M517-E190D Digital Angiography System BRANSIST alexa Read this manual thoroughly before you use the product. Keep this manual for future reference. Operation Guide
Transcript
Page 1: Operation Guide - Shimadzu

Operation G

uideBRA

NSIST alexa

Digital A

ngiography System

M517-E190D

Digital Angiography System

BRANSIST alexa

Read this manual thoroughly before you use the product.Keep this manual for future reference.

Operation Guide

Page 2: Operation Guide - Shimadzu

This Operation Guide describes the basic operation of BRANSIST alexa.

For detailed information, refer to the following “Operation Manuals”.

DAR-9500f Operation Manual (M517-E181)

D150GC-40 Operation Manual (M501-E085)

MH-200S Operation Manual (M523-E016)

MH-300 Operation Manual (M523-E014)

Please read and observe all safety precautions.

SymbolsThe following symbols are used in this Operation Guide.

Symbols Meaning

WARNINGIndicates a potentially hazardous situation which, if not avoided, could result in serious injury or possibly death.

CAUTIONIndicates a potentially hazardous situation which, if not avoided, may result in minor to moderate injury or equipment damage.

NOTEEmphasizes additional information that is provided to ensure the proper use of this product.

1, 2, 3, … Indicates the operation in order.

1, 2, 3, … Indicates further details of above operation.

● Indicates the operating description and items.

Indicates the result of operation.

Describes the points, notices and references of operation.

© 2011-2016 Shimadzu Corporation. All rights reserved.

General Description of BRANSIST alexa is the digital angiography system with FPD (Flat Panel Detector).BRANSIST alexa, a combination of high quality digital imaging with low dose exposures, provides fast and efficient operation for PCI and diagnostic studies.

Page 3: Operation Guide - Shimadzu

3

Contents 1 System Components ...............................4Examination Room ..................................................................4Control Room ........................................................................12

2 System Startup and Shutdown ............. 16Startup ...................................................................................16X-ray Tube Warm-up .............................................................17Shutdown ...............................................................................19Daily Inspection .....................................................................21

3 Start Study ............................................22Input Patient Information .......................................................22

4 Fluoroscopy/Radiography .....................23Operation of C-arm ................................................................23Operation of Catheterization Table ........................................27Fluoroscopy/Radiography ......................................................28Change Fluoroscopy/Radiography Program .........................29HV Flip ...................................................................................30Change Settings of Injection Delay .......................................31Recording Fluoroscopy Image ...............................................32Adjusting Image .....................................................................33Adding, Deleting and Displaying Reference Images .............35Related Operation for Study ..................................................36Measuring and Analyzing ......................................................37Operating DSA Tool ...............................................................41Printing Images ......................................................................42Closing Study ........................................................................42

5 Studies Management ............................43Managing Studies ..................................................................43Sending Studies ....................................................................44Writing to CD/DVD .................................................................45

6 Trouble Shooting ...................................46Emergency Halt .....................................................................46Safety Switch .........................................................................46Shutdown Failure ...................................................................47Digital Angiography system application problem ...................48FPD Problem .........................................................................49Power Failure ........................................................................50Error Messages .....................................................................51

7 Appendix ...............................................58IVR NEO buttons/Keyboard Shortcut keys functions ............58

Page 4: Operation Guide - Shimadzu

4

FPD

X-ray Tube

Collimator IVR NEO Catheterization Table

CyberGrip(or Cyber Console)

Local Console

Foot Switch (Examination Room)

Acquisition Monitor

Review Monitor

C-arm Indicator

C-arm

1 System ComponentsExamination Room

Page 5: Operation Guide - Shimadzu

5

1

Syst

em

Com

pone

nts

■ CyberGrip

Stop Switch

Release Switch

Field of View Switch

Catheterization Table Up/Down Switch

Tabletop Unlocking Switch (Located at the rear side of the grip)

C-arm Rotation Selector Switch

Direct Memory SwitchesC-arm Travel Selector Switch

FPD SID Switch

■ Cyber Console

Field of View Switch

Center Reset Switch

C-arm Rotation/Travel

FPD SID Confirmation Switch

Release Switch (on the side of the console) Direct Memory Switches

Stop Switch

Catheterization Table Up/Down Switch

Tabletop Unlocking Switch

Page 6: Operation Guide - Shimadzu

6

■ Local Console

Type A

① ④③②

⑧ ⑥ ⑤⑦

No. Name or Explanation

① C-arm Working Positions Switch

② Direct Memory Switches

③ Operate the C-leaf of the Collimator by moving up and down.

④ Operates the H/V Collimation. The Collimator opens up when this is pressed down.

⑤ Operates the Compensation Filter in the Collimator.

⑥ Field of View Switch

⑦ Center Reset Switch

⑧ C-arm Parking Switch

Page 7: Operation Guide - Shimadzu

7

1

Syst

em

Com

pone

nts

Type B

① ⑨ ⑤⑧

④⑥③⑦②

No. Name or Explanation

① C-arm Working Position Switch

② Direct Memory Switches

③ Operate the C-leaf of the Collimator by moving up and down.

④ Operates the H/V Collimation.

⑤ Operates the Compensation Filter in the Collimator.

⑥ Leaves Oper Switch. The Collimator opens up when this is pressed.

⑦ AP/PA Position Switch

⑧ C-arm Parking Switch

⑨ Set Switch

Page 8: Operation Guide - Shimadzu

8

* How to use Direct Memory Function (CyberGrip)

Type A

1 Select a button from , , and .

2 Select the location of the C-arm from this area.

3 Press and hold to move the C-arm to memory position.

Type B

1 Select a button from .

2 Select the location of the C-arm from this area.

3 Press and hold to move the C-arm to memory position.

Page 9: Operation Guide - Shimadzu

9

1

Syst

em

Com

pone

nts

* How to use Direct Memory Function (Cyber Console)

Type A

1 Select a button from , , and .

2 Select the location of the C-arm from this area.3 Press and hold to move the

C-arm to memory position.

