SensAASpectrometer
Atomic Absorption
What
and
GBC has alwaysplaced a strongemphasis on quality inall aspects of our operation,from design and manufacture tothe provision of service andsupport to our customers, andwe are fully committed tocontinuous evaluation andimprovement in all areas.
The GBC Quality ManagementSystem has been accredited tothe ISO 9001 quality standardby Lloyd’s Register QualityAssurance Limited. Thiscertification is your assurancethat the procedures andprocesses used to produce thegoods and services which GBCprovides comply with therelevant International Standard,and demonstrates commitmentto meeting the needs andexpectations of our customers.
For almost 30 years GBC hasbeen at the forefront of scientifictechnological development,manufacturing and marketing awide range of award winning,quality scientific instruments.
GBC’s product lines...
GBC Scientific Equipment
will advance people’s knowledge and
their capacity to enhance the quality of life
for all humankind.
HPLC ICP-OESAAS ICP-oTOFMS Rheometry UV-Vis XRD
“Sensitive Technologyfor a Sensitive World”
ISO 9001QualityAccreditation
3
Advanced features include:• Storage of signal graphics.
Since graphics traces are stored,the operator can view the graphicsfor previous samples and zoom infor a close-up if necessary.
• Complete report presentation.This includes printing ofoperator-defined text, headings,sample labels, calibration graphs,operating conditions, resultsand statistics.
• Simple operation is ensured withthe easy-to-use Windows® XPenvironment, together with anextensive on-line help facility.
No other AA spectrometer can offer thesimplicity and user-friendly operationfor your specific application.
• Automatic wavelength and slit settingfor maximum convenience anderror-free set-up.
• Optional automatic flamecontrol system with a full rangeof safety interlocks and integrated liquid trap.
• The unique GBC HYPER-PULSEbackground correction system,included as standard withall models.
• The GBC performance guarantee -sensitivity and precision superiorto that of any other instrumentmanufacturer.
• Data storage using a high capacityhard disk. Storage is provided forapplications, results, graphics,sample labels, weights and dilutions,report headings, etc.
GBC makes validation easy.
TheGBC IQ/OQ/PQ procedureprovides qualification of theInstallation, Operation andPerformance of your SensAADual Atomic AbsorptionSpectrometer.
Touch screen technology
GBC offers a full range of AAaccessories to ensure the widestapplication.
- System 3000 Graphite Furnace
- SDS-540 Autosampler
- PS-270 Diluter/Sample Preparation
- FS 3000 Autosampler
- HG 3000 Hydride Generator
- EHG 3000 Electric Hydride Heater
- MC 3000 Mercury Concentrator
The SensAA offers touch screen technology for ease of use.
Unique Turn-KeyAnalytical SolutionsGBC’s innovative technologyprovides optimum performance andsensitivity within a modest budget.
This unique combination of hardwareand software provides efficient solutionsto meet the specific needs of theanalysts. In addition the SensAA andSensAA Dual offers the following:• easy to operate single and
multi-element sampling• priority sample introduction for
manual sampling• fast flexible work flow for high
productivity analysesWith just one mouse click onyour desktop your application isunderway. You can begin with presetoptimised methods.
Application SourceThe heart of the SensAA is thepatented Application Source whichbrings unprecedented levels ofsimplicity, performance and costsavings to atomic absorptionspectrometry. The Application Sourceor “Mini Turret” consists of six individualcathodes aligned in a circle withinthe one unit. The type of cathodesare user selectable and can becustomized for various applicationssuch as:• environmental• food• petroleum• clinicalIt simplifies lamp handling, operationand alignment. Due to the uniquedesign of the GBC Application Sourcethere is no need to change lampswhen changing elements. This providesadditional cost savings.
The SensAA is versatile in that whileoffering unprecedented benefits theapplication source is interchangeablewith older hollow cathode lampsallowing the use of extra lamps ifrequired.
4
Features
Unsurpassed innovation
Automatic wavelength andslit setting for maximumconvenienceThe large, self-calibrating 0.33 metre(focal length) monochromator hasbeen specifically designed to providethe high light throughput and stabilityneeded for atomic absorption.Spectral bandwidth is continuouslyadjustable between 0.1 and 2 nm (in0.1 nm increments) and, for furnacework, reduced slit height is availablewith all slit widths. You simply select therequired element, and the wavelength,lamp current, slit width and slit heightare set automatically. Graphical displaysof wavelength scans are available forflame emission work or as a trouble-shooting aid. An optional wide-rangephotomultiplier tube covers the fullwavelength range (175 - 900 nm).
