OIML and ISO Perturbation testing for Type Approval
Written and Conducted by Martin Schlebach
Presented by Jacob Freeke
Daniel Measurement and Control
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 2
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
3/30/14
2
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 3
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 4
OIML-ISO Test Configurations
OIML 137-1&2: 2012
ISO 17089: 2010
3/30/14
3
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 5
Severe Disturbance Configurations
HMP Half Moon Plate
DBOOP LT Double Bend Out Of
Plane Left Turn
DBOOP RT Double Bend Out Of
Plane Right Turn
Flow
Flow
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 6
Severe Disturbance Configurations
OIML R137-1: 2006 !! Annex B !! B.2 Mild flow
disturbances –! DBOOP + expander
!! B.3 Severe flow disturbances –! DBOOP + HMP +
Expander –! Fewer tests
!!
OIML R137-1&2: 2012 !! Annex B !! B.2 Mild flow
disturbances –! DBOOP without expander
!! B.3 Severe flow disturbances –! DBOOP + HMP without
expander –! More tests
Became more stringent
3/30/14
4
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 7
Outline !! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 8
Piping Configurations
12” Base line testing
3/30/14
5
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 9
Piping Configurations
BDOOP RT
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 10
Piping Configurations
BDOOP LT
3/30/14
6
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 11
Piping Configurations
Single Elbow test
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 12
Piping Configurations
3/30/14
7
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 13
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 14
Meter and Run for testing
3/30/14
8
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 15
Meter Path Configuration 4-Path Chordal “BG” Design
•! Chords arranged in “X” formation
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 16
Testing done with T22 Transducers
3/30/14
9
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 17
Testing done with 3410 Electronics
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 18
Remote Witnessing
3/30/14
10
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 19
Test Grid 1 Flow Upstream Meter Test Velocities Fig Line Date Notes TCC
conditioner spool orientation Ft/sec # Size Completed Summary1 50D BASELINE YES 10D 0? 7 5,10,25,40,70,100,120 1 12” !"#"$%&$ '()*+,-,./01-2/'+344/&5% 67892 SINGLE BEND YES 10D 0? 4 5,10,40,70 2 12” !"6"$%&$ '()*+,-,./01-2/'+344/&5% 678:3 SINGLE BEND YES 10D 90? 4 5,10,40,70 2 12” !"6"$%&$ '()*+,-,./01-2/'+344/&5% 679%4 REDUCER YES 10D 0? 4 5,10,40,70 3 12” !"#"$%&$ '()*+,-,./01-2/'+344/&5% 67875 EXPANDER YES 10D 0? 4 5,10,40,70 4 12” !"#"$%&$ '()*+,-,./01-2/'+344/&5% 678!6 DBOOP RT YES 10D 0? 4 5,10,40,70 5 12” !"6"$%&$ '()*+,-,./01-2/'+344/&5% 679&
6A DBOOP LT YES 10D 0? 5 5,10,40,71 6 12” !"9"$%&$ '()*+,-,./01-2/'+344/&5% 6798
7 DBOOP RT YES 10D 90? 4 5,10,40,70 5 12” !"8"$%&$ '()*+,-,./01-2/'+344/&5% 679#7A DBOOP LT YES 10D 90? 4 5,10,40,70 5 12” &%":"$%&$ '()*+,-,./01-2/'+344/&5% 8%&%
11 DBOOP, HMP, RT YES 10D 0? 3 25,40,100 8 12” !"8"$%&$ '()*+,-,./01-2/'+344/&5% 679$
11A DBOOP, HMP, LT YES 10D 0? 3 25,40,100 9 12” !"9"$%&$ '()*+,-,./01-2/'+344/&5% 6799
11B DBOOP, HMP, LT YES 10D 90? 3 25,40,100 9 12” &%":"$%&$ '()*+,-,./01-2/'+344/&5% 8%&&
12 DBOOP, HMP, RT YES 5D 0? 3 25,40,100 8 12” !"8"$%&$ '()*+,-,./01-2/'+344/&5% 679613 DBOOP, HMP, LT YES 5D 0? 3 25,40,100 8 12” &%"&%"$%&$ '()*+,-,./;<= 8%&614 50D BASELINE YES 10D 0? 7 5,10,25,40,70,100,120 1 12” !"9"$%&$ '()*+,-,./01-2/'+344/&5% 679!
Test Disturbance # of Test points
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 20
Round 2 Test Plan
Test Disturbance Flow Upstream Meter # of Test points
Test Velocities Org Line Notes
conditioner spool orientation Ft/sec Size !!
1 50D Reproducibility and Pulse Test
YES 10D 0˚ 7 2.5,5,10,25,40,70,100 Both 12” "#$%!&'(')!*$%+,)$-./!(.!*.01$)'!2#%3'!4')3'!5/6+*$('6!75&8!9:;<;:=!5>7!<;?;@!
