188
User Guide OI /TB82EC–EN Rev. B Type TB82EC Advantage Series TM 2-wire, 4-electrode conductivity transmitter
Transcript
Page 1: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

User Guide OI/TB82EC–EN Rev. B

Type TB82EC Advantage SeriesTM

2-wire, 4-electrode conductivity transmitter

Page 2: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1

Preface

This publication is for the use of technical personnel responsi-ble for installation, operation, maintenance and repair of theABB Type TB82EC Advantage Series™ Conductivity/Concen-tration Transmitter.

The Type TB82EC transmitter is delivered with default hard-ware and software configurations. These settings may need tobe changed depending on the application requirements.

Some sections of this instruction have been prepared in proce-dure format. There is a sequence flowchart or table that fol-lows the introduction to the section and any nonproceduralinformation. This flowchart directs personnel to the appropri-ate procedure located in the back of this instruction. By treat-ing each task as a separate entity, the procedures provide aneasy method for finding the information needed to performeach task.

The procedures have check boxes in the margin by each step.When performing a procedure, check each box as each step iscompleted.

It is important for safety and operation that this instruction beread and understood before attempting anything related toinstallation, operation, maintenance or repair.

Page 3: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1

List of Effective Pages

List of Effective Pages

Page No. Change DatePreface Original

List of Effective Pages Originaliii through xi Original

1-1 through 1-11 Original2-1 through 2-3 Original3-1 through 3-6 Original4-1 through 4-4 Original5-1 through 5-3 Original6-1 through 6-4 Original7-1 through 7-2 Original8-1 through 8-10 Original9-1 through 9-2 Original

10-1 through 10-2 Original11-1 through 11-4 Original12-1 through 12-9 Original13-1 through 13-2 Original14-1 through 14-2 Original15-1 through 15-3 OriginalA-1 through A-5 OriginalB-1 through B-5 OriginalC-1 through C-4 Original

Worksheet 1 OriginalPR Index-1 through PR Index-2 Original

PR1-1 through PR1-2 OriginalPR2-1 through PR2-2 OriginalPR3-1 through PR3-2 OriginalPR4-1 through PR4-2 OriginalPR5-1 through PR5-2 OriginalPR6-1 through PR6-2 OriginalPR7-1 OriginalPR8-1 through PR8-2 OriginalPR9-1 through PR9-3 Original

PR10-1 through PR10-2 OriginalPR11-1 through PR11-2 OriginalPR12-1 OriginalPR13-1 through PR13-2 OriginalPR14-1 OriginalPR15-1 OriginalPR16-1 through PR16-2 OriginalPR17-1 Original

When an update is received, insert the latest changed pages and dispose of the superseded pages.

NOTE: On an update page, the changed text or table is indicated by a vertical bar in the outer margin of the pageadjacent to the changed area. A changed figure is indicated by a vertical bar in the outer margin next to the figurecaption. The date the update was prepared will appear beside the page number.

Page No. Change DatePR18-1 OriginalPR19-1 OriginalPR20-1 OriginalPR21-1 OriginalPR22-1 OriginalPR23-1 OriginalPR24-1 OriginalPR25-1 OriginalPR26-1 OriginalPR27-1 through PR27-3 OriginalPR28-1 OriginalPR29-1 through PR29-3 OriginalPR30-1 OriginalPR31-1 through PR31-2 OriginalPR32-1 OriginalPR33-1 OriginalPR34-1 OriginalPR35-1 through PR35-2 OriginalPR36-1 through PR36-2 OriginalPR37-1 through PR37-4 OriginalPR38-1 OriginalPR39-1 through PR39-2 OriginalPR40-1 OriginalPR41-1 through PR41-2 OriginalPR42-1 OriginalPR43-1 OriginalPR44-1 through PR44-2 OriginalPR45-1 through PR45-2 OriginalPR46-1 through PR46-2 OriginalPR47-1 OriginalPR48-1 OriginalPR49-1 OriginalPR50-1 OriginalPR51-1 through PR51-2 OriginalPR52-1 OriginalPR53-1 OriginalPR54-1 OriginalPR55-1 OriginalPR56-1 OriginalIndex-1 through Index-4 Original

Total number of pages in this instruction is 184, consisting of the following:

Page 4: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

Table of Contents

Page

WBPEEUI520003A1 iii

SECTION 1 - INTRODUCTION....................................................................................................1-1OVERVIEW ..................................................................................................................1-1INTENDED USER.........................................................................................................1-1FEATURES...................................................................................................................1-2EQUIPMENT APPLICATION ..........................................................................................1-3INSTRUCTION CONTENT .............................................................................................1-3HOW TO USE THIS INSTRUCTION...............................................................................1-5DOCUMENT CONVENTIONS ........................................................................................1-5REFERENCE DOCUMENTS..........................................................................................1-5NOMENCLATURE ........................................................................................................1-5SPECIFICATIONS.........................................................................................................1-6ACCESSORIES...........................................................................................................1-10

SECTION 2 - DESCRIPTION AND OPERATION........................................................................2-1INTRODUCTION...........................................................................................................2-1FUNCTIONAL OPERATION ...........................................................................................2-1USER INTERFACE .......................................................................................................2-1MODULAR ELECTRONICS ...........................................................................................2-1TEMPERATURE COMPENSATION ................................................................................2-2DAMPING.....................................................................................................................2-2DIAGNOSTICS .............................................................................................................2-2

Transmitter Diagnostics.........................................................................................2-3Sensor Diagnostics ................................................................................................2-3Spike Output .........................................................................................................2-3

SECTION 3 - INSTALLATION .....................................................................................................3-1INTRODUCTION...........................................................................................................3-1SPECIAL HANDLING ....................................................................................................3-1UNPACKING AND INSPECTION ....................................................................................3-2LOCATION CONSIDERATIONS .....................................................................................3-2

Hazardous Locations..............................................................................................3-3Radio Frequency Interference.................................................................................3-4

WIRING CONSIDERATIONS .........................................................................................3-4OTHER EQUIPMENT INTERFACE ................................................................................3-4TRANSMITTER ROTATION ...........................................................................................3-4INSTALLATION SEQUENCE .........................................................................................3-5

SECTION 4 - OPERATING PROCEDURES................................................................................4-1INTRODUCTION...........................................................................................................4-1OPERATOR INTERFACE CONTROLS............................................................................4-1

Liquid Crystal Display............................................................................................4-1Smart Keys ............................................................................................................4-2

MODES OF OPERATION ..............................................................................................4-3HOLD ICON .................................................................................................................4-4FAULT ICON ................................................................................................................4-4SPIKE ICON .................................................................................................................4-4

SECTION 5 - MEASURE MODE..................................................................................................5-1INTRODUCTION...........................................................................................................5-1BOREDOM SWITCH.....................................................................................................5-1PRIMARY DISPLAY.......................................................................................................5-1

Page 5: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

iv WBPEEUI520003A1

Table of Contents (continued)

Page

SECTION 5 - MEASURE MODE (continued)SECONDARY DISPLAY................................................................................................. 5-1

Fault Information Smart Key ................................................................................. 5-2Menu Smart Key.................................................................................................... 5-2

SECTION 6 - CALIBRATION.......................................................................................................6-1INTRODUCTION .......................................................................................................... 6-1CALIBRATE STATES.................................................................................................... 6-1CALIBRATION SEQUENCE .......................................................................................... 6-2

SECTION 7 - OUTPUT/HOLD MODE..........................................................................................7-1INTRODUCTION .......................................................................................................... 7-1OUTPUT/HOLD STATES OF OPERATION .................................................................... 7-1

Hold/Release Hold Output State............................................................................ 7-1Rerange State........................................................................................................ 7-1Damping State ...................................................................................................... 7-2Spike State............................................................................................................ 7-2

SECTION 8 - CONFIGURATION.................................................................................................8-1INTRODUCTION .......................................................................................................... 8-1PRECONFIGURATION DATA REQUIRED ..................................................................... 8-1CONFIGURE VIEW/MODIFY STATE ............................................................................ 8-1BASIC/ADVANCED PROGRAMMING MODE ................................................................ 8-2MODIFY CONFIGURE STATES OF OPERATION ........................................................... 8-3

Analyzer State (Basic/Advanced) ........................................................................... 8-3Conductivity Analyzer State (Basic/Advanced) ....................................................... 8-3Concentration Analyzer State (Advanced)............................................................... 8-5Temperature Sensor State (Basic/Advanced) ......................................................... 8-5Temperature Compensation State (Basic/Advanced) .............................................. 8-5Output State (Basic/Advanced) ............................................................................. 8-6Damping State (Basic/Advanced)........................................................................... 8-7Diagnostics State (Basic/Advanced)....................................................................... 8-7Safe Mode State (Basic/Advanced)......................................................................... 8-7Spike State (Advanced) .......................................................................................... 8-7

CONFIGURATION LOCKOUT ....................................................................................... 8-7CONFIGURATION SEQUENCE..................................................................................... 8-7

SECTION 9 - SECURITY MODE .................................................................................................9-1INTRODUCTION .......................................................................................................... 9-1SECURITY STATE........................................................................................................ 9-1ENTERING PASSWORD ............................................................................................... 9-1SECURITY SEQUENCE................................................................................................ 9-1

SECTION 10 - SECONDARY DISPLAY MODE ........................................................................10-1INTRODUCTION ........................................................................................................ 10-1SECONDARY DISPLAY STATES OF OPERATION........................................................ 10-1SECONDARY DISPLAY OPERATION........................................................................... 10-1

Page 6: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1 v

Table of Contents (continued)

Page

SECTION 11 - UTILITY MODE ..................................................................................................11-1INTRODUCTION.........................................................................................................11-1FACTORY AND USER STATES ...................................................................................11-1

User State............................................................................................................11-1Advanced/Basic Programming Mode User State ...................................................11-3Reset Configuration User State ............................................................................11-3Reset Security User State.....................................................................................11-3Reset All User State .............................................................................................11-3Soft Boot User State.............................................................................................11-4

SECTION 12 - DIAGNOSTICS AND TROUBLESHOOTING ....................................................12-1INTRODUCTION.........................................................................................................12-1FAULT CODES...........................................................................................................12-1

Problem Codes.....................................................................................................12-1Error Codes .........................................................................................................12-6

CALIBRATION DIAGNOSTICS.....................................................................................12-7ADDITIONAL DIAGNOSTICS ......................................................................................12-8TROUBLESHOOTING SEQUENCE..............................................................................12-8

SECTION 13 - MAINTENANCE.................................................................................................13-1INTRODUCTION.........................................................................................................13-1PREVENTIVE MAINTENANCE SCHEDULE .................................................................13-1

SECTION 14 - REPAIR AND REPLACEMENT.........................................................................14-1INTRODUCTION.........................................................................................................14-1REPAIR AND REPLACEMENT SEQUENCE .................................................................14-1

SECTION 15 - SUPPORT SERVICES.......................................................................................15-1INTRODUCTION.........................................................................................................15-1RETURN MATERIALS PROCEDURES .........................................................................15-1REPLACEMENT PARTS ..............................................................................................15-1RECOMMENDED SPARE PARTS KITS........................................................................15-2

APPENDIX A - TEMPERATURE COMPENSATION.................................................................. A-1INTRODUCTION.......................................................................................................... A-1TEMPERATURE COMPENSATION TYPES .................................................................... A-1

Standard KCl ........................................................................................................ A-1Temperature Coefficient ........................................................................................ A-2NaOH.................................................................................................................... A-2NaCl ..................................................................................................................... A-3HCl ....................................................................................................................... A-3H2SO4..................................................................................................................................................A-4User-Defined......................................................................................................... A-4Pure Water............................................................................................................ A-4

APPENDIX B - CONCENTRATION PROGRAMMING...............................................................B-1INTRODUCTION.......................................................................................................... B-1USER-DEFINED CONDUCTIVITY-TO-CONCENTRATION CURVES............................... B-1PREPROGRAMMED CONDUCTIVITY-TO-CONCENTRATION CURVES ......................... B-2

NaOH.................................................................................................................... B-2

Page 7: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

vi WBPEEUI520003A1

Table of Contents (continued)

Page

APPENDIX B - CONCENTRATION PROGRAMMING (continued)NaCl......................................................................................................................B-3HCl .......................................................................................................................B-4H2SO4 ...................................................................................................................B-4

APPENDIX C - PROGRAMMING TEXT STRING GLOSSARY ................................................. C-1INTRODUCTION ..........................................................................................................C-1TEXT PROMPTS...........................................................................................................C-1FLOW TREE ................................................................................................................C-3

List of Procedures

No. Title

PR1 - PIPE MOUNTINGPR2 - PANEL MOUNTINGPR3 - WALL MOUNTINGPR4 - HINGE MOUNTINGPR5 - SIGNAL/POWER WIRINGPR6 - SENSOR WIRINGPR7 - GROUNDINGPR8 - TEMPERATURE CALIBRATIONPR9 - CONDUCTIVITY/CONCENTRATION CALIBRATIONPR10 - OUTPUT CALIBRATIONPR11 - EDIT CALIBRATIONPR12 - RESET CALIBRATIONPR13 - DAMPING ADJUSTMENTPR14 - HOLD OUTPUTPR15 - RELEASE HOLD OUTPUTPR16 - RERANGE OUTPUTPR17 - OUTPUT SPIKE TOGGLEPR18 - SELECTING VIEW OR MODIFY CONFIGURATION STATEPR19 - MODIFYING CONFIGURATION WHILE VIEWINGPR20 - PROGRAMMING MODE CHANGEPR21 - ANALYZER TYPE SELECTIONPR22 - SENSOR GROUP SELECTIONPR23 - CONFIGURATION OPTION SELECTIONPR24 - ENGINEERING UNITS SELECTIONPR25 - DECIMAL POINT LOCATION SELECTIONPR26 - TEXT STRING DEFINITIONPR27 - LINEAR CURVE FIT PROGRAMMINGPR28 - TEMPERATURE SENSOR TYPE SELECTIONPR29 - TEMPERATURE COMPENSATION TYPE SELECTIONPR30 - TEMPERATURE COMPENSATION COEFFICIENT SETTINGPR31 - USER-DEFINED TEMPERATURE COEFFICIENT PROGRAMMINGPR32 - OUTPUT RANGE SETTINGPR33 - DIAGNOSTICS SELECTIONPR34 - SAFE MODE SELECTION

Page 8: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1 vii

List of Procedures (continued)

No. Title

PR35 - SPIKE MAGNITUDE SETTINGPR36 - CONFIGURATION LOCKOUTPR37 - SECURITY AND PASSWORD ASSIGNMENTPR38 - UTILITY MODE ADVANCED/BASIC PROGRAMMINGPR39 - CONFIGURATION RESETPR40 - SECURITY RESETPR41 - RESETTING ALL PARAMETERSPR42 - TRANSMITTER SOFT BOOTPR43 - SENSOR INSPECTIONPR44 - ELECTRONIC SENSOR TESTPR45 - TRANSMITTER TROUBLESHOOTING WITHOUT SENSORPR46 - SENSOR CLEANINGPR47 - KEYPAD CLEANINGPR48 - FRONT BEZEL REMOVALPR49 - POWER SUPPLY PCB REMOVALPR50 - CONDUCTIVITY INPUT PCB REMOVALPR51 - MICROPROCESSOR PCB REPLACEMENTPR52 - CONDUCTIVITY INPUT PCB INSTALLATIONPR53 - POWER SUPPLY PCB INSTALLATIONPR54 - FRONT BEZEL INSTALLATIONPR55 - REAR COVER REPLACEMENTPR56 - SHELL REPLACEMENT

List of Figures

No. Title Page

1-1. Load Limits..........................................................................................................1-103-1. Transmitter Dimensions ........................................................................................3-33-2. Transmitter Rotation..............................................................................................3-53-3. Installation Sequence.............................................................................................3-64-1. Liquid Crystal Display............................................................................................4-25-1. Operating Mode Screen Flow..................................................................................5-36-1. Calibrate Mode Screen Flow ...................................................................................6-36-2. Calibration Sequence .............................................................................................6-47-1. Output/Hold Mode Screen Flow.............................................................................7-28-1. Modify/View and Basic/Advanced States Screen Flow............................................8-28-2. Modify Configure States Screen Flow......................................................................8-48-3. User-Defined Temperature Compensation Curve....................................................8-68-4. Configuration Sequence .........................................................................................8-88-5. Advanced Configuration Sequence .........................................................................8-98-6. Basic Configuration Sequence..............................................................................8-109-1. Security State Screen Flow.....................................................................................9-210-1. Secondary Display Mode Screen Flow...................................................................10-211-1. User State Screen Flow ........................................................................................11-212-1. Troubleshooting Sequence ...................................................................................12-914-1. Repair and Replacement Sequence.......................................................................14-215-1. Exploded View .....................................................................................................15-2

Page 9: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

List of Figures (continued)

No. Title Page

viii WBPEEUI520003A1viii WBPEEUI520003A1

B-1. NaOH Conductivity-To-Concentration Curve..........................................................B-3B-2. NaCl Conductivity-To-Concentration Curve ...........................................................B-3B-3. HCl Conductivity-To-Concentration Curve .............................................................B-4B-4. H2SO4 Conductivity-To-Concentration Curve.........................................................B-5C-1. Function Flow Tree................................................................................................C-4PR1-1. Pipe Mounting ...................................................................................................PR1-2PR2-1. Panel Mounting .................................................................................................PR2-2PR3-1. Wall Mounting ...................................................................................................PR3-2PR4-1. Hinge Mounting.................................................................................................PR4-2PR5-1. Load Limits .......................................................................................................PR5-2PR5-2. Signal and Power Wiring....................................................................................PR5-2PR6-1. Four-Electrode Sensor Connections...................................................................PR6-2PR7-1. Earth Grounds ..................................................................................................PR7-1PR10-1. Output Calibration Setup ................................................................................ PR10-1PR27-1. Linear Curve Fit .............................................................................................. PR27-2PR36-1. Configuration Lockout Jumper ........................................................................ PR36-2PR45-1. Sensor Simulation Setup ................................................................................. PR45-1PR49-1. Power Supply PCB Removal ............................................................................. PR49-1PR50-1. Conductivity Input PCB Removal ..................................................................... PR50-1PR51-1. Microprocessor PCB Replacement.................................................................... PR51-1

List of Tables

No. Title Page

1-1. Reference Documents ............................................................................................ 1-51-2. Nomenclature........................................................................................................ 1-61-3. Specifications ........................................................................................................ 1-71-4. Accessory Kits ..................................................................................................... 1-101-5. Sensors ............................................................................................................... 1-104-1. Smart Key Functions............................................................................................. 4-34-2. Modes of Operation ............................................................................................... 4-45-1. Engineering Units.................................................................................................. 5-16-1. Calibrate States..................................................................................................... 6-17-1. Output/Hold States............................................................................................... 7-18-1. User-Defined Temperature Compensation Data ..................................................... 8-611-1. User States.......................................................................................................... 11-111-2. Factory Software Defaults.................................................................................... 11-312-1. Common Problem Codes...................................................................................... 12-212-2. Uncommon Problem Codes.................................................................................. 12-512-3. Error Codes......................................................................................................... 12-713-1. Preventive Maintenance Schedule........................................................................ 13-215-1. Spare Parts Kits .................................................................................................. 15-2A-1. Standard KCl Data Breakpoints.............................................................................A-2A-2. NaOH Data Breakpoints ........................................................................................A-3A-3. NaCl Data Breakpoints ..........................................................................................A-3A-4. HCl Data Breakpoints............................................................................................A-4A-5. H2SO4 Data Breakpoints .......................................................................................A-4

Page 10: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1 ix

No. Title Page

List of Tables (continued)

A-6. Pure Water Coefficient Values ............................................................................... A-5B-1. NaOH Conductivity-To-Concentration Data........................................................... B-2B-2. NaCl Conductivity-To-Concentration Data............................................................. B-3B-3. HCl Conductivity-To-Concentration Data .............................................................. B-4B-4. H2SO4 Conductivity-To-Concentration Data.......................................................... B-4C-1. Text Prompt Equivalents ....................................................................................... C-1PR6-1. Pin Terminal Sensor Connections...................................................................... PR6-1PR16-1. Output Ranges.................................................................................................PR16-1PR21-1. Analyzer States ................................................................................................PR21-1PR22-1. Sensor Ranges .................................................................................................PR22-1PR27-1. Conductivity-to-Concentration Data.................................................................PR27-2PR29-1. Temperature Compensation States...................................................................PR29-2PR39-1. Factory Software Defaults ................................................................................PR39-1PR45-1. Sensor Simulation Values ................................................................................PR45-2

Page 11: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

x WBPEEUI520003A1

Safety Summary

SPECIFIC

WARNINGSUse this equipment only in those classes of hazardous locationslisted on the nameplate. Uses in other hazardous locations can leadto unsafe conditions that can injure personnel and damageequipment. (p. 3-3)

All error conditions are considered catastrophic and require trans-mitter replacement. Replace the transmitter with a transmitter that isknown to be operable. Leaving an inoperable transmitter in the pro-cess could cause process upsets. Some process upsets can lead todangerous conditions that can harm personnel and damageequipment. (p. 12-7)

Allow only qualified personnel (refer to INTENDED USER inSection 1) to commission, operate, service, or repair this equip-ment. Failure to follow the procedures described in this instruction orthe instructions provided with related equipment can result in anunsafe condition that can injure personnel and damage equipment.(p. 13-1)

Consider the material compatibility between cleaning fluids and pro-cess liquids. Incompatible fluids can react with each other causinginjury to personnel and equipment damage. (p. PR46-1)

Acids and bases can cause severe burns. Use hand and eye protec-tion when handling. (p. PR46-1)

Use solvents only in well ventilated areas. Avoid prolonged orrepeated breathing of vapors or contact with skin. Solvents cancause nausea, dizziness, and skin irritation. In some cases, overex-posure to solvents has caused nerve and brain damage. Solventsare flammable - do not use near extreme heat or open flame.(p. PR46-1)

Page 12: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520002A1 xi

Trademarks and Registrations

Registrations and trademarks used in this document include:

™ Advantage Series Trademark of ABB.

® HART Registered trademark of HART Communication Foundation.

® Mylar Registered trademark of E. I. DuPont de Nemours Company, Inc.

™ Next Step Trademark of ABB.

Page 13: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

OVERVIEWWBPEEUI520003A1 1 - 1

SECTION 1 - INTRODUCTION

OVERVIEW

The Type TB82EC Advantage Series™ Conductivity/Concen-tration Transmitter is an advanced, microprocessor-based,two-wire, four to 20-milliamp compatible transmitter. It fea-tures internal and external diagnostic functionality, an inno-vative user interface with flexible smart key design, twouser-selectable modes of operation, and DIN size packaging.

Diagnostic checks on the internal circuitry and external sen-sor are done continuously. This insures accuracy and immedi-ate problem notification. Detection of sensor integrity includes:fouled or dirty sensor, ground loop detection, and shorted sen-sor cabling. Additional software functions monitor slope, pro-cess variable over or under range, and temperature over orunder range. The transmitter can be programmed to produce,on the output current, a repetitive modulation of a given mag-nitude when these diagnostic conditions occur.

The transmitter packaging conforms to DIN standards. Mount-ing options accommodate pipe, wall, hinge, and panelinstallations.

Changing the transmitter sensing capability to other analyticalproperties such as pH/ORP/pION is quick and easy due to themodular design.

The user interface is an innovative, patent-pending technologythat facilitates a smooth and problem-free link between theuser and transmitter functionality. The programming struc-ture and smart keys reduce programming difficulties by pro-viding a toggle between basic and advanced functions.

INTENDED USER

Installation Personnel Should be an electrician or a person familiar with the NEC(National Electrical Code) and local wiring regulations. Shouldhave a strong background in installation of analyticalequipment.

Application Technician Should have a solid background in conductivity and concen-tration measurements, electronic instrumentation, and pro-cess control, and be familiar with proper grounding and safetyprocedures for electronic instrumentation.

Operator Should have knowledge of the process and should read andunderstand this instruction before attempting any procedurepertaining to the operation of the transmitter.

Page 14: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

FEATURES1 - 2 WBPEEUI520003A1

Maintenance Personnel Should have a background in electricity and be able to recog-nize shock hazards. Personnel must also be familiar with elec-tronic process control instrumentation and have a goodunderstanding of troubleshooting procedures.

FEATURES

Diagnostic sensor capability. Full compatibility with allfour-electrode Model TB4 Series Conductivity Sensors andthree types of RTDs (resistive temperature devices). Diagnosticcapability includes ground loop detection, process and temper-ature variable under or over range, and dirty sensor detection.

Multiple applications. Accepts inputs from all ABB four-elec-trode conductivity sensors. Isolated analog output allows usein grounded or floating circuits.

Temperature compensation. Compatible with either Pt (plati-num) 100, three-kilohm Balco RTD, or ABB 4.75-kilohm RTD net-work. Manual temperature compensation is 0.1N KCl (potassiumchloride) based. Automatic temperature compensation is based oneither: standard (0.1N KCl based), user-entered coefficient adjustablefrom zero to 9.99-percent per degree Celsius, zero to 15-percentNaOH (sodium hydroxide), zero to 20-percent NaCl (sodium chlo-ride), zero to 18-percent HCl (hydrochloric acid), zero to 20-percentH2SO4 (sulfuric acid), a user-defined function generator, and purewater with trace acid, pure water with trace base, and neutral salt forpure water.

Wide rangeability. Analog output span does not affect displayrange of 0.000 microsiemens per centimeter to1,999 millisiemens per centimeter for conductivity and zero to1,999 digits, specified in the configured engineering units, forconcentration. Minimum output span is 100 microsiemens percentimeter for Group A sensors, 10 microsiemens per centime-ter for Group B sensors and, one microsiemen per centimeterfor Group C sensors.

Innovative user interface. Four smart keys and custom LCDallow assignment of multiple functions to each smart key, dis-played at appropriate time depending on programming mode.Patent-pending technology reduces number of keys and allowsfor larger, more visible LCD.

Easy calibration. One-point smart calibration routines forprocess variable and temperature automatically determineappropriate level of sensor gain (slope) and offset. Includesprovisions for viewing and modifying sensor calibration data.

NEMA 4X/IP65 housing. Suitable for corrosive environment.Electronics enclosure is corrosion resistant, low copper alumi-num alloy. Chemical-resistant polyurethane powder coatingprovides external protection.

Page 15: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

EQUIPMENT APPLICATIONWBPEEUI520003A1 1 - 3

Suitable for hazardous locations. Complies with industrystandards for intrinsically safe and nonincendive installations.

Diagnostic indication. Custom LCD has dedicated icons thatact as visible indications of output hold, fault, and diagnosticspike conditions.

Secure operation. Hardware lockout prevents unauthorizedaltering of configuration parameters while allowing access toother functions. Software security codes assignable to config-ure, calibrate, and output/hold modes.

Compact packaging. Industry standard ½-DIN size main-tains standard panel cutouts. Increases installation flexibilityby providing pipe, wall, hinge, or panel mounting.

Nonvolatile memory. Stores and retains configuration andcalibration data in the event of power failure.

Diagnostics. Built-in electronic circuitry and firmware rou-tines perform series of self-diagnostics, monitoring memoryand input circuit integrity. Irregularities indicated for mainte-nance purposes.

EQUIPMENT APPLICATION

Use the Type TB82EC transmitter anywhere conductivity orconcentration measurements are desired.

INSTRUCTION CONTENT

Introduction Provides product overview, physical description of product,possible applications, and description of instruction and howto use it. Contains list of reference documents on relatedequipment and subjects, nomenclature, comprehensive list ofhardware performance specifications including applicable cer-tification information, accessories, and compatible sensors.

Description andOperation

Provides short description on transmitter functionality.

Installation Contains special handling procedures for circuit boards withsemiconductor devices, unpacking and inspection instruc-tions, and location, safety, and wiring and cabling consider-ations. Installation sequence flowchart directs installationpersonnel to the appropriate installation procedures.

Operating Procedures Addresses operator interface controls and their function. Listsmode of operation and transmitter condition icons anddescribes their functions.

Page 16: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

INSTRUCTION CONTENT1 - 4 WBPEEUI520003A1

Measure Mode Describes normal transmitter mode of operation including pri-mary and secondary display, fault information smart key, andmenu smart key functions. Contains screen flow diagram.

Calibration Provides information on sensor and transmitter calibration,and calibration data descriptions. Contains screen flow dia-grams. Calibration sequence flowchart directs calibration per-sonnel to appropriate calibration procedures.

Output/Hold Mode Lists and describes output/hold states of operation includinghold, rerange, damping, and spike features. Contains screenflow diagram. Includes table that directs personnel to desiredoutput/hold procedures.

Configuration Defines required actions to establish and program transmitterconfiguration. Contains screen flow diagrams. Configurationsequence flowcharts direct configuration personnel to appro-priate configuration procedures.

Security Mode Provides information about transmitter security codes. Con-tains screen flow diagram. Directs personnel to proper securityand password procedure.

Secondary Display Mode Provides information about secondary display that appearsduring measure mode. Contains screen flow diagram. Directspersonnel to appropriate secondary display procedure.

Utility Mode Defines reset options and basic/advanced programming tog-gle. Contains screen flow diagram. Directs personnel to properutility mode procedures.

Diagnostics andTroubleshooting

Provides description of diagnostic tools available. Containsproblem code and error code tables. Troubleshooting sequenceflowchart directs personnel to appropriate troubleshootingprocedures.

Maintenance Provides preventive maintenance table that directs personnelto various maintenance procedures.

Repair and Replacement Contains repair and replacement sequence flowchart thatdirects repair personnel to proper repair and replacementprocedures.

Support Services Provides list of replacement parts.

Appendix A Provides temperature compensation information.

Appendix B Provides information on concentration programming.

Appendix C Provides glossary of text prompts used in secondary displayduring transmitter programming.

Procedures PR1 throughPR56

Provide procedures for each task.

Page 17: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

HOW TO USE THIS INSTRUCTIONWBPEEUI520003A1 1 - 5

HOW TO USE THIS INSTRUCTION

Read this entire instruction in sequence before attempting toinstall, maintain, or repair the transmitter. After gaining acomplete understanding of this instruction and the transmit-ter it can be used as a reference.

Some sections of this instruction have been prepared in proce-dure format. There are flowcharts that follow the introductionto the section and any nonprocedural information. These flow-charts direct personnel to the appropriate procedure. By treat-ing each task as a separate entity, the procedures provide aneasy method for finding the information needed to performeach task. The procedures can be removed and placed intoseparate folders or notebooks, or carried to the job site.

Each procedure lists the recommended tools to perform thatprocedure. Specific tool sizes are listed when required, such asAllen wrench size, socket size, wrench size, etc. Screwdriversare listed as long or short when necessary.

DOCUMENT CONVENTIONS

This document uses standard text conventions to representkeys, display items, and user data inputs:

Display item Any item displayed on a screen appears as italic text. Example:

Running

REFERENCE DOCUMENTS

Table 1-1 lists ABB documents referenced in this instruction.

NOMENCLATURE

Table 1-2 presents the nomenclature for the Type TB82ECtransmitter. Use a single digit or letter in each position.

Table 1-1. Reference Documents

Number Document

P-E21-001 Installing a 4 to 20 mA transmitter in a hazardous location

WBPEEUI110502A0 Type STT smart transmitter terminal instruction

Page 18: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

SPECIFICATIONS1 - 6 WBPEEUI520003A1

SPECIFICATIONS

Table 1-3 lists the specifications for the Type TB82ECtransmitter.

