2014 Copyright Peak Scientific - Rev. 5 Peak Scientific Gas Generators OEM Application List
Transcript
Page 1: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

2014 Copyright Peak Scientific - Rev. 5

Peak Scientific Gas Generators OEM Application List

Page 2: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 2

Contents

AB Sciex 3

Agilent 5

Biotage 7

Bruker 8

Dionex 9

Jasco 10

Jeol 11

Leco 12

Perkin Elmer 13

Shimadzu 14

Thermo Scientific 16

Varian 18

Waters 19

All other applications 21

Sample Evaporation/Concentration/ SPE 25

Multiple MC/MS selection guide 27

Gas Chromatographygeneral application guide 28

Page 3: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 3

Application Requirements Solution

AB SCIEX API Q-STAR

Curtain + Vent gas 4 L/min; 60 psi maxSource 12L/min, 105 psi maxExhaust 8L/min 50 psi

CQ-Star w/o MALDI:CAD gas; 1 L/min 60 psi

Infinity 1031Genius ABN2ZAGenius AB- 3G

AB SCIEX API Q-TRAPCurtain 0.5 – 3 L/min; 50 psi maxSource 12L/min, 90 to 100 psigExhaust 8L/min 90 to 100 psig

Infinity 1031Genius ABN2ZAGenius AB- 3G

AB SCIEX API 100/ 150/ 165Curtain 2 L/min; Source 13L/min, Exhaust 4L/min

Infinity 1031Genius ABN2ZA Genius AB- 3G

AB SCIEX API 300/ 365Curtain 2 L/min; Source 15L/minExhaust 4L/min

Infinity 1031Genius ABN2ZAGenius AB- 3G

AB SCIEX API 2000 LC/MS/MSCurtain 0.5 - 3 L/min; 50 psigSource 12L/min, 90 to 100 psigExhaust 8L/min 50 psig

Infinity 1031Genius ABN2ZA Genius AB- 3G

AB SCIEX API 3000 LC/MS/MS

Curtain 3.5 L/min; 60 psiSource 12 L/min, 90 psiExhaust 8 L/min 50 psi

Infinity 1031Genius ABN2ZA Genius AB- 3G

AB SCIEX API 3200 LC/MS/MS 3200 QTRAP

Curtain 5 L/min; 60 psi (max.)Source 22 L/min, 100–105 psiExhaust 6.4 L/min 55 (min.)–60 psi (max.)

Infinity 1031Genius ABN2ZA Genius AB- 3G

AB SCIEX API 4000 LC/MS/MS 4000 QTRAP

Curtain 5 L/min; 60 psi maxSource 17L/min, 90–105 psiExhaust 6L/min 55 max

Infinity 1031Genius ABN2ZA Genius AB- 3G

AB SCIEX API 4500 LC/MS/MS 4500 QTRAP

Curtain 10 L/min; 60 psi maxSource 22L/min, 100 psi maxExhaust 8L/min 55 max

Infinity 1031Genius ABN2ZAGenius AB- 3G

AB SCIEX API 5000 LC/MS/MSCurtain 10 L/min; 60 psiSource 22.5 L/min, 55–60 psi (max.)Exhaust 6 L/min 55–60 psi (max.)

Infinity 1031Genius ABN2ZAGenius AB- 3G

AB SCIEX API 5500 LC/MS/MS API 5500 QTRAP

*Without SELEXION

Curtain 10 L/min; 60 psi maxSource 22 L/min, 100 psi maxExhaust 8 L/min 55 psi minimum

Infinity 1031Genius ABN2ZAGenius AB-3G

AB SCIEX API 5500 LC/MS/MS API 5500 QTRAP

*With SELEXION

Curtain 13 L/min; 60 psi maxSource 22 L/min, 100 psi maxExhaust 17 L/min 55 psi minimum

Infinity 1031Genius 3031

AB SCIEX Triple TOF 5600 Triple TOF 5600 +

Curtain 10 L/min; 60 psi Source 22.5 L/min, 105 psiExhaust 8 L/min 55 psi to 105 psi

CAD gas uhp N2@ 1ml/min, 60psi

Infinity 1031Genius ABN2ZAGenius AB-3G

AB SCIEX Triple TOF 5600 + *With SELEXION

TBA (No info from AB on this yet)

Infinity 1031Genius ABN2ZA, Genius AB-3G?Genius 3031?

AB SCIEX API 6500 LC/MS/MS API 6500 QTRAP

*With or without SELEXION

Curtain 18 L/min; 80 psi maxSource 26 L/min, 100 psi maxExhaust 25 L/min 70 psi minimum

Infinity 1030Genius 3031

Page 4: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 4

Application Requirements Solution

API w/ Photospray Application Same flow rates and pressures as above equivalent models.

Infinity 1031Genius ABN2ZAGenius AB-3GGenius 3031?

Note:For previous AB Sciex models up to and including API5000, AB would state in their manuals that Zero-Air OR Nitrogen can be used for SOURCE GAS. This meant that you could operate an API LC/MS/MS with a single nitrogen-only gas source.

For the newest AB Sciex models API5500, API4500 and the Triple TOF 5600, AB SCIEX recommend only Zero Grade Air to be used for SOURCE GAS (Gas 1/Gas 2) when operating a Nanospray Ion Source. When using UHP N2 with this ion source, there is an increased risk of corona discharge, which can damage the emitter tip. However; if customer is using the DuoSpray™ or Turbo V™ or PhotoSpray ion source, they can still use UHP N2 for Gas 1 and Gas 2.

AB Sciex continued

Page 5: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 5

Application Requirements Solution

Agilent 1100 Series LC/MSD Nitrogen up to 15L/min, 80 - 100 psi( At least 3L/min is required at all times to prevent air entering the instrument)

Infinity NM18LInfinity NM32LGenius 1050Genius NM32LA,Genius NM3G

Agilent 1200 Series HPLC with ELS Detector

Nitrogen up to 4L/min, 50 psi( for G4218A ELSD)

Infinity NM18LGenius N118LANG4000A

Agilent 385 ELSD Nitrogen 0.9 - 3.25 L/min @ 60 - 100 psi Infinity NM18L Genius N118LA NG3000A

Agilent 6100 SeriesNitrogen up to 15L/min, 80 – 100 psi( At least 3L/min is required at all timesto prevent air entering the instrument)

Infinity NM18LInfinity NM32LGenius 1050Genius NM32LA,Genius NM3G

Agilent 6200 Series Nitrogen up to 18L/min, 80 – 100 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Agilent 6300 Series Ion Trap LCMS Nitrogen up to 18L/min, 80 – 100 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Agilent 6400 Series6410 / 6420/ 6430

Nitrogen up to 18L/min, 80 – 100 psi( At least 3L/min is required at all times to prevent air entering the instrument)

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Agilent 6460Nitrogen up to 30L/min, 80 – 100 psi( At least 3L/min is required at all times to prevent air entering the instrument)

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Agilent 6490 LCMS QQQiFunnel

Nitrogen up to 50L/min, 80 – 100 psi( At least 9L/min is required at all times to prevent air entering the instrument)

Genius 3010Infinity 5010

Agilent CHIP CUBE

Chip Cube is available on all 6000 series LC/MS models.-------------------------------------

CHIP CUBE on iFunnel models 6490 and 6550

Up to 30 L/min of nitrogen standard+ Requires 4 L/min of air to be mixed in with the nitrogen supply to achieve 10% oxygen content for Chip Cube operation--------------------------------------------------------Up to 60 L/min of nitrogen standard.+ Requires 10 L/min of air to be mixed in with the nitrogen supply to achieve 10% oxygen content for Chip Cube operation

Generator models with Switch for Chip-Cube operation:Genius 1023 Genius 3023 (2 x)-----------------------------------

Genius 3013

Agilent 6500 Series QTOF LC/MSNitrogen up to 20 L/min for 6510, 6520, 6538Nitrogen up to 30 L/min for 6530 and 6540( At least 3L/min is required at all timesto prevent air entering the instrument)

Infinity NM32L Genius NM32LAGenius 1050Genius NM3G

6550 iFunnel Q-TOF LC/MSNitrogen up to 60 L/min for 6550( At least 12L/min is required at all timesto prevent air entering the instrument)

Infinity 5010 Genius 3010

Agilent models available with Jet Stream technology: Nitrogen 30 L/min @ 100 psi

Infinity NM32LGenius NM32LAGenius 1050 Genius NM3G

Page 6: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 6

Agilent continued

Application Requirements Solution

Agilent 4100 MP-AESMicrowave Plasma-Atomic Emis-sion Spectrometer

Nitrogen for Plasma Operation:22.5 L/min at 72.5 psi ( 99.5%+ )

Nitrogen for Optics purge:10 L/min at 65 psi (99.95%+)

Clean, Dry, Oil-free Compressed Air for:Preoptics protection: 25 L/min at 72.5 psiOptional ECGM: 1.5 L/min at 72.5 psi

To be advised.

