tissue
� The first- and second-order statistics and spectral
characteristics of echoes acquired from tissue carry more
information about the tissue than the random-walk model
suggests
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 11 (134/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.2 Speckle
Speckle for tissue characterisation
� The observations concerning tissue characteristics are the
motivation behind ongoing efforts to develop quantitative
tissue characterization methods that employ ultrasound
imaging
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.2 Slide 12 (135/148)
IAEA
13.6 IMAGE ARTIFACTS AND
QUALITY ASSURANCE 13.6.3 QUALITY ASSURANCE PHANTOMS AND
METHODS
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 1 (136/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6
IMAGE ARTIFACTS AND QUALITY ASSURANCE
� 13.6.1 B-mode Image Artifacts
� 13.6.2 Speckle
� 13.6.3 Quality Assurance Phantoms and Methods
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 2 (137/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Quality Assurance Phantoms and Methods
� Tissue-mimicking phantoms
� Spatial resolution phantoms
� Contrast resolution phantoms
� Flow phantoms
� Training and testing phantoms
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 3 (138/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Tissue-mimicking phantoms (1 of 2)
� Consist principally of a material designed to reproduce the
bulk acoustic properties of soft tissue
• simple laboratory phantoms can be made from suspensions of
gelatine or agar
• commercially available phantoms use proprietary polymer
materials with greater shelf life than gelatine- or agar-based
materials
� The sound speed and attenuation coefficients of a
phantom is usually carefully controlled
• 1540 ms-1
• 0.5 or 0.75 dB/(cm×MHz)
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 4 (139/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Tissue-mimicking phantoms (2 of 2)
� The phantoms also contain suspensions of small
scatterers to produce visually realistic speckle when the
phantom is imaged
� Fabricated from materials such as graphite, polystyrene, or
collagen
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 5 (140/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Spatial resolution phantoms (1 of 2)
� Background material plus
� Spatial resolution targets
• usually small-diameter metal wires or monofilament fibres
suspended horizontally through the phantom
• produce bright point-like targets when imaged in cross section
� A set of wire targets is positioned at the same depth below
the phantom’s acoustic window
• in a regular pattern
• each pair of adjacent wires presents a progressively decreasing
separation from several millimetres down to 0.5 mm or less
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 6 (141/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Spatial resolution phantoms (2 of 2)
� When the pattern of targets is imaged
• lateral spatial resolution can be estimated by observing which pairs
of targets are resolved in the image
• additional wire targets are oriented in a vertical pattern to enable
evaluation of the axial spatial resolution
� A phantom may include several laterally and axially
oriented groups of wires at different depths to enable
resolution to be assessed throughout the scanner’s field of
view
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 7 (142/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Contrast resolution phantoms (1 of 2)
� Contrast resolution targets are usually
• spherical inclusions
• several millimetres to several centimetres in diameter
• similar to the background material but with a mean backscattering
coefficient greater than or less than the background material
� The targets are designed to produce a specific backscatter
contrast relative to the background material
� The contrast of the inclusions can be varied by changing
• size of scattering particles
• composition of scattering particles
• concentration of scattering particles
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 8 (143/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Contrast resolution phantoms (2 of 2)
� Usually easiest to manipulate contrast by changing the
concentration of scattering particles relative to the
background material
• including a few targets with no scatterers to mimic anechoic lesions
� Targets of differing depth will be distributed throughout the
phantom to enable image contrast to be evaluated as a
function of lesion size and depth
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 9 (144/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Flow phantoms
� Provide a useful means of evaluating the performance of
spectral, colour, and power Doppler systems
� In a typical flow phantom design
• tissue-mimicking material is moulded to include one or more hallow
channels that mimic blood vessels of varying diameter and/or
orientation
• channels are connected via tubing to a calibrated pump that is
used to pump blood-mimicking fluid through the flow channels at
controlled flow rates
� Blood-mimicking fluid, is a suspension of small scatterers
designed to reproduce the acoustic properties of blood
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 10 (145/148)
IAEA
13.6 IMAGE ARTIFACTS AND QUALITY ASSURANCE 13.6.3 Quality Assurance Phantoms and Methods
Training and testing phantoms
� Phantoms are also available for
• training sonographers for specific applications
• testing emerging ultrasound imaging methods
� Examples of training phantoms include anthropomorphic
breast or prostate phantoms designed for practicing
ultrasound-guided biopsy or brachytherapy procedures
� An example of a phantom intended for testing emerging
imaging methods is the elastography phantom
• similar in design to contrast phantoms, except the emulated lesions
differ from the background material in elastic modulus as well as
backscattering coefficient
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13.6.3 Slide 11 (146/148)
IAEA
13. BIBLIOGRAPHY13.
Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13. Bibliography Slide 1 (147/148)
IAEA Diagnostic Radiology Physics: A Handbook for Teachers and Students – 13. Bibliography Slide 2 (148/148)
13. BIBLIOGRAPHY13.
� EVANS, D.H., McDICKEN, W.N., Doppler Ultrasound: Physics,
Instrumentation and Signal Processing, Wiley, New York (2000)
� JENSEN, J.A., Estimation of Blood Velocities Using Ultrasound: A Signal
Processing Approach, Cambridge, UK: Cambridge University Press (1996)
� KREMKAU, F.W., Diagnostic Ultrasound: Principles and Instruments, 7th edn,
Saunders/ Elsevier, St. Louis, MO (2006)
� QIN, S., CASKEY, C.F., FERRARA, K.W., Ultrasound Contrast Microbubbles
in Imaging and Therapy: Physical Principles and Engineering, Phys. Med. Biol.
54:R27-R57 (2009)
� SHUNG, K.K., Diagnostic Ultrasound: Imaging and Blood Flow
Measurements, CRC Press, Boca Raton, FL (2006)
� SZABO, T.L., Diagnostic Ultrasound Imaging: Inside Out, Elsevier Academic
Press, Boston (2004)
� ZAGZEBSKI, J.A., Essentials of Ultrasound Physics, Mosby, St. Louis, MO
(1996)