+ All Categories
Home > Documents > Platelet adsorption and hemolytic properties of liquid ...

Platelet adsorption and hemolytic properties of liquid ...

Date post: 27-Dec-2021
Category:
Upload: others
View: 3 times
Download: 0 times
Share this document with a friend
10
International Journal of Pharmaceutics 327 (2006) 117–125 Platelet adsorption and hemolytic properties of liquid crystal/composite polymers Mei-Fen Shih a , Min-Da Shau a , Meng-Ying Chang a , Se-Kai Chiou a , Jiunn-Kae Chang a , Jong-Yuh Cherng b,a Chia Nan University of Pharmacy and Science, 60 Erh-Jen Rd., Sec. 1, Jen-Te, Tainan 717, Taiwan, ROC b National Chung Cheng University, 168 University Rd., Chia-Yi 621, Taiwan, ROC Received 26 January 2006; received in revised form 13 June 2006; accepted 25 July 2006 Available online 29 July 2006 Abstract The aim of this study is to investigate how the presence of liquid crystal, cholesteryl oleyl carbonate, embedded into polymers (PMMA, Eb270, PU) affects the biocompatibility of composite membranes with human blood. The effects of different surface textures of composite membranes on platelet adhesion and platelet activation were evaluated as well. The adhesion and geometric deformation of platelets were demonstrated by SEM. The quantitative assay of platelet activation was determined by measuring the production of P-Selectin, and by measurement of the blood clotting index when PRP blood was incubated with pure polymer films and composite membranes. Moreover, the hemolysis studies on the damage to red blood cells were performed to gain information on the hemocompatibility of these biomaterials. The results showed that inclusion of cholesteryl oleyl carbonate (COC) embedded in composite membranes, improves their biocompatibility with respect to a substantial reduction of platelet adhesion and the controlled decrease of platelet activation. As the COC content of composite membranes was increased, the value of the blood clotting index increased and the production of P-Selectin decreased. The results also showed that the presence of COC resulted in a decrease of hemolysis ratios. Comparing among three different composite membranes, the best biocompatibility is achieved when PU/COC Eb270/COC > PMMA/COC. The in vitro studies performed in this work suggest that it may be reasonable to use liquid crystal COC as a mean of surface modification to improve the blood compatibility of biopolymers. © 2006 Elsevier B.V. All rights reserved. Keywords: Polymer; Liquid crystal; Cholesteryl oleyl carbonate; Platelet; Blood compatibility 1. Introduction Polymeric materials used in medical and related applications have safety issues that are related to their compatibility with blood, notably by the occurrence of thrombosis (Gorbet and Sefton, 2004). Many efforts have been employed to improve the blood compatibility of biomaterials via surface modification, e.g. DNA-blend polysulfone (Zhao et al., 2003) or inclusion of polyethylene oxide (PEO) or negatively charged side chains (Lee and Oh, 2002; Kim et al., 2003). The key point of these modi- fications is to prevent thrombosis initiated by platelet adhesion in normal and pathologic states of hemostasis. Corresponding author. Tel.: +886 958074911; fax: +886 5 2721040. E-mail address: [email protected] (J.-Y. Cherng). Hemostasis is a physiological process for blood coagulation involving plasma coagulation factors, platelets and endothelial cells of blood vessels. The coagulation of blood consists of a cascade of reactions dividing into two pathways: the extrinsic and intrinsic pathways. Both pathways finally lead to the for- mation of an active form of factor X (factor Xa). The factor Xa catalyzes the conversion of prothrombin to thrombin that cleaves the fibrinogen to fibrin monomer. The polymerization and subsequent cross-linking of fibrin monomer produce fibers that enmesh platelets to form hemostatic clots. During this pro- cess of blood coagulation the platelet adhesion on the tissue surface is very important. Platelets can accelerate thrombosis through several actions: (1) secretion of bulk phase agonists (e.g. ADP, Thromboxane A2) which attract more platelet adhe- sion, (2) fibrinogen-mediated platelet–platelet aggregation and (3) by the acceleration of thrombin production (Grunkemeier et al., 1998). Moreover, P-Selectin excreted by activated platelets 0378-5173/$ – see front matter © 2006 Elsevier B.V. All rights reserved. doi:10.1016/j.ijpharm.2006.07.043
Transcript
Page 1: Platelet adsorption and hemolytic properties of liquid ...