(Agilent offer own-brand 4107 PSA Nitrogen gen-erator at the moment)

Agilent 70 series ICP-OESHigh Purity Nitrogen 0.45 L/min in standby modeUp to 2.1 L/min in operational mode

NG Range

Agilent 7500 ICP-MS

Argon Gas SupplyMinimum Purity: 99.99%Maximum Flow Rate: 20 L/minSupply Pressure: 700 kPa

Reaction Gas Supply (Agilent 7500cs, 7500cx)Gas: Helium, Hydrogen (optional for 7500cx)Minimum Purity: 99.999%MaxFlow Rate: He 10 mL/min, H2 7 mL/minTypical Pressure: 40 kPa

Precision Hydrogen 100cc

Agilent Cary Eclipse FluorescenceSpectrophotometer

99.999% pure nitrogen at 10 psiFlow rate 0 – 30 L/min

High Flow PSA systemi-Flow 6010

Agilent Cary 600 Series FTIR spectrometer

10 L/min of clean, dry air or nitrogen < 60 psiDewpoint: - 70C / -94FShould be CO2 free

PG14L

Agilent Cary 610/620 FTIR Microscope

10 L/min of clean, dry air or nitrogen <60 psiDewpoint: - 70C / -94FShould be CO2 free

PG14L

Agilent MP-AES 10L/min of Nitrogen 99.5% Purity36.5L.min of Dry Air

Genius 3055

Page 7: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 7

Application Requirements Solution

Biotage TURBOVAP LV(CALIPER / ZYMARK)

At 10 psi: 1.7 L/min/ per nozzleAt 15/20 psi: 2.83 L/min/ per nozzle

Max gas usage for total of 50 nozzles: 140 L/minInlet gas pressure min. 60 psi/max 100 psi

Infinity 90*Infinity 50**

Biotage TURBOVAP II(CALIPER / ZYMARK)

At 11 psi: 2.3 L/min/nozzleInlet gas pressure should be min. 30 psi/max 80 psi

Infinity 90**

Biotage TURBOVAP 96 Up to 45 L/min at 70psi min

Infinity 90*Infinity 50** RangeInfinity NM45LGenius 3010

Biotage TURBOVAP 500 No Gas required N/A

Biotage ISOLERA ELSD-1080 0.9 L/min to 3.25 L/min @ 60 – 100 psi

Infinity NM18LGenius N118LANG range

Zephyr SPE Workstation Clean dry oil-free air or nitrogen at up to 65 psiInfinity 90*Infinity 50** Range

Zymark RapidTrace SPE Clean, dry, regulated air or nitrogenMin 10 psi / Max 40 psi

Infinity 90*Infinity 50** Range

Evaporex /Biotage /ArgonautSPE Dry 96 Dual Microplate Evaporator Concentrator

Inlet Gas Pressure: Min 15psi, Max 50psiGas Flow Min: 25 L/min - Max: 60 L/minClean, dry, oil-free nitrogen

Genius 3010Infinity 90*Infinity 50** Range

Page 8: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 8

Application Requirements Solution

Bruker- MICROTOF Q MICROTOF II (Focus) MICROTOF MAXIS

Nitrogen between 5 to 12L/min at 80 -90 psiAprox 3 to 4L/min is required during standby-mode

Infinity NM18LInfinity NM32LGenius NM32LAGenius NM3G

Bruker Solarix hybrid Qq-FTMS

Nitrogen between 5 to 12L/min at 90 psiAprox 3 to 4L/min is required during standby-mode

Infinity NM18LInfinity NM32L Genius NM32LAGenius NM3G

Bruker MAXIS QQ-TOF MSNitrogen between 5 to 12L/min at 90 psiAprox 3 to 4L/min is required during standby-mode

Infinity NM18L Infinity NM32LGenius NM32LAGenius NM3G

Bruker Esquire Ion Trap MS HCT Ion Trap MS

Nitrogen between 5 to 15 L/min at 90 psiAprox 3 to 4L/min is required during standby-mode

Infinity NM18LInfinity NM32LGenius NM32LAGenius NM3G

Bruker AmaZon ion trap MSNirogen between 5 to 12 L/min at 90 psiAprox 3 to 4L/min is required during standby-mode

Infinity NM18LInfinity NM32LGenius NM32LA, Genius NM3G

Bruker Fourier Transform Ion Cyclotron Resonance(FT-ICR) mass spec-trometer APEX III

Nitrogen for both nebulizer gas and drygas for ESI (Up to 12L/min?)

Infinity NM18LInfinity NM32LGenius NM32LA, Genius NM3G

Bruker EVOQ TQ LC/MSELITE + CUBE

Nitrogen up to 32L/min @ 80 psiClean, Dry Air up to 50 L/min @ 80 psi

Genius 3045

Bruker APEX IV FTMSSpectrometer

N2 for electrospray ion source: 57 L/min @ 60-100 psi

Dry (-40F DP), clean and oil-free air or N2@ 60-80 psi for pneumatic valve operationVery small quantity, not continuous.

Infinity 5010Genius-3010

Same source as above.(Note: Not Tested

Bruker Ultraflex II TOF/TOF UHP Zero Nitrogen at 20 cc/min @ 72 psiPrecision Nitrogen Trace 250cc

Bruker Aurora ICP/MS Varian 820 MS

Hydrogen for CRI gas (Collision Reaction Interface)Up to 200 cc/min at 60 psi Typical flow rate: 60 – 90 cc/min

Hydrogen for Initialization purge cycle: 2.2 L/min

No Peak solution available.

Requires hydrogen generator with Buffer tank.

Bruker TENSOR series FT-IR Moisture and CO2-free purge gas or nitrogen at 2.5 L/min

PG Range

Bruker EMX EPR Spectrometer Moisture and O2-free purge gasUHP nitrogen at max 30 L/min

i-Flow 6010

NOTE: There are some models of Bruker LCMS that require less than 18 L/min of Nitrogen. Peak Scientific has tested these options and will NOT recommend to connect an N118LA to these applications. The N118LA is not designed to operate continuously at full capacity.

Page 9: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 9

Application Requirements Solution

DIONEX ASE-350Accelerated Solvent Extractor

Nitrogen 99.99% at 150 – 200 psi

A source of 99.9% pure nitrogen gas, regulated to between 140 and 200 psi; 150 psi is recom-mended. UHP gas may be required for apps using a very clean baseline electron capture detector (ECD).

Previously: ASE-HP-220Do not offer this anymore!

Requires inert N2 pressure booster

DIONEX ASE-350Accelerated Solvent Extractor

Nitrogen 99.99% at 150 – 200 psi

A source of 99.9% pure nitrogen gas, regulated to between 140 and 200 psi; 150 psi is recom-mended. UHP gas may be required for apps using a very clean baseline electron capture detector (ECD).

Previously: ASE-HP-220Do not offer this anymore!

Requires inert N2 pressure booster

DIONEX ASE-200Accelerated Solvent Extractor

Nitrogen 99.99% at 150 – 200 psi

A source of 99.9% pure nitrogen gas, regulated to between 140 and 200 psi; 150 psi is recom-mended. UHP gas may be required for apps using a very clean baseline electron capture detector (ECD).

Previously: ASE-HP-220Do not offer this anymore!

