doi:10.1136/hrt.2008.156372 2009;95;334-342 Heart Kim Rajappan part II Permanent pacemaker implantation technique: http://heart.bmj.com/cgi/content/full/95/4/334 Updated information and services can be found at: These include: Data supplement http://heart.bmj.com/cgi/content/full/95/4/334/DC1 "Education module notice" References http://heart.bmj.com/cgi/content/full/95/4/334#BIBL This article cites 7 articles, 3 of which can be accessed free at: Rapid responses http://heart.bmj.com/cgi/eletter-submit/95/4/334 You can respond to this article at: service Email alerting the top right corner of the article Receive free email alerts when new articles cite this article - sign up in the box at Topic collections (8936 articles) Clinical diagnostic tests (1 articles) Alcohol-induced amnestic disorder (Korsak's syndrome) (3131 articles) Interventional cardiology (1 articles) Arrhythmias Articles on similar topics can be found in the following collections Notes http://journals.bmj.com/cgi/reprintform To order reprints of this article go to: http://journals.bmj.com/subscriptions/ go to: Heart To subscribe to on 16 June 2009 heart.bmj.com Downloaded from
Transcript

doi:10.1136/hrt.2008.156372 2009;95;334-342 Heart

  Kim Rajappan  

part IIPermanent pacemaker implantation technique:

http://heart.bmj.com/cgi/content/full/95/4/334Updated information and services can be found at:

These include:

Data supplement http://heart.bmj.com/cgi/content/full/95/4/334/DC1

"Education module notice"

References

  http://heart.bmj.com/cgi/content/full/95/4/334#BIBL

This article cites 7 articles, 3 of which can be accessed free at:

Rapid responses http://heart.bmj.com/cgi/eletter-submit/95/4/334

You can respond to this article at:

serviceEmail alerting

the top right corner of the article Receive free email alerts when new articles cite this article - sign up in the box at

Topic collections

(8936 articles) Clinical diagnostic tests � (1 articles) Alcohol-induced amnestic disorder (Korsak's syndrome) �

(3131 articles) Interventional cardiology � (1 articles) Arrhythmias �

  Articles on similar topics can be found in the following collections

Notes  

http://journals.bmj.com/cgi/reprintformTo order reprints of this article go to:

http://journals.bmj.com/subscriptions/ go to: HeartTo subscribe to

on 16 June 2009 heart.bmj.comDownloaded from

ARRHYTHMIAS

Permanent pacemaker implantationtechnique: part IIKim Rajappan

Correspondence to:Dr Kim Rajappan, CardiacDepartment, John RadcliffeHospital, Headley Way,Headington, Oxford OX3 9DU,UK; [email protected]

Having addressed equipment requirements andsome of the early steps in the process of permanentpacemaker (PPM) implantation in part I of thistwo part series, this section continues with the restof the PPM implantation procedure and someaspects of post-procedural management/care.

LEAD PLACEMENT TECHNIQUESBefore discussing lead placement itself, it isimportant to briefly explain the structure of apermanent pacing lead. The leads themselves arevery floppy and intrinsically have very littlestiffness. This means that as the lead is movedaround, the tip moves freely without any signifi-cant ability to steer it. To overcome this there is acentral lumen to the lead which will allow passageof a stiffer thin wire known as a ‘‘stylet’’. Thefurther the stylet is passed down the lead(potentially almost to the tip), the more of thelead body is stiffened (fig 1). These stylets may alsobe ‘‘reshaped’’ easily to allow the tip of the lead tobe further steered in a specific direction (fig 1). It isimportant to keep this stylet clean and free ofdebris, particularly blood, as this can block thecentral lumen and prevents the stylet from passingfar enough down the lead to give any usefulsupport. Also, the different lead positioning tech-niques described below are not mutually exclusive.A competent operator will be comfortable withmost of them to adapt to different situations,although they may have a preference for which onethey use first. The fixation method of the lead alsohas important implications. Lead tips may fixate‘‘passively’’ or ‘‘actively’’. Passive fixation leadshave ‘‘tines’’ at the end of the lead (fig 1) which actas an anchor to hold the lead tip in place acutely.Over a period of time (weeks to months) the tip ofthe myocardium around the lead tip will fibrose tosecure it further.

