+ All Categories
Home > Documents > PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

Date post: 10-Apr-2018
Category:
Upload: najib-belnajib
View: 214 times
Download: 0 times
Share this document with a friend

of 50

Transcript
  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    1/50

    Achieving Ultra Fast, Low Cost

    Elemental Analyses in Compliance with

    EPA Protocols

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    2/50

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    3/50

    3

    Polling question 1

    What is your typical sample analysis time for a soil or sludge sample?

    Less than 1 minute

    Between 1 and 2 minutes

    Greater than 2 minutes

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    4/50

    4

    Todays Speakers

    Matthew Cassap

    Senior Application Specialist

    Thermo Fisher Scientific

    Jayme CuretApplications Scientist

    Thermo Fisher Scientific

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    5/50

    5

    Polling question 1 Results

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    6/50

    6

    What will be covered

    The sample introduction systems for ICP-OES and ICP-MS

    Simple steps for increasing sample throughput

    New developments in sample introductions systems and ICP-OES dataacquisition technology

    Advanced sample introduction systems becoming routine

    How these advances can be applied to your analysis High speed trend analysis

    Analysis using the 6010C EPA protocol

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    7/507

    ICP-OES and ICP-MS

    Have similar sample introduction systems which turn a liquid sample into

    an aerosol

    The aerosol is transported to the plasma where:

    The sample is atomized and then ionized

    Atoms and ions emit light

    ICP-MS quantifies the concentration of an element based on the amountof ions in the sample

    ICP-OES quantifies the concentration of an element based on theintensity of light emitted from an atom or ion

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    8/508

    Similarities in sample introduction

    The purpose of the sample

    introduction is to convert a liquid toan aerosol

    Via a nebulizer

    Typically the sample is pumped tothe nebulizer from its vessel by a

    peristaltic pump

    The sample is transferred to theplasma from the nebulizer via aspray chamber and transfer tube

    The spray chamber removes large

    droplets in the aerosol to preventplasma loading

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    9/50

    9

    Limitations of sample introduction pre nebulizer

    Manual movements between samples

    Use of a scientist

    Non consistent timings for wash and sampleuptake

    Long sample uptake times

    Distance from the sample probe to the nebulizer

    Limited pump speed

    Large amounts of sample used

    Memory effects from sticky and highconcentration elements

    Hg and B are considered sticky

    Unexpected high concentrations can cause carryover

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    10/50

    10

    Limitations of the nebulizer

    Small capillaries within the nebulizer

    Prone to clogging

    Delicate

    Samples with different physical properties create different amounts ofaerosol

    Viscosity

    Volatility

    Too many to choose from

    Is the correct nebulizer being used for solids content of the sample?

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    11/50

    11

    Limitations of sample introduction post nebulizer

    How efficient is the spray

    chamber at removinglarge droplets

    More important for ICP-MS where too muchsolvent in the plasma will

    cause interferences How large is the spray

    chamber

    Memory effects withsticky elements

    Sample introduction andwash out time

    Double-Pass Spraychamber

    Cyclonic Spraychamber

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    12/50

    12

    Limitations of the system as a whole

    Time consuming

    Can be over 50% of the total sample cycle time

    Wastes sample

    Sample remains in uptake tubing

    Typically less than 3% of the sample ends up in the plasma

    Source of every day problems Poor RSDs

    Poor sensitivity

    Contamination

    Drift

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    13/50

    13

    A fully optimized sample introduction system will;

    Have a short sample uptake time

    Use minimal sample

    Remove large droplets from the aerosol efficiently

    Have minimal memory effects

    Have a short sample uptake time

    Be stable Resistant to blockage

    And hopefully the correct result!!

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    14/50

    14

    Use of an auto sampler

    Will automate sample changing

    Provide consistent sample timings

    Sample uptake

    Sample wash

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    15/50

    15

    Close coupling of the auto-sampler accessory

    Ensure the auto-sampler is as close to the sample introduction systemas possible and reduce sample tubing lengths.

    Minimising sample uptake time

    Minimising wash time

    150 cm47 cm30 cm

    115 cm15 cm

    Typical

    Minimum

    Total = 227 cm

    Total = 177 cm

    Difference = 50 cm

    Saves ~15 s per sample at 2 mL/min uptake/wash with iCAP but the principle applies to alltechniques

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    16/50

    16

    Pump speed

    Peristaltic pump range

    Most pumps will have more thatone speed

    Use different speeds for differentparts of the analysis

    Normal analysis rate 50 rpm

    Fast pump capability for flush

    Saves up to 40 s per sample(including uptake and wash)

