Portex® Thermovent® T2 HMEProviding a clear choice
RESPIRATORY CARE
Portex® Thermovent® T2 HMEProviding a clear choice
Portex® Themovent® T2 HME• Sterile,singleuseheatandmoistureexchanger.
• Provideseffectivehumidificationforspontaneously
breathingpatientswhoseupperairwaysare
bypassedbyatracheostomyortrachealtube.
• Devicecanbeusedforupto24hours.
Oxygen Line Accessory• NewoxygenaccessorylineavailableforthePortex®
Thermovent®T2.
• Allowsacomfortableandeasywayofproviding
oxygenenrichedhumidification.
Indications for use:
• Foruseonadultandpaediatricpatientswithatidal
volumemorethan70ml.
• Reductionofheatlossthroughthebroncho-
pulmonarytree.
• Helpingtopreventthickenedsecretionsand
changesinlungfunction.
• Supplyingadditionaloxygenwhenrequired.
• Facilitatingtheremovalofsecretions.
• Preservinghumidityandminimisingdamageto
trachealepithelialcells.
Smooth,roundededgesforpatientcomfort.
OriginalPortex®Thermovent®T2papermediaprovideshighlevelperformanceheatandmoistureexchangetopatient.
Simpleopeningportallowsquickandeasyaccessforasuctioncathetertoaidtheremovalofsecretions.
Clearhousinggivesanaestheticappearanceandallowsvisualisationofcontamination.
Integratedoxygenportcanprovideupto69%oxygenconcentrationstothepatientviaasimpleconnection.
RESPIRATORY CARE
ThePortex®Thermovent®T2deviceshouldberemovedfromthetracheostomytubepriortodisconnectingtheoxygenlinetoavoidtheriskofdecannulation.
3
8
Formoredetailedinformation,pleaserefertotheinstructionsforuse.
9
Insertthesuctioncatheterandcommencesuctioning.Takecarenottoscrapesecretionsontotheportdoor.
Whensuctioningiscomplete,removethesuctioncatheterandclosethesuctionportdoor.
54
Thedeviceshouldbeorientatedwiththeoxygenportonitsundersidewhenplacedonthe15mmconnectorofthetracheostomytube.
Whensuctioningisrequired,openthesuctionportdoortoallowinsertionofthesuctioncatheter.
Ifsupplementaryoxygenisrequired,thePortex®Thermovent®T2OxygenLineAccessorycanbeconnectedtoasuitableoxygensource.
1 2
7
6
Thesuctionportshouldremainclosedwhennotinuse.
Theoxygenflowshouldbecontrolledaccordingtothepatient’srequirements.
Using the Portex® Thermovent® T2 HME with a Tracheostomy Tube
Moisture Output25mgl-1H2O@15breath/minTidalvolume500ml
O2 l/min 0 1 2 3 4 5 6 7 10 15O2 % 21 27 33 38 42 46 50 53 62 69
Description Product CodePortex®Thermovent®T2Heat&MoistureExchanger 100/570/022
Portex®Thermovent®T2OxygenLineAccessory 100/575/022
For further information on our full range of HME devices, please visit www.smiths-medical.com
Portex® Thermovent® T2 HME
FlexibletubingforconvenientsupplyofoxygentothePortex®Thermovent®T2
Oxygen Flow Rates / Inspired Oxygen Concentration
Specifications
Dead Space11ml
ThePortex®Thermovent®T2hasinbuiltsuctionandoxygenports.
Tidal Vol. RangeTidalVolumeRange 70-500ml
Resistance to Flow0.3hPa(cmH2O)atgasflowrate0.5lsec-1(30lmin-1)[0.6cmH2Ol-1sec-1]
0.7hPa(cmH2O)atgasflowrate1.0lsec-1(60lmin-1)[0.7cmH2Ol-1sec-1]
1.2hPa(cmH2O)atgasflowrate1.5lsec-1(90lmin-1)[0.9cmH2Ol-1sec-1]
Ordering Information
Smiths Medical International Ltd.1500 Eureka Park, Lower Pemberton, Ashford, Kent, UK, TN25 4BFPhone: +44 (0)1233 722351www.smiths-medical.com
Smiths Medical is part of the global technology business Smiths Group plc. Product(s) described may not be licensed or available for sale in all countries. Portex,ThermoventandtheSmithsMedicaldesign marks are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. ©2011 Smiths Medical. All rights reserved. LIT/AM315404/11