An active fixation lead has a retractable screw atthe tip of the lead which is deployed when the leadis in position (fig 1). To do this the lead is firstplaced in the desired position (see below), using astylet to support the lead tip against the myocar-dium. A clip-on tool (‘‘A-frame’’) is then attachedto the distal pin of the lead and slow clockwiserotation of the distal pin transmits torque toextend the screw. This can be directly visualisedunder fluoroscopy (for some leads there is ring tipseparation, and on others the rings come together)and slight resistance may be felt when rotating thedistal pin as the myocardium is entered. Rotation

of the distal tip too rapidly may build up torquewithin a lead such that the whole lead simply coilsup and the screw is not extended. To check thestability of the lead tip, the stylet is then pulledback to see that the lead stays in place before leadtesting. Specific instances where an active fixationlead may be used is when placing the rightventricular (RV) lead in the outflow tract/on thehigh septum, or in patients where the right atrial(RA) lead needs to be secured to the tissue forstability (some operators will use an active fixationRA lead in any patient who has previously hadcardiac surgery, as the RA appendage may havebeen ligated when the patient is placed oncardiopulmonary bypass). It is felt that activefixation leads are easier to extract (although whenthe lead has been in for a long period of time anylead extraction may be difficult); therefore manyoperators will electively use these in youngerpatients, knowing that over time there is amoderate chance that new leads will be neededand the old ones will need to be removed.

RIGHT VENTRICULAR LEAD PLACEMENTTraditionally most RV leads have been positionedin the apex; however, there is a growing trend toplace the RV lead on the septum or in the RVoutflow tract (RVOT). The discussion of therelative merits of these different positions has beenperformed elsewhere1–3 and comparisons are out-lined in table 1. The first component of placing theRV lead is common to both positions and entailscrossing the tricuspid valve (TV) to enter the RVitself. Broadly speaking, three techniques for thisare commonly practised and are shown in fig 2diagrammatically. The first is by ‘‘prolapsing’’ thelead across the TV. The lead is advanced into theRA with the stylet withdrawn 5–10 cm from thetip. As it is further advanced reasonably rapidly thetip may ‘‘catch’’ on the tricuspid annulus. If not,then slight reshaping of the stylet can help. Oncecaught, the lead is advanced with the stylet stillwithdrawn 5–10 cm from the tip and a loop of leadgenerated in the right atrium (RA). A straightstylet is then advanced (and initially the lead itselfis slightly retracted), prolapsing a portion of thelead 5–10 cm from the tip through the TV. As thestylet is advanced further the lead tip may thenflick through the TV or it may get caught on theTV with the rest of the body of the lead continuingto prolapse through. If the latter occurs the styletmay be advanced further to try to force the tip off

Education in Heart

334 Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372

on 16 June 2009 heart.bmj.comDownloaded from

of the TV; however, often it is firmly caught ontrabeculation. In this instance the stylet can bewithdrawn a few centimetres within the lead andthe lead retracted and advanced to try to free thetip, still maintaining the prolapsed configuration—a small degree of caution is needed here becausevigorous manipulation and particularly rotation ofthe lead can cause the tip to become entangled inthe TV apparatus, and make freeing the tip moredifficult. Ultimately if the tip is caught it may benecessary to pull the entire lead back into the RAand start again; if this occurs repeatedly, the‘‘direct crossing’’ or ‘‘drop down’’ techniquesshould be used.

These techniques may be used in the firstinstance, particularly with larger French leads.With the direct crossing technique the stylet isreshaped and advanced towards the tip of the leadto help point it at the TV. The lead is then passeddirectly across the TV. This may take smalladjustments in the direction of the tip, both byrotating the lead clockwise and anticlockwise toaim anteriorly and posteriorly, respectively, as wellas advancing or retracting the stylet to aimsuperiorly or inferiorly. If the lead still does notdirectly cross it may be necessary to reshape thestylet again. Once across the TV the stylet may beslightly withdrawn to allow the tip of the lead tofall towards the apex. Sometimes the tip of thelead will continue towards the RVOT once the TVhas been crossed, particularly if a curved stylet isadvanced all the way into the lead before crossingthe TV. If the final desired position of the RV leadis apical then the stylet is replaced with a straightone; as it is advanced through the lead across theTV, the lead itself may be gradually withdrawncausing the tip to ‘‘drop down’’ to the apex. It canthen be advanced into a more apical position ifnecessary. If the final position of the RV lead is tobe in the RVOT, or high septum at least, then theTV should be crossed with a shaped stylet akin tothe direct crossing technique, but with the styletadvanced all the way into the lead. The lead maythen naturally cross the TV and go straight ontothe high septum/RVOT as it is advanced further.However, if the tip of the lead goes further up intothe RVOT it may be necessary to withdraw the