Requires inert N2 pressure booster

DIONEX ASE-150Accelerated Solvent Extractor

Nitrogen 99.99% at 150 – 200 psi

A source of 99.9% pure nitrogen gas, regulated to between 140 and 200 psi; 150 psi is recom-mended. UHP gas may be required for apps using a very clean baseline electron capture detector (ECD).

Previously: ASE-HP-220Do not offer this anymore!

Requires inert N2 pressure booster

DIONEX ESA CORONA CADCharged Aerosol Detector 99% Nitrogen up to 5 L/min @35 psi

Infinity NM18LGenius N118LANG5000NG5000A

Dionex CORONA ULTRA RSCharged Aerosol Detector

Clean, Dry and oil-free air or nitrogen at 60 psi, 4-5 L/min

Infinity NM18LGenius N118LANG5000NG5000A

Dionex Autotrace 280 SPE

Nitrogen supply pressure up to 100 psiOperation pressure up to 20 psi

Only inert gases such as nitrogen, argon, or compressed air (free from oil) should be used for the Dry with Gas step.

Infinity 90**Infinity 50**

Dionex SE400 / SE500Solvent Evaporators

Nitrogen at 20 psiInfinity 90**Infinity 50**

Dionex ICS seriesIon Chromatography System900, 1000, 2000, 3000, 5000 series

UHP nitrogen or Helium at 25 – 50 psi for eluent blanketing.

Should be CO2-free

Note: PSA nitrogen generators not suitable because they do not remove CO2!

Page 10: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

peakscientific.comPage 10

Peak Scientific Applications List

Application Requirements Solution

JASCO J-815 CD SpectrometerCircular Dichroism

UHP, Dry Nitrogen up to 3 L/min NG Range

JASCO FT/IR-6000 seriesPurge gas CO2 and H2O-free airFlow rate unknown

PG Range

Page 11: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

peakscientific.comPage 11

Peak Scientific Applications List

Application Requirements Solution

JEOL LCMS Nitrogen up to 20 L/min

Infinity NM32L Genius 1050Genius NM32LAGenius NM3G

AccuTOF™ LC JMS T1000LC/TOF/MS Nitrogen up to 10L/min at 100 psi

Infinity NM18LInfinity NM32LGenius 1050Genius NM32LAGenius NM3G

Page 12: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

peakscientific.comPage 12

Peak Scientific Applications List

Application Requirements Solution

LECOPegasus 4D GC X GCLN2 Modulators

2 x Cold Jets : N2 at 30 psi with Dewpoint -50°. 6 L/min for each Jet – Total 12 LPM

2 x Hot Jets: clean dry air or N2 at 30 psi with Dewpoint -50°.20 L/min for each Jet – Total 40 LPM

Secondary oven cooling gas: 100 L/minClean, dry air or N2 at 30 psi(used for 5 mins after each run)

Genius 1052

N/A*

*NOTE: Above LECO Pegasus Modulation gas requirements are as stated in LECO’s manual. However; actual testing carried out by Peak at LECO’s Test-Lab in South Africa confirmed that:

1. Hot Jets actually do not consume more than 10 LPM each2. Additional 100 L/min of compressed air for oven cooling is not required for operation. It only

speeds up the system cool-down time between runs, but has no effect on instrument perfor-mance. The system automatically uses the Hot-Jet air/N2 flow for oven cooling when the Hot-jets are not in use. The 20 LPM as supplied by the Genius 1052 will suffice for this purpose.

Page 13: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

peakscientific.comPage 13

Peak Scientific Applications List

Application Requirements Solution

Perkin Elmer –LCMS range Nitrogen up to 30L/min

Infinity NM32L Genius 1050Genius NM32LAGenius NM3G

Perkin Elmer ICP OES Optima 2000 series Optima 2100 DV Optima 3000 series Optima 4000 series Optima 5000 series Optima 7000 DV Optima 7300 V Optima 8000 Optima 8300

Plasma Gas: Argon at 9 L/min

Purge gas: N2 or Argon Nitrogen purge gas up to 5L/min, 50 psi99.99% – 99.999% purity(Continuous flow of 1 L/min required)

Shear gas: Clean, dry and oil-free compressed air or N2 at 25 L/min , 80 psi

NG5000*NG5000A*NG10L-HP

House Air or small oil-free compressor with dryer and filtration.

Infinity NM32LGenius 1050Genius NM32LA

Perkin ElmerNexION 300 ICP-MS 14 – 20L min Argon Gas only N/A

Perkin Elmer Flexar SQ 300 MS Nitrogen 98% @ 30 L/min 100 psi

Infinity NM32LGenius 1050Genius NM32LAGenius NM3G

Perkin Elmer GC

Carrier Gas: 60 ml/min @ 90 psi(N2 for packed columns)ECD makeup gas: N2 > 30 ml/min @ 30 psiFID flame gas: H2 > 50 ml/min @ 30 psiFID zero air:: ZA > 500 ml/min @ 30 psi NPD: H2 > 3 ml/min @ 30 psiNPD: ZA > 90 ml/min @ 30 psi

Precision Hydrogen TracePrecision Nitrogen 250ccPrecision Zero Air 1.5LFusion 1010

Perkin Elmer Clarus GC 400-480

Carrier Gas: H2 or N2 30 ml/min 90 psiFID Air: 450 ml/min @ 30 psiFID H2: 45 ml/min @ 30 psiECD make up gas: 60 – 100 mL/min

The ECD is shipped with charcoal traps (P/N N6600037) installed in the make-up and injector pneumatics lines to remove contamina-tion from the needle valve, flow controller, or pressure regulator. The traps should be replaced periodically

Precision Hydrogen TracePrecision Nitrogen 250ccPrecision Zero Air 1.5LFusion 1010

Perkin Elmer Clarus GC 500

Carrier Gas: H2 or N2 30 ml/min 90 psiFID Air: 450 ml/min @ 30 psiFID H2: 45 ml/min @ 30 psiECD make up gas: 60 – 100 mL/min

The ECD is shipped with charcoal traps (P/N N6600037) installed in the make-up and injector pneumatics lines to remove contamina-tion from the needle valve, flow controller, or pressure regulator. The traps should be replaced periodically

Precision Hydrogen TracePrecision Nitrogen 250ccPrecision Zero Air 1.5LFusion 1010

Perkin Elmer Clarus GC 680Carrier Gas: H2 or N2 30 ml/min 40 psiFID Air: 450 ml/min @ 30 psiFID H2: 45 ml/min @ 30 psi

Precision Hydrogen TracePrecision Nitrogen 250ccPrecision Zero Air 1.5LFusion 1010

Perkin Elmer Spectrum 100 Series FT-IR & FT-NIR N2 purge gas flow: 28 L/min at max 100 psi PG28L

Perkin Elmer PYRIS DSC & TGA instruments Purge Gas 99.9% + nitrogen up to 65 psi Precision Nitrogen 250cc

Page 14: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 14

Application Requirements Solution

Shimadzu LCMS2010EV Nitrogen at 12 L/min standard rateNitrogen up to 22.5L/min, 110 psi

Infinity NM32LInfinity 5010Genius NM32LAGenius 1050Genius 3020

Please note: some Shimadzu LCMS Instruments such as the LCMS2010EV require 110 psi input pressure for optimum operation of the instruments gas flow controllers. 100 – 116 psi range is recommended by Shimadzu as acceptable nitrogen operating pressure, but at the lower end of the scale (100 psi) the flow controls will be slower and less effective, hence the preference for 110 psi.

The Genius1022 or the NM32L can be offered for Shimadzu LC/MS; however note that NM32L will need to be con-nected to an external compressed air supply with a minimum outlet pressure of 125psi in order to increase nitrogen output pressure up to 110psi.

