1
PH20 Value pH Tester
Instruction Manual
APERA INSTRUMENTS, LLC
www.aperainst.com v 3.4
2
1.BatteryInstallation
Please install batteries according to the following steps. *Please note the correct direction of battery installation:
The Positive Side (“+”) OF EVERY SINGLE Battery MUST FACE UP.
(WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE
TESTER AND POTENTIAL HAZARDS!)
2.KeypadFunctions
■Shortpress-------<2seconds
■Longpress-------->2seconds
1.Shortpresstoturnon,longpresstoturnoff;
2:Whenturnedoff,longpresstoentersetup;3:Inmodesetting,shortpresstochangeparameter;
1.Whenturnedon,longpresstoentercalibrationmode.2.Incalibrationmode,shortpresstoconfirmcalibration;3.Inmodesetting,shortpresstoconfirmparameterselection. Diagram-1
3
3.CompleteKit
Thingsneededinadditiontowhat’sinthebox
Acleancup,distilledwater(8-16oz),andtissuepapersforrinsinganddryingthe
probe.
4.Calibration
4.1Ifit’sfirsttimeuseorthetesterhasn’tbeenusedforalongtime,poursome
pH4.00solutiontotheFilllineintheprobecap,soakprobeforabout15minutes
tohydratethepHsensor.
4.2Shortpress toturnon.Rinseindistilledwater;shakethemeterinthe
airandusetissuepapertodapoffexcesswater.
4.3Longpress toentercalibrationmode;Shortpress toexit.
4.4Inserttheprobeinthe7.00pHcalibrationsolution;Stirgently;
leaveittostand;Waitforthesmileyface toappearand
stayonthescreen(seeDiagram3);shortpress tocomplete
1stpointcalibration,testerreturnstomeasurementmode;
calibrationicon displaysonbuttonleftsideofthescreen.
4.5Rinseprobeindistilledwater.Longpress toentercalibration
mode;inserttheprobeinthepH4.00calibrationsolution,stirgently;leaveit
Diagram-2
Diagram-3
4
tostand;waitforthesmileyface toappearandstayonthescreen;then
shortpress tocompletethe2ndpointcalibration,
testerreturnstomeasuringmode,calibrationicons displayon
bottomleftsideofLCD.
■ Notes
a) TesterwillautomaticallyrecognizepHbuffersolution,userscanchoose
calibrationpoints:1point,2points,or3points.Butthe1stpointcalibration
mustbeincalibrationsolutionof7.00pH,thenfollowedby2ndor3rdpoint
calibration.Fordetails,pleaserefertothefollowingtable:
CalibrationSolution CalibrationIndicationicon
RecommendedAccuracyandRange
1-pointCalibration 7.00pH Accuracy≥0.1pH
2-pointCalibration
7.00pHand4.00pH MeasuringRange<7.0pH
7.00pHand10.01pH MeasuringRange>7.0pH
3-pointCalibration
7.00pH,4.00pH and10.01pH
Widemeasuringrange
b) Automaticself-diagnosticinformation:ifmeasuredvalueisfarofftothe
presetrange,LCDwilldisplay“Er1”;Incalibrationmode,ifthemeasured
valuehasn`tbeenstable,i.e. hasnotbeenstayingonLCD,pressing
willcauseLCDtodisplay“Er2”.
5.Measurement
5.1Shortpress toturnonthetester.Rinseprobeindistilledwater,shake
themeterintheairanddapitwithtissuepapertoremoveexcesswater.
5.2Stirprobeinsamplesolutiongently,leaveittostand.Getreadingsafter
comesupandstays.
5
■ Notes
n IfyouseesomewhitecrystallinesolidleakedoutofthepHprobe,that’sthereference
solution(3MKCL)insidetheprobe.ItisNOTasignofanydefectiveissue.Itisanormal
phenomenonwhentheprobeisstoreddryforaperiodoftime.Itprovesthatthe
junctionoftheprobeisworkingwell.Userscansimplyrinsetheprobeindistilledwater
toremovethesolidsandusethetesterasusual.
n Aftereachtest,usersshouldrinsethepHprobethoroughlywithdistilledwateror
purifiedwater.
n ForthepremixedpHcalibrationbuffersolutions,werecommendreplacingthemafter
10to15timesofusetokeepitsaccuracy.
n ThismeterwillNOTgiveaccurateorstablepHreadingswhentestingdistilledordeionizedwater.Thisisbecausedistilledordeionizedwaterdonothaveenoughionspresentfortheelectrodetofunctionproperly.Tomeasuredistilledordeionizedwater’spH,usersneedtouseaspecializedinstrument.Contactusatinfo@aperainst.comformoredetails.Whentestingpurifiedwaterlikespringwaterordrinkingwater,itwilltakelongerforthereadingstogetstabilized(typically3-5minutes)becausethereisveryfewionslefttobedetectedbythesensorinthosepurifiedwater.