lead gradually and then advance again when at thecorrect level. This may need a slightly differentshaped stylet. Once in position in the RVOT/highseptum it is important to ensure the lead is actuallyposterior and septal rather than anterior and on thefree wall. This is done by moving to an LAOprojection and confirming the posterior position(fig 3).

There are a few important points and pitfallsthat are common to RV lead positioning to beaware of. The proximity of the coronary sinus (CS)os to the TV means that it is possible to place thelead in the CS inadvertently; however, there areways to avoid this. The occurrence of ventricularectopics is a useful indicator that the lead is acrossthe TV and in contact with RV myocardium ratherthan in the CS. Also the lead position may bechecked in the LAO projection—if in the CS or abranch of it, the lead will pass more posteriorlythan expected. Another visual clue is that the leadtip appears to be at the RV apex, but the lead bodyappears excessively straight in the RV cavity andadvancing it forward does not result in a classical‘‘shoulder’’ of the lead across the TV. This suggeststhe lead may be in the middle cardiac vein or aposterolateral branch of the CS. Pacing on a lead inthe CS will also reveal a right bundle branch blockpattern rather than a left bundle pattern, but as theprecordial leads are generally not available atimplant, a clue can be the presence of a notablyleftward axis in the limb leads which are usuallyused. When testing the lead, if parameters are notgood despite excellent lead–myocardium contactand an apparently good anatomical position, checkall connections from the analyser to the lead forany correctable problems (see below); however, ifnone are found the lead should be moved as thisreflects a position that will chronically be poor.Also, if parameters unexpectedly deteriorate with-out any apparent lead dislodgement, consider thepossibility of lead migration and perforationthrough the RV. Often this has no clinical sequelaeand the lead can simply be pulled back andpositioned again, but it is worth ensuring haemo-dynamic stability and performing echocardiogra-phy to confirm the absence of any pericardialeffusion. Finally, one should pull back the stylet

Figure 1 Lead types, stylets, and use of the stylet to change the shape of the lead tip. A passive fixation lead tip (A), and active fixation lead tip withscrew retracted (B) and deployed (C) are shown. A J shaped (D) and straight (E) stylet may be used to stiffen the lead. Panels F, G and H show a pre-shaped J lead initially with the stylet advanced all the way to the tip (F) straightening the lead, but as the stylet is withdrawn the lead flexes (G) andeventually takes up its pre-formed shape (H).

Education in Heart

Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372 335

on 16 June 2009 heart.bmj.comDownloaded from

into the portion of the lead that is in the superiorvena cava (SVC) or subclavian before assessing thepacing characteristics of the lead to minimise thechance of false readings—this will also help toconfirm the final position of the lead, as the styletwill affect the lead shape.

RIGHT ATRIAL LEAD PLACEMENTApart from passive and active fixation RA leads,there are also those that have a pre-shaped ‘‘J’’shape, which is straightened by introducing astraight stylet, and those leads that are inherentlystraight and require a ‘‘J’’ shaped stylet to curve thetip (fig 1). The passive fixation RA leads are all pre-shaped, while active fixation leads come in eitherform (depending on manufacturer). Traditionallythe RA lead is placed in the RA appendage (RAA).In some patients where the RAA position is notsuitable (usually because of poor pacing character-istics) the lead may be placed anywhere else withinthe RA, but most commonly on the lateral wall.Like RV lead positioning there are standardmanoeuvres that are used to position the RA lead,and again one should be aware of and have someexperience of them all (fig 4).