Shimadzu LCMS IT TOFNitrogen up to 21.5L/minPressure between 100 - 116 psi

Infinity NM32LInfinity 5010Genius NM32LAGenius 3020

Shimadzu LCMS-2010ANitrogen up to 30L/minPressure between 100 - 116 psi

Shimadzu LCMS-2020Nitrogen up to 21.5 L/min,Pressure between 100 - 116 psi

Shimadzu LCMS-8030 Nitrogen up to 25L/min

Shimadzu LCMS-8040 Nitrogen up to 25L/min

Shimadzu LCMS-8080Nitrogen at 15 L/min (Curtin & Collision gas)Dry Air (nebuliser gas) at 34 L/minDry Air (heating gas) 31l L/min

Shimadzu GC2010

Carrier Gas: Helium only

FID Fuel Gas: Hydrogen 40ml/ minFID Air: Zero Air 400ml/ min

ECD – nitrogen for Make-up gas

N/A

Precision Hydrogen 100ccPrecision Zero Air 1.5L

Precision Nitrogen Trace 250ccFUSION 1010

Shimadzu GCMS-QP2010 Carrier Gas: Helium only N/A

Shimadzu GCMS-TQ8030 Carrier Gas: Helium only N/A

Shimadzu GC2014

Carrier Gas: Helium only

FID Fuel Gas: Hydrogen 40ml/ minFID Air: Zero Air 400ml/ min

ECD – nitrogen for Make-up gas

N/A

Precision Hydrogen 100ccPrecision Zero Air 1.5L

Precision Nitrogen Trace 250ccFUSION 1010

Shimadzu TOCVcs/Vcp/Vw/VE TOC Gas up to 500cc/min

SCTOCATOC1500HP

Shimadzu TOC-LTOC-L H Type High Sensi-tivityTotal Organic Carbon Analyzer

Ultra Pure Air air at 28-31 psig •CO, CO2, HC content: Each 1 ppm max.•dew point: -50 °C max.

Gas consumption: (Variable flow rate)150 mL/min (230 to 250 mL/min during sparging)

SCTOCA / TOC1500HP

Page 15: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 15

Application Requirements Solution

Shimadzu TOC-LTOC-L N-Type Regular SensitivityTotal Organic Carbon Analyzer

Ultra Pure Air air at 28-31 psig •CO, CO2, HC content: Each 1 ppm max.•dew point: -50 °C max.

Gas consumption: (Variable flow rate)230 mL/min (A separate 100 mL/min is required for sparging with ASI-L)

SCTOCA / TOC1500HP

Shimadzu TNM-L Total Nitrogen Analyzer

For Ozone Source Gas:Compressed air /or house air between28-31 psig and flow at 500mL/min

* Can also be ran w/ Ultra Pure Air for TOC-L s

House airTOC1500TOC1500HPSCTOCA

Shimadzu SSM-5000A Solid Sample Combustion Unit

99.9 % O2 at 500 mL/minHigh-purity O2 gas is required forhigh-sensitivity measurement.

N/A

Shimadzu ELSD LT-II N2 or Air at 3 L/min, max 65 psiInfinity NM18LGenius N118LANG3000A

Shimadzu UV VIS NIR 99.999% nitrogen 50 – 100 L/minHigh Flow PSA systemi-Flow 60**

Shimadzu 7510 ICPSSequential Plasma Emission Spectrometer

Argon gas of >99.95% purity at 50 psiCoolant gas 2 to 20 L/minPlasma gas 0.2 to 1.4 L/min Carrier gas 0.1 to 1.5 L/minPurge gas 3.5 L/min

N/A

Shimadzu IR prestige 21 FTIR

Ultra dry CO2-free Air for puge gas PG Range

Page 16: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 16

Application Requirements Solution

Thermo** LTQ Range Nitrogen up to 21 L/min at 100 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Thermo Quest LCQ DECA XP + Nitrogen up to 30 L/min at 100 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

THERMO LTQ VELOS ETD VELOS PRO ETD LTQ XL ETD

Nitrogen up to 15 L/min 100 psi

ETD option: 1cc/min 99.999% N2 @ 40 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

(separate bottled gas supply)

THERMO LTQ ORBITRAP VELOS With ETD Module(Electron Transfer Dissociation)

Nitrogen for the API source and C-Trap:99.5% pure, flow rate 15 L /min at 100 ± 20 psi

Nitrogen supply for ETD option: UHP 99.999% w/less than 1 ppm H2O & O2.Pressure at:40 ±5 psi

ETD module consumes less than 1 ATM cc/m of reagent carrier gas, so a standard largebottle (245 ft3) of gas should last approximately two years.

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

(separate bottled gas supply)NG Range is only a cost- ef-fective option if UHP N2 is used for other apps, besides just ETD.

Thermo LTQ SeriesLTQ XL / LTQ Orbitrap Velos

Nitrogen up to 21 L/min 100 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Thermo Finnigan TSQ QuantumTSQ Quantum UltraTSQ Quantum Discovery

Nitrogen up to 21 L/min at 100 psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Thermo Exactive Nitrogen up to 30 L/min at 100 psi +- 10 psi NM32L, NM32LA, NM3G

Thermo Q- ExactiveThermo Q- Exactive Plus Nitrogen up to 30 L/min at 116 psi Genius 1022

Thermo LXQ Linear Nitrogen up to 15 L/min, up to 100psi

Infinity NM32LGenius NM32LAGenius 1050Genius NM3G

Thermo LCMS with HESI or HESI-II probe attachment.

Typical settings, as recommended by Thermo do not exceed 25 L/min, however, at max gas settings with HESI probe, nitrogen flow requirement was measured up to 40L/min, 100 psi. Please note; Thermo state that these settings are unlikely to be used for real analysis as they are not optimum and would degrade the signal for any given sample. Verify with customer.

Infinity NM32LInfinity NM45LInfinity 5010Genius NM32LAGenius 1050Genius NM3G

Thermo/ Dionex*** MSQ+ Nitrogen up to 15L/min, 100 psiInfinty NM18LGenius N118LAGenius N418LA

Thermo iCAP 6500

Argon Torch Gas at 15 L/min

Nitrogen Purge gas at 10 L/min, 80 psi 99.995% purity

N/A

NG10L-HP

Page 17: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 17

Thermo Nicolet 380 FTIR 4700 FTIR 5700 FTIR 6700 FTIR 8700 FTIR

Standard: Purge gas 14 L/min at 40 psi

Instrument with Smart Purge option:Purge gas up to 50L/min at 80 psi

PG Range

Thermo NicoletNEXUS 870 FTIRThermo NicoletNEXUS 670 FTIR

Standard: Purge gas 14 L/min at 40 psi

Instrument with Smart Purge option:Purge gas up to 50L/min at 80 psi

PG Range

Thermo Nicolet AVATAR FTIR

Standard: Purge gas 7 L/min at 20 psi

Instrument with Smart Purge option:Purge gas up to 50L/min at 80 psi

PG Range

Thermo Nicolet 560 Magna FTIRThermo Nicolet 760 Magna FTIR

Standard: Purge gas 12 L/min at 40 psi

Instrument with Smart Purge option:Purge gas up to 50L/min at 80 psi

PG Range

Thermo Nicolet iS50FT-IR Spectrometer

Standard: Purge gas 14L/min at 40 psi

Instrument with Smart Purge option:Purge gas up to 50L/min at 80psi

PG Range

Thermo 51i Total HydroCarbon Analyzer with FID

Makeup Air Flow Rate: 150 - 300 cc/min Hydrocarbon free air

Fuel Flow Rate:10 to 35 cc/min Hydrogen

Precision Zero Air 1.5L

Thermo 55CHydrocarbon Analyzer

Support Gas Pressures Zero Air: 20 - 30 PSIHydrogen: 20 - 30 PSI (hydrogen)UHP Zero Nitrogen: 25 - 35 PSI

HC Free Air (200 - 300 cc/min)N2 carrier (35 cc/min)H2 Fuel (25 cc/min)Note: additional pressure regulator required!

