Pneumatech Medical Gas Solutions • Email: [email protected] • www.p-mgs.com• The right is reserved to change this technical specification without prior notice
Technical Datasheet81
0234
1150
Ed1
Pag
e 1
DescriptionThe Pneumatech MGS modular Oxyplant system shall provide medical quality oxygen into the hospital pipeline system, deliv-ered at a pressure of 400 kPa (4 bar) gauge.
The system shall use Atlas Copco compressors, dryer and filters, galvanised steel air receiver, PSA oxygen generator and steel oxygen receiver. The generator shall include a graphic LCD con-troller and oxygen sensor/analyser.
OperationAll components shall be self-controlled, no central controller is required. The compressor shall run to maintain pressure in the air receiver with a typical set pressure of 700 kPa (7 bar) gauge. The generator shall be run in a way to maintain pressure in the oxygen receiver at 400 kPa (4 bar) gauge, or up to a maximum pressure of 500 kPa (5 bar) gauge with an oxygen purity of 93%. The controller shall provide functionality to easily adjust the oxy-gen purity level up to 95%.
In cases using dual independent production lines (according to ISO 10083) the system will work with the following logic:
One generation line shall be determined as duty and the other shall in turn be set up as a back-up. It is recommended to switch between the lines once a week to ensure each generator is run. The system shall be pressure controlled. Each line has a differ-ent set pressure. The duty line has a start pressure of 430 kPa (4.3 bar) gauge and the reserve line at a fixed pressure of 400 kPa (4 bar) gauge. If the duty line is shut-down or fails the main line pressure will fall. When the reserve set pressure is reached (400 kPa, 4 bar) the reserve line will come on-line to maintain working line pressure.
Pressure shall be measured in the oxygen buffer tank. When the pressure depletes it reaches the start set point and the genera-tor starts producing oxygen. When the pressure stop set-point is reached the generator shall go in stand-by mode.
The oxygen generator shall automatically stop if the oxygen level falls below the set purity level (e.g. >92% as standard). In this case the reserve line will come on-line due to the pressure falling to 400 kPa (4 bar) gauge.
The compressors shall run on their own pressure band and shall operate independently to the oxygen generators. As soon as the pressure in the tank reaches it’s set point (e.g. 700 kPa, 7 bar) the compressor shall go into unload/stop mode.
In the case of generator failure a general alarm fault shall be trig-gered at 91% (adjustable). If the oxygen quality falls lower than 90% the generator shall stop and the reserve generator will start production of oxygen due to depletion of line pressure to 400 kPa (4 bar) gauge.
Oxygen Generator Systems
Oxygen GeneratorThe oxyplant generator shall incorporate pressure swing absorp-tion (PSA) technology and supplied as a plug and play compo-nent for simple installation on site, including oxygen monitoring, service indications and relevant alarms.
Non return valves shall be included to avoid back flow from the outlet of the generator. The vessels shall include chemically produced zeolite to adsorb specific types of molecules, such as water vapour or nitrogen.
The oxyplant generator shall include an oxygen sensor and analyser for continuous, online measurement of oxygen purity and shall be provided with a calibration report delivered with every unit. The sensor shall be insulated by solenoid valves during standstill to provide longer sensor lifetime. An inlet pres-sure regulator shall be included to reduce the inlet pressure to a maximum working pressure of 650 kPa (6.5 bar ) gauge.
The oxygen purity produced shall be between 90-95% with a compressed air inlet quality in accordance with ISO 8573-1 clause 1-4-1 and an air outlet quality in accordance with ISO 8573-1 clause 1-2-1.
Control SystemThe oxyplant T4 control system shall provide an intelligent human machine interface including a real-time clock for recording op-erational parameters in the event log. The central control system shall include a BMS connection for a general fault condition.