The first technique involves placing the distal tipof the lead in the middle of the RA. If a pre-shapedlead is being used, the straight stylet is graduallywithdrawn to allow the tip to rise into the RAA. Ifa straight lead is being used, a J shaped stylet isintroduced at this point—be aware that if thecephalic vein has been used for access and/or thereis tortuosity in the subclavian, then the lead mayneed to be gently advanced and retracted as the Jshaped stylet is inserted to allow smooth passage.The lead is then gently withdrawn to try tointubate the RAA os. If too low the lead mayprolapse on itself and simply withdraw into thesuperior vena cava (SVC). If this happens the pre-shaped lead is straightened again with the stylet(or the J shaped stylet is withdrawn from thestraight lead) and the procedure repeated, butstarting a little higher in the RA itself. If too highinitially, then as the straight stylet is withdrawnform the pre-shaped lead (or a J shaped stylet isinserted into a straight lead) the tip catches in aflatter orientation. By advancing the lead it mayintubate the RAA. It is important to be aware thatthe RAA position varies and can appear different inthe standard postero-anterior (PA) projection.Therefore, it may be necessary to check the leadposition in different fluoroscopic projections (forexample, in LAO the lead tip should generally

point anteriorly). Furthermore, the classical ‘‘wind-screen wiper’’ or ‘‘figure-of-eight’’ movement ofthe lead tip suggests a good position in the RAA—but be aware that this motion is absent in atrialfibrillation. When it is decided to place the RA leadin a position other than the RAA, it is generallyadvisable to use an active fixation lead. Positioningis often on the lateral wall. Again the lead is placedin the mid RA and then rotated clockwise to spinthe lead laterally. Sometimes to keep the lead tipfree in the RA, the whole lead will need to beadvanced or retracted as the lead is rotated. Ideallythe tip of the lead should be as close toperpendicular to the wall as possible to maximisethe chance of getting good fixation, but care isneeded as there is a small risk of perforation.

PACING CHECKSOnce the lead is in what is felt to be a goodanatomical position, the pacing parameters need tobe checked to ensure that these are also satisfac-tory. The testing cables are connected at one end toan analyser. At the other end a pair of colouredcrocodile clips are used to test the lead. With abipolar lead there are two rings on the lead and theblack clip goes to the distal ring, the red to theproximal (fig 5). If a unipolar lead is being used(rare at implant these days for single/dual chamberpacemakers but may be encountered at the time ofgenerator changes (see below)) then the lead onlyhas a single distal ring so the black clip is connectedto this, and the red clip needs to be connected tosubcutaneous tissue to reproduce the effect of thegenerator acting as the other electrode. Thesimplest way to do this is either to attach the clipto tissue within the pocket, to use a speciallydesigned plate which is included in some pacingpacks, or my own preference is to attach an arteryclip to the inside of the pocket and then attach thered crocodile clip to this. Once connected, andbefore formally testing the pacing characteristics ofthe lead, the ‘‘injury current’’ is checked (fig 5). Ifthis is large it suggests good contact between thelead tip electrode and the myocardium.

In the presence of a large injury current, aninitially high pacing threshold (see below) will tendto come down substantially within 5 min, so it isworth waiting before repositioning a lead thatotherwise appears to be in a good anatomicalposition. Routinely the following parameters arethen tested (acceptable values are shown in table 2):

c Sensing—Measured in millivolts (mV), this isthe intrinsic signal that the pacemaker willdetect on the lead. It is accepted that the largerthe signal is, the better. In the case of the RVlead if the R wave (the sensed signal on the RVlead) is too small then the pacemaker may notsense intrinsic rhythm and start to paceinappropriately, with a small risk of R-on-Tarrhythmias. More importantly, if the R waveis too small and the device is programmed totry to account for this, it is then possible thatthe lead will detect artefact that is not

Table 1 A comparison of right ventricular (RV) apical and septum/outflow tract pacing

RV apical pacing RV septum/outflow tract pacing

Ease of positioning +++ ++Risk of perforation More likely with RV apical pacing but still low overall incidence, and mainly because

of thin tissue at the apex in some patients

Displacement risk More common with septal/outflow tract leads even when active fixation leads areused

