PERMA PUREPERMA PURE
GASSGASS™™-2040-2040
GGasas AAnalysis nalysis SSamplingampling SSystemsystems
2040-0012040-001
PERMA PUREPERMA PURE
GASS-2040GASS-2040
2040-0022040-002
• New product introduction at AWMA
• Combines capabilities of Probe / Mini-GASS™
and GASS™-II systems
• Can be mounted anywhere, including directly on stack flange with optional integral sample probe
• Most complete sample conditioning system on the market
• GASS-II discontinued, all other GASS systems still available
PERMA PUREPERMA PURE
The General ProblemThe General Problem
2040-0032040-003
Gas analysis for Emissions or Process Monitoring Gas analysis for Emissions or Process Monitoring involves three steps:involves three steps:
1. Sampling 2. Conditioning 3. 1. Sampling 2. Conditioning 3. AnalysisAnalysis
Samples are typically:Samples are typically: HotHot Wet Wet DirtyDirty
Analyzers require samples to be: Analyzers require samples to be: Cool Dry CleanCool Dry Clean
Sample conditioning receives little attention,Sample conditioning receives little attention,but is often the most difficult step of the three.but is often the most difficult step of the three.
PERMA PUREPERMA PURE
Sampling MethodsSampling Methods
2040-0042040-004
1.1. In-situ analysisIn-situ analysis
2. Extraction2. Extraction
a. Wet (filtered only)a. Wet (filtered only)b. Dilutionb. Dilutionc. Dryingc. Drying
1) Absorption1) Absorption 2) Condensation2) Condensation 3) Selective permeation3) Selective permeation
PERMA PUREPERMA PURE
Typical Problems in Gas SamplesTypical Problems in Gas Samples
2040-0052040-005
GasesGases• Rarely a problemRarely a problem
LiquidsLiquids• WaterWater• Acid mists (sulfuric acid)Acid mists (sulfuric acid)
SolidsSolids• Ammonia saltsAmmonia salts• Excessive particulatesExcessive particulates
PERMA PUREPERMA PURE
Water ProblemsWater Problems
2040-0062040-006
• Clogging or damage to analyzers or components
• Requirement for heated lines or analyzers
• Long calibration times due to adsorption/desorption
• Interference in infrared (or other) measurements
• Loss of water-soluble analytes
• Participation in unwanted chemical reactions
PERMA PUREPERMA PURE
Sulfur Dioxide Solubility in WaterSulfur Dioxide Solubility in Water
2040-0072040-007
Temperature (Deg. C)Temperature (Deg. C)
100100
9090
8080
7070
6060
5050
4040
3030
2020
1010
0000 1010 2020 3030 4040 5050 6060 7070 8080 9090 100100
SO
SO
22 S
olu
bilit
y (
ml S
O S
olu
bilit
y (
ml S
O22/1
00 m
l H
/100 m
l H
22O
)O
)
PERMA PUREPERMA PURE
Sulfur Dioxide Solubility in WaterSulfur Dioxide Solubility in Water
2040-0082040-008
TemperatureTemperature Degree CDegree C
SOSO22 Solubility in H Solubility in H22OOml SOml SO22 / 100 ml H / 100 ml H22OO
00112233445566778899
10101111121213131414151516161717
1818191920202121222223232424252526262727282829293030353540409090
100100
79.879.877.277.274.774.772.272.269.869.867.567.565.265.263.063.060.860.858.758.756.656.654.754.752.752.750.850.849.049.047.347.345.645.643.943.9
42.442.440.840.839.439.438.038.036.836.835.335.334.034.032.832.831.631.630.430.429.329.328.228.227.227.222.522.518.818.82.12.1.0.0
TemperatureTemperature Degree CDegree C
SOSO22 Solubility in H Solubility in H22OOml SOml SO22 / 100 ml H / 100 ml H22OO
PERMA PUREPERMA PURE
Unwanted ReactionsUnwanted Reactions
2040-0092040-009
1. Formation of Sulfuric Acid1. Formation of Sulfuric Acid
a. 2SOa. 2SO22 + O + O2 2 2SO2SO33
SOSO33 + H + H22O HO H22SOSO44
b. NOb. NO22 + SO + SO22 + H + H22O HO H22SOSO44 + + NONO
2. Formation of Ammonia Salts2. Formation of Ammonia Salts
a. NHa. NH33 + SO + SO22 + H + H22OO NH NH44HSOHSO33
b. NHb. NH33 + SO + SO33 + H + H22OO NH NH44HSOHSO44
PERMA PUREPERMA PURE
Acid Mist ProblemsAcid Mist Problems
2040-0102040-010
• Mists are very corrosive, viscous, clogging
• Difficult to remove; slow to coalesce into sizeable droplets when cooled
• Special coalescing filters helpful
PERMA PUREPERMA PURE
Ammonia Salts ProblemsAmmonia Salts Problems
2040-0112040-011
Present special problemsPresent special problems
• Ammonia is gas, passes through probe filter
• Forms clogs if allowed to cool, react withSO2 or SO3 and H2O to form ammonium salts
• Requires special absorbents to remove ammonia before cooling to prevent clogs
PERMA PUREPERMA PURE
Excessive ParticulatesExcessive Particulates
