PROVEN PERFORMANCE FOR COST-EFFECTIVE STERILIZATION Using our original, innovative gas plasma technology, the STERRAD ® 100S System has been a pioneer in low-temperature sterilization for nearly twenty years. Recognized for its proven performance, the STERRAD ® 100S System is your cost-effective choice for dry, low-temperature sterilization. SAFETY • Low temperature gas plasma efficiently eliminates traces of residual hydrogen peroxide from materials and devices and the process by-products are non-toxic oxygen and water vapor. SIMPLICITY • Ease of use: one-button operation • No costly plumbing, ventilation, or monitoring required COMPATIBILITY • Proven claims* for a wide range of material and lumens • Capable of sterilizing rigid and flexible endoscopes, cameras, light cables, batteries, lenses, probes, microsurgical and general surgery instruments
Transcript
Page 1: Proven Performance for cost-effective sterilizationonbase.visole.net/base/DATASHEET_STERRAD_100S_-2017-06... · 2017. 6. 9. · Proven Performance for cost-effective sterilization

Proven Performance for cost-effective sterilization Using our original, innovative gas plasma technology, the sterraD® 100s system has been a pioneer in low-temperature sterilization for nearly twenty years. recognized for its proven performance, the sterraD® 100s system is your cost-effective choice for dry, low-temperature sterilization.

SAFETY

• Lowtemperaturegasplasmaefficientlyeliminatestracesofresidualhydrogenperoxidefrommaterialsanddevicesandtheprocessby-productsarenon-toxicoxygenandwatervapor.

SIMPLICITY

• Easeofuse:one-buttonoperation

• Nocostlyplumbing,ventilation,ormonitoringrequired

COMPATIBILITY

• Provenclaims*forawiderangeofmaterialandlumens

• Capableofsterilizingrigidandflexibleendoscopes,cameras,lightcables,batteries,lenses,probes,microsurgicalandgeneralsurgeryinstruments

Page 2: Proven Performance for cost-effective sterilizationonbase.visole.net/base/DATASHEET_STERRAD_100S_-2017-06... · 2017. 6. 9. · Proven Performance for cost-effective sterilization

Allrightsreserved.STERRADisaregisteredtrademarkofAdvancedSterilizationProducts.

*PleaserefertotheSTERRAD100SSterilizationSystemUser’sGuidepriortouse.Consultwiththemedicaldevicemanufacturerofinstrumentsorendoscopesyouintendtosterilizewithanyquestionsregardingmaterialscompatibility.

†PleasecontactyourASPrepresentativetodeterminemaximumchambercapacitybasedonyourfacility’sparticularsterilizationloadconfiguration(s). AD-100058-01-CT_A

FOR MORE INFORMATION OR TO PLACE AN ORDER:PLEASECONTACTYOURLOCALASPREPRESENTATIVE

TECHNICALSPECIFICATIONS

ShortCycle:54minutesLongCycle:72minutes

Lessthan55°C

Height 1664mm Width 762mm Depth 1046mm Weight 350kg

Confi guration Circular Chamber Volume 173L†

Shelf Dimensions Uppershelf:700mmx425mm Lowershelf:760mmx325mm

POWERUSAGE Connections 380-415VAC,50Hz,10A,3-phasecircuitwithseparate neutralandgroundconductors.CBAphaserotation

SERVICEREQUIREMENTS Inoperation,theSTERRAD®100SSystemshouldnotbeplacedcloserthan5cmtoawallattherearand/orsidesofthesystem.Thesterilizershouldbeinstalledinaspaceofsufficientsizetopermitaccesstoallfoursidesofthesystemwhenitisrolledfromtherearand/orsidewallsareatadistanceof3m.Serviceaccessrequiresaminimumclearanceof1monallsidesofthesystem.

Temperature 10°Cto40°C Humidity 30-75%relativehumidity(non-condensing)

TOTALCYCLETIME

CYCLETEMPERATURE

DIMENSIONS&WEIGHT

STERILIZATIONCHAMBER

OPERATIONALENVIRONMENT

HYDROGENPEROXIDEGASPLASMATECHNOLOGYPlasma is generated to safely eliminate residual H2o2

SAFE FOR INSTRUMENTS

system byproducts are oxygen and water

SAFE FOR ENVIRONMENT

vaporized Hydrogen Peroxide (H2o2) is introduced into a vacuum chamber

Sterility Assurance Level: 10-6

Contact for Ireland :

ASPc/oJ&JMedicalIrelandAirtonRd,TallaghtDublin24,Ireland

T:+35314665200F:+35314665340

Contact for South Africa :

ASPc/oJ&JMedicalPty.Ltd.SAPOBox273,MidrandHalfwayHouse1685RepublicofSouthAfrica

T:+27112651120F:+27112651189

Contact for Middle East :

ASPc/oJ&JMedicalMiddleEastDubaiHealthcareCity,J&Jbuilding,3rdfloorPOBox505080,Dubai,UnitedArabEmirates

T:+97144297235F:+97144297250

Contact for UK :

ASPc/oJ&JMedicalLtd.PinewoodCampus,NineMileRideWokinghamBerkshireRG403EW,England

T:+441344871081F:+441344871171

Contact for Egypt:

Johnson&JohnsonMedicalEgyptFloridaMall5thFloor1229SquareElSheikhAliGadElHakSt.HeliopolisCairo,Egypt

T:+20222685026F:+20222684674


Recommended