A

PpTibcaiCit©

K

1

hbSbepafii

0d

International Journal of Pharmaceutics 327 (2006) 117–125

Platelet adsorption and hemolytic properties of liquidcrystal/composite polymers

Mei-Fen Shih a, Min-Da Shau a, Meng-Ying Chang a, Se-Kai Chiou a,Jiunn-Kae Chang a, Jong-Yuh Cherng b,∗

a Chia Nan University of Pharmacy and Science, 60 Erh-Jen Rd., Sec. 1, Jen-Te, Tainan 717, Taiwan, ROCb National Chung Cheng University, 168 University Rd., Chia-Yi 621, Taiwan, ROC

Received 26 January 2006; received in revised form 13 June 2006; accepted 25 July 2006Available online 29 July 2006

bstract

The aim of this study is to investigate how the presence of liquid crystal, cholesteryl oleyl carbonate, embedded into polymers (PMMA, Eb270,U) affects the biocompatibility of composite membranes with human blood. The effects of different surface textures of composite membranes onlatelet adhesion and platelet activation were evaluated as well. The adhesion and geometric deformation of platelets were demonstrated by SEM.he quantitative assay of platelet activation was determined by measuring the production of P-Selectin, and by measurement of the blood clotting

ndex when PRP blood was incubated with pure polymer films and composite membranes. Moreover, the hemolysis studies on the damage to redlood cells were performed to gain information on the hemocompatibility of these biomaterials. The results showed that inclusion of cholesteryl oleylarbonate (COC) embedded in composite membranes, improves their biocompatibility with respect to a substantial reduction of platelet adhesionnd the controlled decrease of platelet activation. As the COC content of composite membranes was increased, the value of the blood clotting indexncreased and the production of P-Selectin decreased. The results also showed that the presence of COC resulted in a decrease of hemolysis ratios.

omparing among three different composite membranes, the best biocompatibility is achieved when PU/COC�Eb270/COC > PMMA/COC. The

n vitro studies performed in this work suggest that it may be reasonable to use liquid crystal COC as a mean of surface modification to improvehe blood compatibility of biopolymers.

2006 Elsevier B.V. All rights reserved.

d com

iccamXcat

eywords: Polymer; Liquid crystal; Cholesteryl oleyl carbonate; Platelet; Bloo

. Introduction

Polymeric materials used in medical and related applicationsave safety issues that are related to their compatibility withlood, notably by the occurrence of thrombosis (Gorbet andefton, 2004). Many efforts have been employed to improve thelood compatibility of biomaterials via surface modification,.g. DNA-blend polysulfone (Zhao et al., 2003) or inclusion ofolyethylene oxide (PEO) or negatively charged side chains (Leend Oh, 2002; Kim et al., 2003). The key point of these modi-

cations is to prevent thrombosis initiated by platelet adhesion

n normal and pathologic states of hemostasis.

∗ Corresponding author. Tel.: +886 958074911; fax: +886 5 2721040.E-mail address: [email protected] (J.-Y. Cherng).

cst(s(a

378-5173/$ – see front matter © 2006 Elsevier B.V. All rights reserved.oi:10.1016/j.ijpharm.2006.07.043

patibility

Hemostasis is a physiological process for blood coagulationnvolving plasma coagulation factors, platelets and endothelialells of blood vessels. The coagulation of blood consists of aascade of reactions dividing into two pathways: the extrinsicnd intrinsic pathways. Both pathways finally lead to the for-ation of an active form of factor X (factor Xa). The factora catalyzes the conversion of prothrombin to thrombin that

leaves the fibrinogen to fibrin monomer. The polymerizationnd subsequent cross-linking of fibrin monomer produce fibershat enmesh platelets to form hemostatic clots. During this pro-ess of blood coagulation the platelet adhesion on the tissueurface is very important. Platelets can accelerate thrombosishrough several actions: (1) secretion of bulk phase agonists

e.g. ADP, Thromboxane A2) which attract more platelet adhe-ion, (2) fibrinogen-mediated platelet–platelet aggregation and3) by the acceleration of thrombin production (Grunkemeier etl., 1998). Moreover, P-Selectin excreted by activated platelets
Page 2: Platelet adsorption and hemolytic properties of liquid ...

118 M.-F. Shih et al. / International Journal of Pharmaceutics 327 (2006) 117–125

D rylatep diacr

inwpbB

immwaysplpcbwlToaImeshm

2

2

sf(oMbpSufr

2m

oCmtw

iagram 1. The structures of tecoflex polyurethane (PU), poly-methyl methacolyurethane, PU; (b) poly-methyl methacrylate, PMMA; (c) aliphatic urethane

s also involved in the adhesion of platelets to monocytes andeutrophils, playing a central role in neutrophil accumulationithin thrombi (Tedder et al., 1995). Therefore, the extent oflatelet adhesion to biomaterials is often used as an index oflood compatibility (Aldenhoff and Koole, 2003; Skarja andrash, 1997).