Pacingcharacteristics

Satisfactory in the majority May be more physiological

Education in Heart

336 Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372

on 16 June 2009 heart.bmj.comDownloaded from

ventricular activity and inappropriately NOTpace the ventricle when the patient is poten-tially asystolic. In the case of the RA lead it isnot only the size of the atrial signal (the Pwave) that is important but also the presenceof ventricular activity that is distant but stilldetected by the atrial lead (so called ‘‘far field’’sensing). If the far field activity is too largerelative to the P wave on the atrial lead thenthe pacemaker may see both of these and counttwo atrial signals where there is only one. Most

pacemakers to varying extents can be pro-grammed around this problem. However, care-ful positioning of the lead at the time ofimplantation to minimise the far field compo-nent can save significant future problems,specifically the possible need to reposition thelead at a later date. Ultimately a smalleramplitude signal may be accepted if multiplesites have been tested, but this decisiongenerally needs experience to know when tostop searching for a better position.

Figure 2 Diagramsshowing three differenttechniques for rightventricular lead placement(see main text).

Education in Heart

Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372 337

on 16 June 2009 heart.bmj.comDownloaded from

Figure 3 Diagrams showing two different techniques for placing the right ventricular lead on the septum/in the right ventricular outflow tract (seemain text). The fluoroscopic images show the right ventricular lead in the apical position on the left (in the right anterior oblique (RAO) and left anterioroblique (LAO) projections), and in the high septal position on the right. Particularly note the LAO projection of the septally placed lead—it is importantthat the lead points as posteriorly as possible to ensure placement on the septum (marked with a dashed line) rather than the free wall, whereperforation and displacement risk are higher.

Education in Heart

338 Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372

on 16 June 2009 heart.bmj.comDownloaded from

c Impedance—Measured in ohms (V), the impe-dance measured though the lead is a reflection ofall the factors that oppose flow of electric currentthrough the lead. This includes the lead conductor(conductor resistance), the resistance to currentflow from the electrode to the myocardium(electrode resistance), and the accumulation ofcharges of opposite polarity in the myocardium atthe electrode–tissue interface (polarisation impe-dance). Again acceptable values are shown intable 2. This measurement may be more impor-tant during PPM follow-up where a high impe-dance value suggests a fracture in the leadconductor or a loose set-screw at the proximalconnector, while a low lead impedance valuesuggests a break in the lead conductor insulation,either of which may necessitate replacement.

c Pacing threshold—Measured in volts (V) and pulsewidth (in milliseconds (ms)), this is the minimumamount of energy needed to capture the myo-cardial tissue electrically. The lower this value isthe better, as the battery depletion is proportionalto the amount of energy needed. In most casesthe pulse width is fixed (at 0.5 ms, for example)and then the voltage is gradually reduced untilcapture is lost (fig 5). The threshold in this case is0.6 V at 0.5 ms. The specifics of voltage and pulsewidth measurement and their implications maybe studied in more detail elsewhere.4 Maximaloutput pacing (usually 10 V) is also performed oneach lead to ensure there is no stimulation ofextracardiac structures (normally indicated bydiaphragmatic twitching).

c Stability testing—The pacing characteristics ofthe leads are tested under conditions thatmight cause a loss of capture. To perform thisat the time of the implant it is most straight-forward to ask the patient to breathe in deeply,cough, sniff, or pant. Some operators will usefluoroscopy to check the amount of movementof the lead during deep inspiration and ensurethere is enough slack in the lead to allow forstraightening during this manoeuvre withoutdisplacing the lead tip.

LEAD CONNECTION, GENERATOR PLACEMENT INPOCKET, AND WOUND CLOSUREOnce a satisfactory position and lead parametershave been achieved, the lead is secured with

non-absorbable sutures around the collar. It isimportant to secure the leads to muscle—if securedto subcutaneous tissue the lead may slip. If thecephalic vein has been used the silk tie around theproximal end and the lead(s) is tied tight enoughonto the lead(s) to stop any residual blood flowfrom the vein, but not so tight as to damage theouter lead insulation or coils. Once the leads aresecured the pacemaker generator is connected tothe leads and placed in the pocket. Caution shouldbe taken at this stage to make sure that:

1. If there is more than one lead—for example,right atrial and ventricular—the correct lead isplaced in the correct port on the header.Checking the serial number on the lead andwhere it has been inserted with the techni-cian/physiologist is the most reliable method.

2. The leads are inserted all the way into theheader—check that the distal end of the pin isshowing past the distal connector.

