A simple colourimetric method for determining seawater alkalinity using bromophenol
blue.
Vikashni Nand1 and Michael J Ellwood1
1 Research School of Earth Sciences, Australian National University, Canberra, ACT 2601,
Australia.
Running Head:
Seawater alkalinity using bromophenol blue
Key words:
Seawater alkalinity, seawater pH, dissolved inorganic carbon, bromophenol blue.
Corresponding author:
Vikashni Nand
Research School of Earth Sciences
The Australian National University
Building 142 Mills Road Acton
Canberra, ACT 2601 Australia
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Abstract
The development of small portable USB-spectrophotometer systems makes monitoring
alkalinity and pH possible in the field and remote locations. Here we present a method
utilising purified Bromophenol Blue (BPB) as an end-point indicator for making simple one-
point alkalinity measurements with spectrophotometric detection. The approach utilises
purified BPB dye whose absorbance characteristics have been determined over a range of
temperatures and salinities. The end-point pH for titrated samples was determined using the
BPB absorbance ratio (R(t) = 25 A590/A436) for the acid and base forms via the following
equation: pH=pK a+ log [ (R [25]−e1)(e2−R [25] . e3) ] where e1, = 0.00533, e2 = 2.232, e3 = 0.0319. A pKa
of 3.513 was determined for the dissociation of the second proton from the BPB dye. The
temperature (t) dependence of R can be expressed using the following relationship:
R(25)=R(t ) [1+(0.006774 ± 0.000009)(25−t)]. The dependence of the pKa on salinity (S) was
weak and can be expressed as p Ka (S )=p Ka (35)+[(0.00174 ± 0.00008) (35−S ) ]. Application of
the method for determining the alkalinity of in-house and certified standards typically
produced an uncertainty of ± 1.5 μmol kg−1 for purified BPB dye. When the impure BPB dye
was used as an end-point indicator the uncertainty for alkalinity measured was slightly higher
at approximately ± 3-4 μmol kg−1. Hence, if high precision alkalinity measurements are not
required (≥ 4 µmol kg-1) then utilisation of the unpurified BPB maybe suitable. We also
compared the use of BPB to two other dyes: bromocresol purple (BCP) and bromocresol
green (BCG). The utilisation of all three dyes for end-point determination produced
comparable results with an overall precision of ± 4 µmol kg-1. The one-point titration method
using BPB was utilised at a remote field location, One Tree Island, Australia, and was found
to be suitable for producing accurate and precise alkalinity data in a timely manner;
2
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
approximately 10-15 samples can be determined per hour. When combined with seawater pH
measurements, the one-point titration method allows the full marine carbonate system to be
fully constrained without the need for high-tech spectrophotometric equipment and
comprehensive laboratory facilities.
3
45
46
47
48
49
Introduction
Comprehensive studies of global oceanic inorganic carbon cycling are essential to understand
the oceanic uptake, transport, and storage of anthropogenic carbon dioxide. High-precision
carbonate system measurements are required to monitor the changes in seawater carbonate
chemistry. The characterization of the seawater carbonate system requires the measurement
of at least two key system parameters, e.g. total dissolved inorganic carbon (DIC), alkalinity,
carbon dioxide fugacity, and pH (Millero et al., 1993a){Millero, 1993 #154@@author-
year;Yao, 2007 #228}. Alkalinity and DIC are the preferred parameters for field
measurements and form the basis of analytical assessments of oceanic CO2 cycling and ocean
acidification. Although DIC and alkalinity are preferred parameters for constraining the
marine carbonate system, their measurement requires high-tech equipment, e.g. expensive
spectrophotometers with temperature control, to obtain high-precision data. Typically,
alkalinity measurements require sophisticated titration systems and high-quality pH
electrodes, which can be finicky to use, i.e. requiring an understanding of the Nernstian
behaviour of the electrode (Millero et al., 1993b), and with sometimes short-life spans.
More recently seawater pH has become a common parameter to measure and to help
constrain the uptake of anthropogenic atmospheric CO2 by the ocean (Clayton and Byrne,
1993; Martz et al., 2010; Rérolle et al., 2012). High precision pH can be measured using UV-
visible spectrophotometer and indicator dyes (Bellerby et al., 2002, Clayton and Byrne, 1993,
Tapp et al., 2000). Spectrophometric indicator-based methods, in theory, are simple, fast, and
precise. Spectrophotometric pH typically has a precision of 0.004−0.001 pH units
(DeGrandpre et al., 2014, Easley and Byrne, 2012); although the accuracy of
spectrophotometric pH measurements can be adversely affected by impurities in the indicator
dyes (Lai et al., 2016, Yao et al., 2007).
4
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
Recent work has demonstrated that indicator purification is required to eliminate systematic
pH measurement errors (Liu et al., 2011, Yao et al., 2007); errors as large as 0.018 units have
been measured for impurified batches of meta-cresol purple (mCP) (Liu et al., 2011, Yao et
al., 2007). Meta-cresol purple is commonly used for measurements of oceanic seawater pH
over a range of 7.2−8.2 (Easley and Byrne, 2012, Patsavas et al., 2013). The work by Yao et
al., (2007) describes the effect of indicator impurities on spectrophotometric pH
measurements, and Liu et al., (2011) demonstrates that purified mCP could be used to
determine seawater pH on the total hydrogen ion concentration scale (pHT) accurately and
precisely over a wide range of temperatures (278.15 ≤ T ≤ 308.15) and salinities (20 ≤ S
≤40).
