    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentation

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentation

    OPC and CBR break systems

    SCHOTT AG

    January 2009/OL

  • 7/23/2019 OPC_CBR Break Systems

    2/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentation

    OPC/CBR Ampoules presentation

    Break system description

    The One Point Cut principle (OPC)

    After the forming process and before the annealing oven,

    a small cut is applied with a carbide disc to theconstriction of the ampoule.

    The depth of the cut must be constant and

    dot

    Chart 2

    contro e very care u y to prevent t e ampou esfrom leaking and to ensure a regular break force.In the cut there are micro cracks which weakens theglass at this position

    As the cut is very small and is not visible withoutmagnification, there is a small colored dot abovethe cut on the bulb of the ampoule which marks theposition of the cut.

    cut

  • 7/23/2019 OPC_CBR Break Systems

    3/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentation

    OPC/CBR Ampoules presentation

    Break system description

    Color ring

    The Color Break Ring principle (CBR)

    The CBR ampoules have a ceramic color ring on the

    constriction. This ring made of ceramic enamel and organicoil is applied after forming and before annealing theampoules. The ceramic enamel has a higher mean linearcoefficient of expansion than the glass.

    Chart 3

    In the annealing oven the organic oil decomposes and theenamel is melted onto the glass surface.After cooling down, due to the higher mean linear coefficientof expansion the enamel contracts more than the glass and

    as it is fixed onto the glass surface, it produces micro crackson the surface which has contact with the glass and whichweakens the glass at this position.

  • 7/23/2019 OPC_CBR Break Systems

    4/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    The 2 systems use the same break force test set-up

    used for both systems

    Chart 4

  • 7/23/2019 OPC_CBR Break Systems

    5/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    OPC - ISO Breaking force

    Nennvolumen

    Nominal

    Volume

    ml

    Auflageabstand

    Length

    Bruchkraft

    Breaking Force

    l

    mm

    l1

    mm

    l2

    mm

    Fmin

    N

    Fmax

    N

    1 - 3

    36 18 18

    25 65

    5 30 70

    10 - 30 60 22 38 30 80

    Chart 5

  • 7/23/2019 OPC_CBR Break Systems

    6/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    CBR - ISO Breaking force

    Nennvolumen

    Nominal

    Volume

    ml

    Auflageabstand

    Length

    Bruchkraft

    Breaking Force

    l

    mm

    l1

    mm

    l2

    mm

    Fmin

    N

    Fmax

    N

    1 - 5 36 18 18 30 80

    10

    60 22 38

    30 90

    20 - 30 30 100

    Chart 6

  • 7/23/2019 OPC_CBR Break Systems

    7/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    Influence on the break force

    The break force of Color Break Ampoules is -in addition to other

    parameters dependent on the viscosity, the thickness of the enamelapplied and on the temperature profile of the annealing oven.As these three parameters are extremely difficult to control in technical

    roduction there is alwa s a bi er s read of the mean value of the

    Origins

    Chart 7

    break force of Color Break Ampoules than of OPC Ampoules.

    As the influencing parameters for the break force of OPC Ampoules,which are the cut depth and the cut position into the constriction, are

    easier to control, the standard deviation of the break force of OPCAmpoules is lower than that of Color Break Ampoules. The productionof OPC ampoules requires the use of sophisticated equipment.

  • 7/23/2019 OPC_CBR Break Systems

    8/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    OPC values: Reduced distribution

    within the ISO range (25 to 65 N)

    Influence on the break force

    Chart 8

    Color Break values: Inflexible withinthe whole ISO range (30 to 80 N)

  • 7/23/2019 OPC_CBR Break Systems

    9/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    Influence on particles

    Hereafter are the limits defined for the test Particles at break on 100 ampoulesThe ampoules are broken, the particles are collected in a recipient and thenweighedThis test is only performed in case of doubt after the usual hand breaking test or

    in case of bad aspect of the ringThese values are not contractual

    Chart 9

  • 7/23/2019 OPC_CBR Break Systems

    10/11

    PHARMACEUTICAL PACKAGING OPC/CBR Ampoules presentationOPC/CBR Ampoules presentation

    Lower mean value

    Reduced standard deviation

    OPC system properties support your move from

    Color break to OPC providing:

    Product safetyfor the end user

    Chart 10

    Better hand break quality

    No enamel particle when opening

    Less glass particles when opening

    Product quality