2
PSI CO2-Pro TM The PSI CO2-Pro TM is a compact, light-weight, plug-n- play pCO2 sensor that is designed for use on moorings (to 1000m), on profilers, in under-way mode, and in the laboratory. The PSI CO2-Pro TM is fitted with an IR detector and the new PSI pump-driven fast interface (patent pending). This revolutionary new interface provides an equilibrated gas stream to the detector. The PSI CO2-Pro TM is factory calibrated from 0 – 600 ppm with calibration burned into EPROM. To maintain accuracy, the detector module has a power interrupt or contact closure initiated automatic zero point calibration (AZPC). When the AZPC is initiated, the gas stream is routed through a CO2 absorbent to provide a zero ppm CO2 measurement. The CO2 zero measurement com- pensates for changes in optical cell performance, and is used in determinations of ppm CO2 until a new AZPC is performed. Features: - high precision and accuracy - low cost - light weight, compact size - unique antifouling protection - RS-232 data interface - data stream includes raw data and ppm CO2 - power saving and onboard logging options available PSI CO2-Pro TM with mounting bracket and Sea-Bird 5T pump Pro-Oceanus Systems Inc.; email: [email protected]; Tel: (902) 530-3550; Fax: (902) 530-3551; www.pro-oceanus.com
Transcript
Page 1: PSI CO2-ProTM - mdsys.co.kr · PSI CO2-ProTM The PSI CO2-ProTM is a compact, light-weight, plug-n- play pCO2 sensor that is designed for use on moorings (to 1000m), on profilers,

PSI CO2-ProTM

The PSI CO2-ProTM is a compact, light-weight, plug-n-play pCO2 sensor that is designed for use on moorings(to 1000m), on profilers, in under-way mode, and in thelaboratory. The PSI CO2-ProTM is fitted with an IRdetector and the new PSI pump-driven fast interface(patent pending). This revolutionary new interfaceprovides an equilibrated gas stream to the detector.

The PSI CO2-ProTM is factory calibrated from 0 – 600ppm with calibration burned into EPROM. To maintainaccuracy, the detector module has a power interrupt orcontact closure initiated automatic zero point calibration(AZPC). When the AZPC is initiated, the gas stream isrouted through a CO2 absorbent to provide a zero ppmCO2 measurement. The CO2 zero measurement com-pensates for changes in optical cell performance, and isused in determinations of ppm CO2 until a new AZPC isperformed.

Features:

- high precision and accuracy- low cost- light weight, compact size- unique antifouling protection- RS-232 data interface- data stream includes raw data and ppm CO2- power saving and onboard logging options available

PSI CO2-ProTM with mountingbracket and Sea-Bird 5T pump

Pro-Oceanus Systems Inc.; email: [email protected]; Tel: (902) 530-3550; Fax: (902) 530-3551; www.pro-oceanus.com

Page 2: PSI CO2-ProTM - mdsys.co.kr · PSI CO2-ProTM The PSI CO2-ProTM is a compact, light-weight, plug-n- play pCO2 sensor that is designed for use on moorings (to 1000m), on profilers,

Performance:

accuracyCO2 concentration ±Gas stream humidity ±Gas stream pressure ±

precisionCO2 concentration 0gas stream humidity 1gas stream pressure 1

calibration range 0temperature range 0

Physical Characteristics *

size lweight 1

2housing h

p

Electrical Characteristics

input voltage 1

power consumption 007

data storage adata output Rsample rate 0equilibration time constant a

*

0

100

200

300

400

500

600

700

me

asu

red

(pp

m)

0 100 200 300 400 500 600 700

standard (ppm)

calibration of CO2-Pro

y = -0.12767 + 0.998275 x

r^2 = 0.999969

Pro-Oceanus Systems Inc.; email: sales@pro-oce

calibration of a PSI CO2-ProTM after 6

~1 ppm1 mb2 mb

.01 ppmmbmb– 600 ppm*– 35 C

ength 48.6 cm (add 10 cm for connectors); 19.05 cm diameter1 kg in air.2 kg in waterard anodized aluminumroof to 1000 meters

2 ± 0.5 Volts DC through 4 pin Impulse male connectorif equipped with voltage regulator module: 9.5 – 18 V DC

.8 A at 12 Volts for warm-up;

.4 A during operation; not including pump power;0 ma in sleep mode; available with logger/ controller optionvailable by special orderS-232 serial output.6 hertzbout 3 minutes depending on pump flow rate

size or electrical specifications may change without notice

months of laboratory and field use

anus.com; Tel: (902) 530-3550; Fax: (902) 530-3551; www-pro-oceanus.com

Bryan Schofield
TextBox
(alternate ranges available)

Recommended