Work by Breland II and Byrne (1993) shows that the pH-sensitive dye bromocresol green
(BCG) is ideal for determining the pH of solutions in the range between 3.4 and 4.6. Thus the
dye is a useful indicator for determining the pH of seawater samples following the addition of
acid for alkalinity determination (Breland II and Byrne, 1993, Martz et al., 2006). Using
BCG as a dye, Breland II and Byrne (1993) were able to reproducibly determine seawater
alkalinity to a precision of ± 1.5 mol kg-1. More recently, Liu et al. (2015), developed an
automated system to measure alkalinity using the dye bromocresol purple (BCP). This
method obtains a precision ±1 mol kg-1 utilising volumetric addition of acid to the sample.
Like BCG and BCP, Bromophenol blue (BPB) can be used for spectrophotometric pH
measurements for the determination of total alkalinity (Haraldsson et al., 1997, Okamura et
al., 2010). Bromophenol blue is a diprotic sulfonephthalein indicator dye with a pKa value
~3.7 (King and Kester, 1989, Okamura et al., 2010), thus making it suitable for monitoring
pH for one-point alkalinity titrations. Here we build on the earlier work of Okamura et al.
5
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
(2010) to investigate the impact impurities have on the use BPB as an end-point indicator for
alkalinity measurements and determining extinction coefficients and its temperature and
salinity dependence.
Theory
In seawater, alkalinity is defined as the number of moles of hydrogen ion required to convert
bicarbonate to carbonic acid (Dickson and Goyet, 1994, Dickson et al., 2007) and can be
expressed as follows:
HC O3−¿+H +¿→ H2 CO 3¿ ¿ (1)
Seawater also contains additional acid-base compounds thus total alkalinity (TA) is expressed
as:
TA=¿
(2)
where [H+]F represents the hydrogen ion concentration
The addition of H+ changes alkalinity by one equivalent unit (i.e. [HCO3-]), while two protons
are required to titrate CO32- to H2CO3. In equation (2), the addition of [H+] and [HSO4
−] will
decrease alkalinity as they are the sources of protons.
If equation (2) is inverted, the total analytical concentration of hydrogen ions can be obtained
relative to the reference proton condition for pure H2CO3. Following this conversion, the
total hydrogen ion concentration (CH) can be defined as:
6
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
CH=¿ (3)
The addition of a mass (m) of hydrogen ions to the solution reduces TA, thus after the
addition of a mass (m) of acid to an initial mass of seawater(m¿¿0)¿, the following equation
can be expressed:
CH=mC−m0 TA
m0+m (4)
where C in the concentration of the acid added to the solution.
If equation (3) is substituted into equation (4) the following equation can be written:
mC−m0 TAm0+m
=¿
(5)
At the end of most open cell titrations, the pH is usually between 3.0 – 4.5 depending on the
initial alkalinity of the sample and the mass of acid added to the sample. At pH values below
~ 4.5, virtually all the carbonate species are converted to H2CO3 and CO2 and lost to the
atmosphere following purging. Thus for CH evaluation, the majority of the species can be
neglected apart from total sulfate and fluoride and equation (5) can be reduced to:
mC−m0 TA
m0+m=¿ (6)
where ¿ is the free hydrogen ion concentration.
7
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
The TA of the sample can be determined by measuring [H+]T, if m0, mC and m are all known.
TA+¿) +¿) + m0+m+mI
m0 ¿) C =0 (7)
The ST and FT are the total sulfate and total fluoride and the Ks and Kf are the dissociation
constant of [HSO4-] and [HF]. Thus are dependent on the salinity and temperature (Dickson et
al. 2003).
The measurement of pH (expressed on the total proton scale) can either be done with a well-
calibrated pH electrode or with pH-sensitive indicator dyes. Two such indicator dyes useful
for determining pH in the range between 3 and 4.5 are BCG and BPB (Breland II and Byrne,
1993, Okamura et al., 2010). The relative colour of the dye is dependent on the proportion of
the dye acid and dye base present in solution (Figure 1), which can be expressed as:
H 2 I ⇋H+¿+H I−¿ ¿¿ (8)
H I−¿⇋H+¿+ I2−¿¿¿ ¿ (9)
Thus, the dissociation constant (Ka) of indicator species can be expressed as:
K a=¿¿ (10)
8
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
The pH of the solution containing the indicator dye can be determined by measuring the
relative amounts of the acid and base present in solution using the following equation:
pH=pK a+ log [ (R−e1)(e2−R .e3) ] (11)
where R is the absorbance ratio of acid (HI-) and base (I2-) forms of the indicator (R=λ2 Aλ1 A ),
at wavelengths λ1 (436 nm) and λ2 (590 nm), respectively for BPB. The symbols e1, e2 and e3
are the molar absorbance ratios for each indicator species (Liu et al., 2011, Yao and Byme,
2001, Yao et al., 2007):
e1=λ2∈HI
λ1∈HI
(12)
e2=λ2∈I
λ1∈HI
(13)
e3=λ 1∈I
λ 1∈HI
(14)
where, λ1∈I and λ2∈I denote the molar absorbance of I2−, and λ1∈HI and λ2∈HI refers to the
molar absorbance of HI− at wavelengths λ1 and λ2 respectively.
