Date post: | 22-Nov-2015 |
Category: |
Documents |
Upload: | rrheman-kumar |
View: | 15 times |
Download: | 6 times |
11
Rob LeindersSlides: Tim Jongen
Medtronic Bakken Research Center
Bioelectronica en Nanotechnologie
2008 - 2012
Module: Pacemaker Basics
2Recommended Literature Foundation of Cardiac Pacing. Sutton, Bourgeois, 1991, ISBN 0-
87993-337-2
Design of Cardiac Pacemakers. Webster 1993, ISBN 0-7803-1134-5
Rapid Interpretation of EKGs. Dubin 1970, SBN 0-912912-00-6 The evolution of pacemakers: An electronics perspective, from the
hand crank to advanced wavelet analysis. Haddad AP, Houben RP, Serdijn WA. IEEE Eng Med Biol Mag. 2006 May-Jun; 25(3):38-48
3Pacemaker Basics EGM / ECG
Electrophysiology Measuring Signals 12 Lead ECG
Diseases Bradycardia Tachycardia
Pacemaker: Basics Components Overview Simplified Pacemaker Output Pulse: Width / Amplitude / Capacitor
Sensing Why Sensing Oversensing Undersensing Sensing Hardware
Frequency spectrum Filters Analog to Digital Conversion
Pacemaker: Current Devices ICDs Leads
Unipolar / Bipolar Low voltage / High voltage Active / Passive Fixation Lead Design Lead Implantation
24Practicum
Building a Pacemaker
Program a Pacemaker Simulations with Patient Simulator
11EGM / ECG Measuring Signals
ECG
EGM
2
3Cardiac Conduction - Sinus Node
The Hearts Natural Pacemaker Rate of 60-100 bpm at rest
Sinus Node(SA Node)
24Cardiac Conduction - AV Node
Receives impulses from SA node Delivers impulses to the His-
Purkinje System Delivers rates between 40-60
bpm if SA node fails to deliver impulses
Atrioventricular Node (AV Node)
5Cardiac Conduction - HIS Bundle
Begins conduction to the ventricles
AV Junctional Tissue: Rates between 40-60 bpm
Bundle of His
6Cardiac Conduction - Purkinje Fibers
Bundle Branches and Purkinje Fibers
Moves the impulse through the ventricles
Provides Escape Rhythm: 20-40 bpm
Purkinje Network
37Normal Sinus Rhythm
8Impulse Formation in SA Node
9Atrial Depolarization
410Delay at AV Node
11Conduction through Bundle Branches
12Conduction through Purkinje Fibers
513Ventricular Depolarization
14Plateau Phase of Repolarization
15Final Rapid (Phase 3) Repolarization
616AutomaticityCardiac Cells are unique
because they spontaneously depolarize
Upper (SA Node) 60-80 bpm
Middle (AV Junction) 40-60 bpm
Lower (Purkinje Network) 20-40 bpm
17Measuring Signals
18ECG
719ECG
20EGM / ECG
21ECG
Precordial Leads: V1 V2 V3 V4 V5 V6
82212 Lead ECG
23EGM Electrogram
Measure locally in Atria or Ventricles
Lead II and RV electrogram Lead II and HRA electrogram
11Diseases Bradycardia
Tachycardia
2Bradycardia
Patient has a slow (Brady: slowness) heart rhythm: Sinus Arrest Brady/Tachy Syndrome Sinus Bradycardia
Exit Block Bi/Trifascicular Block AV Block
3Sinus Arrest
Failure of sinus node discharge Absence of atrial depolarization Periods of ventricular asystole May be episodic as in vaso-vagal syncope, or carotid sinus hypersensitivity
May require a pacemaker
24Sinus Bradycardia Sinus Node depolarizes very slowly If the patient is symptomatic and the rhythm is persistent and irreversible, may
require a pacemaker
5Brady/Tachy Syndrome Intermittent episodes of slow and fast rates from the SA node or atria Brady < 60 bpm Tachy > 100 bpm AKA: Sinus Node Disease
Patient may also have periods of AF and chronotropic incompetence 75-80% of pacemakers implanted for this diagnosis
6Exit Block Transient block of impulses from the SA node
Sinus Wenckebach is possible, but rare Pacing is rare unless symptomatic, irreversible, and persistent
37First-Degree AV Block Standard PR interval 120 200 ms
Delayed conduction through the AV
In most cases no symptoms. No medications or PM therapy is needed. Correcting electrolyte imbalances is standard therapy
8Second-Degree AV Block
Type 1: Wenkebach prolongation of PR interval Type 2: no prolongation
9Third-Degree AV Block No impulse conduction from the atria to the
ventricles Atrial rate = 130 bpm, Ventricular rate = 37 bpm Complete A V disassociation Usually a wide QRS as ventricular rate is
idioventricular
410Fascicular Block
Right bundle branch block and left anterior hemiblock
Right bundle branch block and left posterior hemiblock