2A SINGLE BEND YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;:=!5>7!<;?;B!
3 REDUCER YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;:=!5>7!<;?;B!
4 EXPANDER YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;:=!5>7!<;?;B!
5B DBOOP RT YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;:=!5>7!<;?;B!
5C DBOOP, HMP, RT YES 10D 0˚ 3 25,40,100 OIML 12” 75&8!A;@!
6B DBOOP LT YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;@=!5>7!<;?;B!
6C DBOOP, HMP, LT YES 10D 0˚ 3 25,40,100 OIML 12” 75&8!A;@!
7 +3% STEP YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;:=!5>7!<;?;B!
8 -3% STEP YES 10D 0˚ 5 2.5,10,25,40,70,100 Both 12” 75&8!A;:=!5>7!<;?;B!
10 Pressure Tests 9 Bar YES 10D 0˚ 7 2.5,5,10,25,40,70, 100 OIML 8” C#).%..1=!75&8!9:;<;<!
10A Pressure Tests 60 Bar YES 10D 0˚ 7 2.5,5,10,25,40,70, 100 OIML 8” C#).%..1=!75&8!9:;<;<!
3/30/14
11
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 21
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 22
Baseline and Performance Tests
12” Base line testing
3/30/14
12
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 23
Baseline and Performance Tests
4” Base line testing
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 24
Baseline and Performance Tests
36” Base line testing
3/30/14
13
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 25
Baseline Results for all sizes
-2.0
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
0 20 40 60 80 100 120 140
Test 15 4inch WME Test 16 8inch WME Test 1 12inch WME Test 17 18inch WME Test 18 24inch WME Test 19 36inch WME
% of FS
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 26
Single Bend, Reducer and Expander Test
16 to 12 inch reducer testing
3/30/14
14
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 27
Single Bend, Reducer and Expander Test
10 to 12” expander testing
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 28
Single Bend, Reducer and Expander Test
Single Bend testing
3/30/14
15
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 29
Minor Perturbation Results
-2.00
-1.50
-1.00
-0.50
0.00
0.50
1.00
1.50
2.00
0 10 20 30 40 50 60 70 80
Single Bend 0 vs BL Single Bend 90 vs BL Reducer vs BL Expander vs BL -3% Step vs BL
% of FS
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 30
Double Bends Out Of Plane (DBOOP) and Half Moon Plate (HMP)
DBOOP & HMP testing
3/30/14
16
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 31
Double Bends Out Of Plane (DBOOP) and Half Moon Plate (HMP)
DBOOP & HMP testing head at 90 degree
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 32
Double Bends Out Of Plane (DBOOP) and Half Moon Plate (HMP)
DBOOP LT & HMP testing head at 0 degree
3/30/14
17
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 33
Severe Perturbation Results
-2.00
-1.50
-1.00
-0.50
0.00
0.50
1.00
1.50
2.00
0 10 20 30 40 50 60 70 80 90 100
DBOOP RT 90 vs BL DBOOP RT 0 vs BL DBOOP LT 90 vs BL
DBOOP LT 0 vs BL DBOOP HMP RT 0 vs BL DBOOP HMP LT 0 vs BL
% of FS
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 34
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
3/30/14
18
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 35
CFD vs actual results
R-137: DBOOP + HMP + 10D
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 36
Velocity Trajectories –10D Installation
3/30/14
19
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 37
Velocity Trajectories –10D Installation
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 38
Velocity Trajectories –10D Installation
3/30/14
20
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 39
FlowVelA (m/s) FlowVelB (m/s) FlowVelC (m/s) FlowVelD (m/s)
Per Path Velocities Test 11A Test 11B .00
.00
.00
.00
.00
.00
.00
-6.00 -4.00 -2.00 0.00 2.00 4.00 6.00
CPA50E 10&10 OIMLR137 25m 11500 Horizontal 15D Downstream CPA50E 10&10 OIMLR137 25m 11500 Vertical 15D Downstream
Theoretical and Measured Profiles
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 40
Theoretical and Measured Profiles
FlowVelA (m/s) FlowVelB (m/s) FlowVelC (m/s) FlowVelD (m/s)
3/30/14
21
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 41
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 42
Final Results
-2.00
-1.50
-1.00
-0.50
0.00
0.50
1.00
1.50
2.00
0 10 20 30 40 50 60 70 80 90 100
Transdcuer Swap vs BL Electronics and transducer swap vs BL Transducer Fails vs BL
Single Bend 0° vs BL Single Bend 90° vs BL Reducer vs BL
Expander vs BL DBOOP RT 90° vs BL DBOOP RT 0° vs BL
DBOOP HMP RT 0° vs BL DBOOP LT 90° vs BL DBOOP LT 0° vs BL
DBOOP HMP LT 0° vs BL -3% Step vs BL
% of FS
3/30/14
22
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 43