Table 1-2. Nomenclature

Position 5 6 7 8 9 10 11 12 13Advantage Series Transmitter

T B 8 2 _ _ _ _ _ _ _ _ _

Input

P H _ _ _ _ _ _ _ pH/ORP/pION

E C _ _ _ _ _ _ _ Four-electrode conductivity

T E _ _ _ _ _ _ _ Two-electrode conductivity

T C _ _ _ _ _ _ _ Toroidal conductivity

Programming

1 _ _ _ _ _ _ Basic

2 _ _ _ _ _ _ Advanced

Digital Communications

0 _ _ _ _ _ None

1 _ _ _ _ _ HART1

Lightning Arrestor

0 _ _ _ _ None

1 _ _ _ _ Included

Housing Type

0 _ _ _ Powder coated, alodined aluminum

Mounting Hardware

0 _ _ None

1 _ _ Pipe

2 _ _ Hinge

3 _ _ Panel

4 _ _ Wall

Agency Approvals

0 _ None

1 _ FM

2 _ CSA

3 _ CENELEC

4 _ SAA

Label

0 None

1 Stainless steel

2 Mylar®

NOTE:1. This instruction covers the standard Type TB82EC transmitter. Consult factory for availability of HART version.

Page 19: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

SPECIFICATIONSWBPEEUI520003A1 1 - 7

Table 1-3. Specifications

Property Characteristic/Value

Type 2-wire conductivity transmitter

Input types

Conductivity and concentration ABB 4-electrode conductivity sensors

Temperature Pt 100, 3-kΩ Balco RTD, 4.75-kΩ RTD network

Process display range

Conductivity 0.000 µS/cm to 1,999 mS/cm

Concentration 0.000 to 1,999 digits (engineering unit configurable)

Temperature display range -20°C to +300°C (-4°F to +572°F)

Sensor full scale measurement range

Group A 0 to 1,999 mS/cm

Group B 0 to 1,999 µS/cm

Group C 0 to 199.9 µS/cm

Display resolution

Conductivity

Group A 0.1 µS/cm

Group B 0.01 µS/cm

Group C 0.001 µS/cm

Concentration 0.001 digits (configuration dependent)

Temperature 1°C or 1°F

Accuracy

Display

Conductivity ±0.5% of measurement range per decade

Temperature 1°C or 1°F

Output ±0.02 mA for output range set to full scale values

Nonlinearity

Display

Conductivity ±0.5% of measurement range per decade

Temperature 1°C or 1°F

Output ±0.02 mA for output range set to full scale values

Repeatability

Display

Conductivity ±0.5% of measurement range per decade

Temperature 1°C or 1°F

Output ±0.02 mA for output range set to full scale values

Stability

Display

Conductivity ±2 LSD typical, ±5 LSD max.

Temperature 1°C or 1°F

Output ±0.01 mA for output range set to full scale values

Page 20: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

SPECIFICATIONS1 - 8 WBPEEUI520003A1

Temperature compensation Manual (0.1N KCl based) and automatic. Automatic configurable as:

Standard (0.1N KCl based) 0 to 20% H2SO4

Coefficient (0 to 9.99%/°C adjustable) Pure water - neutral salt

0 to 15% NaOH Pure water - trace acid

0 to 20% NaCl Pure water - trace base

0 to 18% HCl User-defined

Dynamic response 3 secs for 90% step change at 0.00-sec damping

Ambient temperature effect at 95% relative humidity

Conductivity ±0.1%/°C full scale, ±0.2%/°C displayed value

Output ±0.01 mA/°C

Output minimum span

Conductivity

Group A 100.0 µS/cm

Group B 10.00 µS/cm

Group C 1.000 µS/cm

Concentration 5% max. concentration range

Output maximum span (full scale settings)

Conductivity

Group A 1,999 mS/cm

Group B 1,999 µS/cm

Group C 199.9 µS/cm

Concentration 1,999 digits

Damping Continuously adjustable from 00.0 to 99.9 secs

Supply voltage 13 to 53 VDC (13 to 42 VDC for agency certified applications)

Load limits Refer to Figure 1-1.

Power supply effect ±0.005% of full scale span per volt

Turn on time 2 secs. typical, 4 secs. max.

Maximum sensor cable length

Group A 30.5 m (100 ft)

Group B 15.2 m (50 ft)

Group C 7.6 m (25 ft)

Sensor diagnostics Fouled or dirty sensor, ground loop detection, shorted cabling

Diagnostic notification Local indication via FAULT or SPIKE icon

Analog mode Programmable output pulse, 0 to 16 mA for 1 sec on 6-sec cycles

Environmental

Temperature

Operating -20°C to +60°C (-4°F to +140°F)

LCD -20°C to +60°C (-4°F to +140°F)

Storage -40°C to +70°C (-40°F to +158°F)

Table 1-3. Specifications (continued)

Property Characteristic/Value

Page 21: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

SPECIFICATIONSWBPEEUI520003A1 1 - 9

Environmental (continued)

Humidity Will meet specifications to 95% RH (operating and storage)

Enclosure Classification NEMA 4X/IP65

Mounting position effect None

Size ½ DIN

h x w x d 144.0 by 144.0 by 171.0 mm (5.67 by 5.67 by 6.73 in.)

Minimum panel depth 144.8 mm (5.70 in.)

Maximum panel cutout 135.4 (+1.3, -0.8) by 135.4 (+1.3, -0.8) mm(5.33 (+0.05, -0.03) by 5.33 (+0.05, -0.03)) in.

Weight 1.9 kg (4.2 lb) without mounting hardware3.4 kg (7.5 lb) with pipe mounting hardware

EMC requirements Complies with all applicable European Community product requirements, specifically those required to display the CE marking on the nameplate.

Electromagnetic emission - EN50081-2:1994

EN55011:1991 (CISPR11:1990) Class A

Electromagnetic immunity - EN50082-2:1996

EN61000-4-2:1995 EN61000-4-3:19976 kV contact 10 V/m (unmodulated, rms)6 kV indirect 80 to 1,000 MHz

EN61000-4-4:1995 EN61000-4-8:19941 kV signal lines 50 Hz5/50 Tr/Th nS 30 A (rms)/m5 kHz

ENV50141:1994 ENV50204:19960.15 to 80 MHz 900, ±5 MHz10 V (unmodulated, rms) 10 V/m (unmodulated, rms)80% AM (1 kHz) 50% duty cycle150 Ω, source impedance 200 Hz

Agency certifications1

Nonincendive (nonsparking)

CSA Class I; Division 2; Groups A, B, C, and DClass II; Division 2; Groups E, F, and GClass III; Division 2

FM Class I; Division 2; Groups A, B, C, and DClass II; Division 2; Groups F and GClass III; Division 2

SAA Ex n, Zone 2; Group IIC, T6

Intrinsic safety When used with appropriate barriers per application guide Installing a 4 to 20 mA transmitter in a hazardous location.

CENELEC EEX ib, Zone 1; Group IIC, T6

CSA Classes I, II, III; Division 1, Applicable Groups A, B, C, D, E, F, and G; T4

FM Classes I, II, III; Division 1; Applicable Groups A, B, C, D, E, F, and G; T6

SAA Ex ia, Zone 0; Group IIC, T6NOTE:1. Hazardous location approvals for use in flammable atmospheres are for ambient conditions of -25°C to +40°C (-13°F to +104°F), 86to 108 kPa (12.5 to 15.7 psi) with a maximum oxygen concentration of 21%.

SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.

Table 1-3. Specifications (continued)

Property Characteristic/Value

Page 22: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

ACCESSORIES1 - 10 WBPEEUI520003A1

ACCESSORIES

Table 1-4 lists the accessory kits for the Type TB82EC trans-mitter and Table 1-5 lists compatible sensors.

Figure 1-1. Load Limits

13 18 23 28 33

S U PP LY VO LTAG E (V D C )

LO A DR E SIS TA N C E

(O H M S)

38 43 48 53

T 01964B

2000

1800

1600

1400

1200

1000

800

600

400

200

0

Table 1-4. Accessory Kits

Part Number Description

1948385?1 Contains static-dissipative work surface (mat), ground cord assembly, wrist bands, and alligator clip for person-nel working on devices containing semiconductor components.

4TB9515-0123 Panel mounting

4TB9515-0124 Pipe mounting

4TB9515-0125 Hinge mounting

4TB9515-0156 Wall mounting

Table 1-5. Sensors

ModelSensor Group

Fitting Type

TB451 A Kynar sanitary/sterilizable

TB452 B

TB454 A Inline, twist lock, submersible

TB456 A Inline, submersible

TB457 A 316 stainless steel sanitary/sterilizable

TB458 B

TB459 C

TB461 A Inline threaded, submersible ball valve insertion, hot tap

TB465 B

TB466 C High purity, flow cell

TB467 C Inline threaded, submersible ball valve insertion, hot tap

Page 23: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTION

ACCESSORIESWBPEEUI520003A1 1 - 11

TB468 A Inline threaded, corrosion-resistant, hot tap

TB471 A High pressure hot tap

TB475 B

TB477 C

Table 1-5. Sensors (continued)

ModelSensor Group

Fitting Type

Page 24: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 2 - 1

SECTION 2 - DESCRIPTION AND OPERATION

INTRODUCTION

This section contains an overview of the functionality of theType TB82EC transmitter.

FUNCTIONAL OPERATION

The Type TB82EC transmitter provides a four to 20-milliampoutput signal that is proportional to solution conductivity asmeasured by the transmitter and the sensor. The transmitteris compatible with all ABB four-electrode conductivity sensors.

The transmitter has internal diagnostic capabilities. Thesedetect any potential electronics or firmware operation prob-lems. Diagnostic capabilities also include detection of sensorintegrity such as:

• Ground loop detection.• Shorted cabling.• Process variables out of range.• Incorrect calibration values.

USER INTERFACE

The user interface consists of a tactile keypad with fournondedicated smart keys and a custom LCD.

The LCD has a three and one-half digit numeric region thatdisplays the process variable, a six-digit alphanumeric regionthat displays secondary information and programmingprompts, and several status-indicating and programmingicons.

Using a patent-pending novel approach, each of the four smartkeys is located under a given set of icons. In each of the instru-ment modes and mode states, one icon over any given smartkey illuminates and represents that smart key function. Thesesmart key assignments vary as different programming modesand states are entered. In addition to the smart key assign-ments, text strings located in the six-character alphanumericfield (secondary display) are used as programming prompts.

MODULAR ELECTRONICS

The transmitter consists of three separate PCB (printed circuitboard) assemblies that concentrate specific circuit functional-ity. This modular design allows for the ability to change theinstrument from one of four types of instruments to

Page 25: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DESCRIPTION AND OPERATION

TEMPERATURE COMPENSATION2 - 2 WBPEEUI520003A1

another: pH/ORP/pION, four-electrode conductivity, two-elec-trode conductivity, and toroidal conductivity. In addition,instrument repair is made quick and easy by replacing thenonfunctioning PCB.

TEMPERATURE COMPENSATION

The process temperature is monitored using one of three typesof RTD inputs: three-kilohm Balco, Pt 100, and the ABB4.75-kilohm RTD network. It is possible to program the sec-ondary display to show the temperature in degrees Celsius ordegrees Fahrenheit when in the measure mode.

Temperature affects the activity of the disassociated ions insolution and hence the conductivity of that solution. There-fore, manual and automatic temperature compensation func-tions are available. Temperature compensation functions forconductivity and concentration configurations include manual(0.1N KCl based) and 10 types of automatic compensation rou-tines. Refer to Table 1-3 for a list of the automatic types.

DAMPING

Input damping can be adjusted from 00.0 to 99.9 seconds.This feature is useful in noisy process environments. It helpsminimize the displayed process variable and output currentbounce.

Damping simulates a capacitive type lag where reaction to anysignal change is slowed according to an entered time constant.For example, the output response to a step change in inputreaches approximately 63.2 percent of its final value in fiveseconds for five seconds of damping.

DIAGNOSTICS

Diagnostics are provided for both the transmitter and sensor.Diagnostic detection of a serious condition that prevents theinstrument from properly functioning enables a preset safemode state. This configurable safe mode state forces theinstrument output to be either high or low.

Some problems do not keep the instrument from functioning.A diagnostic spike output feature is used for these conditions.Once enabled, this feature modulates the output for onesecond out of every six seconds. The magnitude of thesemodulations can be set from zero to 100 percent of the maxi-mum output. Detection of over 40 problem conditions can beenabled.

In both cases, diagnostic conditions cause the FAULT andFAULT info icons on the display to energize. Interrogation ofeach fault condition is available using a single keystroke.

Page 26: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DESCRIPTION AND OPERATION

DIAGNOSTICSWBPEEUI520003A1 2 - 3

Section 12 provides diagnostics information.

Transmitter Diagnostics

Five critical errors in operation are monitored and linked tothe safe mode feature. These conditions include: inoperable orincorrect input circuit, bad RAM, and damaged EE memory.

Sensor Diagnostics

The transmitter continually performs diagnostics on sensorintegrity. When configured to do so, the FAULT and FAULT infoicons and the spike output feature notify the operator of incon-sistencies in sensor performance.

Sensor faults that activate the diagnostics are: fouled or dirtysensor, shorted cable, shorted or open temperature sensor,high and low PV (process variable), high and low temperature,and many more.

Spike Output

Using the spike state in the configure mode initiates remoteproblem notification. The spike output option allows program-ming of a one to 100-percent (0.16 to 16-milliamp) pulseimpressed on the four to 20-milliamp output for one secondout of a six-second repeating cycle upon detection of a problemcondition. Should the actual output of the transmitter bebelow 12 milliamps, the pulse adds current; if above 12 milli-amps, it subtracts current.

Page 27: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 3 - 1

SECTION 3 - INSTALLATION

INTRODUCTION

This section contains special handling procedures for circuitboards with semiconductor devices, inspection instructions,and special location and safety considerations.

Following these topics is an installation sequence flowchartthat guides personnel, seeking to perform a specific installa-tion task, to the proper procedures to perform that task.

SPECIAL HANDLING

In addition to the normal precautions for storage and handlingof electronic equipment, the transmitter has special semicon-ductor handling requirements. This equipment contains elec-tronic components that can be damaged from discharges ofstatic electricity. If at all possible, do not touch the compo-nents on the circuit board. Ordinarily, the circuit will not bedamaged if the circuit board is handled by the edges.

Semiconductor devices are subject to damage by static elec-tricity. Therefore, observe the following techniques during ser-vicing, troubleshooting, and repair.

1. Remove assemblies containing semiconductor devices fromtheir protective containers only under the following conditions:

a. When at a designated static-free workstation.

b. Only after firm contact with an antistatic mat and/orfirmly gripped by a grounded individual.

2. Personnel handling assemblies with semiconductor devicesshould be neutralized to a static-free work station by agrounding wrist strap that is connected to the station or to agood ground point at the field site.

3. Do not allow clothing to make contact with semiconductordevices. Most clothing generates static electricity.

4. Avoid touching edge connectors and components.

5. Avoid partial connection of semiconductor devices. Semi-conductor devices can be damaged by floating leads, especiallythe power supply connector. If an assembly must be insertedin a live system, it should be done quickly. Do not cut leads orlift circuit paths when troubleshooting.

6. Ground the test equipment.

Page 28: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INSTALLATION

UNPACKING AND INSPECTION3 - 2 WBPEEUI520003A1

7. Avoid static charges during maintenance. Make sure thecircuit board is thoroughly clean around its leads but do notrub or clean with an insulating cloth.

NOTE: An antistatic kit (refer to Table 1-4) is available for personnelworking on devices containing semiconductor components.

UNPACKING AND INSPECTION

Examine the equipment upon receipt for possible damage intransit. File a damage claim with the responsible transporta-tion company if necessary and notify the nearest ABB salesoffice.

Carefully inspect packing material before discarding it to makesure that all mounting equipment and any special instructionsor paperwork has been removed. Careful handling and instal-lation insures satisfactory performance of the transmitter.

Use the original packing material and container for storage.The storage environment should be protected and free fromextremes of temperature and humidity, and fall within theenvironmental constraints listed in Table 1-3.

NOTE: Remove the protective film from the transmitter lens afterplacing it in its final installed location.

LOCATION CONSIDERATIONS

When mounting the transmitter, leave ample clearance forremoval of the front bezel and rear cover. Signal wiring shouldnot run in conduit or open trays where power wiring or heavyelectrical equipment could contact or interfere with the signalwiring. Use twisted, shielded pairs for best results.

Figure 3-1 shows the overall dimensions of the transmitter.Mounting hardware attaches to two or more of the four sets ofthreaded holes on the housing.

The transmitter design allows for panel mounting, pipe mount-ing, hinge mounting, or wall mounting. The installation siteshould be vibration free and conform to the environmentalconstraints listed in Table 1-3. Careful placement of the trans-mitter insures proper operation as well as overall safety.

NOTE: Temperature is an important consideration. Allow for adequateair flow, especially if installing the transmitter in an enclosed area.

Page 29: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INSTALLATION

LOCATION CONSIDERATIONSWBPEEUI520003A1 3 - 3

Hazardous Locations

Table 1-3 lists the agencies and types of hazardous locationcertifications for the transmitter.

Figure 3-1. Transmitter Dimensions

WARNING

Use this equipment only in those classes of hazardous loca-tions listed on the nameplate. Uses in other hazardous loca-tions can lead to unsafe conditions that can injure personneland damage equipment.

FRO N T

S ID E

B OTTO M

143.85.66

114 .04 .49

143.55 .65

143.85.66

41.91 .65

22.20 .88

(2 TY P)

D IA

T 01688A

50.82 .00

10.20.40

26.71 .05

14 .00.55

(4 TY P) 32.81.29

(4 TY P )

D IM E N S IO N S

M ILLIM ETE R SIN C H E S

Page 30: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INSTALLATION

WIRING CONSIDERATIONS3 - 4 WBPEEUI520003A1

Refer to Installing a 4 to 20 mA transmitter in a hazard-ous location for additional information when using equip-ment in a hazardous area.

Radio Frequency Interference

Most electronic equipment is influenced by RFI (radio fre-quency interference). Exercise caution with regard to the useof portable communications equipment in the area. Postappropriate signs in the plant.

WIRING CONSIDERATIONS

NOTE: To prevent possible signal degradation, use a separate metalconduit run for both the sensor and signal/power wiring.

Transmitter power passes through the signal leads. Underideal conditions, the use of conduit and shielded wire may notbe required. However, to avoid noise problems, enclose thesensor and signal/power wiring in separate conduit. Just priorto entering the housing, terminate rigid conduit and install ashort length of flexible conduit to reduce any stress to thehousing.

Signal/power wiring must bear a suitable voltage rating, atemperature rating of 75-degrees Celsius (167-degrees Fahren-heit), and must be in accordance with all NEC requirementsfor the installation site.

OTHER EQUIPMENT INTERFACE

The transmitter has an isolated output and controls the loopcurrent between four and 20 milliamps depending on therange and process variable values. Since the output is iso-lated, the instrument loop may have a maximum of onenonisolated device within its circuit. The maximum load on thecurrent loop must not exceed that shown in Figure 1-1.

TRANSMITTER ROTATION

The transmitter has four pairs of threaded mounting holes.Since these holes are located at the corners of the transmitter,it can be mounted in any of the four positions as shown inFigure 3-2.

Page 31: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INSTALLATION

INSTALLATION SEQUENCEWBPEEUI520003A1 3 - 5

INSTALLATION SEQUENCE

Refer to Figure 3-3 for the transmitter installation sequence.Each block of the flow represents a single task that must becompleted before continuing with the sequence.

In some cases, more than one path can be taken during instal-lation. For paths that are in parallel, either complete all of thetasks in all of the paths before continuing or complete all ofthe tasks in only those paths that apply before continuing. Atleast one path must be completed.

Some blocks contain alphanumeric codes. These codes identifythe procedure that describes the steps to complete an indi-cated task. Complete all of the steps given in a procedurebefore continuing to the next procedure.

The procedures have check boxes in the margin by each proce-dural step. When performing a procedure, check each box aseach step is completed.

By treating each task as a separate entity, the procedures pro-vide an easy method for finding the information needed to per-form each task in the installation sequence.

Figure 3-2. Transmitter Rotation

N O ROTATIO N

90 R OTATIO No

270 R OTATIONo 180 R OTATIONo

T 01 6 87 A

Page 32: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INSTALLATION

INSTALLATION SEQUENCE3 - 6 WBPEEUI520003A1

Figure 3-3. Installation Sequence

M O U N TIN GS C H E M E ?

T 02434A

S TA RT

H IN G E

W A LL

PA N E L

P IPE

S EN S O RW IR IN G

D O N E

P R 6

G RO U N D IN G

P R 7

S IG N A L/P OW E RW IR IN G

P R 5W A LLM O U N TIN G

P IPEM O U N TIN G

H IN G EM O U N TIN G

PA N E LM O U N TIN G

P R 3

P R 1

P R 4

P R 2

Page 33: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 4 - 1

SECTION 4 - OPERATING PROCEDURES

INTRODUCTION

The Type TB82EC transmitter has six main operating modes:measure, calibrate, output/hold, configure, security, and sec-ondary display. An operating mode has several programmingstates that contain functions specific to that mode.

The transmitter has a built-in user interface through which alltransmitter functions are programmed or monitored. In orderto maximize the viewing area and minimize the space neededfor the keypad, the patent-pending interface uses a customLCD and four-button keypad. Instrument functions and pro-gramming prompts are available through two regions on theLCD. These regions include a primary area that shows the pro-cess variable (conductivity) and a secondary area that displaystext prompts for programming or auxiliary information.

In addition to the user friendly interface, the transmitter has agroup of icons that alert the user of an existing fault condition,diagnostic spike output, or a held output. These icons, locatedat the top of the LCD, only energize under the specified condi-tion. Pressing the FAULT info smart key while in the measuremode allows interrogation of any fault condition.

OPERATOR INTERFACE CONTROLS

The operator interface consists of the LCD and the smart keys.

Liquid Crystal Display

The LCD contains nine regions that provide information on theprocess variable, engineering units, mode of operation, outputhold condition, fault indication, secondary variable, and softkey assignments. A view of the fully energized LCD with smartkey and mode text included is shown in Figure 4-1.

The top set of icons indicates abnormal operating conditionssuch as the Hold, Fault, or Spike state. These icons only ener-gize when the transmitter detects such a condition. They areactive in all modes of operation.

The mode of operation indicators, shown as right arrowsgrouped next to the mode text, indicate the current mode ofoperation. Only one indicator is lit at a time. The appropriateindicator energizes when moving from one mode to the other.The mode of operation indicators are active in all modes ofoperation.

Page 34: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

OPERATING PROCEDURES

OPERATOR INTERFACE CONTROLS4 - 2 WBPEEUI520003A1

The process variable appears in the three and one-half digit,seven-segment region. This display region is supported by theengineering unit region. These regions are normally active inall modes of operation; however, some programming states usethese regions for data entry.

The secondary variable is displayed in the six-character,14-segment region. This display region displays secondaryinformation and fault information in the measure mode andtextual prompting in all other modes of operation. Due to thelimited number of characters for this display region, much ofthe prompting takes the form of text abbreviations. Refer toAppendix C for programming text abbreviations. This region isactive in all modes of operation.

The smart key assignments are grouped into four sets of icons,each group directly positioned above one of the four smartkeys. These icons are textual representations of the functionfor the associated smart key. Only one assignment will beenergized per smart key at any given time.

Smart Keys

A four-button, tactile keypad is located on the front panel ofthe instrument. The gray buttons are embossed to easily showtheir location. A fifth hidden button, located at the top of the

Figure 4-1. Liquid Crystal Display

M E A S U R E

C A LIB R ATE

O U T/H O LD

C O N FIG U R E

S E C U R ITY

D IS P LAY

N OM E N U

M ohm s-cmppmppbS /m

m S /m

%m VpH

m S /cmS /cmµ

Y E SN E X T

M E A S U R EFAU LT

exit to

in fo

S E LE C TE N TE R

FAU LTS P IK EH O LD

T01686B

Page 35: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

OPERATING PROCEDURES

MODES OF OPERATIONWBPEEUI520003A1 4 - 3

NT in the text ADVANTAGE, provides access to infrequentlyused functions.

The four embossed keys are called smart keys since their func-tions are dependent on the mode and state of the instrument.Since these four keys do not have a preassigned function,icons energize over the key to indicate its function. If a smartkey does not have an icon energized above it, this smart keydoes not have a function and initiates no action when pressed.Using this smart key method, a reduced number of keys canbe used without complicating instrument functionality.

Pressing the smart key initiates the displayed function of thatsmart key for each operating mode and state. For example, theNEXT function enables the cycling through of a series of pro-gramming states. The SELECT function enables entering into agiven mode of operation or programming state. Using thismethod, the transmitter guides the user through the necessarysteps to program or monitor the desired functions.

A general description of each smart key function is given inTable 4-1.

MODES OF OPERATION

The measure mode is the normal operating mode of the trans-mitter and is the default mode upon power up. The measuremode is the starting point for entry into other modes. Eachmode contains a unique set of transmitter functions or states.These modes and their related functions are shown inTable 4-2.

Table 4-1. Smart Key Functions

Icon Function

ENTER Stores configured items and alphanumeric data into permanent memory.

exit to MEASURE Escapes to measure mode from all other modes and programming states. Not avail-able in measure mode.

FAULT info Accesses information on diagnostic problem or error conditions. Displays information as short text string and code. Only available in measure mode.

MENU Increments through modes of operation.

NEXT Increments through series of programming states.

NO Denies action about to take place.

SELECT Selects mode of operation or programming state shown in secondary display.

YES Affirms action about to take place.

Decrements numeric values or moves through a series of parameters.

Increments numeric values or moves through a series of parameters.

Steps to right moving from one digit to the next.

Page 36: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

OPERATING PROCEDURES

HOLD ICON4 - 4 WBPEEUI520003A1

HOLD ICON

The HOLD icon energizes when a hold condition is active.Holding the output can only be manually enabled. Manualactivation is accessible in the output/hold mode of operation.In this mode, the hold state permits the output to be held atthe captured level or at a manually set level.

FAULT ICON

The FAULT icon energizes when the transmitter detects a faultcondition. Fault conditions include all problem and errordetection as outlined in Section 12.

SPIKE ICON

When enabled, the spike output function induces a stepchange in the level of the output current. The magnitude of thestep change is configured as a percentage of the output cur-rent. When the transmitter detects a fault condition and hasthe spike output function enabled, the transmitter outputbegins to modulate and the SPIKE icon energizes. This pro-vides local and remote indication of a measurement loop faultcondition. Refer to Section 12 for more information on spikeoutput and fault conditions.

Table 4-2. Modes of Operation

Mode Function

Calibrate Calibration of input and output functions.

Configure Configuration of transmitter functions such as type of conductivity sensor, temperature compensation types, temperature sensor, temperature units, etc.

Display Selection of variable displayed in secondary display when in measure mode.

Measure Display of process and secondary variables. Normal transmitter operating mode.

Output/hold Online tuning of output parameters or manual setting of transmitter output. Useful during instrument maintenance, for example.

Security Entering of password protection for calibrate, output/hold, and configure modes.

Page 37: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 5 - 1

SECTION 5 - MEASURE MODE

INTRODUCTION

The measure mode is the mode of operation upon transmitterpower up and is the normal operating state of the transmitter.In this mode, the process variable, output state, fault condi-tion state, spike state, and secondary display information aredisplayed. All other modes of operation and fault informationare accessible from the measure mode.

BOREDOM SWITCH

When any operating mode or state is entered and the measuremode is not returned to after the final step, the transmitterautomatically returns to the measure mode of operation after20 minutes of unattended use. This feature insures the trans-mitter always ends up in its normal mode of operation.

PRIMARY DISPLAY

The primary display shows the process variable. The value ofthis variable is dependent on the configured analyzer, temper-ature compensation type, temperature value, sensor output,and damping value. The engineering units for the process vari-able are dependent only on the configured analyzer. Table 5-1lists the analyzer types and corresponding engineering units.

SECONDARY DISPLAY

The secondary display has the ability to show a large array ofinformation. Since the display area only has six characters,only one item can be shown at any given time. Typically, thisregion displays the process temperature in degrees Celsius;however, it can be changed to display the process temperaturein degrees Fahrenheit, output current in milliamps, sensortype, sensor group, conductivity value, and solute name for aconcentration analyzer, or firmware revision. Refer toSection 10 for more information.

Table 5-1. Engineering Units

Analyzer Type Engineering Unit

Concentration ppm, ppb, percent, or user-defined

Conductivity mS/cm, µS/cm

Page 38: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

MEASURE MODE

SECONDARY DISPLAY5 - 2 WBPEEUI520003A1

Fault Information Smart Key

Fault information is only accessible from the measure mode ofoperation. It is interrogated through the FAULT info smart key.A fault condition causes the FAULT icon to blink and theFAULT info smart key to appear. These indicators continue tobe present as long as the fault condition exists.

When pressing the FAULT info smart key, the faults appear ina FIFO (first in, first out) order and the first fault condition isshown in the secondary display. A short text string followed bythe fault code is sequentially shown. Depressing the FAULTinfo smart key progressively moves from one fault to the nextuntil all faults have been shown. Once all faults have beencycled through, the FAULT icon no longer blinks but remainson until removal of all fault conditions. If a new fault conditionis detected, the FAULT icon begins to blink to indicate thenewly detected condition. For more information on fault condi-tions and codes, refer to Section 12 and its related procedures.

Menu Smart Key

The MENU smart key provides access to all other modes ofoperation. By pressing the MENU smart key, the transmittermoves from one mode of operation to the next. Visual feedbackis provided in two manners: the mode indication arrow movesto the next mode, such as CALIBRATE, and the secondary dis-play shows the text string representative of that mode, such asCALIBR. Access into the displayed mode of operation isallowed by pressing the SELECT smart key. The exit to MEA-SURE smart key provides an escape function to the measuremode.

As seen in the screen flow diagram shown in Figure 5-1, press-ing the MENU smart key when in the measure mode moves thetransmitter into the calibrate mode. Once in the calibratemode, pressing the exit to MEASURE smart key returns thetransmitter back to the measure mode. Pressing the SELECTsmart key moves the transmitter into the calibrate states ofoperation. Pressing the MENU smart key moves the transmitterto the output/hold mode. Use Figure 5-1 to identify the smartkey assignments and the resulting action.

Page 39: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

MEASURE MODE

SECONDARY DISPLAYWBPEEUI520003A1 5 - 3

Figure 5-1. Operating Mode Screen Flow

T02404A

M E A S U R E

C A LIB R

O U TH LD

C O N FIG

S E C U R

SE C D S P

M EA S U R EC ALIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD ISP LAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD ISP LAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD ISP LAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD ISP LAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD ISP LAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD ISP LAY

exit toM EA SU R E

exit toM EA SU R E

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

23 .5°C

C A LIB R

O U TP U T

C O N FIG

S E C U R

S E C.D S P

68.2

68.2

68.2

68.2

68.2

68.2

S E LE C T

S E LE C T

S E LE C T

S E LE C T

S E LE C T

M EN U

M EN U

M EN U

M EN U

M EN U

M EN U

m S/cm

m S/cm

m S/cm

m S /cm

m S /cm

m S /cm

Page 40: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 6 - 1

SECTION 6 - CALIBRATION

INTRODUCTION

The calibrate mode provides the ability to calibrate the sensorinput, temperature input, and transmitter output. These func-tions, referred to as calibrate states, include conductivity/con-centration calibration, temperature calibration, editcalibration, reset calibration, and output calibration.

CALIBRATE STATES

The calibrate mode consists of five states. Table 6-1 describesthe function of each state.

When in the calibrate mode, pressing the NEXT smart keycauses the display to sequentially move through each calibratestate. This cycle repeats until either selecting a calibrate stateusing the SELECT smart key or choosing the escape functionby pressing the exit to MEASURE smart key.

When performing the conductivity/concentration calibration,invalid new calibration values generate the text stringBAD.CAL on the display, and the calibration value is notaccepted. If the new value is valid, the slope is shown. Pressingthe NEXT smart key displays the offset value. At this point,pressing the NEXT smart key returns the transmitter to theconductivity/concentration calibrate state. Pressing the exit toMEASURE smart key returns the transmitter to the measuremode.

