©2003 Waters Corporation Performance of an ultra low elution volume 96-well plate Claude R. Mallet, Ziling Lu, Jeff R. Mazzeo, Uwe D. Neue Waters Corporation PittCon 2003 March 10-14 2003 Orlando, Florida
Transcript
Page 1: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Performance of an ultra low elutionvolume 96-well plate

Claude R. Mallet, Ziling Lu, Jeff R. Mazzeo, Uwe D. NeueWaters Corporation

PittCon 2003March 10-14 2003Orlando, Florida

Page 2: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Today’s Challenges Faced by Analytical Chemists in Methods Development

• Sample Process– Faster methods development– Generic and simple extraction protocol– Increased sensitivity and selectivity

• Instrumentation– Faster analysis (1000-analysis-per-day barrier)

Page 3: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Method Development

Sample Preparation Chromatography

MassSpectrometry

Polarity:Silica- C18, C8, C4, C2Hybrid- C18, C8, C4, C2Polymer- C18, C8, C4, C2Embedded polar groupCyano, PhenylParticle size:2.5, 3.5, 5 or 7 µmInternal diameter:4.6, 3.9, 2.1, 1.0, 0.32 mm and 75 µmLength:150, 100, 50, 30, 20 mm

Source:ESIAPcINano-ESIMass analyzers:magnetic sectorselectric sectorstime of flightquadrupoleion trapFT-ICR

Raw sample:- CACO2, microsomes, P450,hepatocytes … etc- tissue, CSF, plasma, serumurine, tears … etc- water, sediment, food … etc

Extracted sampleFor LC/MS/MS

Ideally, the final sample should be as clean as a non extracted standard.

Page 4: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Extraction Techniques

“Many extractions practices are based on classical methodologiesof liquid-liquid or liquid-solid extraction using practices which have not changed for the last one hundred years”

Extraction Methods in Organic Analysis – A.J. Handley 1998

Protein precipitationLiquid-liquid extractionMembrane extractionSoxhlet

Solid phase extractionSolid phase micro extractionAccelerated fluid extractionSupercritical fluid extractionMicrowave-assisted extraction

Classical New technologies

Page 5: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Drug Development Overview

DRUG DISCOVERYDRUG DISCOVERY

DRUG DEVELOPMENT

DRUG DEVELOPMENT

DRUG MANUFACTURE

DRUG MANUFACTURE

Target ID, Lead Generation, Lead Optimization

Pre-Clinical, Clinical Trials, Pilot Plant

QA/QC, Patent Protection

Matrix: CaCO2, microsomes P450, hepatocytes, rat plasmaSample volume: 50 – 100 µLLinearity range: 5 – 250 ng/mL

Matrix: plasma/serum from rat, rabbit, dog, monkey, humanSample volume: 50 µL up to 1 mLLinearity range: 0.1 ng/mL – 500 ng/mL

Page 6: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

96-Well Plates and Cartridges

96-well plate

barrel

Page 7: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Generic reversed phase SPE Method

Condition/Equilibrate500 µL methanol / 500 µL water

Load500 µL spiked sample solution

Wash500 µL 5% methanol in water

Elute500 µL methanol

Evaporate and Reconstitute

Prepare Sample Solution

Typical values for a 10 mg bed packing

Note: For larger bed packing (e.g. 30 or 60 mg), increase the condition, load, wash and elution volumesPre concentration step

Page 8: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Choice of Sorbent Weight Based on Sample Size

Sorbent per Well

Maximum Mass Capacity

Typical Sample Volume

Typical Elution Volume

5 mg 0.15 to 1 mg 10 to 100 µL < 150 µL

10 mg 0.35 to 2 mg 50 to 400 µL < 250 µL

30 mg 1 to 5 mg 100 µL to 1 mL > 400 µL

60 mg 2 to 10 mg 200 µL to 2 mL > 800 µL

With decreasing limit of quantification (LOQ’s), it will be necessary to include an evaporation and reconstitution step, which is extremely time consuming.

Page 9: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

30 mg10 mg 60 mg5 mg

Enabling technology for 96-well plates - Oasis®

sorbent in the amount required to meet your capacity and elution volume needs.