Type B

1 Select a button from .

2 Select the location of the C-arm from this area.3 Press and hold to move the

C-arm to memory position.

Page 10: Operation Guide - Shimadzu

10

■ IVR NEO The IVR NEO features 16 buttons with LED indicators and a multi-function Joystick with button.

Default button settings are as following table. Refer to Appendix for the other functions available.

Joystick

Category Button LED Meaning A R

Acquisition

Select Fluoro Program ON=Fluoro menu displayed.

Select Rad Program ON=Rad menu displayed.

Select Image Comment ON=Comment menu displayed.

Add Reference Image ON=can add Reference Image.

Fluoro Record (Last N Seconds mode)

ON=can record the latest fluoro loop (unavailable for Direct Fluoro Record mode 

"Recording Fluoroscopy Image" P.32)

Start/Stop/Close Timer ON=timer displayed on Acq. monitor.

Image display

Monitor Select ON=Acquisition monitor selected.

Single/Split Display ON=Review monitor split vertically.

Left/Right Focus ON=right half of split Review monitor selected.

Reference Mode ON=Reference Mode.

Cine Area ZoomON= Zoom the images only in the Examination

room.

Page 11: Operation Guide - Shimadzu

11

1

Syst

em

Com

pone

nts

A= Relevant to Acquisition monitor R= Relevant to Review monitor

Category Button LED Meaning A R

Other

Save Still Image ON=can save still image (must be paused).

16-up Selector ON=Review monitor in 16-image select mode.

LIVE ON=display DSA LIVE image.

Linkage with C-arm

PositioningON= send projection angle of currently

displayed image to the C-arm.

Image FilterON= filter and display the projection angle of

image as current C-arm angle.

The function keys of the digital system keyboard can be configured as IVR NEO buttons (except for [Monitor Select] and [Reference Mode] functions).For details, contact Shimadzu service representative.

NOTE

Page 12: Operation Guide - Shimadzu

12

Review Monitor X-ray Generator Console

Foot Switch

Acquisition Monitor

Keyboard Digital Operation Cabinet

Mouse C-arm Remote Console

Control Room

Page 13: Operation Guide - Shimadzu

13

1

Syst

em

Com

pone

nts

■ X-ray Generator Console

Hand Switch

Fluoroscopy ParametersMessage

Power ON ButtonPower OFF Button

Radiography Parameters

Heat Unit Indicator

Fluoroscopy Accumulated Time

Fluoroscopy Density ButtonIBS Button

BH Filter NumberC-arm Information

X-ray OFF ButtonSetting Button

Accumulated Fluoroscopy Time Reset Button

Warm-up Button

[Tool] menu

C-arm Collision Indicator

Focus Switch Button

Field of View Switch Button

Auto/Manual Switch Button

Radiography Density Button

The field of view switch window can remain open during a study to switch field of view size quickly.

ex.) [12] ⇔ [8]

NOTE

Page 14: Operation Guide - Shimadzu

14

■ Main Window

Acquisition Monitor

Study Information is displayed in reference mode.

Side Menu (displayed when the mouse pointer is put on the left edge of the screen)

Parameters

Acquisition Active

Image Selector

Status Messages ProgramsComment

Cine Control

Exit Button Modes/Info Network Activity

Timer

Image Viewer

Page 15: Operation Guide - Shimadzu

15

1

Syst

em

Com

pone

nts

Review Monitor

Side Menu

Image Viewer

Probe Signal

Digital Program Menu

is displayed in reference mode.

Status Messages ProgramsComment

Cine Control

Exit Button Modes/Info Network Activity

Study Information Study Information

Parameters Image Selector

Page 16: Operation Guide - Shimadzu

16

Startup

Be sure to perform daily inspection before patient use."Daily Inspection" P.21 NOTE

X-ray Generator Console

1 Press (ON) Starting up whole system.

Power ON Button

2 System Startup and Shutdown

Page 17: Operation Guide - Shimadzu

17

2

Syst

em

Star

tup

and

Shut

dow

n

Be careful!During warm-up, X-ray exposure is performed automatically without operating hand switch or foot switch.

CAUTION

The warm-up procedure can be performed when C-arm is at the park position even though the "fluoroscopy at park position" setting is set to OFF. In this setting X-ray exposure by the foot switches and the hand switch is not available.

NOTE

X-ray Generator Console

1 Press on the [Tool] menu Warm-up window is displayed.

2 Press Warm-up procedure is started.

The progress of warm-up is displayed.

When warm-up is complete, the progress bar displays full.

X-ray Tube Warm-up

Page 18: Operation Guide - Shimadzu

18

If you need to pause warm-up during procedure,

・ Press to pause warm-up.

Press to resume.

・ Press to cancel warm-up, and press

to confirm.

Page 19: Operation Guide - Shimadzu

19

2

Syst

em

Star

tup

and

Shut

dow

n

Shutdown

C-arm (Type A)

1 Set the C-arm to the center position

Press and hold the Center Reset Switch until travel is completed.

2 Move the C-arm to the park position

Press and hold the C-arm Parking Switch until travel is completed.

C-arm (Type B)

1 Set the C-arm to the center position

Press the Center Reset Switch to select the operation. Then press and hold the Set Switch until travel is completed.

2 Move the C-arm to the park position

Press the C-arm Parking Switch to select the operation. Then press and hold the Set Switch until travel is completed.

