+ All Categories
Home > Documents > Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Date post: 23-Feb-2016
Category:
Upload: merton
View: 22 times
Download: 0 times
Share this document with a friend
Description:
Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia. Scope of Opening Presentation. Technology basics – review Limitations of LD Technology in general Limitations of LISST instruments What can go wrong in data/operations - PowerPoint PPT Presentation
Popular Tags:
28
OPENING PRESENTATION LISST-TECHNOLOGY AND PERFORMANCE – CHALLENGES AND SOLUTIONS -TEAM SEQUOIA
Transcript
Page 1: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

OPENING PRESENTATIONLISST-TECHNOLOGY AND PERFORMANCE – CHALLENGES AND SOLUTIONS

-TEAM SEQUOIA

Page 2: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Scope of Opening Presentation

Technology basics – review Limitations of LD Technology in general Limitations of LISST instruments What can go wrong in data/operations What can go wrong in interpretation Analysis of your experiences –

Tim Straub; Mark Landers

Page 3: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

TECHNOLOGY BASICS

Page 4: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

LD Technology Basics

All LD technology provides PSD and Concentration of an ensemble of particles ‘A view in a flash’

LISST and all other LD instruments are NOT particle counters, and NOT samplers.

Page 5: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

ISO 13320:2009

ISO-13320 was written only for PSD Summing PSD gives concentration over the size range

Page 6: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

How Laser Diffraction Works

0 10 20 300

0.2

0.4

0.6

0.8

1

1.2

Size

Input PSD

0 10 20 30 400

1

2

3

4

5

6

7x 10-3

Ring no.

Rings Output

0 10 20 300

0.2

0.4

0.6

0.8

1

1.2

Size

Inverted PSD

0 10 20 300

0.2

0.4

0.6

0.8

1

1.2

Size

Input PSD

0 10 20 30 400

1

2

3

4

5

6

7x 10-3

Ring no.

Rings Output

0 10 20 300

0.2

0.4

0.6

0.8

1

1.2

Size

Inverted PSD

0 10 20 300

0.2

0.4

0.6

0.8

1

1.2

Size

Inverted PSD

0 10 20 300

0.2

0.4

0.6

0.8

1

1.2

Size

Input PSD

0 10 20 30 400

0.002

0.004

0.006

0.008

0.01

0.012

0.014

Ring no.0 10 20 30

0

0.05

0.1

0.15

Size

Input PSD

0 10 20 30 400

0.5

1

1.5

2

2.5x 10-3

Ring no.

Rings Output

0 10 20 300

0.05

0.1

0.15

Size

Inverted PSD

Particles of ANY size produce light on ALL rings Light on a particular ring does NOT represent concentration of any

particular size particle. Common misunderstanding

Page 7: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Randomly Shaped Particles Scattering from randomly shaped

grains differ slightly from same-size spheres; this changes the size scale

All pre-Sequoia work produced PSD of equivalent spheres Only Sequoia offers inversion to give

PSD of ‘sieved sizes’. Using the randomly shaped

matrices enables comparison with existing, historical SIEVE data sets

Randomly shaped

Agrawal et al. 2008 (JGR)

Page 8: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

LIMITATIONS OF LD TECHNOLOGY IN GENERAL

- Upper and lower size limits- Cannot discriminate types of

particles

Page 9: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Upper and Lower Size Limits

Upper size limit derives from smallest ring Lower size limit derives from largest ring

0 5 10 15 20 25 30 350

0.002

0.004

0.006

0.008

0.01

0.012

0.014

0.016

Ring Detector #

Scattering signatures

Smallest ParticleLargest Particle

Page 10: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

How LD sees particles outsidethis range

Particles Leak into the nearest size bins This often produces rising tails in PSD

More so for particles smaller than the smallest size bin of the LISST

Page 11: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Almost everything scatters

Particles, bubbles, algae, etc. but also density fluctuations caused by temperature variance

Density-related scattering (‘scintillation’) invents large particles. Scintillation is avoided by letting water and LISST reach identical

temperature. This is common to all LD systems.