Designed for Optimal Flow Property

Waters two-Stage Well Designin 96-Well Extraction Plate

Page 10: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Oasis® µElution 96-well plate

Page 11: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

µElution tip

Top ball frit

Bottom ball frit

Sorbent

750 µLreservoir

Plate tip

Page 12: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Fat LovingLipophilic monomer

NO

Water LovingHydrophilic monomer

Hydrophilic-Lipophilic Balanced copolymer

N-Vinyl-Pyrrolidone Di-Vinyl-Benzene

- Provides wetting properties- No impact of sorbent drying

-Provides reversed phase properties for analyte retention

Oasis® HLB Sorbent

Page 13: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Strong cation-exchange mode

Best retention at least 2 pH units below pKa

O

OH

NH

CH3

CH3

Propranolol

H+

Reversed-phase mode

Basic drug

NO

SO3HSO3H

Oasis® MCX Sorbent

Page 14: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Generic micro-elutionreversed phase SPE Method

Condition/Equilibrate200 µL methanol / 200 µL water

Load100 µL spiked sample solution

Wash200 µL 5% methanol in water

Elute25 µL ACN:IPA 40:60 + 2% FA

Dilute with 50 µl water

Prepare Sample Solution

Pre-concentration step is avoided

Smaller conditionvolumes

Dilute plasma1:1 ratio with water

HILIC

C18

Page 15: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Generic micro-elutionmixed mode SPE Method

Condition/Equilibrate200 µL methanol / 200 µL water

Load100 µL spiked sample solution

Prepare Sample Solution

Wash 2200 µL MeOH

Elute25 µL ACN:IPA 40:60 + 2% NH4OH

Dilute with 50 µl waterPre-concentration step is avoided

Smaller conditionvolume

Wash 1200 µL Water + 2 % FA Locks basic drug

on ion exchanger

Removes polar interferences

Dilute plasma1:1 ratio with water

Page 16: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Discovery Application

NCH3

CH3

ON

CH3CH3

HOHPropranolol

Antihypertensive

AmitriptylineAntidepressant

Page 17: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation.

LC/MS/MS conditions

MS: Micromass Quattro Ultima LC: Waters Alliance® 2795Ion source: ESI (+) Flow rate: 0.2 mL/minSource temperature: 150 °C Mobile phase A: Water + 0.5 % NH4OHGas cell: 2.0 e-3 bar Argon Mobile phase B: ACN + 0.5 % NH4OHDesolvation temperature: 350 °C Column: XTerra MS C18Capillary voltage: 3.5 kV 2.1 x 30 mm, 3.5 µmDrying gas flow: 500 L/hr LC conditions: 5% to 95% in 1 min. Cone gas flow: 50 L/hr Column temperature: ambientCone voltage: 35 volts

MRM transition: Metoclopramide (IS) m/z 299.8 → 226.7 Propranolol m/z 259.9 → 154.9 Amitriptyline m/z 278.1 → 232.9

Page 18: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00Time2

100

%

7

100

%

16

100

%

MRM of 2 Channels ES+259.9 > 154.9

3.79

MRM of 2 Channels ES+259.9 > 154.93.83

MRM of 2 Channels ES+ 259.9 > 154.93.83

Ppt Propranolol 1 ng/mL

Reversed Phase Propranolol 1 ng/mL

Mixed Mode Propranolol 1 ng/mL

Area: 278

Area: 1329

Area: 13 534

Comparison of Ppt vs HLB vs MCX

Page 19: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00Time2

100

%

4

100

%

17

100

%

MRM of 2 Channels ES+278.1 > 232.9

9.71e34.754.02

2.98

3.38

5.18

MRM of 2 Channels ES+ 278.1 > 232.9

3.62e44.67

MRM of 2 Channels ES+ 278.1 > 232.9

8.91e44.68

Ppt Amitriptyline 0.1 ng/mL

Reversed Phase Amitriptyline 0.1 ng/mL

Mixed Mode Amitriptyline 0.1 ng/mLArea: 5 904

Area: 1 624

Area: 434

Comparison of Ppt vs HLB vs MCX

Page 20: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Development Application

N

OH

OH

CH3CH3

CH3

TerfenadineAntihistaminic

N

OH

OH

CH3CH3

OH

N

OH

OH

CH3CH3

OHO

Terfenadine-alcohol metabolite

Terfenadine-carboxylate metabolite

Page 21: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation.

LC/MS/MS analysis of Terfenadine and Metabolites on mixed mode

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50Time0

100

%

0

100

%

0

100

%

0

100

%

MRM of 4 Channels ES+ 502.2 > 466.2

1.52e52.45

MRM of 4 Channels ES+488.2 > 452.2

2.97e52.61

MRM of 4 Channels ES+472.2 > 436.2

2.18e52.91

MRM of 4 Channels ES+263.9 > 190.8

2.20e62.45

Terfenadine 0.5 ng/mL

IS

Terfenadine-alcohol 0.5 ng/mL

Terfenadine-carboxylate 0.5 ng/mL

Area: 20 599

Area: 29 366

Area: 16 034

Page 22: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation.

Conc. ng/mL N=6

Average

Standard Deviation

RSD %

0.5 1.0 5.0 10.0 20.0 100.0 200.0

0.50 0.96 4.99 10.11 19.26 103.25 197.18

0.008 0.048 0.25 0.51 0.38 3.68 1.92

2 5 5 5 2 4 1

Concentration (ng/mL)

Res

pons

e

Calibration curve of Terfenadine on mixed mode

Coefficient of Determination: 0.997836Calibration curve: 0.00190507 * x^2 + 0.617981 * x + -0.146974Response type: Internal Std ( Ref 1 ), Area * ( IS Conc. / IS Area )Curve type: 2nd Order, Origin: Exclude, Weighting: 1/x^2, Axis trans: None

0.0 20.0 40.0 60.0 80.0 100.0 120.0 140.0 160.0 180.0 200.0-0.147

200

Page 23: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation.