Liquid crystals are a phase of matter whose molecule order isntermediate between that of a liquid and that of a crystal. This

eans that a liquid crystal may flow like a liquid but have theolecules in the liquid arranged and oriented in a crystal-likeay. Liquid crystal has been used in medical applications such

s embedding liquid crystal molecules in membranes for an anal-sis of thermo-responsive systems (Lin et al., 2001). Similarly,everal surfactants with liquid crystal properties are used forrolonged drug delivery, as well as topical delivery systems viaiquid crystals (Makai et al., 2003; Nesseem, 2001). Due to thehase transition property (solid to cholesteric phase) of liquidrystal molecules, they can behave like mobile plasma mem-ranes. Also, thermotropic nematic liquid crystal molecules,hose thickness is just twice the length of lipid molecules, are

ike the structural units of cell membranes (Tu et al., 2001).hese characteristics of liquid crystals [especially cholesterylleyl carbonate (COC)] make them biocompatible and usefuls anticoagulant biomaterials (Tu et al., 2001; Lin et al., 2002).n this work, the blood compatibility of three series of poly-er/liquid crystal composite membranes is investigated. The

xtent of platelet adhesion and platelet activation on polymerurfaces was determined for establishing a correlation to theemocompatible properties of liquid crystal-embedded poly-ers.

lptp

(PMMA) and aliphatic urethane diacrylate (Eb270): (a) Structure of tecoflexylate, Eb270.

. Materials and methods

.1. Materials

Chemical structures of three polymers were plotted and arehown in Diagram 1. Tecoflex polyurethane (PU) was purchasedrom Fluka, Buchs, Switzerland. Polymethyl methylacrylatePMMA) with molecular weight of 15,000 Da and cholesterylleyl carbonate (COC) were obtained from Aldrich chemicals,ilwaukee, WI. Aliphatic urethane diacrylate was supplied

y Double Bond Chem. Ind. Co. Ltd. 1-hydroxy-cyclohexyl-henyl-ketone (HCPK) was obtained from Ciba-Geigy, Basel,witzerland. For the quantitative determination of human sol-ble P-Selectin in plasma, an immunoassay kit was purchasedrom R&D systems, Inc., MN, USA. All other chemicals andeagents used were of analytical grade.

.2. Preparation of polymer/liquid crystal compositeembrane via photo-polymerization

A polymer, Eb270, was synthesized by photo-polymerizationf aliphatic urethane diacrylate in the presence of liquid crystalOC. The aliphatic urethane diacrylate (prepolymer) was firstixed with 3% radical initiator (HCPK). A clear melting solu-

ion was formed around 60◦–70◦ and various amounts of COCere added to obtain weight ratios of COC from 0 to 30%. A hol-

ow spacer (80 �m in depth) on a bottom lining (all composed ofolypropylene) was set on a glass plate and then the mixed solu-ion was cast into a mold. After a polypropylene cover lining waslaced over the mold, a glass cover was put on and attached to

Page 3: Platelet adsorption and hemolytic properties of liquid ...

al of

abatu

2m

Pc

cwfhaf

2m

towwwwafa

m1sopaob

2m

wmf0fs

3awsftTrt0wie

(abs

I

2P

iSapjcwioPlck

2m

wsbtrodtbant

M.-F. Shih et al. / International Journ

nother glass plate with clips. Subsequently, the whole assem-ly was moved into a UV-lamp box for photo-polymerizationnd composite membranes were formed after 60 s of exposureime. After polymerization was finished, the polymer was vac-um dried in dessicator.

.3. Preparation of polymer/liquid crystal compositeembrane via solvent evaporation

Composite membranes composed of polymer (PU orMMA) and liquid crystal (COC) were prepared by a solvent-asting method. Briefly, the solution of polymer (PU or PMMA)

ontaining COC in various ratios (0, 10, 20 and 30%, w/w)as prepared by stirring the components in solvent chloroform

or 1 h at room temperature. The resulting solution is clear andomogeneous. The solution was cast into an aluminum mold andfter 48 h in a clean environment the solvent was evaporated toorm composite membranes.