Complete left bundle branch block
11Trifascicular Block Complete block in the right bundle branch, and Complete or incomplete block in both divisions of
the left bundle branch Identified by EP Study
12Chronotopic Incompetence
CO = SV x HR Healthy hearts are able to increase peak CO by
up to 5x baseline with exercise Chronotopic Incompotence patients are only able
to increase CO x2 over baseline
513Tachycardia
Sinus Tachycardia
Premature Contractions Atrial Tachycardia
Accelerated Junctional Rhythm Accelerated Idioventricular Rhythm (AIVR)
Atrial Flutter Atrial Fibrillation AVRT/AVNRT Ventricular Tachycardia Ventricular Fibrillation
14Sinus Tachycardia
Origin: Sinus Node Rate: 100-180 bpm Mechanism: Abnormal or Hyper Automaticity (for example, exercise)
15Atrial Tachycardia
Origin: Atrium - Ectopic Focus
Rate: >100 bpm
Mechanism: Abnormal Automaticity
616Premature Beats - PAC
Origin: Atrium (outside the Sinus Node) Mechanism: Abnormal Automaticity
Characteristics: An abnormal P-wave occurring earlier than expected, followed by compensatory pause
17
Origin: AV Node Junction
Mechanism: Abnormal Automaticity
Characteristics: A normally conducted complex with an absent P-wave, followed by a compensatory pause
Premature Beats - PJC
18
Origin: Ventricles
Mechanism: Abnormal Automaticity Characteristics: A broad complex occurring earlier than expected,
followed by a compensatory pause
Premature Beats - PVC
719PVC Patterns
Bigeminy- Every other beat
Trigeminy- Every third beat
Quadrigeminy- Every fourth beat
20
Origin: Varies within the Ventricle
Mechanism: Abnormal Automaticity Characteristics: Each premature beat changes axis; implies a different focus
of origin for each beat Note: PVCs by themselves are not a predictor of VT/VF, nor do they imply
the need for a defibrillator
Multifocal PVC
21Accelerated Junctional Rhythm
Origin: AV Node or Junctional Tissue Mechanism: Abnormal Automaticity Characteristics: Occurs when AV nodal cells depolarize at a rate
faster than the sinus node
822Accelerated Idioventricular Rhythm
Origin: Ventricle Mechanism: Abnormal Automaticity Rate: Ventricular rate >sinus rate, but
925Reentry
26Reentry
27Atrial Flutter
Origin: Right and Left Atrium
Mechanism: Reentry, circus tachycardia
Rate: 250 400 bpm
Characteristics: Rapid, regular P-waves, regular R-waves
10
28
Origin: Right and/or left atrium, pulmonary veins Mechanism: Multiple wavelets of reentry Atrial Rate: > 400 bpm Characteristics: Random, chaotic rhythm; associated with irreg. ventricular rhythm
Atrial Fibrillation (AF)
29Atrial Fibrillation (AF)
30AF Mechanism
Paroxysmal: Sudden onset and spontaneous cessation Persistent: Requires intervention to terminate, usually > 24-48
hour duration Permanent or Chronic: Unable to terminate AF begets AF
The more frequent the AF the more frequently it will re-occur and episodes tend to last longer
11
31
Mutifocal Atrial Tachycardia Mechanism: Abnormal Automaticity (multi-
sites) Characteristics: Many depolarization waves;
activation occurs asynchronouslyNot commonly used terms anymore, usually
just called AF
Other AF Mechanisms
Single Foci Mechanism: Abnormal Automaticity (single-
focus, usually in the Posterior Left Atrium) Characteristics: Rapid discharge; single
ectopic site Parasystole rare
32Atrial Flutter vs. Atrial FibrillationSummary of Disease Characteristics
Atrial Flutter Atrial Fibrillation
Atrial Rate 250 to 400 bpm 400 bpm
Ventricular Rate/Rhythm Usually regular Varies with conductionGrossly irregular
Pattern Saw tooth baseline Irregular or almost flat baselineIrregularly irregular
Underlying Mechanism Reentry via macro re-entrant circuit
Typically multiple wavelet reentry
33AVRT - AV Re-entrant TachycardiaAn SVT caused by the existence of an extra
pathway from the atria to the ventricles Extra pathway + AV Node = reentry
Two Types Orthodromic Antidromic
12
34
Orthodromic Mechanism: Reentry
Rate: 180 - 260 bpm+
Characteristics: Extra electrical pathway to ventricles
AVRT
Accessory PathwayConduction to the ventricles via the AV node (normal conduction) - then from Ventricles to the Atria via the accessory pathway. Produces narrow complex SVT.