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    17/50

    17

    Just for ICP-MS

    As the ion optics and detector are in a vacuum chamber an interface is

    present in the instrument Use of two cones with small orifices

    Can be prone to clogging with high sample matrix

    Dilution of the sample

    Minimize the time the cones are exposed to the matrix

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    18/50

    18

    Software optimization

    The instrument software can also help to optimize the sampleintroduction time

    The Thermo Scientific iTEVA Software has the following features Intelligent Rinse

    Auto sampler Step A Head

    Software automatically detectswashout to baseline forselected analytes

    Non-productive time reduced;analysis time optimized

    Washout completed sooner

    Maybe no wash is needed?1

    10

    100

    1000

    10000

    100000

    000000

    0 20 40 60 80 100 120 140 160 180

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    19/50

    19

    Step-Ahead Sampling

    Step-ahead autosampler control

    send the autosampler probe to wash beforethe sample analysis is complete

    Residual sample used to complete theanalysis

    Sample line already primed with washsolution for washout

    Can save more than 1 minute per sample innon-productive time

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    20/50

    20

    Autosampler Step-Ahead

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    21/50

    21

    Recap simple optimization

    Simple steps can be taken to optimize sample introduction

    Shorten all tubing lengths

    Use a autosampler

    Use fast sample uptake and washout pump speeds

    Use a low volume spray chamber

    Use the software features

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    22/50

    22

    Advanced optimization of the system

    Use of vacuum to load the sample very quickly on to a loop

    Inject the sample from the loop in to the spray chamber

    Significantly shortens sample introduction times

    Reduces the amount of matrix that enters the plasma

    Key for ICP-MS, this will prevent cones from clogging

    Systems are commercially available and in use in routine environments

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    23/50

    23

    Introducing the dual loop system

    As mentioned single loop systems are used in routine labs

    There is still dead time associated with this configuration

    For a fully optimized system a dual loop system can be used inconjunction with the auto sampler Step A Head feature of the the iTEVASoftware.

    Whilst one loop the solution from one loop is being analyzed the otherloop is being filled ready for analysis

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    24/50

    24

    What happens during the analysis with two loops

    ESI SC4 DX FAST iTEVA Software

    Autosampler moves to sample 1

    Autosampler moves to sample 2

    Sample 1 flush time

    Sample 1 integration

    Sample 1 loaded on to loop

    Sample 1 injected to nebulizer

    Autosampler moves to wash

    Sample 2 loaded on to loop

    Sample 2 injected to nebulizer

    Autosampler moves to wash

    Autosampler moves to sample 3

    Sample 3 loaded on to loop

    Sample 3 injected to nebulizer

    Autosampler moves to wash

    Sample 2 flush time

    Sample 2 integration

    Sample 3 flush time

    Sample 3 integration

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    25/50

    25

    Two sample loops

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    26/50

    26

    Two sample loops

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    27/50

    27

    The Fast method

    Additional software is used to control the ESI SC 4 DX FAST

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    28/50

    28

    Smart use of the detector

    The duty of cycle of an instrument can also effect the analysis times

    Many instruments have optimized systems for different regions of thespectrum under analysis (mainly split in to UV and Vis)

    The in order that these regions are read by the detector and how thedetector operates is know as the duty of cycle

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    29/50

    29

    Traditional data acquisition for ICP

    Traditional acquisition mode reads the wavelengths like this:

    Vis replicate 1

    UV replicate 1

    Vis replicate 2

    UV replicate 2

    Vis replicate 3

    UV replicate 3

    Each time the ICP-OES switches view there is overhead time.

    For a three replicate analysis there are 6 transitions

    Therefore six lots of overhead time

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    30/50

    30

    Speed mode for optimized data acquisition

    Removal of the overhead time

    UV replicate 1,2,3

    One transition

    Visible replicate 1,2,3

    Allows for up to 30% more samples to be analysed

    Traditional

    Speed

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    31/50

    31

    Data Acquisition Mode Sprint mode

    Accessible on ALL iCAP 6500 models Uses the same sample replicate cycle as Speed mode

    Reduced detector overhead time by use of CSPI

    3 seconds per slit (for 0-20 wavelengths)

    4 seconds per slit (for 20-75 wavelengths)

    Typically achieves 3% RSDs on replicate measurements

    Enables fast data acquisition when using both UV & Vis slits

    High throughput wear metal applications

    High throughput Argri/Enviro applications

    High throughput Geochem screening

    Enhanced data acquisition for high speed analysis

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    32/50

    32

    How Cumulative Set Pattern Integration works

    Each wavelength is read in order

    If the counts are reaching the limit which the pixel can hold the countsare recorded and the pixel is rested and allowed to accumulate morecounts, if not the next wavelength is interrogated

    High intensity wavelengths (high concentration) will cause pixels to fillup quickly

