5
pH Transmitter pH INDICATOR MANUAL
Transcript
Page 1: pH INDICATOR MANUAL - Bizzporto.compH Transmitter 5 DIGITAL MEASUERMENT SOLUTIONS A) Basic Hardware Connection 1. Make the connections as per the below diagram for PT100 temperature

pH Transmitter

1

DIGITAL MEASUERMENT SOLUTIONS

pH INDICATOR MANUAL

Page 2: pH INDICATOR MANUAL - Bizzporto.compH Transmitter 5 DIGITAL MEASUERMENT SOLUTIONS A) Basic Hardware Connection 1. Make the connections as per the below diagram for PT100 temperature

pH Transmitter

2

DIGITAL MEASUERMENT SOLUTIONS

A) INSTRUMENT DESCRIPTION

1. A general purpose pH/Conductivity/TDS/temperature meter for use in panel / filed / lab as per models recommended by us.

2. The instrument includes Features include: automatic range selection, calibration on standard solutions or direct cell constant entry, automatic temperature compensation

B) SAFETY

1. Please read this information carefully prior to installing or using this equipment. 2. The unit described in this manual is designed to be operated only by trained personnel. Any adjustments,

maintenance and repair must be carried out as defined in this manual, by a person qualified to be aware of the hazards involved.

3. It is essential that both operating and service personnel employ a safe system of work, in addition to the detailed instructions specified in this manual.

4. References should always be made to the Health and Safety data supplied with any chemicals used. Generally accepted laboratory procedures for safe handling of chemicals should be employed.

5. If it is suspected that safety protection has been impaired in any way, the unit must be made inoperative and secured against any intended operation. The fault condition should immediately be reported to the appropriate servicing authority

C) GOOD PRACTICE GUIDELINES

1. For greatest accuracy ensure no particulate matter is suspended in the solution under test. If necessary, filter or allow the particles to settle prior to use. Do not allow the cell to come into contact with any sediment which may be present.

2. Ensure no air bubbles are trapped in the cell area between the plates. 3. Ensure the cell plates or cell electrodes are completely immersed in the solution under test. 4. Ensure no salt deposits or particulate matters are allowed to build up around the cell points or on the probe

body as this may produce a conductivity path lower than that through the solution. It is recommended that such deposits be removed by soaking the cell in demonized water. No attempt should be made to wipe off these deposits as this may cause damage to the cell electrodes.

5. Ensure the correct reference temperature is selected for the operating procedures being used.

For more information on our products installation guide lines, please contact our channel partner

or mail us on [email protected]

Page 3: pH INDICATOR MANUAL - Bizzporto.compH Transmitter 5 DIGITAL MEASUERMENT SOLUTIONS A) Basic Hardware Connection 1. Make the connections as per the below diagram for PT100 temperature

pH Transmitter

3

DIGITAL MEASUERMENT SOLUTIONS

Introduction:

DMS Solutions provide wide range of instruments with different types as per the industrial requirements, we provide Sensor housing is designed to mount at by-pass line, thereby gives you convenient mounting of sensor and housing. We supply industrial standard pH sensor along with this meter to give you best performance.

pH Meters made available by DMS Solutions are applicable in every area such as agricultural, chemical, pharmaceutical, research institutes, bio-technology laboratories, hospitals etc. These devices have been designed with respect to the scientific precision that is accepted in analytical instruments requirements. The offered pH Meters come with advanced microprocessors as well as provision of auto temperature compensation that is highly desired in analytical instruments. Microcontroller based pH meters and is considered to be an ideal choice for carrying out high end pH measurements. Further, these pH Meter is scientifically designed for achieving maximum stability and quick results. The range offered by us as pH Controllers Transmitters & indicators in different requirements Like Panel, field, flameproof & Bench top.

Salient Features:

Advanced micro controller based design Full 4 Digit, 14mm red LED display for two different parameters with Error code for self diagnosis for easy trouble shutting. User Pass code provision for Authorized settings for change as per application requirement Isolated proportional Retransmitting Output options 4-20 mA, 0 – 2 Volt, 0 – 5 Volt, 0 – 10 Volt Temperature Compensation: Auto and Manual auto switching mode Software settings allow you for your regular calibration of instruments in the easiest way. Available in panel type, field type, and Flame proof and weather proof version on customer request

Page 4: pH INDICATOR MANUAL - Bizzporto.compH Transmitter 5 DIGITAL MEASUERMENT SOLUTIONS A) Basic Hardware Connection 1. Make the connections as per the below diagram for PT100 temperature

pH Transmitter

4

DIGITAL MEASUERMENT SOLUTIONS

Technical Specification chart

Parameter pH mV Temperature Range Range 0 - 14.00 pH 0 - ± 1999.9 mV 0 - 100 °C

Resolution 0.01 pH 0.1mV 0.5 °C

Accuracy ±0.01 pH ± 0.1 mV ± 0.1 °C

Temperature Compensation Auto PT100 0 - 100 °C

Auto PT1000 0 - 100 °C

Manual -- 0 - 100 °C

Calibration 4.00 pH, 7.00 pH, 9.20 pH any Std. Solution

Display pH Full four Digit LED 7 Segment

TEMP Full four Digit LED 7 Segment

Maximum 0 to 9999 Counts

Keyboard 4 Keys, Soft Touch Membrane Type

Input impedance Less than 100 mega Ohm

Accuracy ±0.5% full scale

Slope 98-100%

Response time 0.33 sec

Dimensions Panel 96 x 96 x 85 mm

Filed 120 x 140 x 70 mm

Table 100 x 140 x 85 mm

Weight Instrument 0.4 Kg (Approx.)

Power 230 V ± 10% AC, 50 Hz

Accessories Combination pH Electrode, Temperature Probe(PT100 / PT1000), pH Electrode Stand, Buffer Tablets (7.0 & 4.0 pH), Operation Manual

Page 5: pH INDICATOR MANUAL - Bizzporto.compH Transmitter 5 DIGITAL MEASUERMENT SOLUTIONS A) Basic Hardware Connection 1. Make the connections as per the below diagram for PT100 temperature

pH Transmitter

5

DIGITAL MEASUERMENT SOLUTIONS

A) Basic Hardware Connection 1. Make the connections as per the below diagram for PT100 temperature sensor and PH sensor.

2. Connect pH sensor to terminal No. 1 & 2.

3. Connect temperature sensor (Pt100 or PT 1000) to terminal No. 4, 5 & 6.

4. Connect 4-20 mA output to terminal No. 8 & 9 as per polarity.

5. Connect 230Volt AC 50 Hz Power supply to terminal No. 10, 11 & 12 as per designation.

6. Connect RELAY-1 Output to terminal No. 13, 14 & 15 as per requirement of NO / NC combination.

7. Connect RELAY-2 Output to terminal No. 16, 17 & 18 as per requirement of NO / NC combination.

8. As for Preamplifier +5 volt & ground is also provided in case of Preamplifier use terminal No. 3 & 7


Recommended