NOTE: If HOLD appears on the display, the transmitter inquires ifthis condition should be released.

More information on sensor calibration diagnostics and trou-bleshooting techniques can be found in Section 12.

Table 6-1. Calibrate States

State Display Description

Conductivity/concen-tration calibration

CON.CAL Calibration of process sensor input via one-point smart calibration that adjusts offset, slope, or both based on sensor calibration history.

Edit calibration EDT.CAL Manual adjustment of process sensor and temperature offset and slope values.

Output calibration OUT.CAL Calibration of transmitter output values to measured values using exter-nal validation device.

Reset calibration RST.CAL Restores calibration values for process and temperature sensors to fac-tory settings.

Temperature calibration

TMP.CAL Calibration of temperature sensor input via one-point smart calibration that adjusts offset, slope, or both based on sensor calibration history.

Page 41: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CALIBRATION

CALIBRATION SEQUENCE6 - 2 WBPEEUI520003A1

Figure 6-1 is a screen flow diagram for the calibrate mode ofoperation.

CALIBRATION SEQUENCE

Refer to Figure 6-2 for the calibration sequence for the trans-mitter. Each block of the flow represents a single task thatmust be completed before continuing with the sequence.

In some cases, more than one path can be taken during cali-bration. For paths that are in parallel, either complete all ofthe tasks in all of the paths before continuing or complete allof the tasks in only those paths that apply before continuing.At least one path must be completed.

Some blocks contain alphanumeric codes. These codes identifythe procedure that describes the steps to complete an indi-cated task. Complete all of the steps given in a procedurebefore continuing to the next procedure.

The procedures have check boxes in the margin by each proce-dural step. When performing a procedure, check each box aseach step is completed.

By treating each task as a separate entity, the procedures pro-vide an easy method for finding the information needed to per-form each task in the calibration sequence.

Page 42: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CALIBRATION

CALIBRATION SEQUENCEWBPEEUI520003A1 6 - 3

C

Figure 6-1. Calibrate Mode Screen Flow

T02405A

M E A S U R E

C O N C A L

TM P C A LR E TU R N

ED TC ALR E TU R N

R S TC A LR E TU R N

PA SS W D

C O N C A LR E TU R N

C A LIB R

N O PA S SW O R DPROTEC T FO R C AL

O U TC A LR E TU R N

TM P.C AL

E D T.C A L

R S T.C A L

O U T.C A L

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ESE C U R ITYD ISP LAY

exit toM E A SU R E

exit toM E A SU R E

exit toM E A SU R E

exit toM E A SU R E

exit toM E A SU R E

C O N .C A L

TM P.C A L

E D T.C A L

R S T.C A L

O U T.C A L

68 .2

68.2

68.2

68.2

68.2

S E LE C T

S E LE C T

S E LE C T

SE LEC T

SE LEC T

N E X T

N E XT

N E XT

N E XT

N E XT

m S/cm

m S/cm

m S/cm

m S/cm

m S/cm

Page 43: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CALIBRATION

CALIBRATION SEQUENCE6 - 4 WBPEEUI520003A1

Figure 6-2. Calibration Sequence

T02435A

O U TP U TC ALIB R AT IO N

O K?

PA S SW O R DP ROTEC TE D ?

E D ITC A LIB R AT IO N ?

S TART

Y ES

YE S

Y ES

D O N E

C O N D U C TIV ITY /C O N C EN TR ATIO N

C A LIBR ATIO N

TEM P E R ATU R EC A LIBR ATIO N

P R 9

P R 8

O U TPU TC A LIBR ATIO N

P R 10

E D ITC A LIB R ATIO N

PR 11

R E S ETC ALIB R AT IO N

P R 12

E N TE RPA SS W O R D

C O N D U C TIV ITY /C O N C E N TR ATIO N

A N D TE M P E R ATU R EC A LIB R AT IO N

O K ?

C O N D U C TIV ITY /C O N C EN TR ATIO N

A N D TEM P ER ATU R EC A LIB R ATIO N

O K ?

Y ES

Y ES

N O

N O

N O

N O

N O

Page 44: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 7 - 1

SECTION 7 - OUTPUT/HOLD MODE

INTRODUCTION

The output/hold mode of operation provides the ability to setthe output to a fixed level, change the output range, damp theoutput signal, or enable the diagnostic spike.

This section describes the states in the output/hold mode ofoperation (Table 7-1).

OUTPUT/HOLD STATES OF OPERATION

When in the output/hold mode, pressing the NEXT smart keycauses the display to move sequentially through each output/hold state. This cycle repeats until either selection of an out-put/hold state using the SELECT smart key or choosing theescape function by pressing the exit to MEASURE smart key.

The output/hold mode consists of five states of operation.Table 7-1 describes the function of each state of operation andthe related procedures. There is no particular sequence forthese procedures. Figure 7-1 is a screen flow diagram for theoutput/hold mode of operation.

Hold/Release Hold Output State

The hold output state allows the transmitter output to be fixedat a level captured when the hold was initiated. This capturelevel can also be manually adjusted to any value between zeroand 100 percent (four and 20 milliamps).

Rerange State

The rerange output/hold state provides the ability to changethe output range. One or both endpoint values can be changedto any value or range of values that are within the specifica-tions listed in Table 1-3.

Table 7-1. Output/Hold States

State Display Function Procedure

Damping DAMPNG Reduces fluctuation in output signal. PR13

Hold HOLD Fixes output level at value captured upon initiation of hold or at manually entered level.

PR14

Release hold REL.HLD Releases existing output/hold state. PR15

Rerange RERANG Changes output range. PR16

Spike SPIKE Enables or disables spike output function if configured. PR17

Page 45: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

OUTPUT/HOLD MODE

OUTPUT/HOLD STATES OF OPERATION7 - 2 WBPEEUI520003A1

Damping State

The damping state applies a lag function on the output signaland reduces fluctuations caused by erratic process conditions.The damping value can be set from 00.0 to 99.9 seconds. Thisvalue represents the time required for the output to reachapproximately 63.2 percent of its final value in response to astep change in the process variable.

Spike State

The spike state toggles the operational state of the spike out-put function. The spike function modulates the current outputby an amount set in the transmitter configuration. This func-tion is only available on transmitters with the advanced pro-gramming option.

Figure 7-1. Output/Hold Mode Screen Flow

T02407A

M E A S U R E

H O LD

PAS S W D

O U TH LD

N O PA S S W O R D P ROTE C TFO R O U T/H O LD

A DVA N C E D C O N FIG O N LY

R E R N G E

DA M P N G2

SP IKE 2

M EA S U R EC ALIB R AT EO U T /H O LDC O N FIG U R ES E C U R ITYD IS P LAY

M EA S U R EC ALIB R AT EO U T /H O LDC O N FIG U R ES E C U R ITYD IS P LAY

M E A S U R EC A LIB R ATEO U T/H O LDC O N F IG U R ES EC U R ITYD IS P LAY

M E A S U R EC A LIB R ATEO U T/H O LDC O N F IG U R ES EC U R ITYD IS P LAY

exit toM E A S U R E

exit toM E A S U R E

exit toM E A SU RE

exit toM E A SU RE

H O LD

R E R N G E

DA M P N G

S P IKE

68.2

68.2

68.2

68.2

S ELE C T

S ELE C T

SE LE C T

SE LE C T

N EX T

N E XT

N EX T

N E XT

m S/cm

m S/cm

m S /cm

m S /cm

Page 46: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 8 - 1

SECTION 8 - CONFIGURATION

INTRODUCTION

The configure mode of operation establishes the operatingparameters of the Type TB82EC transmitter. These parametersinclude: programming type, analyzer type, sensor type, tem-perature compensation type, output range, damping value,diagnostic functionality, safe mode level, and spike magnitude.Review each of the following sections and related proceduresbefore configuring the transmitter.

This section contains screen flow diagrams and brief descrip-tions of the configure states of operation. Refer to CONFIGU-RATION SEQUENCE for procedures needed to perform theconfiguration tasks.

PRECONFIGURATION DATA REQUIRED

Before attempting to configure the transmitter, define thefollowing:

• Analyzer parameters.• Output range values.• Security requirements.• Sensor diagnostic functionality.

Use the worksheet at the end of these sections to help estab-lish the proper settings for any given application. Use thisworksheet during the configuration entry procedure and retainit as a historical record for future reference.

CONFIGURE VIEW/MODIFY STATE

When the configure mode is entered, a decision point isreached to determine whether to modify or view the configura-tion. The modify configure state enables transmitter options tobe set and saved into memory. In order to provide the ability tosecure the modify configure state, yet leave the ability to viewconfiguration information, the view configure state can beentered without using a security code.

As seen in Figure 8-1, the transmitter allows the opportunityto modify the configuration. Pressing the YES smart key movesthe transmitter into the modify configure state. Pressing theNO smart key moves the transmitter to the view configurationinquiry. Pressing the exit To MEASURE smart key escapes tothe measure mode. To view a configuration, press the YESsmart key when queried to view the configuration.

Page 47: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

BASIC/ADVANCED PROGRAMMING MODE8 - 2 WBPEEUI520003A1

BASIC/ADVANCED PROGRAMMING MODE

The configure mode is split into two groups of programming:basic and advanced. These two options are specified bynomenclature and control the number of configuration optionsavailable in the modify configure mode.

Figure 8-1. Modify/View and Basic/Advanced States Screen Flow

T02408A

C O N FIG

PA SS W DC O N FIG

E X IT N O PAS SW O R DP ROTEC T FO R C O N FIG .

ADVA N C EDTO G G LE SE T

BA S ICTO G G LE

N OTE: W H E N E D IT IN G C O N FIG U R ATIO N /M O D IFY E N V IRO N M E N T AN D O N E O R M O R E C O N FIG U R ATIO N ITE M S H AV E BE EN C H AN G E D, A S AVE SC R EE N A P PE AR S.

S A M P LE

TO PLE V

TO P LE V

TO PLE V

M E AS U R EC A LIB R AT EO U T/H O LDC O N FIG U R ESE C U R ITYD IS PLAY

M E AS U R EC A LIB R AT EO U T/H O LDC O N FIG U R ESE C U R ITYD IS PLAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N F IG U R ES E C U R ITYD IS P LAY

M EA S U R EC ALIBR ATEO U T/H O LDC O N F IG U R ES E C U R ITYD IS P LAY

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

M O D IFY

V IE W

B A SIC

A DV N C D

68.2

68.2

68.2

68.2

YE S

YE S

YE S

YE S

N O

N O

N O

N O

m S /cm

m S /cm

m S /cm

m S /cm

M E A S U R E

M E A S U R E

C O N F IGE X IT

S AV ES C H AN G ES

D O ES N OT S AV EC H A N G E S

M E A SU R EC A LIB R AT EO U T/H O LDC O N FIG U R ESE C U R IT YD IS PLAYexit to

M E AS U R E

S AV E ?

68 .2

YE S N O

m S/cm

Page 48: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

MODIFY CONFIGURE STATES OF OPERATIONWBPEEUI520003A1 8 - 3

The basic programming mode contains a subset of configura-tion options found in the advanced mode. Separation into twoprogramming groups is advantageous when limited functional-ity is desired. Fewer options reduces confusion and the possi-bility of configuration errors.

When advanced programming is ordered, the basic/advancedprogramming toggle must be set in two locations:

• User state in the utility mode.• Modify configure state in the configure mode.

In order to select either the basic or advanced programmingmode in the modify configure state, the programming modemust be set to advanced in the user state.

Refer to Section 11 for more information on setting the userstate programming mode to advanced.

MODIFY CONFIGURE STATES OF OPERATION

Since the view configure state only displays the configuredoptions, the following sections focus on each modify configurestate and available options for these states. Refer to CONFIGU-RATION SEQUENCE for the procedures necessary to performthe modifications.

The modify configure state contains all the available settingsthat establish the functionality of the transmitter. Upon receiptof the transmitter, the default configuration (unless otherwisespecified by the customer) is used once the transmitter hasbeen powered.

Before installing the transmitter, modify the configuration toreflect the final installed application. The modify configurestates of operation define the sensor interface, output parame-ters, and diagnostic functionality.

The following sections contain descriptions of each modify con-figure state of operation. Figure 8-2 shows the screen flow forthe modify configure states of operation.

Analyzer State (Basic/Advanced)

The analyzer state determines the transmitter and sensor type.

Conductivity Analyzer State (Basic/Advanced)

The conductivity analyzer state contains three choices forfour-electrode sensor groups: A, B, and C.

Page 49: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

MODIFY CONFIGURE STATES OF OPERATION8 - 4 WBPEEUI520003A1

Figure 8-2. Modify Configure States Screen Flow

T02409A

A N A LZR

A N A LZRR E TU R N

TM P S NSR E TU R N

TC.TY P ER E TU R N

TM P U N TR E TU R N

O U TP U TR E TU R N

S AFE M DR E TU R N

SP IKER E TU R N

DA M P N GR ETU R N

TO P LE V C O N FIGEX IT

A DVA N C EDP RO G R AM M IN G

O N LYBA SIC

P RO G R AM M IN G O N LY

D IAG SR ETU R N

TM P.S N S

TC.TY P E

TM P U N T

O U TP U T

DA M P N G

D IAG S

S A FE M D

S PIK E

M E AS U R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS PLAY

M E AS U R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS PLAY

M E AS U R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS PLAY

M E AS U R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS PLAY

M E AS U R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS PLAY

M EA S U R EC A LIBR ATEO UT/H O LDC O N FIG U R ESE C U R ITYD IS PLAY

M EA S U R EC A LIBR ATEO UT/H O LDC O N FIG U R ESE C U R ITYD IS PLAY

M EA S U R EC A LIBR ATEO UT/H O LDC O N FIG U R ESE C U R ITYD IS PLAY

M EA S U R EC A LIBR ATEO UT/H O LDC O N FIG U R ESE C U R ITYD IS PLAY

exit toM E AS U R E

exit toM E AS U R E

exit toM E AS U R E

exit toM E AS U R E

exit toM E AS U R E

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

A N A LZR

TM P.S N S

TC.TY P E

TM P.U N T

O U TP U T

DA M P N G

D IAG S

S A FE .M D

S P K .O FF

68 .2

68.2

68.2

68.2

68.2

68.2

68.2

68.2

68.2

SE LEC T

SE LEC T

SE LEC T

SE LEC T

SE LEC T

S ELE C T

S ELE C T

S ELE C T

S ELE C T

N E X T

N EX T

N EX T

N EX T

N EX T

N E XT

N E XT

N E XT

N E XT

m S /cm

m S /cm

m S /cm

m S /cm

m S /cm

m S /cm

m S /cm

m S /cm

m S /cm

Page 50: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

MODIFY CONFIGURE STATES OF OPERATIONWBPEEUI520003A1 8 - 5

Concentration Analyzer State (Advanced)

The concentration analyzer state converts conductivity valuesto concentration units. This state applies temperature com-pensated conductivity measurements to a predefined oruser-defined function.

Temperature Sensor State (Basic/Advanced)

The temperature sensor state configures the temperatureinput for no temperature sensor (none), a Pt 100 RTD,three-kilohm Balco RTD, or 4.75-kilohm RTD network.

When none is selected, the temperature is fixed at 25-degreesCelsius and the temperature compensation is fixed at manual.Changing the fixed temperature value requires a temperaturecalibration.

Temperature Compensation State (Basic/Advanced)

Temperature has a marked effect on the conductance of manysolutions. The effect is generally nonlinear and dependent onthe particular ionic species and their concentration.

The transmitter software has a number of preprogrammed cor-rection algorithms that compensate the effect of temperatureon conductivity to a reference temperature of 25-degrees Cel-sius. This results in accurate and stable measurements whentemperature varies.

Options for temperature compensation are grouped into twosets: manual and auto. Manual temperature compensationcontains no additional options. It is locked to a specific pro-cess temperature independent of the selected temperaturesensor. If desiring a different process temperature, adjust itduring a temperature calibration.

Auto temperature compensation options are sensor groupdependent. They use temperature values measured by thetransmitter temperature input. Compensation algorithmsinclude: standard KCl (0.1N KCl based), solution coefficient,zero to 15-percent NaOH, zero to 20-percent NaCl, zero to18-percent HCl, zero to 20-percent H2SO4, pure water neutral,pure water trace acid, pure water trace base, anduser-defined.

The user-defined temperature compensation option requiresuncompensated conductivity data from the reference tempera-ture of 25-degrees Celsius to the maximum process tempera-ture on a representative sample of process solution. This datais used to calculate the ratio of uncompensated conductivity toconductivity at the reference temperature of 25-degrees Cel-sius. These ratios are then plotted against temperature.

Page 51: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

MODIFY CONFIGURE STATES OF OPERATION8 - 6 WBPEEUI520003A1

An example of a typical user-defined temperature compensa-tion plot is shown in Figure 8-3 and Table 8-1.

The curve is a nonlinear function divided into five line seg-ments. The endpoints represent points one and six. The break-points represent points two through five.

Refer to Appendix A for more information on temperaturecompensation.

Output State (Basic/Advanced)

The output state sets the output type and range. The defaultoutput range values are 10-percent of the full scale processvariable for Group A sensors, and full scale process variablefor Groups B and C sensors. If desiring a reverse-acting out-put, reverse the four and 20-milliamp process variable values.

Figure 8-3. User-Defined Temperature Compensation Curve

T02414A

200175150125100

TEM P ER ATU R E (°C )

K /K S TD

7550

1.40

1 .20

1 .00

0 .80

0 .60

0 .40

0 .20

0 .00250

B R E A K P O IN TS

AC TUAL N O N -L IN E A R R E LATIO N S H IP

Table 8-1. User-Defined Temperature Compensation Data

PointTemp.(°C)

Uncompensated Conductivity (mS/cm)

Conductivity Ratio (K/KSTD)

1 0 7.21 0.70

2 25 10.30 1.00

3 50 12.25 1.19

4 75 12.97 1.26

5 100 12.82 1.24

6 200 9.06 0.88

Page 52: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

CONFIGURATION LOCKOUTWBPEEUI520003A1 8 - 7

Damping State (Basic/Advanced)

The damping state applies a lag function on the output signaland reduces fluctuations caused by erratic process conditions.The damping value can be set from 00.0 to 99.9 seconds. Thisvalue represents the time required for the output to reachapproximately 63.2 percent of its final value in response to astep change in the process variable.

Diagnostics State (Basic/Advanced)

The diagnostics state allows disabling of the built-in sensordiagnostics.

Safe Mode State (Basic/Advanced)

The safe mode state determines the output level of the trans-mitter if an error condition occurs that renders the transmitterinoperable. The available states are fail low and fail high. Formore information on error conditions, refer to Section 12.

Spike State (Advanced)

The spike state sets the diagnostic spike level as a percent ofoutput. This level determines the magnitude of the spike.

When the spike has been set for any level greater than zeropercent and is enabled in the spike output state, the transmit-ter modulates the output signal by the configured level for oneout of every six seconds. Using this modulation, the transmit-ter informs the operator of a detected diagnostic condition. Formore information on error conditions, refer to Section 12.

CONFIGURATION LOCKOUT

The transmitter has a hardware lockout feature that, onceengaged, prohibits access to the configure mode. This featuredoes not affect parameters that can be changed in other modesof operation. Refer to CONFIGURATION SEQUENCE for theconfiguration lockout procedure.

CONFIGURATION SEQUENCE

Refer to Figures 8-4, 8-5, and 8-6 for the transmitter configu-ration sequence. Each block of the flow represents a singletask that must be completed before continuing with thesequence.

In some cases, more than one path can be taken during config-uration. For paths that are in parallel, either complete all ofthe tasks in all of the paths before continuing or complete all

Page 53: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

CONFIGURATION SEQUENCE8 - 8 WBPEEUI520003A1

of the tasks in only those paths that apply before continuing.At least one path must be completed.

Some blocks contain alphanumeric codes. These codes identifythe procedure that describes the steps to complete an indi-cated task. Complete all of the steps given in a procedurebefore continuing to the next procedure.

The procedures have check boxes in the margin by each proce-dural step. When performing a procedure, check each box aseach step is completed.

By treating each task as a separate entity, the procedures pro-vide an easy method for finding the information needed to per-form each task in the configuration sequence.

Figure 8-4. Configuration Sequence

V IEW IN GO R M O D IFY IN G

C O N FIG U R AT IO N ?

BA S IC O RA DVAN C ED

P RO G R A M M IN G ?

T02429A

PA SS W O R DPROTE C TE D ?

PAS S W O R DP ROTE C TE D ?

C H AN G EP RO G R A M M IN G

M O D E ?

M O D IFY IN G APA R A M E TER ?

V IEW IN G

B AS IC

M O D IFY IN G

ADVAN C ED

Y ESYE S

Y ES

N O

D O N E

P RO G R AM M IN GM O D E

C H A N G EP R 20

E N TE RPAS S W O R D

E N TE RPAS S W O R D

M O D IF Y IN GC O N FIG U R AT IO NW H ILE V IEW IN G

P R 19

N ON O

N O

Y ES

A B

STA RT

SE LEC TIN G V IEWO R M O D IFY

C O N FIG U R ATIO NS TATE PR 18

Page 54: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

CONFIGURATION SEQUENCEWBPEEUI520003A1 8 - 9

Figure 8-5. Advanced Configuration Sequence

T02432A

C O N D U C TIV ITYO R

C O N C E N TR ATIO NA N ALY ZER ?

TC.C O EFSE LEC TED ?

M A N UA LS E LE C TE D ?

C H A N G ETEM P ER ATU R E

SE TTIN G ?

U S R .D E FSE LEC TED ?

S E TH AR DWA R E

C O N FIG U R ATIO NLO C KO U T?

PR ED EFIN E DO R

U S ER -D E FIN E D ?

C O N C EN TR ATIO N

N O N ON O

N O

Y E SY E SYE S

Y ES

Y ES

N O

C O N D U C T IV ITY

P R E D EFIN E D

U S E R -D E FIN E D

D O N E

A

SE N S O RG RO U P

SE LEC TIO N

A N A LY ZERTY P E

SE LEC TIO N

PR 22

PR 21

C O N FIG U R ATIO NO PTIO N

S E LE C TIO NP R 23

TE M P ER ATUR EC O M PEN SATIO N

TYP E S ELE C TIO NP R 29

TE M P ER ATUR EC ALIB RATIO N

PR 8

TE M P E R ATU R ESE N S O R TYP E

SE LEC TIO NPR 28

U S ER -D EF IN E DTEM P ER ATU R EC O E FF IC IE N T

P RO G R A M M IN GPR 31

TEM P ER ATU R EC O M PE N S ATIO N

C O E FFIC IE N TS E TTIN G

P R 30

O U TP U TR AN G E

SE TTIN GP R 32

D IAG N O STIC SS E LE C TIO N

PR 33

C O N FIG U R ATIO NLO CKO U T

P R 36

SA FE M O D ESE LEC TIO N

S P IK EM AG N ITU D E

SE T TIN GP R 34 PR 35

DAM PIN GAD JU STM E N T

P R 13

E N G IN EE R IN GU N ITS

SE LEC TIO N

TE X T S TR IN GD EFIN IT IO N

PR 24

PR 26

D E C IM A L P O IN TLO C ATIO N

S E LE C TIO N

LIN E A R C U RV EFIT

PRO G R A M M IN G

PR 25

PR 27

Page 55: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION

CONFIGURATION SEQUENCE8 - 10 WBPEEUI520003A1

Figure 8-6. Basic Configuration Sequence

C H A N G ETE M P ER ATU R E

S ET TIN G ?

S ETH A R DWAR E

C O N FIG U R ATIO NLO C KO U T ?

T02433A

M AN UALS ELE C TE D ?

TC.C O E FSE LEC TE D ?

YE S

N O

N O N O

N O

D O N E

YE S YE S

C O N FIG U R ATIO NLO C KO U T

PR 36

S E N S O RG RO U P

S E LE C TIO NP R 28

TEM PE R ATU R EC O M P EN S ATIO N

TY PE SE LEC T IO N

TEM PE R ATU R ES E N S O R TY P E

S ELE C TIO NP R 29

TEM PE R ATU R EC O M PE N S ATIO NC O E FFIC IE N TSE TTIN G P R 30

YE S

T EM PE R ATU R EC A LIB R ATIO N

PR 8

O U TP U TR A N G E

SE TTIN GPR 32

DAM PIN GAD JU ST M E N T

P R 13

D IAG N O S TIC SS ELE C TIO N

P R 33

SA FEM O D E

SE LEC T IO NP R 34

B

A N A LYZE RTYP E

SE LEC T IO N

P R 22

P R 21

Page 56: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 9 - 1

SECTION 9 - SECURITY MODE

INTRODUCTION

The security mode of operation establishes password protec-tion against unauthorized changes to transmitter functions byunqualified personnel. Password protection can be assigned tothe calibrate and output/hold modes of operation, and modifyconfigure state of operation.

SECURITY STATE

The security mode of operation contains one state of operation.This state provides password protection of critical operatingenvironments. Each password protected mode or state of oper-ation can have its security state toggled on or off in the associ-ated security screen. All security assignments must be madebefore assigning a password.

The security of the security state itself is automatically set toon when one or more mode or state has the security on. Onepassword assignment applies to all secured modes and states.Figure 9-1 shows the screen flow for the security state ofoperation.

ENTERING PASSWORD

When the security state has been set, the password must beentered to gain access to the modes and states that have beenpassword protected.

1. When the password inquiry screen (PASSWD) appears:

a. The display reads _ _ _. Use the smart key to incre-ment the value of the blinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 1a and 1b for each digit.

2. Press the ENTER smart key to accept the password.

SECURITY SEQUENCE

There is only one procedure associated with the security mode.Refer to PR37 to set the security states and password.

Page 57: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SECURITY MODE

SECURITY SEQUENCE9 - 2 WBPEEUI520003A1

Figure 9-1. Security State Screen Flow

T02075B

M E AS U R E

N OPA SS W O R DP R OTE C TFO RSE C U R ITY PAS S W D

SE C U R

PA SS W D2

M E A S U R EC ALIB R ATEO U T/H O LDC O N F IG U R ES EC U R ITYD IS P LAY

M E A S U R EC ALIB R ATEO U T/H O LDC O N F IG U R ES EC U R ITYD IS P LAY

M E A S U R EC ALIB R ATEO U T/H O LDC O N F IG U R ES EC U R ITYD IS P LAY

M EA S U R EC ALIB RATEO U T /H O LDC O N F IG U R ES E C U R ITYD IS P LAY

M EA S U R EC ALIB RATEO U T /H O LDC O N F IG U R ES E C U R ITYD IS P LAY

M EA S U R EC ALIB RATEO U T /H O LDC O N F IG U R ES E C U R ITYD IS P LAY

exit toM E A S U R E

exit toM E A S U R E

exit toM EA S U R E

exit toM E A S U R E

exit toM EA S U R E

exit toM EA S U R E

C A LIB R

O U T/H O LD

C O N FIG

C A LIB R

O U T/H O LD

C O N FIG

O FF

O FF

O FF

O N

O N

O N

EN TE R

EN TE R

E N TE REN TE R

E N TE R

E N TE R

Page 58: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 10 - 1

SECTION 10 - SECONDARY DISPLAY MODE

INTRODUCTION

The Type TB82EC transmitter has two display regions activewhile in the measure mode of operation. The primary displayregion shows the measured variable. The secondary displayregion can show a multitude of process, sensor, or transmitterinformation: process temperature, current output value, sen-sor type, sensor group, compensated conductivity (useful forconcentration configurations), user-defined text description(concentration configuration only), spike mode status, andsoftware revision. All of these are viewable in the secondarydisplay region using the secondary display mode. Any of thesecan be set as the displayed value while in the measure mode ofoperation.

SECONDARY DISPLAY STATES OF OPERATION

The secondary display mode contains eight states. These pro-vide information on the process temperature, transmitter set-tings, and transmitter status. As shown in Figure 10-1, eachsecondary state can be sequentially viewed by pressing theNEXT smart key. Any given secondary display state can becontinually shown in the measure mode by pressing theENTER smart key when the desired state is shown. The trans-mitter proceeds to the measure mode and displays the enteredsecondary display state in the secondary display region.

SECONDARY DISPLAY OPERATION

To use the secondary display mode and states of operation:

1. Press the MENU smart key until DISPLAY is highlighted onthe display.

2. Press the SELECT smart key to enter the secondary displaymode.

3. Press the NEXT smart key to sequentially view each sec-ondary display state.

4. Press the ENTER smart key to have the desired secondarydisplay state appear during the measure mode.

Page 59: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SECONDARY DISPLAY MODE

SECONDARY DISPLAY OPERATION10 - 2 WBPEEUI520003A1

Figure 10-1. Secondary Display Mode Screen Flow

T02410A

S E C D SP

M EA SU R EC A LIB R AT EO U T /H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M EA SU R EC A LIB R AT EO U T /H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M EA SU R EC A LIB R AT EO U T /H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M EA SU R EC A LIB R AT EO U T /H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M EA SU R EC A LIB R AT EO U T /H O LDC O N FIG U R ESE C U R ITYD ISP LAY

M E A SU R EC A LIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD IS P LAY

M E A SU R EC A LIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD IS P LAY

M E A SU R EC A LIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD IS P LAY

M E A SU R EC A LIBR ATEO U T/H O LDC O N FIG U R ES E C U R ITYD IS P LAY

exit toM EA SU R E

exit toM EA SU R E

exit toM EA SU R E

exit toM EA SU R E

exit toM EA SU R E

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

exit toM EA S U R E

24°C

74°F

12.0 M A

4 E LE C 1

G R P A 1

100 M S

N AC L1

S P K .O FF 1

R E V.A 10

68.2

68.2

68.2

68.2

68.2

68.2

8 .2

8 .2

68.2

E N TE R

E N TE R

E N TE R

E N TE R

E N TE R

EN TER

EN TER

EN TER

EN TER

N E XT

N E X T

N E X T

N E X T

N E X T

N E XT

N E X T

N E X T

N E X T

m S/cm

m S/cm

m S/cm

m S/cm

m S/cm

m S/cm

%

%

m S/cm

M E AS U R E

JU M P TO LA ST S ELEC TE D, M AY BEA N Y O F TH E FO LLOW IN G S C R E EN S.

N OTE :1. O P TIO N C H O S EN D U R IN G C O N FIG U R ATIO N .

BA S ICC O N FIG U R ATIO N O N LY

Page 60: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 11 - 1

SECTION 11 - UTILITY MODE

INTRODUCTION

The Type TB82EC transmitter has a utility mode of operationthat provides access to powerful functions not usually neededduring normal operating conditions. These functions havebeen separated into two categories: factory and user. Factoryfunctions are reserved to ABB personnel. User functionsinclude: programming mode selection, reset configuration todefault values, remove security, reset all parameters to defaultvalues, and a self-test.

This section contains descriptions of each state in the utilitymode of operation. Refer to Table 11-1 for the proceduresneeded to perform utility mode tasks.

FACTORY AND USER STATES

Access the factory and user states of operation by using thehidden fifth key located directly above the NT in the ADVAN-TAGE text on the keypad. Once pressed, the hidden key causesthe textual prompt USER to display in the secondary displayregion. Pressing the SELECT smart key moves the transmitterinto the user state. Pressing the NEXT smart key moves thetransmitter to the factory selection. Pressing the exit to MEA-SURE smart key escapes back to the measure mode.

User State

The user state consists of five states of operation. Table 11-1describes the function of each state of operation and lists therelated procedures. There is no particular sequence for theseprocedures.

Table 11-1. User States

State Display Function Procedure

Mode MODE Sets programming mode (basic or advanced) that can be selected in the modify configure mode of operation.

PR38

Reset configuration RST.CON Resets configuration to factory defaults. PR39

Reset security RST.SEC Resets security to OFF state for all applicable modes and modify configure state.

PR40

Reset all RST.ALL Resets all programming parameters such as configuration, calibration, output/hold, security, and secondary display functions to factory defaults.