SensAA for maximum energyThe GBC SensAA uses the simplicityand high energy of reflective optics toprovide the best possibleperformance/price ratio. An automaticbaseline zero between samples allowsthe SensAA to be used for extendedperiods with flame or furnaceautosamplers without the problem ofbaseline drift.
SensAA Dual for true doublebeam stabilityThe genuine double beam opticalsystem of the SensAA Dual measuresthe sample and reference signalsevery mains cycle to give the bestcorrection for lamp noise and themost stable baseline possible.
Protected optical systemensures long-term performance
The SensAA Dual use mirrorsthroughout the optical system to ensurethat correct focus and high energythroughput are maintained at allwavelengths. To protect the opticalsurfaces, a dust cover is fitted over theoptical system and additional protectionis provided by coating the mirrorswith quartz.
Asymmetric Modulationreduces noise
GBC introduced asymmetricmodulation for double beaminstruments and continues it with theSensAA. Today, the technique hasbeen optimised for the best possibledouble-beam performance, andfurther refined for use in the SensAA.
Most of the noise in an atomicabsorption measurement comesfrom the sample beam. By spendingmore time measuring the samplebeam (twice that of most conventionalsystems) the GBC SensAA reducesnoise by up to 40%.
SensAA
5
Choice of flame
Guaranteed performance
Hyper-Pulse backgroundcorrection
The Hyper-Pulse background corrector,one of the fastest systems available,has been further improved by pulse interpolation. This allows formore accurate interpolation of“Transient Signals” such as GF.
With all background correctionsystems, there is a small time delaybetween the measurement ofbackground and total absorbance.When the background is changingvery rapidly, as is often the case withgraphite furnace work, this delay canlead to an error in the backgroundcorrected reading. Systems with slowersampling rates and longer delay timesshow greater errors.
Most background correctors measurethe background absorbance 50 or 60times per second and the delaybetween the measurement ofbackground and total absorbance canbe as much as 10 milliseconds.
The GBC Hyper-Pulse system takes200 (50 Hz) or 240 (60 Hz) samplemeasurements per second and thedelay between the measurement ofbackground and total absorbance isabout 1 millisecond. This produces adramatic reduction in backgroundcorrection errors. Accuracy is furtherimproved by interpolating betweenbackground measurements tocalculate the background when theatomic signal is measured.
This system also allows correction forhigher background levels than mostsystems-up to 2.5 total absorbance.
Interlocked flame control
This system allows the flame to beignited only if a burner is installed, theliquid trap is full and an air-acetyleneflame is selected. Changeover tothe nitrous oxide-acetylene flame isallowed only if a nitrous oxide-acetyleneburner is present. Push-button flameignition is included.
GCM (Gas Control Module)
This system is interlocked to protectagainst faults in gas supplies, burnertype, flame detector, mains power, liquidtrap, nebulizer bung and relief bung. Itprovides a programmed ignition andshutdown sequence - if nitrous oxide-acetylene is selected, an air-acetyleneflame is first established and then theacetylene flow is automaticallyboosted before automatic changeoverto the nitrous oxide-acetylene flame.
Extended Lamp Life
In addition, the design of the highintensity D2 lamp and associated powersupplies ensures unequalled lamplifetime. The GBC lamp will operate for1000 hours even at full current.Compare this to some alternative brandinstruments where the D2 lamp isguaranteed to last only 90 hours ofoperation, and add up the savings!
Guarantee of performance
GBC is so confident of instrumentquality that it guarantees both sensitivityand precision. GBC's guaranteedperformance is superiorto that of any other company in theworld - greater than 0.8 absorbancefor 5 mg/L copper with an RSD lessthan 0.45% for ten 10-secondintegrations. This performance isachieved by an efficient blend of gooddesign and durable materials.
Flame atomizationsystem delivers reliabilityand performance
All materials in contact with the samplehave been selected for maximumcorrosion resistance.The spray chamber is made from inertpolypropylene. The burner is madeentirely of titanium. The nebulizer has aplatinum-iridium capillary and atitanium venturi.