Pneumatech Medical Gas Solutions • Email: [email protected] • www.p-mgs.com• The right is reserved to change this technical specification without prior notice
Technical Datasheet81
0234
1150
Ed1
Pag
e 2
Specifications• Supply voltage: 115-230V 50/60Hz (oxygen generator)• Power consumption: 200W max (oxygen generator)• CSA/UL approval available• Protection degree: IP54 (Nema 3S)• Operation at a maximum ambient temperature of 45°C. • Compressor 750 kPa (7.5 bar) gauge including EFF1 (CE MEP) rated TEFC, IP55 class F electric motors. Incorporating maintenance-free greased for life bearings. • Dryer 3°C PDP at 45°C• Air receivers DIRECTIVE 97/23/EC, supplied with relevant test certificates. • Oxygen vessel of steel construction and approved according to Fluid Group 1• PSA oxygen modules are able to produce oxygen compliant with the United States Pharmacopeia (USP) monograph for Oxygen 93 or the European Pharmacopeia (EurPh) mono graph for Oxygen 93. Both standards were created expressly to permit the use of PSA produced oxygen.
ClassificationOxyplant system is designed to ISO10083 standards and air quality to the European Pharmacopoeia monograph.
Features• Pre-piped, factory tested and certified• 91%, 93% or 95% oxygen purity• Zeolite providing infinite lifetime (if the dryers and filters are properly serviced) • Valves degreased for oxygen use are provided with an oxygen certificate, delivered with every unit• Pressure controlled – no central controller for simple operation• Silencer on exhaust to reduce noise levels• Touch screen controller on generator with 4” high-definition display• Zirconia oxygen sensor included on generator for long life and low maintenance• Atlas Copco’s premium GA oil flooded rotary screw compressor• Atlas Copco’s FX refrigerated dryer module• Atlas Copco DD+ and PD+ high efficiency filters• Accessories
Options• Advanced T5 controller including E-mail and SMS event notification for additional convenience. Graphical log for analogue values and user friendly touch screen 6” high-definition colour display
• High pressure booster System using oil free, high pressure pistons compressors shall be provided suitable for the on site filling of 15,000 kPa (150 bar) gauge or 20,000 kPa (200 bar) gauge oxygen cylinders. Compressors shall operate on an inlet pressure auto-cycling system (auto start/stop).
• Filling Ramp for on site cylinder filling consisting of a high pressure bar with individual isolation valves at each cylinder connection. The cylinders shall be protected from over pressurising by a safety pressure relief valve and shall include a pressure gauge. The filling ramp shall be tested up to 30,000 kPa (300 bar) gauge and shall be supplied complete with test certificate (cylinders are NOT included).
• Purge Control. An additional purge line to ensure that only the appropriate level of oxygen quality is supplied to the hospital distribution pipeline.
• Dew Point and Flow Monitoring. The dryer shall be able to incorporate a dew point sensor with an accuracy of ±10C in the range -20 to -800C atmospheric dew point and 4-20mA analogue output. The PDP sensor shall ensure the air to the generator inlet is suitable in order to safeguard the generator from high moisture content.