2040-0122040-012
Easiest to solveEasiest to solve
• Probe filter with heated blowback
• Secondary filter with finer porosity
• Inertial bypass filters if necessary
PERMA PUREPERMA PURE
The SolutionThe Solution
2040-0132040-013
• Retain water-soluble analytes (HCl, SO2, NOX)
• Achieve low final sample dew points (down to -25oC)
• Accept high sample flow rates (up to 25 L/minute)
• Accommodate nearly any sample (customized)
• Eliminate heated sample lines (stack mounting)
GASSGASS™™-2040-2040 GGas as AAnalysis nalysis SSampling ampling SSystemsystems
PERMA PUREPERMA PURE
GASSGASS™™
-2040 Gas Analysis Sampling System-2040 Gas Analysis Sampling System
2040-0142040-014
PERMA PUREPERMA PURE
2040-0152040-015
• Stainless steel NEMA-4X enclosure
• Integral probe with heated filter and blowback
• Hastelloy heat exchanger
• Coalescing/particulate filter with automatic drain
• Ammonia scrubber(s)
• Wide range of dryers, double dryers
• PLC with Ethernet for remote control
• Safety interlocks and alarms
• Sample pump, other options to customize system
GASSGASS™™
-2040 Overview-2040 Overview
PERMA PUREPERMA PURE
Integral Probe, Filter and BlowbackIntegral Probe, Filter and Blowback
2040-0162040-016
PERMA PUREPERMA PURE
Integral Probe, Filter and BlowbackIntegral Probe, Filter and Blowback
2040-0172040-017
• Range of probe lengths and materials Directly connected, no heated line
• Probe filter stainless steel or glass-coated Temperature controlled by PLC Avoids losses of analytes in filter
• Blowback heated, closely coupled Blowback timing controlled by PLC Avoids condensation during blowback
PERMA PUREPERMA PURE
Coalescing / Particulate Filter and DrainCoalescing / Particulate Filter and Drain
2040-0182040-018
PERMA PUREPERMA PURE
Coalescing / Particulate FilterCoalescing / Particulate Filter
2040-0192040-019
PERMA PUREPERMA PURE
Coalescing / Particulate FilterCoalescing / Particulate Filter
2040-0202040-020
PERMA PUREPERMA PURE
Coalescing / Particulate FilterCoalescing / Particulate Filter
2040-0212040-021
• Two-layer, coalescing and particle element Temperature controlled by PLC Removes particles >0.1 micron and mists
• Hastelloy heat exchanger Improves acid mist removal
• Available in stainless steel or PVDF Kynar®
Corrosion resistant
PERMA PUREPERMA PURE
Automatic DrainAutomatic Drain
2040-0222040-022
PERMA PUREPERMA PURE
Automatic DrainAutomatic Drain
2040-0232040-023
• Coalesced liquid removed by eductor Drain timing controlled by PLC Reliable, no peristaltic pump
• Eductor entirely PTFE Teflon®
Full corrosion resistance • Teflon check valve Positive seal when not operating
PERMA PUREPERMA PURE
Ammonia ScrubberAmmonia Scrubber
2040-0242040-024
PERMA PUREPERMA PURE
Ammonia ScrubberAmmonia Scrubber
2040-0252040-025
PERMA PUREPERMA PURE
Ammonia ScrubberAmmonia Scrubber
2040-0262040-026
PERMA PUREPERMA PURE
Ammonia ScrubberAmmonia Scrubber
2040-0272040-027
• Proprietary scrubbing media Long-lasting, selective for ammonia
• No tools needed for media replacement Easy maintenance
• Stainless steel with stainless steel shell or polysulfone shell for visibility Corrosion resistant
PERMA PUREPERMA PURE
NafionNafion®® Gas Dryer Gas Dryer
2040-0282040-028
PERMA PUREPERMA PURE
NAFIONNAFION®® Chemical Structure Chemical Structure [ (CF[ (CF22 CF CF22 ) )mm CF CF CF CF22 ] ] nn
O
CF2
CF CF3
O
CF2
CF2
WHEN:WHEN:m = 5 to m = 5 to 13.513.5n = ca 1,000n = ca 1,000z = 1,2,3...z = 1,2,3...xx = 1 to 13 = 1 to 13
ZZ
2040-0292040-029
SOSO33 H ● H ● [[HH22OO]]XX
PERMA PUREPERMA PURE
NAFIONNAFION Gas Dryer Principle Gas Dryer Principle
2040-0302040-030
PERMA PUREPERMA PURE
NAFIONNAFION Dryer Schematic Dryer Schematic
2040-0312040-031
PERMA PUREPERMA PURE
PD-Series PD-Series Polytube DryersPolytube Dryers
2040-0322040-032
PERMA PUREPERMA PURE
NAFIONNAFION Selectivity Selectivity
2040-0332040-033
• Nafion permeation selectivity is based onchemical reactivity, not size
• Only compounds that chemically associate with
sulfonic acid permeate through Nafion
• Bases associate with sulfonic acid and permeate
through Nafion
• Very few bases are gases at typical operatingtemperatures, so very few compounds permeate
PERMA PUREPERMA PURE
2040-0342040-034
Effect on AnalytesEffect on Analytes
• Removes water directly as water vapor.No condensate forms to dissolve analytes.