In order to determine pH, the addition of the dye to the sample can perturb the pH of the
system, accounted for in the calculation of alkalinity. Thus equation (6) can be extended to
include the HI-:
mC−m0 TAm0+m+mI
=¿ (15)
9
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
where mI is the mass of the indicator added to the sample.
Equation (15) shows that once all the relevant bases present in solution have been neutralised,
the number of moles of the protons in solution will be proportional to the concentration of
acid added to the solution. This expression is akin to a “Gran Function” whereby the end
point of the titration can be obtained from the number of moles of the protons in solution
versus the volume of acid added to the solution (Gran, 1952).
Methods
Dye purification and calibration
The solvents used in the purification of the indicator dyes BPB, BCG and BCP consisted of
water, acetonitrile (MeCN) and trifluoroacetic acid (TFA). The initial separations involved
using thin-layered chromatography (TLC) silica gel 60 F254 (Lot: HX248024) 500 glass
10
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
plates (2.5 x 7.5 cm) plates. Using TLC, the mobile phase composition and concentration
were determined by trial and error, starting at 30% MeCN with 0.05% TFA and 0.1 % TFA
and then increasing the MeCN by 10% increments until the organic phase concentration was
80% MeCN. The optimum solvent composition dye-impurity separation consisted of 80%
MeCN plus 20% MQ-water (H2O) and 0.1% TFA. Once the composition of mobile phase
was optimized, preparative column chromatography using the silica gel 60 (0.040-0.063 mm)
was prepared for purification of the three dyes.
Stock solutions of unrefined BPB, BCG and BCP were prepared by dissolving the sodium
salt (Acros Organics, Lot: A0276342) in 80% MeCN, 20% MQ-water, and with 0.1% TFA.
To separate impurities, a 10 mmol L-1 solution of BPB in the mobile phase solution was
prepared and used for chromatographic separation. Pure BPB was collected at its
characteristic band once impurities had passed through the column. The excess solvent was
removed by rotary evaporation at 35oC under partial evacuation. Portions of the purified BPB
stock was re-dissolved in mobile phase for further High–Performance Liquid
Chromatography (HPLC) characterization. This procedure was repeated for BCG and BCP.
The purity of the BPB, BCG and BCP stock solution was confirmed by HPLC (Thermo
Finnigan Surveyor with a PDA detector) using an anion exchange column (PhenoSphere 5µ
SAX 80A, Column 150 x 4.6 mm) under isocratic conditions using a flow rate of 1.0 mL min-
1 and a 20 µL injection volume.
Characterisation of bromophenol blue extension coefficients
Standard solutions of BPB at pH values of 0.5 and pH 7.5 were prepared for characterisation
with an ionic strength of 0.7 mol kg-1 using sodium chloride (NaCl). pH adjustments were
11
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
made by addition of either sodium hydroxide (NaOH) or hydrochloric acid (HCl). The pH of
each buffer solution was checked using Ross combination electrode (ThermoFisher). Dye
absorbance measurements were made using a UV-Visible spectrophotometer (Varian Cary
300). All BPB characterisation measurements were performed within a water jacketed 10 cm
path-length cell at 25 ± 0.1°C. Measurements were made at wavelengths 436 nm and 590 nm
and corrected at 690 nm. The absorbance measurements were corrected to produce values for
e1, e2 and e3.
Determination seawater alkalinity
The initial HCl solution was prepared to a concentration of ≈0.05 mol kg-1 with the BPB
indicator dye (and using BCG and BCP for comparison experiments) present at a
concentration of approximately 2 mmol L-1. The exact concentration of the acid was
determined by titrating against a standard sodium carbonate solution prepared in 0.7 mol kg-1
NaCl and against a certified reference material (CRM) (Batch 140; alkalinity = 2232 ± 0.8
µmol kg-1). The equivalence point for the titration was determined using the Gran function
with BPB as the pH indicator (Breland II and Byrne, 1993, Gran, 1988).
For laboratory-based alkalinity determination, 100 g (± 0.0002 g) of an in-house seawater
standard (ANU Std) was accurately weighed out into a wide-neck volumetric flask.
Typically, 5 g (± 0.0002 g) of HCl (0.05 mol L-1) containing BPB was added, and the sample
was accurately re-weighed. The amount of acid added is such that the final pH of the solution
was ~3.5. The sample was capped and well mixed before being purged with air for 5 minutes
to drive-off evolved CO2. The absorbance spectrum of each sample was collected at 25 ± 0.2
°C. Similar conditions were used for the BCG and BCP dye experiments except that the final
alkalinity pH was calculated using the pKa for the two respective dyes.