3. ALL set screws are tightened adequately—thescrewdriver is turned clockwise to tighten.Some manufacturers have one set screw for abipolar lead, while others have two.

4. When placing the generator in the pocketensure the redundant length of the leads arecoiled and placed behind the generator,otherwise there is a greater risk of damageat the time of future generator changes.Ideally, the generator should be placed inthe pocket with the logo face up. Sometimesthe pulse generator is secured to the muscleusing a stitch through a hole on the header.This may help prevent migration in certainpatients—for example, elderly patients withloose subcutaneous tissue or where thegenerator has migrated caudally at the timeof a generator change.

Wound closure may be performed in differentways—interrupted sutures, continuous suture,locking stitch, etc. However, common to alltechniques is that one or two layers of absorbablesutures are used to close the deeper subcutaneouslayer, while a single layer (either absorbable or non-absorbable) is used in the subcuticular plane. Someoperators use tissue glue—for example,Dermabond—to replace the subcuticular suture,

Figure 4 Diagramsshowing three differenttechniques for right atrial(RA) lead placement (seemain text). RAA, right atrialappendage.

Education in Heart

Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372 339

on 16 June 2009 heart.bmj.comDownloaded from

but this can only be used if there is good haemostasisand skin apposition with the deeper sutures. It isimportant to get the tension in these sutures correct(too tight will cause puckering of the skin, and tooloose will allow gaping of the wound).

POST-PROCEDURE CAREControversy remains over the use of topicalantibiotics in the pocket (commonly gentamicin)

and the routine use of antibiotics parenterally and/or orally after the procedure. This will, like the pre-procedure antibiotic regimen, be guided by localprotocols/guidelines and individual operator pre-ference. Chest radiography should be performed inany patient who has had a subclavian puncture toexclude a significant pneumothorax, and in manyinstitutions may be performed in all patientsroutinely irrespective of venous access route.Also, increasingly these procedures are performedin patients who are discharged the same day (daycase) as evidence for many years has shown thatthis is safe,5 with or without further testing beforedischarge.

ANTICOAGULATION ISSUESA comprehensive review of the management ofanticoagulation at the time of pacemaker implantis outside the remit of this article, and reviews on

Figure 5 Panel A shows the IS-1 connection (the current standard connector tip) of a bipolar pacing lead. (Theproximal ring electrode is indicated by the red arrow and the distal electrode by the black arrow; this corresponds to theelectrodes at the tip of the lead itself.) Below this the cable for testing the lead parameters at implant have beenconnected. The ‘‘crocodile’’ clips used are colour coded with red being attached to the proximal electrode and black tothe distal electrode. Panel B shows recordings of the signal from the pacing lead known as the intracardiac electrogram(EGM). The top strip is taken immediately at implant and attention is drawn to what may be described as ST elevation(circled) if this were a normal ECG—this is known as the ‘‘injury current’’ and suggests the lead is in good contact withthe myocardial surface. Within a relatively short period of time (minutes to hours) this ‘‘injury current’’ disappears and amore simple EGM spike (circled in the lower strip) is left. In panel C a printout is shown from the analyser duringventricular lead testing. The top row shows the pacing output in volts (V) at that particular time, and can be seen togradually decrement from 2.0 V on the left to 0.3 V on the right. The next line shows the intracardiac EGM recorded (asin panel B); the next line is the marker channel which shows when a pacing impulse is produced by the generator(denoted by a VP in this case); and the final line is an ECG recording. On the left of the strip with every pacing impulse(VP on the marker channel) there is a corresponding intracardiac EGM and a QRS complex on the ECG (solid blackarrows). This continues down to 0.6 V, but at 0.5 V there is a pacing impulse (VP again on the marker channel) but nointracardiac EGM and no QRS on the ECG (red arrows), indicating failure to capture the ventricular tissue at this output.Subsequently there is the patient’s own intrinsic QRS complex (VR on the marker channel indicates that this signal fallsin the ‘‘refractory’’ period of the pacemaker setup) and different intracardiac EGM and QRS on the ECG are seen (dashedblack arrows). The threshold—that is, the minimum pacing output at which the ventricular tissue is captured with thepacing lead in this position—is in this instance 0.6 V (with a pulse width of 0.5 ms).