Fusion 1010

Precision Nitrogen Trace 250cc

Precision Zero Air 1.5L

Thermo Trace GC/FID

Carrier gas:He, H2 or N2Capillary Column: Typically 2ml/minPacked Column: 40ml.min

Make up gas:N2 or He10-60ml.min / Typically 30ml.min(only used with capillary columns)

Detector gas:H2 for fuel gas 30-50ml/minZA for flame support 300-600ml/min

Precision Hydrogen Trace

Precision Nitrogen Trace 600cc

Precision Zero Air 1.5L

*** Thermo/ Dionex MSQ: • N118LA model number is used for general sales• N418LA model number is ONLY used for direct sales to Thermo/ Dionex. • N418LA is rebadged by Dionex to MSQ18LA

Page 18: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 18

Application Requirements Solution

Varian 300 MS (GC/MS/MS) Helium only for carrier gas N/A

Varian 320 MS (GC/MS) Helium only for carrier gas N/A

Varian 310 MS (LC/MS)Nitrogen 12 L/min 80 psiAir 2 L/min 80 psi

Genius 1053Precision Zero Air 3.5L

Varian 320 MS (LC/MS)Nitrogen 12 L/min 80 psiAir 2 L/min 80 psi

Genius 1053Precision Zero Air 3.5L

Varian 325 MS (LC/MS)Nitrogen 12 L/min 80 psiAir 2 L/min 80 psi

Genius 1053Precision Zero Air 3.5L

Varian 500 MS (LC Ion Trap)Nitrogen 12 L/min 80 psiAir 2 L/min 80 psi(APCI requires additional 5 L/min of nitrogen)

Genius 1053Precision Zero Air 3.5L

Varian NMR 60 – 500L/min of clean dry air or nitrogenMinimum Dewpoint: -40C / - 40 F

AD rangeInfinity 50**

Varian GC 3900

FID Detector:Zero Air up to 500 cc/minHydrogen up to 50 cc/minMake up gas up to 50 cc/min

TCD Detector:Make up gas and reference gas up to 50 cc/min (Helium or Nitrogen)

Precision Zero AIr 1.5L

Precision Nitrogen 250CC

Precision Nitrogen Trace 250CC

Page 19: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 19

Application Requirements Solution

Waters LCT Premier Range Nitrogen up to 30L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters ZQ /EMD Nitrogen up to 30 L/min, 100 psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters Acquity Range Nitrogen up to 30L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters 3100 Mass Detector Nitrogen up to 30L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters LCT Premier Range Nitrogen up to 30L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters ZQ /EMD Nitrogen up to 30 L/min, 100 psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters Acquity Range Nitrogen up to 30L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters 3100 Mass Detector Nitrogen up to 30L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters Synapt Range

API gas: Nitrogen up to 20L/min, 100psi

Synapt G2 HDMS also requires IMS gas:>99.5% N2 @ 14 psi, 10 – 100 ml/min*Common gas supply for API & IMS is ok if higher purity is used.

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Recommendation TBA*

Waters XEVO TQS Nitrogen at 23.5 L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters G2 QTOF Nitrogen at 23.5 L/min, 100 psi

Infinity NM32LGenius 1050Genius NM32lAGenius NM3G

Waters QTOF API-USAPI-LC/MS/MS

Desolcation Gas: N2 @ 3 L/min - 12.5 L/min

Cone Gas: N2 @ 1 L/min - 5 L/min

Nebulising Gas: 0.3 L/minMax Total N

2 18 L/min at 90-100 psi

Infinity NM32LGenius 1050Genius NM32LAGenius NM3G

Page 20: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 20

Application Requirements Solution

Waters QTOF API-USAPI-LC/MS/MS

Desolvation Gas: N2 @ 3 L/min - 12.5 L/minCone gas: N2 @ 1 L/min – 5 L/minNebulising gas: 0.3 L/minMax Total N2: 18 L/min at 90 – 100 psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters Quattro Micro LC/MS/MS Nitrogen up to 20L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters Quattro Premier Nitrogen up to 25L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters Quattro Micro GC/MS/MS Nitrogen up to 20L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters ELSD 2420 Nitrogen up to 4 L/min @ min 65 psiInfinity NM18LNG4000AGenius N118LA

Waters ELSD 2424 Nitrogen up to 4 L/min @ min 65 psiInfinity NM18LNG4000AGenius N118LA

Waters QTOF Micro Nitrogen between 10 – 20 L/min, 100psi

Infinty NM32LGenius NM32LAGenius 1050Genius NM3G

Waters APGC/XEVO TQ MSDAtmospheric Pressure GC/MS/MS

Carrier Gas: He 2.0 mL/minMake-up Gas: N2 350 mL/min

N/APrecision Nitrogen 600cc

Waters APGC/XEVO QTOFAtmospheric Pressure GC/MS-TOF

Carrier Gas: He 1.2 mL/minMake-up Gas: N2 250 mL/min

N/APrcision Nitrogen 250cc

Waters QDa Nitrogen up to 20 L/min @ min 98 psi

Infinity NM32LGenius 1050Genius NM32LAGenius NM3G

Page 21: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 21

Other ApplicationsApplication Requirements Solution

DESI- OmniSpray ion SourceDesorption Electrospray Ionization (DESI)

(Prosolia.com)

DESI ion Source requires high velocity nitrogen spray at 150 psi for optimum results.

This pressure is currently not obtainable with gas generator solution.

N/A

FT-IRFourier Transform – Infra Red Spectroscopy

CO2-free Dry Air or N2 10 – 100 L/min PG Range

NDIR CO2-free Air or N2 10 – 100 L/min PG Range

ICP UHP N2 for Purge up to 10 L/minNG RangeNG10L-HP, NG5000A*

Teledyne Leeman labsProdigy H ICP Spectrometer

UHP N2, up to 2 L/min @ 20 psi NG Range

Teledyne Leeman labsProfile ICP Spectrometer

UHP N2 for Optics purge10 L/min @ 60 – 90 psi

NG RangeNG10L-HPNG5000A*

NMR200 – 600 MHz or Solid State

Dry Air or N2 for lifting/spinning /Ejecting N2 for magnet cooling

Minimum Dewpoint: -40C / - 40 FFlow rates vary: 45 – 300 L/min

AD RangeInfinity 50**

AA Atomic Absorption Spectroscopy

Dry Air for oxidant gas 20 – 200 L/min AD Range

Eksigent 1D Nano LCDry compressed air or nitrogen, 80-100 psi (Pressure in the system is generated by connecting laboratory air or nitrogen to a pneumatic pressure amplifier. The MFC controller regulates this pressure to generate the required flow rate.)

TBA

UV Fluo UHP N2 for purge gas 1 – 5 L/min NG Range

TOC Analyzer TOC Gas up to 1200 cc/minTOC1500TOC1500HPSCTOCA

Sedere SEDEX LT ELSDLT- ELSD 65, 75 & 85

Up to 3.5 L/min at 50 psi (N2 or AIR)

Infinity NM18LGenius N118LANG range,Precision Zero Air 3.5L

Polymer Laboratories ELSDPL-ELS 2100 IcePL-ELS 1000PL-EMD 960

Nitrogen (98% purity or better)Up to 3.5L/min at 60 psi - max 100 psi

Infinity NM18LGenius N118LAPrecision Zero AIr 3.5L

Microsaic 3500 MiDPressure 2- 6 bar (29 – 87 psi)Purity (99.5%)

Flow rate TBA

Teledyne Fusion TOCTOC gas up to 500cc/min at 65 to 100 psiHydrocarbon and Carbon Dioxide (CO2 ) free air with TOC content <1ppm

TOC1500TOC1500HPSCTOCA

Teledyne Torch TOCTOC gas up to 500cc/min at 65 to 100 psiHydrocarbon and Carbon Dioxide (CO2 ) free air with TOC content <1ppm

TOC1500TOC1500HPSCTOCA

OI Analytical Aurora TOC RangeTOC gas up to 500cc/min at 65 to 100 psiHydrocarbon and Carbon Dioxide (CO2 ) free air with TOC content <1ppm

TOC1500TOC1500HPSCTOCA

OI Analytical Aurora TOC Range TOC gas up to 700 cc/min at 50 -60 psiTOC1500TOC1500HP

Page 22: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 22

Other ApplicationsApplication Requirements Solution

OI Analytical TOC 9210 Range TOC gas up to 200 cc/min at 50 -60 psiTOC1500TOC1500HPSCTOCA

OI Analytical 5380 PFPDPulsed Flame Photometric Detector

Carrier Gas: N2, or H2 up to 10ml/min 80 psiFPD Zero Air : 20 – 30 ml/min at 60 psiFPD Hydrogen up to 15 ml/min at 60 psi

Precision Nitrogen Trace 250ccPrecision Hydrogen TracePrecision Zero Air 1.5L

OI Analytical 4450 Tandem PID/FID

PID Sweep Gas H2 up to 40 ml/minPID Make-up gas N2 up to 25 ml/minFID Air ZA up to 200 ml/min

Precision Nitrogen Trace 250ccPrecision Hydrogen TracePrecision Zero Air 1.5LFusion 1010

ALPHA MOS / FOX Electronic Nose

TOC gas at minimum 150 ml/min @ 30 psiSC-TOC-ATOC1500HP

IonSense DART ion sourceHelium, or 99.998% nitrogen at 80 psi.Flow rate: 2.5 L/min up to 7 L/min

NG RangeNG10L-HP

Chirascan™ Circular Dichroism Spectrometer

Total UHP nitrogen flow: 5 L/min.3 separate flow controlled inlets:

• lamp housing : 1 l/min• monochromator : 3 l/min• light path sample chamber : 1 l/min

NG5000NG5000A, NG10LHP

AVIV Circular Dichroism Spec-trometer

Model 62 DSModel 202Model 420

UHP dry nitrogen at 20 - 35 psi

Flow rate between 10 – 20 L/min

Note: Oxygen will damage the lamp and optics of the CD Spectrometer.

