+ All Categories
Home > Documents > Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum...

Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum...

Date post: 21-Jun-2020
Category:
Upload: others
View: 0 times
Download: 0 times
Share this document with a friend
10
© A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z‐magnetization Knowledge of T1 enables you to make an informed choice of the duration of delay between scans. If too short a relaxation delay is chose, then the peaks of slowly recovering protons will be smaller than they should be, and it will not be possible to compare peak areas to learn about proton numbers. Also, we use the T1 as our first estimate for choosing mix times in NOESY spectra, and 3xT1 as our relaxation delay in NOESY spectra. Hence the need to know T1. Start with a beautiful 1d spectrum, and a calibrated pw90. (It turns out that the default value in the probe file is a good compromise for people in a rush or not bothered about perfection.) Under the Experiments choose 'convert current parameters to do... > Relax. Measurements > T1' This will retain your current sw , tof and pw90 , but it will add a few features to the experiment to support measurement of the magnetization's recovery after inversion. To see the pulse sequence click Sequence [1] in Figure 1 . The first pulse is a 180 ° pulse (p1=2*pw90) that will invert magnetization [2]. This places it in a high‐energy state with respect to the field, so it will 'decay' back to the resting state. Time is allowed for this process, in the form of a delay called d2 [3]. So far all magnetization is still along Z, but its amplitude is recovering with time. To determine what the amplitude is after a given time d2, we use a 90 ° pulse (pw) [4] to 'read' our magnetization by rotating it into the XY plane where our receiver coils detect it [5]. We cannot monitor recovery continuously, instead we repeat the experiment 6 or more times, waiting a different amount of time d2 in each repeat before executing the read pulse. Thus we end up with a series of 1d spectra, each representing a different duration of recovery after the inversion pulse. In Acquire>Defaults, a set of three boxes invites you to enter first‐guesses of the T1s you sample may have, and Varian uses these to set the delays in the experiment to perform a measurement (you will have to use the scroll bar on the RHS of the panel to reveal these) [6]. Set T1 Mode to inversion recovery [7] (don't worry if the bar stays blank), then set T1 min to the shortest T1 you expect to be present among your resonances (say 0.2 sec.: enter 0.2) [8], the longest T1 you anticipate (err on the long side here, say 3 (seconds)) and the time you are willing to devote to the experiment (eg. 0.1 (hours)). The min. and especially the max times will be used to choose an array of d2 values ranging from very short to something on the order of your maximum estimate. You can see these by clicking on 'Arrays'. [9]. The experiment takes a long time in part
Transcript
Page 1: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

©A.‐F.Miller2010 T1Measurement 1

Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1.

Strychnine,dissolvedCDCl3TheT1isthecharacteristictimeofrelaxationofZ‐magnetizationKnowledgeofT1enablesyoutomakeaninformedchoiceofthedurationofdelaybetweenscans.Iftooshortarelaxationdelayischose,thenthepeaksofslowlyrecoveringprotonswillbesmallerthantheyshouldbe,anditwillnotbepossibletocomparepeakareastolearnaboutprotonnumbers.Also,weusetheT1asourfirstestimateforchoosingmixtimesinNOESYspectra,and3xT1asourrelaxationdelayinNOESYspectra.HencetheneedtoknowT1.

Startwithabeautiful1dspectrum,andacalibratedpw90.(Itturnsoutthatthedefaultvalueintheprobefileisagoodcompromiseforpeopleinarushornotbotheredaboutperfection.)

UndertheExperimentschoose'convertcurrentparameterstodo...>Relax.Measurements>T1'

Thiswillretainyourcurrentsw,tofandpw90,butitwilladdafewfeaturestotheexperimenttosupportmeasurementofthemagnetization'srecoveryafterinversion.ToseethepulsesequenceclickSequence[1]inFigure1.

Thefirstpulseisa180°pulse(p1=2*pw90)thatwillinvertmagnetization[2].Thisplacesitinahigh‐energystatewithrespecttothefield,soitwill'decay'backtotherestingstate.Timeisallowedforthisprocess,intheformofadelaycalledd2[3].SofarallmagnetizationisstillalongZ,butitsamplitudeisrecoveringwithtime.Todeterminewhattheamplitudeisafteragiventimed2,weusea90°pulse(pw)[4]to'read'ourmagnetizationbyrotatingitintotheXYplanewhereourreceivercoilsdetectit[5].Wecannotmonitorrecoverycontinuously,insteadwerepeattheexperiment6ormoretimes,waitingadifferentamountoftimed2ineachrepeatbeforeexecutingthereadpulse.Thusweendupwithaseriesof1dspectra,eachrepresentingadifferentdurationofrecoveryaftertheinversionpulse.