Page 20: Operation Guide - Shimadzu

20

Digital Radiography System

3 Click A confirmation message is displayed.

4 Confirm the message by clicking

X-ray Generator Console

5 Press (OFF) when System Display window is changed

Make sure the OFF button is lit.

To restart the system, wait 1 minute after turning off power to the X-ray generator, then press (ON) again. NOTE

Page 21: Operation Guide - Shimadzu

21

2

Syst

em

Star

tup

and

Shut

dow

n

Check the following items before using the system.

1 Before turning on the power switch. Ensure there is:

• No damage on the exterior.

• No Condensation or Water on the system.

• No twisting of the cables.

2 After turning on the power switch. Ensure there are:

• No abnormal sounds.

• No abnormal smells.

• No error messages on X-ray generator console, Acquisition monitor or Review monitor.

3 Check the operation of the C-arm and Catheterization Table.

Confirm that each operation functions correctly and there are no abnormal sounds.

• LAO-RAO / CAUD-CRAN operation of the C-arm.

• Change the position of the C-arm and move the FPD forward/backward.

• Up-and-down operation of Catheterization Table and Tabletop Lock Switches.

4 Check the Safety Switch operation.

• Actuation of the Safety Switch of the Collimator and FPD.

• Operation of Stop Switches.

5 Confirm the Field of View Switch.

• While performing fluoroscopy, change and confirm the field of view size of the FPD.

6 Confirm the operation of the Collimator.

• Check to be sure that the shutter of the Collimator is moving relative to the FPD field of view change.

7 Check to be sure that images are displayed.

• Make sure that fluoroscopy/radiography images are displayed on monitors properly.

Daily Inspection

Page 22: Operation Guide - Shimadzu

22

1 Click on the side menu [New Study] window is displayed.

2 Input each itemsBe sure to input the items with a dot.

3 Click Patient information will appear on the upper area of the main window.

If loops/images have not yet been recorded/saved, study information can be modified by clicking on the side menu.

Indicates Mandatory Items.

Patient Information

3 Start StudyInput Patient Information

Page 23: Operation Guide - Shimadzu

23

4

Fluo

rosc

opy/

Rad

iogr

aphy

■ CyberGrip

1 Backward/Forward Movement of FPDMove the CyberGrip in the up and down direction while pressing .

FPD moves according to the direction of the grip.

2 Rotation of C-arm (RAO/LAO)Move the grip in the up/down direction while pressing .

C-arm rotates according to the direction of the grip.

4 C-arm TravelMove the grip in the up/down direction or twist it in the left/right direction while pressing .

C-arm moves according to the direction of the grip.

3 Rotation of C-arm (CRAN/CAUD)Twist the grip in the left/right direction while pressing .

C-arm rotates according to the direction of the grip.

4 Fluoroscopy/RadiographyOperation of C-arm

Page 24: Operation Guide - Shimadzu

24

■ Cyber Console

2 Rotation of C-arm (RAO/LAO) Move the C-arm rotation/travel lever forward/backward while pressing C-arm rotation selector switch.

C-arm rotates according to the direction of the lever.

C-arm Rotation Selector Switch (upper on the far side)

3 Rotation of C-arm (CRAN/CAUD)Move the C-arm rotation/travel lever in the left/right direction while pressing C-arm rotation selector switch.

C-arm rotates according to the direction of the lever.

C-arm Rotation Selector Switch (upper on the far side)

4 C-arm TravelMove the C-arm rotation/travel lever forward/backward or in the left/right direction while pressing C-arm travel selector switch.

C-arm moves according to the direction of the lever. C-arm Travel Selector Switch

(lower on the far side)

1 Backward/Forward Movement of FPDMove the FPD SID lever forward/backward while pressing confirmation switch.

FPD moves according to the direction of the lever.

FPD SID Confirmation Switch (on the far side)

Page 25: Operation Guide - Shimadzu

25

4

Fluo

rosc

opy/

Rad

iogr

aphy

■ Local Console (C-arm position change)

MH-300 (Floor mount type)

No. Switch C-arm position

① Home position 0°

② Femoral position CW45°

③ Peripheral position CW40°

Movable range of the C-arm rotation is "LAO90-RAO-90" in the femoral position, however limited to "LAO30-RAO30" in the peripheral position.

NOTE

C-arm Working Positions Switch

③②

Page 26: Operation Guide - Shimadzu

26

MH-200S (Ceiling suspended type)

C-arm Working Positions Switch

②③

④⑥

No. Switch C-arm position

① Head position

② Left position

③ Left sheltered position

④ Right position

⑤ + Left diagonal position (press simultaneously)

⑥ + Right diagonal position (press simultaneously)

Page 27: Operation Guide - Shimadzu

27

4

Fluo

rosc

opy/

Rad

iogr

aphy

Horizontal movement or raising/lowering of catheterization table.

1 Horizontal Movement of Catheterization TableMove the table manually while pressing the unlocking switch.

CyberGrip Grip Switch Table Console

2 Raising of Catheterization TablePress (Up) switch.

Tabletop rises.CyberGrip Table Console

3 Lowering of Catheterization TablePress (Down) switch.

Tabletop lowers.CyberGrip Table Console

Operation of Catheterization Table

Page 28: Operation Guide - Shimadzu

28

Fluoroscopy

1 Confirm the fluoroscopy program・ Modify the condition as needed.

2 Press the fluoroscopy foot switch

Examination Room Control Room

When accumulated fluoroscopy time reaches to 5 minutes, alarm buzzer sounds and alarm stop button appears on the screen.

To stop the buzzer, press [alarm stop] button.