PR41

Soft boot RST.SFT Initiates self-test. PR42

Page 61: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

UTILITY MODE

FACTORY AND USER STATES11 - 2 WBPEEUI520003A1

The NEXT smart key sequentially moves through each of thefive user states. This cycle repeats until a state is selected orthe escape function is chosen using the exit to MEASUREsmart key. To select a state, press the SELECT smart key whenthe desired user state is shown in the secondary displayregion. Figure 11-1 identifies the smart key assignments andresulting action.

Figure 11-1. User State Screen Flow

T02411A

M E AS U R E

M O D E

M O D ER E TU R N

R S TC O NR E TU R N

R S TS ECR E TU R N

R S TA LLR E TU R N

U S E R

R STS FTR E TU R N

R S TC O N

R S TS EC

R S TA LL

R STS FT

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS P LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS P LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS P LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS P LAY

M E A SU R EC A LIB R ATEO U T/H O LDC O N FIG U R ES EC U R ITYD IS P LAY

exit toM E A SU R E

exit toM E A SU R E

exit toM E A SU R E

exit toM E A SU R E

exit toM E A SU R E

M O D E

R S T.C O N

R S T.S E C

R S T.A LL

R S T.S FT

68.2

68.2

68.2

68.2

68.2

S ELE C T

S ELE C T

S ELE C T

S ELE C T

S ELE C T

N E XT

N E XT

N E XT

N E XT

N E XT

m S/cm

m S/cm

m S/cm

m S/cm

m S/cm

Page 62: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

UTILITY MODE

FACTORY AND USER STATESWBPEEUI520003A1 11 - 3

Advanced/Basic Programming Mode User State

In order to simplify the configuration process when needing alimited amount of functionality, the transmitter contains twotypes of programming modes: basic and advanced. The pro-gramming mode is a nomenclature option.

The basic programming mode contains a subset of the featuresfound in the advanced programming mode. Reducing the avail-able features streamlines the configuration process. If thetransmitter is ordered with advanced programming, the basicor advanced programming mode can be used.

Contact ABB for information on advanced programming con-version for transmitters purchased with only basicprogramming.

Reset Configuration User State

The reset configuration user state returns the configuration ofthe transmitter back to factory default settings. Table 11-2summarizes the default software settings.

Reset Security User State

The reset security user state returns the security of the trans-mitter back to factory default settings. The factory defaults aresecurity off for all applicable modes and the modify configurestate.

Reset All User State

The reset all user state returns all transmitter values back tofactory defaults. This includes calibration, output/hold, con-figuration, security, and secondary display values.

Table 11-2. Factory Software Defaults

Parameter Default

Analyzer type Conductivity, Sensor Group A

Damping value 00.5 secs

Instrument mode Basic

Output range 0.00 to 199.9 mS/cm

Safe mode fail output state Low

Sensor diagnostics state Disable

Spike output level1 0%

Temperature compensation Manual

Temperature sensor 3-kΩ BalcoNOTE:1. This function only available in advanced programming mode.

Page 63: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

UTILITY MODE

FACTORY AND USER STATES11 - 4 WBPEEUI520003A1

Soft Boot User State

The soft boot user state initiates a self-test. All programmableinstrument parameters will be the same as those before initiat-ing a self-test.

Page 64: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 12 - 1

SECTION 12 - DIAGNOSTICS AND TROUBLESHOOTING

INTRODUCTION

The Type TB82EC transmitter performs a number of diagnos-tic checks on hardware, software, and sensor functions. Upondetection of a nonconforming condition, the FAULT icon alertsthe operator. Configurable remote indication is performed byusing the spike output option to modulate the output current.Pressing the FAULT info smart key interrogates the transmitteras to the cause of the fault. The display of a short text stringand fault code alternate on the secondary display. If multiplefaults exist, the FAULT info smart key moves the operator tothe next fault. Upon interrogation of all faults, the transmitterreturns to the measure mode and the MENU smart key iconenergizes.

This section describes the type of fault conditions and theirapplicability to transmitter functionality. If the evaluation ofthe fault codes does not resolve the problem, refer to TROU-BLESHOOTING SEQUENCE to find additional troubleshootingprocedures.

FAULT CODES

Fault conditions are grouped into two categories based ontheir severity. Conditions that result in degradation of trans-mitter performance are reported as problem codes (PC). Condi-tions that render the transmitter inoperable are reported aserror codes (EC).

Fault codes are reported in the secondary display region in aFIFO (first in, first out) order. All active fault conditions can beviewed at any time while in the measure mode by using theFAULT info smart key. A flashing FAULT icon indicates a newfault condition that has not been interrogated. Upon resolu-tion of all fault conditions, the FAULT icon and FAULT infosmart key are disabled.

Problem Codes

Problem codes result from fault conditions that impact theperformance of the transmitter. These conditions are usuallyresolved using standard practices.

The occurrence of a problem code fault condition energizes theFAULT icon and modulates the spike output. These diagnosticindicators provide local and remote reporting capability.

Page 65: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

FAULT CODES12 - 2 WBPEEUI520003A1

Table 12-1 lists common problem codes and Table 12-2 listsuncommon problem codes. Each entry lists the problem codenumber, displayed text string, a short description of the faultcondition, and corrective action. Most problem codes havemore than one corrective action listed. Perform the correctiveactions in the order they appear until the problem isresolved.

Table 12-1. Common Problem Codes

Code Text String Description Corrective Action

PC3 DRTY.SN Dirty sensor detected. Check cleanliness of sensor. If not clean, remove any foreign material. Clean sensor (PR46) and verify response to con-ductivity standards.

Electronically test sensor (PR44). Replace if it does not meet requirements.

Verify sensor wiring connections.

PC4 GND LP Ground loop present or shorted sensor cable

Verify sensor wiring connections.

Verify sensor does not have any exposed wire from nicks, etc. If it does, repair (if possible) or replace.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Verify sensor responds to conductivity standards. Replace sensor and/or sensor extension cable (if present) if sensor does not respond.

Electronically test sensor (PR44). Replace if it does not meet requirements.

PC6 HI.LOOP Current loop above upper range value (+0.4 mA hysteresis)

Verify process conditions are within configured output range. If PV is outside configured range, increase output range.

Verify transmitter is configured for correct temperature com-pensation type.

Verify sensor wiring connections.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Clean sensor and perform process calibration.

Conduct temperature calibration. If not using temperature sensor, verify configuration for TMP.SNS is NONE and proper process temperature is set.

Electronically test sensor and temperature compensator (PR44). Replace sensor if it does not meet requirements.

Page 66: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

FAULT CODESWBPEEUI520003A1 12 - 3

PC7 LO.LOOP Current loop below lower range value (-0.2 mA hysteresis)

Verify process conditions are within configured output range. If PV is outside configured range, increase output range.

Verify transmitter is configured for correct temperature com-pensation type.

Verify sensor wiring connections.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Clean sensor and perform process calibration.

Conduct temperature calibration. If not using temperature sensor, verify configuration for TMP.SNS is NONE and proper process temperature is set.

Electronically test sensor and temperature compensator (PR44). Replace sensor if it does not meet requirements.

PC8 HI.PV PV above transmitter range

Verify process conditions are within transmitter range. PV must be within transmitter range.

Verify sensor wiring connections.

Verify sensor does not have any exposed wire from nicks, etc. If it does, repair (if possible) or replace.

Remove any liquids, oils, scale, or corrosion from transmitter terminal block or extension cable junction box terminals.

Verify sensor responds to conductivity standards. Replace sensor and/or sensor extension cable (if present) if sensor does not respond.

Electronically test sensor (PR44). Replace if it does not meet requirements.

PC9 LO.PV PV below transmitter range

Verify process conditions are within transmitter range. PV must be within transmitter range.

Verify sensor wiring connections.

Verify sensor does not have any exposed wire from nicks, etc. If it does, repair (if possible) or replace.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Verify sensor responds to conductivity standards. Replace sensor and/or sensor extension cable (if present) if sensor does not respond.

Electronically test sensor (PR44). Replace if it does not meet requirements.

Table 12-1. Common Problem Codes (continued)

Code Text String Description Corrective Action

Page 67: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

FAULT CODES12 - 4 WBPEEUI520003A1

PC10 HI.TEMP Temperature above transmitter range

Verify process conditions are within transmitter range. PV must be within transmitter range.

Verify sensor wiring connections.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Clean sensor and perform process calibration.

Conduct temperature calibration. If not using temperature sensor, verify configuration for TMP.SNS is NONE and proper process temperature is set.

Electronically test sensor and temperature compensator (PR44). Replace sensor if it does not meet requirements.

PC11 LO.TEMP Temperature below transmitter range

Verify process conditions are within transmitter range. PV must be within transmitter range.

Verify sensor wiring connections.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Clean sensor and perform process calibration.

Conduct temperature calibration. If not using temperature sensor, verify configuration for TMP.SNS is NONE and proper process temperature is set.

Electronically test sensor and temperature compensator (PR44). Replace sensor if it does not meet requirements.

PC12 HI.T.AD Open or missing tem-perature sensor

Verify process conditions are within transmitter range. PV must be within transmitter range.

Verify sensor wiring connections.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Clean sensor and perform process calibration.

Conduct temperature calibration. If not using temperature sensor, verify configuration for TMP.SNS is NONE and proper process temperature is set.

Electronically test sensor and temperature compensator (PR44). Replace sensor if it does not meet requirements.

Replace conductivity input PCB assembly.

PC13 LO.T.AD Shorted temperature sensor

Verify process conditions are within transmitter range. PV must be within transmitter range.

Verify sensor wiring connections.

Remove any liquids, oils, scales, or corrosion from transmitter terminal block or extension cable junction box terminals.

Clean sensor and perform process calibration.

Conduct temperature calibration. If not using temperature sensor, verify configuration for TMP.SNS is NONE and proper process temperature is set.

Electronically test sensor and temperature compensator (PR44). Replace sensor if it does not meet requirements.

Replace conductivity input PCB assembly.

Table 12-1. Common Problem Codes (continued)

Code Text String Description Corrective Action

Page 68: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

FAULT CODESWBPEEUI520003A1 12 - 5

Table 12-2. Uncommon Problem Codes

Code Text String Description Corrective Action

PC20 BAD.SEE Bad SEEPROM or conductivity input PCB assembly

Input PCB assembly factory calibration con-stants can not be loaded. Calibrate sensor and order replacement conductivity input PCB assembly. Existing assembly should properly function until new assembly is received.

PC21 NO.F.CAL Missing factory calibration and func-tional SEEPROM

Contact ABB for factory calibration procedure. Calibrate sensor for short-term usage until factory calibration is performed or a new con-ductivity input PCB assembly is installed.

PC22 BLNK.uP Blank microprocessor EEPROM Cycle transmitter power.

Contact ABB.

PC25 ROM.SUM Incorrect EPROM checksum Cycle transmitter power.

Contact ABB.

PC30 R0.F.CAL Out of range or missing factory calibra-tion for conductivity circuit range zero

Contact ABB for factory calibration procedure. Calibrate sensor for short-term usage until factory calibration is performed.PC31 R1.F.CAL Out of range or missing factory calibra-

tion for conductivity circuit range one

PC32 R2.F.CAL Out of range or missing factory calibra-tion for conductivity circuit range two

PC33 R3.F.CAL Out of range or missing factory calibra-tion for conductivity circuit range three

PC34 R4.F.CAL Out of range or missing factory calibra-tion for conductivity circuit range four

PC35 G0.F.CAL Out of range or missing factory calibra-tion for ground loop circuit range zero

Contact ABB for factory calibration procedure. If ground loop fault is continuously active and can not be tolerated, disable diagnostics (PR33) until factory calibration is performed.

PC36 G1.F.CAL Out of range or missing factory calibra-tion for ground loop circuit range one

PC37 G2.F.CAL Out of range or missing factory calibra-tion for ground loop circuit range two

PC38 G3.F.CAL Out of range or missing factory calibra-tion for ground loop circuit range three

PC39 G4.F.CAL Out of range or missing factory calibra-tion for ground loop circuit range four

PC45 BA.F.CAL Out of range or missing factory calibra-tion for 3-kΩ Balco temperature compensator

Contact ABB for factory calibration procedure. Calibrate temperature sensor for short-term usage until factory calibration is performed.

PC46 PT.F.CAL Out of range or missing factory calibra-tion for Pt 100 temperature compensator

PC47 RT.F.CAL Out of range or missing factory calibra-tion for 4.75-kΩ RTD network tempera-ture compensator

Page 69: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

FAULT CODES12 - 6 WBPEEUI520003A1

Error Codes

Error codes result from fault conditions that render the trans-mitter inoperable. These conditions can not usually beresolved using standard practices.

PC50 R0.CHKS Incorrect or missing conductivity circuit range zero checksum

Contact ABB for factory calibration procedure. Calibrate sensor for short-term usage until factory calibration is performed.PC51 R1.CHKS Incorrect or missing conductivity circuit

range one checksum

PC52 R2.CHKS Incorrect or missing conductivity circuit range two checksum

PC53 R3.CHKS Incorrect or missing conductivity circuit range three checksum

PC54 R4.CHKS Incorrect or missing conductivity circuit range four checksum

PC55 G0.CHKS Incorrect or missing ground loop circuit range zero checksum

Contact ABB for factory calibration procedure. If ground loop fault is continuously active and can not be tolerated, disable diagnostics (PR33) until factory calibration is performed.

PC56 G1.CHKS Incorrect or missing ground loop circuit range one checksum

PC57 G2.CHKS Incorrect or missing ground loop circuit range two checksum

PC58 G3.CHKS Incorrect or missing ground loop circuit range three checksum

PC59 G4.CHKS Incorrect or missing ground loop circuit range four checksum

PC65 BA.CHKS Incorrect or missing 3-kΩ Balco tem-perature compensator checksum

Contact ABB for factory calibration procedure. Calibrate temperature sensor for short-term usage until factory calibration is performed.

PC66 PT.CHKS Incorrect or missing PT 100 tempera-ture compensator checksum

Contact ABB for factory calibration procedure. Calibrate temperature sensor for short-term usage until factory calibration is performed.PC67 RT.CHKS Incorrect or missing 4.75-kΩ RTD net-

work temperature compensator checksum

PC70 HI.G.L.AD Ground loop signal above transmitter A/D range

Conductivity input PCB assembly ground loop circuit failure exists. Disable diagnostics (PR33) and order replacement conductivity input PCB assembly. Existing conductivity input PCB assembly should function properly until new conductivity input PCB assembly is installed.

PC71 LO.G.L.AD Ground loop signal below transmitter A/D range

PC72 HI.D.S.AD Dirty sensor above transmitter A/D range

Conductivity input PCB assembly dirty sensor diagnostic circuit failure exists. Disable diag-nostics (PR33) and order replacement con-ductivity input PCB assembly. Existing conductivity input PCB assembly should func-tion properly until new conductivity input PCB assembly is installed.

PC73 LO.D.S.AD Dirty sensor below transmitter A/D range

Table 12-2. Uncommon Problem Codes (continued)

Code Text String Description Corrective Action

Page 70: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

CALIBRATION DIAGNOSTICSWBPEEUI520003A1 12 - 7

The occurrence of an error code fault condition energizes theFAULT icon and enables the safe mode output. When in thesafe mode, the current output is fixed high or low based on theconfiguration of the safe mode. These diagnostic indicatorsprovide local and remote reporting capability.

Table 12-3 contains all the error codes supported by the trans-mitter. Each entry lists the error code number, displayed textstring, and a short description of the fault condition.

When an error code appears on the transmitter display, thetransmitter must be replaced with one that is known to beoperable. Return the nonfunctional transmitter to ABB forrepair. Contact ABB for processing instructions.

CALIBRATION DIAGNOSTICS

The transmitter performs automatic efficiency and offset cal-culations. These calculations are relative to a theoretically per-fect conductivity and temperature sensor during eachcalibration cycle. Calibration history is retained for futureinterrogation using the edit calibrate state. The calibrationconstants displayed are slope and offset for the process vari-able and temperature.

A slope of less than 0.2 or greater than five for the processvariable indicates a potentially bad process calibration point orpoorly performing sensor. In these cases, the text stringBAD.CAL appears on the secondary display. The transmitterreturns to the beginning of the calibration cycle after it reportsthe bad calibration.

WARNING

All error conditions are considered catastrophic and requiretransmitter replacement. Replace the transmitter with a trans-mitter that is known to be operable. Leaving an inoperabletransmitter in the process could cause process upsets. Someprocess upsets can lead to dangerous conditions that canharm personnel and damage equipment.

Table 12-3. Error Codes

Error Code Text String Description

EC1 HI.PV.AD Overrange PV A/D

EC2 LO.PV.AD Underrange PV A/D

EC3 PH.PCB pH/ORP/pION input PCB with 4-electrode conductivity firmware

EC4 TC.PCB Toroidal conductivity input PCB with 4-electrode conductivity firmware

EC6 TE.PCB 2-electrode conductivity input PCB with 4-electrode conductivity firmware

EC7 EC.PCB Electrode conductivity input PCB with 4-electrode conductivity firmware

Page 71: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

ADDITIONAL DIAGNOSTICS12 - 8 WBPEEUI520003A1

An offset value of less than -20 or greater than +20 microsie-mens per centimeter for Sensor Group A, less than -4 orgreater than +4 microsiemens per centimeter for SensorGroup B, or less than -0.800 or greater than +0.800 microsie-mens per centimeter for Sensor Group C also indicates apotentially bad process calibration or poorly performing sen-sor. In these cases, the transmitter reports the bad calibrationand returns to the beginning of the calibration cycle.

The transmitter reports a bad temperature calibration andrejects calibration values for slope values less than 0.2 orgreater than 1.5 and offset values less than -40-degrees Cel-sius or greater than +40-degrees Celsius. Temperature calibra-tions use smart software routines that automatically adjustthe value for slope, offset, or both based on the calibrationvalue being entered and the calibration history if it exists.

ADDITIONAL DIAGNOSTICS

Other diagnostic messages may appear during transmitterprogramming. These messages include BAD.VAL (bad value),DENIED, and RAM.ERR (RAM error).

BAD.VAL indicates the attempted numeric entry of a value outof the allowed range of the transmitter. Table 1-3 lists thetransmitter range limits.

DENIED indicates incorrect entry of a security password.Section 9 contains information on security.

RAM.ERR indicates a RAM read/write error. The transmitterautomatically resets when this error has been encountered. Ifthe transmitter continues to reset, contact ABB for problemresolution.

TROUBLESHOOTING SEQUENCE

Refer to Figure 12-1 for the transmitter troubleshootingsequence. Each block of the flow represents a single task thatmust be completed before continuing with the sequence.

In some cases, more than one path can be taken during trou-bleshooting. For paths that are in parallel, either complete allof the tasks in all of the paths before continuing or completeall of the tasks in only those paths that apply before continu-ing. At least one path must be completed.

Some blocks contain alphanumeric codes. These codes identifythe procedure that describes the steps to complete an indi-cated task. Complete all of the steps given in a procedurebefore continuing to the next procedure.

Page 72: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DIAGNOSTICS AND TROUBLESHOOTING

TROUBLESHOOTING SEQUENCEWBPEEUI520003A1 12 - 9

The procedures have check boxes in the margin by each proce-dural step. When performing a procedure, check each box aseach step is completed.

By treating each task as a separate entity, the procedures pro-vide an easy method for finding the information needed to per-form each task in the troubleshooting sequence.

Figure 12-1. Troubleshooting Sequence

S E N S O RIN S P E C TIO N

P R 43

R E FE R TO P RO B LE MA N D E R RO R C O D E

TA B LE S IN TH ISS E C TIO N

E LE C TRO N ICS E N S O R

T ES TP R 44

R E S E TC A LIB R AT IO N

P R 12

TR A N S M ITTE RTRO U B LE S H O OTIN GW ITH O U T S E N S O R

P R 45

C O N TAC TTBI-B A ILE Y

C A LIBR ATETR A N S M ITTE R

R E P LAC ES E N S O R

FAU LTIC O N

D IS P LAY E D ?

IN S P E C TIO NPA S S ED ?

T02431A

S TA RT

Y E S

Y E S

N O

N O

D O N E

D O N E

D O N E

S TILLP RO B LE M S ?

S TILLP RO B LE M S ?

N O

N O

Y E S

Y E S

Page 73: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 13 - 1

SECTION 13 - MAINTENANCE

INTRODUCTION

This section contains a preventive maintenance schedule forthe Type TB82EC transmitter (Table 13-1). This table has aprocedure reference next to the task when applicable. The ref-erence indicates the procedure number where the procedurefor that task can be found.

Be sure to follow all warnings, cautions, and notes. Put boardscontaining semiconductors into antistatic bags when stored orshipped back to the factory. Do not repair printed circuitboards in the field. All repairs and adjustments should be per-formed by qualified personnel.

The maintenance of any stand-alone product or control systemaffects the reliability of that product. ABB recommends that allequipment users practice a preventive maintenance programthat will keep the equipment operating at an optimum level.

The procedures referred to in this section contain instructionsthat the customer should be able to perform on site. Thesepreventive maintenance procedures should be used as aguideline to assist in establishing good preventive mainte-nance practices. Select the minimum steps required to meetthe cleaning needs of your system.

Personnel performing preventive maintenance should meet thefollowing qualifications:

• Maintenance personnel should be qualified electrical tech-nicians or engineers that know the proper use of testequipment.

• Maintenance personnel should be familiar with the trans-mitter and have experience working with process controlsystems.

PREVENTIVE MAINTENANCE SCHEDULE

Table 13-1 is the preventive maintenance schedule for theType TB82EC transmitter. The table lists the preventive

WARNING

Allow only qualified personnel (refer to INTENDED USER inSection 1) to commission, operate, service, or repair thisequipment. Failure to follow the procedures described in thisinstruction or the instructions provided with related equipmentcan result in an unsafe condition that can injure personnel anddamage equipment.

Page 74: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

MAINTENANCE

PREVENTIVE MAINTENANCE SCHEDULE13 - 2 WBPEEUI520003A1

maintenance tasks in groups according to their specifiedmaintenance interval. Some tasks in Table 13-1 are selfexplanatory. Instructions for tasks that require further expla-nation are found in the procedures or in the documentationsupplied with any associated equipment.

Table 13-1. Preventive Maintenance Schedule

Task Procedure Frequency (months)

Check and clean all wiring and wiring connections. — 12

Calibrate transmitter output. PR10

Inspect sensor. PR43 As required

Clean sensor. PR46

Clean keypad. PR47

Calibrate transmitter sensor input. Fig. 6-2

Clean and lubricate all gaskets and O-rings. — Each time seals are broken

Complete all tasks in this table. — Shutdown

Page 75: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 14 - 1

SECTION 14 - REPAIR AND REPLACEMENT

INTRODUCTION

Due to the modular design of the Type TB82EC transmitter,the replacement of an assembly can be easily completed.Replacements are available for each major assembly. Theseinclude the input PCB, microprocessor PCB, power supplyPCB, front bezel, shell, and rear cover assemblies.

This section does not contain repair instructions for the sen-sor. Refer to Section 12 for diagnostics and troubleshootinginformation. Due to the nature of its design, if the trouble-shooting procedures do not solve the problem, complete sensorreplacement is required when it has been damaged or does notproperly function.

REPAIR AND REPLACEMENT SEQUENCE

Refer to Figure 14-1 for the repair sequence for the transmit-ter. Each block of the flow represents a single task that mustbe completed before continuing with the sequence.

In some cases, more than one path can be taken during repair.For paths that are in parallel, either complete all of the tasksin all of the paths before continuing or complete all of thetasks in only those paths that apply before continuing. At leastone path must be completed.

Some blocks contain alphanumeric codes. These codes identifythe procedure that describes the steps to complete an indi-cated task. Complete all of the steps given in a procedurebefore continuing to the next procedure.

The procedures have check boxes in the margin by each proce-dural step. When performing a procedure, check each box aseach step is completed.

WARNING

Do not substitute any components other than those listed inthe appropriate procedures. Doing so will compromise the cer-tification listed on the transmitter nameplate. Invalidating thesecertifications can lead to unsafe conditions that can injure per-sonnel and damage equipment.

Do not disconnect equipment unless power has been switchedoff at the source or the area is known to be nonhazardous. Dis-connecting equipment in a hazardous location with sourcepower on can produce an ignition-capable arc that can injurepersonnel and damage equipment.

Page 76: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

REPAIR AND REPLACEMENT

REPAIR AND REPLACEMENT SEQUENCE14 - 2 WBPEEUI520003A1

By treating each task as a separate entity, the procedures pro-vide an easy method for finding the information needed to per-form each task in the repair sequence.

Figure 14-1. Repair and Replacement Sequence

C O N D U C TIV ITYIN PU T P C BR EM OVA L

POW E R S U P PLYP C B

IN STA LLATIO N

C O N D U C TIV ITYIN PU T P C B

IN STA LLATIO N

M IC RO P RO C E SS O RP C B

R E PLACE M EN T

FRO N T BE ZE LIN STA LLATIO N

FRO N T BE ZE LR EM OVA L

P R 48

FRO N T B EZELR E M OVA L

PR 48

R E AR C OV E RR E PLAC EM E N T

P R 55

P R 50

P R 53

POW E R S U P PLYPC B R E M OVA L

P R 49

P OW E R SU PP LYP C B R EM OVA L

P R 49

P OW E R SU PP LYP C B

IN S TA LLATIO NP R 53

FRO N T B E ZE LIN STA LLATIO N

P R 54

C O N D U C TIV ITYIN PU T P C BR E M OVA L

C O N D U C TIV ITYIN PU T P C B

IN STA LLATIO N

P R 50

P R 52

P R 52

P R 51

P R 54

P OW E RS U P PLY O R

C O N D U C TIV ITYIN P U T P CB ?

R E P LAC EW H IC H

C O M P O N E N T?

T02430AD O N E

P OW E RS U P PLY P C B

IN P U T P C B

A B C D

R E AR C OV E RR EP LAC E M EN T

FRO N T BE ZE LR E M OVA L

FRO N T BE ZE LIN STALLATIO N

S H ELLR EP LAC E M EN T

PR 56

PR 55

PR 48

PR 54

STA RT

R EA R C OV ER

S H E LL

P OW E R S U PP LY P C B

IN PU T P C B

M IC RO PRO C E S SO R P C B

FRO N T B EZEL

C

D

B

B

A

A

R EM OV EP OW E R

Page 77: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 15 - 1

SECTION 15 - SUPPORT SERVICES

INTRODUCTION

Figure 15-1 is an assembly drawing of the Type TB82EC transmitter. Whenordering replacement parts, specify nomenclature type, part name, and partnumber of spare parts kits.

ABB is ready to assist in the use and repair of its products at any time.Requests for sales and/or application service should be made to the nearestsales or service office.

Factory support in the use and repair of the Type TB82EC transmitter can beobtained by contacting:

ABB Inc.9716 S. Virginia St., Ste.EReno, Nevada 89511USAPhone: +1 (775) 850-4800FAX: +1 (775) 850-4808Web site: http://www.abb.com

RETURN MATERIALS PROCEDURES

If any equipment should need to be returned for repair or evaluation, pleasecontact ABB at (775) 883-4366, or your local ABB representative for theRMA (return materials authorization) number. At the time the RMA numberis given, repair costs will be provided, and a customer purchase order will berequested. The RMA and purchase order numbers must be clearly marked onall paperwork and on the outside of the return package container.

Equipment returned to ABB with incorrect or incomplete information mayresult in significant delays or nonacceptance of the shipment.

REPLACEMENT PARTS

When making repairs at your facility, order spare parts kits from an ABBsales office. Provide this information.

1. Spare parts kit description, part number, and quantity.

2. Model and serial number (if applicable).

3. ABB instruction number, page number, and reference figure that identi-fies the spare parts kit.

Page 78: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SUPPORT SERVICES

RECOMMENDED SPARE PARTS KITS15 - 2 WBPEEUI520003A1

When ordering standard parts from ABB, use the part num-bers and descriptions from RECOMMENDED SPARE PARTSKITS. Order parts without commercial descriptions from thenearest ABB sales office.

NOTE: Contact ABB for replacement sensors. Due to the specialnature of these items, factory consultation is required.

RECOMMENDED SPARE PARTS KITS

Table 15-1 lists the recommended spare parts kits.

Figure 15-1. Exploded View

FRO N T B EZ ELK IT

SH E LL K IT

1 /2 -IN . L IQ U ID T ITEFITT IN G K IT

M IC RO PR O C ES S O RPC B A S SE M B LY K IT

POW ER S U P PLYPC B A S SE M B LY

C O N D U C TIV ITY IN P U TP C B A S S EM BLY K IT T02412A

Table 15-1. Spare Parts Kits

Part Number Description

4TB5001-0150 Rear label, TB82EC

4TB9515-0123 Panel mount kit

4TB9515-0124 Pipe mount kit

4TB9515-0125 Hinge mount kit

4TB9515-0155 Power supply PCB assembly kit

4TB9515-0156 Wall mount kit

4TB9515-01571 Power supply PCB assembly kit for HART compatible transmitters

4TB9515-0158 Power supply PCB assembly with lightning arrestor kit

4TB9515-01591 Power supply PCB assembly with lighting arrestor kit for HART compatible transmitters

4TB9515-0160 Front bezel kit

4TB9515-0162 Rear cover kit

Page 79: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SUPPORT SERVICES

RECOMMENDED SPARE PARTS KITSWBPEEUI520003A1 15 - 3

4TB9515-0163 ½-in. liquid-tite cable grip fitting kit

4TB9515-0175 Shell kit

4TB9515-0176 Input PCB assembly kit (4-electrode conductivity)

4TB9515-0177 ROM (read only memory) chip, TB82EC

4TB9515-0178 Microprocessor PCB/display board

4TB9515-0181 Front bezel kit, FM approvedNOTE:1. Contact factory for availability.

Table 15-1. Spare Parts Kits (continued)

Part Number Description

Page 80: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 A - 1

APPENDIX A - TEMPERATURE COMPENSATION

INTRODUCTION

The Type TB82EC transmitter has a variety of standard con-ductivity temperature compensation options. These include:

• Manual (0.1N KCl based).• Standard automatic (0.1N KCl based).• Temperature coefficient (0.00 to 9.99%/°C).• 0 to 15% NaOH.• 0 to 20% NaCl.• 0 to 18% HCl.• 0 to 20% H2SO4.• User-defined.

Additionally, three specialized types of automatic temperaturecompensation are available for the measurement of pure waterusing a Sensor Group C configuration. These types includeneutral salt, trace base, and trace acid.

The concentration analyzer type configuration offers the sametemperature compensation options as does the standard con-ductivity analyzer configuration.

TEMPERATURE COMPENSATION TYPES

Both the conductivity and concentration analyzer types offer11 types of temperature compensation. Manual temperaturecompensation is based on 0.1N KCl and has a reference tem-perature of 25-degrees Celsius.

Automatic temperature compensation can be set to one of sev-eral types. When configured for automatic temperature com-pensation, the transmitter measures the process temperaturevia the temperature compensator located either in the sensoror external to the sensor. It automatically adjusts the raw con-ductivity to a conductivity referenced to 25-degrees Celsius.

Standard KCl

The standard KCl temperature compensation option charac-terizes the temperature effect of 0.1N KCl. The data break-points are listed in Table A-1. The value KREF in Table A-1 isthe conductivity at 25-degrees Celsius. The Value KT in TableA-1 is the conductivity at temperature T.

Page 81: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE COMPENSATION

TEMPERATURE COMPENSATION TYPESA - 2 WBPEEUI520003A1

Temperature Coefficient

The temperature coefficient option allows for a fixed correctionbased on a percentage change of the reference conductivity(conductivity at 25-degrees Celsius) per degree Celsius. Thetemperature compensation factor is derived from the equation:

where:

Typical ranges for temperature compensation coefficients are:

• Acids: 1.0 to 1.6%/°C.• Bases: 1.8 to 2.0%/°C.• Salts: 2.2 to 3.0%/°C.• Neutral water: 2.0%/°C.