Pneumatech Medical Gas Solutions • Email: [email protected] • www.p-mgs.com• The right is reserved to change this technical specification without prior notice
Technical Datasheet81
0234
1150
Ed1
Pag
e 3
Typical Layout - Schematic
BA
LL
V
AL
VE
BA
LL
VA
LV
E
AIR TANK 1
PR
ES
SU
RE
R
EG
UL
AT
OR Du
st
FIL
TE
R
WA
TE
R D
RA
IN
BA
LL
VA
LV
E
PR
ES
SU
RE
S
AF
ET
Y V
AL
VE
OX
YG
EN
GE
NE
RA
TO
R
BA
LL
VA
LV
E
PR
ES
SU
RE
S
AF
ET
Y V
AL
VE
BA
LL
VA
LV
EN
ON
-RE
TU
RN
V
AL
VE
BA
LL
V
AL
VE
BA
LL
VA
LV
E
AIR TANK 2
PR
ES
SU
RE
R
EG
UL
AT
OR D
ust
FIL
TE
R
WA
TE
R D
RA
IN
BA
LL
VA
LV
E
PR
ES
SU
RE
S
AF
ET
Y V
AL
VE
BA
LL
VA
LV
E
PR
ES
SU
RE
S
AF
ET
Y V
AL
VE
BA
LL
VA
LV
EN
ON
-RE
TU
RN
V
AL
VE
Fill
ing
ra
mp
INL
ET
-
DD
FIL
TE
R
INL
ET
-
PD
FIL
TE
RQ
D
AC
10
1
SR
10
1S
R1
02
SR
10
3
AD
10
1
BA
10
1
TK
10
1
BA
10
3
BA
10
2
SA
10
1
OG
10
1
BA
10
4
TK
10
2
SA
10
2
BA
10
5
O
XY
GE
N
BU
FF
ER
TA
NK
RV
10
1
SR
10
4N
R1
01
SR
20
4N
R2
01
RV
20
1
TK
20
2
BA
20
5
O
XY
GE
N
BU
FF
ER
TA
NK
BA
20
4
BA
20
3
BA
20
2
BA
20
1TK
20
1
SA
20
1S
A2
02
FR
10
1
B0
30
1
HIG
H
PR
ES
SU
RE
No
t in
clu
de
d in
sco
pe
of su
pp
ly
A/A
HE
AT
EX
CH
AN
GE
R
WA
TE
R
SE
PE
RA
TO
R
REFRIGERANT
SEPERATOR
R/A
HE
AT
EX
CH
AN
GE
R
RE
FR
IGE
RA
NT
CO
ND
EN
SO
R
RE
FR
IGE
RA
NT
CO
MP
RE
SS
OR
WA
TE
R
DR
AIN
FX
DR
YE
R
G
A
CO
MP
RE
SS
OR
WD
10
1
WD
20
1
INL
ET
-
DD
FIL
TE
R
INL
ET
-
PD
FIL
TE
R
SR
20
1S
R2
02
SR
20
3
AD
20
1
A/A
HE
AT
EX
CH
AN
GE
R
WA
TE
R
SE
PE
RA
TO
R
REFRIGERANT
SEPERATOR
R/A
HE
AT
EX
CH
AN
GE
R
RE
FR
IGE
RA
NT
CO
ND
EN
SO
R
RE
FR
IGE
RA
NT
CO
MP
RE
SS
OR
WA
TE
R
DR
AIN
FX
DR
YE
R
CO
MP
RE
SS
OR
INL
ET
FIL
TE
R
AIR
IN
TA
KE
TE
MP
ER
AT
UR
ET
RA
NS
DU
CE
R
AF
TE
R
CO
OL
ER
P
TE
MP
ER
AT
UR
E
TR
AN
SD
UC
ER
PR
ES
SU
RE
TR
AN
SD
UC
ER
NO
N-R
ET
UR
N
VA
LV
E
PR
ES
SU
RE
S
AF
ET
Y V
AL
VE
AU
TO
MA
TIC
D
RA
IN
BA
LL
VA
LV
ES
FL
EX
IBL
E
CO
NN
EC
TIO
N
TO
WE
RT
OW
ER
A
B
OX
YG
EN
GE
NE
RA
TO
R
OG
20
1
TO
WE
RT
OW
ER
A
B
AC
20
1
CO
MP
RE
SS
OR
INL
ET
FIL
TE
R
AIR
IN
TA
KE
TE
MP
ER
AT
UR
ET
RA
NS
DU
CE
R
AF
TE
R
CO
OL
ER
P
TE
MP
ER
AT
UR
E
TR
AN
SD
UC
ER
PR
ES
SU
RE
TR
AN
SD
UC
ER
NO
N-R
ET
UR
N
VA
LV
E
PR
ES
SU
RE
S
AF
ET
Y V
AL
VE
AU
TO
MA
TIC
D
RA
IN
BA
LL
VA
LV
ES
FL
EX
IBL
E
CO
NN
EC
TIO
N
G
A
CO
MP
RE
SS
OR
BA
LL
VA
LV
E
BA
LL
VA
LV
E
AIR
IN
LE
T
OX
YG
EN
OU
TL
ET
AIR
IN
LE
T
OX
YG
EN
OU
TL
ET
QD
(*)
Not applic
able
when o
nly
one o
xygen g
enera
tor
is insta
lled
BA106
BA
107
BA
207
BA206
Manifold
(*
)
BA310
T
O H
OS
PIT
AL
F
EE
D
NO
N-R
ET
UR
N
VA
LV
E (
*)
GA
10