InAcquire>Defaults,asetofthreeboxesinvitesyoutoenterfirst‐guessesoftheT1syousamplemayhave,andVarianusesthesetosetthedelaysintheexperimenttoperformameasurement(youwillhavetousethescrollbarontheRHSofthepaneltorevealthese)[6].SetT1Modetoinversionrecovery[7](don'tworryifthebarstaysblank),thensetT1mintotheshortestT1youexpecttobepresentamongyourresonances(say0.2sec.:enter0.2)[8],thelongestT1youanticipate(erronthelongsidehere,say3(seconds))andthetimeyouarewillingtodevotetotheexperiment(eg.0.1(hours)).

Themin.andespeciallythemaxtimeswillbeusedtochooseanarrayofd2valuesrangingfromveryshorttosomethingontheorderofyourmaximumestimate.Youcanseethesebyclickingon'Arrays'.[9].Theexperimenttakesalongtimeinpart

Page 2: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

©A.‐F.Miller2010 T1Measurement 2

duetothelongrelaxationtimeallowedbetweenscans[10].Thisischosentobe≈5timesyourestimatedlongestT1.Thedurationoftheexperimentalsoincreasesinproportiontothenumberofscanscollectedforeachvalueofthedelayd2.Therefore,wewillcutbackonthenumberofscansifpossible(Iwillchange4to1[11]).

Youmayalsowanttogointothe'Array'panelanddecreasethenumberofvaluesofd2used.Becauseweareexpectinganexponentialrecovery,weanticipatethatanexponentialarray[1]willbeabetterchoicethanalineararray(Figure2).Thusinthiscasesuccessivevaluesofd2are2xthepreviousvalue,ratherthanbeingsomevalueplusthepreviousvalue.Theincrementistheln(2).Ifyouwanttotripleeachvaluetogetthenextone,useanincrementofln(3)=1.1.IfIamrushingIchooseanarraysizeof6(theminimumallowed)andanincrement1.1.Theresultingarraycoversafactorofapproximately700.Thusforalongestestimateof3secondsIwantashortestvalueof0.004.IgobackandmakeadjustmentstothefirstvalueandincrementasneededtogetthelastvalueIwant.

MakesurethatGainisanumber,notautogain(Acquire>Acquisition)

Asalways,closetheArraybox,checkSequence,ShowTimeandthenAcquire.

Theresultshouldbeaseriesofspectrainwhichweseethatspinsareinitiallyinvertedandthengraduallyrecoverasthedelayd2ismadelonger(Figure3).

IntheLHSpanel,withthe'ArrayedSpectra'tabactive,changevaluesin'ChartDimensions'.ForVert.Pos.use20[1](from90)VertHeight300[2](from900).UnderOffsetVerticalchangethisinteractively(stepsizeof10)[3]with+10,itgives17.7(agooddisplay)andin"Show"checkthe'values'box[4].

Youseethatdifferentpeaksrecoverindifferenttimes.ThenullrepresentsapproxT1*0.7.Dividenulltimeby0.7toestimateT1≈1sec/.7=1.4secforpeakat6ppm.0.5s/.7=0.7Sforaliphatics.

InProcess>T1Analysis(Figure4[1]),clickonDisplayLastSpectrum[2].Itshouldbeuprightasshown.Ifitisnot,youwillhavetophaseitmanually(afteractivatingtheadditionaloptionsontheright‐hand‐sideofthegraphicaldisplaywindow[3]).Figure4showsacasewherethesoftwarepickedaridiculousnumberoflines.ActivatetheactionbuttonsontheRHSofthegraphicsdisplaywindow[3]andactivatethreshold(Figure5[1]).Movethethresholduptobemoreselective[2].Youcanalsousethecursorstorestrictyouranalysistooneregionofthespectrumbyplacingcursorsarounditandclickingthe'+'magnifyingglass[3].Thenclickon'DoT1Analysis'[4].AllthelinestheinstrumentpickedwillbeinthetableintheRHSbox[5].Ifyouscrollallthewaytotherightedgeoftheboxwiththenumbersinit[6],asecondverticalscrollbarwillappearinsidetheonevisibleinthefigure.Youmayonlybeabletoseethesmalldownarrowheadatthebottom.Clickonthattoscrollupanddown.First,allthepickedpeaksarelistedandeachisassignedan'index'number,whichwillbeusedbelowtoprovidetheT1forthecorrespondingpeak.Nextthereisasummaryoftheanalysisinwhicheachpeak'sT1isprovided

Page 3: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

©A.‐F.Miller2010 T1Measurement 3

alongwiththeassociatederror.Finallyadetailedanalysisfollowsforeachpickedpeak.