• Removes compounds with hydroxyl groups (-OH)
H2O, some organics
• Retains important compounds O2, O3, CO, CO2, NOX, SO2, SO3, H2S, Hg, HCl
PERMA PUREPERMA PURE
PD-Series PD-Series Performance ChartPerformance Chart
2040-0352040-035
DEW
PO
INT (
DEW
PO
INT (
ooC
)C
)
-40-40
-50-50
-30-30
FLOW (liters/min.)FLOW (liters/min.)
PD-50T PD-50T ...0 ...0 11 22 33 44 55 66 77 88 99 1010
-45-45
-35-35
12”12”
24”24”
48”48”
72”72”
-25-25-20-20-15-15-10-10-5-50055
101015152020
PD-100T PD-100T ...0 ...0 22 44 66 88 1010 1212 1414 1616 1818 2020PD-200TPD-200T ...0 ...0 44 88 1212 1616 2020 2424 2828 3232 3636 4040
MODELMODEL
PERMA PUREPERMA PURE
Low Final Dew PointsLow Final Dew Points
• Dew point as low as -45oC can be achieved.
• Dew point of -10oC to -25oC is typical.
• Performance is a function of dryer size and sample flow rate.
• Analytical results and reliability improved by reducing water interferences.
• Acid mist problems overcome.2040-0362040-036
PERMA PUREPERMA PURE
PLCPLC
2040-0372040-037
PERMA PUREPERMA PURE
PLCPLC
2040-0382040-038
• Digital selection of operating conditions Precise control of operating parameters
• Standard unit, easy programming Easy maintenance and updating
• Optional Ethernet connection Remote control and data display
PERMA PUREPERMA PURE
Optional Heatless DryerOptional Heatless Dryer
2040-0392040-039
PERMA PUREPERMA PURE
Optional Heatless DryerOptional Heatless Dryer
2040-0402040-040
PERMA PUREPERMA PURE
Optional Heatless DryerOptional Heatless Dryer
2040-0412040-041
PERMA PUREPERMA PURE
Optional Heatless DryerOptional Heatless Dryer
2040-0422040-042
• Steady supply of dry air for system use Continuous 60 L/min. to -45ºC dewpoint
• No line pressure surges Inactive canister repressurized before reuse
• Very reliable with no maintenance Electronic switching, permanent spring- loaded molecular sieve
PERMA PUREPERMA PURE
Optional Sample PumpOptional Sample Pump
2040-0432040-043
PERMA PUREPERMA PURE
Optional Sample PumpOptional Sample Pump
2040-0442040-044
• Pump head mounted in heated zone Prevents sample condensation
• Pump motor mounted in unheated zone Improves pump reliability
• Pump head stainless steel or Teflon Corrosion-resistant, long life
• Sample transported under pressure Avoids contamination from minor leaks
PERMA PUREPERMA PURE
2040-0452040-045
GASS-2040 OptionsGASS-2040 Options
• Integral ProbeIntegral Probe
• Probe Filter Probe Filter
• BlowbackBlowback
• Bypass FilterBypass Filter
• Coalescing FilterCoalescing Filter
• Automatic DrainAutomatic Drain
• Ammonia Ammonia ScrubbersScrubbers
• Dryer SizesDryer Sizes
• Purge Heatless Purge Heatless DryerDryer
• Ethernet Ethernet ConnectionConnection
• X- or Z-PurgeX- or Z-Purge
• Kynar® FittingsKynar® Fittings
• Sample PumpSample Pump
• Purge EductorPurge Eductor
PERMA PUREPERMA PURE
2040-0462040-046
SummationSummation• GASS™-2040 combines all necessary sampling and conditioning in one package
•Proven technology, years of experience
• Opportunity to eliminate heated lines
• Many options; buy only what is needed
• Smaller systems available for less demanding applications