Date post: | 30-Dec-2015 |
Category: |
Documents |
Upload: | ciaran-oneil |
View: | 118 times |
Download: | 1 times |
pNeuton Transport Ventilator
mini Ventilator
Self Guided Tour
Program Objectives
This program is a self guided tour of the
pNeuton mini Ventilator.
At the end of this tour you will be able to:
Describe the main operating features of the ventilator Identify the controls and connections Describe the patient circuit Set-up the ventilator for patient operation Define various clinical environments for ventilator application Clean and maintain the ventilator
Features and Controls – Side Panel
Driving Gas Inputs
Connections for high pressure oxygen and medical grade compressed air source gas
Always use a high flow gas source that can provide 55 + 15 psi (3.8 + 1 bar) pressure at a minimum of 20 L/min
Insure that the compressed air is clean and free of water
Features and Controls – Side Panel
Remote Alarm Output
Remote Alarm connection A properly configured
Remote Alarm will indicate an active ventilator alarm condition in a remote location
Features and Controls – Side Panel
Patient Connection
Connection for the large bore hose of the patient circuit
Features and Controls – Side Panel
Prox Pressure
Patient circuit medium hose connection
Connects to the blue tube from the proximal patient pressure port on the patient circuit
Features and Controls – Side Panel
Exp Valve
Patient circuit small hose connection
Connects to the clear tube for the expiratory valve on the patient circuit
Features and Controls – Side Panel
Audible Alarm Audible alarm sound generator Used for patient disconnect, high
pressure and low gas supply alarms
Will not turn off during low gas supply alarm
Do NOT put tape over the holes, push sharp objects into the holes or interfere in any way with the alarm
The alarm is for patient protection. If you suspect the alarm is not operating properly have the unit serviced immediately
Features and Controls – Front Panel
PEEP / CPAP Control
Adjusts the PEEP / CPAP from 0 to 20 cm H2O
Operates in either mode Automatic sensitivity Oxygen concentration is
maintained at set % Oxygen control
Features and Controls – Front Panel
Peak Pressure Control
Limits peak airway pressure from 15 to 60 cm H2O
Operates only in the IMV + CPAP mode
Used to set the pressure limit during mandatory breath delivery
Features and Controls – Front Panel
Continuous Flow Control
Sets baseline continuous flow for 6, 8, 10, 15 or 20 L/min
Operates in either mode Used to set the flow used for both
spontaneous breaths and mandatory breath delivery
Features and Controls – Front Panel
Inspiratory Time Control
Sets inspiratory time for mandatory breaths from 0.25 to 2.0 seconds
Operates only in the IMV + CPAP mode
Can be used to preset tidal volume using the tidal volume chart on the top of the ventilator
Features and Controls – Front Panel
Expiratory Time Control
Sets expiratory time of mandatory breaths from 0.25 to 20 seconds
Operates only in the IMV + CPAP mode
Use with the respiratory rate chart on the top of the ventilator to determine mandatory breath breath rate
Features and Controls – Front Panel
% Oxygen Control
Delivered oxygen choice from 21 to 100%
Operates in either mode Sets oxygen
concentration of the continuous flow
Features and Controls – Front Panel
Visual Alarm Indicator
Illuminates for an active alarm condition
Alarm warns of patient disconnect after 10 seconds
Alarms for high pressure and low gas supply
Will turn off when alarm is corrected or Reset / Silence Button is pressed
Will not turn off during low gas supply alarm
Features and Controls – Front Panel
Alarm Reset / Silence Button
Press button to silence the alarm system for 25 seconds
Every time the button is pressed the alarm silence period will reset for 25 seconds
Features and Controls – Front Panel
Mode Control
Activates IMV + CPAP mode for delivery of ventilator breaths
Mode is the same as CMV + PEEP
CPAP mode provides continuous flow and CPAP for spontaneous breathing patients
No ventilator breaths in the CPAP mode
Features and Controls – Front Panel
Pressure Gauge
Represents pressure in patient circuit at the patient wye of the circuit
All patient alarms are sensed at the patient wye
Features and Controls – Front Panel
High Pressure Alarm Control
Sets the high pressure alarm from 15 to 70 cm H2O
Independent audible and visual alarm from the Peak Pressure Control
Patient Circuit
Patient circuit is a disposable, dual limb system All hoses attach to the side of the ventilator Patient end connects for clinical intervention
Directly to an endotracheal tube Use NIV kit to provide nasal prong or nasal mask
application in neonates and infants Use full face CPAP mask (non-vented)
Patient Circuit
Which size circuit - Neonatal or Pediatric? The neonatal circuit uses 10 mm hoses and is designed for
patients up to 3 kg The pediatric circuit uses 15 mm hoses and is designed for
patients above 3 kg If your patient needs 10 L/min or more continuous flow to
meet spontaneous flow demand, you should use the pediatric circuit to minimize expiratory flow resistance
Patient Circuit
Use only Airon Corporation patient circuits Calibration of the PEEP / CPAP and Peak Pressure controls is
based on the Airon circuit and exhalation valve Using other manufacturers circuits / exhalation valve may
cause: errors in pressure delivery high expiratory flow resistance interference with CPAP setting
Program Objectives
pNeuton mini Ventilator self guided tour, objectives completed so far
Describe the main operating features of the ventilator
Identify the controls and connections Describe the patient circuit
With an understanding of pNeuton’s capabilities, let’s look at patient application.