The nebulizer also has an adjustablesample uptake rate, essential foroptimum performance with refractoryelements or organic solvents. The wholeflame atomization system is easilyremoved for cleaning or for changeoverto graphite furnace.
6
GBC SensAA software
Fast and powerful
GBC has earned a world-widereputation for superlative AAS software.Easy for the first-time user to operate,it has the power to handle the mostdemanding applications.
The software comprises six individualmodules which simplify analysis andsetup, so that only one mouse click isrequired to initiate an analysis.Operation is further simplified by anextensive context-sensitive on-line helpsystem which includes tutorials onanalytical technique.
The system provides fully integratedsoftware for instrument and accessorycontrol, graphics, data acquisition,storage and reporting - there isno need to switch to add-onsoftware packages.
Advanced features of SensAAsoftware
• High resolution graphic displayfor methods development andquality control
The graphics signal is linked toindividual sample readings, allowingatomic and background traces to bedisplayed simultaneously.For graphite furnace work, graphicsare overlaid with the furnacetemperature program, simplifyingmethod development.
• Choice of calibration routines toprovide the most accurate results
The SensAA Dual allows for automaticpreparation of calibration curves usingup to ten standards. Choose an exactfit to the calibration points, rational orlinear least squares fit, or linear least squares fit through zero. You may alsocalibrate using standard additions orbracketing standards.
• The software is available in a widerange of international languages.
Element selection is made easy with SensAA software
7
GBC SensAA software
Fully integrated software
• Storage of signal graphics provides a valuable quality assurance aid
To aid in method development and quality assurance, signal traces for all analyzed samples are stored withthe corresponding numerical result. If a problem is suspected, simply clicking on the suspect replicate willdisplay the graphical result on the same page. Zoom and cross-hair facilities also enable a close upinspection that can be used to quickly diagnose any problem.
• Choice of measurement mode enhances productivity
Choose from peak height, peak area, integration or running mean. The automated running meanmode provides a significant reduction in measurement time since the measurement is completedwhen the required precision is achieved.
• Results editing facility means that samples need to be run only once
To ensure that samples and calibrations do not need to be repeated, the software features full editing capabilities for individual replicates as well as the ability to delete complete samples or standards. To prevent unauthorized editing, all results are passwordprotected to ensure data integrity.
• Status Panels
The user can monitor or access information on their samples or instrument parameters and conditions by using status panels. The user can select from nearly 400 parameters to incorporate into their own personalized status panels.
Each parameter can be displayed as editable or non editable text, dial or slide meters, switches, buttons, graphs or scans further adding to the flexibility of the SensAA software. These paratmeters can be assigned to customizable status button.
Results and graphics display
Customize selection via the status panel
8
GBC SensAA Software
Valuable quality assurance
• Analytical applications are storedfor instant recall
When you select the element to bedetermined, the software automaticallyselects the recommended wavelength,lamp current, slit width and flametype. After selecting the otherparameters for your application, allinformation may be stored in anapplication file for instant recallat any time.
• All data is preserved on hard disk
All results are stored on disk andcan be recalled for viewing orprinting at any time, or transferredto another disk for archiving.The time and date of analysis isstored with the results, enabling youto monitor the progress of samples,identify batches of results andcomply with the statutory regulations.
• Reporting facility enables fullydocumented analyses
Produce printed reports incomprehensive single element orsummarized multi-element formats.Single element reports can containany combination of informationincluding replicates, sample labels,calibration graphs, methodparameters and weights andvolumes. Reports can include pagefooters, headers and individualmargins, and all columns and rowscan be individually sized. A fullselection of fonts and font sizes maybe used. Connect an electronicbalance to directly transfer sampleweights to the weight and dilutiontable, or use a bar code reader todirectly enter sample labels.
• Help is instantly available
Context-sensitive on-line helpexplains all aspects of the hardwaresetup, including operation, safetyand maintenance. Learning to usethe instrument is made easy viamulti-media tutorials. While runninga tutorial or displaying context-sensitivehelp, you can work through thesoftware to create an applicationusing as much or as little help asis required.
• Continuous display of concentrationfor rapid sampling
The digital concentration read-outon the results and graphics screensis continually updated, invaluablefor optimization and rapidmanual sampling.
•Touch Screen Technology
Intuitive icons make operating thesoftware both easy and user friendly.
Works when operator is wearing gloves.
Comprehensive quality control protocols.