35
Antidromic Mechanism: Reentry
Rate: 180 - 260 bpm+
Characteristics: Extra electrical pathway to ventricles. Wide-complex QRS.
AVRT
Accessory PathwayConduction to ventricles via the accessory pathway. The impulse is then conducted retrograde to atrial via the AV node. Produces a wide-complex SVT.
36
Origin: A - V conduction outside the AV node (bundle of Kent). The Wolff pathway conducts faster than the AV node
Characteristics: Short PR Interval (< 120 ms), wide QRS (> 110 ms), delta wave Sudden cardiac death in these individuals is due to the propagation of an atrial arrhythmia to the ventricles at a very high rate.
Wolff-Parkinson-White
13
37
Origin: AV Node Mechanism: Reentry Rate: 150 - 230 bpm, faster in teenagers Characteristics: Normal QRS with absent P-waves
AVNRT - AV Node Re-entrant Tachycardia
38AVRT vs. AVNRT AVRT
180 260 bpm Narrow QRS if
orthodromic Wide QRS if antidromic Delta wave + in SR PR < 120 msec 1:1 Conduction
AVNRT 150 230 bpm Narrow QRS Short RP No delta waves Initiating PR long P-waves buried in QRS Conduction 1:1, or 2:1
when distal block presentTreatment: Ablation Rarely is a ICD implanted
39Monomorphic VT Origin: Ventricles (Single Focus) Mechanism: Reentry initiated by abnormal automaticity
or triggered activity Characteristics: Rapid, wide and regular QRS. A-V
disassociation
14
40
Origin: Ventricles (Wandering Single Focus) Mechanism: Reentry with movement in the circuit initiated by
abnormal automaticity or triggered activity Characteristics: Wide and irregular QRS Complex that changes in axis
Polymorphic VT
41
Origin: Ventricle Mechanism: Reentry (movement in focus) Rate: 200 250 bpm Characteristics: Associated with Long QT interval; QRS changes axis
and morphology with alternating positive/negative complexes
Torsades de Pointes - Twisting of the points
42
Origin: Ventricle Mechanism: Multiple wavelets of reentry Characteristics: Irregular with no discrete QRS
Ventricular Fibrillation (VF)
11Pacemaker Basics Components Overview
Simplified Pacemaker
Output Pulse Capacitor Width Amplitude Strength Duration Curve+
2Pacemaker Overview Pulse Generator
Battery, sensing, pulse formation Control (counter), telemetry
Wires (Leads)
3Pacemaker Principle of Operation
24Pacemaker Output
Output pulse Anodal/Cathodal Pulse width Amplitude
5Pacemaker Output
Delivered energy: Strength duration curve
6Pacemaker Output
11Sensing Why Sensing Noise Oversensing Undersensing Sensing Hardware
Frequency spectrum Filters Analog to Digital Conversion
2Sensing Sensing is the ability of the pacemaker to see
when a natural (intrinsic) depolarization is occurring Pacemakers sense cardiac depolarization by
measuring changes in electrical potential of myocardial cells between the anode and cathode
3EGM / ECG
Intrinsic deflection on an EGM occurs when a depolarization wave passes directly under the electrodes
Two characteristics of the EGM are: Signal amplitude Slew rate
24Slew Rate - Change in Voltage with Respect to the Change in Time
The longer the signal takes to move from peak to peak: The lower the slew rate The flatter the signal
Higher slew rates (number in mV) translate to greater sensing Measured in volts per second Vo
ltage
Time
Slope
Slew rate= Change in voltageTime duration ofvoltage change
5Sense and Respond to Cardiac Rhythms Accurate sensing enables the pacemaker to
determine whether or not the heart has created a beat on its own
The pacemaker is usually programmed to respond with a pacing impulse only when the heart fails to produce an intrinsic beat
6Accurate Sensing...