Patient Application
Set Main Controls PEEP / CPAP can be set
to minimum Determine peak
ventilating pressure Set desired continuous
flow to meet patient needs Set Inspiratory Time Set Expiratory Time Choose % Oxygen Adjust High Pressure
Alarm for safety
Patient Application
Initial Set-up Attach Airon patient circuit
to ventilator Attach a test lung to the
patient connection on the circuit
Attach high pressure oxygen and air hoses to the side of the ventilator
Attach these hoses to the gas supplies
Patient Application
Begin Ventilation Turn ON gas supplies Alarm system will
immediately activate Press Alarm Silence button if
desired For IMV or CMV, set Mode
control to IMV+CPAP Ventilator will start to provide
breaths at the settings on the controls
Patient Application
Set PEEP / CPAP Adjust the PEEP / CPAP
control to the level desired Adjust continuous flow to
meet patient demand – not too high to create expiratory flow resistance – not to low to cause pressure swings and high patient effort Use 6 – 8 for neonates Use 8 - 10 for infants Use 10 – 20 for children
Ventilator Operation
Alarm System Functionality Alarm preset to activate within 10 seconds Pressing the Alarm Silence button equals 25
second delay Disconnect Alarm is looking for a specific
minimum pressure of at least 3 cm H2O at all times
If there is not at least 3 cm H2O CPAP or mandatory breaths pressure of 3 cm H2O, the alarm will sound
High pressure alarm activates whenever the pressure in the circuit exceeds the alarm control setting.
Input Gas Pressure below 40 psi (2.8 bar) will activate the alarm
Ventilator Operation
Peak Pressure Limit Operates to limit the peak airway
pressure during mandatory breaths
For volume ventilation, set the Peak Pressure 10 cm H2O above the pressure needed to deliver the set Tidal Volume
Peak Pressure control is not an alarm. Use the High Pressure Alarm control for this function
Ventilator Operation
% Oxygen Control Set % Oxygen control to the
desired oxygen level Accuracy is plus / minus 3% The ventilator uses the oxygen
gas supply for internal operation so high oxygen concentrations may deplete tanks quicker
Clinical Issues
Neonatal Ventilation Inspiratory time setting 0.25 to 2.0 sec Use low continuous flow for proper
breath delivery and minimize expiratory flow resistance
Inspiratory time, flow and pressure (along with lung compliance / resistance and ET-Tube leak) will determine tidal volume to the patient
Use the respiratory rate chart on top of the ventilator to estimate ventilator rate
Use NIV kit for nCPAP Monitor closely for patient synchrony
Clinical Issues
Non-invasive Ventilation (NIV) Use Airon NIV kit with nasal prongs or
masks Proximal pressure measured will
display on pressure gauge and be used by alarm system
Effective patient interface seal will insure proper alarms
Rotate from nasal masks to prongs every few hours for patient comfort and proper skin care
Clinical Issues
Pediatric Volume Ventilation
Set the Peak Pressure to a safe pressure level
Refer to the Tidal volume chart on the top of the ventilator for I-Time / Flow combinations to determine the proper settings for the tidal volume desired
Adjust Inspiratory Time control / flow to increase or decrease tidal volume - more flow or time equals larger tidal volume
Adjust PEEP / CPAP to established baseline pressure
Clinical Issues
MRI Operation pNeuton mini is compatible with MRI
scanners up to 3 Tesla in strength While tests show that pNeuton functions at
the bore of the MRI unit, a minimum proximity of 12 inches (0.3 meter) from the bore is recommended
Safe MRI practice calls for all devices used in the proximity of a MRI scanner, including pNeuton mini, be anchored to prevent inadvertent movement
Clinical Issues
Altitude Operation The pNeuton mini may be used for air
transports in unpressurized cabins up to 15,000 feet
Inspiratory time will increase and expiratory time will decrease as altitude increases
Pressure settings will not change as altitude changes
Always use external monitors and pulse oximeters to monitor your patient
Patient Safety
pNeuton - Patient Safety Features
Audible and Visual Alarms for patient disconnect, high pressure and low supply gas pressure
High Pressure Release – pressure is relieved at the Peak Pressure control setting
Internal Safety Pressure Release – limits circuit pressure to 70 cm H2O, regardless
of the setting of the Peak Pressure control Anti-Suffocation System – in the event of
ventilator malfunction an internal valve will allow ambient air to enter the patient circuit
Cleaning and Maintenance
Clean ventilator between patient use applications using a soft cloth and a mild detergent / disinfectant
Do not immerse the ventilator in water Patient circuit is disposable, single use
application and should not be cleaned, disinfected or reused
Ventilator should receive preventative maintenance every 2 years
Program Objectives
pNeuton mini Ventilator self guided tourcompleted; you should now be able to:
Describe the main operating features of the ventilator Identify the controls and connections Describe the patient circuit Set-up the ventilator for patient operation Define various clinical environments for ventilator application Clean and maintain the ventilator
Next Step – consider reviewing the pNeuton Case Study and completing the competency evaluation
Contact Information
Thank you for you interest in the pNeuton mini Ventilator
For more information visit www.AironUSA.com
or call Airon Corporation at
USA + Canada toll free 888-448-1238International +1-321-821-9433