NaOH

The zero to 15-percent NaOH compensation option character-izes an average temperature correction required to cover a zero

Table A-1. Standard KCl Data Breakpoints

Temperature (°C) KREF/KT

0 1.80

5 1.57

10 1.38

15 1.22

20 1.10

25 1.00

30 0.91

50 0.69

75 0.50

100 0.38

128 0.30

156 0.25

306 0.18

α TC.COEF

KT

KREF-------------- 1.0– 100.0×

T 25.0–-------------------------------------------------------= =

α and TC.COEF Percentage change in reference con-ductivity per degree Celsius.

KT Conductivity at temperature T (°C).

KREF Conductivity at 25°C.

T Temperature of solution (°C)

Page 82: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE COMPENSATION

TEMPERATURE COMPENSATION TYPESWBPEEUI520003A1 A - 3

to 15-percent NaOH concentration range. Since NaOH has arelatively constant set of temperature coefficients over a largerange of concentrations, use this compensation type for weakas well as concentrated solutions of NaOH. The data breakpoints are listed in Table A-2. The value KREF is the conductiv-ity at 25-degrees Celsius. The value KT is the conductivity attemperature T.

NaCl

The zero to 20-percent NaCl compensation option character-izes an average temperature correction required to cover a zeroto 20-percent NaCl concentration range. Since NaCl has a rela-tively constant set of temperature coefficients over a largerange of concentrations, use this compensation type for weakas well as concentrated solutions of NaCl. The data break-points are listed in Table A-3. The value KREF is the conductiv-ity at 25-degrees Celsius. The value KT is the conductivity attemperature T.

HCl

The zero to 18-percent HCl compensation option characterizesan average temperature correction required to cover a zero to18-percent HCl concentration range. Since HCl has a relativelyconstant set of temperature coefficients over a large range ofconcentrations, use this compensation type for weak as well asconcentrated solutions of HCl. The data breakpoints are listed

Table A-2. NaOH Data Breakpoints

Temperature (°C) KREF/KT

0 1.79

25 1.00

50 0.69

75 0.53

100 0.43

156 0.30

Table A-3. NaCl Data Breakpoints

Temperature (°C) KREF/KT

0 1.75

25 1.00

50 0.66

75 0.47

100 0.35

140 0.25

156 0.23

Page 83: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE COMPENSATION

TEMPERATURE COMPENSATION TYPESA - 4 WBPEEUI520003A1

in Table A-4. The value KREF is the conductivity at 25-degreesCelsius. The value KT is the conductivity at temperature T.

H2SO4

The zero to 20-percent H2SO4 compensation option character-izes an average temperature correction required to cover a zeroto 20-percent H2SO4 concentration range. The data break-points are listed in Table A-5. The value KREF is the conductiv-ity at 25-degrees Celsius. The value KT is the conductivity attemperature T.

User-Defined

The user-defined temperature compensation option allows theentering of six break point values for KT/KREF. The value KREFis the conductivity at 25-degrees Celsius. The value KT is theconductivity at temperature T. Choose each break point valueto provide the closest fit of each linear segment to the actualtemperature in degrees Celsius versus the KT/KREFrelationship.

Pure Water

There are three pure water temperature compensation optionsavailable when using a Sensor Group C configuration. Theseinclude neutral salt (NEUTRL), trace acid (ACID), and trace

Table A-4. HCl Data Breakpoints

Temperature (°C) KREF/KT

0 1.75

25 1.00

50 0.66

75 0.47

100 0.35

140 0.25

156 0.23

Table A-5. H2SO4 Data Breakpoints

Temperature (°C) KREF/KT

0 1.37

25 1.00

50 0.84

75 0.73

100 0.67

156 0.61

Page 84: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE COMPENSATION

TEMPERATURE COMPENSATION TYPESWBPEEUI520003A1 A - 5

base (BASE). The reference temperature for these types of tem-perature compensation is 25-degrees Celsius.

Temperature compensation for pure water is a polynomialbased on data from T.S. Light. This equation compensates forvariations in conductivity due to pure water. Temperaturecompensation for the effect of pure water becomes insignifi-cant compared to the effects brought about by the solute forwater having a conductivity value of greater than 0.5 microsie-mens per centimeter. Using pure water temperature compen-sation for water with a conductivity greater than0.5 microsiemens per centimeter will not cause errors; how-ever, the compensation on solute effects (neutral salt, traceacid, trace base) may not accurately adjust for effects causedby the process liquid.

The polynomial is:

The pure water coefficient values (K0, K1, K2, etc.) are listed inTable A-6.

Factor K0 K1 T K2 T 2 K3 T 3 K4 T 4 K5 T 5 K6 T 6×+×+×+×+×+×+=

Table A-6. Pure Water Coefficient Values

Coefficient Pure Water Salt Trace Acid Trace Base

K0 1.170848E-02 0.532688 0.7000 0.5700

K1 9.101055E-04 1.439182E-02 0.0120 0.0172

K2 2.132244E-07 2.852080E-04 — —

K3 4.548839E-07 -6.504617E-06 — —

K4 -4.042016E-11 9.640603E-08 — —

K5 0.0 -6.982205E-10 — —

K6 0.0 1.887667E-12 — —

Page 85: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 B - 1

APPENDIX B - CONCENTRATION PROGRAMMING

INTRODUCTION

The concentration analyzer configuration has four specific sol-ute options and one user-defined option. The specific soluteoptions are:

• 0 to 15% NaOH.• 0 to 20% NaCl.• 0 to 18% HCl.• 0 to 20% H2SO4.

When needing a custom concentration configuration, selectthe user-defined option. This provides a six-point, five-seg-ment linear approximation of a specific conductivity-to-con-centration curve. When using the user-defined option, choosecustom units by either selecting one of the three engineeringunit icons or enter a six-character, alphanumeric string.

USER-DEFINED CONDUCTIVITY-TO-CONCENTRATION CURVES

The transmitter may be used in any range and with any ABBconductivity sensor when using the concentration analyzeroption. This allows infinite programming capability.

The user-defined option allows the characterization of conduc-tivity-to-concentration curves. These curves are determinedseparately in a laboratory or from published data such as thatin the International Critical Tables. These curves are thensegmented into five straight lines and programmed into thetransmitter.

During this procedure, some rules must be followed.

• Point 1 for both conductivity and concentration is alwaysthe zero-percent (four-milliamp) output point.

• Point 6 for both conductivity and concentration is alwaysthe 100-percent (20-milliamp) output point.

• All conductivity points must be ascending. Concentrationpoints can be either ascending or descending.

• If a reverse acting output is desired, swap the output rangevalues either in the modify configure state or the outputmode.

• The output range (four and 20-milliamp output range) cannot exceed the point one and point six concentration

Page 86: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONCENTRATION PROGRAMMING

PREPROGRAMMED CONDUCTIVITY-TO-CONCENTRATION CURVESB - 2 WBPEEUI520003A1

range; however, the output range is compressible using thererange function while in the output/hold mode.

The engineering units that appear on the primary display areuser-defined. Select either PPM, PPB, or % to appear in the pri-mary display, or enter a six-character, alphanumeric string.This string appears permanently or temporarily on the second-ary display.

PREPROGRAMMED CONDUCTIVITY-TO-CONCENTRATION CURVES

Processes with fixed solute types allow the selection of one ofthe four predefined solute options.

The information provided in the remainder of this appendixpertains to these four options. It was compiled and extrapo-lated from the International Critical Tables. All data andcurves are referenced at 25-degrees Celsius. The informationwas curve-fit to simple equations for use in the transmitter.They only approximate the actual concentration curves. Ifrequiring improved accuracy, especially in a narrow region ofconcentration, ABB recommends use of the user-definedoption and manual entering of the data.

NaOH

The zero to 15-percent NaOH option is characterized by thedata given in Table B-1 and the curve shown in Figure B-1.

Table B-1. NaOH Conductivity-To-Concentration Data

Conductivity(mS/cm)

Concentration (Weight %)

0.0 0.0

140.0 3.0

255.0 6.0

331.0 9.0

398.0 12.0

410.0 15.0

Page 87: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONCENTRATION PROGRAMMING

PREPROGRAMMED CONDUCTIVITY-TO-CONCENTRATION CURVESWBPEEUI520003A1 B - 3

NaCl

The zero to 20-percent NaCl option is characterized by thedata given in Table B-2 and the curve shown in Figure B-2.

Figure B-1. NaOH Conductivity-To-Concentration Curve

T02415A

500

C O N D U C TIV ITY (m S /C M )

W E IG H TP ER C EN T

N aO H

400300200

16

14

12

10

8

6

4

2

01000

Table B-2. NaCl Conductivity-To-Concentration Data

Conductivity (mS/cm) Concentration (Weight %)

0.0 0.0

34.3 2.0

119.7 8.0

172.6 12.0

207.6 16.0

231.9 20.0

Figure B-2. NaCl Conductivity-To-Concentration Curve

T02416A

250

C O N D U C TIV ITY (m S/C M )

W E IG H TP ER C E N T

N aC l

200150100

20

18

16

14

12

10

8

6

4

2

0500

Page 88: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONCENTRATION PROGRAMMING

PREPROGRAMMED CONDUCTIVITY-TO-CONCENTRATION CURVESB - 4 WBPEEUI520003A1

HCl

The zero to 18-percent HCl option is characterized by the datagiven in Table B-3 and the curve shown in Figure B-3.

H2SO4

The zero to 20-percent H2SO4 option is characterized by thedata given in Table B-4 and the curve shown in Figure B-4.

Table B-3. HCl Conductivity-To-Concentration Data

Conductivity (mS/cm) Concentration (Weight %)

0.0 0.0

365.0 4.0

625.0 8.0

755.0 12.0

820.0 15.0

850.0 18.0

Figure B-3. HCl Conductivity-To-Concentration Curve

T02417A

1000

C O N D U C TIV ITY (m S /C M )

W E IG H TP E R C E N T

H C l

800600400

18

16

14

12

10

8

6

4

2

02000

Table B-4. H2SO4 Conductivity-To-Concentration Data

Conductivity (mS/cm) Concentration (Weight %)

0.0 0.0

190.0 4.0

355.0 8.0

499.0 12.0

618.0 16.0

710.0 20.0

Page 89: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONCENTRATION PROGRAMMING

PREPROGRAMMED CONDUCTIVITY-TO-CONCENTRATION CURVESWBPEEUI520003A1 B - 5

Figure B-4. H2SO4 Conductivity-To-Concentration Curve

T02418A

800700600500

C O N D U C T IV ITY (m S /C M )

W EIG H TP E R C E N T

H SO2 4

400300200

20

18

16

14

12

10

8

6

4

2

01000

Page 90: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 C - 1

APPENDIX C - PROGRAMMING TEXT STRING GLOSSARY

INTRODUCTION

When programming the transmitter, the six-character, alpha-numeric region displays a wide variety of text prompts. Inmany cases, these prompts are abbreviations or portions ofwords.

TEXT PROMPTS

Table C-1 lists the text prompts and their full text equivalents.

Table C-1. Text Prompt Equivalents

Text Prompt Equivalent

20MA.PT 20-mA point

3K.BLCO 3-kΩ Balco (temperature compensation)

4.75K.RTD 4.75-kΩ RTD network

4MA.PT 4-mA point

A GRP Sensor Group A

AAAAAA Alphanumeric entry

ACID Acid

ADVNCD Advanced programming state

ANALZR Analyzer state

AUTO Automatic temperature compensation (Nernstian)

B GRP Sensor Group B

BAD.CAL Bad calibration - entered value caused calculated values to exceed maximum values

BAD.VAL Bad value - entered value exceeded maximum or minimum allowable value for entered parameter

BASE Base

BASIC Basic programming state

C GRP Sensor Group C

CALIBR Calibrate mode

CON.CAL Conductivity or concentration calibration state

CONCEN Concentration

COND Conductivity

CONFIG Configure mode

D.P. POS Decimal point position

DAMPNG Damping state

DENIED Incorrect security password entered

DIAGS Diagnostic state

DISABL Disable

EDT.CAL Edit calibration state

Page 91: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PROGRAMMING TEXT STRING GLOSSARY

TEXT PROMPTSC - 2 WBPEEUI520003A1

FAIL.HI Fail high (20 mA)

FAIL.LO Fail low (4 mA)

H2SO4 Sulfuric acid

HCL Hydrochloric acid

HLD.LVL Hold level

HOLD Hold state

K1/K25 Conductivity at temperature Point 1 to reference conductivity at 25°C. Points 2 through 6 represented in same manner.

MODIFY Modify configure state

NACL Sodium chloride

NAOH Sodium hydroxide

NEUTRL Neutral

NEW.VAL New calibration value. The PV or temperature value expected during a PV or temperature calibration.

NEW.VL.C New value in °C

NO D.P. No decimal point

NO.ICON No icon desired in primary display

NONE None

OUT.CAL Output calibration state

OUTPUT Output mode

PASSWD Security password

PT 100 Pt 100 RTD

PUR.H2O Pure water

REL.HLD Release hold

RERANG Rerange state

RESET? Conduct a reset operation?

REV.A10 Software revision A10

RST.ALL Reset all parameters to factory settings

RST.CAL Reset calibration constant and data to factory settings

RST.CON Reset configurations to factory settings

RST.SEC Reset security — remove any existing security

SAFE.MD Safe mode state

SAVE? Save the configuration?

SEC.DSP Secondary display mode

SECS Seconds

SECUR Security mode

SLF.TST Self test

SLOPE Slope

SPIKE Spike output state

SPK.MAG Spike output magnitude

SPK.OFF Spike output function set to off (disabled)

Table C-1. Text Prompt Equivalents (continued)

Text Prompt Equivalent

Page 92: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PROGRAMMING TEXT STRING GLOSSARY

FLOW TREEWBPEEUI520003A1 C - 3

FLOW TREE

Figure C-1 is a function flow tree for the Type TB82ECtransmitter.

STABL? Is the displayed process variable stable?

T.OFF°C Temperature offset in °C

TC.COEF Temperature compensation coefficient

TC.TYPE Temperature compensation type state

TMP.CAL Temperature calibration state

TMP.SLP Temperature slope

TMP.SNS Temperature sensor type state

U.D.UNIT User-defined engineering units

UNITS Engineering units

VIEW View configure state

X1.COND Conductivity independent variable value for break point 1 in conductivity units. Points 2 through 6 represented in same manner.

Y1.COND Concentration dependent variable value for break point 1 in concentration units. Points 2 through 6 represented in same manner.

Table C-1. Text Prompt Equivalents (continued)

Text Prompt Equivalent

Page 93: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PROGRAMMING TEXT STRING GLOSSARY

FLOW TREEC - 4 WBPEEUI520003A1

Figure C-1. Function Flow Tree

T02419A

M E A S U R E

C A LIB R ATE

C O N D U C TIV ITY /C O N C E N TR ATIO N

C A LIBR ATIO N

T E M P E R AT U R ESE N S O R

N O N E, 3 -k B ALC O,PT 100 , 4 .75-k

N E TW O R K

ΩΩ

T E M P E R AT U R EC O M P E N S ATIO N

M A N UA L, AU TO,S TD KC l, N aO H , N aC l,

H C l, H S OU SE R -D E FIN ED

P U R E W ATE RN E U TR A L S ALT, TR AC E AC ID,

TR AC E B AS E

2 4

O U T P U T R A N G E4-m A PO IN T

20-m A P O IN T

A N A LY ZE RC O N D U C TIV ITY,

S E NS O R G RO U P

C O N C EN T R ATIO NN aO HN aC lH C l

H S OU SE R -D E FIN ED

2 4

A , B , C

R E R A N G E O U T P U T4 m A P O IN T

20 m A P O IN T

O U T/H O LD

H O LD /R E LE A S EH O LD O U TP U T

DA M P IN GS E C O N D S

DA M P IN GS E C O N D S

D IAG N O S T IC SO N /O FF

S A F E M O D EH IG H /LOW

S P IK E N OTIF IC ATIO NP ER C EN T M AG N ITU D E

S P IK E O U TP U TO N /O FF

TE M P ER ATU R EC A LIBR ATIO N

O U T PU TC A LIBR ATIO N

N OTE : FU N C TIO N S IN ITA LIC S A R E O N LY AVA ILA BLE O N V E R SIO N S W ITH A DVAN C ED P RO G R A M M IN G N O M EN C LATU R E O PT IO N .

C O N FIG U R E

M O D IF Y /V IE W

B A S IC /A DVA N C E D

S E C U R ITY

S E C U R E D M O D ESC A LIBR ATE

O U TPU TC O N F IG U R E

D IS PLAY

TE M P E R ATU R E (C )

E D IT C A LIB R AT IO N

R E S E T C A LIBR AT IO N

TE M PE R AT U R E (F )

PA S SW O R DC U R R E N T O U TP U T

(m A )

D IAG N O S TICA LA R M S

S E N S O R T Y P E

S E N S O R IN P U T(m S /cm , uS /cm )

S E N S O R G RO U P

S P IK E O U TP U TS TATE

C O N C E N T R ATIO NA LP H A-N U M E R IC

S T R IN G

S O F TW A R ER E V IS IO N

Page 94: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WorksheetWorksheetWorksheetWorksheet TB82EC

Type TB82EC Advantage SeriesConductivity/Concentration Transmitter

TAG ______________ DATE ______________

PROGRAMMING MODE Basic Advanced

ANALYZER TYPE _____________________

Conductivity

Sensor Group A B C

Concentration

Sensor Group A B C 0 to 15% NaOH 0 to 20% NaCl 0 to 18% HCl 0 to 20% H2SO4 User-defined

Engineering units _________________COND1 _______ CONC1 _______COND2 _______ CONC2 _______COND3 _______ CONC3 _______COND4 _______ CONC4 _______COND5 _______ CONC5 _______COND6 _______ CONC6 _______

TEMPERATURE SENSOR None 3-kΩ Balco Pt 100 4.75-kΩ RTD network

TEMPERATURE COMPEN-SATION TYPE

Manual Auto 0 to 15% NaOH 0 to 20% NaCl 0 to 18% HCl 0 to 20% H2SO4 User-defined

T1 _______ KSTD/K1 _______T2 _______ KSTD/K2 _______T3 _______ KSTD/K3 _______T4 _______ KSTD/K4 _______T5 _______ KSTD/K5 _______T6 _______ KSTD/K6 _______

OUTPUT RANGE 4 mA _______ 20 mA _______

DAMPING VALUE _______sec

SAFE MODE LEVEL Fail Low Fail High

SPIKE MAGNITUDE _______%

DIAGNOSTICS Enabled Disabled

SECURITY Password _______ Configure Calibrate Output/Hold

Page 95: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

INTRODUCTIONWBPEEUI520003A1 PR Index - 1

PROCEDURE INDEX

INTRODUCTION

This index is provided as a quick reference for those with athorough knowledge of the Type TB82EC transmitter, relatedsensors, and this instruction. Procedures referenced in thisindex are part of an overall sequence. Going directly to a pro-cedure without consulting the sequence flowcharts presentedearlier in this instruction will not give an indication of whatcomes before and after in the sequence.

Title Procedure

Analyzer type selection PR19

Conductivity input PCB installation PR49

Conductivity input PCB removal PR47

Conductivity/concentration calibration PR8

Configuration lockout PR36

Configuration option selection PR23

Configuration reset PR39

Damping adjustment PR13

Decimal point location selection PR25

Diagnostics selection PR33

Edit calibration PR11

Electronic sensor test PR44

Engineering units selection PR24

Four-electrode conductivity sensor wiring PR6

Front bezel installation PR54

Front bezel removal PR48

Grounding PR7

Hinge mounting PR4

Hold output PR14

Keypad cleaning PR47

Linear curve fit programming PR27

Microprocessor PCB replacement PR51

Modifying configuration while viewing PR19

Output calibration PR10

Output range setting PR32

Output spike toggle PR17

Panel mounting PR2

Pipe mounting PR1

Power supply PCB installation PR53

Power supply PCB removal PR49

Page 96: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PROCEDURE INDEX

INTRODUCTIONPR Index- 2 WBPEEUI520003A1

Programming mode change PR20

Rear cover replacement PR55

Release hold output PR15

Rerange output PR16

Reset calibration PR12

Resetting all parameters PR41

Safe mode selection PR34

Security and password assignment PR37

Security reset PR40

Selecting view or modify configuration state PR18

Sensor cleaning PR46

Sensor group selection PR22

Sensor inspection PR43

Shell replacement PR56

Signal and power wiring PR5

Spike magnitude setting PR35

Temperature calibration PR8

Temperature compensation coefficient setting

PR30

Temperature compensation type selection PR29

Temperature sensor type selection PR28

Text string definition PR26

Transmitter soft boot PR42

Transmitter troubleshooting without sensor PR45

User-defined temperature coefficient programming

PR31

Utility mode advanced/basic programming PR38

Wall mounting PR3

Title Procedure

Page 97: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR1 - 1

PROCEDURE PR1 - PIPE MOUNTING

PURPOSE/SCOPE

This procedure describes how to mount the transmitter to apipe using pipe mounting kit 4TB9515-0124.

Parts

Tools • Crescent wrench.

PROCEDURE

The pipe mounting kit contains a pipe mounting bracket, aninstrument mounting bracket, and associated hardware. Thepipe mounting bracket accommodates pipe diameters as largeas two inches.

1. Select the location and orientation of the transmitter.

2. Refer to Figure PR1-1 and use four -inch by ¾-inchbolts, -inch flatwashers, -inch lockwashers, and -inchnuts to attach the instrument mounting bracket to the pipemounting bracket.

3. Tighten the hardware using the crescent wrench.

4. Use the two U-bolts, and four each of the -inch flatwashers, -inch lockwashers, and -inch nuts to attach thepipe mounting bracket to the pipe.

5. Tighten the hardware using the crescent wrench.

30 min.

Number Qty Description

4TB4704-0086 4 Bolt, -in. x ¾-in.

4TB4704-0096 2 U-bolt, -in.

4TB4704-0119 4 Bolt, -in. x -in.

4TB4710-0022 8 Lockwasher, -in.

4TB4710-0023 4 Lockwasher, -in.

4TB4710-0025 4 Flatwasher, -in.

4TB4710-0028 8 Flatwasher, -in.

4TB4711-0013 4 Nut, -in.

4TB4711-0020 4 Nut, -in.

4TB5008-0022 1 Bracket, pipe mounting

4TB5008-0071 1 Bracket, instrument mounting

Customer-supplied A/R Fitting, liquid tight

38 ⁄516⁄

38 ⁄ 58 ⁄38 ⁄516⁄516⁄38 ⁄

516⁄38 ⁄

38 ⁄38 ⁄ 38 ⁄ 38 ⁄

516⁄516⁄ 516⁄

Page 98: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PIPE MOUNTING

PROCEDUREPR1 - 2 WBPEEUI520003A1

6. Use the four -inch x -inch bolts, -inch flatwashers,

and -inch lockwashers to attach the transmitter to theinstrument mounting bracket.

NOTE: The transmitter has four sets of threaded mounting holes toaccommodate positioning the transmitter in any of four orientations.

7. Tighten the hardware using the crescent wrench.

Figure PR1-1. Pipe Mounting

T01089C

5/16-IN . U -B O LT(2 TYP )

3/8-IN . x 3 /4-IN . BO LT (4 TY P)

5/16-IN .FLATWASH ER (4 )

P IPE

3/8-IN .FLATWAS H ER (8)

5 /16-IN . N U T (4 )

5 /16-IN .LO C K WASH ER (4)

3 /8 -IN .LO C KWA SH ER (8)

L IQ U ID T IG H TFITTIN G (2 TYP)3/8-IN . x 5 /8-IN .

BO LT (4)

3 /8-IN . N U T (4)

IN STRU M E N TM O U N TIN GBR AC KET

P IPEM O U N TIN GBR AC KET

38 ⁄ 58 ⁄ 38 ⁄38 ⁄

Page 99: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR2 - 1

PROCEDURE PR2 - PANEL MOUNTING

PURPOSE/SCOPE

This procedure describes how to mount the transmitter into apanel using panel mounting kit 4TB9515-0123.

Parts

Tools • Tools for making panel cutout (dependent on installation).• Flat-blade screwdriver.• Crescent wrench.

PROCEDURE

The panel mounting kit contains four panel mounting bracketassemblies and a panel gasket. The transmitter enclosure con-forms to DIN sizing. Figure PR2-1 shows the required panelcutout, maximum panel thickness, and minimum panel depth.

1. Select the location and orientation of the transmitter.

2. Use suitable tools (dependent on installation) to make a135.4 (+1.3, -0.8) by 135.4 (+1.3, -0.8) mm (5.33 (+0.05, -0.03)by 5.33 (+0.05, -0.03)) in. cutout with diagonal corners asshown in Figure PR2-1.

3. Install the panel gasket onto the transmitter.

4. Install the transmitter into the panel cutout.

5. Use the crescent wrench and four -16 x ½-in. hex screwsand -in. lockwashers to attach the panel mounting bracketsto all four corners of the transmitter.

6. Use the flat-blade screwdriver to tighten the panel mount-ing screws on the panel mounting bracket until the transmit-ter seats tightly against the panel.

30 min.

Number Qty Description

4TB4704-0048 4 Screw, hex, -16 x ½-in.

4TB4704-0118 4 Screw, panel mounting

4TB4710-0022 4 Lockwasher, split, -in.

4TB4906-0019 1 Gasket, panel

4TB5205-0292 4 Bracket, panel mounting

38 ⁄

38 ⁄

38 ⁄38 ⁄

Page 100: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PANEL MOUNTING

PROCEDUREPR2 - 2 WBPEEUI520003A1

Figure PR2-1. Panel Mounting

PA N E LC U TO U T

R E A R C OV E RR EM OVE D

E X TE R N A LG RO U N D S C R EWS E E N OTE 2

PA N E L M O U N TIN GBR AC K E T (4)

3 /8-IN .LO C K W AS H E R (4)

3 /8-16 x 1 /2-IN .H EX S C R E W (4)

PAN EL M O U N TIN GS C R E W S (4)S EE N OTE 1

T01088D

R E A R V IE WPA N E L G A SK E T

PA N E LM IN U M U M PA N E L D E P TH = 144.8 m m (5 .70 in .)

11 .90.47

135.4 +1.3–0.8

5 .33 +0.05–0 .03

135.4 +1.3–0 .8

5 .33 +0.05–0 .03

11.90.47

9.50.38

M A X IM U M PA N E LTH IC K N E S S

N OTE S :1. T IG H TE N S C R E W S O N M O U N TIN G BR AC K E T A S S EM B LY TO D R AW IN STRU M EN T U P AG AIN S T WA LL A N D G A S K E T TO E N S U R E WATE R T IG H T S EA L.

2 . R EM OV E EX TE R N A L G RO U N D S C R E W A N D H A R DWA R E B E FO R E IN S TA LLIN G PA N E L M O U N T B R AC K E T AS S E M B LY.

3. IN S E RT TR A N S M ITTER IN TO PA N E L C U TO U T A N D IN S TA LL PA N E L M O U N TIN G B R AC K E TS.

D IM E N S IO N S

M ILLIM E TER SIN C H ES

S E E N OTE 3

Page 101: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR3 - 1

PROCEDURE PR3 - WALL MOUNTING

PURPOSE/SCOPE

This procedure describes how to mount the transmitter to awall or other suitable surface using wall mounting kit4TB9515-0123.

Parts

Tools • Tools for mounting instrument mounting bracket to wall(dependent on installation).

• Crescent wrench.

PROCEDURE

The wall mounting kit contains an instrument mountingbracket and associated hardware. Wall mounting accommo-dates installations where the transmitter can be positioned fora clear line of sight and free access to the rear terminations.These types of installations include supporting beams, flangebrackets, and wall ends.

1. Select the location and orientation of the transmitter.

2. Refer to Figure PR3-1 and attach the instrument mountingbracket to the selected location using the appropriate type offastener based on the material of the wall.

3. Use four -inch x -inch bolts, -inch flat washers, and-inch lockwashers to attach the transmitter to the instru-

ment mounting bracket.

NOTE: The transmitter has four sets of threaded mounting holes toaccommodate positioning the transmitter in any of four orientations.

4. Tighten the hardware using the crescent wrench.

30 min.

Number Qty Description

4TB4704-0119 4 Bolt, -in. x -in.

4TB4710-0022 4 Lockwasher, -in.

4TB4710-0028 4 Flatwasher, -in.

4TB5008-0071 1 Bracket, instrument mounting

Customer-supplied A/R Fitting, liquid tight

Customer-supplied A/R Fasteners for wall

38 ⁄ 58 ⁄38 ⁄38 ⁄

38 ⁄ 58 ⁄ 38 ⁄38 ⁄

Page 102: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WALL MOUNTING

PROCEDUREPR3 - 2 WBPEEUI520003A1

Figure PR3-1. Wall Mounting

LIQ U ID T IG H TF ITT IN G (2 TY P)

W ALL

IN S TRU M E N TM O U N TIN GBR AC KE T

3/8-IN .FLATW A SH ER (4)

3 /8 -IN . x 5 /8-IN .B O LT (4 )

3 /8 -IN . LO C KW AS H ER (4 )

FA S TE N ER S FO R W A LL(C U S TO M E R SU PP LIE D )

T01092C

Page 103: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR4 - 1

PROCEDURE PR4 - HINGE MOUNTING

PURPOSE/SCOPE

This procedure describes how to mount the transmitter to awall or other suitable surface using hinge mounting kit4TB9515-0125.

Parts

Tools • Tools for mounting L-bracket to mounting surface (depen-dent on installation).

• Crescent wrench.

PROCEDURE

The hinge mounting kit contains an L bracket, an instrumentmounting bracket, a stainless steel hinge, and associatedhardware. The hinge mounting kit allows free access to therear of the transmitter.

1. Select the location and orientation of the transmitter.

2. Refer to Figure PR4-1 and attach the L-bracket to theselected location using the appropriate type of fastener basedon the material of the mounting surface.

3. Use four of the -inch x ¾-inch bolts, -inch flat wash-ers, -inch lockwashers, and -inch nuts to attach thehinge to the L-bracket.

4. Tighten the hardware using the crescent wrench.

5. Use four -inch x ¾-inch bolts, -inch flat washers,-inch lockwashers, and -inch nuts to attach the instru-

ment mounting bracket to the hinge.

30 min.

Number Qty Description

4TB4704-0048 4 Bolt, -in. x -in.

4TB4704-0086 8 Bolt, -in. x ¾-in.

4TB4710-0022 12 Lockwasher, -in.

4TB4710-0028 12 Flatwasher, -in.

4TB4711-0020 8 Nut, -in.

4TB5008-0071 1 Bracket, instrument mounting

4TB5008-0073 1 Bracket, L

4TB5010-0005 1 Hinge, stainless steel

Customer-supplied A/R Fitting, liquid tight

Customer-supplied A/R Fasteners for mounting surface

38 ⁄ 58 ⁄38 ⁄

38 ⁄38 ⁄

38 ⁄

38 ⁄ 38 ⁄38 ⁄ 38 ⁄

38 ⁄ 38 ⁄38 ⁄ 38 ⁄

Page 104: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

HINGE MOUNTING

PROCEDUREPR4 - 2 WBPEEUI520003A1

6. Tighten the hardware using the crescent wrench.

7. Use four -inch x -inch bolts, -inch flat washers, and-inch lockwashers to attach the transmitter to the instru-

ment mounting bracket.

NOTE: The transmitter has four sets of threaded mounting holes toaccommodate positioning the transmitter in any of four orientations.

8. Tighten the hardware using the crescent wrench.

Figure PR4-1. Hinge Mounting

T 01090C

3/8-IN . x 3 /4-IN .B O LT (8 )

3 /8-IN . x 5 /8 -IN .B O LT (4 ) 3 /8-IN .