9
Regulatory compliance
Compliance
GMP/GLP compliance
The SensAA and SensAA Dualhardware and software offer fullGMP/GLP compliance for thoseanalyzes requiring certification.
• AA certification procedurepackage - ISO9001 accredited
• Automatic wavelengthcalibration software
• Calibration failurecontrol using calibrationalgorithm statistics
• Comprehensive Qualitycontrol software suite
• Concentrationrange checking
• QC check samplesincluding a fullrange of fail actions
• Automatic spikerecoveries allowingenhancement &suppressiveinterferencechecking
• Password protectionallowing useraccess control
• Full date & timestamping of results
• Method linked to results- eliminates confusionand enforces traceability
• Real time data export to any LIMS,allowing enhanced flexibility
• Y2K compliance certificateprovided with software
Regulatory Compliance
• Full compliance with USA Food andDrugs Authority, CFR 21, part 11“Electronic Records and Signatures”
• Security Menu to provide ElectronicRecords and Signatures
• Tamperproof Audit Trail File
• User levels with password log on toprovide method security
10
System 3000 graphite furnace for ultra-trace analysis
Performance
The System 3000 graphitefurnace suits applicationsrequiring the measurement ofvery low concentrations.
This system achieves state-of-the-artperformance enabling most elements tobe determined at sub-part-per-billionconcentrations. It blends all of the latestdevelopments in graphite furnacetechnology into a fully integrated system.The GF3000 power supply and PAL3000autosampler are both controlled from theAA computer. The superb System 3000incorporates all of the features expected inan advanced graphite furnace:
Rapid heating (2000°C/sec)under full control
The unique GBC temperature controlsystem uses true power feedback to ensurereproducible analytical conditions. Thissystem has a fast response and isindependent of fluctuations in mainsvoltage or conductivity of the graphitecomponents and covers the fulltemperature range.
Flexible temperature program
The temperature program allows anynumber of steps, enables performanceto be optimized for even the mostcomplex samples.
High performancefurnace assembly
The furnace enclosure allows the graphitetube to be surrounded by a carefullycontrolled flow of inert gas.
The tube itself is designed for optimumperformance, with raised sections ensuringthat all samples are safely confined to thecentral part of the tube.
Total pyrolytic graphiteplatform for reducedinterferences
Chemical interferences can be aproblem in many graphite furnaceapplications. Sensitivity is often reducedand complex sample treatment schemesare required to overcome theinterferences. The platform for theSystem 3000 eliminates or reduces mostchemical interferences.
Programmable gas selection
The provision of two gas supplies givesmaximum flexibility for handling thewidest range of samples.
The GF3000 graphite furnace workhead
Simple furnace program interface
11
Automatic furnace sampler
Accuracy and productivity
Performance
The System 3000 is designed toachieve state-of-the-art sensitivities anddetection limits. Approximately 50elements can be determined, most atsub part-per-billion concentrations.The typical characteristic mass for someof the elements is shown in thefollowing table. (Characteristic mass isthe mass of the element required togive an absorbance of 0.0044).
Element Charac. Element Charac.Mass (pg) Mass(pg)
Ag 0.9 Fe 1.5Al 5 Mn 0.7As 6 Mo 8Au 4 Ni 5Ba 12 Pb 2.5Cd 0.25 Se 20Co 4 Ti 45Cr 1.5 V 20Cu 2.5
The PAL3000 Programmable AutoLoader provides automaticcalibration with up to 10 standardsand automatic analysis of up to 40samples. Any sample volume between1 and 100 μL may be selected.
Automatic mixing of standardssaves time and improvesreliability
The PAL3000 will automaticallyprepare a series of up to 10 standardsfrom a single stock solution and canalso automatically prepare standardadditions from the stock solution. Thisprocedure eliminates errors in thepreparation of standards, reduces thelikelihood of contamination and savesconsiderable time.
Automatic dispensing ofchemical modifier eliminatesthe need for manual sampletreatment
With many applications, the addition ofa chemical modifier to the sampleimproves sensitivity, reducesbackground absorbance or reduceschemical interferences. The PAL3000can automatically add a specifiedvolume of chemical modifier to eachsample and standard.
Automatic dilution
Manual dilution is time and labour-intensive and is also prone to errorsand contamination. The PAL3000eliminates these problems byautomatically diluting any over-rangesamples to bring them into thecalibration range.