Radiography

1 Confirm the radiography program・ Modify the condition as needed.

・ Positioning shall be set by fluoroscopy. Perform fluoroscopy more than 3 seconds.

2 Press the radiography foot switch

Examination room

Radiography can be performed with the hand switch on the X-ray generator console. Depress the hand switch until the X-ray is exposed.

Hand Switch

Fluoroscopy/Radiography

Page 29: Operation Guide - Shimadzu

29

4

Fluo

rosc

opy/

Rad

iogr

aphy

Fluoroscopy

1 ConfirmMake sure that selected Fluoroscopy Program is displayed.

Radiography

1 ConfirmMake sure that selected Radiography Program is displayed.

Radiography

1 Click or on the IVR NEO

Preset Radiography parameters will appear.

2 Select a parameter from either of the following operation・ Select with , , or key on

keyboard and fix with key.

・ Select with IVR NEO joystick and fix with joystick button.

・ Click parameter directory with mouse.

Selected parameter is set.

Fluoroscopy/Radiography program presets and corresponding parameters are changeable. These parameters can be changed at the digital processing system in the control room or at the IVR NEO.

"IVR NEO" P. 10

■ Change Preset ParametersFluoroscopy

1 Click or on the IVR NEO

Preset Fluoroscopy parameters will appear.

2 Select a parameter with either of the following operation・ Select with , , or key on

keyboard and fix with key.

・ Select with IVR NEO joystick and fix with joystick button.

・ Click parameter directory with mouse.

Selected parameter is set.

■ Confirmation of Parameters

Change Fluoroscopy/Radiography Program

Page 30: Operation Guide - Shimadzu

30

Flip and acquire the fluoroscopy and radiography images in the Horizontal or Vertical direction.

1 Click the flip buttons on the side menu of Acquisition monitor to flip images

Flip horizontal

(Click the button again to change the setting back to normal)

Flip vertical

(Click the button again to change the setting back to normal)

2 Perform fluoroscopy or radiography

Reversed image is acquired.

It is selectable during fluoroscopy or radiography, but it is valid on next acquisition. Note that the acquired image can be reversed on Review monitor. However, the image sent to the image server will be reversed only if it is reversed before acquisition.

NOTE

HV Flip

Page 31: Operation Guide - Shimadzu

31

4

Fluo

rosc

opy/

Rad

iogr

aphy

Injection-Acquisition Delay (the delay in seconds between injection and beginning of acquisition), Mask-Acquisition Delay (the delay in seconds between end of mask acquisition and beginning of live acquisition)and Radiation Times are set in advance for each acquisition program. These values can be changed temporary during study according to the following procedure.

1 Click on the review monitor side menu

[Acquisition Parameters] window appears.

2 Enter the delay time by seconds to one place of decimal and the radiation times by seconds and click

3 Click

When select the acquisition program again, default values will be restored. NOTE

Change Settings of Injection Delay

Page 32: Operation Guide - Shimadzu

32

Last N seconds fluoroscopy record or Either Direct fluoroscopy record can be used.

Default setting is Last N seconds fluoroscopy record. If you want to use Direct fluoroscopy record, system setting is required beforehand.

NOTE

■ Last N seconds Fluoroscopy Record

1 Perform fluoroscopy

2 Click or press on IVR NEO The latest fluoroscopy loop will be saved.

■ Direct Fluoroscopy Record

1 Click Make sure that selected Fluoroscopy Program is displayed.

2 Perform fluoroscopy Fluoroscopy loop will be saved.

Direct fluoroscopy record can be operated with the side menu button, but not with the IVR NEO. NOTE

Recording Fluoroscopy Image

Page 33: Operation Guide - Shimadzu

33

4

Fluo

rosc

opy/

Rad

iogr

aphy

■ Display

1

2

4 3

Adjusting Image

No. Button Function

1

Adjust image brightness.Right-click: + (increase)Left-click: - (decrease)Click both mouse buttons simultaneously: 0

2

Adjust image contrast.Right-click: + (increase)Left-click: - (decrease)Click both mouse buttons simultaneously: 0

3

Enhance the image.

Sharpness Right-click: + (increase), Left-click: - (decrease),Click both mouse buttons simultaneously: 0

Convolution Right-click: + (increase), Left-click: - (decrease), Click both mouse buttons simultaneously: 0

Negative/positive conversion

H Flip Images can be flipped horizontally.

V Flip Images can be flipped vertically.

Deactivate all filters

Optimize the brightness and contrast frame by frame.

Optimize the image brightness and contrast on the monitor.

Select any one of buttons.

4

Zoom in and out the displayed image.Right-click: + (increase)Left-click: - (decrease)Click both mouse buttons simultaneously: Return to default

Page 34: Operation Guide - Shimadzu

34

■ Controlling Image Playback

IVR NEO JoystickKeyboardControl Bar

■ Playback

① ②

Control Bar Keyboard IVR NEO Joystick Function

Play/show next loop/image.

Play/show previous loop/image.

or (Press button) Pause

Advance to next frame. (when paused)

Return to previous frame. (when paused)

( Press more than 2 seconds)

(Press button)Resume forward play. (same speed as acquisition)

( Press more than 2 seconds)

Resume backward play. (same speed as acquisition)

No. Button Function

1

Adjust the loop playback speed.Right-click: + (increase)Left-click: - (decrease)Click both mouse buttons simultaneously: Return to default

2

Change loop playback mode.Play Single Loop with Synchronize:Plays back selected single loop until a new loop is received. Then, the new loop is selected and played back continuously.Play Single Loop:Plays back the selected loop continuously.