Usingthechemicalshiftsprovidedwiththespectrum(Figure6[1])inconjunctionwiththenumericaloutputtellingyouwhichindexnumbercorrespondstoeachchemicalshiftvalue[2]youcanselectafewlinesandforthemplotthepeakamplitudevs.durationofrecoveryd2.Nextto'DisplaySelectedFits'enterthelinenumbersforthelinesyouwouldlikeinformationon,forexample"1,10,36"[3].Then<return>.

Atimecourseofmagnetizationrecoveryisproduced(Figure7).

Don'tbelieveT1valuesderivedfromcurvesthatdon'tfitthedatawell.Don'tbelieveanyvalueswhoseerrorsaremorethan10%thevalue(readthenumericaloutputinthesummaryofanalysissection).Don'tbelieveaT1thataremorethanonethirdoftherecoverytimeyouusedbetweenscans(=d1+at).IftheT1syougetarelongerthan(d1+at)/3,repeattheexperimentwithalongerd1,roughly5timesthelongestT1yougot.

Togetacopyofthegraphicaloutput,usetheLHpanel's1Dtab(Figure8[1]),click'PrintScreen..."[2].

Inwindowthatopens(Figure9)selectPSland,(•)file,givenamewithpath(eg.che555/data/yourname/yourfile(omitthe.ps),POSTSCRIPT,Mono/Color...Save,thenClose

Asusual,youcansaveyourdatasetasa.fidfileusingthedisketteiconinupperleft.(Figure10)

Illustration of why the relaxation delay must be longer than T1 Ifyoudonotallowsufficienttimebetweenscansforresonancestorecover,theywillbesmallerinsubsequentscans.Ifthetotalrecycletime(d1+at)isshorterthantheT1sofsomeresonancesbutlongerthantheT1sofothers,thentheformerresonanceswillbeshrunkenmorethantheothers,inyourspectra.Thus,peakareaswillnolongerbeproportionaltothenumbersofprotonsassociatedwitheachpeak.Figure11comparestheresultsofusingfourdifferentd1valuesallincombinationwithanatof0.5.Thus,thetotalrecycletimeswere11.5,3.8,1.5and0.8s(ford1valuesof11,3.3,1.0and0.3s).Theresultsshowthatthe0.8srecycletimecausesaromaticresonances,inparticular,tobeproportionaltelysmaller(Figure12).NotethattheTMSlineisalsomuchsmallercomparedtotheotherlineswhenrecyclingisfast.

Theabovepartialsaturationofslow‐relaxingresonancescanbepreventedeitherbychoosingalonger(d1+at)orbyusingasmallertip‐anglepulse.Ifyouusea45°pulsethenyoudonothavetowaitaslongbetweenscans,becausemagnetizationdoesnotneedtorecoverasmuch.

Page 4: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

©A.‐F.Miller2010 T1Measurement 4

Usingasmallertipangleincombinationwithashorterrecycletimeallowsyoutogetthebestdataperhourofspectrometertime.Thisisinvaluableforweakspectrasuchas13C1ds,orsamplesthatarenotstable.

OnceyouknowT1thesoftwarecancalculatetheoptimalpulsewidthandtipanglegivenachoiceofd1.Inthecommandlinetype'ernst(3.4,13.6)'fortheexampleofacasewheretheT1is3.4s.andthepw90is13.6us.Thusthegeneralformofthecommandis'ernst(t1_est,pw90)'.Thiscausesthesoftwarewillupdatethepulsewidthandtipangleoftheexperimentformaximalsensitivityperhourofspectrometertime,giventhed1valueinyourparameterset.

Page 5: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

© Anne-Frances Miller, 2010

T1, Figure 1

[1]

[2]

[3] [4] [5]

[6]

[7]

[8]

[9][10][11]

© Anne-Frances Miller, 2010

T1, Figure 2

[1]

[2]

Page 6: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

© Anne-Frances Miller, 2010

T1, Figure 3

[1]

[2]

[3]

[4]

© Anne-Frances Miller, 2010

T1, Figure 4

[1]

[2]

[3]

Page 7: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

© Anne-Frances Miller, 2010

T1, Figure 5

[1][2]

[3]

[4]

[5]

[6]

[7]

© Anne-Frances Miller, 2010

T1, Figure 6

[1]

[2]

[3]

Page 8: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

© Anne-Frances Miller, 2010

T1, Figure 7

© Anne-Frances Miller, 2010

T1, Figure 8

[1]

[2]

Page 9: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

© Anne-Frances Miller, 2010

T1, Figure 9

© Anne-Frances Miller, 2010

T1, Figure 10

Page 10: Practicum 3, Fall 2010 Measuring the longitudinal ...A.‐F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved

© Anne-Frances Miller, 2010

T1, Figure 11

© Anne-Frances Miller, 2010

T1, Figure 12


Recommended