If nitrogen supply runs out during experiment, it will permanently damage the instrument.

A buffer/storage tank downstream of the generator, or a back-up bottled gas supply with automatic switch-over is recommended.

Nitrogen supply should be fitted with an ad-ditional 5 ppb O2 scrubber and indicating trap. (99.9999995% pure nitrogen)

NG10L-HP -with N2 buffer tank

iFlow 6010

SRI Instruments GC

FPD H2: 60mL/min (30 mL/min for each of the two hydrogen flows)

FPD air: 100mL/min

Precision Hydrogen Trace

Precision Zero Air 1.5L

Total Hydrocarbon Analyzer (THA) with FID detector

Hydrogen for Fuel Gas (up to 50cc/min)Zero Air for flame support (up to 500 cc/min)

Precision Hydrogen 100ccPrecision Zero Air 1.5L

ESA CORONA CAD detector99% Nitrogen up to 5 L/min @35 psi(or clean dry oil-free air)

Infinity NM18LNG5000NM500A

ALLTECH ELSD 800 Up to 3L/min 98-99% pure Nitrogen at 15-90 psiInfinty N0M18LGenius N118LANG3000A

ALLTECH ELSD 3300 Up to 4L/min 98-99% pure Nitrogen at 60-80 psiInfinty N0M18LGenius N118LANG4000A

CO² AnalyzerCalibration Gas up to 1000 cc/minCO²-free /Hydrocarbon-free gas

CG15L

TOD (Total Oxygen Demand Analyzer)

UHP Zero grade nitrogen up to 300cc/min Fusion 1010

Entech Canister Cleaner Zero Air at 50 psi CG range

Page 23: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 23

Other ApplicationsApplication Requirements Solution

Buck Sci. 210/211 VGPAtomic Absorption Spectropho-tometer

Acetelyne or Hydrogen at 15 L/min

Clean Dry, Oil Free Air: 20L/min @ 50 psi

Nitrogen UHP. Flow rate varies depending on application

N/A

TBA

Cyclic Voltammetry (CV)UHP Nitrogen for cell purge and blanketing5 – 10 cc/min

Precision Nitrogen Trace 250cc

RheometryClean, dry, oil-free compressed air or nitrogen at 50 -60 psi.Flow rate: 30 to 60 L/min

Genius 3010Infinity 9010Infinity 5010

TA Instruments RheologyQ series

UHP nitrogen up to 100 cc/min at 10-20 psi Precision Nitrogen Trace 250cc

TA Instruments Q5000 SA Dy-namic Vapor Sorption

Purge Gas (Dry UHP N2) for humidity chamber:200 ml/min at 10-20 psi

Purge Gas for furnace/sample area:10 ml/min at 10-20 psi

Precision Nitrogen Trace 250cc

TA Instruments AutosamplerQ5000IR TGA

Purge Gas (UHP N2) for balance area:10 ml/min at 10-20 psi

Purge Gas for furnace/sample area:25 ml/min at 10-20 psi

Precision Nitrogen Trace 250cc

TA Instruments Q50 / Q500 TGAThermogravimetric AnalyzerWith Standard Furnace

Purge Gas (UHP N2) for balance area:40 ml/min at 10-20 psi

Purge Gas for furnace/sample area:60 ml/min at 10-20 psi

Precision Nitrogen Trace 250cc

TA Instruments Q50 / Q500 TGAThermogravimetric AnalyzerWith EGA Furnace

Purge Gas (UHP N2) for balance area:10 ml/min at 10-20 psi

Purge Gas for furnace/sample area:90 ml/min at 10-20 psi

Precision Nitrogen Trace 250cc

TA Instruments Q600 SDTSlimultaneous TGA/DSC Analyser

Purge Gas (UHP N2) flow: 100ml/min Precision Nitrogen Trace 250cc

TA Instruments Qseries DSCDifferential Scanning Calorimeter

Purge Gas (UHP N2) for balance area:10 ml/min at 10-20 psi

Purge Gas for furnace/sample area:90 ml/min at 10-20 psi

Precision Nitrogen Trace 250cc

Titration Oven(Coulometric Karl Fisher) 0.1 – 0.5 L/min clean, dry N2 Precision Nitrogen Trace 250cc

GEA NIRO SD Micro Spray Dryer

Liquid N2 at 30 Kg/hourOr Gaseous N2 at 15 – 20 cfm / 280 – 560 LPMPressure: 80 to 90 psi

Infinity 50**Infinity 90**

GEA NIRO PSD-1 Spray Dryer

Liquid N2 at 80 Kg/hourOr Gaseous N2 at 35 – 40 cfm / 990 – 1100 LPMPressure: 80 to 90 psi

Infinity 50**Infinity 90**

Page 24: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 24

Other ApplicationsApplication Requirements Solution

BUCHIMini Spray Dryer B-290

Spray Gas Compressed Air or Nitroigen200-1000 L/hr, 5-8 bar(4-17 L/min at 70-116 psi)

Note: The Inert Loop B-295 attachment used for organic solvents has an 0

2 analyzer that shuts

down the system when it detects 6% Oxygen in the Nitrogen.

Ensure membrane Nitrogen system is used that can achieve 98%-99% consistently to avoid shut-downs during runs

Infinity NM32LInfinity NM45LGenius NM32LA Genius 1050

Alcor Micro Carbon ResidueTester

Clean Dry CO2-free UHP Nitrogen 20-145 psig); cyl-inder preferred, consumption approx 37L 91.3ft3

per 95 minute test

Note: No available unit with CO2 removal

CEM Microwave PeptideSynthesizer

Requires UHP Nitrogen at 20 L/min at 25 psi i-flow 6010

Applied Biosystems 433Apeptide Synthesizer

Requires UHP Nitrogen at 65 psiNG Rangei-Flow 6010

Perseptive Biosystems /Applied BiosystemsMAriner BiospectrometryWorkstation

UHP (4.8 grade purity) Nitrogen at 80-0100 psiNebulizer gas flow: 0.2-1 L/minCurtain gas flow: 0.5 -2 L/minAuxiliary gas flow: 3 -8 L/min

NG Range

Page 25: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 25

Sample Evaporation /Concentration/ SPEApplication Requirements Solution

Sample PrepSolvent Evaporat

Nitrogen 95% - 99% up to 140 L/minInfinity 90**Infinity 50**

Organomation AquavapNitrogen evaporator

Nitrogen flow rate up to 4 L/min6 positions X 660 ml/position/minute

Infinity NM18LGenius N118LAInfinity 90**

Organomation N-EVAP

Max Nitrogen consumption at 30 psi:6 -12 positions = up to 5 LPM24 positions = up to 10 LPM34 – 45 positions = up to 20 LPM

Infinity 90**Infinity 50**

Organomation MULTIVAP

Max Nitrogen consumption at 30 psi:64 -100 positions = up to 30 LPM96 positions = up to 15 LPM30 positions = up to 15 LPM 48 positions = up to 20 LPM