Ensures that undersensing will not occur
Ensures that oversensing will not occur
Provides for proper timing of the pacing pulse
37Undersensing . . . Pacemaker does not see the intrinsic beat, and
therefore does not respond appropriately
Intrinsic beat not sensed
Scheduled pace delivered
VVI / 60
8Oversensing
An electrical signal other than the intended P or R wave is detected
Marker channel shows intrinsic
activity...
...though no activity is present
VVI / 60
9Sensitivity
The Greater the Number, the Less Sensitive the Device to Intracardiac Events
410SensitivityAm
plitu
de (m
V)
Time
5.0
2.5
1.25
11Sensitivity
Ampl
itude
(mV)
Time
5.0
2.5
1.25
12Sensitivity
Ampl
itude
(mV)
Time
5.0
2.5
1.25
513Noise
Sensing amplifiers use filters that allow appropriate sensing of P waves and R waves and reject inappropriate signals
Unwanted signals most commonly sensed are: T waves Far-field events (R waves sensed by the atrial channel) Skeletal myopotentials (e.g., pectoral muscle myopotentials)
14Noise
Electrophysiological signals and interferenceMyopotentials in unipolar EGM resulting from pectoral muscles=> Intermittent, baseline variations
Electromagnetic interference (main 50/60 Hz) picked up by electrode tip- IPG can loop=> Superimposed variations
RV current of injury, (artificial) ST-segment shift. Disappears after approximately 30 min after lead implant
15Accurate Sensing is Dependent on . . . The electrophysiological properties of the
myocardium The characteristics of the electrode and its
placement within the heart The sensing amplifiers of the pacemaker
616Factors That May Affect Sensing Are: Lead polarity (unipolar vs. bipolar) Lead integrity
Insulation break Wire fracture
EMI Electromagnetic Interference
17Unipolar Sensing
Produces a large potential difference due to: A cathode and anode
that are further apart than in a bipolar system
_
18Bipolar Sensing
Produces a smaller potential difference due to the short interelectrode distance Electrical signals from
outside the heart such as myopotentials are less likely to be sensed
719An Insulation Break
May Cause Both Undersensing or Oversensing Undersensing occurs when inner and outer conductor
coils are in continuous contact Signals from intrinsic beats are reduced at the sense
amplifier and amplitude no longer meets the programmed sensing value
Oversensing occurs when inner and outer conductor coils make intermittent contact Signals are incorrectly interpreted as P or R waves
20A Wire Fracture
Can Cause Both Undersensing and Oversensing
Undersensing occurs when the cardiac signal is unable to get back to the pacemaker intrinsic signals cannot cross the wire fracture
Oversensing occurs when the severed ends of the wire intermittently make contact, which creates potentials interpreted by the pacemaker as P or R waves
21Sensing Hardware
11
Pacemakers
2Pacemaker Parts
PacemakerCan
Connector Block
Leads
SensingCircuit
OutputStage
Battery
BatteryEOL circuit
TelemetrySystem
Antenna
Sensor(s)
SensorInterfaceFeed-throughs
Bonding wires
Mechanical Electronics
Timing/ControlCircuits
TelemetryProtocol SW
SoftwareAlgorithms
Memory
Microprocessor
AD Converter
Software Other
Hybrid
3
Telemetry
24Pacemaker Telemetry
Pacemaker TelemetryVoltage
regulators Transceiver
DecodingCoding
Logic control unitTimebase
Telemetry front-end
Coupled coils175 KHz AM, PModulatedMicroWatts power uplink
5RF Telemetry products
TEL A/B/V, Used by CRDM
TEL N, used by Neuro
Range: Less then 50 cm
RF freq: 175 kHz
6RF Telemetry products -2
TEL C Conexus currently in use by CRDM
Later also by Neuro
Range: Up to 10 meters
RF band: 402 - 405 MHzMICS: Medical Information Communication Service
37
RAMware
8RAMwhere?
