www.greencrossms.com
hemoglobin A1c in human whole blood. GREENCARE A1c is a fully automated boronate a�nity assay for determination of
GREENCARE A1c provides the most convenient.
Cassette optical reading area
Hemoglobin A1c Test KitSpecification
One Step HbA1c measuring system
Fast & accurate result CV3%, in 4.5min.
RELIABLELCD touch screenMulti-lingual voice & video guideBuilt-in Thermal Printer
EASY-TO-USESingle test cartridge containing synthetic materials(Bead, Washing solution)
SIMPLE TEST STEPS
• Save freight cost• Reduce concern for stability due to temp
ROOM TEMPERATURE STABLE TEST KIT
Model RT-100 Storage Condition
Analyzer: 10~60 ℃, 15~75% R.H.Operating Type Full automatic (1-Step) Test kit: 2~30 ℃Measurement Time 4.5 min. Shelf life of Test kit 12 monthsSample Type Capillary/Venous blood Rated Voltage 12 V DC - 3.33 ASample Volume 3.5 ul Weight 1.6 kgs
Measurement RangeHbA1c 4.0~15.0 % (20~140 mmol/mol) Dimensions 205(W)x200(L)x150(H) mmeAG 68~385 mg/dL (mmol/L)
Additional Function
Full color touch screen
Operating ConditionTemperature: 17~32 ℃ Internal printerHumidity: 15~75% R.H. USB port, EthernetAir pressure: 860~1,060 hpa External barcode reader available (optional)
One Step HbA1c measuring system
One Step HbA1c measuring system
One Step HbA1c measuring system
Built-in Thermal Printer
Communication Interface
SIMPLE TEST STEPS
GREENCARE A1c Hemoglobin A1c Test Kit
Manufactured by GREEN CROSS MEDIS CORP.Distributed by GREEN CROSS MEDICAL SCIENCE CORP.H.Q. : 107, Ihyeon-ro 30beon-gil, Giheung-gu, Yongin-si, Gyeonggi-do,446-770, Korea
Tel. :+82-31-260-0965,0969, 9663Fax. : +82-31-260-9416E-mail : [email protected]
http : //www.greencrossms.com R10IF0C00 (09/17)
No pipetting required
The GREENCARE A1c System was evaluated with 40 different samples provided by the NGSP.The results show high accuracy and excellent correlation.
Easy blood sampling No sample preparation
Reduced test errors by operator
Accuracy
Within-Run Precision (IFCC sample)Each of the samples was measured 20 times per each for precision using 2 meters across 3 lots.
IFCC calibrator Normal AbnormalMean HbA1c% 5.5 8.9
CV (%) 2.4 1.9
Difference, % GREENCARE A1c Systemwithin±3%within±6%within±10%
50%(20/40)92.5%(37/40)100%(40/40)
Day-to-day Precision (IFCC sample)Each of the samples was measured 1 time a day for 20 days using 2 meters across 3 lots.
IFCC calibrator Normal AbnormalMean HbA1c% 5.4 9.0
CV (%) 2.6 1.9
y = 0.9769x + 0.1592R² = 0.978
0.0
2.0
4.0
6.0
8.0
10.0
12.0
0.0 2.0 4.0 6.0 8.0 10.0 12.0
GREE
NCAR
E A1c
(A1c
%)
Regression Analysis
-12.0
-7.0
-2.0
3.0
8.0
0.0 2.0 4.0 6.0 8.0 10.0 12.0
Di�e
renc
e (%
)
Tosoh G7 (A1c%) Tosoh G7 (A1c%)
Accuracy analysis