Multiple injections forautomatic pre-concentration
Several aliquots of a sample may beinjected and dried prior to atomization.This allows the effective samplevolume to be increased and permitsvery low concentrations to be measuredwithout the need for chemicalpre-concentration procedures.
Heated injection reducesanalysis time
The PAL3000 can be programmed toinject the sample after the graphite tubetemperature has been raised abovethe boiling point of the solvent.The sample then dries as it is injectedand the time-consuming drying stageis eliminated.
Simple set-up and storage ofprobe co-ordinates
The sampler capillary tip is convenientlyaligned with the furnace tube samplinghole with the use of the cursor keys onthe keyboard. Once established, theco-ordinates are stored in the computer,even when the power is switched off.
Top view of the PAL3000 autoloader
12
Specifications
SensAA
Optics
Double beam (SensAA Dual) withbackground correction and flameemission capability. The SensAA Dualhas asymmetric modulation with 2:1sample-to-reference ratio for noisereduction. All-reflective system withquartz overcoating on mirrors.Dust cover over optical system.
Monochromator
Ebert-Fastie design with 333 mm focallength and 175-650 nm wavelengthrange. Optional wide rangewavelength detector available 175 -900 nm. 1800 line/mm holographicgrating with dual-blazed profile and1.6 linear reciprocal dispersion.Automatic wavelength selection andpeaking. Continuously adjustable slitswith 0.1 to 2 nm spectral bandwidth(0.1 nm increments). Reduced heightfor furnace operation available withall slit widths. Automatic setting of slitwidth and height. Automatic scanning.Selected wide range multi-alkaliphotomultiplier tube.
Choice of lamp types
Plug-in mounting for hollow cathodelamp or Application Source withvertical and horizontal adjustments foroptimizing light throughput.
Lamp supply
Allows warm-up for next elementusing the Application Source.
Background correction
Hyper-Pulse background correctortakes 200 (50 Hz) or 240 (60 Hz)sample readings per second forcorrection of fast background peaks.With approximately 1 ms betweenpulses and interpolation betweenmeasurements, the best possible
Footprint
Minimum bench space of 455 mm x810 mm is all that is required.
Performance guarantee
Greater than 0.8 abs for 5 mg/Lcopper with an RSD less than 0.45%for ten 10-second integrations.
Interlocked flame control
Flame ignition only on air basedflames. Nitrous oxide ignitioninhibited. Push button actuated glowplug ignition system. Safety interlockson burner, nebulizer, pressure reliefbung and liquid trap. Safety interlocksystem also prevents nitrous oxideselection withair-acetylene burner. Fast changeselection of nitrous oxide-acetyleneflame.
Integrated liquid trap
The liquid trap is now integrated intothe spray chamber to enhance thesafety of the instrument.
Gas control module GCM(optional)
Microprocessor controlled. Monitorsgas pressures, flame and safetyinterlocks. Keypress selection of air ornitrous oxide-acetylene flames withautomatic fuel enrichment for nitrousoxide flames. Automatic flameshutdown upon mains power failure.
Flame atomization system
Premix design with solidpolypropylene mixing chamber. All-titanium burner construction.Nebulizer has platinum-iridiumcapillary and titanium Venturi. Liquidtrap with liquid level interlock. Quick-change mounting to enable easychangeover to graphite furnace.
SensAA
accuracy is assured. High intensitydeuterium arc lamp provides 175 -423 nm correction range. Corrects to2.5 total absorbance.
13
Options
High resolution display of atomicabsorbance and background signals,furnace temperature programs,calibration curves, peaking meters andwavelength scans. Different scales maybe used for atomic and backgroundsignals.Signals may be displayed in the singleshot, overlay or overview mode.Consecutive signals may be overlaid.Graphics cursor may be used to obtainnumerical information from graphicstraces. Zoom function allows graphicstraces to be expanded.
Data storage
Storage is provided for signal graphics,applications, results, sample labels,weights and dilutions, and reportheaders.
Quality
Quality control software satisfying GLPand GMP is supplied as standard.
Report generation
Reports may be printed during a run orfrom stored result files. They mayinclude headings, notes, operatingparameters, calibration graphs, samplelabels, results and statistics. Resultsfrom different runs (e.g., flame, furnaceand vapour generation) may becombined into a single report. Reportsmay be printed sequentially by elementor in a multi-element format.