12
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
For field-based alkalinity determination, 50 g (± 0.0002 g) of seawater was accurately
weighed out into a wide-neck volumetric flask. Typically, 2 g (± 0.0002 g) of HCl (0.05 mol
L-1) containing BPB was added to the sample so that final pH was around ~3.5 and was also
purged with air for 5 minutes. A USB UV-VIS spectrophotometer (USB4000, OceanOptics,
USA) coupled to a multi-LED light source (Ocean Optics BluLoop, Ocean Optics, USA) was
used for collecting absorbance spectra. The absorbance measurements were collected at 436
nm, 590 nm and 690 nm. The spectrophotometer, the sample cell holder and light source
were coupled via fibre optic cables. The temperature for each field sample was determined at
the time of measurement using a thermocouple (FEP coated thermocouple, TC Direct,
Australia coupled to a NI USB-9211, National Instruments, Australia) and used for
temperature correction of measured R values. The spectrophotometer and termeprature
sensor were controlled using custom software written in LabVIEW (National Instruments).
13
264
265
266
267
268
269
270
271
272
273
274
275
276
Results and Discussion
Characterisation of bromophenol dye
The purification and characterisation experiments reveal that the off-the-shelf batches of the
BPB, BCG and BCP indicator dyes all contain impurities (Figure 2). We also tested various
brands of off-the-shelve-batches for BPB. All batches contained impurities to varying
degrees. Purification of these off-the-shelve-batches dyes removed >95% of the impurities
based on HPLC analysis (Figure 2). Work by Liu et al. (2011) showed that for the natural
seawater pH indicator dye mCP, impurities could lead to pH errors up to 0.018 pH units for
seawater pH values of ~8.1. We tested whether impurities can also influence the use of BPB
as an end-point pH indicator dye. By using impure BPB across multiple alkalinity
determinations using the one-point titration method, an average alkalinity of 2134 ± 2.7 µmol
kg-1 was obtained for an in-house seawater standard (Table 2). In contrast, using the same in-
house seawater standard and approach, but with the purified dye we obtained a value of 2138
± 1.2 µmol kg-1 (n = 10) (Table 2, Figure 3A). Consequently, there appears to be a small
offset in the “true” value obtained using the impure dye and a decrease in the precision of the
method. For titration of the alkalinity CRM (2232 ± 0.8 µmol kg-1), we obtained a value of
2232 ± 1.3 µmol kg-1 (n = 5) for the purified dye and 2233 ± 3.9 µmol kg-1 for the unpurified
dye (n = 5) (Figure 3B). The standard deviation for the unpurifed BPB dye is larger that for
both the in-house standard and the CRM standard thus it appears that dye impurities can
influence the absorbance spectrum resulting in subtle differences in measured R values,
which will translate into subtle pH variations and hence alkalinity. Depending on the nature
of the impurity, it may disproportionately influence relative size of the absorbance peaks for
acid and based forms of the dye thereby leading to an offset in the calculated solution pH.
This has been observed for impure mCP when used to determine natural seawater pH (Liu et
al., 2011). Our results would suggest that purification of the BPB indicator dye is required
14
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
before it is used for high precision (<3 µmol kg-1) alkalinity determination. However, for
alkalinity measurements that do not require such high precision (>4 µmol kg-1) then the
impure BPB dye is suitable.
Initial investigations aimed at characterising BPB molar absorption coefficients were carried
out by Okamura et al. (2010) at 25°C and a salinity of 35 using unpurified BPB. The molar
absorption coefficients obtained for this dye were from absorbance spectra measured at pH
0.5 and 7.5. The molar absorption coefficients obtained for the purified BPB (this study,
Table 1) are slightly lower than values obtained by Okamura et al. (2010). These differences
for molar absorption coefficients may be because Okamura et al. (2010) did not purify their
dye.
An in-house seawater standard was titrated across a range of acid additions and the alkalinity
at each point computed and compared to its known alkalinity. Note that the alkalinity of the
in-house seawater standard should be constant and should not correlate to the mass of acid
added as long as all of the carbonate and bicarbonate ions present in solution have been fully
titrated (e.g. Figure 4). Using this approach, the in-house seawater standard was titrated
multiple times across a range of pH values between 3.2 and 3.7 (Figure 5A and 5B). The
alkalinity for each titration was computed and then plotted versus the residual H+
concentration (Figure 5A). The pKa of the dye was then adjusted so that the alkalinity across
the pH range was constant (Figure 5B). Using this pKa adjustment procedure, we obtained a
pKa of 3.513 for the in-house seawater standard with a mean alkalinity 2138 ± 1.2 μmol kg−1
(Figure 4). The pKa obtained for BPB in this study is significantly lower than the values of
15
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
320
321
322
323
324
3.7037 and 3.695 obtained by Okamura et al. (2010) and King and Kester (1989),
respectively.
As a check, we also used the “Gran Function” approach to calculate the total alkalinity of the
in-house seawater standard (Gran, 1952, Gran, 1988) (Figure 5A). The alkalinity obtained
using this approach was 2138 ± 10 μmol kg−1 and compares favourably to our earlier value of
2138 ± 1.2 μmol kg−1 although the error associated with the Gran Function approach is
approximately eight times larger.
