VARIAN Medical Systems
Portaldosimetrie mit ´Portal Vision´
© Varian Medical Systems, 2005
Varian’s QA solution
VARIAN Medical Systems March 2005 Slide 2
Die Vision
Goal
Verify the dose distribution Each IMRT field, without patient in beam
PortalVision LC250 and aS500
Increase efficiency for the Physicist Commission IMRT sooner
Ongoing pre-treatment QA
Reduce time to evaluate a plan
Less time/plan = Verify more plans
VARIAN Medical Systems March 2005 Slide 3
Die Zukunft...Asymmetric Jaws
Focus
MLC / CustomField Shaping
Patient
TreatmentCouch
Quantitative dosimetric QA ofPatient Treatment, including3D reconstruction of multi-fieldIMRT
Central Dose
Absolute dose=100 cGy delivered?
VARIAN Medical Systems March 2005 Slide 4
Traditional Dosimetry Tools
Measurement tools Ion chamber
Diodes
Film
Analysis software
Challenges
Non-linear sensitometry of film
Image registration; accuracy variations
User errors
Labor intensive, time consuming (2-4 hr. / plan)
VARIAN Medical Systems March 2005 Slide 5
Traditional Film Dosimetry
Expose film
Process film
Scan film
H&D curvecalibration
Expose film
Process film
Scan film
Compareresults
Import dose image
Align images
Calibration Planned Field Analysis
VARIAN Medical Systems March 2005 Slide 6
Portal Dosimetry
CalibrateImager
Exposeimager
Compareresults
Importdose image
Align images
Expose film
Process film
Scan film
Process film
Scan film
Calibration Planned Field Analysis
PV
Vision
Eclipse
PV
VARIAN Medical Systems March 2005 Slide 7
Portal Dosimetry
Dosimetry is a Process
‘Portal Dosimetry’ is a system capability (TPS, PV,VarisVision)
‘Dosimetry Workspace’access in PV (standalone)and Vision Review
Portal Dosimetry is a set of capabilities whichprovides Accumulation of relative dose at the PV imager position
Dose evaluation, acquired dose vs. TPS-predicted dose
For the application of Pre-treatment QA
Beam through air only (thru phantom = future capability)
No patient in beam
VARIAN Medical Systems March 2005 Slide 8
Portal Imager
Asymmetric Jaws
Focus
DMLC
Dosimetry
Patient
TreatmentCouch
Portal Dosimetry
Application
Pre-Treatment Verificationwithout Phantom
Qualitative Comparison
FDA 510(k)
PredictedPortal Dose
AcquiredPortal Dose
VARIAN Medical Systems March 2005 Slide 9
The Future?Machine QA pre-Treatment QA in vivo QA
Portal Dosimetry Applications
EPID/Film
Q
A
Test
Detector
Asymmetric Jaws
Focus
MLC / CustomField Shaping
Patient
TreatmentCouch
PortalDose
Tx
Plan
Planvs
ActualConsistency
PortalDose
VARIAN Medical Systems March 2005 Slide 10
Treatment Methods Supported
Treatment Type Overall Dose DosePrediction Acquisition
IMRT sliding windows Yes Yes Yes
Step-and-shoot Yes Yes Yes
Static field for machine QA Yes Yes Yes
Dynamic wedge No No Yes
VARIAN Medical Systems March 2005 Slide 11
EPID
TPS
Varis/Vision
Dosimetric Image
Visio
n
Visio
nTP
S
Fluence map+ dose matrix
Portal Dose Prediction
USERPROCESS
VARIAN Medical Systems March 2005 Slide 12
Eclipse (TPS) role in Dosimetry
Patient
Treatment Plan
Verification Plan
Dose prediction file for each field
Gantry 0 degrees for all fields (Plan angles possible)
User-defined imager position (SAD)
Result–predicted dose matrix ( image ) absorbed inPV cassette in ABSOLUTE values
VARIAN Medical Systems March 2005 Slide 13
Portal Vision ( EPID) role in Dosimetry
Calibration (Treat or Standalone PV) Calibration of detector at each energy, dose rate, and
2 distances
Dose normalization for LC250
Acquisition (Treat or Standalone PV) Integrated Image acquisition mode and sequence template
Automated Density Image to Dose Image conversion
Off-line integration of split fields (multiple image)
VARIAN Medical Systems March 2005 Slide 14
FDA 510(k) - Images courtesy of KFJ Hospital, Vienna Austria
Portal Vision (EPID) Acquisition
VARIAN Medical Systems March 2005 Slide 15
Portal Vision (EPID) AcquisitionIntegrated DoseReference Image
VARIAN Medical Systems March 2005 Slide 16
VarisVision (R&V) role in Dosimetry
Dosimetry workspace in Review (Vision orStandalone PV)
Dose Image Display
Dose Images Alignment
Dose Measurement & Evaluation
Point dose
Line Profile
Dose difference
Gamma evaluation
Off-line integration of split fields (multiple image)
VARIAN Medical Systems March 2005 Slide 17
Dosimetry Review: Dose EvaluationTraditional Dose Evaluation Tools: Line Profile + Point Dose
VARIAN Medical Systems March 2005 Slide 18
Dosimetry Review: Dose EvaluationGamma Evaluation showing good correlation
Gamma > 1
VARIAN Medical Systems March 2005 Slide 19
Dosimetry Review: Dose EvaluationGamma Evaluation showing mismatch
Gamma > 1Gamma > 1
VARIAN Medical Systems March 2005 Slide 20
Dosimetry Summary
IMRT adoption easier and faster Facilitates IMRT QA
Confirms that Clinac delivers the planned dose
Pre-treatment QA in a fraction of the time vs. traditional methods
Advantages Improved staff efficiency
Costs , billings
Facilitates good patient care
VARIAN Medical Systems March 2005 Slide 21
Conclusion
Acceptance testing, commissioning and QA of IMRTinvolves new concepts specific to IMRT.
Standards and guidelines are not yet established.
Your clinical judgment is important to patient safety.
In general, you should expect IMRT planning anddelivery systems to perform as well as 3DCRTplanning and delivery systems.
VARIAN Medical Systems
Vielen Dank für Ihre Aufmerksamkeit