1G
A1
02
GA
20
2G
A2
01
1 2
No
te:
Oxygen g
enera
tor
options c
an b
e insta
lled
See d
edic
ate
d flo
w d
iagra
m o
f genera
tor
Not all
deta
ils for
main
com
ponents
( c
om
pre
ssor,
genera
tor,
dry
er
)show
n. S
ee d
eta
ils for
indiv
idual com
ponents
*
Optional equip
ment
RV
30
1
AIR TANK 3
TK
30
1
BA
LL
V
AL
VE
BA
30
1
p
p
Pu
rge
fu
nctio
n(
op
tio
n )
p
p
Pu
rge
fu
nctio
n(
op
tio
n )
Pneumatech Medical Gas Solutions • Email: [email protected] • www.p-mgs.com• The right is reserved to change this technical specification without prior notice
Technical Datasheet81
0234
1150
Ed1
Pag
e 4
Typical Simplex Installation Typical Duplex Installation
1 Compressor 6 Purity measurement line
2 Refrigerant dryer 7 Ball valve
3 Filters DD-PD-QD 8 Oxygen tank
4 Air receiver 9 Pressure regulator
5 Oxygen generator 10 Dust filter
1 Compressor 6 Purity measurement line
2 Refrigerant dryer 7 Ball valve
3 Filters DD-PD-QD 8 Oxygen tank
4 Air receiver 9 Pressure regulator
5 Oxygen generator 10 Dust filter
11 Non-return valve
Pneumatech Medical Gas Solutions • Email: [email protected] • www.p-mgs.com• The right is reserved to change this technical specification without prior notice
Technical Datasheet81
0234
1150
Ed1
Pag
e 5
Typical Duplex Installation With Booster And Filling Ramp
1 Compressor 8 Oxygen tank
2 Refrigerant dryer 9 Pressure regulator
3 Filters DD-PD-QD 10 Dust filter
4 Air receiver 11 Non-return valve
5 Oxygen generator 12 High pressure booster
6 Purity measurement line 13 Pressure switch
7 Ball valve 14 Filling station
Pneumatech Medical Gas Solutions • Email: [email protected] • www.p-mgs.com• The right is reserved to change this technical specification without prior notice
Technical Datasheet81
0234
1150
Ed1
Pag
e 6
Typical Simplex Containerised Installation With Booster And Filling Ramp
OXYGEN FILLING STATION WITH 6 OUTLETS113 T5 CONTROL BOX112 FUSE BOX111 RIX 4V4 AL BOOSTER110 PARKER TGA108 FILTER19 FILTER REGULATOR18 OXYGEN RECEIVER17 EXHAUST PIPE (SILENCER)16 OGP35 GENERATOR15 AIR RECEIVER14 DD-PD-QD130+ FILTER SET 13 FX17 DRYER12 GA30+ COMPRESSOR11
CommentsNameQtyPos
Drawn by Reviewed by Approved by - date Scale
VersionDrawing no.
Title
Paper size
Gazcon A/SOldvej 1A - DK 3540 LyngePhone5 44 99 69 00Fax +45 44 99 69 03 A3
SS
CONTAINERIZED O35 OXYGEN GENERATOR
1:20
Pos Amount Description Weight Dimensions (LxWxH)
1 2 3 4 5 6 7 8 9
A
10
B
C
D
E
F
G
agreed by us this document
may constitute a contractual
obligation on our part.
