On-line Analysis – A Water Company Perspective Peter Boruszenko – R&D Engineer John Haley – Water Quality Compliance Manager K. Clive Thompson ALcontrol Laboratories
Transcript
Page 1: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

On-line Analysis – A Water CompanyPerspective

Peter Boruszenko – R&D Engineer

John Haley – Water Quality Compliance Manager

K. Clive Thompson

ALcontrol Laboratories

Page 2: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Introduction

• Why is on-line monitoring important to YW?

• 3 Case studies - Online Monitoring for:-

– Raw Water

– Water Treatment

– Distribution

Page 3: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

RiverDerwent

Sheffield

ElvingtonArthington

Reservoirs 45% Groundwater 22%

RiverSwale Wiske

RiverWharfe

RiverUre

RiverNidd

RiverOuse

Hull

DoncasterBarnsley

LeedsHull

Scarborough

Bridlington

Richmond

York

Aire

Eccup

Moor Monkton

Elvington

Wakefield

Halifax

Huddersfield

Lobwood

BradfordLoftsome

Rivers 33%

Page 4: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Why on-line analysis?

• Raw waters are variable

• Water treatment is rarely steady-state

• Best possible knowledge of risks to water quality

• Optimise treatment

• Manage risky situations

• Improved customer service

Page 5: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 6: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

• There are over 15 million known organicsubstances.

•Robust technique to detect all ofthese? Impossible !

• Compromise, pragmatism robustness,costs are key

• Not much TLC available – OPEX limited!

CASE STUDY 1: RAW WATER MONITORING

Page 7: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

On-line systems for river intake protection(11 sites) in YWS – position in 1986

• The Worcester & Dee Incidents made intake protection high profile

• Avoid supplying Water unfit for human consumption- Due diligence defence

• Some companies have opted for highly intensive on-line analysis oftenspecifically directed on the more potentially polluted rivers

• YW has relatively unpolluted rivers – more widespread use of ‘broad- band’monitors because we do not have known fixed contaminant risk!

• UV absorption detects a wide range of chemicals but not very sensitive in1986 essentially detecting gross contamination.

• YW therefore developed the UV based system from firstprinciples

Page 8: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Old YWS Uvikon with “Sand Pre-filter”

Page 9: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

On-line systems for river intake protection(11 sites) in YWS – position in 2002

• Decision made to re-examine systems for on-line monitoring from first

principles

• Options considered and conclusion that broad screen monitoring most

appropriate for YW rivers due to no specific risks.

• Some very expensive options on the market but no UV systems

• Decided that UV detection was the best process and to persuade a

commercial supplier to develop a UV system with modern hardware and

software

Page 10: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

The PMS system

Page 11: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Specord S100 spectrophotometer (4cm cell)Asulam standard in Lobwood river

(RO water baseline)

0.30

0.50

0.70

0.90

1.10

1.30

1.50

1.70

1.90

230 240 250 260 270 280 290 300 310 320

nm

Abs

.

100ppb 250ppb 500ppb 1ppm 2ppm 5ppm lobw ood

Page 12: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

On-line Monitors

Key criteria

• Robust

• Low false positives

• Very low false negatives

• Sample pre-treatment for raw waters is key (This is the ‘Achilles

heel’ of many commercial systems)

• Minimum preventative maintenance frequency: 1 visit / Month

KEY ISSUE

• If there’s no aromatic ring then it will not work.

Page 13: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

CASE STUDY2: WATER TREATMENTCoagulation optimisation using on-lineinstrumentation

Page 14: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 15: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 16: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 17: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 18: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 19: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 20: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Coagulation in Brief

• Coagulant metal hydroxide precipitates to give floc.• Coagulant only does this efficiently in a narrow pH band• Colour particles stick to the precipitating coagulant hydrolysis

products in the flash mixer (takes seconds)• If there is not enough coagulant the colour particles give the

precipitate a negative charge– Compromises optimum flocculation (small particles)– Compromises optimum filtration because charged particles

won’t stick to filter media at any time during the filter run.• Overdosing of coagulant wastes money

– Increases coagulant costs– Increases sludge production

Page 21: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Simplified Treatment Process

Mixers ClarifiersRapid Gravity

Filters

Up to 95% flocremoval

Remaining FlocRemoved

Add

• Aluminium or Ferric Sulphate(coagulant)

• Lime or Acid to control the pH of themixture

Floc production

Products

Potable Water (after disinfection)

Sludge

Page 22: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Time

Baseline

This distance is an indicator ofcoagulation optimisation.

Changing the coagulant dose (for agiven raw water quality) changesthis value.

Changing from optimumcoagulation pH range to a differentcoagulation pH increases thisvalue.

Backwash Spike

Wider and higher spike indicateslesser degree of coagulationoptimisation.

Breakthrough

Onset of break-through caused byoverloading filter with small weak floc.Usually occurs in winter/spring at flotationsites.