    P lli i

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    33/50

    33

    Polling question 2

    Which of these is most important for your lab when thinking about

    instruments? Instrument sensitivity

    Speed of analysis

    Operating costs of the instrument

    Ease of use

    P lli i 2 R l

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    34/50

    34

    Polling question 2 Results

    I l i f h h l i l b

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    35/50

    35

    Implementation of the technology in your lab

    Two applications both environmental

    Agricultural trend analysis

    Extremely high sample numbers (1000s of samples a day) High sensitivity not critical

    An EPA analysis using the 6010C

    Moderate sample numbers (100s of samples a day)

    High sensitivity is key

    Examples use The Thermo Scientific iCAP 6500 ICP-OES

    EXI SC 4- DX FAST

    E l 1 T d E i t l l i

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    36/50

    36

    Example 1 - Trend Environmental analysis

    Analysis of surface soil for major, trace and micro-nutrients from arable

    land to determine fertiliser concentrations Analysis of foliage during the growing season to determine the

    concentration of fertiliser to be applied.

    Results of analysis are then mapped and the dose of fertiliser isautomatically changed as the farmer moves across the land

    T d E i t l A l i

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    37/50

    37

    Trend Environmental Analysis

    Sensitivity and speed

    Even 1s integration yields adequate detection limits for this type of

    analysis Stability

    Essential for long runs

    Full wavelength coverage allows selection of appropriate wavelengthfor expected concentration

    Matrix Tolerance

    Solid State generator, swing frequency

    Low cost of ownership

    Th l ti il l

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    38/50

    38

    The sample preparation- soil as an example

    This type of analysis is not regulated but the accepted

    method is

    Sample is dried overnight and ground

    ~5g aliquot is placed in to a 30ml vial

    20ml ammonium acetate of Mehlich 3 solution

    Shaken Filtered

    Analysed

    P t

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    39/50

    39

    Parameters

    Parameter Setting

    Pump speed 40rpm

    Nebulizer gas flow 0.65 L/min

    Spray Chamber Baffled mini cyclonic (25 ml)

    Integration times UV/Vis 5 second per region

    Number of replicates 1

    Analysis mode Sprint

    Loop volume 750l (total vol used 1.5ml)

    Loop fill time 1 second

    Total sample flush time 2.5 seconds

    High nebulizer gas flow reduces flush time

    Small spray chamber volume - reduces flush time and memory

    Small loop volume minimal sample usage

    S l l i

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    40/50

    40

    Sample analysis

    The iCAP 6500 ICP-OES wascalibrated

    60 samples analyzed

    Two different soils

    Determine repeatability

    Precision, accuracy

    Blanks

    To determine detection limits

    Look at carry over after samples

    A QC sample every ten samples

    Element High Std concmg/L

    Correlation

    B 249.773 nm 5 0.994

    Ca 227.547 nm 3000 0.999

    Cu 244.700 nm 10 1.000

    Fe 238.863 nm 100 0.992

    K 769.800 nm 500 0.989

    Mg 279.079 nm 300 0.995

    Mn 293.306 nm 50 0.995

    Na 818.326 nm 75 0.994

    P 213.618 nm 100 0.999S 182.034 nm 50 1.000

    Zn 206.200 nm 5 0.999

    Res lts

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    41/50

    41

    Results

    66 samples analyzed in 15 minutes - 13.6 seconds a sample

    Element Soil A mg/L Soil B mg/L QC Sample

    Measured mg/L True mg/L Recovery %B 249.773 nm 1.21 1.15 2.65 2.5 105.9

    Ca 227.547 nm 330 996 3238 3000 107.9

    Cu 244.700 nm ND 0.88 10.2 10 102.4

    Fe 238.863 nm 22.8 17.44 106 100 106.4

    K 769.800 nm ND ND 493 500 98.6

    Mg 279.079 nm 106 30.6 315 300 104.8

    Mn 293.306 nm 1.60 13.9 53.3 50 106.5

    Na 818.326 nm 11.8 16.9 81.4 75 108.6

    P 213.618 nm 14.9 37.7 98.0 100 98.0

    S 182.034 nm 15.7 9.35 49.2 50 98.3

    Zn 206.200 nm 0.33 49.6 5.08 5 101.7

    Example 2 US EPA Method 6010C

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    42/50

    42

    Example 2 US EPA Method 6010C

    Analysis of waters, soils and sludge following the

    digestion and filtration of soils and sludge

    Acidification of waters

    This method and variations of this method are widely used in NorthAmerica and around the world.