This approach also revealed no systematic variation in alkalinity values with increasing
hydrogen ion concentration, and hence the volume of acid added to the sample. Similarly, the
titration of the unpurified dye did not show any systematic variation with varying hydrogen
ion concentration (Figure 3) although the measured alkalinity concentration was slight offset
from the concentration for the in-house seawater standard determined using the purified dye.
The precision of the method was also slightly reduced when the unpurified dye was used
compared to use of the purified dye (Table 3). Various batches of bromophenol blue dye from
three different vendors (Acros Organic, Sigma-Aldrich and Fisher Chemicals) were also
tested and had no significant difference between the BPB brands (Table 2).
Tests on purified versions of the BCG and BCP dyes were also undertaken to determine
whether they produced results similar to that of BPB. The results show that these two dyes
were comparable to that of BPB indicating that all three dyes are suitable for determining
seawater alkalinity to a precision of ± 4 µmol kg-1 or better (Table 3). Note that the pKa values
for these dyes were slightly adjusted so that they produced consistent alkalinity values across
and the range of hydrogen ion concentrations, as done for BPB in figures 4 and 5 (Table 3).
16
325
326
327
328
329
330
331
332
333
334
335
336
337
338
339
340
341
342
343
344
345
346
347
348
349
Errors associated with weighing, temperature and salinity.
Breland II and Byrne (1993) outlined the main errors associated with the colorimetric
determination of seawater alkalinity with the main errors associated with the weighing of the
sample and the titrant acid. The main errors in our experiments are associated with the weight
of the acid required to titrate the sample and with dye measurement by UV-visible
spectrophotometry. The weighing error was approximately two parts in ten thousand (±
0.0002 g for a 2 g acid addition), which results in an alkalinity error of approximately ± 0.4
mol kg-1. Repeated absorption analysis of fully titrated and purged samples resulted in an
alkalinity error in the range of ± 0.4-1.0 mol kg-1 using the OceanOptics USB4000
spectrophotometer. This range is larger than the error purely associated with weighing the
acid and sample and is likely to represent errors associated with the temperature and salinity
calibration of the dye, and subtle differences in alignment with replacing the sample cell in
and out of the sample cell holder.
Two factors that can influence the accuracy of colourimetric alkalinity determination is that
of temperature and salinity (Breland II and Byrne, 1993, Okamura et al., 2010). To test the
temperature dependence of absorbance ratio R for the acid (HI-) and base (I2-) forms of BPB,
we measured R across a temperature range for a sample with known alkalinity acidified to a
pH of ~3.5. These ratios were then referenced back to 25C (Figure 6). The absorbance ratio
of the acid (HI-) and base (I2-) forms showed strong temperature dependence whereby R
increases as the temperature of the sample decreases. For instance, we observed a 10%
increase in the R when the solution temperature was decreased from 25C to 10C (Figure 6).
To correct for this temperature effect we defined the following equation:
17
350
351
352
353
354
355
356
357
358
359
360
361
362
363
364
365
366
367
368
369
370
371
372
373
374
R(25)=R(t ) [1+(0.006774 ± 0.000009)(25−t)] (16)
where R(25) is the absorbance ratio at 25C and R(t) is the absorbance ratio at a given
temperature (t) between 10C and 35C. The slope of this relationship is slightly less than for
BCG (Breland II and Byrne, 1993).
To determine the influence of salinity on pKa for BPB, we measured the pH of an in-house
standard seawater sample across a salinity range from 36.4 to 21.8. The initial pH of the
solution was measured at 3.30. After each subsequent dilution using Milli-Q water, the pH of
the sample was re-determined (Figure 7). The effect of changing salinity on sample pH can
be described using the following equation:
p Ka=p Ka (35)+[(0.00174 ± 0.00008) (35−S )] (17)
where pKa is the subsequent pH for each of the diluted samples and pKa(35) is the calculated
pH of our acidified and purged seawater sample at salinity 35. In equation (17) S represents
salinity. The change in the pKa for BPB associated with changing salinity was found to be
relatively small (Figure 7), which is consistent with the work of Breland II and Byrne (1993)
where they also noted that salinity had a minor influence on the pKa for BCG.
Field application
As a test, we utilised BPB alkalinity method, in conjunction with natural seawater pH
measurements utilising purified mCP dye (Liu et al., 2011), to track changes in carbonate
system parameters on the algal reef flat at One Tree Island, Queensland Australia. At low-tide
seawater samples were collected from the algal reef flat (Latitude 23°30.695 S, Longitude
152°05.271 E) across a 4-5 hour sampling window. For field-based alkalinity determination,
18
375
376
377
378
379
380
381
382
383
384
385
386
387
388
389
390
391
392
393
394
395
396
397
398
399
60 mL seawater samples were collected periodically (~30 minutes) from isolated pools
during low tide and filtered through a 0.2 µm syringe filters (Pall, Australia). Two dome
experiments were also conducted in parallel during the low-tide period to quantify carbonate
production and dissolution within the algal pools when isolated from the atmosphere and
surrounding waters. During these experiments, the domes were placed on the algal turf, thus
isolating a section of benthos and allowing changes in water chemistry that attribute to the
inorganic and biological processes within. The domes were fully filled with ambient
seawater and any remaining air was removed from the domes to prevent CO2 gas exchange.