Sediment flocs in the Po River, Italy

Schlieren / Scintillation

Bubbles Phyto- and zoo-plankton

Page 12: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Converting volume to mass All LD systems measure volume distribution Converting to mass requires the user to ASSUME an

effective mass density and then apply that Easier in the lab, where all particles can be dispersed

(ISO 13320:2009). Not easy with field data where aggregates, bubbles, low-

density biological particles can be present

Page 13: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

LIMITATIONS OF LISST INSTRUMENTS

Concentration limits Limits exists in the field Optimum accuracy at optical transmission between 0.3 and 0.98 For –SL: upper limit (mg/l) = 300*d[µm] For –SS: 200*d[µm] Lower limit limited by signal-to-noise ratio

Size range may not cover all particles in suspension Clogging of intakes for pumped instruments

LISST-StreamSide LISST-SL

Velocity limits for isokinetic sampling and general operation (-SL only)

Page 14: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

WHAT CAN GO WRONG WHEN COLLECTING

DATA

Page 15: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Misalignment of optics Low zscat laser power, increased zscat scattering on inner rings

constant in time Dirty windows

Increased zscat scattering on middle rings Poor temperature equilibrium - scintillation

Increased scattering on inner rings, decreasing in time Clogging of intake by debris Clogging of pitot tube by debris (-SL only)

Page 16: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

LISST-SL pump not locked in, still adjusting during sampling Lock-in ~ 2 minutes after immersion when powered After lock-in isokinetic control is better than 10% as set at factory Minimum river velocity for isokinetic operation is 0.5 m/s [1.65 ft/s]

Purging of bleed port not adequate Erratic velocity readings on TCB

Page 17: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Is swimming a problem? Eddies change instantaneous flow

direction The component of drag normal to

river flow produces side-ways drift of -SL

Because USGS samplers weight >> drag, they are pushed less in sideways direction

A look at video of –SL always shows wake of ‘sail’ aligned with flow.

Even so, for operation in higher velocity rivers, extra weight will be needed. Sequoia to design.

Page 18: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

WHAT CAN GO WRONG WHEN INTERPRETING DATA

Page 19: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Short LISST-SL time series is an issue Makes interpretation and troubleshooting more difficult Measurement duration should be >> time-scale of local eddies

Presence of loosely aggregated particles Density << 2.65 Affects conversion from volume to mass!

Particles outside the size range Will influence computation of size parameters (mean, median etc.) Will overestimate total concentration

Page 20: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

A common source of error - Mixing

Concentration gradients exist everywhere, in the field and lab [Rouse (1937) showed C ~ z –ku*/wf ]

To avoid, turbulent velocity fluctuations >> settling velocity

Location of intake pump in a gauging station will affect measured PSD [see Cowlitz/Puyallup river data]

Even in the lab, concentration gradients make calibrations and comparisons very difficult.

Page 21: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Incomplete mixing in a river

Vertical variation of size distribution seen in Cowlitz (also Puyallup) river

Page 22: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

YOUR EXPERIENCES

Page 23: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Comparing size distributions

Size distributions from different methods CANNOT be compared.

The pipette method will always produce smaller diameters than LD for the same sample: Konert & Vandenberghe 1997

Same applies to Sedigraph and Sieving LD PSD should be compared to LD PSD only Discrepancies WILL occur (except for spheres)

Use literature to evaluate if observed differences > expected Sample analysis in the lab changes the size distribution

compared to the in-situ data (settling, sonification, in-situ aggregation etc.)

Page 24: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Converting to mass

Mass density of in-situ particles is not the same as the density of the same particles in the lab when disaggregated

Calculating an effective in-situ density using water samples is an effective method for converting in-situ volume to mass

Page 25: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Effective Density: Definition

Effective Density = It is NOT the density of the individual

particles It changes in time and space

depending upon conditions

Page 26: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Effective Density: Example

Sediment Grains Flocculated particles

Bubbles

Particle Size (µm) 30 30 30Volume Conc (µl/l) 100 100 100

Filtered and Weighed Mass Conc (mg/l) 260 40 0

Effective Density 2.6 0.4 0

Page 27: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Flocculation effect on size distribution

Arctic Rivers (Droppo et al., 1998)

No single grains > ~10 µm

All in situ particles > 10 µm were flocs

Page 28: Opening Presentation LISST-Technology and performance – challenges and solutions -Team Sequoia

Conclusions

LD has strengths and limitations.

Data suggest the possibility of flocs in-situ.

Remains to examine data to be presented.


Recommended