FLATWA SH ER (4 )

3 /8-IN .LO C KW AS H ER (4)

3 /8-IN .N U T (8) 3 /8-IN .FLATWA SH E R (8)

3 /8-IN .LO C KW A SH E R (8)

IN S TRU M E N TM O U N TIN GBR AC K ET

LBR AC K ET

H IN G E

TO P V IE W FRO N T V IE W

FA S TEN E R SFO R W ALL

(C U STO M E R -S U P PLIE D )

M O U N TIN GS U R FAC E

LIQ U ID T IG H TFITT IN G(2 TY P)

38 ⁄ 58 ⁄ 38 ⁄38 ⁄

Page 105: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR5 - 1

PROCEDURE PR5 - SIGNAL/POWER WIRING

PURPOSE/SCOPE

This procedure describes how to connect the signal/power wir-ing to the transmitter.

Parts None.

Tools • Flat-blade screwdriver.• Small flat-blade screwdriver.

PROCEDURE

Use shielded wire and separate conduit for the signal/powerwiring, and the sensor wiring. Under ideal conditions, this maynot be required; however, it minimizes the chance of problemsfrom noise and signal degradation.

Just prior to entering the housing, terminate rigid conduit andinstall a short length of flexible conduit. This reduces stress tothe housing.

The power and output signals share the same pair of wires.This wiring must bear a suitable voltage rating and be rated toat least 75-degrees Celsius (167-degrees Fahrenheit). All wir-ing and wiring practices must be in accordance with theNational Electric Code (NEC), Canadian Electrical Code (CEC),or other applicable local or international codes for the countrywhere the transmitter will be installed.

The signal terminals, located at the rear of the transmitter,accept wire sizes from 12 to 24 AWG. ABB recommendspin-style terminals for all connections.

A terminal block (TB1) label is marked POWER for the signalconnections and shows the polarity. Wiring should not be runin conduit or open trays where power wiring or heavy electricalequipment could contact or physically and electrically interferewith the signal wiring. Twisted, shielded pairs should be usedfor cabling to insure the best performance. Reverse polarityprotection, built into the transmitter, protects it against acci-dental reversal of the field wiring connections.

All power passes over the signal leads. The maximum supplyvoltage is 53 VDC (42 VDC for certified applications). Minimumsupply voltage is determined by the loop resistance(Fig. PR5-1) as follows:

10 min.

min supply voltage (VDC) 13 VDC 0.020 A total R in ohms×( )+=

Page 106: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SIGNAL/POWER WIRING

PROCEDUREPR5 - 2 WBPEEUI520003A1

The load resistance must include any meters external to thetransmitter, the wiring, and the system input.

1. Use the flat-blade screwdriver to loosen the four captivescrews that secure the rear cover to the transmitter.

2. Use the small flat-blade screwdriver to connect the signaland power wiring to TB1-1 (+) and TB1-2 (-) as shown inFigure PR5-2.

3. Install the rear cover and tighten the captive screws withthe flat-blade screwdriver.

Figure PR5-1. Load Limits

13 18 23 28 33

S U P P LY VO LTAG E (V D C )

LO A DR E S IS TA N C E

(O H M S )

38 43 48 53

T01964B

2000

1800

1600

1400

1200

1000

800

600

400

200

0

Figure PR5-2. Signal and Power Wiring

T02468C

S EN S O R C O N N EC TIO N S POW E R /O U TPU TC O N N E C TIO N S

TB 2

4-20m AS IG N A L

+1

–2

+3

–4

G R E EN

R E D

W H ITE

B LAC K

BLU E

YE LLOW

H V Y G RN

1

2

3

4

5

6

7

8

D R IV E

D R IV E

SE N SE

S H IE LD

RTD

RTD

SE N SE

AM M E TE RTER M IN A LS

TB 1

POW E R

TE ST

POW ER S U P P LY

+–

C O N TRO L H O U SER EA R V IE W

TH E M IN IM U M POW ER SU P P LY VO LTAG E IS13V D C + (0 .02 A M P S x TOTA L LO O P

R ES ISTAN C E IN O H M S ). T H E M AX IM U MPOW ER S U P P LY VO LTAG E IS 53V D C (42 V D C

F O R C ERTIF IE D AP P LIC ATIO N S).

C O N TRO LLER4-20m A

+–

R E C O R D E R4-20m A

AD D IT IO N A LELE M E N TS

4-20m A+ +– –

N OTE S:

1 . VO LTAG E AT P OW E R T ER M IN ALS T B1-1 A N D TB 1-2 M U ST B E B E TW E E N 13 A N D 53 V D C (13 A N D 42 V D C FO R C ERTIF IE D A PP LIC ATIO N S ).

2 . RU N 22 AW G (O R H E AVIE R ) S H IELD E D C A B LE IN C O N D U IT.

S E EN OTE 1

S E EN OTE 2

SH IELD

Page 107: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR6 - 1

PROCEDURE PR6 - SENSOR WIRING

PURPOSE/SCOPE

This procedure describes how to connect ABB four-electrodeconductivity sensors to the transmitter.

Parts None.

Tools • Flat-blade screwdriver.• Small flat-blade screwdriver.

PROCEDURE

The sensor wiring connects to the rear of the transmitter. Theterminals accept wire sizes from 12 to 24 AWG. ABB recom-mends pin-style terminals for all connections. Run sensor wir-ing in shielded conduit, or similar, to protect it fromenvironmental influences. Do not allow the wires to becomewet or lie on the ground or over any other equipment. Insurethere is no abrading, pinching, or bending of the cables atinstallation.

The sensor leads are color coded and have the functions listedin Table PR6-1.

1. Use the flat-blade screwdriver to loosen the four captivescrews that secure the rear cover to the transmitter.

2. Use the small flat-blade screwdriver to connect the sensorwiring as shown in Figure PR6-1.

3. Install the rear cover and tighten the captive screws withthe flat-blade screwdriver.

10 min.

Table PR6-1. Pin Terminal Sensor Connections

Terminal Color Code Function

TB2-1 Green Drive

TB2-2 Red Sense

TB2-3 White Sense

TB2-4 Black Drive

TB2-5 Blue1 RTD

TB2-6 Yellow1 RTD

TB2-7 Heavy Green Shield

TB2-8 No connection No connectionNOTE:1. Blue and yellow conductors are only present when using a temperaturecompensator.

Page 108: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SENSOR WIRING

PROCEDUREPR6 - 2 WBPEEUI520003A1

Figure PR6-1. Four-Electrode Sensor Connections

T02469B

S EN S O R C O N N E C TIO N S P OW ER /O U TPU TC O N N EC TIO N S

TB 2

4-20m ASIG N AL

+1

–2

+3

–4

G R E EN

R E D

W H ITE

B LAC K

B LU E

YE LLOW

H V Y G R N

1

2

3

4

5

6

7

8

D R IV E

D R IV E

SE N S E

SH IELD

RTD

RTD

SE N S E

AM M ETERTER M IN A LS

TB1

P OW ER

TES T

R E A R V IE W

TY P IC A L

Page 109: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR7 - 1

PROCEDURE PR7 - GROUNDING

PURPOSE/SCOPE

This procedure describes how to properly ground thetransmitter.

Parts None.

Tools • Flat-blade screwdriver.

PROCEDURE

Signal wiring should be grounded at any one point in the sig-nal loop, preferably before signal processing occurs. It may beleft ungrounded (floating) if electrical noise is minimal. Groundthe transmitter enclosure to an earth ground having less than0.2 ohms of resistance. Internal and external earth groundterminals are provided and are shown in Figure PR7-1.

10 min.

Figure PR7-1. Earth Grounds

T02470B

S EN SO R C O N N E C T IO N S POW E R /O U T P U TC O N N E C TIO N S

TB2

4-20m AS IG N AL

+1

–2

+3

–4

G R EE N

R ED

W H IT E

B LAC K

B LU E

YE LLOW

H VY G R N

1

2

3

4

5

6

7

8

D R IV E

D R IV E

S EN S E

S H IE LD

RTD

RTD

S EN S E

A M M ET ERTE R M IN ALS

TB 1

PO W E R

T ES T

R EA R V IEW

EA RTH G RO U N D S

Page 110: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR8 - 1

PROCEDURE PR8 - TEMPERATURE CALIBRATION

PURPOSE/SCOPE

This procedure describes how to perform a temperaturecalibration.

Parts None.

Tools • Temperature measuring device.

PROCEDURE

The temperature calibration state is a smart calibration rou-tine that allows for single-point and dual-point calibrations.Calibrating the temperature at two points that are at least20-degrees Celsius apart causes the transmitter to automati-cally adjust the temperature sensor offset, slope, or both.Since this routine only uses the most recent calibration data,calibration can be conducted throughout the life of the sensor.This ensures consistent performance. The reset calibrationstate restores the calibration to factory settings in the event ofbad calibration data. The reset calibration state is discussed inPR12.

NOTE: The reset calibration state resets all calibration valuesincluding the process sensor; therefore, the process sensor requirescalibration after performing the reset calibration procedure.

1. If this is a new installation, allow the sensor to reach ambi-ent temperature. If this is an existing installation, wait untilthe process temperature stabilizes.

2. Press the MENU smart key until CALIBRATE is highlightedon the display.

3. Press the SELECT smart key to enter the calibrate state.

4. Press the NEXT smart key until TMP.CAL appears on thedisplay.

5. Press the SELECT smart key to start the temperature cali-bration procedure.

6. The transmitter displays STABL?. Observe the displayedreading. If it is stable, press the YES smart key. If it is not sta-ble, press the NO smart key. If NO is selected, the transmitterreturns to the temperature calibration state. Perform Steps 7through 9. If YES is selected, go to Step 10.

20 min.

Page 111: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE CALIBRATION

PROCEDUREPR8 - 2 WBPEEUI520003A1

7. If the ambient or process liquid temperature has becomeunstable, wait until the process temperature stabilizes.

8. Check to see if the transmitter has detected a fault condi-tion by looking for the FAULT icon on the display. Interrogatethe fault by escaping to the measure mode through the exit toMEASURE smart key and then pressing the FAULT info smartkey in that order.

9. Refer to Section 12 for information on problem and errorcodes.

10. The transmitter asks for the NEW VAL. Use the tempera-ture measuring device to measure the ambient temperature fora new installation or the process liquid for an existing installa-tion and enter that value into the NEW VAL. screen.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 10a and 10b for each digit.

d. Press the ENTER smart key to accept the new value.

11. For an existing installation where a two-point calibration isdesired, wait for the process liquid temperature to change to20°C. When the process liquid temperature has stabilized, goto Step 6. For a new installation, mount the sensor in its finalinstalled location and allow it to reach the process liquid tem-perature. When the sensor and process liquid have stabilized,go to Step 5.

Page 112: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR9 - 1

PROCEDURE PR9 - CONDUCTIVITY/CONCENTRATIONCALIBRATION

PURPOSE/SCOPE

This procedure describes how to perform a conductivity/con-centration calibration.

Parts None.

Tools • External instrument having the same type of temperaturecompensation as the transmitter.

• Grab sample of process liquid.

PROCEDURE

The conductivity/concentration calibration state is a smartone-point calibration routine. It allows for single-point anddual-point calibrations. Initiating calibrations at two differentconductivity values having ample separation allows the trans-mitter to automatically adjust the offset, slope, or both. Thisinsures the best sensor performance. Since this routine usesonly the latest calibration data, calibration can be conductedthroughout the life of the sensor. This insures consistent sen-sor performance.

If an incorrect calibration has been entered, the reset calibra-tion state provides the ability to return transmitter calibrationto those set at the factory.

NOTE: The reset calibration state resets all calibration values,including the temperature calibration values. A temperature calibra-tion must be performed after a calibration reset.

The transmitter is configurable as a conductivity or concentra-tion transmitter. The smart one-point calibration routinesautomatically set the units of calibration to be the same asthose for the measured process variable.

1. Make sure the sensor is in its final installed location andorientation.

2. Measure the process variable value, using an externalinstrument having the same type of temperature compensa-tion as the transmitter, and a grab sample.

3. Record the value displayed on the transmitter at the timethe grab sample was taken and the value displayed on theexternal instrument.

4. Press the MENU smart key until CALIBRATE is highlightedon the display.

10 min.

Page 113: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONDUCTIVITY/CONCENTRATION CALIBRATION

PROCEDUREPR9 - 2 WBPEEUI520003A1

5. Press the SELECT smart key to enter the calibrate state.

6. Press the NEXT smart key until CON.CAL appears on thedisplay.

7. Press the SELECT smart key to start the one-point calibra-tion procedure.

8. The transmitter displays STABL?. Observe the displayedreading. If it is stable, press the YES smart key. If it is not sta-ble, press the NO smart key. If NO is selected, the transmitterreturns to the process sensor calibration state. PerformSteps 9 through 11. If YES is selected, go to Step 12.

9. Wait until the process liquid composition stabilizes.

10. Check to see if the transmitter has detected a fault condi-tion by looking for the FAULT icon on the display. Interrogatethe fault by escaping to the measure mode through the exit toMEASURE smart key and then pressing the FAULT info smartkey in that order.

11. Refer to Section 12 for information on problem and errorcodes.

12. The transmitter displays NEW VAL.

a. Calculate the new process variable value by subtract-ing the transmitter value recorded in Step 3 from the exter-nal instrument value recorded in Step 3 and then addingthat result to the current value displayed on thetransmitter.

b. Use the smart key to increment the value of theblinking digit.

c. When the first digit value is correct, use the smartkey to move to the next digit.

d. Repeat Steps 12b and 12c for each digit.

e. Press the ENTER smart key to accept the new value.

13. If the entered calibration value is not valid, the transmitterdisplays BAD.CAL, and the calibration value is rejected. If theentered calibration value is valid, the slope (sensor efficiency)appears on the display. Press the NEXT smart key to displaythe offset.

Page 114: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONDUCTIVITY/CONCENTRATION CALIBRATION

PROCEDUREWBPEEUI520003A1 PR9 - 3

14. Press the NEXT smart key to return to the conductivity/concentration calibration state or press the exit to MEASUREsmart key to go to the measure mode.

NOTE: If an output hold condition is present, the display inquires ifthis condition should be released.

Page 115: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR10 - 1

PROCEDURE PR10 - OUTPUT CALIBRATION

PURPOSE/SCOPE

This procedure describes how to calibrate the transmitter out-put values using an external validation device.

Parts None.

Tools • Digital multimeter (DMM).• Flat-blade screwdriver.• Small flat-blade screwdriver.

PROCEDURE

The output calibrate state trims the output signal to maintainprecise transmission of the process variable to the final moni-toring system. The transmitter output current is factory cali-brated; however, the output can be trimmed to compensate forother input and output devices.

1. Use the flat-blade screwdriver to loosen the four captivescrews that secure the rear cover to the transmitter.

2. Use the small flat-blade screwdriver to remove the shortingjumper from the test terminals, TB1-3 (+) and TB1-4 (-), asshown in Figure PR10-1.

10 min.

Figure PR10-1. Output Calibration Setup

S EN S O R C O N N E C TIO N S P OW E R /O U TPU TC O N N EC TIO N STB2

4 -20 m AS IG N A L

+1

-2

+3

-4

G R E E N

R E D

W HITE

B LAC K

B LU E

Y E LLO W

H V Y G R N

12345678

D R IV E

D R IV E

S E N S E

SH IE LD

RTD

RTD

S E N S E

A M M E TE RTE R M IN A LS

TB 1

P OW E R

TE S T

P OW ER SU PPLY

+–

C O N TRO L H O U S E

1.999

R E AR V IE W

T02471C

TY P IC A L

D M M

THE M IN IM U M POW E R SU PP LY VO LTAG E IS13VD C + (0 .02 A M PS x TOTAL LO O P

R ES ISTA N C E IN O H M S). TH E M AX IM U MPOW E R S U PPLY VO LTAG E IS 53V D C (42 VD C

FO R C E RTIF IE D AP PLIC AT IO N S ).

C O N TRO LLE R4-20m A

+–

R EC O R D ER4-20m A

AD D IT IO N A LELE M EN TS

4-20m A+ +– –

N OT ES :

1. S H O RTIN G JU M P E R M U S T B E R E M OV E D W H E N M E A S U R IN G LO O P C U R R E N T A N D R E TU R N E D A FTE R D M M H A S B E E N R E M OV E D.

2. VO LTAG E AT P OW E R TE R M IN ALS TB 1-1 A N D TB 1-2 M U S T B E BE TW E E N 13 A N D 53 V D C (13 A N D 42 V D C F O R C ERTIF IE D A P P LIC ATIO N S ).

3 . RU N 22 AW G (O R H E AV IE R ) S H IE LD E D C A BLE IN C O N D U IT.

S E EN OTE 2

S EEN OTE 3

S H IE LD

Page 116: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

OUTPUT CALIBRATION

PROCEDUREPR10 - 2 WBPEEUI520003A1

3. Connect the DMM, set to measure mA, to the TEST termi-nals, TB1-3 (+) and TB1-4 (-).

4. Press the MENU smart key until CALIBRATE is highlightedon the display.

5. Press the SELECT smart key to enter the calibrate state.

6. Press the NEXT smart key until OUT.CAL appears on thedisplay.

7. Press the SELECT smart key to start the output calibrationprocedure.

8. Use the and smart keys to adjust the output so thatthe DMM reads 4.0 mA.

9. Press the ENTER smart key to enter the new value and pro-ceed to the 20-mA output.

10. Use the and smart keys to adjust the output so thatthe DMM reads 20.0 mA.

11. Press the ENTER smart key to enter the new value. Thetransmitter returns to the output calibration state.

NOTE: Once the output level has been permanently stored usingthe ENTER smart key, the output calibration procedure must berepeated to rectify a bad calibration.

12. Press the exit to MEASURE smart key to return to the mea-sure mode.

Page 117: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR11 - 1

PROCEDURE PR11 - EDIT CALIBRATION

PURPOSE/SCOPE

This procedure describes how to edit the process sensor andtemperature sensor offset and slope values.

Parts None.

Tools None.

PROCEDURE

The edit calibration state allows manual adjustment of theprocess sensor and temperature sensor slope and offset val-ues. This function may not be suitable for many applications,but it facilitates quick and easy access to these calibration val-ues for troubleshooting purposes.

1. Press the MENU smart key until CALIBRATE is highlightedon the display.

2. Press the SELECT smart key to enter the calibrate state.

3. Press the NEXT smart key until EDT.CAL appears on thedisplay.

4. Press the SELECT smart key to start the edit calibrationprocedure.

5. The transmitter displays the process sensor slope. Validslope values range from 0.20 to 5.00.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 5a and 5b for each digit.

d. Press the ENTER smart key to accept the new valueand edit the sensor offset. To continue, go to Step 6. Toend the procedure, press the exit to MEASURE smart key.

6. The transmitter displays the process sensor offset. Thevalid offsets range from -20 to +20 µS/cm for Group A sensors,

5 min.

Page 118: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

EDIT CALIBRATION

PROCEDUREPR11 - 2 WBPEEUI520003A1

-4 to +4 µS/cm for Group B sensors, and -0.8 to +0.8 µS/cmfor Group C sensors.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 6a and 6b for each digit.

d. Press the ENTER smart key to accept the new valueand edit the temperature slope. To continue, go to Step 7.To end the procedure, press the exit to MEASURE smartkey.

7. The transmitter displays the temperature sensor slope.Valid slope values range from 0.2 to 1.5.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 7a and 7b for each digit.

d. Press the ENTER smart key to accept the new valueand edit the temperature offset. To continue, go to Step 8.To end the procedure, press the exit to MEASURE smartkey.

8. The transmitter displays the temperature sensor offset.Valid offset values range from -40°C to +40°C.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 8a and 8b for each digit.

d. Press the ENTER smart key to accept the new valueand go back to the edit calibration state. To end the proce-dure, press the exit to MEASURE smart key.

Page 119: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR12 - 1

PROCEDURE PR12 - RESET CALIBRATION

PURPOSE/SCOPE

This procedure describes how to restore process sensor andtemperature sensor calibration values to the factory values.

Parts None.

Tools None.

PROCEDURE

The reset calibration state sets all process sensor and temper-ature sensor calibration data to the values set at the factory.This state allows the purging of all calibration history and thestart of a new history. The reset sets the process sensor andtemperature sensor slope values to 1.000, and the processsensor and temperature sensor offset values to 0.000.

1. Press the MENU smart key until CALIBRATE is highlightedon the display.

2. Press the SELECT smart key to enter the calibrate state.

3. Press the NEXT smart key until RST.CAL appears on thedisplay.

4. Press the SELECT smart key to start the reset calibrationprocedure.

5. The display reads RESET?. Press the YES smart key to con-firm the reset or the NO smart key to refuse the reset.

NOTE: The reset calibration state resets all sensor and temperaturecalibration values; therefore, the process sensor and temperaturesensor require calibration after performing the reset calibrationprocedure.

1 min.

Page 120: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR13 - 1

PROCEDURE PR13 - DAMPING ADJUSTMENT

PURPOSE/SCOPE

This procedure describes how to change the damping on thetransmitter output from the output/hold mode and the config-ure mode.

Parts None.

Tools None.

PROCEDURE

The damping state applies a lag function on the output signaland reduces fluctuations caused by erratic process conditions.The damping value is adjustable from 0.0 to 99.9 seconds.This value represents the time required for the output to reachapproximately 63.2 percent of its final value in response to astep change in the process variable input.

Output/Hold Mode

1. Press the MENU smart key until OUT/HOLD is highlightedon the display.

2. Press the SELECT smart key to enter the output/holdmode.

3. Press the NEXT smart key until DAMPNG appears on thedisplay.

4. Press the SELECT smart key to start the dampingprocedure.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 4a and 4b for each digit.

d. Press the ENTER smart key to accept the new value.

5. When the process is complete, the transmitter automati-cally goes to the measure mode.

1 min.

Page 121: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

DAMPING ADJUSTMENT

PROCEDUREPR13 - 2 WBPEEUI520003A1

Configure mode

1. Press the SELECT smart key to modify the DAMPNG state.

2. The transmitter displays the current damping value.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 2a and 2b for each digit.

d. Press the ENTER smart key to accept the new value.

Page 122: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR14 - 1

PROCEDURE PR14 - HOLD OUTPUT

PURPOSE/SCOPE

This procedure describes how to hold the transmitter outputat a captured or manually adjusted value.

Parts None.

Tools None.

PROCEDURE

The hold state allows the transmitter output to be fixed at alevel captured upon initiation of the hold or to be manuallyadjusted to any value between zero and 100 percent (four and20 milliamps).

1. Press the MENU smart key until OUT/HOLD is highlightedon the display.

2. Press the SELECT smart key to enter the output/holdmode.

3. Press the NEXT smart key until HOLD appears on thedisplay.

4. Press the SELECT smart key to start the hold outputprocedure.

5. To accept the current hold value, press the ENTER smartkey and the transmitter automatically goes to the measuremode. To adjust the current hold value, do not press theENTER smart key and continue with Step 6.

6. Press the smart key to increment the blinking digit tothe desired value.

7. Press the smart key to move to the next digit.

8. Repeat Steps 6 and 7 for each digit.

9. Press the ENTER soft key to accept the new value.

10. When the process is complete, the transmitter automati-cally goes to the measure mode.

1 min.

Page 123: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR15 - 1

PROCEDURE PR15 - RELEASE HOLD OUTPUT

PURPOSE/SCOPE

This procedure describes how to release the transmitter out-put from a hold condition.

Parts None.

Tools None.

PROCEDURE

The hold state is used to release a hold condition that alreadyexists.

1. Verify that OUTPUT HELD appears in the upper left cornerof the display.

2. Press the MENU smart key until OUT/HOLD is highlightedon the display.

3. Press the SELECT smart key to enter the output/holdmode.

4. Press the NEXT smart key until HOLD appears on thedisplay.

5. Press the SELECT smart key to start the release hold out-put procedure.

6. The transmitter display reads REL.HLD. Press the YESsmart key to release the hold output condition or the NO smartkey to continue to hold the output.

7. When the process is complete, the transmitter automati-cally goes to the measure mode.

1 min.

Page 124: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR16 - 1

PROCEDURE PR16 - RERANGE OUTPUT

PURPOSE/SCOPE

This procedure describes how to change the output range.

Parts None.

Tools None.

PROCEDURE

The rerange state provides the ability to change the outputrange. Change one or both endpoint values to any value orrange of values that are within those listed in Table PR16-1.

1. Press the MENU smart key until OUT/HOLD is highlightedon the display.

2. Press the SELECT smart key to enter the output/holdmode.

3. Press the NEXT smart key until RERNGE appears on thedisplay.

4. Press the SELECT smart key to start the rerangeprocedure.

5. Edit the process variable value for the four-mA point.

a. Press the smart key to increment the blinking digitto the desired value.

b. Press the smart key to move to the next digit or unitof conductivity.

c. Repeat Steps 5a and 5b for each digit or unit ofconductivity.

2 min.

Table PR16-1. Output Ranges

Input Type Output Range

Concentration 0.000 to 1,999 digits (engineering unit configurable)

Conductivity

Group A 0.000 to 1,999 mS/cm

Group B 0.000 to 1,999 µS/cm

Group C 0.000 to 199.9 µs/cm

Page 125: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

RERANGE OUTPUT

PROCEDUREPR16 - 2 WBPEEUI520003A1

6. Press the ENTER smart key to accept the new value andcontinue to the process variable value for the 20-mA point.

7. Press the exit to MEASURE smart key to escape to the mea-sure mode or continue with the procedure to adjust the pro-cess variable value for the 20-mA point.

NOTE: If the four-mA value is changed and accepted using theENTER smart key, the value is valid per those shown inTable PR16-1, and the transmitter is returned to the measure modeby pressing the exit to MEASURE smart key without adjusting the20-mA value, the output range will reflect the new four-mA point.

8. Edit the process variable value for the 20-mA point.

a. Press the smart key to increment the blinking digitto the desired value.

b. Press the smart key to move to the next digit or unitof conductivity.

c. Repeat Steps 8a and 8b for each digit or unit ofconductivity.

9. Press the ENTER smart key to accept the new value.

10. When the process is complete, the transmitter automati-cally goes to the measure mode.

Page 126: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR17 - 1

PROCEDURE PR17 - OUTPUT SPIKE TOGGLE

PURPOSE/SCOPE

This procedure describes how to enable or disable the diagnos-tic spike. This setting is available in the Advanced program-ming mode.

Parts None.

Tools None.

PROCEDURE

The spike state toggles the operational state of the spike out-put function. The spike function modulates the current outputby the amount set in the transmitter configuration.

1. Press the MENU smart key until OUT/HOLD is highlightedon the display.

2. Press the SELECT smart key to enter the output/holdmode.

3. Press the NEXT smart key until SPIKE appears on thedisplay.

4. Press the SELECT smart key to start the spike outputprocedure.

5. Toggle the spike output function to the desired state (ON orOFF) by using the smart key to toggle between ON and OFF.

6. Press the ENTER smart key to select the desired state.

7. When the process is complete, the transmitter automati-cally goes to the measure mode.

NOTE: Once the spike state is OFF, changing the configured spikelevel in the configure mode will not enable the spike state. The spikestate can only be turned ON or OFF in the output/hold mode.

1 min.

Page 127: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR18 - 1

PROCEDURE PR18 - SELECTING VIEW OR MODIFYCONFIGURATION STATE

PURPOSE/SCOPE

This procedure describes how to select whether to view ormodify the configuration.

Parts None.

Tools None.

PROCEDURE

When the configure mode is selected, a decision point isreached to determine whether to view or modify theconfiguration.

1. Press the MENU smart key until CONFIGURE is highlightedon the display.

2. Press the SELECT smart key to enter the configure mode.

3. The MODIFY screen appears. Press the YES smart key tomodify the configuration and go to the next procedure in theflow. Press the NO smart key to view the configuration and goon to Step 4.

4. The VIEW screen appears. Press the YES smart key to viewthe configuration.

5. To view the configuration only, perform Steps 6 and 7. Tomodify the configuration, go to the next procedure in the flow.

6. Press the NEXT smart key to scroll through theconfiguration.

7. At any time during the viewing of the configuration, pressthe exit to MEASURE smart key to go back to the measuremode.

1 min.

Page 128: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR19 - 1

PROCEDURE PR19 - MODIFYING CONFIGURATION WHILEVIEWING

PURPOSE/SCOPE

This procedure describes how to modify the configurationwhile in the view configure state.

Parts None.

Tools None.

PROCEDURE

1. Press the NEXT smart key to scroll through the configura-tion until the parameter that requires changing appears.

2. Press the ENTER smart key to modify that parameter.

3. A MODIFY? screen appears. Press the YES smart key tomodify the parameter.

4. If the configure mode is password protected, perform thisstep and then go on to Step 5. If it is not password protected,go on to Step 5. When the password inquiry screen (PASSWD)appears:

a. The display reads _ _ _. Use the smart key to incre-ment the value of the blinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 4a and 4b for each digit.

d. Press the ENTER smart key to accept the password.

5. Modify the desired parameter using the proper procedure.

6. Press the exit to MEASURE smart key.

7. When the SAVE? screen comes up, press the YES smartkey to accept the change. Press the NO smart key to abort thechange. In either case, the transmitter goes to the measuremode.

1 min.

Page 129: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR20 - 1

PROCEDURE PR20 - PROGRAMMING MODE CHANGE

PURPOSE/SCOPE

This procedure describes how to change the programmingmode from advanced to basic or from basic to advanced fortransmitters with the advanced programming option.

Parts None.

Tools None.

PROCEDURE

The configure mode is split into two groups of programming:basic and advanced. These two options are specified bynomenclature and control the number of configuration optionsavailable in the modify configure mode.

The basic programming mode contains a subset of configura-tion options found in the advanced programming mode. Sepa-ration into two programming groups is advantageous whenlimited functionality is desired. Fewer options reduces confu-sion and the possibility of configuration errors.

Transmitters ordered with the advanced programming optioncan be changed between basic and advanced programming.Transmitters ordered with the basic programming optionrequire an update password to change to advanced program-ming. Contact ABB to obtain the password.

The programming toggle (BASIC or ADVNCD) for transmitterswith the advanced programming option must be set in twolocations: user state in the utility mode and modify configurestate in the configure mode. In order to select either the basicor advanced programming mode in the modify configure state,the programming mode must be set to advanced in the userstate.

NOTE: ADVNCD is the factory default setting in both modify config-ure state in the configure mode and user state in the utility mode fortransmitters ordered with the advanced programming option.

1. The BASIC screen appears. Press the ENTER smart key toset the programming to basic and advance to the modify con-figure states or press the NEXT smart key to advance to thenext screen and go on to Step 2.

2. The ADVNCD screen appears. Press the ENTER smart keyto set the programming to advanced and advance to the modifyconfigure states.

1 min.

Page 130: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR21 - 1

PROCEDURE PR21 - ANALYZER TYPE SELECTION

PURPOSE/SCOPE

This procedure describes how to configure the analyzer state.This setting is available in both the basic and advanced pro-gramming modes.

Parts None.

Tools None.

PROCEDURE

The analyzer state determines the transmitter and sensor type.Table PR21-1 describes the function and programming modeof each state.

1. Press the SELECT smart key to modify the analyzer state.

2. If this is a basic configuration, the transmitter goes directlyto the sensor group options. Go to the next procedure in theflow. If this is an advanced configuration, continue withStep 3.

3. The currently configured analyzer state appears. Press theNEXT smart key until the desired analyzer state appears.

4. Press the ENTER smart key to accept the new value.

1 min.

Table PR21-1. Analyzer States

State Display Programming Mode Function

Concentration CONCEN Basic and Advanced Used to measure conductivity of a solution and con-vert nonspecific measurement to specific solute concentration. Process variable engineering units are %, ppm, ppb, and user-defined.