Samples were periodically removed from the domes for pH (Liu et al. 2011), alkalinity and
dissolved oxygen determination. The alkalinity samples were filtered upon collection and
then stored on ice and transported back to the One Tree Island research station where they
were analyzed. Samples were measured within 2 to 8 hours of collection. Samples were not
poisoned before analysis. Before analysis, alkalinity samples were accurately weighed into 50
mL wide-neck volumetric flasks and then acidified with HCl containing BPB and re-
weighed. The final pH of each sample was around ~3.5. Samples were warmed to room
temperature (approximately 25C), purged and then analyzed using the USB-
spectrophotometer system noting that the temperature at the time of measurement was
carefully determined and accounted for in the final alkalinity results. The results from these
field measurements are presented in Figures 8A, 8B and 8C. The alkalinity and natural
seawater pH measurements made on samples collected from the isolated algal reef flat pools
and dome experiments showed significant variations. For samples collected during the
daytime period (4 January 2015), seawater alkalinity decreased while natural seawater pH
increased, which is consistent with calcium carbonate production and photosynthesis by
crustose coralline algae. During the nighttime period (5 January 2015), seawater alkalinity
increased while natural seawater pH decreased, which is consistent with calcium carbonate
19
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
dissolution and organic matter respiration. The diurnal variations in seawater alkalinity are
large (up to ± 300 µmol kg-1) and were well outside procedural errors for both alkalinity and
natural pH measurement. The beauty alkalinity of the method presented here is that alkalinity
measurements can be made quickly, approximately 10-15 samples per hour, with a high
precision thus allowing the study of ecosystem changes with high temporal resolution, such
as at One Tree Island.
Marine ecosystems have become a significant research focus especially in coastal areas
where chemical and biological variations can be large and dynamic, such as at One Tree
Island (e.g. Silverman et al. (2012)). In this study, a simple but accurate alkalinity method has
been developed to determine seawater alkalinity. This method is well suited to field
experiments and sites where access to high-tech instrumentation and laboratories is limited.
Moreover, when this method is combined with similar dye methods for constraining natural
seawater pH, it provides an easy way to constrain the marine carbonate system with a high
degree of accuracy and precision (Figure 8C). The method is applicable especially when there
are dynamic diurnal changes in calcification, dissolution, photosynthesis and respiration.
20
425
426
427
428
429
430
431
432
433
434
435
436
437
438
439
440
441
442
443
Conclusions
In this study, we developed a method utilising BPB as an end-point indicator dye for making
simple one-point alkalinity measurements. HPLC analysis of the BPB dye reveals impurities
that can result in a slight loss of precision (e.g. 4 µmol kg-1) of the procedure compared to
alkalinity measurements made with the purified dye (±1.2 µmol kg-1). The absorbance
characteristics of the BPB dye is strongly influenced by changes in temperature (10°C - 35
°C) but is weakly influenced by changes in salinity (20-36). High precision (± 1.2 µmol kg-1)
alkalinity measurements are obtainable using a low-cost and low-powered USB-
spectrophotometer system. If high precision alkalinity measurements are not required, i.e. ≥
±4 µmol kg-1 then unpurified BPB maybe suitable. The one-point alkalinity method utilising
BPB as an end-point indicator appears to be ideal for use in remote field locations where
quick, but precise alkalinity measurements are required to help constrain the marine
carbonate system.
Acknowledgments
VN was supported through an ANU funded PhD scholarship. We acknowledge managers at
One Tree Island for their support during the undertaking of fieldwork on the island. Funding
Support from the Australian Research Council (DP170103067) is also acknowledged.
21
444
445
446
447
448
449
450
451
452
453
454
455
456
457
458
459
460
461
462
Table 1. Measured and calculated molar absorption coefficient ratios for e1, e2, e3 at 25oC and
salinity 35 for this study compared with Okamura et al., (2010).
Lab experiment Okamura et al. (2010)
e1 0.00533 0.0027
e2 2.232 2.458
e3 0.0319 0.1566
Table 2: Measured alkalinity using the purified and impure bromophenol blue from different
vendors for a sample size of n =10.
Bromophenol Blue Pure dye
Average ± SD
(µmol kg-1)
Impure dye
Average ± SD
(µmol kg-1)
P value$
Acros Organic 2138 ± 1.2 2134 ± 2.7 <0.001
Sigma-Aldrich 2138 ± 1.9 2132 ± 3.6 <0.001
Fisher Chemicals 2138 ± 1.4 2134 ± 3.9 0.0069
$ P values indicate that the alkalinity measurements utilising the impure BPB dye are
significantly different to alkalinity measurements utilising the purified BPB dye.
22
463
464
465
466
467
468
469
470
471
472
473
474
475
Table 3. Average alkalinity measurement for ANU standard seawater using the purified
bromophenol blue, bromocresol green and bromocresol purple dyes.