.4. Observation of platelet adhesion by scanning electronicroscopy (SEM)

Fresh blood in addition with the anticoagulant citrate dex-rose (ACD) (9:1) was centrifuged at 100× for 10 min at 4 ◦C tobtain platelet-rich plasma (PRP). The composite membranesere rinsed three times with deionized water and then coveredith 3 ml PRP (average platelet number is 5.4 × 105 ml−1), pre-armed to 37 ◦C. After 1 h of incubation at 37 ◦C, membranesere washed with PBS to remove non-adherent platelets. The

dhered platelets were fixed with 2% (w/v) glutaraldehyde/PBSor 5 min at 4 ◦C. After thorough washing with PBS, the platelet-ttached membranes were vacuum dried prior to SEM studies.

Non-treated composite membranes and platelet-attachedembranes were shadowed with Pt–Pd alloy at 15 mA for

00 s. Base on the observation of electron scanning micro-cope (Hitachi S-3000N), the distribution of COC on the surfacef composite membranes was evaluated. Also, the degree oflatelet adhesion on different membranes and the morphology ofdhesive platelets were investigated to understand the influencef the COC-embedded polymer surface on the blood compati-ility.

.5. In vitro blood compatibility test: blood clottingeasurement

The prepared polymer/liquid crystal composite membranesere placed into flat-bottom bottles. These bottles were ther-ostated in water bath at 37 ◦C for 5 min. Then 0.27 ml

BCI index = 100 ×

rom blood sample (0.3 ml ACD-whole blood by addition of.024 ml CaCl2 (0.2 mol/l)) was slowly dropped on the sur-ace until completely covered. The bottles containing bloodamples were further incubated in a thermostate incubator at

ttad

Pharmaceutics 327 (2006) 117–125 119

7 ◦C. After 10 min, 10 ml of deionized water were carefullydded by dripping water down the inside wall of the bottlesithout disturbing the clotted blood. Subsequently 10 ml of

olution were taken from bottles and centrifuged at 100 × gor 30 s. The supernatant was decanted into a tube with addi-ional 40 ml deionized water and kept in 37 ◦C for 60 min.he blood clotting test was carried out by spectrophotomet-

ic measurement of the relative absorbance of blood sampleshat had been diluted to 50 ml at 542 nm. The absorbance of.25 ml ACD-whole blood in 50 ml deionized water at 542 nmas assumed to be 100 as a reference value. The blood clotting

ndex (BCI) of biomaterials can be quantified by the followingquation:

of blood which had been in contact with sample at 542 nm)

abs of ACD whole blood in water at 542 nm

t is clear that as the BCI index rises, blood clotting decreases.

.6. Quantitative evaluation of activation of platelets by-Selectin measurements

The P-Selectin assay employs the quantitative sandwichmmunoassay technique. A monoclonal antibody specific for P-electin was pre-coated onto a microplates. Standards, samplesnd control were pipetted into microwells and then added to aolyclonal antibody specific for P-Selectin, which had been con-ugated with horseradish peroxidase. After removal of unboundonjugated antibody, a substrate was added and color developed,hich is proportional to P-Selectin concentration. PRP without

ncubation of polymers is taken as a control for the comparisonf samples in presence of polymers. A known concentration of-Selection (36.86 ng/ml) included in kit was measured at wave-

ength at 450 nm as a standard and for establishing of a standardurve. Experimental procedures were followed as listed in theit brochure.

.7. In vitro blood compatibility test: hemolysis ratioeasurement

The prepared polymer/liquid crystal composite membranesere rinsed three times with deionized water and normal

aline before being transferred and placed into flat-bottomottles. Then 10 ml normal saline were poured into the bot-les and kept at 37 ◦C in a shaking water bath (shakingate = 100 times per hour). After 60 min incubation, 0.2 mlf diluted ACD-whole blood (8 ml ACD-whole blood wasiluted by addition of 10 ml normal saline) was dropped intohe bottles allowing the test membranes to be soaked in thelood solution for another 60 min. Next, the solutions werespirated and centrifuged at 100 × g for 5 min. The super-atant was measured at the absorbance of 542 nm by spec-rophotometer. The hemolysis ratio (HR) was obtained by

he equation HR = 100 × (AS − AN)/(AP − AN), where AS ishe absorbance of sample supernatant. AP and AN are thebsorbance of the positive controls (10 ml pure water + 0.2 mliluted ACD-whole blood), and the negative control (0.2 ml
Page 4: Platelet adsorption and hemolytic properties of liquid ...