n DoNOTstoreprobeinpurifiedwaterbecausethatwillcausepermanentdamagetothepHprobe.Purifiedwaterisonlyrecommendedforrinsingtheprobe.Theprobeshouldbestoredin3MKCLpHelectrodestoragesolution(SKUAI1120)forbestaccuracyorstoredinthepH4.00calibrationsolutionasanalternativeifstoragesolutionisnothandy.
n DoNOTuseanyotherbrand’sstoragesolutionsbecausedifferentchemicalsmaybeusedandpotentialpermanentdamagecouldbecausedtothemeter.
n StoringtheprobedrywillNOTcausepermanentdamagetoit.Itwillonlytemporarily
causetheprobetoloseitssensitivity,whichcanalwaysberestoredbysoakinginthestoragesolutionorpH4.00calibrationsolution.
6.ParameterSetting
6.1SettingSchedule
Prompt
MarkParameterSettingItems Code
Factory
Default
P1 SelectpHbuffer USA–NIST USA
P2 SelectTemperatureUnit ˚F–˚C ˚F
P3 BacktoFactoryDefault No–Yes No
6
6.2ParameterSetting
Whenturnedoff,longpress toentersetup → shortpress toswitch
P1-P2-P3→ Shortpress ,parameterflashing→shortpress tochoose,
shortpress toconfirmparameterselection→Longpress togobackto
measurementmode.
6.3ParameterSettingInstruction
SelectstandardpHbuffersolution(P1):Therearetwooptionsofstandardbuffer
solutions:USAseriesandNISTseriesasthefollowingchart:
IconspHstandardbuffersolutionseries
USAseries NISTseries
Three-Pointcalibration
1.68pHand4.00pH 1.68pHand4.01pH
7.00pH 6.86pH
10.01pHand12.45pH 9.18pHand12.45pH
6.4Fortheself-diagnosisinformation,pleaserefertothetablebelow:
Symbol Self-Diagnosis information How to fix
Wrong pH calibration solution, which exceeds the recognizable range of the meter.
1. Check if calibration solution is correct 2. Check if probe is damaged. 3. Check if there is any air bubble in the glass bulb sensor
Is pressed before measurement is stable ( appears and stays)
Wait for the smile icon to appear and stay, then press
* If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few times to remove it. The existence of an air bubble in the glass bulb will significantly decrease the accuracy of measurement.
7
* The 1st point calibration must be 7.00 pH. Perform the 2nd point calibration (4.00 pH) immediately after the 1st point. Do NOT turn off the meter before you conduct 2nd point calibration. If the meter is turned off after 1st point calibration, users will need to restart the calibration process with the 7.00 pH and the 4.00 pH following after. Calibrating directly in pH 4.00 after turning meter off and back on will cause Er1.
7.TechnicalSpecifications
8.TesterFunctions
8.1Calibrationpointsindication:
8.2StableMeasurement: appearsandstaysonthescreen
8.3Self-Diagnosticinformation:Er1,Er2
8.4Low-Voltagewarning: flashes,reminderofbatteryreplacement
8.5AutoPower-Offin8minutesifnooperation.
Gotosupport.aperainst.comtofindtutorialvideosandFAQarticles,orsubmita
ticketifyouhaveanyquestionsregardingtheusingmeter.
pH
Range 0-14.0pH
Resolution 0.1pH
Accuracy ±0.1pH
CalibrationPoints 1-3points
AutomaticTemperatureCompensation 0–50˚C(0–122˚F)
Temp.
Range 0–50˚C(0–122˚F)
Resolution 0.1˚C
Accuracy ±0.5˚C
8
9.Warranty
Wewarrantthisinstrumenttobefreefromdefectsinmaterialandworkmanship
andagreetorepairorreplacefreeofcharge,atoptionofAPERAINSTRUMENTS,
LLC,anymalfunctionedordamagedproductattributabletoresponsibilityofAPERA
INSTRUMENTS,LLCforaperiodofTWOYEARS(SIXMONTHSfortheprobe)fromthe
delivery.
Thislimitedwarrantydoesnotcoveranydamagesdueto:
Transportation,storage,improperuse,failuretofollowtheproductinstructionsor
toperformanypreventivemaintenance,modifications,combinationorusewithany
products,materials,processes,systemsorothermatternotprovidedorauthorized
inwritingbyus,unauthorizedrepair,normalwearandtear,orexternalcausessuch
asaccidents,abuse,orotheractionsoreventsbeyondourreasonablecontrol.