PacemakerCan
Connector Block
Leads
SensingCircuit
OutputStage
Battery
BatteryEOL circuit
TelemetrySystem
Antenna
Sensor(s)
SensorInterfaceFeed-throughs
Bonding wires
Mechanical Electronics
Timing/ControlCircuits
TelemetryProtocol SW
SoftwareAlgorithms
Memory
Microprocessor
AD Converter
Firmware Other
Hybrid
9Device Memory
ROM(Read Only Memory)
EEPROM(Electrically Erasable Programmable ROM)
RAM(Random Access Memory)
Firmware
Programmed Parameters RAMware
Working data Diagnostic data RAMware
410RAMware download
(1) RAMware download(2) call RAMware
initializationfunction
&
RAMware Media
11
Algorithms
12Algorithms Rate-Responsiveness
Based on activity (also possible on QT interval) Day and night difference
Managed Ventricular Pacing If not needed, pacing should be avoided Switch between DDD(R) and AAI(R)
Mode Switch What happens during AF when PM is programmed in DDD
mode Switch between DDD(R) and DDIR
513Algorithms (cont)
Additional chamber: LV pacing Rate dependant AV interval Rate-drop response PVC response
11
ICD
2ICD
Implantable Cardioverter Defibrillator
Whats the function Sense Detect Therapy Pace
3
CPU/Memory The processor chip analyses electrical signals from the heart and determines whether any shocks are necessary. The chip runs at less than 100 kHz
ICD ComponentsICD Components
GEM DR Dual Chamber, Rate Responsive ImplantabDefibrillator
BatteryThe special batteries, made of lithium silver vanadium oxide, last six years or more, even though they have less energy than a standard laptop battery
TransformerFor serious heart disturbances requiring a higher shock, a transformer converts low-voltage battery power into a higher voltage.
CasingThe I.C.D. is encased in titanium. Scar tissue grows around it, locking it in place. The device is sealed shut to prevent leakage, so when the battery dies, the entire device must be replaced
ConnectorsA lead with three connections provides sensing capabilities and a pathway for high-energy shocks to one of the hearts lower chambers. A second lead with only one connection provides sensing and pacing to an upper chamber.
BeeperA warning beeper indicates a low battery or another problem that a doctor should check out. There are no warnings to signal a large shock to the patient.
CapacitorsSimilar to the ones used in camera flashes, capacitors can take up to 10 seconds to draw enough energy (30-40 J) from the battery for a large shock. Charge times for smaller shocks are much shorter.
AntennaCommunication between the I.C.D. and the programmer is made through low-frequency radio waves sent from the unit's antenna to a doughnut-shaped receiver held over the patient's chest
LeadsFlexible wires convey sensing information to the I.C.D. and carry electric shocks to the heart. A patient may receive one or two leads depending on the nature of the heart problem. They are covered with silicone insulation.
Marquis DR Implantable Defibrillator
24ICD Specs
Combined with: Pacemaker Cardiac Resynchronization
Longevity 7-10y
37 cc / 68 g
5ICD Sensing
6ICD Detection
Detection of VT/VF based on: Rate Duration (Morphology)
37ICD Therapy
8ICD Settings
9ICD - Shock
Shocking: 35 J charged in 8 s Defibrillation (VF) Cardioversion (VT)
Shock Delivery waveform
410ICD Shock (cont)
11ATP
11
Pacing Leads
2Pacing Lead Lifetime Activity 70 bpm 100,000 beats/day 37,000,000 beats/year 500,000,000 beats/13.5 years
3
Conductor Tip Electrode Insulation Connector Pin
Pacing Lead Components Conductor Connector Pin Insulation Electrode Lead Assembly
24Conductor Purpose
Deliver electrical impulses from IPG to electrode Return sensed intracardiac signals to IPG
Conductor
5Conductor -- Types Types
Unifilar
Multifilar
Cable
6Conductor -- Construction Unipolar Construction
37Conductor -- Unipolar Construction Unipolar lead
1 pacing conductor IPG case (can)
for sensing
8Conductor -- Unipolar Construction Unipolar Lead Characteristics
Larger pacing spikes on EKG Small diameter lead body Less rigid lead body More susceptible to oversensing May produce muscle and nerve stimulation
9Conductor -- Construction Bipolar Construction
Co-axial
Co-radial
Outer insulation
Tip electrode coilIndifferent electrode
coilIntegral insulation
Tip electrode coil
Indifferent electrodecoil
410Conductor -- Construction Bipolar Construction
Parallel Coils
Coil/Cables