120 M.-F. Shih et al. / International Journal of Pharmaceutics 327 (2006) 117–125

Table 1Contact angle (◦) and surface energy (mN/m) on PMMA and PU composite polymers

PMMA PMMA + 10% COC PMMA + 30% COC PU PU + 10% COC PU + 30% COC

Contact angle in water 95.2 ± 0.8 78.7 ± 0.5 67.1 ± 1.6 96.2 ± 5.6 84.9 ± 0.4 83.9 ± 0.6C 18.S 54.

V

dr

2

aTdsAs

2

wtr

3

3

pitwC

htwfeiastbtrmtt

3a

Cbpiol

Fa

ontact angle in diiodomethane 27.3 ± 2.8 21.5 ± 1.6E (surface energy) 46.1 ± 0.3 49.9 ± 0.3

alues are means of at least 10 measurements ± standard deviation.

iluted ACD-whole blood in presence of 10 ml normal saline),espectively.

.8. Contact angle and surface energy determination

Contact angles methods were used to determine the wettingbility and surface energy of PMMA and PU composite films.he contact angles subtended by two kinds of liquid (water andiiodomethane) were measured with a sessile drop method andurface energy was then calculated (DSA100, Kruss, Germany).t least five drops of each liquid were used on each sample

urface to get good statistical results.

.9. Statistical analysis

Data from at least three independent polymers in each groupere analyzed. Results are expressed as mean values (±S.D.) of

hree experiments. Results of SEM experiments are shown withepresentative micrographs.

. Results and discussion

.1. General properties of the composite membranes

The pure polymer films and the polymer/liquid crystal com-osite membranes in which the liquid crystal (COC) content

s less than 10% are colorless and transparent at room tempera-ure. The composite membranes were observed to be translucenthen the COC content is more than 10%. With the increase ofOC these composite membranes become more hydrophlic and

pWid

ig. 1. SEM micrographs of platelets adhesion on the surface of PMMA film afternd flattening of adhered platelets were observed.

5 ± 1.2 42.4 ± 6.6 37.5 ± 3.3 25.2 ± 2.25 ± 0.5 38.6 ± 0.3 42.6 ± 1.0 47.5 ± 1.0

igher wetting ability (Table 1). This might be due to the facthat the presence of amphipathic COC molecules in a polymerould shield the biomaterial surface with their polar moieties

ace outwards to the hydrophilic environment (non-polar moi-ties of COC are more close to the hydrophobic polymer) asndicated in contact angel studies. The higher wetting ability ofsurface causes a decrease in contact angle and an increase of

urface energy. The transition temperature of COC from crys-al state to liquid state is lower than room temperature. This isecause the temperature of smectic–cholesteric transition andhe cholesteric–isotropic transition of COC are 18.9 and 38 ◦C,espectively (Lin et al., 2001). Therefore, the liquid crystalolecules were mobile in the composite membranes. It is likely

hat these COC molecules are also mobile, behaving in vivo likehe phospholipid component of cell membranes.

.2. Observation of surface morphology and plateletdhesion on pure polymer films and composite membranes

SEM micrographs were employed to assess the effect ofOC contents, embedded in artificial membranes, on humanlood compatibility. The extent of adhesion and deformation oflatelets were observed and correlated to the blood–membranenteraction. Fig. 1A and B shows that the morphology of plateletsn pure PMMA films, spreads flatten and extends into irregu-ar shapes indicating activation of platelets. Also, coagulated

latelets were found to be less compatible with human blood.hen COC content is increased to 10% in PMMA compos-

te membranes, the number of adhesive platelets on the surfaceecreases (Fig. 2A and B). However, the shape of most adhesive

60 min: (A) 1000× magnification and (B) 2000× magnification. Pseudopodia

Page 5: Platelet adsorption and hemolytic properties of liquid ...

M.-F. Shih et al. / International Journal of Pharmaceutics 327 (2006) 117–125 121

F COCm s).

ptbmdw

mts(Btlmimpp

siEsgfmttfmvtom

Fa

ig. 2. SEM micrographs of platelets adhesion on the surface of PMMA–10%agnification. Platelet (arrowheads) and crystalline of COC (double arrowhead

latelets was shown to stay regular and did not flatten, indicatinghat the biocompatibility of the composite membrane to humanlood is improved. The SEM micrographs of PMMA compositeembranes with COC content of 20 and 30% could not be taken

ue to the motion of COC molecules on composite membranes,hich is caused by high-energy electron beams.Fig. 3A and B shows the morphology of platelets on pure PU