Initial versus Final Baseline (difference)
V [ft/s]
-2.00
-1.50
-1.00
-0.50
0.00
0.50
1.00
1.50
2.00
0 5 10 15 20 25 30 35 40
Initial vs Final BAseline
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 44
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Weighted Mean Error Different from BaseLine
Final Results 10 D Lmin
1/3rd MPE OIML / MID Class 2.0
1/3rd MPE OIML / MID Class 1.0
3/30/14
23
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 45
Baseline Shifts with Additional 10D TEST 1 BL 2 3 4 5 6 6A 7 7A 11 11A 11B
WME Lmin 0.105 0.146 0.197 0.109 0.144 0.277 0.203 0.173 0.219 0.226 0.139 0.083 BL Shift Lmin 0 0.041 0.092 0.004 0.039 0.172 0.098 0.068 0.114 0.121 0.034 -0.022 WME L+10D 0.341 0.429 0.42 0.387 0.398 0.564 0.386 0.492 0.437 0.413 0.326 0.437
10D shift 0.236 0.283 0.223 0.278 0.253 0.287 0.183 0.319 0.218 0.187 0.187 0.355 BL Shift +10 0 0.088 0.079 0.046 0.056 0.223 0.045 0.15 0.096 0.072 -0.015 0.096
10D BL shift 0.047 -0.013 0.042 0.017 0.05 -0.053 0.083 -0.018 -0.049 -0.049 0.118
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
1A
Sin
gle
Ben
d
Sin
gle
Ben
d 90
Red
ucer
Exp
ande
r
DB
OO
P R
T 90
DB
OO
P LT
90
DB
OO
P R
T 0
DB
OO
P LT
0
DB
OO
P H
MP
RT
DB
OO
P H
MP
LT
-3%
Ste
p
3% S
tep
9 B
L
Weighted Mean Error Difference with additional 10D
1/3rd MPE OIML / MID Class 2.0
1/3rd MPE OIML / MID Class 1.0
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 46
Severe Disturbance Configurations
OIML R137-1: 2006 !! Annex B !! B.2 Mild flow
disturbances –! DBOOP + expander
!! B.3 Severe flow disturbances –! DBOOP + HMP +
Expander –! Fewer test
OIML R137-1&2: 2012 !! Annex B !! B.2 Mild flow
disturbances –! DBOOP without expander
!! B.3 Severe flow disturbances –! DBOOP + HMP without
expander –! More tests
Became more stringent
Keep this in mind!
3/30/14
24
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 47
Meter and Run for testing
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 48
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
3/30/14
25
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 49
2 x 4 or 1 x 8 path British Gas Design !! Combines redundant 4
path British Gas meters, effectively eliminating sensitivity to swirl and other installation effects
!! The 3818’s interlocked redundant 4 path design was patented in 2010
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 50
Transducer Firing Synchronization Patent
!! Redundant electronics are not free running !! Firing of the 16 transducers is synchronized
to avoid signal collisions between the transducer paths
!! Having CPU boards connected also allows the combining of 4 path meter information into an 8 path meter output from either electronics
3/30/14
26
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 51
MeterLink for 1 x 8 or 2 x 4 path design •!Meters can be fully independent
and combined into an 8 path meter. •!Diagnostic information displays both
4 path and 8 path information
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 52
Outline
!! Standards !! Pipe configuration examples !! Used meter and equipment !! Tests & Results !! CFD versus actual results !! Final Results !! What’s next !! Conclusions
3/30/14
27
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 53
Summary and Conclusions
!! OIML and ISO Perturbation testing for Type Approval - although a costly and time consuming effort - brings a wealth of test results that helps understanding and reducing the effects of flow disturbances and hence reduce measurement uncertainty.
!! Results are well within the OIML and MID Class 1.0% when using 5D – FC – 10D – USM – 5D, most are within Class 0.5%
!! Further improvement can be expected with a more efficient FC is used and or longer Lmin.
!! Dual multipath meters (2 x 4, 1 x 8) further help to reduce flow disturbances effects, but need to be synchronized to avoid signal collisions between the transducer paths
!! OIML 137-1&2: 2012 is more stringent compared to OIML 137-1: 2006)
Thanks for your attention Questions?
3/30/14
28
[File Name or Event] Emerson Confidential 27-Jun-01, Slide 55
Martin’s Acknowledgements !! CPA for the donation of 36” and 24” Flow Conditioners !! Danny Sawchuk for all the CFD models and time spent
explaining them to me!! !! Wayne Haner and the crew from TransCanada Calibrations
for their diligent efforts in completing all the configurations and testing in a timely manner.
!! Klaus Zanker for analyzing the data and assuring me that I wasn’t the only one confused by the specs!!
!! David Brown and Kyle Kergen of Daniel for helping fill in whenever I’d leave during testing, usually on very short notice!!