Genius NM32LAGenius 1050Infinity 90**Infinity NM32L

Organomation MICROVAPMax Nitrogen consumption at 30 psi:up to 30 LPM

Genius NM32LAGenius 1050Infinity 90**Infinity NM32L

SPEware System 96Positive Pressure Manifold

Moisture, Hydrocarbon and particulate-free gas at 40 – 50 psi ( N2 or CDA)

Average Flow rate: 3 L/minAdjustable up to 10 L/min

Infinity 90**Infinity NM18LGenius N118LA

Cerex SPEware 48 port Processor & Evaporator

Processor: N2 at 2 cfh (1 L/min) 70 psiEvaporator: N2 at 25 cfh (12 L/min) 70 psi

Infinity 90**Infinity NM18LGenius N118LA

SPEware Cerex 96 portProcessor & Evaporator

Processor: N2 at 2 cfh (1 L/min) 70 psiEvaporator: N2 at 25 cfh (12 L/min) 70 psi

Infinity 90**Infinity NM18LGenius N118LA

Evaporex / Biotage / ArgonautSPE Dry 96 Dual MicroplateEvaporator Concentrator

8 place RapidVap N2: 17 L/min 20psi maxRapidVap N2/48: 99 L/min 20psi max

Genius 3010Infinity 90**Infinity50**

LABCONCO RapidVapVACUUM N2 & N2/48

8 place RapidVap N2: 17 L /min 20 psi maxRapidVap N2/48: 99 L /min 20 psi max

Infinity 90**Infinity 50**

LABCONCO RapidVap VERTEX73200 Series

Max N2 demand for all 50 nozzles:6.5 cfm / 185 LPM nitrogen at 50 psi5 cfm / 142 LPM nitrogen at 24 psi4.5 cfm / 127 LPM nitrogen at 20 psi

N2 consumption = 1 cfm (28.3 L/min) per row of 10 nozzles at 24 PSI.

N2 consumption = 0.74 cfm (21 L/min) per row of 10 nozzles at 15 PSI.

Infinity 90**Infinity 50**

Gilson SPE 215 Nitrogen up to 20 L/min

Genius NM32LAGenius 1050Infinity 90**Infinity 50**

TECHNE Sample Concentrator Nitrogen 99% 7 - 15 L/min at max 2 psiInfinity NM18LGenius NM32LA

Apricot Evaporex EVX96 Nitrogen up to 60 L/min at 15 – 50 psiInfinity 90**Infinity 50**Genius 3010

Porvair SciencesUltraVap RC + Minivap

Nitrogen min 28 L/min - MAx 80 L/min@ max 6 bar / 87 psi

Infinity 90**Infinity 50**Genius 3010

Page 26: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 26

Sample Evaporation /Concentration/ SPE cont.

Zymark RapidTrace SPEClean, dry, regulated air or nitrogenMin 10 psi / Max 40 psi

Infinity 90**Infinity 50**

UCT Positive Pressure ManifoldSPE N2 or Compressed Air at 80 – 120 psi

Infinity 90**Infinity 50**

Biotage TURBOVAP LV(CALIPER / ZYMARK)

At 10 psi: 1.7 L/min/ per nozzle At 15/20 psi: 2.83 L/min/ per nozzle

Max gas usage for total 50 nozzles: 140 L/minInlet gas pressure min. 60 psi/max 100 psi

Infinity 90**Infinity 50**

Biotage TURBOVAP II(CALIPER / ZYMARK)

At 11 psi: 2.3 L/min/nozzleInlet gas pressure should be min. 30 psi/max 80 psi

Infinity 90**Infinity NM32LAGenius 1050

Biotage TURBOVAP 96 Up to 45 L/min at 70psi minInfinity 90**Infinity NM45LGenius 3010

Biotage TURBOVAP 500 No Gas required N/A

Biotage ISOLERA ELSD-1080 0.9 L/min to 3.25 L/min @ 60 – 100 psiInfinity NM18LGenius N118LANG Range

Zephyr SPE WorkstationClean dry oil-free air or nitrogenup to 65 psi

Infinity 90**Infinity 50**

Zymark RapidTrace SPEClean, dry, regulated air or nitrogenMin 10 psi / Max 40 psi

Infinity 90**Infinity 50**

Genevac EZ-2 Evaporator Nitrogen up to 30 L/min at 15 – 20 psi

Genius NM32LAGenius 1050Infinity 90**Infinity 50**

NOTE: When sizing a nitrogen generator to supply both Mass spec instruments as well as Evaporation equipment, please keep in mind that in some cases, with high N2 demand evaporators, it may be more efficient and cost effec-tive to offer a separate NM**L and CNMG**L generator.

If you are adding the evaporator flow rate to the NM**L capacity, also consider the added purge loss through the membranes.(Purge loss = 120 L/min per membrane)

In the majority of labs, Evaporation equipment is only used intermittently, not continuous.With a separate CNMG**L generator, the unit will only have a demand for compressed air when the evaporator is in operation, therewith preventing waste of compressed air.Furthermore, as the CNMG unit has fewer membranes than the NM**L equivalent, it has a significantly lower purge loss, and therefore places less demand on compressed air supply.Ultimately this will result in more efficient and cost effective operation of the existing compressed air system, or it may enable you to offer a smaller capacity compressor for labs with no house-air supply.

Page 27: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 27

MULTIPLE LCMS INSTRUMENTSLCMS Manufacturer Number of Instruments Solution

AB SCIEX up to 2 Infinity 1032, Genius 3030

AB SCIEX up to 3 Infinity 1033

AB SCIEX up to 4 Infinity1034

AB SCIEX + OTHER1 x standard API +1 x standard LCMS

Genius 3040

AB SCIEX 6500 + 550 w/SelexionUp to 2Up to 3Up to 4

Infinity 1032Infinity 1033Infinity 1034

THERMO Q Exactive Up to 2 ( 116 psi) Infinity 5010 / Genius 3022

Shimadzu LCMS2010 Up to 2 (110 psi Infinity 5010 / Genius 3022

AGILENT CHIP CUBEUp to 2 (Not for iFunnel models!)

Genius 3023

Low Flow LC/MS 2 x @max 15 L/min each Infinity NM32L

Low Flow LC/MS 2 x @max 20 L/min each Infinity NM45L

ALL OTHERS * up to 2 @ max 30 Lpm eachInfinity 5010Genius 3020

ALL OTHERS * up to 4 @ max Lpm each Infinity 5020

ALL OTHERS * up to 6 @ max 30 Lpm each Infinity 5030

ALL OTHERS * up to 8 @ max 30 Lpm each Infinity 5040

ALL OTHERS * up to 10 @ max 30 Lpm each Infinity 5050

ALL OTHERS * up to 12 @ max 30 Lpm each Infinity 5060

ALL OTHERS * up to 14 @ max 30 Lpm each Infinity 5070

ALL OTHERS * up to 16 @ max 30 Lpm each Infinity 5080

• For the Infinity 1034, it may be possible to supply more than 4 x API instruments at the same time, as long as the total curtain, source and exhaust gas requirements do not exceed the total capacity of the generator.

• Standard LCMS (ALL OTHERS) are calculated on a maximum requirement of 30L/min for each instrument.

When sizing an Infinity 50** system for a variety of instruments with different gas requirements, please add up the total gas requirement for each of the instruments.

It is recommended to leave sufficient margin when sizing the generator to make sure the generator does not continuously run at its maximum flow capacity and leave some margin for potential future additions.

• For multi-instrument set-ups including Evaporator equipment, please see comments on previous page.

Page 28: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 28

GAS CHROMATOGRAPHY GENERAL APPLICATIONS NOTES:• Capillary and packed column carrier gas flow rates are variable for all GC’s, mainly depending on the column

size, the type of gas and the type of injector.

• The standard flow rates are shown below, but please note that for SPLIT/SPLITLESS INJECTORS carrier gas flow rates have to be individually calculated based on the operators’ maximum Split Flow Rate. Split Vent flow rates up to 500 cc/min are possible, therefore the hydrogen generator should be appropriately sized.

PLEASE ASK CUSTOMERS FOR MAX SPLIT FLOW RATE AND ADD 20CC TO THAT FLOW TO GET TOTAL CARRIER

GAS USE!