Indicator pKa Alkalinity (Average ± SD, n = 10)
Experimental Literature Literature pKa Adjusted pKa
Bromophenol blue 3.513 3.7037$ 2325 ± 46.6 2138 ± 1.2†
Bromocresol green 4.4099 4.4160# 2143 ± 4.9 2138 ± 3.8†
Bromocresol purple 5.9718 5.9770& 2139 ± 3.8 2136 ± 4.3†
References: $ Okamura et al. 2010, #Breland II and Byrne, 1993; &Yao and Byrne, 2001.
†P values for these alkalinity measurements utilising purified dyes are not significant (P >
0.2) i.e. different from each other.
23
476
477
478
479
480
481
482
Figure captions
Figure 1. Absorbance spectra of BPB dye in acidic (HI-) and basic (I2-) forms.
Figure 2. Plot showing the 3D UV-Vis spectra (300 to 600 nm) for HPLC profiles of the A.
unpurified BPB dye (10 mmol L-1), B. purified BPB dye (10 mmol L-1), C. non-purified BCG
dye (~10 mmol L-1) and D. purified BCG dye (1~0 mmol L-1). For each chromatogram the
total injection volume was 20µL at a flow rate of 1 mL min-1.
Figure 3. A. Multiple alkalinity determinations using impure and pure BPB (average
alkalinity of 2134 µmol kg-1 ± 2.7 µmol kg-1 and 2138 ± 1.2 µmol kg-1 (n = 10) for an in-
house seawater standard (ANU Std) respectively. B. Multiple alkalinity determinations using
impure and pure BPB (an average alkalinity of 2233 µmol kg-1 ± 3.9 µmol kg-1 and 2232 ±
1.3 µmol kg-1 (n = 5) for alkalinity CRM, respectively. The certified alkalinity for the CRM
(batch 140) was 2232.58 ± 0.80 µmol kg–1.
Figure 4. Plot showing changes in alkalinity versus mass of HCl for various pKa values for
BPB. The pKa for BPB was optimised using an in-house seawater standard to a value of
3.513 where the variation in alkalinity concentration was at a minimum.
Figure 5. A. Amount of hydrogen ion versus mass of HCl added to the sample. B. Total
alkalinity measurements for the CRM and an in-house seawater standard versus sample
hydrogen ion concentration measured using the USB UV-VIS spectrophotometer system. The
mean alkalinity and the standard deviation for the CRM (batch 140) and an in-house seawater
24
483
484
485
486
487
488
489
490
491
492
493
494
495
496
497
498
499
500
501
502
503
504
505
506
standard (ANU Std) are 2232 ± 1.3μmol kg−1 (n = 5) and 2138 ± 1.2 μmol kg−1 (n = 10),
respectively.
Figure 6. A. Plots of R(25)/R(t) versus the temperature range from 10-35oC of the CRM and in-
house seawater standard (ANU Std). The best fit linear regression line for the data is given
as: R(25)=R(t ) [1+(0.006774 ± 0.000009)(25−t)]. B. Plot of residuals versus the temperature
for the CRM and the in-house seawater standard.
Figure 7. Plot of subsequent pH of diluted and in-house seawater standard (ANU Std) versus
salinity at 25°C. The best fit linear regression line for the data is given as:
p Ka=p Ka (35)+[(0.00174 ± 0.00008) (35−S )]
Figure 8. Examples of A. alkalinity and B. pH measurements versus time of day for samples
collected on the algal ridge at low tide from One Tree Island, Queensland, Australia. Samples
were collected within a 24-hour window and measured within 2-8 hours of collection. C.
Transformation of data from A and B into alkalinity versus DIC plot overlaid with aragonite
saturation state ().
25
507
508
509
510
511
512
513
514
515
516
517
518
519
520
521
522
523
524
525
526
527
References
BELLERBY, R. G. J., OLSEN, A., JOHANNESSEN, T. & CROOT, P. 2002. A high
precision spectrophotometric method for on-line shipboard seawater pH
measurements: the automated marine pH sensor (AMpS). Talanta, 56, 61-69.
BRELAND II, J. A. & BYRNE, R. H. 1993. Spectrophotometric procedures for
determination of sea water alkalinity using bromocresol green. Deep Sea Research
Part I: Oceanographic Research Papers, 40, 629-641.
CLAYTON, T. D. & BYRNE, R. H. 1993. Spectrophotometric seawater pH measurements:
total hydrogen ion concentration scale calibration of m-cresol purple and at-sea
results. Deep Sea Research Part I: Oceanographic Research Papers, 40, 2115-2129.
DEGRANDPRE, M. D., SPAULDING, R. S., NEWTON, J. O., JAQUETH, E. J.,
HAMBLOCK, S. E., UMANSKY, A. A. & HARRIS, K. E. 2014. Considerations for
the measurement of spectrophotometric pH for ocean acidification and other studies.
Limnology and Oceanography: Methods, 12, 830-839.
DICKSON, A. G., AFGHAN, J. D. & ANDERSON, G. C. 2003. Reference materials for
oceanic CO2 analysis: a method for the certification of total alkalinity. Marine
Chemistry, 80, 185-197.
DICKSON, A. G. & GOYET, C. 1994. Handbook of methods for the analysis of the various
parameters of the carbon dioxide system in seawater
DICKSON, A. G., SABINE, C. L. & CHRISTIAN, J. R. 2007. Guide to Best Practices for
Ocean CO2 Measurements. In: DICKSON, A. G., SABINE, C. L. & CHRISTIAN, J.