Conc. ng/mL N=6

Average

Standard Deviation

RSD %

0.5 1.0 5.0 10.0 20.0 100.0 200.0

0.49 1.02 5.05 10.12 19.71 101.69 196.39

0.013 0.052 0.26 0.49 0.46 3.25 4.41

3 5 5 5 2 3 2

Concentration (ng/mL)

Res

pons

e

Calibration curve of Terfenadine-Carboxylate on mixed mode

Coefficient of Determination: 0.996306Calibration curve: 0.300987 * x + -0.0391382Response type: Internal Std ( Ref 1 ), Area * ( IS Conc. / IS Area )Curve type: Linear, Origin: Exclude, Weighting: 1/x^2, Axis trans: None

0.0 20.0 40.0 60.0 80.0 100.0 120.0 140.0 160.0 180.0 200.0-0.0391

63.5

Page 24: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation.

Conc. ng/mL N=6

Average

Standard Deviation

RSD %

0.5 1.0 5.0 10.0 20.0 100.0 200.0

0.51 0.97 5.19 10.24 19.80 102.05 196.55

0.008 0.041 0.27 0.44 0.77 3.15 7.92

2 4 5 4 4 3 4

Concentration (ng/mL)

Res

pons

e

Calibration curve of Terfenadine-Alcohol on mixed mode

Coefficient of Determination: 0.993727Calibration curve: 0.854217 * x + -0.196977Response type: Internal Std ( Ref 1 ), Area * ( IS Conc. / IS Area )Curve type: Linear, Origin: Exclude, Weighting: 1/x^2, Axis trans: None

0.0 20.0 40.0 60.0 80.0 100.0 120.0 140.0 160.0 180.0 200.0-0.197

183

Page 25: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Conc. (ng/mL) 0.5 1.0 5 10 20 100 200

Terfenadine

Day 1 (N=6) (C.V.) 0.50 (3) 0.97 (8) 4.96 (4) 9.93 (2) 19.89 (4) 105.29 (4) 195.87 (2)Day 2 (N=6) 0.51 (2) 0.92 (2) 5.29 (2) 10.51 (4) 19.41 (3) 100.56 (4) 200.70 (1)Day 3 (N=6) 0.50 (3) 0.98 (5) 5.09 (6) 9.64 (5) 19.22 (3) 106.38 (3) 194.47 (2)Day 4 (N=6) 0.51 (2) 0.89 (1) 5.21 (3) 9.94 (5) 19.20 (3) 107.04 (2) 197.1 (1)

Terfenadine-carboxylate

Day 1 (N=6) (C.V.) 0.50 (2) 0.96 (5) 4.88 (6) 10.45 (5) 20.74 (6) na naDay 2 (N=6) 0.50 (3) 1.01 (6) 4.79 (1) 9.68 (4) 21.11 (3) 97.12 (5) 204.06 (4)Day 3 (N=6) 0.45 (3) 0.99 (5) 5.21 (4) 10.08 (5) 20.79 (4) 98.36 (4) 194.85 (4)Day 4 (N=6) 0.50 (2) 1.00 (4) 5.10 (6) 10.35 (3) 20.34 (3) 96.02 (5) 192.70 (3)

Terfenadine-alcohol

Day 1 (N=6) (C.V.) 0.52 (1) 0.92 (2) 4.99 (2) 10.30 (4) 20.53 (2) 96.74 (4) 195.23 (5)Day 2 (N=6) 0.51 (1) 0.94 (2) 5.01 (4) 10.03 (4) 20.55 (3) 100.83 (3) 191.65 (3)Day 3 (N=6) 0.50 (2) 0.97 (5) 4.99 (5) 9.88 (2) 20.27 (5) 99.34 (5) 204.10 (3)Day 4 (N=6) 0.50 (2) 1.00 (5) 5.03 (4) 10.36 (4) 20.23 (6) 97.43 (7) 196.04 (3)

Long term stability of Terfenadine and Metabolite on mixed mode

Page 26: Performance of an ultra low elution volume 96-well plate › webassets › cms › library › docs › wa20746.pdf©2003 Waters Corporation Performance of an ultra low elution volume

©2003 Waters Corporation

Conclusions

Wide volume range for loading (50 to 800 µL)

Low elution volume (25 – 50 µL)

Generic protocols for reversed phase and mixed mode

Pre-concentration factor up to 10x

Sub ng/mL quantitation limits

96-well plate format for high throughput

No evaporation and reconstitution step needed


Recommended