Table 2 Pacing parameters and normal ranges

Sensing (millivolts, mV) Impedance (ohms, V) Threshold (volts, V)

P wave >1.5 mV 400–1000 V ,1 V at 0.5 ms

R wave >5.0 mV

May accept lower values ifmultiple sites have already beentried and found unacceptable

Will vary for different leads—refer to manufacturer’s guidancefor the specific lead

The lower the threshold is thebetter, but like sensing, sometimesone will need to accept highervalues

Education in Heart

340 Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372

on 16 June 2009 heart.bmj.comDownloaded from

this subject are available.6 7 The main issuesurrounds how to manage a patient who is takingwarfarin. Clearly the indication for warfarin andthe risks of stopping it will largely determine whatis done in this respect. An increasing number ofoperators are performing these procedures withoutstopping the warfarin, aiming to keep the interna-tional normalised ratio (INR) between 2 and 3, aswell as being meticulous with haemostasis. Resultswith this approach suggest it is safe with fewercomplications as a result of under anticoagulation,but with no increase in bleeding complications.8

GENERATOR CHANGES, IMPLANTABLECARDIOVERTER DEFIBRILLATORS, AND CARDIACRESYNCHRONISATION THERAPYThe elective generator change is often incorrectlyperceived as a more straightforward procedurethan the implant itself. Infection rates are higher,lead damage is possible, and unexpected findingsare not uncommon—for example, sudden changesin pacing thresholds/impedances. Some simple

steps can be taken to minimise the risk ofproblems. First is to be absolutely certain aboutthe implanted hardware (number of leads, unipolaror bipolar). Second is to check if the patient is‘‘pacing dependent’’—if there is absolutely nounderlying rhythm (often one can only programa pacemaker to a minimum of 30 beats/min) thena decision has to be made: one may try to removethe lead quickly and connect to an analyser whichis already set to pace (but this may be a nervousprocess); or a temporary pacing wire is inserted(usually from the groin) before the generatorchange which can then be removed at the end ofthe procedure with probably negligible infectionrisk; or one can use high dose isoprenaline infusionsto try to encourage underlying rhythm. Third is touse fluoroscopy where necessary to help identifythe course of the leads in relation to the generator.Finally, to make the incision to remove thegenerator in a position that makes this easy (forexample, just over the most cephalic portion of thegenerator) which may be distant to the originalimplant incision. Where it is recognised beforehandthat a new lead needs to be implanted at the timeof a generator change, then performing a venogramwill confirm vein patency, and if the generator hasmigrated caudally it may even be necessary tomake two incisions—one to remove the generatorand one to gain venous access.

Implantable cardioverter-defibrillator (ICD)implantation and cardiac resynchronisation ther-apy (CRT) are seen to be more complex deviceprocedures. In fact, implantation of an ICD issimilar to a standard pacemaker, particularly nowthat ventricular leads with shock coils are of asimilar French size to standard pace/sense leads,albeit slightly heavier and stiffer. However, the factthat these patients often have structurally abnor-mal hearts (dilated and scarred) may make findinga position with adequate pacing and, moreimportantly, sensing characteristics difficult. Alsothe generators are larger, and the position and sizeof the pocket needs to be adjusted accordingly.CRT similarly shares a number of commoncomponents with standard bradycardia pacing,but placement of a lead to pace the left ventriclerequires training and experience.

SUMMARYPPM implantation remains a core skill of traineecardiologists, despite increasing subspecialisation.This two part article is aimed at giving theinexperienced implanter a framework on whichto develop their skills. There are now also anumber of courses available that offer the oppor-tunity to learn about PPM implantation in acontrolled teaching environment, many usingvirtual technology for components such as leadplacement. Ultimately the technique for implanta-tion can be best learnt in a pacing theatre/laboratory under the guidance of an experiencedimplanter.

Permanent pacemaker implantation II: key points

c When positioning a lead, be careful not to introduce blood, etc, into the centrallumen—eventually the lumen will occlude and make passage of the styletimpossible.

c The presence of far field sensing can be problematic and is best dealt with atthe time of implant by careful lead positioning.

c Once the leads are in position, use fluoroscopy intermittently while securingthem to ensure they do not move. This includes briefly screening withfluoroscopy after the generator is in the pocket.

c The use of peri-procedure antibiotics and anticoagulation management varieswidely—there are often local protocols and it is important to be familiar withthese.