Page 35: Operation Guide - Shimadzu

35

4

Fluo

rosc

opy/

Rad

iogr

aphy

Keyboard

Insert

Delete

Home

End

PageUp

PageDown

IVR NEO Joystick IVR NEO Button

Adding, Deleting and Displaying Reference Images

Control Bar Keyboard IVR NEO Joystick Function

( Tilt and hold to the right)

Play frame-by-frame forward according to the speed of joystick angle when paused. (Larger the tilt angle, higher the speed)

( Tilt and hold to the left)

Play frame-by-frame backward according to the speed of joystick angle when paused. (Larger the tilt angle, higher the speed)

Keyboard IVR NEO Function

Add Reference Image.

Delete Reference Image.

Go to next loop with image selected as Reference and pause on the image.

(Tilt downward)In Reference mode, go to next loop with image selected as Reference and pause on the image.

Go to previous loop with image selected as Reference and pause on the image.

(Tilt upward)In Reference mode, go to previous loop with image selected as Reference and pause on the image.

Switch Playback mode/Reference mode.

Page 36: Operation Guide - Shimadzu

36

1 2

3

5 6

4

No. Button Function

1 New study (Start examination)

2 Re-synchronize the Review monitor display to the acquiring study.

3 Open previous study.

4 Close the study.

5 Display study information.

6 Display comment menu.

Related Operation for Study

Page 37: Operation Guide - Shimadzu

37

4

Fluo

rosc

opy/

Rad

iogr

aphy

■ Quantitative Coronary Analysis (QCA)

QCA Tool Bar

1 Click and select QCA tool bar will appear.

2 Perform calibrationThe default value of calibration factor is calculated as the measured object is located on the ISO center. Perform calibration according to the following procedure if needed.

To perform calibration, display the image of reference object (such as a catheter) previously acquired with the same magnification as the analysis object. The smaller reference object (the thinner the catheter), the greater the analysis error.

1 Click button from [Calibration] menu.

2 Select a catheter diameter on the image from the list, or input a value (mm).

Measuring and Analyzing

Page 38: Operation Guide - Shimadzu

38

3 Click on 3 points along the catheter.

Calibration factor (mm/pixel) will appear.

Click button from side menu to modulate magnification rate of Ref image. NOTE

3 Specify the stenosed artery

1 Click button from [Manual] menu.

2 Click points on opposite sides of the narrowest part of the stenosed artery.

The stenosis is measured and indicated on screen.

4 Specify the reference value

1 Click button from [Manual] menu.

2 Click points on opposite sides of the part of the un-stenosed artery.

The reference distance and percent stenosis are displayed.

Measuring a vessel diameter from angiography causes several errors. Therefore, do not diagnose or treat based solely upon this analysis result. Use the analysis result as a reference value.

WARNING

Page 39: Operation Guide - Shimadzu

39

4

Fluo

rosc

opy/

Rad

iogr

aphy

■ Left Ventricle Analysis (LV) Perform LV analysis with a projected image of RAO 30 degrees (or images of RAO 30 degree and

LAO 60 degree).

LV Tool Bar

1 Click and select LV Tool bar will appear.

2 Perform calibrationThe default value of calibration factor is calculated as the measured object is located on the ISO center. Perform calibration according to the following procedure if needed.

To perform calibration, display the image of reference object (such as a ball) previously acquired with the same magnification as the analysis object. The smaller reference object (the smaller the ball), the greater the analysis error.

1 Click from [Calibration] menu.

2 Input the ball diameter (mm) which is on the image.

3 Outline the ball by clicking 3 points.

Calibration factor (mm/pixel) is displayed.

Page 40: Operation Guide - Shimadzu

40

3 Outline the ventricular wallOutline each image of End-of-Diastole (ED) and End-of-Systole (ES). Outline both ED and ES image repeating the following procedure.

1 Click the projection icon that corresponds to the image.

RAO ED RAO ES LAO ED LAO ES

2 Using the mouse, click multiple points along the ventricular wall starting from the base of the aorta.

An outline curve passing through clicked points is created.

Move a point by clicking and dragging it to a new position.

Delete a point by clicking it with the left and right mouse buttons simultaneously.

3 Once the outline is specified to the opposite side of aortal base, click the icon of the next projection to be drawn.

The outline that has been drawn to the current image will be recorded.

4 Click The result of LV analysis is displayed.

This analysis is calculated with an approximation formula called "Area Length", so the calculation result includes approximation errors.

CAUTION

Page 41: Operation Guide - Shimadzu

41

4

Fluo

rosc

opy/

Rad

iogr

aphy

1 Click and select [DSA Tools] bar will appear.

2 Click [Pixel Registration] tool bar will appear.

Operating DSA Tool

No. Button Function

1

3

2

4

5

7

6

1 Setting mask image.

2 Switching ON/OFF of subtraction.

3 Create peak hold image.

4 Set ROI of MAP.

5 Resetting mask image.

6Executing reregistration. (Shift the mask image)

7Create a contour enhanced blood-vessel image from DSA image.

No. Button Function

1

2 5

3 4

6

1 Moves the mask image upward.

2 Moves the mask image downward.

3 Moves the mask image left.

4 Moves the mask image right.

5Return the mask image to the default (0, 0) position.

6 Select the value when moving less than 1 pixel.

Page 42: Operation Guide - Shimadzu

42

1 Click and select

Printer dialog will appear.