R. (eds.). PICES Special Publication 3.
EASLEY, R. A. & BYRNE, R. H. 2012. Spectrophotometric calibration of pH electrodes in
seawater using purified m-cresol purple. Environmental science & technology, 46,
5018-5024.
26
528
529
530
531
532
533
534
535
536
537
538
539
540
541
542
543
544
545
546
547
548
549
550
551
552
GRAN, G. 1952. Determination of the equivalence point in potentiometric titrations. Part II.
Analyst, 77, 661-671.
GRAN, G. 1988. Equivalence volumes in potentiometric titrations. Analytica Chimica Acta,
206, 111-123.
HARALDSSON, C., ANDERSON, L. G., HASSELLÖV, M., HULTH, S. & OLSSON, K.
1997. Rapid, high-precision potentiometric titration of alkalinity in ocean and
sediment pore waters. Deep Sea Research Part I: Oceanographic Research Papers,
44, 2031-2044.
KING, D. W. & KESTER, D. R. 1989. Determination of seawater pH from 1.5 to 8.5 using
colorimetric indicators. Marine Chemistry, 26, 5-20.
LAI, C.-Z., DEGRANDPRE, M. D., WASSER, B. D., BRANDON, T. A., CLUCAS, D. S.,
JAQUETH, E. J., BENSON, Z. D., BEATTY, C. M. & SPAULDING, R. S. 2016.
Spectrophotometric measurement of freshwater pH with purified meta-cresol purple
and phenol red. Limnology and Oceanography: Methods, 14, 864-873.
LIU, X., BYRNE, R. H., LINDEMUTH, M., EASLEY, R. & MATHIS, J. T. 2015. An
automated procedure for laboratory and shipboard spectrophotometric measurements
of seawater alkalinity: Continuously monitored single-step acid additions. Marine
Chemistry, 174, 141-146.
LIU, X., PATSAVAS, M. C. & BYRNE, R. H. 2011. Purification and Characterization of
meta-Cresol Purple for Spectrophotometric Seawater pH Measurements.
Environmental Science &Technology, 45, 4862-4868.
MARTZ, T. R., DICKSON, A. G. & DEGRANDPRE, M. D. 2006. Tracer Monitored
Titrations: Measurement of Total Alkalinity. Anal. Chem., 78, 1817-1826.
MARTZ, T.R., CONNERY, J.G. & JOHNSON, K.S. 2010. Testing the Honeywell Durafet®
for seawater pH applications. Limnology and Oceanography: Methods 8, 172-184.
27
553
554
555
556
557
558
559
560
561
562
563
564
565
566
567
568
569
570
571
572
573
574
575
576
577
MILLERO, F. J., ZHANG, J.-Z., FIOL, S., SOTOLONGO, S., ROY, R. N., LEE, K. &
MANE, S. 1993a. The use of buffers to measure the pH of seawater. Marine
Chemistry, 44, 143-152.
MILLERO, F.J., ZHANG, J.-Z., LEE, K. & CAMPBELL, D.M., 1993b. Titration alkalinity
of seawater. Marine Chemistry, 44, 153-165.
OKAMURA, K., KIMOTO, H. & KIMOTO, T. 2010. Open-cell Titration of Seawater for
Alkalinity Measurements by Colorimetry Using Bromophenol Blue Combined with a
Non-linear Least-squares Method. Analytical Sciences, 26, 709-713.
PATSAVAS, M. C., BYRNE, R. H. & LIU, X. 2013. Physical–chemical characterization of
purified cresol red for spectrophotometric pH measurements in seawater. Marine
Chemistry, 155, 158-164.
RÉROLLE, V.M.C., FLOQUET, C.F.A., MOWLEM, M.C., CONNELLY, D.P.,
ACHTERBERG, E.P. & BELLERBY, R.R.G.J. 2012. Seawater-pH measurements for
ocean-acidification observations. TrAC Trends in Analytical Chemistry 40, 146-157.
SILVERMAN, J., KLINE, D.I., JOHNSON, L., RIVLIN, T., SCHNEIDER, K., EREZ, J.,
LAZAR, B. & CALDEIRA, K. 2012. Carbon turnover rates in the One Tree Island
reef: A 40-year perspective. Journal of Geophysical Research: Biogeosciences 117,
G03023.
TAPP, M., HUNTER, K. A., CURRIES, K. & MACKASKILL, B. 2000. Apparatus for
continuous-flow underway spectrophotometric measurement of surface water pH.
Marine Chemistry, 72, 193-202.
YAO, W. & BYME, R. H. 2001. Spectrophotometric Determination of Freshwater pH Using
Bromocresol Purple and Phenol Red. Environmental Science &Technology, 35, 1197-
1201.
28
578
579
580
581
582
583
584
585
586
587
588
589
590
591
592
593
594
595
596
597
598
599
600
601
YAO, W., LIU, X. & BYRNE, R. H. 2007. Impurities in indicators used for
spectrophotometric seawater pH measurements: Assessment and remedies. Marine
Chemistry, 107, 167-172.
29
602
603
604