OPERATION SPECIFICATION
FOR
TESCAN MIRA3 ELECTRON BEAM LITHOGRAPHY
Rev. Date Prepared by
01 03/05/2018 Manouchehr Teimouri
TESCAN MIRA3 Electron Beam Lithography
Page 2 of 17
1 SCOPE
1.1 The purpose of this document is to describe requirements and basic
operating instructions for the TESCAN MIRA 3, E-Beam Lithography tool.
This tool is intended for nano-patterning by electron beam.
2 SAFETY
2.1 Must be trained and signed off to use this equipment.
2.2 Record the tool malfunction, breakage, etc. in the log-sheet and send an
email to staff. Never try to fix anything by yourself.
3 APPLICABLE DOCUMENTS, MATERIALS AND REQUIREMENTS
3.1 For more information about the detailed operation of this tool refer to the
TESCAN factory manual – “DRAWBeam for Electron Beam Lithography” File
name: TESCAN MIRA 3 Manual.pdf.
4 OPERATION
4.1 Loading the sample
4.1.1 Log into Mira-TC software by clicking on MIRA3-TESCAN icon and entering user name
and password.
TESCAN MIRA3 Electron Beam Lithography
Page 3 of 17
4.1.2 Vent the chamber by clicking on the VENT button on the Vacuum Panel and wait to
reach “Venting finished” status. (Note: make sure the stage is at home position (X=0;
Y=0; Z=40; Rot & Tilt = 0) before you vent the chamber. If not, click on Home button in
stage control panel).
4.1.3 Open the chamber gently and load the sample onto sample holder. User must have
clean gloves on while loading and unloading the sample.
4.1.4 Gently close the chamber and hold the chamber’s door.
4.1.5 Pump the chamber by clicking on the PUMP button on the Vacuum Panel and wait to
reach “Vacuum ready” status.
TESCAN MIRA3 Electron Beam Lithography
Page 4 of 17
4.2 Turning on the electron beam
4.2.1 Move the stage above gold standard position
4.2.2 Open the “Beam Blanker” panel in the main SEM menu and select “Enable beam on
acquisition”.
4.2.3 Select SE detector in the SEM Detectors & Mixer panel. Make sure the BSE detector is
retracted.
TESCAN MIRA3 Electron Beam Lithography
Page 5 of 17
4.2.4 Set the exposure parameters including acceleration voltage (HV; 30 kV is recommended)
and Beam Index (BI; less than or equal to 5 is recommended) in Info Panel. (Note: for
large patterns where both dimensions are in the range of few tens of microns and
resolution is not as critical, use BI 8-10 to reduce the exposure time (higher BI is
equivalent to higher beam current)).
4.2.5 Turn the electron beam on by clicking on the BEAM ON button on the Electron Beam
panel.
4.2.6 Perform focus on gold standard at 50kx by turning WD knob on Control Panel. Once the
gold standard is focused, gradually decrease WD&Z in stage control panel (WD&Z = 15,
10, and then 6.0) and re-perform focus until gold standard is focused at WD 6.0.
TESCAN MIRA3 Electron Beam Lithography
Page 6 of 17
4.3 Electron Column Alignment
4.3.1 Perform fine focusing at magnifications ≥200kx.
4.3.2 Perform gun centering by automatic procedure. Click on AUTO GUN CENTERING on the
Electron Beam panel.
4.3.3 Perform Column Centering (Centering the aperture) manually. Click on icon on
the toolbar and use control panel knobs ( for X and for Y) or the trackball to
center the aperture.
4.3.4 Correct Astigmatism (Correcting non-circularity in electron beam). Click on icon
on the toolbar and use control panel knobs or the trackball.
4.3.5 Re-Perform focusing.
4.4 Exposure
4.4.1 Open the DrawBeam panel in the main Tools menu.
TESCAN MIRA3 Electron Beam Lithography
Page 7 of 17
4.4.2 Create a new project/open a saved project/ import from GDSII format by clicking on the
/ / icon.