• When calculating total required gas flows, take into account that clients may be operating dual columns in the GC-oven, and each GC instrument can have up to three detectors.

• Hydrogen can often be used as an alternative to Helium as carrier gas in most GC’s, depending on the detec-tor and the specimen to be analyzed. For Carrier Gas, hydrogen has to be very dry, therefore the Precision Hydrogen Trace is the preferred generator model. PH range generators should only be used with an additional moisture trap in the gas stream.

• Hydrogen may react with compounds in the specimen that are in the same thermal conductivity range as hydrogen. ‘False peaks’ a.k.a. ‘ghost peaks’ will occur on the Chromatogram. Also, a chromatographer has to select the correct column stationary phase for use with hydrogen, as the gas may react with components in certain types of stationary phase, which will ‘look like’ hydrocarbon contamination.

• • Trace levels of moisture and oxygen in the carrier gas decompose the stationary phase inside the GC-column,

therefore; all GC manufacturers recommend installing Moisture, Oxygen and Hydrocarbon traps on any GC gas line, even if the source gas is already very pure. In the event of a leak occurring on the gas supply line, the traps will ensure that ambient contaminants do not enter the system.

• • > A hydrocarbon-trap does NOT remove methane from nitrogen carrier/make-up gas. For total hydrocarbon-

free nitrogen you need a catalyst to remove CH4 (Methane). The Peak Zero-Air and Zero-Nitrogen generators feature a catalyst.

• Stainless steel or copper tubing is recommended for GC gases. Do not use nylon or Teflon/PTFE tubing as they are permeable and allow impurities to enter the gas stream.

• For combination N2/Air generators; only offer ANG-range generators if clean, dry air is sufficient. For GC appli-cations that require Zero Air (Hydrocarbon-Free Air) in combination with Zero Nitrogen, please offer the Fusion 1010 model, or separate ZA + NG units, as the ANG-range does not produce zero air

• Gas purity requirements for GC vary depending on the ‘limit of detection’ the chromatographer is looking for. In the case of Trace-detection (0-1 ppm) only Ultra pure/ Research grade gases should be used (99.9995% - 99.9999%)

Page 29: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

Peak Scientific Applications List

peakscientific.comPage 29

GAS CHROMATOGRAPHY GENERAL APPLICATION cont.

INSTRUMENT GAS REQUIREMENT GENERATOR

GC/FID

Hydrogen for fuel gas (up to 50 cc/min per FID)Zero Air for flame support gas (up to 600cc/min per FID)Hydrogen for capillary column carrier gas (up to 10cc/min)Nitrogen for packed column carrier gas (up to 60 cc/min)

Precision Hydrogen TracePrecision Nitrogen Trace 250ccPrecision Zero Air 1.5L

TOTAL FID Hydrogen for fuel gas (up to 50cc/min)Zero Air for flame support gas (up to 500cc/min per FID) Precision Zero Air 1.5L

GC/FPD

Hydrogen for capillary column carrier gas (up to 50cc/min)Nitrogen for packed column carrier gas (up to 50 cc/min)Hydrogen for fuel gas (up to 90 cc/min)Zero Air for flame support gas (up to 200cc/min)

Precision Hydrogen TracePrecision Nitrogen Trace 250ccPrecision Zero AIr 1.5LFusion 1010

GC/NPDHydrogen for capillary gas (up to 50 cc/min)Nitrogen for make up gas (up to 30 cc/min)Zero Air for flame support gas (up to 200cc/min)

Precision Hydrogen TracePrecision Nitrogen 250ccPrecision Zero AIr 1.5LFusion 1010

GC/ECDUHP Hydrogen, Zero Nitrogen or Argon / CH4 for Carrier Gas:Capillary Column flow: up to 20 cc /minPacked Colimn flow: up to 60 cc/ min

Precision Hydrogen TracePrecision Nitrogen Trace 250cc

GC/TCD H2 or N2 as carrier/reference gas (up to 50 cc/min)Precision Hydrogen TracePrecision Nitrogen Trace 250cc

GC/ATD Dry air purge (DP -70°C ) ( up to 2 L/min) PG Range

GC/AED UHP Zero Grade Nitrogen for carrier gas (up to 1L/min) N/A

GC/ELCD Hydrogen as reaction gas (up to 200 cc/min)

GC/HALL Hydrogen as reaction gas (up to 200 cc/min)

GC/PID Hydrogen for Carrier Gas (up to 60 cc/min)

GC/MSCarrier Gases: Capillary column ; Hydrogen up to 10 cc/minPacked column: Nitrogen up to 40 cc/min

Precision Hydrogen TracePrecision Nitrogen Trace 250cc

Thermo 55CHydrocarbon Analyzer

Support Gas Pressures Zero Air: 20 - 30 PSIHydrogen: 20 - 30 PSI (hydrogen)UHP Nitrogen: 25 - 35 PSI

HC Free Air (200 - 300 cc/min)Zero N2 carrier (35 cc/min) H2 Fuel (25 cc/min)Note: additional pressure regulator required!

Precision Hydrogen TracePrecision Nitrogen Trace 250ccPrecision Zero AIr 1.5L

Perkin Elmer Clarus SQ8 GCMS

Carrier Gas 5 ml/minHelium or Hydrogen

Precision Hydrogen Trace

Oi AnalyticalTandem PID/FID Detector for GC

PID Sweep gas - Hydrogen 99.999% up to 40 cc/minPID Make up gas – N2 99.999% or Helium at 25 cc/minFID Air – Zero Air up to 200 cc/min

Precision Nitrogen 250ccPrecision Zero AIr 1.5L

SIS TD 5Short Path Thermal Desorption

Carrier Gas Helium or NitrogenMax 100 c/min at 40 – 60 psi

Compressed Air for gas valve operation, 100 cc/min at 60 psi

Precision Nitrogen Trace 250ccPrecision Zero AIr 1.5LFusion 1010

Page 30: Peak Scientific Gas Generators OEM Application · PDF fileGenerator models with ... Gas: Helium, Hydrogen (optional for 7500cx) Minimum Purity: 99.999% MaxFlow Rate: ... Peak Scientific

peakscientific.com2013 Copyright Peak Scientific

Peak Scientific Taiwan4F.-6, No.736Zhongzheng Rd.Zhonghe Dist.New Taipei City 23511Taiwan (R.O.C.)

Tel: +886 2 8226 2383Fax: +886 2 8226 9499

Peak Scientific BrazilAv. Queiroz Filho1700 – Torre Sunny – Conj. 602São Paulo SP , Brasil CEP: 05319-000

Tel: +55 11 3644-4246Fax: +55 11 3831-7323

Peak Scientific Japan K.K2-7-26, 2F Fuji Building 28Kita AoyamaMinato-KuTokyo, Japan107-0061

Tel: +81 3-6864-0468

Peak Scientific GermanyHerriotstrasse 160528 Frankfurt

Germany

Tel: +49 (0)69 677 33 490Fax: +49 (0)69 677 33 200

Peak Scientific ChinaRoom 606, Building 1Lane 2277 Zuchongzhi RoadPudong New areaShanghaiChina 201203

Tel: +86 21 5079 1190 Fax: +86 21 5079 1191

Peak Scientific MexicoPaseo de la Reforma 505-31Cuauhtemoc, 06500 Mexico D.F.Mexico

Tel: +52 1 771 772 5757

Peak Scientific AustrailiaPO BOX 65BelmontVictoria 3216

Australia

Tel: +61 1300 965 352

Peak Scientific UKFountain CrescentInchinnan Business ParkInchinnanPA4 9REScotland, UK

Tel: +44 (0)141 812 8100Fax: +44 (0)141 812 8200

Peak Scientific North America19 Sterling RoadSuite #1Billerica, MA 01862USA

Tel: +1 866 647 1649Fax: +1 978 608 9503

Peak Scientific India202, Amsri Shamira Old Lancer LineOpp. St. Mary’s Degree CollegeS.D. Road, Secunderabad500025, India

Tel: +91 (0)40 2780 0663Fax: +91 (0)40 2780 0663

Peak Scientific AfricaPO Box 478Somerset Mall7137South Africa

Tel: +27 (0)83 296 4516Fax: +27 (0)86 546 4473


Recommended