Conductivity COND Basic Used to measure conductivity of a solution. Process variable engineering units are µS/cm and mS/cm.

Page 131: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR22 - 1

PROCEDURE PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPE

This procedure describes how to select the sensor group.

Parts None.

Tools None.

PROCEDURE

There are three sensor groups: A, B, and C. The sensor groupmust coincide with the conductivity ranges specified inTable PR22-1. The physical orientation and surface area of theelectrode, and the measured solution voltage set by the elec-tronics of the Type TB82EC transmitter determine the cell con-stants of four-electrode sensors. Therefore, ABB recommendsthe use of only ABB conductivity sensors with the TypeTB82EC transmitter.

1. The currently configured sensor group appears. Press theNEXT smart key until the desired sensor group appears.

2. Press the ENTER smart key to accept the new value.

3. If this is a conductivity configuration, press the NEXTsmart key to go to the next configuration parameter as shownin the basic configuration sequence in Figure 8-6. If this is aconcentration configuration, the display advances to the con-figuration option selection as shown in the advanced configu-ration sequence flowchart in Figure 8-5.

1 min.

Table PR22-1. Sensor Ranges

Group Maximum Range

A 0 to 1,999 mS/cm

B 0 to 1,999 µS/cm

C 0 to 199.9 µS/cm

Page 132: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR23 - 1

PROCEDURE PR23 - CONFIGURATION OPTION SELECTION

PURPOSE/SCOPE

This procedure describes how to select either a predefined oruser-defined concentration configuration. This setting is avail-able in the advanced programming mode and applies to theconcentration analyzer state.

Parts None.

Tools None.

PROCEDURE

The concentration state converts conductivity values to con-centration units. This state applies temperature compensatedconductivity measurements to a predefined or user-definedfunction. These functions convert conductivity measurementsto concentration values having a fixed decimal point locationso that decimal point ranging will not occur. The concentrationstate provides the following predefined configurations:

• 0 to 15% NaOH.• 0 to 20% NaCl.• 0 to 18% HCl.• 0 to 20% H2SO4.

NOTE: The user-defined configuration option is the only one avail-able for Sensor Groups B and C.

These predefined configurations are based on data containedin the International Critical Tables. Refer to Appendix B formore information on concentration programming.

The user-defined configuration provides the capability ofselecting an engineering unit icon, the decimal point position,a custom text description, and a six-point linear curve fit.These topics are discussed in subsequent procedures.

1. Press the NEXT smart key until the desired configurationoption appears.

2. Press the ENTER smart key to accept the new value.

1 min.

Page 133: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR24 - 1

PROCEDURE PR24 - ENGINEERING UNITS SELECTION

PURPOSE/SCOPE

This procedure describes how to configure the engineeringunit icon. This setting is available in the advanced program-ming mode and applies to user-defined configurations in theconcentration analyzer state.

Parts None.

Tools None.

PROCEDURE

The concentration state allows for conversion of conductivityinputs to concentration units such as percent, parts per mil-lion (ppm), and parts per billion (ppb). It is also possible toselect no engineering unit icon.

1. Press the NEXT smart key until the desired engineeringunit icon appears.

2. Press the ENTER smart key to accept the new value.

1 min.

Page 134: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR25 - 1

PROCEDURE PR25 - DECIMAL POINT LOCATION SELECTION

PURPOSE/SCOPE

This procedure describes how to configure the decimal pointlocation. This setting is available in the advanced program-ming mode and applies to user-defined configurations in theconcentration analyzer state.

Parts None.

Tools None.

PROCEDURE

1. Press the smart key to move the decimal point to thedesired location.

2. Press the ENTER smart key to accept the new position.

1 min.

Page 135: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR26 - 1

PROCEDURE PR26 - TEXT STRING DEFINITION

PURPOSE/SCOPE

This procedure describes how to define and enter a customtext description. This setting is available in the advanced pro-gramming mode and applies to user-defined configurations inthe concentration analyzer state.

Parts None.

Tools None.

PROCEDURE

The text string can be up to six characters consisting of anycombination of A through Z, 1 through 9, a space, and a dash.

1. Use the smart key to increment the character.

2. When the first character is correct, use the smart key tomove to the next character.

3. Repeat Steps 1 and 2 for each character.

4. Press the ENTER smart key to accept the text string.

2 min.

Page 136: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR27 - 1

PROCEDURE PR27 - LINEAR CURVE FIT PROGRAMMING

PURPOSE/SCOPE

This procedure shows how to define the linear curve fit for aconductivity-to-concentration conversion. This setting is avail-able in the advanced programming mode and applies touser-defined configurations in the concentration analyzerstate.

Parts None.

Tools None.

PROCEDURE

The six-point linear curve fit sets the endpoint and breakpointvalues for the conductivity-to-concentration conversion. Theend point values define the full scale output range. The break-points identify the transition points between the five line seg-ments defining the conductivity-to-concentration curve.

A plot of temperature compensated conductivity versus soluteconcentration is needed before beginning the procedure. Anexample of a linear approximation for zero to 45-percentammonium nitrate (NH4NO3) is shown in Table PR27-1 andFigure PR27-1. Table PR27-1 shows the data andFIgure PR27-1 shows the linear approximation.

The curve is a nonlinear function divided into five line seg-ments. The endpoints represent points one and six. The break-points represent points two through five. The transmitteroutput is linear relative to the NH4NO3 concentration.

The endpoints define the full scale output range; therefore,rerange of the output is restrained to the range between pointsone and six (zero and 45-percent NH4NO3). Refer toAppendix B for more information on concentrationprogramming.

1. Obtain a plot of temperature compensated conductivityversus solute concentration.

2. Divide the plot into five line segments that best approxi-mate the shape of the conductivity-to-concentration curve. Thebeginning of the first and end of the fifth line segment identifythe endpoints of the linear approximation and output range.

10 min.

Page 137: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

LINEAR CURVE FIT PROGRAMMING

PROCEDUREPR27 - 2 WBPEEUI520003A1

3. Set the first end point conductivity value (X1.COND).

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 3a and 3b for each digit.

d. Press the ENTER smart key to accept the new value.

4. Set the first end point concentration value (Y1.CONC) thatrepresents the endpoint conductivity value entered in Step 3.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 4a and 4b for each digit.

d. Press the ENTER smart key to accept the new value.

Table PR27-1. Conductivity-to-Concentration Data

Point Conductivity (mS/cm) Concentration (%) Output (mA)

1 0 0 4.0

2 55 5 5.8

3 105 9 7.2

4 195 16 9.7

5 310 28 14.0

6 400 45 20.0

Figure PR27-1. Linear Curve Fit

T02413A

400350300250200

C O N D U C TIV ITY (m S /C M )

A M M O N IU MN ITR ATE

(% )

150100

50

45

40

35

30

25

20

15

10

5

0500

B R E A K P O IN TS

AC TUA L C O N D U C TIV ITY-TO -C O N C E N TR ATIO N C U RV E

Page 138: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

LINEAR CURVE FIT PROGRAMMING

PROCEDUREWBPEEUI520003A1 PR27 - 3

5. Repeat Steps 3 and 4 for the four breakpoints and the lastendpoint (X2.COND through X6.COND, and Y2.CONC throughY6.CONC).

6. Press the NEXT smart key to go on to the next configura-tion parameter.

Page 139: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR28 - 1

PROCEDURE PR28 - TEMPERATURE SENSOR TYPESELECTION

PURPOSE/SCOPE

This procedure describes how to configure the temperaturesensor state. This setting is available in both the basic andadvanced programming modes.

Parts None.

Tools None.

PROCEDURE

The temperature sensor state configures the temperatureinput for a Pt 100, three-kilohm Balco RTD, 4.75-kilohm RTDnetwork, or for no temperature input (none).

1. Press the SELECT smart key to modify the TMP.SNS state.

2. The currently configured temperature sensor state appearsfirst. Press the NEXT smart key until the desired temperaturesensor state appears. Choose between NONE, 3K.BLCO,PT100, and 4.75K.RTD.

3. Press the ENTER smart key to accept the new value.

1 min.

Page 140: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR29 - 1

PROCEDURE PR29 - TEMPERATURE COMPENSATION TYPESELECTION

PURPOSE/SCOPE

This procedure describes the types of temperature compensa-tion available and how to choose them.

Parts None.

Tools None.

PROCEDURE

Temperature has a marked effect on the conductance of manysolutions. The effect is generally nonlinear and dependent onthe particular ionic species and their concentration.

The transmitter software has a number of preprogrammed cor-rection algorithms that compensate the effect of temperatureon conductivity to a reference temperature of 25-degrees Cel-sius. This results in accurate and stable measurements whentemperature varies.

Options for temperature compensation are grouped into twosets: manual and auto. Manual temperature compensationcontains no additional options. It is locked to a specific pro-cess temperature independent of the selected temperaturesensor. If desiring a different process temperature, adjust itduring a temperature calibration.

Auto temperature compensation options are sensor groupdependent. They use temperature values measured by thetransmitter temperature input. Compensation algorithmsinclude: standard KCl (0.1N KCl based), solution coefficient,zero to 15-percent NaOH, zero to 20-percent NaCl, zero to18-percent HCl, zero to 20-percent H2SO4, pure water neutralsalt, pure water trace acid, pure water trace base, anduser-defined.

The user-defined temperature compensation option requiresuncompensated conductivity data from the reference tempera-ture of 25-degrees Celsius to the maximum process tempera-ture on a representative sample of process solution. This datais used to calculate the ratio of uncompensated conductivity toconductivity at the reference temperature of 25-degrees Cel-sius. These ratios are then plotted against temperature.

Refer to Appendix A for more information on temperaturecompensation.

1 min.

Page 141: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE COMPENSATION TYPE SELECTION

PROCEDUREPR29 - 2 WBPEEUI520003A1

Table PR29-1 describes the function and programming modeof each state.

1. Press the SELECT smart key to modify the TC.TYPE state.

2. Press the NEXT smart key to toggle between MANUAL andAUTO.

3. Press the ENTER smart key to accept the desired value. Ifmanual was selected, press the NEXT smart key to go on to thenext configuration parameter as shown in the advancedconfiguration sequence in Figure 8-5 and the basic configura-tion sequence in Figure 8-6. If AUTO was selected, go on withStep 4.

Table PR29-1. Temperature Compensation States

State DisplayProgramming

ModeFunction

Manual MANUAL Basic and Advanced

Employed when using fixed temperature value instead of measured value. Initial value is 25°C. Use temperature calibrate state to change fixed temperature value. Com-pensation is 0.1N KCl based.

Standard potassium chloride (automatic)

STD.KCL Basic and Advanced

Used when temperature sensor provides measured tem-perature value. Compensation is 0.1N KCl based (standard).

Temperature com-pensation coefficient (automatic)

TC.COEF Basic and Advanced

Used when temperature sensor provides measured tem-perature value. Compensation is based on percent change of conductivity at reference temperature (25°C) per °C. The value is adjustable between 0 and 9.99%/°C.

Sodium hydroxide (automatic)

NAOH Advanced Used when temperature sensor provides measured tem-perature value. Compensation is 0 to 15% NaOH based.

Sodium chloride (automatic)

NACL Advanced Used when temperature sensor provides measured tem-perature value. Compensation is 0 to 20% NaCl based.

Hydrochloric acid (automatic)

HCL Advanced Used when temperature sensor provides measured tem-perature value. Compensation is 0 to 18% HCl based.

Sulfuric acid (automatic)

H2SO4 Advanced Used when temperature sensor provides measured tem-perature value. Compensation is 0 to 20% H2SO4 based.

Pure water neutral salt (automatic)

NEUTRL Advanced Used when temperature sensor provides measured tem-perature value. Compensation is pure water with neutral salt based. Restricted to Sensor Group C.

Pure water trace acid (automatic)

ACID Advanced Used when temperature sensor provides measured tem-perature value. Compensation is pure water with acid based. Restricted to Sensor Group C.

Pure water trace base (automatic)

BASE Advanced Used when temperature sensor provides measured tem-perature value. Compensation is pure water with base based. Restricted to Sensor Group C.

User-defined (automatic)

USR.DEF Advanced Used when temperature sensor provides measured tem-perature value. Compensation is defined as ratio of uncompensated conductivity over compensated conduc-tivity for a specific set of temperatures.

Page 142: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TEMPERATURE COMPENSATION TYPE SELECTION

PROCEDUREWBPEEUI520003A1 PR29 - 3

4. Press the NEXT smart key until the desired type of auto-matic temperature compensation appears.

5. If TC.COEF or USR.DEF were selected, go to the next proce-dure in the sequence as shown in Figures 8-5 and 8-6. Other-wise, press the ENTER smart key to accept the desired valueand go on to the next configuration parameter in the sequence.

Page 143: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR30 - 1

PROCEDURE PR30 - TEMPERATURE COMPENSATIONCOEFFICIENT SETTING

PURPOSE/SCOPE

This procedure describes how to configure the temperaturecompensation coefficient. This setting is available in both thebasic and advanced programming modes.

Parts None.

Tools None.

PROCEDURE

The temperature compensation coefficient is used when a tem-perature sensor provides the measured temperature value.Compensation is based on percent change of conductivity at areference temperature (25-degrees Celsius) per degree Celsius.The value is adjustable between zero and 9.99 percent perdegree Celsius. Refer to Appendix A for more information ontemperature compensation.

1. Use the smart key to increment the digit.

2. When the first digit is correct, use the smart key tomove to the next digit.

3. Repeat Steps 1 and 2 for each digit.

4. Press the ENTER smart key to accept the value.

5. Press the NEXT smart key to go on to the next configura-tion parameter.

1 min.

Page 144: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR31 - 1

PROCEDURE PR31 - USER-DEFINED TEMPERATURECOEFFICIENT PROGRAMMING

PURPOSE/SCOPE

This procedure describes how to define the linear curve fit for auser-defined temperature compensation plot. This setting isavailable in the advanced programming mode.

Parts None.

Tools None.

PROCEDURE

The user-defined temperature compensation option requiresuncompensated conductivity data from the reference tempera-ture of 25-degrees Celsius to the maximum process tempera-ture on a representative sample of process solution. This datais used to calculate the ratio of uncompensated conductivity(K) to conductivity at the reference temperature of 25-degreesCelsius (KSTD). These ratios are plotted against temperature.

The curve is a nonlinear function divided into five line seg-ments. The endpoints represent points one and six. The break-points represent points two through five.

Refer to Appendix A for more information on temperaturecompensation.

1. Obtain a plot of temperature versus conductivity ratio.

2. Divide the plot into five line segments that best approxi-mate the shape of the curve. The start of the first and end ofthe fifth segments are the end points of the approximation.

3. Set the first endpoint temperature value (TMP1°C).

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 3a and 3b for each digit.

d. Press the ENTER smart key to accept the new value.

10 min.

Page 145: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

USER-DEFINED TEMPERATURE COEFFICIENT PROGRAMMING

PROCEDUREPR31 - 2 WBPEEUI520003A1

4. Set the first endpoint ratio value (K1/K25) that representsthe endpoint temperature value entered in Step 3.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 4a and 4b for each digit.

d. Press the ENTER smart key to accept the new value.

5. Repeat Steps 3 and 4 to enter the four breakpoints and lastendpoint (TMP2°C through TMP6°C, and K2/K25 throughK6/K25).

6. Press the NEXT smart key to go on to the next configura-tion parameter.

Page 146: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR32 - 1

PROCEDURE PR32 - OUTPUT RANGE SETTING

PURPOSE/SCOPE

This procedure describes how to configure the output state.This setting is available in both the basic and advanced pro-gramming modes.

Parts None.

Tools None.

PROCEDURE

The output state sets the output type and range. The defaultoutput range values are 10 percent of the full scale processvariable for Group A sensors, and full scale process variablefor Groups B and C sensors. If requiring a reverse acting out-put, reverse the four and 20-milliamp process variable values.

1. Press the SELECT smart key to modify the OUTPUT state.

2. The transmitter displays the process variable value for thefour-mA point.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 2a and 2b for each digit.

d. Press the ENTER smart key to accept the new value.

3. The transmitter displays the process variable value for the20-mA point.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 3a and 3b for each digit.

d. Press the ENTER smart key to accept the new value.

4. Press the NEXT smart key to go on to the next configura-tion parameter.

1 min.

Page 147: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR33 - 1

PROCEDURE PR33 - DIAGNOSTICS SELECTION

PURPOSE/SCOPE

This procedure describes how to configure the diagnosticsstate. This setting is available in both the basic and advancedprogramming modes.

Parts None.

Tools None.

PROCEDURE

The diagnostics state allows disabling of the built-in sensordiagnostics.

1. Press the SELECT smart key to modify the DIAG state.

2. Use the smart key to toggle between ON and OFF.

3. Press the ENTER smart key to accept the new value.

1 min.

Page 148: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR34 - 1

PROCEDURE PR34 - SAFE MODE SELECTION

PURPOSE/SCOPE

This procedure describes how to configure the safe mode state.This setting is available in both the basic and advanced pro-gramming modes.

Parts None.

Tools None.

PROCEDURE

The safe mode state determines the output level of the trans-mitter if an error condition occurs that renders the transmitterinoperable. The available states are FAIL.LO (fail low) andFAIL.HI (fail high). More information about error conditions iscontained in Section 12.

1. Press the SELECT smart key to modify the SAF.MOD state.

2. Use the NEXT smart key to toggle between FAIL.HI andFAIL.LO.

3. Press the ENTER smart key to accept the new value.

4. If this is an advanced configuration, go to Step 6. If this isa basic configuration press the exit to MEASURE smart key andgo on to Step 5.

5. When the SAVE? screen comes up, press the YES smartkey to accept the configuration. Press the NO key to abort theconfiguration. In either case, the transmitter goes to the mea-sure mode. Do not perform Step 6.

6. Press the NEXT smart key to go on to the next configura-tion parameter.

1 min.

Page 149: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR35 - 1

PROCEDURE PR35 - SPIKE MAGNITUDE SETTING

PURPOSE/SCOPE

This procedure describes how to configure the spike magni-tude state. This setting is available in the advanced program-ming mode.

Parts None.

Tools None.

PROCEDURE

The spike state sets the diagnostic spike level as a percent ofoutput. This level determines the magnitude of the spike.

When the spike magnitude has been set to any level greaterthan zero percent and is enabled in the spike output state, thetransmitter modulates the output signal by the configuredlevel for one second out of every six seconds. Using this modu-lation, the transmitter informs the operator of a detected diag-nostic condition.

Enter the spike magnitude as a percentage of the 16-milliampoutput range. A 10-percent spike magnitude generates a1.6-milliamp spike, a 20-percent spike magnitude generates a3.2-milliamp spike, etc.

NOTE: Once the spike state is OFF, changing the configured spikelevel in the configure mode will not enable the spike state. The spikestate can only be turned ON or OFF in the output/hold mode.

1. Press the SELECT smart key to modify the SPK.MAG state.

2. The transmitter displays the current spike magnitudevalue.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 2a and 2b for each digit.

d. Press the ENTER smart key to accept the new value.

3. Press the exit to MEASURE smart key.

1 min.

Page 150: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SPIKE MAGNITUDE SETTING

PROCEDUREPR35 - 2 WBPEEUI520003A1

4. When the SAVE? screen comes up, press the YES smartkey to accept the configuration. Press the NO key to abort theconfiguration. In either case, the transmitter goes to the mea-sure mode.

Page 151: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR36 - 1

PROCEDURE PR36 - CONFIGURATION LOCKOUT

PURPOSE/SCOPE

This procedure describes how to set the hardware configura-tion lockout jumper.

Parts None.

Tools • Flat blade screwdriver.• Needle nose pliers.

PROCEDURE

The transmitter has a lockout feature that, once engaged, pro-hibits access to the configure mode. This feature does notaffect parameters that can be changed in the other modes ofoperation: calibrate, output/hold, security, and secondarydisplay.

NOTE: The printed circuit boards contain semiconductor devicesand are subject to damage by static electricity. Refer to SPECIALHANDLING in Section 3 before performing this procedure.

1. Turn off power to the transmitter and allow at least oneminute for it to discharge.

2. Use the flat-blade screwdriver to loosen the four captivescrews that secure the front bezel assembly to the transmittershell.

3. Pull gently on the front bezel assembly to remove it fromthe shell.

4. The microprocessor PCB assembly, which is attached tothe front bezel, contains the configuration lockout jumper.Position A (jumper W1 on pins 1 and 2) is the factory defaultposition and disables the configuration lockout. Position B(jumper W1 on pins 2 and 3) enables the configuration lock-out. Refer to Figure PR36-1 and use the needle nose pliers tochange the jumper to the desired position.

5. Place the front bezel assembly into the shell and pressgently.

6. Use the bladed screwdriver to tighten the four captivescrews.

5 min.

Page 152: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION LOCKOUT

PROCEDUREPR36 - 2 WBPEEUI520003A1

Figure PR36-1. Configuration Lockout Jumper

11

1 2 3

2 3 2 3

W 1 C O N FIG U R ATIO N LO C KO U T

E N A B LE LO C KO U T

D IS AB LE LO C KO U T(FAC TO RYD E FAU LT S E TTIN G )

C U S TO M E R SE TT IN G

W 1

123AB

T01690B

Page 153: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR37 - 1

PROCEDURE PR37 - SECURITY AND PASSWORD ASSIGNMENT

PURPOSE/SCOPE

This procedure describes how to define which modes andstates of operation are security protected. It also describes howto set the password for the protected states and modes of oper-ation, how to remove all security, and how to change securityand the password.

Parts None.

Tools None.

PROCEDURE

This procedure contains three actions. Setting the securityand password, removing all security, and changing the secu-rity and password.

NOTE: In the unlikely event that the password cannot be retrievedand the secured modes and states must be accessed, a reset secu-rity state exists. Refer to PR40 for the reset security procedure.

Setting Security and Password

1. Press the MENU smart key until SECURITY is highlightedon the display.

2. Press the SELECT smart key to enter the security mode.

3. The CALIBR screen appears first. To set the security, pressthe smart key to change the security from OFF to ON andcontinue with Step 4. To leave the security OFF, press theENTER key and go on to Step 5.

4. Press the ENTER smart key to accept the selection.

5. The OUTPUT screen appears. To set the security, press the smart key to change the security from OFF to ON and con-

tinue with Step 6. To leave the security OFF, press the ENTERkey and go on to Step 7.

6. Press the ENTER smart key to accept the selection.

7. The CONFIG screen appears. To set the security, press the smart key to change the security from OFF to ON and con-

tinue with Step 8. To leave the security OFF, press the ENTERkey and go on to Step 9.

8. Press the ENTER smart key to accept the selection.

5 min.

Page 154: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SECURITY AND PASSWORD ASSIGNMENT

PROCEDUREPR37 - 2 WBPEEUI520003A1

9. The PASSWD screen appears with _ _ _ shown. Define thepassword for all secured modes and states.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 9a and 9b for each digit.

d. Press the ENTER smart key to accept the password.

10. The password must now be verified. The PASSWD screenappears with _ _ _ shown.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 10a and 10b for each digit.

d. Press the ENTER smart key to accept the password.

NOTE: The password must be defined as three digits and verified toenable security on the modes and states entered in Steps 3 through8. If security is not ON for any of the modes and states, the transmit-ter bypasses the password screen.

Removing All Security

1. Press the MENU smart key until SECURITY is highlightedon the display.

2. Press the SELECT smart key to enter the security mode.

3. If the security mode has been secured, the transmitterrequires the password and displays _ _ _. To enter thepassword:

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 3a and 3b for each digit.

d. When the password is correct, press the ENTER smartkey.

Page 155: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SECURITY AND PASSWORD ASSIGNMENT

PROCEDUREWBPEEUI520003A1 PR37 - 3

4. The CALIBR screen appears first. Press the smart key tochange the security from ON to OFF.

5. Press the ENTER smart key to accept the selection.

6. The OUTPUT screen appears. Press the smart key tochange the security from ON to OFF.

7. Press the ENTER smart key to accept the selection.

8. The CONFIG screen appears. Press the smart key tochange the security from ON to OFF.

9. Press the ENTER smart key to accept the selection.

Changing Security or Password

1. Press the MENU smart key until SECURITY is highlightedon the display.

2. Press the SELECT smart key to enter the security mode.

3. If the security mode has been secured, the transmitterrequires the password and displays _ _ _. To enter thepassword:

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 3a and 3b for each digit.

d. When the password is correct, press the ENTER smartkey.

4. The CALIBR screen appears first. To change the security,press the smart key to toggle the security between OFF andON and continue with Step 5. To leave the security unchanged,press the ENTER key and go on to Step 6.

5. Press the ENTER smart key to accept the selection.

6. The OUTPUT screen appears. To change the security, pressthe smart key to toggle the security between OFF and ONand continue with Step 7. To leave the security unchanged,press the ENTER key and go on to Step 8.

7. Press the ENTER smart key to accept the selection.

8. The CONFIG screen appears. To change the security, pressthe smart key to toggle the security between OFF and ON

Page 156: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SECURITY AND PASSWORD ASSIGNMENT

PROCEDUREPR37 - 4 WBPEEUI520003A1

and continue with Step 9. To leave the security unchanged,press the ENTER key and go on to Step 10.

9. Press the ENTER smart key to accept the selection.

10. The PASSWD screen appears with _ _ _ shown. Eitherchange the password or enter the existing password to acceptthe changes to the security.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 10a and 10b for each digit.

d. Press the ENTER smart key to accept the password.

11. If the password was changed, it must now be verified. If thepassword was not changed, this procedure is complete. ThePASSWD screen appears with _ _ _ shown.

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 11a and 11b for each digit.

d. Press the ENTER smart key to accept the password.

Page 157: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR38 - 1

PROCEDURE PR38 - UTILITY MODE ADVANCED/BASICPROGRAMMING

PURPOSE/SCOPE

This procedure describes how to change programming modeswhile in the user state.

Parts None.

Tools None.

PROCEDURE

The basic programming mode contains a subset of configura-tion options found in the advanced programming mode. Sepa-ration into two programming groups is advantageous whenlimited functionality is desired. Fewer options reduces confu-sion and the possibility of configuration errors.

Transmitters ordered with the advanced programming optioncan be changed between basic and advanced programming.Transmitters ordered with the basic programming optionrequire an upgrade password to change to advanced program-ming. Contact ABB to obtain the password.

The programming toggle for transmitters with the advancedprogramming option must be set in two locations: user state inthe utility mode and modify configure state in the configuremode. In order to select either the basic or advanced program-ming mode in the modify configure state, the programmingmode must be set to advanced in the user state.

NOTE: ADVNCD is the factory default setting in the modify configurestate in the configure mode and the user state in the utility mode fortransmitters ordered with the advanced programming option.

1. Press the hidden key located above the NT in the ADVAN-TAGE text on the keypad. The prompt USER appears in thesecondary display region.

2. Press the SELECT smart key. The text MODE appears inthe secondary display.

3. Press the SELECT smart key and the BASIC screenappears. Press the ENTER smart key to set the programming tobasic, and advance to the next user state or press the NEXTsmart key to advance to the next screen and go on to Step 4.

4. The ADVNCD screen appears. Press the ENTER smart keyto set the programming to advanced and advance to the nextuser state.

1 min.

Page 158: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR39 - 1

PROCEDURE PR39 - CONFIGURATION RESET

PURPOSE/SCOPE

This procedure describes how to reset the configuration backto the factory default settings.

Parts None.

Tools None.

PROCEDURE

Table PR39-1 lists the factory software default configurationsettings.

1. Press the hidden key located above the NT in the ADVAN-TAGE text on the keypad. The prompt USER appears in thesecondary display region.

2. Press the SELECT smart key. The text MODE appears inthe secondary display.

3. Press the NEXT smart key until RST.CON appears in thesecondary display.

4. Press the SELECT smart key to reset the configuration.

1 min. 1 min.

Table PR39-1. Factory Software Defaults

Parameter Default

Instrument mode Basic

Analyzer type Conductivity, Sensor Group A

Temperature sensor 3-kΩ Balco

Temperature compensation Manual

Output range 0.00 to 199.9 mS/cm

Damping value 0.5 secs

Sensor diagnostics state Disable

Safe mode fail output state Low

Spike output level1 0%NOTE:1. This function only available in advanced programming mode.

Page 159: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

CONFIGURATION RESET

PROCEDUREPR39 - 2 WBPEEUI520003A1

5. If the configure mode has been secured, the transmitterrequires the password and displays _ _ _. To enter thepassword:

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value is correct, use the smartkey to move to the next digit.

c. Repeat Steps 5a and 5b for each digit.

d. When the password is correct, press the ENTER smartkey.

6. The text RESET? appears. Press the YES smart key to con-firm the reset operation or the NO smart key to abort theprocedure.

Page 160: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR40 - 1

PROCEDURE PR40 - SECURITY RESET

PURPOSE/SCOPE

This procedure describes how to reset the security back to thefactory default settings. The factory defaults are security off forall applicable modes (calibrate, output/hold, and modifyconfigure).

Parts None.

Tools None.

PROCEDURE

1. Press the hidden key located above the NT in the ADVAN-TAGE text on the keypad. The prompt USER appears in thesecondary display region.

2. Press the SELECT smart key. The text MODE appears inthe secondary display.

3. Press the NEXT smart key until RST.SEC appears in thesecondary display.

4. Press the SELECT smart key to reset the security.

5. The transmitter displays _ _ _. The security reset passwordis 732. To enter the password:

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value reaches 7, use the smartkey to move to the next digit.

c. Use the smart key to increment the value of theblinking digit.

d. When the second digit value reaches 3, use the smart key to move to the next digit.

e. Use the smart key to increment the value of theblinking digit.

f. When the third digit value reaches 2, press the ENTERsmart key.

6. The text RESET? appears. Press the YES smart key to con-firm the reset operation or the NO smart key to abort theprocedure.

1 min.

Page 161: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR41 - 1

PROCEDURE PR41 - RESETTING ALL PARAMETERS

PURPOSE/SCOPE

This procedure describes how to reset all transmitter valuesback to the factory default settings. This includes calibration,output/hold, configuration, security, and secondary displayvalues.

Parts None.

Tools None.

PROCEDURE

1. Press the hidden key located above the NT in the ADVAN-TAGE text on the keypad. The prompt USER appears in thesecondary display region.

2. Press the SELECT smart key. The text MODE appears inthe secondary display.

3. Press the NEXT smart key until RST.ALL appears in thesecondary display.

4. Press the SELECT smart key to reset all transmitterparameters.

5. The transmitter displays _ _ _. The reset all password is255. To enter the password:

a. Use the smart key to increment the value of theblinking digit.

b. When the first digit value reaches 2, use the smartkey to move to the next digit.

c. Use the smart key to increment the value of theblinking digit.

d. When the second digit value reaches 5, use the smart key to move to the next digit.

e. Use the smart key to increment the value of theblinking digit.

f. When the third digit value reaches 5, press the ENTERsmart key.

1 min.

Page 162: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

RESETTING ALL PARAMETERS

PROCEDUREPR41 - 2 WBPEEUI520003A1

6. The text RESET? appears. Press the YES smart key to con-firm the reset operation or the NO smart key to abort theprocedure.

Page 163: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR42 - 1

PROCEDURE PR42 - TRANSMITTER SOFT BOOT

PURPOSE/SCOPE

This procedure describes how to perform a transmitterself-test without affecting existing parameters.

Parts None.

Tools None.

PROCEDURE

The soft boot user state initiates a self-test. All programmabletransmitter parameters remain unchanged after performingthe soft boot.

1. Press the hidden key located above the NT in the ADVAN-TAGE text on the keypad. The prompt USER appears in thesecondary display region.

2. Press the SELECT smart key. The text MODE appears inthe secondary display.

3. Press the NEXT smart key until RST.SFT appears in thesecondary display.

4. Press the SELECT smart key to initiate the self-test.

5. The text RESET? appears. Press the YES smart key to con-firm the self-test or the NO smart key to abort the procedure.