4.4.3 Click on the icon in the DrawBeam panel to open the DrawBeam Process panel.
Enter the suitable Dose(s) and Spacing by double click on Dose window. (Note: keep the
spacing at minimum possible value;
(1) Exp pitch should be ≥ 4×DAC resolution (DAC resolution depends on view field)
(2) Exp pitch = spot size × spacing
TESCAN MIRA3 Electron Beam Lithography
Page 8 of 17
4.4.4 Move the stage to the faraday cup. Magnify to 50kx (the entire screen should be black;
i.e. all the electrons should be collected by faraday cup). Measure the beam current by
clicking on icon in the DrawBeam Process panel. The correct value of the beam
current will be automatically inserted in the panel.
4.4.5 Move the stage to the edge of the sample and find the scratch you made on sample
after spin coating. Click on Define UV and then Align sample to align sample with x and y
directions.
4.4.6 Go to the tip of the scratch. Take caution not to expose the sample. Find a tiny particle
and try to bring the particle on focus without changing WD. In order to do that you
need to change Z. Keep changing Z using buttons until you see a clear image
of the particle.
4.4.7 Open the Analysis & Measurement panel in the main Tools menu (steps 4.4.7-4.4.9 are
optional if step 4.4.6 is performed properly).
TESCAN MIRA3 Electron Beam Lithography
Page 9 of 17
4.4.8 Click on the New Point/line button and place the point (Magnification should be 400kx).
Start and stop the burning by clicking on the icon. The burning time is 10-20
seconds.
4.4.9 Perform final focusing and astigmatism correction on the spot to achieve an ideal
circular spot. Repeat burning and focusing, if necessary.
TESCAN MIRA3 Electron Beam Lithography
Page 10 of 17
4.4.10 Blank the beam by clicking on the Single button on the Info Panel.
4.4.11 Move the stage to exposure location by entering new X and/or Y coordinates.
4.4.12 Start the exposure by clicking on button in the DrawBeam Process panel. Move the
stage and start another exposure, if needed.
4.4.13 After exposure is finished, turn the beam off by clicking on the BEAM ON button on the
Electron Beam panel.
4.5 Unloading the sample
4.5.1 Move the stage to home location by clicking on home button in Stage Control Panel.
4.5.2 Vent the chamber by clicking on the VENT button on the Vacuum Panel and wait to
reach “Venting finished” status.
4.5.3 Open the chamber gently and unload the sample.
4.5.4 Gently close the chamber and hold the chamber’s door.
4.5.5 Pump the chamber by clicking on the PUMP button on the Vacuum Panel and wait to
reach “Vacuum ready” status.
4.5.6 Log off the software. Click on the log off in the main tools menu. You will be asked two
question: 1) do you really wish to log off: Yes 2) Go to the STANDBY mode: No
4.5.7 Fill out the log-sheet.
TESCAN MIRA3 Electron Beam Lithography
Page 11 of 17
Appendix A:
Some Examples for writing circles, lines, rings, and texts
TESCAN MIRA3 Electron Beam Lithography
Page 12 of 17
TESCAN MIRA3 Electron Beam Lithography
Page 13 of 17
TESCAN MIRA3 Electron Beam Lithography
Page 14 of 17
TESCAN MIRA3 Electron Beam Lithography
Page 15 of 17
TESCAN MIRA3 Electron Beam Lithography
Page 16 of 17
TESCAN MIRA3 Electron Beam Lithography
Page 17 of 17
Appendix B:
PMMA Spin Coating and Developing
PMMA is purchased from Micro Chem
Spin Coating:
Step 1: 500 rpm, 5 seconds, 100 rpm/s
Step 2: 3000 rpm, 45 seconds, 500 rpm/s
Soft baking: 185 C for 90 seconds
Thickness: 950 PMMA A2: ~ 80 nm ; 950 PMMA A4: ~ 240 nm
Developer: 1:3 MIBK:IPA (MIBK: 4-Methyl-2-pentanone; purchased from Sigma Aldrich)
Develop time: gently shaking for 60 seconds
Wash thoroughly with DI Water after develop
Dry with N2.