embranes. It is clear in the figures there are fewer platelets onhe membranes, but that they are more regular in shape. On theurface of PU composite membrane having more COC content10%), far fewer adhesive platelets were observed (Fig. 4A and). The more COC (20–30%) present in composite membranes,

he fewer adhesive platelets were found (Fig. 5A and B). Vesicle-ike objects are liquid crystal droplets or domains of COC

olecules, which form on the surface texture of PU compos-

te membranes (Figs. 4 and 5). Non-PRP treated PU composite

embranes that exhibit vesicle-like objects. The Eb270 com-osite membrane is prepared by dissolving COC in a suitablerepolymer, and is followed by photo-polymerization instead of

aMob

ig. 3. SEM micrographs of platelets adhesion on the surface of PU film after 60 min:nd crystal of salts (double arrowheads).

composite membrane after 60 min: (A) 1000× magnification and (B) 2000×

olvent evaporation, which is used for PU and PMMA compos-te membranes. In Fig. 6A and B, fewer platelets adhere on theb270 pure film in comparison with PMMA. Also, these adhe-ive platelets all remained in their regular shapes, indicatingood biocompatibility. When 10 or 20% COC was embedded,ar fewer adhesive platelets were found on Eb270 compositeembranes (Figs. 7 and 8, respectively). Also, a distinct surface

exture on Eb270 composite membranes was found, showinghe existence of micelle-like domains of COC molecules. It wasound that as the COC content that is embedded into compositeembranes was increased, the hemocompatibility was improved

ia a marked drop in platelet adhesion. It could be assertedhat the reason behind this phenomenon is that the existencef COC molecules acts like phospholipid moieties of plasmaembranes, resulting in better hemocompatibility (Morimoto et

l., 2002; Nakabayashi and Williams, 2003; Yang et al., 2003).oreover, Kim et al. reported that grafting of polyethylene

xide on a material surface resulted in more blood compati-ility, due to an additional chain motion (molecular cilia) effect

(A) 1000× magnification and (B) 2000× magnification. Platelet (arrowheads)

Page 6: Platelet adsorption and hemolytic properties of liquid ...

122 M.-F. Shih et al. / International Journal of Pharmaceutics 327 (2006) 117–125

Fig. 4. SEM micrographs of platelets adhesion on the surface of PU–10% COC composite membrane after 60 min: (A) 1000× magnification and (B) 2000×magnification. Platelet (arrowheads) and vesicle-like objects (double arrowheads).

Fig. 5. SEM micrographs of platelets adhesion on the surface of PU–20% COC composite membrane after 60 min: (A) 1000× magnification and (B) 2000×magnification.

Fig. 6. SEM micrographs of platelets adhesion on the surface of Eb270 film after 60 min: (A) 1000× magnification and (B) 2000× magnification. Platelet (arrowheads).

Page 7: Platelet adsorption and hemolytic properties of liquid ...

M.-F. Shih et al. / International Journal of Pharmaceutics 327 (2006) 117–125 123

Fig. 7. SEM micrographs of platelets adhesion on the surface of Eb270–10% COC composite membrane after 60 min: (A) 1000× magnification and (B) 2000×magnification.

F COC composite membrane after 60 min: (A) 1000× magnification and (B) 2000×m

(pt1ias

3m

barlsopT

ig. 8. SEM micrographs of platelets adhesion on the surface of Eb270–20%agnification. Micelle-like objects (double arrowheads).

Kim et al., 2003). By means of shielding material surfaces fromlatelet (e.g. platelet glycoprotein Iib-IIIa (Gp Iib-IIIa) recep-ors), thrombotic effects may be retarded (Skarja and Brash,997; Spijker et al., 2002). Furthermore, the inclusion of COCn composite polymers results in less platelet adhesion; this islso likely due to an increase of membrane hydrophilicity onurface (Zhao et al., 2003; Lee et al., 2002).

.3. Blood clotting properties of polymer compositeembranes

The blood clotting study evaluates the actions of antithrom-ogenic activity of a material on human blood. Thentithrombogenic activity is quantitatively expressed by aelative parameter known as the blood clotting index (BCI). Aarger BCI value indicates an increase of compatibility. Fig. 9

hows the influence of COC contents in composite membranesn BCI values. An increase in the BCI value was found forolymers embedded with higher COC content (from 10 to 30%).his indicates that the addition of liquid crystal into polymer

Fig. 9. Effects of COC content embedded in composite membranes on bloodclotting index (BCI).

Page 8: Platelet adsorption and hemolytic properties of liquid ...