    Parameters

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    43/50

    43

    Parameters

    Parameter Setting

    Pump speed 40rpm

    Nebulizer gas flow 0.65 L/min

    Spray Chamber Baffled mini cyclonic (25 ml)

    Integration times UV/Vis 5 second per region

    Number of replicates 3

    Analysis mode Sprint

    Loop volume 2000l (total vol used 3 ml)

    Loop fill time 5 second

    Total sample flush time 8 seconds

    High nebulizer gas flow reduces flush time

    Small spray chamber volume - reduces flush time and memory

    Integration times longer before achieve the required sensitivity

    Method

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    44/50

    44

    Method

    The iCAP 6500 ICP-OES (Duo) was used for the analysis

    Axial view for UV region, Radial view for Visible region A Sc internal standard was used

    Axial view with Sc 227.318 nm internal

    std

    Radial view with Sc 361.384 nm internal

    stdAs 189.042 nm, Cd 226.502 nm,

    Co 228.616 nm, Cr 205.552 nm,

    Hg 194.227 nm, Mo 202.030 nm,

    Ni 221.647 nm, P 177.495 nm,

    Pb 220.353 nm, Sb 206.833 nm,Se 196.090 nm, Ti 190.856 nm,

    Zn 213.856 nm

    Ag 328.068 nm, Al 328.068 nm,

    B 249.678 nm, Ba 493.409 nm,

    Ca 315.887 nm, Cu 34.754 nm,

    Fe 259.940 nm, K 766.490 nm,

    Li 670.784 nm, Mg 279.553 nm,Na 588.995 nm, Si 251.611 nm,

    Sr 407.771 nm, Ti 334.904 nm,

    V 292.402 nm

    Spectral inferences and optical stability

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    45/50

    45

    Spectral inferences and optical stability

    Previous work demonstrates the iCAP 6500 ICP-OES (duo) ability to

    resolve the interferences mentioned in the 6010 method, with the use ofthe relevant interference check solutions

    Application note 40836

    Sample introduction parameters will not effect the spectral interferences

    Linear range

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    46/50

    46

    Linear range

    As the mode of data

    collection is different linearrange was investigates

    Sprint mode uses CSPI

    > 2 mg/L > 50 mg/L >500 mg/L

    Hg 194.227 nmAg 328.068 nm

    As 189.042 nmB 249.678 nm

    Ba 493.409 nm

    Cd 226.502 nm

    Co 228.616 nm

    Cr 205.552 nm

    Cu 34.754 nm

    Li 670.784 nm

    Mg 279.553 nm

    Mo 202.030 nm

    Ni 221.647 nm

    P 177.495 nm

    Pb 220.353 nm

    Sb 206.833 nm

    Se 196.090 nm

    Si 251.611 nmSr 407.771 nm

    Ti 334.904 nm

    Tl 190.856 nm

    V 292.402 nm

    Zn 213.856 nm

    Al 328.068 nmCa 315.887 nm

    Fe 259.940 nm

    K 766.490 nm

    Na 588.995 nm

    Detection limits and Sample timings

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    47/50

    47

    Detection limits and Sample timings

    Analysis time per sample for the 6010method was less than 50 Seconds

    Instrument detection limits compared toEPA estimates

    Element DL ug/L EPA ug/L Element DL ug/L EPA ug/L

    Ag 4 4.7 Mg 15 20

    Al 27 30 Mo 2 5.3

    As 12 35 Na 32 (19) 19

    B 3 3.8 Ni 5 10

    Ba 0.7 0.87 P 11 51

    Ca 17 (7) 6.7 Pb 3 28

    Cd 2 2.3 Sb 6 21

    Co 1 4.7 Se 9 50

    Cr 0.9 4.7 Si 12 17

    Cu 3 3.6 Sr 0.3 0.28

    Fe 9 (3) 4.1 Ti 4 5

    Hg 3 17 Tl 7 27

    K 60 - V 5 5

    Li 7 2.8 Zn 1 1.2

    Summary

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    48/50

    48

    Summary

    Simple optimization of sample introduction can reduce sample analysis

    times for ICP-OES and ICP-MS Use of accessories such as single loop ESI SC FAST can reduce these

    times further

    For ICP-MS there is the added benefit of the cones seeing less matrixreducing user maintenance

    Use of the ESI SC 4 DS FAST dual loop with the Thermo ScientificiCAP 6500 ICP-OES in Sprint mode can reduce sample introductiontimes significantly

    Typically more than 2 minutes per sample to less than 50 seconds for aEPA method!

    Question and Answer

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    49/50

    49

    Question and Answer

    Type your question in the

    Submit Question

    box below your slide window

    Thank You

  • 8/8/2019 PresentatioAchieving Ultra Fast, Low Cost Elemental Analysis

    50/50

    Thank You

    We Need Your Feedback!

    Please complete the post-webcast evaluation form

    that will immediately appear in your window.


Recommended