You can get CPD/CME credits for Education in Heart

Education in Heart articles are accredited by both the UK Royal College ofPhysicians (London) and the European Board for Accreditation in Cardiology—you need to answer the accompanying multiple choice questions (MCQs). Toaccess the questions, click on BMJ Learning: Take this module on BMJLearning from the content box at the top right and bottom left of the onlinearticle. For more information please go to: http://heart.bmj.com/misc/education.dtlc RCP credits: Log your activity in your CPD diary online (http://www.

rcplondon.ac.uk/members/CPDdiary/index.asp)—pass mark is 80%.c EBAC credits: Print out and retain the BMJ Learning certificate once you have

completed the MCQs—pass mark is 60%. EBAC/ EACCME Credits can now beconverted to AMA PRA Category 1 CME Credits and are recognised by allNational Accreditation Authorities in Europe (http://www.ebac-cme.org/newsite/?hit = men02).

Please note: The MCQs are hosted on BMJ Learning—the best available learningwebsite for medical professionals from the BMJ Group. If prompted, subscribersmust sign into Heart with their journal’s username and password. All users mustalso complete a one-time registration on BMJ Learning and subsequently log in(with a BMJ Learning username and password) on every visit.

Education in Heart

Heart 2009;95:334–342. doi:10.1136/hrt.2008.156372 341

on 16 June 2009 heart.bmj.comDownloaded from

Acknowledgements: My thanks to Dr Ernest Lau for allowing meto use his line drawings on lead placement. Thanks also to DrsJoseph De Bono, John Paisey, David Tomlinson, and Zaheer Youseffor their contributions.

Competing interests: In compliance with EBAC/EACCME guide-lines, all authors participating in Education in Heart have disclosedpotential conflicts of interest that might cause a bias in the article.The author has no competing interests.

REFERENCES1. Stambler BS, Ellenbogen KA, Zhang X, et al. Right ventricular

outflow versus apical pacing in pacemaker patients with congestiveheart failure and atrial fibrillation. J Cardiovasc Electrophysiol2003;14:1180–86.

2. de Cock CC, Meyer A, Kamp O, et al. Hemodynamic benefitsof right ventricular outflow tract pacing: comparison with rightventricular apex pacing. Pacing Clin Electrophysiol1998;21:536–41.

3. Victor F, Leclercq C, Mabo P, et al. Optimal right ventricular pacingsite in chronically implanted patients: a prospective randomizedcrossover comparison of apical and outflow tract pacing. J Am CollCardiol 1999;33:311–16.

4. Ellenbogen KA, Wood MA. Cardiac pacing & ICDs. BlackwellScience, 2008:49–58.

5. Haywood GA, Jones SM, Camm AJ, Ward DE. Day casepermanent pacing. Pacing Clin Electrophysiol 1991;14:773–7.

6. Dunn AS, Turpie AGG. Perioperative management of patientsreceiving oral anticoagulants: a systematic review. Arch Intern Med2003;163:901–8.

7. Michaud GF, Pelosi Jr F, Noble MD, et al. A randomized trialcomparing heparin initiation 6 h or 24 h after pacemaker ordefibrillator implantation. J Am Coll Cardiol 2000;35:1915–18.

8. Goldstein DJ, Losquadro W, Spotnitz HM. Outpatient pacemakerprocedures in orally anticoagulated patients. Pacing ClinElectrophysiol 1998;21:1730–4.

Useful reference books

c Ellenbogen KA, Wood MA, eds. Cardiac pacing & ICDs, 5th ed. BlackwellScience Inc, 2008.– the most widely used concise textbook for device therapy.

c Timperley J, Leeson P, Mitchell ARJ, Betts T, eds. Pacemakers and ICDs.Oxford specialist handbooks in cardiology. Oxford University Press, 2008– excellent pocket sized book for quick reference on all aspects of pacing.

c Barold SS, Stroobandt RX, Sinnaeve AF, eds. Cardiac pacemakers step bystep. An illustrated guide. Futura, Blackwell Publishing, 2005– the basics of pacing explained in a clear, illustrative format.

Education in Heart