124 M.-F. Shih et al. / International Journal of Pharmaceutics 327 (2006) 117–125

Fig. 10. Effects of COC content embedded in composite membranes on P-Selectin levels. Various ratios of COC to composite membranes (0, 10, 20 and30%, w/w) were incubated with platelet-rich plasma (PRP). Statistics are shownf#

s

ictoccbp

3m

aecasa

PfiImtewtSCub

3m

p

Fy

otTaib5rpsfCHa

4

pgicb

R

A

G

G

K

or *p < 0.05, **p < 0.01 and ***p < 0.005, compared to control and #p < 0.05 and#p < 0.01 compared to the 0% addictive groups. The P-Selectin in PRP bloodample was taken as a control.

mproves biocompatibility with human blood. This result islearly attributed to the presence of a liquid crystal phase onhe surface of composite membranes that prevents the adhesionf platelets. Moreover, the addition of COC in PU or Eb270omposite membranes shows superior blood compatibility inomparison with PMMA composite membranes. This mighte due to the formation of different surface textures on eacholymer compound, as shown in SEM images.

.4. Functional assay for activation of platelets: P-Selectineasurements

P-Selectin is a surface-component of platelet granules thatppears when platelets are stimulated and activated (McEvert al., 1989). P-Selectin has a positive correlation with plateletount (Rand et al., 2003). Therefore, to directly quantify thentithrombotic effects mediated by platelets, P-Selectin mea-urements were performed to determine the level of plateletctivation.

In Fig. 10, we demonstrate the production of P-Selectin inRP blood samples that were incubated with three pure polymerlms, as well as P-Selectin formation on composite membranes.n pure polymer films, incubation with PMMA resulted in for-ation of extensive amounts of P-Selectin. This suggests that

here is substantial platelet activation, making PMMA the poor-st performer in biocompatibility. Fifty-percent less P-Selectinas produced in incubation with Eb270 and PU, indicating that

he biocompatibility of Eb270 is close to PU. Moreover, the P-electin level was decreased proportionally with embedding ofOC contents, again showing that liquid crystal COC can betilized to improve the compatibility of a material with humanlood.

.5. The hemolysis properties of polymer composite

embranes

Good biocompatibility with human blood not only concernslatelet adhesion and activation, but also involves the hemolysis

L

ig. 11. Effects of COC content embedded in composite membranes on hemol-sis ratio (HR) values.

f red blood cells (RBC). The hemolysis ratio (HR) representshe extent of RBC broken by the sample in contact with blood.he greater the value of HR, the more broken RBC is. Therefore,smaller HR value translates into increased blood compatibil-

ty of biomaterials. It is known that the HR value of acceptableiomaterials, required for medical applications, must be below% (Zhou and Yi, 1999). The effects of COC content on HRatios and comparisons among the three composite membranesrepared (PU, Eb270 and PMMA) are shown in Fig. 11. Thistudy definitively shows that the HR value decreases in three dif-erent composite membranes as COC content increases. GreaterOC contents result in lower HR values. The relationship ofR values and biocompatibility of polymers can be expressed

s: PU/COC > Eb270/COC > PMMA/COC.

. Conclusion

In this work, an analysis of biocompatibility properties ofolymeric materials to human blood is demonstrated and showsood correlation between qualitative and quantitative data. Then vitro studies also suggest that it may be reasonable to use liquidrystal COC as a mean of surface modification to improve thelood compatibility of polymers.

eferences

ldenhoff, Y.B.J., Koole, L.H., 2003. Platelet adhesion studies on dipyridamolecoated polyurethane surfaces. Eur. Cell. Mater. 5, 61–67.

orbet, M.B., Sefton, M.V., 2004. Biomaterial-associated thrombosis: roles ofcoagulation factors, complement, platelets and leukocytes. Biomaterials 25,5681–5703.

runkemeier, J.M., Tsai, W.B., Horbett, T.A., 1998. Hemocompatibility oftreated polystyrene substrates: contact activation, platelet adhesion, and pro-coagulant activity of adherent platelets. J. Biomed. Mater. Res. 41, 657–670.

im, Y.H., Han, D.K., Park, K.D., Kim, S.H., 2003. Enhanced blood compat-ibility of polymers grafted by sulfonated PEO via a negative cilia concept.

Biomaterials 24, 2213–2223.

ee, J.H., Oh, S.H., 2002. MMA/MPEOMA/VSA copolymer as a novel blood-compatible material: effect of PEO and negatively charged side chains onprotein adsorption and platelet adhesion. J. Biomed. Mater. Res. 60, 44–52.