1 min. 1 min.

Page 164: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR43 - 1

PROCEDURE PR43 - SENSOR INSPECTION

PURPOSE/SCOPE

This procedure describes how to visually inspect the sensor.

Parts None.

Tools None.

PROCEDURE

If the sensor is suspected of being the source of problems, aquick visual inspection can identify the problem.

Sensor Body

1. Remove the sensor from the process.

2. Inspect the sensor body for cracks and distortions.

3. If cracks or distortions exist, contact ABB for alternativesensor styles and materials.

Cable and Connectors

1. Inspect the sensor cable for cracks, cuts, or shorts.

2. If using a junction box or extension cable, check for mois-ture, oil, corrosion, or particulates. All connections must bedry and free of oil, corrosion, and particulates. Even slightamounts of these contaminants can short sensor signals andaffect conductivity.

3. Check to see that all wiring is dry and not shorting againstany metal, conduit, or earth grounds.

O-Ring Seals

1. Inspect the sealing O-rings for attack by the process liquid.

2. If the O-rings show evidence of corrosion, deterioration, ordistortion, contact ABB for alternate material choices.

10 min.

Page 165: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR44 - 1

PROCEDURE PR44 - ELECTRONIC SENSOR TEST

PURPOSE/SCOPE

This procedure describes how to run the electronic sensor test.

Parts None.

Tools • Digital multimeter (DMM) with a conductance functioncapable of reading zero to 200 nanosiemens, and a resis-tance function capable of reading zero to 20 kilohms.

• Temperature measuring device.

PROCEDURE

There is an electronic sensor test to verify the integrity of con-ductivity sensor elements and associated cable.

1. Disconnect the sensor leads and automatic temperaturecompensator leads from the transmitter.

2. Use the DMM set to measure ohms to measure the resis-tance between the yellow and blue temperature compensatorleads.

a. If using a three-kΩ Balco RTD, the expected resistancecan be calculated from:

where T is the temperature in degrees Celsius. The mea-sured resistance should be within the expected value by±15%.

b. If using a PT 100 RTD, the expected resistance can becalculated from:

where T is the temperature in degrees Celsius. The mea-sured resistance should be within the expected value by±5%.

c. If using a 4.75-kΩ RTD network, the expected resis-tance can be calculated from:

where T is the temperature in degrees Celsius. The mea-sured resistance should be within the expected value by±20%.

10 min.

RTC T 0°C–( ) 0.0045×( ) 1+( ) 3 000,×=

RTC 100 T 0°C–( ) 0.385×( )+=

RTC 6 329, 104.5 T 0.774 T 2 0.0026 T 3 0.000003 T 4×+×–×+×–=

Page 166: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

ELECTRONIC SENSOR TEST

PROCEDUREPR44 - 2 WBPEEUI520003A1

3. Moisture intrusion behind the sensor electrode seal can bedetected. Remove the sensor from the process and dry itthoroughly.

4. Use the DMM set to measure conductance, to measure theconductance between the yellow temperature compensatorlead and each of the other sensor leads (green, red, white,black, and heavy green). The reading must be less than0.05 nS.

5. Check the conductance between the green drive lead andeach of the other sensor leads (black, white, red, and heavygreen). The reading must be less than 0.05 nS.

6. Check the conductance between the heavy green shieldlead and each of the other sensor leads (blue, yellow, black,green, red and white). The reading must be less than 0.05 nS.

Page 167: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR45 - 1

PROCEDURE PR45 - TRANSMITTER TROUBLESHOOTINGWITHOUT SENSOR

PURPOSE/SCOPE

This procedure describes how to troubleshoot the transmitterwith the sensor disconnected. Sensor simulation is an easyway to check the operation of the transmitter without thesensor.

Parts None.

Tools • Decade resistance box.• 3000-Ω resistor.

PROCEDURE

1. Remove the transmitter from the process and disconnectthe sensor.

NOTE: The transmitter calibration values must be set back to thoseentered at the factory in order for this procedure to be valid.

2. There are two ways to perform sensor simulation. Eitherconnect a 3000-Ω resistor across the temperature sensorinput as shown in Figure PR45-1, or configure the transmitterfor manual temperature compensation with a constant 25°Ctemperature.

3. Connect the transmitter in the setup shown inFigure PR45-1.

10 min.

Figure PR45-1. Sensor Simulation Setup

T 02817C

S E N S O R C O N N E C TIO N S P OW E R /O U T P U TC O N N E C T IO N S

T B 2

4-20m AS IG N A L

+1

–2

+3

–4

G R E E N

R E D

W H ITE

B LAC K

B LU E

Y E LLOW

H V Y G R N

1

2

3

4

5

6

7

8

D R IV E

D R IV E

S E N S E

S H IE LD

RTD

RTD

S E N S E

A M M E TE RTE R M IN A LS

TB 1

P OW E R

T E ST

R EAR V IE W

3,000 ,1 ,000 ,

O R 100

ΩΩ

Ω

D EC A D ER ESIS TA N C E

BOX

Page 168: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

TRANSMITTER TROUBLESHOOTING WITHOUT SENSOR

PROCEDUREPR45 - 2 WBPEEUI520003A1

4. Set the decade resistance box to the values indicated inTable PR45-1 and record the displayed value, comparing it tothe value indicated in the desired display value column.

NOTE: The reset calibration feature resets all calibration values.Before putting the transmitter back into service, be sure to perform atemperature, process, and output calibration.

Table PR45-1. Sensor Simulation Values

Sensor GroupVariable

Resistance (Ω)Desired

Display ValueDisplayed Value

A Open 0.000 µS/cm

5,000.0 100.0 µS/cm

1,000.0 500.0 µS/cm

500.0 1,000 µS/cm

100.0 5.000 mS/cm

50.0 10.00 mS/cm

10.01 50.00 mS/cm

5.01 100.0 mS/cm

1.01 500.0 mS/cm

0.51 1,000 mS/cm

B Open 0.000 µS/cm

50,000 1.000 µS/cm

5,000 10.00 µS/cm

1,000 50.00 µS/cm

500 100.0 µS/cm

100 500.0 µS/cm

50 1000 µS/cm

C Open 0.000 µS/cm

100,000 0.050 µS/cm

50,000 0.100 µS/cm

5,000 1.000 µS/cm

1,000 5.000 µS/cm

500 10.00 µS/cm

100 50.00 µS/cm

50 100.0 µS/cmNOTE:1. Leadwire and contact resistance becomes significant part of simulated conductivity and mustbe considered in order to obtain indications close to desired display values.

Page 169: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR46 - 1

PROCEDURE PR46 - SENSOR CLEANING

PURPOSE/SCOPE

This procedure describes how to clean the sensor.

Parts None.

Tools • Gloves.• Eye protection.• Safety shield.• Other protective items as applicable.• 1% to 5% hydrochloric acid (HCl) solution (for acid dip).• Isopropyl alcohol or other appropriate solvent (for solvent

dip).• Clean cloth.• Rag, acid brush, or toothbrush (for physical cleaning).• Water.

SAFETY CONSIDERATIONS

PROCEDURE

ABB conductivity sensors are cleaned using one or a combina-tion of methods. These are recommendations and may not besuitable for all applications. Other cleaning methods may bedeveloped that better suit particular applications. When clean-ing, observe all safety precautions required for handling chem-icals. When handling chemicals, always use gloves, eyeprotection, safety shields, and similar protective items, andconsult material data safety sheets.

20 min.

WARNING

1. Consider the material compatibility between cleaning fluidsand process liquids. Incompatible fluids can react with eachother causing injury to personnel and equipment damage.

2. Acids and bases can cause severe burns. Use hand and eyeprotection when handling.

3. Use solvents only in well ventilated areas. Avoid prolongedor repeated breathing of vapors or contact with skin. Solventscan cause nausea, dizziness, and skin irritation. In somecases, overexposure to solvents has caused nerve and braindamage. Solvents are flammable - do not use near extremeheat or open flame.

Page 170: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

SENSOR CLEANING

PROCEDUREPR46 - 2 WBPEEUI520003A1

Acid Dip

This method removes scales caused by hard water.

1. Verify that any process fluid on the sensor is compatiblewith HCl.

2. Put on gloves, eye protection, safety shields, and other pro-tective items as needed.

3. Dip the tip of the sensor into a 1-percent to 5-percent solu-tion of HCl until this region is free of the unwanted coating.Minimize exposure of any of the metal on the sensor to thissolution or corrosion may occur.

4. Rinse the sensor with clean water.

Solvent Dip

This method removes organic coatings.

1. Verify that any process fluid on the sensor is compatiblewith isopropyl alcohol or other appropriate solvent.

2. Put on gloves, eye protection, safety shields, and other pro-tective items as needed.

3. Dip the sensor into the solvent. Do not use a solvent that isknown to be incompatible with the sensor.

4. Remove the solvent using a clean cloth.

Physical Cleaning

This method removes especially thick scales andaccumulations.

1. Use a rag, acid brush, or toothbrush to clean the sensor.

W1

W2

W1

W3

Page 171: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR47 - 1

PROCEDURE PR47 - KEYPAD CLEANING

PURPOSE/SCOPE

This procedure describes how to clean the keypad.

Parts None.

Tools • Soft, lint-free cloth.• Mild soap.• Warm water.

PROCEDURE

1. Mix mild soap into warm water according to the soap man-ufacturer’s instructions.

2. Dampen the cloth with the soap and water mixture andwring out excess liquid.

3. Gently wash off the keypad with the cloth.

4. Allow to air dry.

2 min.

Page 172: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR48 - 1

PROCEDURE PR48 - FRONT BEZEL REMOVAL

PURPOSE/SCOPE

This procedure describes how to remove the front bezel.

Parts None.

Tools • Flat-blade screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Remove power from the transmitter and allow at least oneminute for it to discharge.

2. Use the flat-blade screwdriver to loosen the four captivescrews that secure the front bezel to the transmitter shell.

3. Pull gently on the front bezel to remove it from the shell.

2 min.

Page 173: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR49 - 1

PROCEDURE PR49 - POWER SUPPLY PCB REMOVAL

PURPOSE/SCOPE

This procedure describes how to remove the power supplyPCB.

Parts None.

Tools • Phillips screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Use the Phillips screwdriver to remove the two screws thatretain the power supply PCB (Fig. PR49-1).

2. Gently pull on the power supply PCB to disengage it fromthe microprocessor PCB.

2 min.

Figure PR49-1. Power Supply PCB Removal

FRO N T B E Z E LK IT

S H E LL K IT

1/2-IN . L IQ U ID T ITEFIT T IN G K IT

M IC RO PRO C ESS O RP C B A S S E M B LY K IT

P OW E R S U P P LYP C B A S S E M B LY

C O N D U C TIV ITY IN P U TP C B A S S E M B LY K IT T02412A

Page 174: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR50 - 1

PROCEDURE PR50 - CONDUCTIVITY INPUT PCB REMOVAL

PURPOSE/SCOPE

This procedure describes how to remove the conductivity inputPCB.

Parts None.

Tools • Phillips screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Use the Phillips screwdriver to remove the two screws thatretain the conductivity input PCB (Fig. PR50-1).

2. Gently pull on the conductivity input PCB to disengage itfrom the microprocessor PCB.

2 min.

Figure PR50-1. Conductivity Input PCB Removal

FRO N T B E Z E LK IT

S H E LL K IT

1/2-IN . L IQ U ID T ITEFIT T IN G K IT

M IC RO P RO C E S S O RP C B A S S E M B LY K IT

P OW E R S U P P LYP C B A S S E M B LY

C O N D U C TIV ITY IN P U TP C B A S S E M B LY K IT T02412A

Page 175: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR51 - 1

PROCEDURE PR51 - MICROPROCESSOR PCB REPLACEMENT

PURPOSE/SCOPE

This procedure describes how to replace the microprocessorPCB.

Parts

Tools • Phillips screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Release the keypad ribbon cable connector latch by push-ing the outside of the connector and lightly pulling outwards.

2. Use the Phillips screwdriver to remove the two screws thatsecure the microprocessor PCB to the front bezel (Fig. PR51-1).

2 min.

Number Qty Description

4TB9515-0178 1 Microprocessor PCB

Figure PR51-1. Microprocessor PCB Replacement

FRO N T B EZE LKIT

S H E LL K IT

1 /2-IN . L IQ U ID T ITEFITT IN G K IT

M IC RO PRO C ES S O RP C B A SS EM B LY K IT

POW E R SU PP LYPC B A S SE M BLY

C O N D U C TIV ITY IN PU TPC B A S SE M BLY K IT T02412A

Page 176: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

MICROPROCESSOR PCB REPLACEMENT

PROCEDUREPR51 - 2 WBPEEUI520003A1

3. Remove the microprocessor PCB.

4. Install the new microprocessor PCB.

5. Install the two screws to secure the microprocessor PCB tothe front bezel and tighten them with the Phillips screwdriver.

6. Insert the keypad ribbon cable into the mating connectorand push connector locks into place.

Page 177: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR52 - 1

PROCEDURE PR52 - CONDUCTIVITY INPUT PCBINSTALLATION

PURPOSE/SCOPE

This procedure describes how to install the conductivity inputPCB.

Parts

Tools • Phillips screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Insert the header on the conductivity input PCB into theproper connector on the microprocessor PCB.

2. Install the two screws to retain the conductivity input PCBand tighten them with the Phillips screwdriver.

2 min.

Number Qty Description

4TB9515-0176 1 Conductivity input PCB

Page 178: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR53 - 1

PROCEDURE PR53 - POWER SUPPLY PCB INSTALLATION

PURPOSE/SCOPE

This procedure describes how to install the power supply PCB.

Parts

Tools • Phillips screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Insert the header on the power supply PCB into the properconnector on the microprocessor PCB.

2. Install the two screws to retain the power supply PCB andtighten them with the Phillips screwdriver.

2 min.

Number Qty Description

4TB9515-0155 1 Power supply PCB

4TB9515-0157 Power supply PCB for HART compatible transmitters

4TB9515-0158 Power supply PCB with lightning arrestor

4TB9515-0159 Power supply PCB with lighting arrestor for HART compatible transmitters

Page 179: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR54 - 1

PROCEDURE PR54 - FRONT BEZEL INSTALLATION

PURPOSE/SCOPE

This procedure describes how to install the front bezel.

Parts

Tools • Flat-blade screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Insert the front bezel with electronics assembly into theshell and press gently to engage terminal block connectors.

2. Use the flat-blade screwdriver to tighten the four captivescrews.

2 min.

Number Qty Description

4TB9515-0160 1 Front bezel

Page 180: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR55 - 1

PROCEDURE PR55 - REAR COVER REPLACEMENT

PURPOSE/SCOPE

This procedure describes how to replace the rear cover.

Parts

Tools • Flat-blade screwdriver.

PROCEDURE

NOTE: Refer to SPECIAL HANDLING in Section 3 before perform-ing this procedure.

1. Remove power from the transmitter and allow at least oneminute for it to discharge.

2. Use the flat-blade screwdriver to loosen the four captivescrews that secure the rear cover to the transmitter shell.

3. Pull gently on the rear cover to remove it from the shell.

4. Install the rear cover onto the shell.

5. Use the flat-blade screwdriver to tighten the four captivescrews.

2 min.

Number Qty Description

4TB9515-0162 1 Rear cover

Page 181: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION
Page 182: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

PURPOSE/SCOPEWBPEEUI520003A1 PR56 - 1

PROCEDURE PR56 - SHELL REPLACEMENT

PURPOSE/SCOPE

This procedure is for part number reference only. The repairsequence flowchart (Fig. 14-1) includes the procedures neces-sary to replace the transmitter shell.

Parts

Tools None.

1 min.

Number Qty Description

4TB9515-0175 1 Shell

Page 183: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1 Index - 1

Index

A

Accessories .............................................................. 1-10Accuracy..................................................................... 1-7Advanced configuration

Analyzer type................................................. PR21-1Configuration option....................................... PR23-1Damping ........................................................ PR13-2Decimal point location.................................... PR25-1Diagnostics .................................................... PR33-1Engineering units ........................................... PR24-1Linear curve fit ............................................... PR27-1Modifying while viewing ................................. PR19-1Output range.................................................. PR32-1Programming mode change .......................... PR20-1Safe mode ..................................................... PR34-1Sensor group ................................................. PR22-1Spike magnitude............................................ PR35-1Temperature compensation coefficient.......... PR30-1Temperature sensor type............................... PR28-1Text string ...................................................... PR26-1User-defined temperature compensation ...... PR31-1Viewing .......................................................... PR18-1

Ambient temperature effect ........................................ 1-8Analyzer type................................................ 8-3, PR21-1

B

Basic configurationAnalyzer type................................................. PR21-1Damping ........................................................ PR13-2Diagnostics .................................................... PR33-1Modifying while viewing ................................. PR19-1Output range.................................................. PR32-1Programming mode change .......................... PR20-1Safe mode ..................................................... PR34-1Sensor group ................................................. PR22-1Temperature compensation coefficient.......... PR30-1Temperature sensor type............................... PR28-1Viewing .......................................................... PR18-1

Basic/advanced configuration ........................... 8-2, 11-3Boredom switch.......................................................... 5-1

C

CalibrationConductivity/concentration........................ 6-1, PR9-1Diagnostics ......................................................... 12-7Edit.......................................................... 6-1, PR11-1Output ..................................................... 6-1, PR10-1Reset ...................................................... 6-1, PR12-1Sequence.............................................................. 6-2States.................................................................... 6-1

Temperature..............................................6-1, PR8-1Changing password .............................................PR37-3Changing security ................................................PR37-3Cleaning

Keypad ...........................................................PR47-1Sensor............................................................PR46-1

ConcentrationCalibration ........................................................PR9-1

Concentration analyzer type .......................................8-5Concentration programming ...................................... B-1Conductivity

Calibration ........................................................PR9-1Conductivity analyzer type ..........................................8-3Conductivity input PCB installation ......................PR52-1Conductivity input PCB removal ..........................PR50-1Conductivity/concentration calibration ........................6-1Conductivity-to-concentration conversion ............PR27-1Configuration

AdvancedAnalyzer type............................................PR21-1Configuration option .................................PR23-1Damping ...................................................PR13-2Decimal point location ..............................PR25-1Diagnostics ...............................................PR33-1Engineering units......................................PR24-1Linear curve fit ..........................................PR27-1Modifying while viewing............................PR19-1Output range ............................................PR32-1Programming mode change .....................PR20-1Safe mode ................................................PR34-1Sensor group............................................PR22-1Spike magnitude.......................................PR35-1Temperature compensation coefficient ....PR30-1Temperature sensor type .........................PR28-1Text string.................................................PR26-1User-defined temperature compensation .PR31-1Viewing.....................................................PR18-1

Analyzer type ........................................................8-3Basic

Analyzer type............................................PR21-1Damping ...................................................PR13-2Diagnostics ...............................................PR33-1Modifying while viewing............................PR19-1Output range ............................................PR32-1Programming mode change .....................PR20-1Safe mode ................................................PR34-1Sensor group............................................PR22-1Temperature compensation coefficient ....PR30-1Temperature sensor type .........................PR28-1Viewing.....................................................PR18-1

Basic/advanced.....................................................8-2Concentration analyzer type .................................8-5

Page 184: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

Index - 2 WBPEEUI520003A1

Index (continued)

Conductivity analyzer type ....................................8-3Damping................................................................8-7Diagnostics............................................................8-7Lockout...........................................................PR36-1Modifying .................................................8-3, PR18-1Option.............................................................PR23-1Output....................................................................8-6Reset .....................................................11-3, PR39-1Safe mode.............................................................8-7Sequence ..............................................................8-7Spike output ..........................................................8-7Temperature compensation ..................................8-5Temperature sensor ..............................................8-5View/modify ...........................................................8-1Viewing...........................................................PR18-1

D

Damping....................1-8, 2-2, 7-2, 8-7, PR13-1, PR13-2Decimal point location..........................................PR25-1Diagnostics............................................ 2-2, 8-7, PR33-1

Calibration ...........................................................12-7Fault codes..........................................................12-1

Error codes ...........................................12-6, 12-7Problem codes, common...............................12-2Problem codes, uncommon...........................12-5

Spike ..............................................................PR17-1Display resolution........................................................1-7Dynamic response ......................................................1-8

E

Edit calibration...............................................6-1, PR11-1Electronic sensor test...........................................PR44-1Engineering units .................................................PR24-1Entering password ......................................................9-1Environmental .............................................................1-8Error codes.......................................................12-6, 12-7

F

Fault codes ...............................................................12-1Error codes.................................................12-6, 12-7Problem codes ....................................................12-1

Common........................................................12-2Uncommon ....................................................12-5

Fault icon ....................................................................4-4Fault info .....................................................................5-2Features......................................................................1-2Front bezel installation .........................................PR54-1Front bezel removal .............................................PR48-1Function flow tree....................................................... C-3

G

Grounding.............................................................. PR7-1

H

Hardware lockout................................................. PR36-1Hazardous locations ................................................... 3-3Hinge mounting ..................................................... PR4-1Hold icon..................................................................... 4-4Hold output .......................................................... PR14-1

Release.......................................................... PR15-1Hold/release output .................................................... 7-1How to use this instruction.......................................... 1-5

I

Input types .................................................................. 1-7Inspection ................................................................... 3-2Installation

Configuration lockout ..................................... PR36-1Grounding ........................................................ PR7-1Mounting

Hinge.......................................................... PR4-1Panel .......................................................... PR2-1Pipe............................................................ PR1-1Rotation........................................................... 3-4Wall ............................................................ PR3-1

Sequence.............................................................. 3-5Wiring

Sensor........................................................ PR6-1Signal and power ....................................... PR5-1

Instruction content ...................................................... 1-3Intended user.............................................................. 1-1

K

Keypad cleaning .................................................. PR47-1

L

LCD ............................................................................ 4-1Linear curve fit ..................................................... PR27-1Load limits ......................................................... 1-8, 1-10Location considerations.............................................. 3-2

Hazardous locations ............................................. 3-3Radio frequency interference (RFI) ...................... 3-4

M

Maintenance ............................................................. 13-1Keypad cleaning ............................................ PR47-1Schedule............................................................. 13-1

Page 185: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

WBPEEUI520003A1 Index - 3

Index (continued)

Sensor cleaning............................................. PR46-1Sensor inspection .......................................... PR43-1

Measure mode ........................................................... 5-1Boredom switch .................................................... 5-1Fault info ............................................................... 5-2Menu smart key .................................................... 5-2Primary display ..................................................... 5-1Secondary display ................................................ 5-1

Menu smart key.......................................................... 5-2Microprocessor PCB replacement....................... PR51-1Modes of operation..................................................... 4-3Modify configuration ..................................... 8-3, PR18-1Mounting

Hinge ............................................................... PR4-1Panel................................................................ PR2-1Pipe.................................................................. PR1-1Wall.................................................................. PR3-1

Mounting kits ............................................................ 1-10

N

Nomenclature ............................................................. 1-5Nonlinearity ................................................................ 1-7

O

Operating modes........................................................ 4-3Operator interface ...................................................... 4-1

LCD....................................................................... 4-1Smart keys............................................................ 4-2

Other equipment interface.......................................... 3-4Output calibration ......................................... 6-1, PR10-1Output maximum span ............................................... 1-8Output minimum span ................................................ 1-8Output range ......................................... PR16-1, PR32-1Output state................................................................ 8-6Output/hold................................................................. 7-1

Damping ................................................. 7-2, PR13-1Hold output .................................................... PR14-1Hold/release output............................................... 7-1Rerange................................................................ 7-1Spike..................................................................... 7-2

P

Panel mounting ..................................................... PR2-1Password

Changing ....................................................... PR37-3Entering ................................................................ 9-1Setting............................................................ PR37-1

Pipe mounting ....................................................... PR1-1Power supply effect .................................................... 1-8Power supply PCB installation............................. PR53-1

Power supply PCB removal .................................PR49-1Power wiring ..........................................................PR5-1Primary display ...........................................................5-1Problem codes ..........................................................12-1

Common..............................................................12-2Uncommon..........................................................12-5

Process display range.................................................1-7Programming mode change...................PR20-1, PR38-1

R

Radio frequency interference......................................3-4Rear cover replacement.......................................PR55-1Reboot .................................................................PR42-1Reference documents.................................................1-5Release hold output .............................................PR15-1Removing security ...............................................PR37-2Repair

Conductivity input PCB ....................PR50-1, PR52-1Front bezel .......................................PR48-1, PR54-1Microprocessor PCB ......................................PR51-1Power supply PCB ...........................PR49-1, PR53-1Rear cover......................................................PR55-1Sequence ............................................................14-1Shell ...............................................................PR56-1

Repeatability ...............................................................1-7Replacement parts....................................................15-1Rerange ........................................................7-1, PR16-1Reset all parameters...................................11-3, PR41-1Reset calibration ...........................................6-1, PR12-1Reset configuration .....................................11-3, PR39-1Reset security .............................................11-3, PR40-1Return materials procedures.....................................15-1RFI ..............................................................................3-4Rotation.......................................................................3-4

S

Safe mode.....................................................8-7, PR34-1Secondary display..............................................5-1, 10-1

Use......................................................................10-1Security .......................................................................9-1

Changing........................................................PR37-3Procedure.......................................................PR37-1Removing .......................................................PR37-2Reset.....................................................11-3, PR40-1Screen flow ...........................................................9-2Sequence ..............................................................9-1Setting ............................................................PR37-1

SensorCable length ..........................................................1-8Cleaning .........................................................PR46-1

Page 186: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

Index - 4 WBPEEUI520003A1

Index (continued)

Electronic test.................................................PR44-1Full scale measurement range ..............................1-7Group .............................................................PR22-1Inspection .......................................................PR43-1Types...................................................................1-10Wiring ...............................................................PR6-1

Setting password..................................................PR37-1Setting security.....................................................PR37-1Shell replacement ................................................PR56-1Signal wiring...........................................................PR5-1Size .............................................................................1-9Smart keys ..................................................................4-2Soft boot......................................................11-4, PR42-1Special handling..........................................................3-1Specifications..............................................................1-6

Accuracy................................................................1-7Ambient temperature effect ...................................1-8Damping................................................................1-8Display resolution ..................................................1-7Dynamic response.................................................1-8Environmental .......................................................1-8Input types.............................................................1-7Load limits ....................................................1-8, 1-10Nonlinearity ...........................................................1-7Output maximum span ..........................................1-8Output minimum span ...........................................1-8Power supply effect ...............................................1-8Process display range ...........................................1-7Repeatability..........................................................1-7Sensor cable length...............................................1-8Sensor full scale measurement range...................1-7Size .......................................................................1-9Stability..................................................................1-7Supply voltage.......................................................1-8Temperature compensation ..................................1-8Temperature display range ...................................1-7Turn on time ..........................................................1-8Type ......................................................................1-7Weight ...................................................................1-9

Spike .............................................................7-2, PR17-1Spike magnitude ..................................................PR35-1Spike output ................................................. 2-3, 4-4, 8-7Stability .......................................................................1-7Supply voltage ............................................................1-8Support services .......................................................15-1

Replacement parts ..............................................15-1Return materials procedures ...............................15-1

T

TemperatureCalibration................................................. 6-1, PR8-1Compensation.......................... 1-2, 1-8, 2-2, 8-5, A-1

Coefficient ................................................ PR30-1User-defined............................................. PR31-1

Display range........................................................ 1-7Effect..................................................................... 1-8Sensor .................................................................. 8-5

Type ......................................................... PR28-1Text prompts...............................................................C-1Text string............................................................ PR26-1Transmitter troubleshooting................................. PR45-1Troubleshooting

Electronic sensor test .................................... PR44-1Sequence............................................................ 12-8Transmitter (without sensor) .......................... PR45-1

Turn on time ............................................................... 1-8

U

Unpacking................................................................... 3-2User state ................................................................. 11-1

Advanced/basic programming ....................... PR38-1User-defined temperature compensation ............ PR31-1Utility mode............................................................... 11-1

Basic/advanced .................................................. 11-3Reset all parameters........................................... 11-3Reset configuration............................................. 11-3Reset security ..................................................... 11-3Soft boot ............................................................. 11-4User state ........................................................... 11-1

V

View/modify configuration........................................... 8-1Viewing configuration .......................................... PR18-1

Modifying while .............................................. PR19-1

W

Wall mounting........................................................ PR3-1Weight ........................................................................ 1-9Wiring

Considerations...................................................... 3-4Sensor ............................................................. PR6-1Signal and power ............................................. PR5-1

Page 187: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

Products and customer supportAutomation SystemsFor the following industries:— Chemical & Pharmaceutical— Food & Beverage— Manufacturing— Metals and Minerals— Oil, Gas & Petrochemical— Pulp and Paper

Drives and Motors— AC and DC Drives, AC and DC Machines, AC Motors to

1kV— Drive Systems— Force Measurement— Servo Drives

Controllers & Recorders— Single and Multi-loop Controllers— Circular Chart and Strip Chart Recorders— Paperless Recorders— Process Indicators

Flexible Automation— Industrial Robots and Robot Systems

Flow Measurement— Electromagnetic Flowmeters— Mass Flowmeters— Turbine Flowmeters— Wedge Flow Elements

Marine Systems & Turbochargers— Electrical Systems— Marine Equipment— Offshore Retrofit and Refurbishment

Process Analytics— Process Gas Analysis— Systems Integration

Transmitters— Pressure— Temperature— Level— Interface Modules

Valves, Actuators and Positioners— Control Valves— Actuators— Positioners

Water, Gas & Industrial Analytics Instrumentation— pH, Conductivity and Dissolved Oxygen Transmitters and

Sensors— Ammonia, Nitrate, Phosphate, Silica, Sodium, Chloride,

Fluoride, Dissolved Oxygen and Hydrazine Analyzers— Zirconia Oxygen Analyzers, Katharometers, Hydrogen

Purity and Purge-gas Monitors, Thermal Conductivity

Customer supportWe provide a comprehensive after sales service via a Worldwide Service Organization. Contact one of the following offices for details on your nearest Service and Repair Centre.

USAABB Inc.Tel: +1 800 HELP 365 (435 7365)Fax: +1 860 298 7669

UKABB LimitedTel: +44 (0)1453 826661Fax: +44 (0)1453 829671

ChinaABB Engineering (Shanghai) LimitedTel: +86 (0) 21 6105 6666Fax: +86 (0) 21 6105 6992

Client WarrantyPrior to installation, the equipment referred to in this manual must be stored in a clean, dry environment, in accordance with the Company's published specification.Periodic checks must be made on the equipment's condition. In the event of a failure under warranty, the following documentation must be provided as substantiation:— A listing evidencing process operation and alarm logs

at time of failure.— Copies of all storage, installation, operating and

maintenance records relating to the alleged faulty unit.

Page 188: OI Type TB82EC Advantage SeriesTM 2-wire, 4-electrode ... · 2-wire, 4-electrode conductivity transmitter. WBPEEUI520003A1 ... PR21 - ANALYZER TYPE SELECTION PR22 - SENSOR GROUP SELECTION

Contact us

OI/T

B82

EC

–EN

Rev

. B10

.201

2

ABB Inc.Process Automation843 N Jefferson StreetPO Box 831Lewisburg 24901-9509USATel: +1 304 647 4358Fax: +1 304 645 4236

ABB LimitedProcess AutomationOldends LaneStonehouseGloucestershire GL10 3TAUKTel: +44 1453 826 661Fax: +44 1453 829 671

ABB Engineering (Shanghai) Ltd.Process AutomationNo5, Lane 369, Chuangye Road201319, ShanghaiP.R. ChinaPhone: +86 (0) 21 6105 6666Fax: +86 (0) 21 6105 6992

www.abb.com

NoteWe reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.

We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts – is forbidden without prior written consent of ABB.

Copyright© 2012 ABBAll rights reserved

WBPEEUI520002A1

Sales

Service

Software


Recommended