2 Setting any required items such as

3 Add images with

4 Click Printing is started.

1 Click from [Study] menu

Printing Images

Closing Study

Page 43: Operation Guide - Shimadzu

43

5St

udie

s M

anag

emen

t

1 Click from [Study] menu [Studies Management] window is displayed.

1 2 3 54 6 7 8 9 0

5 Studies ManagementManaging Studies

No. Button Function

1 SearchSearches based on the search target set by the [System] and [Drives] buttons, plus all search criteria.

2 View Displays the selected study or studies in the Image Viewer window.

3 RDSR(Available if there is dose report.)Displays the dose report.

4 SeriesWhen a study is selected in the Study List, Series displays all study series types contained in this study.

5 Send Sends selected studies to another server or system.

Page 44: Operation Guide - Shimadzu

44

1 Select the desired studies in the Study List

2 Click A list of destinations will appear.

3 Select the send type and the Process Before Transfer

4 Click the desired destination

5 When the confirmation window appears, click [OK]

Sending Studies

No. Button Function

6 Info Displays detailed information about the selected patient.

7 DeletePermanently deletes the unprotected studies selected from the local system.

8 Write Prepares and writes the selected studies to CD/DVD.

9 Protect Protects the selected studies from deletion.

0 Unprotect Removes protection from selected studies.

Page 45: Operation Guide - Shimadzu

45

5St

udie

s M

anag

emen

t

1 Select the studies

1 Select the desired studies from the Study List.

2 Click .

Prompted to insert a blank disc to the writer.

3 Insert a blank disk.

Media Writing dialog box appears.

Volume name is changeable.

Select the writing item from [Images], [RDSR] or [Both].

2 Click Writing is started.

To cancel a media write, click on the side menu of the Image Viewer window.

Do not write to media during a study, only write to media after study completion. NOTE

Writing to CD/DVD

Page 46: Operation Guide - Shimadzu

46

■ Halting the C-arm

1 Press (Stop switch)

CyberGrip Catheterization TableCyber Console

Stop Switch Stop SwitchStop Switch

■ Resetting the Emergency Halt

1 Turn (Stop switch) clockwise

When resetting, be sure to reset the pressed switch.Reset the STOP switch after the message "Stop Mode: Reset OK" is displayed on the X-ray generator console or the information display to ensure normal operation.

NOTE

Safety SwitchTouch sensors are mounted on the FPD front cover and the collimator cover. When these sensors are tripped, the movement of C-arm and catheterization table stops.To free the C-arm from collision, operate CyberGrip or CyberConsole carefully with pressing the release switch. C-arm can be moved slowly.FPD backward movement is possible without release switch operation.

CyberGrip Cyber Console

Release Switch Release Switch (on the side of the console)

6 Trouble ShootingEmergency Halt

Page 47: Operation Guide - Shimadzu

47

6

Trou

ble

Shoo

ting

If trouble occurs when trying to shut down the Digital Radiography System via [Exit] button (system does not shut down), shut down the system according to the following procedure.

Digital Operation Cabinet (Control Room)

1 Press and hold the Power button for at least 5 seconds The power goes off.

Cabinet (Equipment Room)

2 Open the switch cover in front of the cabinet and press the Power button for at least 5 seconds

Acquisition computer is turned off.

X-ray Generator Console

3 Press (OFF) X-ray high voltage generator and C-arm are shut down.

Shutdown Failure

Page 48: Operation Guide - Shimadzu

48

To restart the system, wait 1 minute after turning off power to the X-ray generator, then press (ON) again. NOTE

If the acquisition monitor is blank when restarting the system after shutdown failure, press the power button on the digital control cabinet in the equipment room to turn on the power.If the review monitor is blank when restarting the system after shutdown failure, press the power button on the digital operation cabinet in the control room to turn on the power.After the application screen is displayed on both acquisition and review monitors, shut down the system again by standard procedure and then restart the system.

NOTE

Digital Angiography system application problem

■ Image processing errors or unknown system problems If image processing errors or unknown system errors occur, the error dialog is displayed automatically.

1 Click The application reboots.

■ Mouse Pointer If the mouse pointer is locked within one monitor, press the [Scroll Lock] key on the keyboard.

■ Monitor If the screen is blank or gray-scale-like image is displayed, press the button shown in the figure below.

Review MonitorAcquisition Monitor

Page 49: Operation Guide - Shimadzu

49

6

Trou

ble

Shoo

ting

1 Select and click on the side menu

FPD calibration window appears.

2 Make sure that [Timeout (Seconds)] is set to 10 seconds

This defines the maximum number of seconds to wait before cancelling if there is a problem communicating with the FPD.

3 Select [Offset-All Modes] in the [Calibration Type] list.

4 Click Progress messages are displayed during calibration.

Completion message is displayed when calibration is completed.

5 Click

Click if you terminate the calibration before completing.If you terminate the calibration, be sure to calibrate again before acquisition.

NOTE

FPD ProblemAutomatic FPD calibration is supported to the system. Manual calibration is not necessary, so follow the instruction below when manually calibrating the system.

Page 50: Operation Guide - Shimadzu

50

Power FailureThis system is protected from power failure with UPS (uninterruptible power supply system). If the power failure occurs suddenly such as thunderbolt, screen display disappears. If power supply is restored within 5 minutes, system will be automatically restored and images on monitors will be redisplayed.So you can use system again.

If the power failure occurs for long time such as for more than 5 minutes or it is difficult to judge whether the momentary power failure occurs or the longer one, once shut down the system, and then re-activate the system according to the Startup procedure ( P.16) or "Chapter 3 on DAR-9500f Operation Manual (M517-E181)".

The system is designed to minimize data loss by power failure, however, it cannot be restored the acquisition image on power failure.After the system restores, the previous study on power failure re-activates automatically.

NOTE

If power failure occurs, FPD calibration data are restored to defaults and any acquisitions performed before automatic FPD calibration is executed cannot be considered to be reliable.Be sure to calibrate FPD manually after power failure.