Dimensions
Spectrometer module: 830 x 580 x 520(WxDxH, mm)
Weight
Spectrometer module: Unpacked 48 kg,Packed 60 kg.
Electrical Requirements110/220/240 volts AC, 50/60 Hz,300 VA.
AASSensAA
In-built computer
Touch screen interface computeroperating Windows. A minimum of256 MB RAM, one parallel port, twoUSB ports, LAN and a mimimum of 40GB data storage capacity.
Software
Microsoft Windows®XP operatingsystem for true multitasking.Controls SensAA Dual Plus AAS,SDS-540 autosampler, PS-270 auto-diluter, GF3000 graphite furnace andPAL3000 furnace autosampler, andMC3000 mercury concentrator.
Data processing
Analysis by atomic absorption oremission. Absorbance range to 2.5.Measurement by integration, runningmean, peak height or peak area.Mean and RSD of up to 50 replicatereadings. Calibration using up to 10standards. Curve correction by linearleast squares, linear least squaresthrough zero, exact fit or concentrationleast squares (polynomial), plusstandard additions or bracketingstandards. Programmable rescale usinga single standard or completerecalibration. Result editing to removeunwanted readings. Weight anddilution correction. All editingavailable either during or post-run.
Specifications
With capacity up to 540 samples and10 standards, it can be used for flameanalysis, hydride generation, mercurycold vapour analysis and 10 standards,it can be used for flame analysis,hydride generation, mercury cold vapouranalysis and with the MC 3000 mercuryconcentrator.
A built in peristaltic pump providestrouble free continuous stream of cleanrinse solution.
Dimensions
260 x 160 x 290 (WxDxH, mm)
Weight
Unpacked 7 kg, packed 10 kg
Electrical requirements
110/220/240 volts AC, 50/60 Hz.
Dispensed volume is 1–100 μL,programmable in 1 μL increments.All-PTFE capillary. 1 L rinse container.Probe set-up controlled by computer withco-ordinates stored in memory.Program options include automaticmixing of standards, automatic mixingof standard additions, automatic dilutionof over-range samples, automaticinjection of chemical modifier, multipleinjection, heated injection, automaticrescale or complete recalibration, checksample and spike recovery.
Inert gas requirement
Argon or nitrogen at a pressure of70–200 kPa (10–30 psi).
Dimensions
GF3000: 380 x 360 x 290 (WxDxH, mm.)PAL3000: 220 x 209 x 140 (WxDxH, mm.)
Weight
GF3000: Unpacked 35 kg, Packed 45 kg.PAL3000: Unpacked 7 kg, Packed 10 kg.
Electrical requirements
208/220/240 volts AC, 50/60 Hz,Rated current 15A, surge current 40A.
SDS-540 sample delivery system
The SDS-540 is an X-Y-Z autosamplersystem for flame AAS, ICP or ICP-MS.The SDS 540 is now 40% faster than thesuperseded model and providesunmatched productivity with full randomaccess capability from a small footprint.
Accessories
System 3000 Graphite Furnace
Automated graphite furnace system.Comprises GF3000 graphite furnacepower supply and workhead plusPAL3000 programmable automaticsample loader. Controlled by theAAS computer.
GF3000 furnace assembly
Furnace assembly includes graphite tube(and platform if required) mounted inenclosure with quartz windows.Permanently connected to powersupply by umbilical cord carrying gas,cooling water and electrical supplies.Inert and auxiliary gas supplies.Temperature range ambient to 3000°C.Computer-controlled maximum heatingrate of 2000°C/sec. Temperature programhas unlimited number of steps, eachwith ramp and hold, gas selection,graphics display option and readoption. Temperature controller monitorscurrent and voltage and uses powerfeedback to provide accurate controlover the full temperature range, andduring both ramp and hold stages.Interlocked to inert gas and coolingwater pressures. Corrects for changes incooling water temperature.
PAL3000 furnace autosampler
Accommodates up to 49 samples andone stock solution or 40 samples and10 pre-mixed standards or one stocksolution for automatic mixing of up to10 standards. Container volumes are2 mL for samples and standards, 5 mLfor automix standard, 10 mL forblank and chemical modifier.