Only to the extent expressly
transmitted or communicated to any other party
property and must not be copied, reproduced,
or used for purposes not expressly perm
itted by us.
This document and its contents are our exclusive
Note:All measurements to be verified at time of installation.
1
2
3
4
5
7
8
9
10
11
6
12
2438
2352
6058
863
863
5897
24217644941023650
8018503501420500
276
1800
800
780
1020 96
893
R1176
2730
2630
2430
2172
1560
910
1800
2350
457
37
450
50
500
50
250
13
1 Compressor 8 Filter Regulator
2 Refrigerant dryer 9 TGA108 Filter
3 Filters DD-PD-QD 10 High pressure booster
4 Air receiver 11 Fuse Box
5 Oxygen generator 12 Controller
6 Exhaust Pipe (Silencer)
13 Oxygen Filling Station
7 Oxygen tank
OXYGEN FILLING STATION WITH 6 OUTLETS113 T5 CONTROL BOX112 FUSE BOX111 RIX 4V4 AL BOOSTER110 PARKER TGA108 FILTER19 FILTER REGULATOR18 OXYGEN RECEIVER17 EXHAUST PIPE (SILENCER)16 OGP35 GENERATOR15 AIR RECEIVER14 DD-PD-QD130+ FILTER SET 13 FX17 DRYER12 GA30+ COMPRESSOR11
CommentsNameQtyPos
Drawn by Reviewed by Approved by - date Scale
VersionDrawing no.
Title
Paper size
Gazcon A/SOldvej 1A - DK 3540 LyngePhone5 44 99 69 00Fax +45 44 99 69 03 A3
SS
CONTAINERIZED O35 OXYGEN GENERATOR
1:20
Pos Amount Description Weight Dimensions (LxWxH)
1 2 3 4 5 6 7 8 9
A
10
B
C
D
E
F
G
agreed by us this document
may constitute a contractual
obligation on our part.
Only to the extent expressly
transmitted or communicated to any other party
property and must not be copied, reproduced,
or used for purposes not expressly perm
itted by us.
This document and its contents are our exclusive
Note:All measurements to be verified at time of installation.
1
2
3
4
5
7
8
9
10
11
6
12
2438
2352
6058
863
863
5897
24217644941023650
8018503501420500
276
1800
800
780
1020 96
893
R1176
2730
2630
2430
2172
1560
910
1800
2350
457
37
450
50
500
50
250
13
OXYGEN FILLING STATION WITH 6 OUTLETS113 T5 CONTROL BOX112 FUSE BOX111 RIX 4V4 AL BOOSTER110 PARKER TGA108 FILTER19 FILTER REGULATOR18 OXYGEN RECEIVER17 EXHAUST PIPE (SILENCER)16 OGP35 GENERATOR15 AIR RECEIVER14 DD-PD-QD130+ FILTER SET 13 FX17 DRYER12 GA30+ COMPRESSOR11
CommentsNameQtyPos
Drawn by Reviewed by Approved by - date Scale
VersionDrawing no.
Title
Paper size
Gazcon A/SOldvej 1A - DK 3540 LyngePhone5 44 99 69 00Fax +45 44 99 69 03 A3
SS
CONTAINERIZED O35 OXYGEN GENERATOR
1:20
Pos Amount Description Weight Dimensions (LxWxH)
1 2 3 4 5 6 7 8 9
A
10
B
C
D
E
F
G
agreed by us this document
may constitute a contractual
obligation on our part.
Only to the extent expressly
transmitted or communicated to any other party
property and must not be copied, reproduced,
or used for purposes not expressly perm
itted by us.
This document and its contents are our exclusive
Note:All measurements to be verified at time of installation.
1
2
3
4
5
7
8
9
10
11
6
12
2438
2352
6058
863
863
5897
24217644941023650
8018503501420500
276
1800
800
780
1020 96
893
R1176
2730
2630
2430
2172
1560
910
1800
2350
457
37
450
50
500
50
250
13