"FPD Problem" P.49

NOTE

Page 51: Operation Guide - Shimadzu

51

6

Trou

ble

Shoo

ting

Error MessagesCommon error messages are described in this sectionFor other error messages, refer to the operation manual of each system.

■ Digital Radiography System The following error messages will appear on the system monitor.

Message Action

Note: "XX" or "X" represents variable information. Values and characters appear according to the error situation.

Burn Operation Failed.Insert a blank CD-R or DVD-R and re-execute the operation.

An error occurred during cache space reservation. If CD-R or DVD-R media are used as patient data archive,

archive all unsaved patient data to new CD-R or DVD-R.No available cache.

Cannot retrieve media information.

Confirm whether the media is inserted properly.

Cannot write DICOMDIR to CD-R (DVD-R). No patient has been added.

Confirm that writable media is inserted.

Character XX is not valid. Try other characters.

Could not find the PPS Manager. Please configure your PPS Manager in the hosts table.

Please contact Shimadzu service representative.

Error: Invalid query criteria entered.

Input an appropriate value.

No DICOM media found in drive X. Insert DICOM media.

Page 52: Operation Guide - Shimadzu

52

Error Messages Related to FPD

When "FPD ERROR" is displayed at upper left of the system monitor, errors in the following list are also displayed.

Stop study immediatelyand contact Shimadzuservice representative.

Action Message Description

[Sensor]:Temperature error

[Cooling Unit]:Water flow error

Error of sensor chassistemperature.

Error of FPD cooling unit water flow.

Page 53: Operation Guide - Shimadzu

53

6

Trou

ble

Shoo

ting

■ X-ray High Voltage Generator When the equipment detects errors, the error messages will appear on the X-ray generator console. When an error message is displayed, press and take the appropriate action listed.

Messages Related to the X-ray high voltage generator

The following message is displayed. Radiography is not possible while this message is displayed.

If "No Filament S" or "No Filament L" appears, the focus is switched to normal one. if this message appears after rebooting the X-ray high voltage generator, contact Shimadzu service representative.

No Filament S/L/Both Abnormal filament.

If this message appears repeatedly though pressing [RESET] key, contact Shimadzu service representative.

Radio Over Current

Measured kV Over

Measured kV Under

Starter Error

I.F. Over

Filament not ready

Measured tube current exceeded preset value by 200 mA or more.

Abnormal filament heating current.

Measured tube voltage exceeded permitted range.

Measured tube voltage fell below permitted range.

Starter failure.

Action Message Description

Page 54: Operation Guide - Shimadzu

54

Contact Shimadzu service representative.

Arcing Trouble

Board status Error

kV adjustment board Error

Line Voltage OVER

Charge Volt Error

Power Down

Supply voltage exceeded permitted range.

Fluo I/F Error Abnormal Fluo I/F Board.

Abnormal charging voltage of primary smoothing capacitor.

Abnormal control circuit supply voltage.

H.V.T. Not ConnectedAbnormal connection to high-voltage transformer.

Repeated arcing occurred.

Abnormal control board.

Abnormal kV Adjust Board.

Action Message Description

Messages Related to the X-ray Tube The following message is displayed.

Radiography is not possible while this message is displayed.

Action Message Description

HU Over Predicted

Value predicted to go out of range Continuing radiography with present setting will cause heat unit overrun.

Change the radiographyconditions or cease operation until heat unit value reduces.

LX-Tube Water Flow ErrorWater flow in heat exchanger of X-ray tube is stopped.

Radiography is not available while this warning displays. If this message appears repeatedly after resetting or rebooting the X-ray high voltage generator, X-ray exposure will be impossible in a few minutes.Stop study immediately.

Stop the operation until the heat unit value reduces.Thermal Over

X-ray tube unit temperature exceeded the permitted limit.

Heat units reached the permitted limit.

HU Full Stored

Page 55: Operation Guide - Shimadzu

55

6

Trou

ble

Shoo

ting

Action Message Description

LX-Tube Water Level WarningWater volume of heat exchanger gets low

Contact Shimadzu Service Representative.(It’s possible to do exposure X-ray while this Warning displays.)

LX Tube Thermal Warning

X-ray tube unit temperature exceeded 70 celusius degrees. The continued use can be detect "Thermal Over".

Cease operation or decrease frequency of exposure X-ray. (It’s possible to do exposure X-ray while this Warning displays.)

Starter is not workingThe starter is in standby status.

Contact Shimadzu service representative.

Error Messages at Power ON

Action Message Description

Contact Shimadzu service representative.

The data recording battery in the control cabinet must be replaced soon.

The data recording battery in the control cabinet is empty.

Cabinet Battery Empty

Cabinet Battery Warning

Power OFF the high voltage system, and reset it after confirming the switch of fluoroscopy and radiography is not ON.

Radiography switch was ON, when the high voltage system is ON.

Fluoroscopy switch was ON, when the high voltage system is ON.

Fluo switch Error

X-ray switch Error

Page 56: Operation Guide - Shimadzu

56

Other Messages

The following message is displayed. Radiography is not possible while this message is displayed.

Action Message Description

Door/InterlockThe examination room door is open.

Close the door.

FPD Water Flow ErrorHeat Exchanger of FPD detects error.

Stop study immediately and Contact Shimadzu service representative. (It’s possible to do exposure X-ray for about 30 minutes while this Warning displays.)

■ C-arm Action Message Description

Reset with the STOP switch or turn off and on the power again. If the equipment cannot be reset, contact Shimadzu service representative.

Motor Error XXXMotor command fail to drive the motor.

Camera Rotation ErrorError of FPD rotation potentiometer.

Collimator Error XXXError of the collimator potentiometer.

Collimator Rot ErrorError of collimator rotation potentiometer.

Limit Error XXXDetects the limit switch activation.