14
Specifications
Full range of AA accessories
SDS-540 Sample Delivery System
15
Requirements
Weight
Unpacked 12 kg, packed 19kg
Electrical requirements
110/220/240 volts AC, 50/60 Hz,120 VA
HG3000 automatic hydride analysis
The HG3000 hydride generator is anautomatic continuous flow system.Coupled with the SDS-540 autosampler,it provides totally automated analysis ofarsenic, selenium, and the other hydrideforming elements at part-per-billionlevels. It can also be used to determinemercury by the cold vapour method.The optional EHG3000 electric heaterprovides controlled heating of thehydride cell to temperatures up to1000°C, allowing hydride analysis tobe performed without the use of aflame. With a precision of around 1%at ppb concentrations and a samplethroughput of 60 samples per hour,the HG3000 provides the ultimatein fast, accurate and convenienthydride analysis.
MC3000 mercury concentrator
Regulations for mercury analysis arepushing the required detection limitswell below the level achievable usingthe standard cold vapour technique.The MC3000 mercury concentratoraccessory for the HG3000 addressesthis problem. By concentrating themercury vapour on a gold foil, detectionlimits of 5 ppt can be readily achieved.The MC3000 accessory is fullycontrolled by the AAS softwareand unattended analysis can be carriedout when used in conjunction with theSDS-540 sample delivery system.
Specifications
Accessories
PS-270 Prep Station
The PS-270 is used in combination withthe SDS-540 auto-sampler, removingthe need for manual standard andsample preparation and sample dilutionfor flame and hydride applications.The PS-270 is totally controlled throughthe AA software enabling truefully-automated analysis. Simple andreliable automation of standardadditions and spike recovery, orsample matrix modification throughthe addition of up to two chemicalmodifier solutions provides hands-offoperation and versatility.
Dimensions
300 x 200 x 370 (WxDxH, mm)
Weight
Unpacked 5 kg, packed 11 kg.
Electrical requirements
110/220 volts AC, 50/60 Hz
FS3000 autosampler
For flame and hydride use only.Carousel holds 10 standards and 60samples. Separate positions for rinse,blank and rescale standard.
Dimensions
420 x 230 x 380 (WxDxH, mm)
Designed and manufactured byGBC Scientific Equipment Pty LtdA.C.N. 005 472 686
GBC reserves the right tochange specifications withoutprior notice.
GBC publication number01-0966-02
August 2007 Australia
GBC SCIENTIFIC EQUIPMENT
Manufacturer of world-classscientific instruments andaccessories - AA, HPLC, ICP-OES, ICP-TOFMS,Rheometry, UV-Vis and XRD
12 Monterey RoadDandenong, Victoria 3175Australia
Telephone 61 3 9213 3666Facsimile 61 3 9213 3677email [email protected] www.gbcsci.com
All trade-marks and trade-namesare the property of theirrespective owners.
Ordering InformationModel Optics Background Flame Control Part NumberSensAA Dual* Double Beam Yes GCM 99-0538-00SensAA Dual* Double Beam Yes Interlocked 99-0539-00SensAA G* Double Beam Yes None 99-0542-00
All SensAAs are supplied with an inbuilt computer, 256 MB RAM, 40 GByte Hard Disk, one parallel port, 2USB ports, LAN and15 inch High Resolution Resistive Touch LCD, Sound Blaster Sound Card and Ethernet Card. The PC is loaded with Windows XP,SensAA software and touchscreen drivers.
The SensAA G is not capable of flame operation and is designed to operate with a graphite furnace or hydride generator only.
All SensAAs are supplied with operation manual and SensAA software.
All flame SensAAs are supplied complete with air-acetylene burner, spray chamber, adjustable nebuliser, gas hoses, and flamemethods manual.