Page 9: Platelet adsorption and hemolytic properties of liquid ...

al of

L

L

L

M

M

M

N

N

R

S

S

T

T

Y

M.-F. Shih et al. / International Journ

ee, H.J., Hong, J.K., Goo, H.C., Lee, W.K., Park, K.D., Kim, S.H., Yoo, Y.M.,Kim, Y.H., 2002. Improved blood compatibility and decreased VSMC pro-liferation of surface-modified metal grafted with sulfonated PEG or heparin.J. Biomater. Sci. Polym. Ed. 13, 939–952.

in, S.Y., Ho, C.J., Li, M.J., 2001. Precision and reproducibility of tempera-ture response of a thermo-responsive membrane embedded by binary liquidcrystals for drug delivery. J. Control. Release 73, 293–301.

in, S.Y., Lin, H.L., Li, M.J., 2002. Manufacturing factors affecting the drugdelivery function of thermo-responsive membrane prepared by adsorptionof binary liquid crystals. Eur. J. Pharm. Sci. 17, 153–160.

akai, M., Csanyi, E., Nemeth, Z., Palinkas, J., Eros, I., 2003. Structure anddrug release of lamellar liquid crystals containing glycerol. Int. J. Pharm.256, 95–107.

cEver, R.P., Beckstead, J.H., Moore, K.L., Marshall-Carlson, L., Bainton,D.F., 1989. GMP-140, a platelet alpha granule protein, is also synthesizedby vascular endothelium and is located in Weibel–Palade bosies. I. Clin.Invest. 84, 92–99.

orimoto, N., Iwasaki, Y., Nakabayashi, N., Ishihara, K., 2002. Physical prop-erties and blood compatibility of surface-modified segmented polyurethane

by semi-interpenetrating polymer networks with a phospholipid polymer.Biomaterials 23, 4881–4887.

akabayashi, N., Williams, D.F., 2003. Preparation of non-thrombogenic mate-rials using 2-methacryloyloxyethyl phosphorylcholine. Biomaterials 24,2431–2435.

Z

Z

Pharmaceutics 327 (2006) 117–125 125

esseem, D.I., 2001. Formulation and evaluation of itraconazole via liquid crys-tal for topical delivery system. J. Pharm. Biomed. Anal. 26, 387–399.

and, M.L., Leung, R., Packham, M.A., 2003. Platelet function assays. Transfus.Apheresis Sci. 28, 307–317.

karja, G.A., Brash, J.L., 1997. Physicochemical properties and platelet inter-actions of segmented polyurethanes containing sulfonate groups in the hardsegment. J. Biol. Mater. Res. 34, 439–455.

pijker, H.T., Busscher, H.J., van Oeveren, W., 2002. Influence of abciximab onthe adhesion of platelets on a shielded plasma gradient prepared on polyethy-lene. Thromb. Res. 108, 57–62.

edder, T.F., Steeber, D.A., Chen, A., Engel, P., 1995. The Selectins: vascularadhesion molecules. FASEB J. 9, 866–873.

u, L., Li, M., Mou, S., Zhou, C., 2001. Preparation and blood compatibil-ity of polysiloxane/liquid-crystal composite membranes. Biomaterials 22,2595–2599.

ang, Z.M., Wang, L., Yuan, J., Shen, J., Lin, S.C., 2003. Synthetic stud-ies on nonthrombogenic biomaterials 14: synthesis and characterization ofpoly(ether-urethane) bearing a Zwitterionic structure of phosphorylcholineon the surface. J. Biomater. Sci. Polym. Ed. 14, 707–718.

hao, C., Liu, X., Nomizu, M., Nishi, N., 2003. Blood compatible aspects ofDNA-modified polysulfone membrane-protein adsorption and platelet adhe-sion. Biomaterials 24, 3747–3755.

hou, C., Yi, Z., 1999. Blood-compatibility of polyurethane/liquid crystal com-posite membranes. Biomaterials 20, 2093–2099.

Page 10: Platelet adsorption and hemolytic properties of liquid ...

本文献由“学霸图书馆-文献云下载”收集自网络,仅供学习交流使用。

学霸图书馆(www.xuebalib.com)是一个“整合众多图书馆数据库资源,

提供一站式文献检索和下载服务”的24 小时在线不限IP

图书馆。

图书馆致力于便利、促进学习与科研,提供最强文献下载服务。

图书馆导航:

图书馆首页 文献云下载 图书馆入口 外文数据库大全 疑难文献辅助工具


Recommended