Made worse by increasing filtered water flow,increasing pre-filt turbidity and by flow surgesthrough the filter.

Observed relationships between filter outlet turbiditytrend and degree of coagulation optimisation

BASELINE

BREAKTHROUGH

BACKWASHSPIKET

urbi

dity

Page 23: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Raw water colour coagulant dose and filteredturbidity

Works D

0

10

20

30

40

50

60

23-Jun 23-Jun 24-Jun 24-Jun 25-Jun 25-Jun 26-Jun 26-Jun 27-Jun 27-Jun

UV

abso

rban

ce(a

bs/m

)

0

0.1

0.2

0.3

0.4

0.5

0.6

Tu

rbid

ityan

dm

g/l

Al/

10

Raw Water UV Coagulant Dose (mg/l)/10 Filter 5 Turbidity Filter 8Turbidity Filter 14 Turbidity

Increase in coagulant dose causesfiltered water turbidity to decrease

Raw water UV increase causesfiltered water turbidity increase

Page 24: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Example ACC Algorithm

0

1

2

3

4

5

6

7

8

9

0 10 20 30 40 50 60 70 80 90

Raw Water UV (abs/m)

mg

/lA

l

0 5 10 15 20

Temp

Page 25: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Automatic Coagulation Control Schematic (ACC).

Raw UV 254Monitor.

Sam

ple

Works first stageRaw water inlet

DosedRawFlashMixer

Inletflowmeter

CoagulantPump

Flowmeter

pHCorrection

PLC

User Interface

Calculates predictedcoagulant dose (mg/l).Converts to coagulant flowsetpoint.

SCADA

Feedback control ofcoagulant flow.

Page 26: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 27: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 28: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90
Page 29: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

17/02/200500:00

22/02/200500:00

27/02/200500:00

04/03/200500:00

09/03/200500:00

14/03/200500:00

19/03/200500:00

24/03/200500:00

29/03/200500:00

03/04/200500:00

08/04/200500:00

Raw

UV

(mg/

lC)/

100

and

filte

red

turb

idity

(FTU

).

0.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

Co

agu

alan

tD

ose

s(m

g/l

Al),

Raw

Tu

rbid

ity/

10(F

TU

)an

dT

emp

Filter 3 Outlet Turbidity Filter 1 Outlet Turbidity Raw UV/100 Actual Coagulant Dose

Predicted Coagulant Dose Temperature Dosed raw pH Spsheet Predicted Dose

ACC Example – River Works (Site F)

Page 30: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Conclusions

• Reduces the need for manual intervention

• Reduces overdosing of Coagulant

• (1500 Tonnes in 11 months)

• Reduces out-of-hours call-outs

• Maintains optimum treated water turbidity

• Minimises risk of Cryptosporidium breakthrough

The ACC System:-

Page 31: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Real Time MonitoringCASE STUDY 3 - DISTRIBUTION

Page 32: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

ImpoundingReservoir

Borehole Raw WaterPumping

RiverAbstraction

WaterTreatment

TreatedWaterStorage

Grid TreatedWaterPumping

DistributionManagement

Areas

Monitoring, Controlling & Optimising

Internal Work Mngt and Scheduling

Real Time Asset Monitoring, Control & Optimisation

Remote Asset Management & Technical Support

Variable Approach to Asset Maintenance

Reactive Customer Service

MONITOR CONTROL OPTIMISE

SourceSource TapTap

Page 33: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

RTnet Drivers

1. Reduce/Remove manual data collection

2. Leakage data on a daily basis

3. Receive customer service data every half hour

‘Respond to failure before the customer is impacted’

Page 34: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Key Issues

• Communications• Battery life

– Currently only 2.5 years

– Power harvesting?

• Managing the data• COST !!!

Page 35: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Future Issues

• Rtnet only measure at zone inlets…– How do we measure rest of the network

• Measure real time water quality:– In distribution– At service reservoirs

• Multi-functioning device– Water quality, pressure, flow– CENSAR

Page 36: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

This is very heavy –a two man operation– also attenuatessignals

Page 37: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

Valving Operations at Daisy Hill - Detail of Initial turns

0

0.5

1

1.5

2

2.5

3

9:47:31 9:50:24 9:53:17 9:56:10 9:59:02 10:01:55 10:04:48 10:07:41 10:10:34 10:13:26 10:16:190

5

10

15

20

25

30

35

40

45

Turbidity Valve turns

Initial Closure Turbidity Effects

Page 38: PB JH On-line Analysis A Water Company PersSpecord S100 spectrophotometer (4cm cell) Asulam standard in Lobwood river (RO water baseline) 0.30 0.50 0.70 0.90 1.10 1.30 1.50 1.70 1.90

CONCLUSIONS

• There is a widespread move away from manualintervention in water treatment process control

• 20 new sites require robust reliable pollution monitors –various new systems are being considered.

• At least one month unattended operation is essential

• Enhanced customer service

• Attain top position in the OFWAT league table !