AAcccceessssoorriieessPremium PMT Upgrade Kit 99-0563-00Nitrous oxide – acetylene burner 99-0075-00Recommended spares and consumables for 2 years operation 95-0637-00Fume extraction system (complete) 220/240V, 50 Hz 99-0012-00Fume extraction system (complete) 110V, 60 Hz 99-0012-01Fume extraction system (complete) 220, 60 Hz 99-0400-00Air compressor (oil filled) 110V, 60 Hz 75-0005-00Air compressor (oil filled) 220/240V, 50 Hz 75-0006-00Air compressor (oil fiIled) 220V, 60 Hz 75-0045-00Air compressor (oil less) for high humidity or high altitude 220/240V 75-0049-00Air compressor (oil less) for high humidity or high altitude 110V 75-0050-00Nitrous oxide cylinder regulator with heater 99-0050-02Air cylinder regulator 99-0099-00Acetylene cylinder regulator 99-0100-00Balston gas purifier 99-0284-00Gas purifier maintenance kit 95-9912-00
Recommended spares and consumables for SensAA 95-0367-00
FFllaammee AAuuttoossaammpplleerrFS3000 Flame Autosampler 99-0106-00
Supplied complete with test tubes, beakers, stainless steel probescables, tubing and 60 position sample rack
SDS-540 Sample Delivery System 99-0326-11Supplied complete with test tubes, PTFE–coated probes, cables,tubing, inbuilt wash pump, fixed wash reservoir and three 60–position sample racks
PS-270 Sample Preparation System 99-0407-11Sampler/accessory trolley 96-0021-00
HHyyddrriiddeeHG3000 Automatic hydride generator 99-0276-00Recommended spares and consumables for HG3000 95-0016-00EHG3000 Electric heater for hydride cell 99-0237-11MC3000 mercury concentrator unit for HG3000 99-0245-11
GGrraapphhiittee FFuurrnnaacceeSystem 3000 automated graphite furnace system (complete) 99-0108-00
Comprises GF3000 power supply and workhead plus PAL3000 automatic samplerloader. Supplied complete with 10 Pyrollytically coated graphite tubes, 5 platforms,500 sample vials, Furnace Methods Manual, beakers, hoses, cables and tubing
Graphite Furnace Tubes, pyrollytically coated (pack of 10) 99-0059-00Pyrolltic graphite platforms (pack of 10) 99-0060-00Pyrollytically coated graphite furnace tubes with integrated platform (pack of 10) 99-0342-00Pair of electrodes 99-0061-00Sample vials for PAL3000 (pack of 500) 99-0022-00Recommended spares and consumables for 2 years operation for System 3000 95-0015-00
/ColorImageDict > /JPEG2000ColorACSImageDict > /JPEG2000ColorImageDict > /AntiAliasGrayImages false /CropGrayImages true /GrayImageMinResolution 300 /GrayImageMinResolutionPolicy /OK /DownsampleGrayImages true /GrayImageDownsampleType /Bicubic /GrayImageResolution 300 /GrayImageDepth -1 /GrayImageMinDownsampleDepth 2 /GrayImageDownsampleThreshold 1.50000 /EncodeGrayImages true /GrayImageFilter /DCTEncode /AutoFilterGrayImages true /GrayImageAutoFilterStrategy /JPEG /GrayACSImageDict > /GrayImageDict > /JPEG2000GrayACSImageDict > /JPEG2000GrayImageDict > /AntiAliasMonoImages false /CropMonoImages true /MonoImageMinResolution 1200 /MonoImageMinResolutionPolicy /OK /DownsampleMonoImages true /MonoImageDownsampleType /Bicubic /MonoImageResolution 1200 /MonoImageDepth -1 /MonoImageDownsampleThreshold 1.50000 /EncodeMonoImages true /MonoImageFilter /CCITTFaxEncode /MonoImageDict > /AllowPSXObjects false /CheckCompliance [ /None ] /PDFX1aCheck false /PDFX3Check false /PDFXCompliantPDFOnly false /PDFXNoTrimBoxError true /PDFXTrimBoxToMediaBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXSetBleedBoxToMediaBox true /PDFXBleedBoxToTrimBoxOffset [ 0.00000 0.00000 0.00000 0.00000 ] /PDFXOutputIntentProfile () /PDFXOutputConditionIdentifier () /PDFXOutputCondition () /PDFXRegistryName () /PDFXTrapped /False
/Description > /Namespace [ (Adobe) (Common) (1.0) ] /OtherNamespaces [ > /FormElements false /GenerateStructure false /IncludeBookmarks false /IncludeHyperlinks false /IncludeInteractive false /IncludeLayers false /IncludeProfiles false /MultimediaHandling /UseObjectSettings /Namespace [ (Adobe) (CreativeSuite) (2.0) ] /PDFXOutputIntentProfileSelector /DocumentCMYK /PreserveEditing true /UntaggedCMYKHandling /LeaveUntagged /UntaggedRGBHandling /UseDocumentProfile /UseDocumentBleed false >> ]>> setdistillerparams> setpagedevice