11Conductor -- Bipolar Construction Bipolar Lead Characteristics
Larger diameter lead body Tend to be stiffer Less susceptible to oversensing Unipolar programmable Less likely to produce muscle and nerve stimulation
12Conductor -- Material Typical Conductor Materials
MP35N (nickel alloy) MP35N silver cored
513Connector Purpose
Connects lead to IPG, and provides a conduit to: Deliver current from IPG to lead Return sensed cardiac signals to IPG
Connector
14Connector -- IS-1 Standard IS-1 Standard Connectors
Sizes Prior to IS-1 Standard 3.2 mm low-profile connectors 5/6 mm connectors
15Insulation Purpose:
Contain electrical current Prevent corrosion
Insulation
616Insulation -- Properties Properties of Insulation Materials
Tensile strength Elongation Tear strength Abrasion Compression set Crush (cyclic compression) Creep
17Insulation -- Type Insulation Types
Silicone Polyurethane Fluoropolymers (PTFE, ETFE)
18Insulation -- Type Silicone
Advantages Inert Biocompatible Biostable
Disadvantages High friction coefficient (sticky) Handling damage Size (for some types of silicone)
719Insulation -- Type Polyurethane
Advantages Biocompatible High tear strength Low friction coefficient Less fibrotic Small lead diameter
Disadvantages
ESC MIO
20Insulation -- Type Fluoropolymers (PTFE, ETFE)
Advantages Inert Most biocompatible High tensile strength Small size
Disadvantages Stiff when >0.003 More prone to creep Difficult to manufacture without pinholes
21Insulation -- Small SizeNew Insulation Materials Facilitate the Benefits of Smaller Lead Diameters Smaller introducer size Easier insertion/passage through smaller veins More flexible lead bodies Two leads through one introducer Less intrusive
822Electrodes Purpose
Deliver a stimulus to myocardium Detect (sense) intracardiac signals
Optimal Performance Factors Low, Stable Thresholds High Pacing Impedance Low Source Impedance Good Sensing
Tip Electrode Ring Electrode
23Electrodes Characteristics and Design Factors that Impact
Electrical PerformanceFixation mechanism Polarity Surface material Size Surface structure Steroid elution
24Electrodes -- Fixation Mechanism Passive Fixation Mechanism Endocardial
Tined
Canted/curved
925Electrodes Fixation Mechanism Active Fixation Mechanism Endocardial
Fixed screw
Extendible/retractable
26Electrodes -- Fixation/Visualization
Fluoroscopic Image of Leads
CapSure CapSure SP
NovusCapSure Z
Novus CapSureFixExtended Retracted Fixed Screw
space
27Electrodes -- Fixation Mechanism Fixation Mechanism Myocardial/Epicardial
Stab-in
Screw-in
Suture-on
10
28Electrodes -- Surface Material Surface Material
Polished platinum Activated carbon Platinized metal
Surface Material Characteristics Corrosion Resistant Biocompatible Reduced Polarization
29Electrodes -- Size
Reducing Electrode Size Increases Impedance Reduces Current Drain Increases Longevity
Disadvantage: Increase polarization
30Electrodes - Size/Polarization
Current Current Tissue
- + -
++
+++ -+
-
-++++ +--
Polarization Layering Effect
11
31Electrodes -- Surface Structure Porous Electrode Surface
CapSure8.0 mm2
Porous ElectrodeCapSure SP Novus5.8 mm2 Platinized Porous Electrode
CapSure Z Novus1.2 mm2 Platinized
Porous Electrode
15KV x2500 12.0V MDT
32Electrodes -- Surface Structure Benefits of a Porous Electrode Surface
Reduces Polarization
Improves Sensing
Promotes Tissue In-Growth
33Electrodes -- Steroid Elution
Tines forStable Fixation
Silicone Rubber PlugContaining Steroid
Porous, Platinized Tipfor Steroid Elution
Type - Steroid in matrix
12
34Electrodes Steroid Elution
IMPLANT CHRONIC(8 weeks or longer)
ExcitableCardiacTissue
Non-ExcitableFibroticTissue
ExcitableCardiacTissue
35Electrodes -- Steroid ElutionBenefits of Steroid Elution Excellent Electrode-tissue Biocompatibility:
Fewer and less active inflammatory cells Less fibrotic development
Improved Electrode Performance: No significant threshold peaking nor chronic threshold
increases, virtually eliminatingexit block
Improved consistent sensing characteristics
36Electrodes -- Steroid Elution Effect of Steroid on Stimulation Thresholds
Pulse Width = 0.5 msec
03 6
Implant Time (Weeks)
Textured Metal Electrode
Smooth Metal Electrode
1
2
3
4
5
Steroid-Eluting Electrode
0 1 2 4 5 7 8 9 10 11 12
Volts
13
37Leads - LV
CollegePM0_EGM_ECG1_Diseases2a_PacemakerBasics3_Sensing2b_Pacemakers2c_ICDBasics4_Leads