Ask Shimadzu service representative to readjust the limit switch.

It is possible to operate the C-arm by holding down the release switch. Reset with the STOP switch or turn off and on the power again. If the equipment cannot be reset, contact Shimadzu service representative.

Position Detect Error XXXMismatching of position detection and motor drive.

Potentio Error XXXDetects disconnection of position detector circuit.

Page 57: Operation Guide - Shimadzu

57

6

Trou

ble

Shoo

ting

Note: "XXX" represents values and characters according to the error situation.

Action Message Description

Contact Shimadzu service representative.

Pos. Calibration ErrorIncorrect installation setting of position detection.

Interlink Error ! XXXDetects disconnection of console.

Lever Calib. ErrorIncorrect installation setting of the console.

Low Battery Drop of battery voltage.

Although the equipment is available, contact Shimadzu service representative.

STOP Switch ErrorBreak in the stop switch circuit is detected.

One of the doubled circuits is out of order. Although the equipment is available for the temporary use, contact Shimadzu service representative immediately.

Initial Input Error !Release All Switches

Any switch may be pressed during startup.

Release all the switches and reset the equipment.

When operating the system by holding down the release switch, the collision prevention feature and safety switches are disabled.This may result in equipment damage or patient contact. Operate carefully to ensure the STOP switch can be pressed immediately.

CyberGrip Cyber Console

Release Switch Release Switch (on the side of the console)

CAUTION

Page 58: Operation Guide - Shimadzu

58

The settings of IVR NEO buttons (default settings: P.10 and P.11) and keyboard function keys can be changed to the functions in the following table.See “Chapter 12.8 Physicians Tab” on DAR-9500f Operation Manual (M517-E181) for changing the settings.

7 AppendixIVR NEO buttons/Keyboard Shortcut keys functions

Category Button Function A R

Fluoro

Fluoro ProgramAssociated predefined Fluoro mode selectedupon selection.

Save Fluoro Last Frame

Fluoroscopy images can be saved mode.

View Fluoro Last acquired Fluoro loop shown upon selection.

Direct Fluoro Record

Direct Record mode enabled upon selection.LED goes OFF upon completion of next Fluoro acquisition.

Rad

Rad ProgramAssociated predefined Rad mode selected upon selection.

Toggle Rad Mode(Biplane only)Change the plane in order of Frontal, Bi and Lateral.

Set Rad Mode BiPlane

(Biplane only)Change the plane to Bi-plane.

Set Rad Mode Frontal

(Biplane only)Change the plane to Frontal.

Set Rad Mode Lateral

(Biplane only)Change the plane to Lateral.

MAP

DSA MAP DSA-MAP mode.

DSA-MAP(Live) DSA-MAP Live mode.

DSA-MAP(Sub) DSA-MAP Sub mode.

FluoroMAP FluoroMAP Mask Creation mode.

FluoroMAP(Live) FluoroMAP Live mode.

FluoroMAP(Sub) FluoroMAP Sub mode.

Blank MAP Register the Blank mask.

MAP mode MAP mode.

Live mode Live mode.

Reset BG Mask Reset the background mask.

TRANS IVR NEO joystick changes MAP Power.

Sketch Display Sketch Display mode.

Sketch EditSketch Edition window is displayed, and the joystick become available.

Page 59: Operation Guide - Shimadzu

59

7

App

endi

x

Category Button Function A R

MAP

Contour Enhancement

Display the contour enhanced image.

Set Mask RegionROI of DSA-MAP is displayed on the image and the joystick is in the operation mode.

DSA Peak Hold Display the peak hold image.

Action

X-ray OFF Ban the X-ray exposure.

Dose Reset Dose meter reset upon selection.

Delete Reference Image

Reference image deleted upon selection.

Add Print Image displayed is added for printing upon selection.

Lateral Operation(Biplane only) Enable to operate from ACQ-L when “Edit Sketch” and “MAP Power” are changed.

Ref Sync RefSync mode.

PCI View mode PCI View mode.

Display

Filter with Reference image

Search and display the added reference images.

Zoom In Zoom level increases slightly upon selection.

Zoom Out Zoom level decreases slightly upon selection.

Flip image Horizontal

Flip an acquisition image horizontally.

Flip image Vertical Flip an acquisition image vertically.

Play

Play/Sync PlayON=Play Single Loop with Synchronize mode.OFF=Play Single Loop mode.

Play Forward Loop plays forward upon selection.

Play Backward Loop plays backward upon selection.

AnalysisLV Mode LV menu shown upon selection.

QCA Mode QCA menu shown upon selection.

Annotations

Text Add text tool selected upon selection.

LineLine tool added at mouse cursor location upon selection.

CurveCurve tool added at mouse cursor location upon selection.

ArrowArrow added at mouse cursor location upon selection.

CaliperCaliper tool added at mouse cursor location upon selection.

Analysis Analysis results added upon selection.

Study Study information added upon selection.

Clear Last Annotation

Last annotation deleted upon selection.

Clear All Annotation All annotations deleted upon selection.

Page 60: Operation Guide - Shimadzu

60

http://www.shimadzu.com

Category Button Function A R

Annotations

Show Annotation Properties

Annotations properties shown upon selection.

Pointer Pointer tool selected upon selection.

A=[relevant to Acquisition monitor], R=[relevant to Reference monitor]

Page 61: Operation Guide - Shimadzu

http://www.shimadzu.com

Page 62: Operation Guide - Shimadzu

Operation G

uideBRA

NSIST alexa

Digital A

ngiography SystemM517-E190D

Digital Angiography System

BRANSIST alexa

Read this manual thoroughly before you use the product.Keep this manual for future reference.

Operation Guide